FRISTADEN LAB B0896VV35S Vacuum Filtration Pump User Manual

June 1, 2024
FRISTADEN LAB

FRISTADEN LAB B0896VV35S Vacuum Filtration Pump

FRISTADEN LAB B0896VV35S Vacuum Filtration Pump

Summary

IMPORTANT!

Read t he User Manual carefully before use and follow all operating and safety instructions!

PRODUCT FEATURES

POWERFUL EFFICIENT MOTOR: The Fristaden Lab Diaphragm Vacuum Pump is designed to reduce filtration processing t ime in order to improve efficiency It is built with a powerful, oil less 20W motor and can pump at a fast rate of 70 liters per minute.
COMPACT DESIGN: The vacuum pump has a small, compact design and can fit easily onto most desktops and workspaces.
The hose easily connects to a vacuum filtration apparatus.
SAFETY FEATURES: The machine has the added safety features of overload protection, short circuit protection and over temperature protection The base has four suction cup feet so the machine won’t slip while in use.
LAB QUALITY: The American-brand Fristaden Lab Vacuum Pump is made from lab-quality materials designed to deliver you accurate results year after year.

PACKING LIST

  • Fristaden Lab Vacuum Pump
  • Rubber Connection Tube
  • Power Cord
  • User Manual

Safety

IMPORTANT WARNINGS!

  • Ensure everyone operating the machine reads the user manual and understands the operating instructions, cautions and dangers associated with the machine Only trained staff should operate the machine.
  • Wear proper safety gear, including but not limited to protective eyewear, gloves, etc, to prevent injury from the liquids or gasses used w ith the vacuum filtration pump.
  • Check the machine and accessories prior to each use for damage Do not use the machine if t he machine or any parts are broken Contact the manufacturer and do not attempt to repair it yourself
  • Do not rep lace power cables, plugs or other parts un less they are directly from the manufacturer.
  • Always supervise t he machine while in use! Never leave the machine alone or unattended w hen it is operating.
  • Before plugging the machine into an electrical outlet, make sure the voltage and frequency matches that of the machine. Do not p lug in or unplug the machine if your hands have liquid on them.
  • Operate the machine in a safe environment away from hazards including conductive dust and explosive and corrosive gas. Do not operate in a flammable or dangerous environment Keep the machine away from pets, children and other hazards.
  • Operate the machine in a safe, dry and well ventilated environment The environment temperature should not exceed 5-40°C and have less than or equal to 80% relative humidity (continued on next page) ..
  • (continued) The altitude should not exceed 2000m.
  • Place the machine on a flat, stable and spacious work surface away from vