Specifications

Specifications are subject to change without notice.

Specification Details
Operating temperature –5 to 105 ºC (23 to 221 ºF)
Storage temperature 4 to 70 ºC (40 to 158 ºF)
Temperature element NTC300 thermistor
Sensor cable pHD: 5-conductor (plus 2 shields), 6 m (20 ft); LCP: 5-conductor

(plus 1 shield), 3 m (10 ft)
Dimensions (length/diameter)| pHD: 271 mm (10.7 in.)/35 mm (1.4 in.); 1-in. NPT; LCP: 187 mm (7.35 in.)/51 mm(2 in.); 1-½ in. NPT
Components| Corrosion-resistant materials, fully- submersible
Pressure limit| 6.9 bar at 105 ºC (100 psi at 221 ºF)
Maximum flow rate| 3 m/s (10 ft/s)

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

Safety information

The manufacturer is not responsible for any damages due tomisapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is soley responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.

Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.

Use of hazard information

DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.

NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.

| This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
---|---
| This symbol indicates that a risk of electrical shock and/or electrocution exists.
| This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and indicates that care must be taken to prevent damage with the equipment.
| Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

Product overview

This sensor is designed to work with a controller for data collection and operation. Multiple controllers can be used with this sensor. This document assumes sensor installation and use with an sc200 controller. To use the sensor with other controllers, refer to the user manual for the controller that is used.

The sensor is available in different styles. Refer to Figure 1.
Sensor styles
Sensor styles

1 Insertion—allows removal without stopping the process flow| 3 Sanitary—for install in a 2-inch sanitary tee
---|---
2 Convertible—for a pipe tee or immersion in an open vessel| 4 Convertible—LCP type

Installation

Mounting

WARNING

Personal injury hazard. Removal of a sensor from a pressurized vessel can be dangerous. Installation and removal of these sensors should be done by individuals trained in proper high pressure and temperature installation. Always use industry approved hardware and safety procedures when dealing with high pressure and/or temperature fluid transport systems.
Figure 2 Mounting examples (1)
Mounting Instruction

1 Sanitary mount 3 Flow-through mount
2 Union mount 4 Flow-through mount—LCP sensor

For examples of sensors in different applications, refer to Figure 2 and Figure 3. The sensor must be calibrated before use. Refer to Calibrate the sensor on page 12.

Figure 3 Mounting examples (2)
Mounting Instruction

1 PVS insertion mount 3 Immersion mount
2 Insertion mount 4 Immersion mount, ball float
Connect the sensor to the module

Warning
 Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical connections.

  Warning
Electrocution Hazard. High voltage wiring for the controller is conducted behind the high voltage barrier in the controller enclosure. The barrier must remain in place except when installing modules, or when a qualified installation technician is wiring for power, relays or analog and network cards.

NOTICE

 Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity, resulting in degraded performance or eventual failure.

To install the module and connect the sensor, refer to the illustrated steps on page 9 and Table 2.

Note: If the sensor cable is not long enough to reach the controller, an interconnect cable and junction box are required to extend the distance.

Table 2 pH and ORP differential sensor wiring

Connector Pin no. Signal Sensor wire
8-pin 1 Reference Green
2 Inner shield Clear
3 –V supply White
4
5
6 Temp + Yellow
7 Temp – /Circuit low Black
8
2-pin 1 Active Red
2

Sensor shield wires – Connect all sensor ground/shield wires to the controller enclosure grounding screws.| Clear with black band

Installation Instruction

Installation Instruction
Installation Instruction
Installation Instruction
Installation Instruction
Installation Instruction
Installation Instruction
Installation Instruction
Installation Instruction

Operation

Guidelines for operation

CAUTION
Personal injury hazard. The glass bulb or shank on the sensor can break. Handle the sensor carefully to prevent injury.

  • Remove the protective cap before the sensor is put into the process.
  • When the sensor is removed from the process for >1 hour, fill the protective cap with pH 4 buffer (recommended) or tap water and put the cap on the sensor. Repeat every 2–4 weeks for extended storage.

User navigation

Refer to the controller documentation for keypad description and navigation information.

Push the RIGHT arrow key on the controller multiple times to show more information on the home screen and to show a graphical display.

Configure the sensor

Use the Configure menu to enter identification information for the sensor and to change options for data handling and storage. The following procedure can be used to configure pH or ORP sensors.

  1. Push the MENU key and select Sensor Setup, [Select Sensor], Configure.
  2. Use the arrow keys to select an option and push ENTER. To enter numbers, characters or punctuation, push and hold the UP or DOWN arrow keys. Push the RIGHT arrow key to advance to the next space.
Op tion Description
EDIT NAME Changes the name that corresponds to the sensor on the top of the

measure screen. The name is limited to 10 characters in any combination of letters, numbers, spaces or punctuation.
SENSOR S/N| Allows the user to enter the serial number of the sensor, limited to 16 characters in any combination of letters, numbers, spaces or punctuation.
DISPLAY FORMAT| For pH sensors only—changes the number of decimal places that are shown on the measure screen to XX.XX (default) or XX.X
TEMP UNITS| Sets the temperature units to °C (default) or °F
TEMP ELEMENT| pH sensors—sets the temperature element for automatic temperature compensation to PT100, PT1000 or NTC300 (default). If no element is used, the type can be set to manual and a value for temperature compensation can be entered (manual default: 25 °C). ORP sensors— temperature compensation is not used. A temperature element can be connected to measure temperature.
FILTER| Sets a time constant to increase signal stability. The time constant calculates the average value during a specified time—0 (no effect, default) to 60 seconds (average of signal value for 60 seconds). The filter increases the time for the sensor signal to respond to actual changes in the process.
PURE H2O COMP| For pH sensors only—adds a temperaturedependent correction to the measured pH value for pure water with additives—None (default), Ammonia, Morpholine or User defined. For process temperatures above 50 °C, the correction at 50 °C is used. For user-defined applications, a linear slope (default: 0 pH/°C) can be entered.
SET ISO POINT| At the isopotential point, the pH slope is independent of temperature. Most sensors have an isopotential point of 7.00 pH (default), however sensors for special applications may have a different isopotential value.
LOG SETUP| Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10, 15 (default), 30, 60 minutes
RESET DEFAULTS| Sets the configuration menu to the default settings. All sensor information is lost.

Calibrate the sensor

About sensor calibration

Calibration adjusts the sensor reading to match the value of one or more reference solutions. The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor must be calibrated regularly to maintain accuracy. The calibration frequency varies with the application and is best determined by experience.

A temperature element is used to provide pH readings that are automatically adjusted to 25 °C for temperature changes that affect the active and reference electrode. This adjustment can be manually set by the customer if the process temperature is constant.

During calibration, data is not sent to the datalog. Thus, the datalog can have areas where the data is intermittent.

pH calibration procedure

Sensors can be calibrated with 1 or 2 reference solutions (1-point or 2- point calibration). Standard buffers are automatically recognized. Make sure that the correct buffer set is used (refer to Change calibration options on page 15).

  1. 1.       1. Put the sensor in the first reference solution. Make sure that the sensor portion of the probe is fully immersed in the liquid **(Figure 4)**.  

    Figure 4 Sensor in reference solution
    Sensor In Reference Solution

    1. Wait for the sensor and solution temperature to equalize. This can take 30 minutes or more if the temperature difference between the process and reference solution is significant.
    2. Push the MENU key and select Sensor Setup, [Select Sensor], Calibrate.
      Option| Description
      ---|---
      2 point buffer| Use 2 buffers for calibration, for example pH 7 and pH 4 (recommended method). The buffers must be from the buffer set that is specified in the Cal Options menu (refer to Change calibration options on page 15).
      1 point buffer| Use 1 buffer for calibration, for example pH 7. The buffer must be from the buffer set that is specified in the Cal Options menu.
      2 point sample| Use 2 samples or buffers of known pH value for calibration. Determine the pH value of samples with a different instrument.
      1 point sample| Use 1 sample or buffer for calibration. Determine the pH value of samples with a different instrument
    3. Select the type of calibration:

If the passcode is enabled in the security menu for the controller, enter the passcode.Select the option for the output signal during calibration:

Option Description
Active The instrument sends the current measured output value during the

calibration procedure.
Hold| The sensor output value is held at the current measured value during the calibration procedure.
Transfer| A preset output value is sent during calibration. Refer to the controller user manual to change the preset value

  1. 1.       1. With the sensor in the first reference solution, push ENTER. The measured value is shown.

Wait for the value to stabilize and push ENTER.

Note:

  1. The screen may advance to the next step automatically.

  2. If the reference solution is a sample, measure the pH value with a secondary verification instrument. Use the arrow keys to enter the measured value and push ENTER.

Note:

  1. If a pH buffer not listed the Cal Options menu is used, refer to the buffer bottle to find the pH value that corresponds to the temperature of the buffer.

  2. For a 2-point calibration, measure the second reference solution (or sample):

  3.  Remove the sensor from the first solution and rinse with clean water.

    1.  Put the sensor in the next reference solution and push

ENTER

  1. 1. .
    1.  Wait for the value to stabilize. Push

ENTER.
Note:

  1. 1. The screen may advance to the next step automatically.
    1.  If the reference solution is a sample, measure the pH value with a secondary verification instrument. Use the arrow keys to enter the measured value and push

ENTER.

    1. Review the calibration result:
  * Passed—the sensor is calibrated and ready to measure samples. The slope and/or offset values are shown.
  * Failed—the calibration slope or offset is outside of accepted limits. Repeat the calibration with fresh reference solutions. Refer to mAINTENANCE n page 16 and Troubleshooting on page 17 for more information.
  1. If the calibration passed, push ENTER to continue.
  2. If the option for operator ID is set to Yes in the Calibration Options menu, enter an operator ID. Refer to Change calibration options on page 15.
  3. On the New Sensor screen, select whether the sensor is new:
    Option| Description
    ---|---
    Yes| The sensor was not calibrated previously with this controller. The days of operation and previous calibration curves for the sensor are reset.
    No| The sensor was calibrated previously with this
  4. Return the sensor to the process and push ENTER.
    The output signal returns to the active state and the measured sample value is shown on the measure screen.
    Note: If the output mode is set to hold or transfer, select the delay time when
    the outputs return to the active state.

ORP calibration procedure

Sensors can be calibrated with an ORP reference solution or with the process sample.

  1.  Put the sensor in the reference solution. Make sure that the sensor portion of the probe is fully immersed in the solution (Figure 5).
    Figure 5 Sensor in reference solution
    Sensor In Reference Solution

  2. Push the MENU key and select Sensor Setup, [Select Sensor], Calibrate.

  3. Push ENTER to select 1 Point Sample calibration.

  4. If the passcode is enabled in the security menu for the controller, enter the passcode.
    Option| Description
    ---|---
    Active| The instrument sends the current measured output value during the calibration procedure.
    Hold| The sensor output value is held at the current measured value during the calibration procedure.
    Transfer| A preset output value is sent during calibration. Refer to the controller user manual to change the preset value

  5. Select the option for the output signal during calibration:

  6.  With the sensor in the reference solution or sample, push ENTER. The measured value is shown.

  7. Wait for the value to stabilize and push ENTER.
    Note: The screen may advance to the next step automatically.

  8. If the process sample is used for calibration, measure the ORP value with a secondary verification instrument. Use the arrow keys to enter the value and push ENTER.

  9. Review the calibration result:

    • Passed the sensor is calibrated and ready to measure samples. The slope and/or offset values are shown.
    • Failed the calibration slope or offset is outside of accepted limits. Repeat the calibration with fresh reference solutions. Refer to Maintenance on page 16 and Troubleshooting on page 17 for more information.
  10. If the calibration passed, push ENTER to continue.

  11. If the option for operator ID is set to Yes in the Calibration Options menu, enter an operator ID. Refer to Change calibration options on page 15.

  12. On the New Sensor screen, select whether the sensor is new:
    Option| Description
    ---|---
    Yes| The sensor was not calibrated previously with this controller. The days of operation and previous calibration curves for the sensor are reset.
    No| The sensor was calibrated previously with this

  13. Return the sensor to the process and push ENTER. The output signal returns to the active state and the measured sample value is shown on the measure screen.
    Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active state.

Temperature calibration

The instrument is calibrated at the factory for accurate temperature measurement. The temperature can be calibrated to increase accuracy.

  1. Put the sensor in a container of water that is at a known temperature. Measure the temperature of the water with an accurate thermometer or independent instrument.
  2. Push the MENU key and select Sensor Setup, [Select Sensor], Calibrate.
  3. Select 1 PT Temp Cal and push ENTER.
  4. Wait for the value to stabilize and push ENTER.
  5. Enter the exact value and push ENTER.
  6. Return the sensor to the process and push ENTER.

Exit calibration procedure

If the BACK key is pushed during a calibration, the user can exit the calibration.

  1. Push the BACK key during a calibration. Three options are shown:
    Option| Description
    ---|---
    QUIT CAL| Stop the calibration. A new calibration must start from the beginning.
    BACK TO CAL| Return to the calibration.
    LEAVE CAL| Exit the calibration temporarily. Access to other menus is allowed. A calibration for a second sensor (if present) can be started. To return to the calibration, push the MENU key and select Sensor Setup, [Select Sensor
  2.  Use the arrow keys to select one of the options and push ENTER.

Change calibration options

The user can set a reminder or include an operator ID with calibration data from the CAL OPTIONS menu.

  1. Push the MENU key and select Sensor Setup, [Select Sensor], Calibrate, Cal Options.
  2. Use the arrow keys to select an option and push ENTER.
Option Description
SELECT BUFFER For pH sensors only—changes the set of buffer solutions that

are recognized for calibration to pH 4.00, 7.00, 10.00 (default set) or DIN 19267 (pH 1.09, 4.65, 6.79, 9.23, 12.75) Note: Other buffers can be used if the 1 point sample or 2 point sample option is selected during calibration.
CAL REMINDER| Sets a reminder for the next calibration in days, months or years—Off (default), 1 day, 7, 30, 60, or 90 days, 6 or 9 months, 1 or 2 years
OP ID on CAL| Includes an operator ID with calibration data— Yes or No (default). The ID is entered during the calibration.

Reset calibration options

The calibration options can be reset to the factory default options.

  1. Push the MENU key and select Sensor Setup, [Select Sensor], Calibrate, Reset Default Cal.
  2. If the passcode is enabled in the security menu for the controller, enter the passcode.
  3. Push ENTER. The Reset Cal? screen is shown.
  4. Push ENTER. All calibration options are set to the default values.
  5. If the option for operator ID is set to Yes in the Calibration Options menu, enter an operator ID. Refer to Change calibration options on page 15.
  6. On the New Sensor screen, select whether the sensor is new:
    Option| Description
    ---|---
    Yes| The sensor was not calibrated previously with this controller. The days of operation and previous calibration curves for the sensor are reset.
    No| The sensor was calibrated previously with this controller
  7. Push the BACK key to return to the measure screen.

Impedance measurements

To increase the reliability of the pH measurement system, the controller determines the impedance of the glass electrodes. This measurement is taken every minute. During diagnostics, the pH measurement reading will be on hold for five seconds. If an error message appears, refer to Error list on page 19 for more details.

To enable/disable the sensor impedance measurement:

  1. Push the MENU key and select Sensor Setup.
  2. Select Diag/test and push ENTER.
  3. Select Imped Status and push ENTER.
  4. Select Enable/Disable and push ENTER

To see both actual and reference sensor impedance readings:

  1. Push the MENU key and select Sensor Setup.
  2. Select Diag/test and push ENTER.
  3. Select sensor signals and push ENTER

Modbus registers

A list of Modbus registers is available for network communication. Refer to the manufacturer’s website for more information.

Maintenance

WARNING

Personal injury hazard. Only qualified personnel should conduct the tasks described in this section of the manual.

Maintenance schedule

The maintenance schedule shows minimum times for regular maintenance tasks. Perform maintenance tasks more frequently for applications that cause electrode fouling.

Maintenance task 90 days Annually
Clean the sensor X
Inspect sensor for damage X
Replace salt bridge and fill solution X
Calibrate sensor Set by regulatory agencies or experience
Clean the sensor

WARNING

Note Icon
| **** Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.

WARNING

Personal injury hazard. Removal of a sensor from a pressurized vessel can be dangerous. Installation and removal of these sensors should be done by individuals trained in proper high pressure and temperature installation. Always use industry approved hardware and safety procedures when dealing with high pressure and/or temperature fluid transport systems.

Pre-requisite: Prepare a mild soap solution with a non-abrasive dishwashing detergent that does not contain lanolin. Lanolin leaves a film on the electrode surface that can degrade the sensor performance. Examine the sensor periodically for debris and deposits. Clean the sensor when there is a buildup of deposits or when performance has degraded.

  1. Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with clean, warm water.
  2. Soak the sensor for 2 to 3 minutes in the soap solution.
  3. Use a soft bristle brush to scrub the entire measuring end of the sensor.
  4. If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5% HCl for a maximum of 5 minutes.
  5. Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
  6. Rinse the sensor with clean water.
    Note: Sensors with antimony electrodes for HF applications may require
    additional cleaning. Contact technical support.

Always calibrate the sensor after maintenance procedures are done.

Replace the salt bridge

Differential sensors have a removable salt bridge (Figure 6). If the sensor has been cleaned but fails calibration, replace the salt bridge and the standard cell solution. Refer to the instructions included with the salt bridge.

Figure 6 Salt bridge
Salt Bridge

1 Salt bridge 2 Differential sensor

Always calibrate the sensor after maintenance procedures.

Troubleshooting

Intermittent data

During calibration, data is not sent to the datalog. Thus, the datalog can have areas where the data is intermittent.

Test the pH sensor

Pre-requisites: Two pH buffers and a multimeter. If a calibration fails, first complete the maintenance procedures in Maintenance on page 16.

  1. Put the sensor in a pH 7 buffer solution and wait for the temperature of the sensor and buffer to reach room temperature.

  2. Disconnect the red, green, yellow and black sensor wires from the module.

  3. Measure the resistance between the yellow and black wires to verify the operation of the temperature element. The resistance should be between 250 and 350 ohms at approximately 25 ºC. If the temperature element is good, reconnect the yellow and black wires to the module.

  4. Measure the DC mV with the multimeter (+) lead connected to the red wire and the (–) lead connected to the green wire. The reading should be between –50 and + 50 mV. If the reading is outside of these limits, clean the sensor and change
    the salt bridge and standard cell solution.

  5. With the multimeter still connected the same way, rinse the sensor with water and put it in a pH 4 or pH 10 buffer solution. Wait for the temperature of the sensor and buffer to reach room temperature.

  6. Compare the mV reading in the pH 4 or 10 buffer to the reading in the pH 7 buffer. The reading should differ by approximately 160 mV. If the difference is less than 160 mV, call technical support.

Test the ORP sensor

Pre-requisites: 200 mV ORP reference solution, multimeter. If a calibration fails, first complete the maintenance procedures in Maintenance on page 16.

  1. Put the sensor in a 200 mV reference solution and wait for the temperature of the sensor and solution to reach room temperature.

  2. Disconnect the red, green, yellow and black sensor wires from the module.

  3. Measure the resistance between the yellow and black wires to verify the operation of the temperature element. The resistance should be between 250 and 350 ohms at approximately 25 ºC.
    If the temperature element is good, reconnect the yellow and black wires to the module.

  4. Measure the DC mV with the multimeter (+) lead connected to the red wire and the (–) lead connected to the green wire. The reading should be between 160 and 240 mV.
    If the reading is outside of these limits, call technical support.

Sensor diagnostic and test menu

The sensor diagnostic and test menu shows current and historical information about the instrument. Refer to Table 3. To access the sensor diagnostic and test menu, push the MENU key and select Sensor Setup, [Select Sensor], DIAG/TEST.

Table 3 Sensor DIAG/TEST menu

Option Description
MODULE INFORMATION Shows the version and the serial number for the sensor

module.
SENSOR INFORMATION| Shows the name and serial number that was entered by the user.
CAL DAYS| Shows the number of days since the last calibration.
CAL HISTORY| Shows a list of and details for each calibration.
RESET CAL HISTORY| Resets the calibration history for the sensor (requires service-level passcode). All previous calibration data is lost.
SENSOR SIGNALS| Shows the current reading in mV, the impedance of the active and reference electrodes and the temperature-compensated impedance. Shows predicted life when Predict Enable is on.
IMPED STATUS| Shows when the impedance of the sensor (measured every 3 hours) indicates malfunction. If enabled (recommended), a warning is shown when the impedance is outside of normal limits.
PREDICT ENABLE| Adds the predicted life of the pH sensor to the Sensor Signals screen. The Impedance Status must be enabled and the sensor must have a 2-point calibration for accurate prediction.When the predicted life has expired, a warning is shown on the measurement screen.

Table 3 Sensor DIAG/TEST menu (continued)

Option Description
SENSOR DAYS Shows the number of days that the sensor has been in operation.
RESET SENSOR DAYS Resets the number of days that the sensor has been in

operation.

Error list

When an error occurs, the reading on the measurement screen flashes and all outputs are held when specified in the controller menu. To show the sensor errors, press the MENU key and select Sensor Diag, [Select Sensor], Error List. A list of possible errors is shown in Table 4.

Table 4 Error list for pH and ORP sensors

Error Description Resolution
PH TOO HIGH The measured pH is > 14 Calibrate or replace the sensor.
ORP TOO HIGH The measured ORP value is > 2100 mV
PH TOO LOW The measured pH is < 0 Calibrate or replace the sensor.
ORP TOO LOW The measured ORP value is < –2100 mV
OFFSET TOO HIGH The offset is > 9 (pH) or 200 mV (ORP) Follow the

maintenance procedures for the sensor and then repeat the calibration, or replace the sensor.
OFFSET TOO LOW| The offset is < 5 (pH) or –200 mV (ORP)
SLOPE TOO HIGH| The slope is >62 (pH)/1.3 (ORP)| Repeat the calibration with a fresh buffer or sample, or replace the sensor.

Table 4 Error list for pH and ORP sensors (continued)

Error Description Resolution
SLOPE TOO LOW The slope is <50 (pH)/0.7 (ORP) Clean the sensor, then repeat

the calibration, or replace the sensor.
TEMP TOO HIGH| The measured temperature is >130 °C| Make sure that the correct temperature element is selected.
TEMP TOO LOW| The measured temperature is < –10 °C
ADC FAILURE| The analog to digital conversion failed| Power off and power on the controller. Call technical support.
A ELEC TOO HIGH| The active electrode impedance is > 900 MΩ| The sensor is in air. Return the sensor to the process.
A ELEC TOO LOW| The active electrode impedance is < 8 MΩ| The sensor is damaged or dirty. Call technical support.
R ELEC TOO HIGH| The reference electrode impedance is > 900 MΩ| Buffer leaked or evaporated. Call technical support.
R ELEC TOO LOW| The reference electrode impedance is < 8 MΩ| The reference electrode is damaged. Call technical support.
SAME BUFFER| The buffers for 2 point buffer calibration have the same value| Complete the steps in Test the pH sensor on page 17.
SENSOR MISSING| The sensor is missing or disconnected| Examine the wiring and connections for the sensor and for the module.

Table 4 Error list for pH and ORP sensors (continued)

Error Description Resolution
TEMP MISSING The temperature sensor is missing Examine the wiring for the

temperature sensor. Make sure that the correct temperature element is selected.
GLASS IMP LOW| The glass bulb is broken or reached end of life| Replace the sensor. Call technical support.

Warning list for sensors

A warning does not affect the operation of menus, relays and outputs. A warning icon flashes and a message is shown on the bottom of the measurement screen. To show the sensor warnings, press the MENU key and select Sensor Diag, [Select Sensor], Warning List. A list of possible warnings is shown in Table 5.

Table 5 Warning list for pH and ORP analog sensors

Warning Description Resolution
PH TOO HIGH The measured pH is > 13 Calibrate or replace the sensor.
ORP TOO HIGH The measured ORP value is > 2100 mV
PH TOO LOW The measured pH is < 1 Calibrate or replace the sensor.
ORP TOO LOW The measured ORP value is < –2100 mV
OFFSET TOO HIGH The offset is > 8 (pH) or 200 mV (ORP) Follow the

maintenance procedures for the sensor and then repeat the calibration.
OFFSET TOO LOW| The offset is < 6 (pH) or –200 mV (ORP)

Table 5 Warning list for pH and ORP analog sensors (continued)

Warning Description Resolution
SLOPE TOO HIGH The slope is >60 (pH)/1.3 (ORP) Repeat the calibration with a

fresh buffer or sample.
SLOPE TOO LOW| The slope is <54 (pH)/0.7 (ORP)| Clean the sensor, then repeat the calibration.
TEMP TOO HIGH| The measured temperature is >100 °C| Make sure the right temperature element is used.
TEMP TOO LOW| The measured temperature is < 0 °C
CAL OVERDUE| The Cal Reminder time has expired| Calibrate the sensor.
REPLACE SENSOR| The sensor has been in operation > 365 days, or the predicted life has expired (Refer to Sensor diagnostic and test menu on page 18)| Replace the sensor.
NOT CALIBRATED| The sensor has not been calibrated| Calibrate the sensor.
FLASH FAILURE| External flash memory failed| Contact technical support.
A ELEC TOO HIGH| The active electrode impedance is > 800 MΩ| The sensor is in air. Return the sensor to the process.
A ELEC TOO LOW| The active electrode impedance is < 15 MΩ| The sensor is damaged or dirty. Call technical support.

Table 5 Warning list for pH and ORP analog sensors (continued)

Warning Description Resolution
R ELEC TOO HIGH The reference electrode impedance is > 800 MΩ Buffer leaked

or evaporated. Call technical support.
R ELEC TOO LOW| The reference electrode impedance is < 15 MΩ| The reference electrode is damaged. Call technical support.
CAL IN PROGRESS| A calibration was started but not completed| Return to calibration.

Event list for sensors

The event list shows current activities such as configuration changes, alarms, warning conditions, etc. To show the events, press the MENU key and select Sensor Diag, [Select Sensor], Event List. A list of possible events is shown in Table 6. Previous events are recorded in the event log, which can be downloaded from the controller. Refer to the controller documentation for data retrieval options.

Table 6 Event list for pH and ORP sensors

Event Description
CAL READY The sensor is ready for calibration
CAL OK The current calibration is good
TIME EXPIRED The stabilization time during calibration expired
NO BUFFER No buffer is detected
SLOPE HIGH The calibration slope is above the upper limit
SLOPE LOW The calibration slope is below the lower limit

Table 6 Event list for pH and ORP sensors (continued

Event Description
OFFSET HIGH The calibration offset value for the sensor is above the upper

limit
OFFSET LOW| The calibration offset value for the sensor is below the lower limit
PTS CLOSE| The calibration points are too similar in value for a 2-point calibration
CAL FAIL| The calibration failed
CAL HIGH| The calibration value is above the upper limit
UNSTABLE| The reading during calibration was unstable
CHANGE IN CONFIG float| The configuration was changed—floating point type
CHANGE IN CONFIG text| The configuration was changed—text type
CHANGE IN CONFIG int| The configuration was changed—integer value type
RESET CONFIG| The configuration was reset to the default options
POWER ON EVENT| The power was turned on
ADC FAILURE| The ADC conversion failed (hardware failure)
FLASH ERASE| The flash memory was erased
TEMPERATURE| The recorded temperature is too high or too low
1PT MANUAL START| Start of 1-point sample calibration
1PT AUTO START| Start of 1-point buffer calibration for pH

Table 6 Event list for pH and ORP sensors (continued)

Event Description
2PT MANUAL START Start of 2-point sample calibration for pH
2PT AUTO START Start of 2-point buffer calibration for pH
1PT MANUAL END End of 1-point sample calibration
1PT AUTO END End of 1-point buffer calibration for pH
2PT MANUAL END End of 2-point sample calibration for pH
2PT AUTO END End of 2-point buffer calibration for pH

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Description Quantity Item no.
Buffer Solution, pH 4 500 mL 2283449
Buffer Solution, pH 7 500 mL 2283549
Buffer Solution, pH 10 500 mL 2283649
Gel Powder 2 g 25M8A1002-101
ORP Reference Solution, 200 mV 500 mL 25M2A1001-115
ORP Reference Solution, 600 mV 500 mL 25M2A1002-115
Standard Cell Solution for pHD 500 mL 25M1A1025-115
Standard Cell Solution for LCP 500 mL 25M1A1001-115

pHD sensors

Description Item no.
Salt Bridge, pHD PEEK/Kynar, with Santoprene gasket SB-P1SV
Salt Bridge, pHD PEEK/Kynar, with perfluoroelastomer gasket SB-P1SP
Salt Bridge, pHD PEEK/Ceramic with Santoprene gasket SB-P2SV
Salt Bridge, pHD Ryton sensor with Santoprene gasket SB-R1SV
Sensor Protector, pHD convertible style sensor, PEEK 1000F3374-002
Sensor Protector, pHD convertible style sensor, Ryton 1000F3374-003

LCP and Ryton encapsulated sensors

Description Item no.
Salt Bridge, LCP/Kynar, with O-ring 60-9765-000-001
Salt Bridge, LCP/Ceramic, with O-ring 60-9765-010-001
Salt Bridge, Ryton/Kynar, with O-ring 60-9764-000-001
Salt Bridge, Ryton/Ceramic, with O-ring 60-9764-020-001

Costumer Support

HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. 970-669-3050(800) 227-4224 (U.S.A. only)
Fax 970-669-2932
[email protected]
www.hach.com
HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
[email protected]
www.de.hach.com
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499
Symbols
Hach Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HACH User Manuals

Related Manuals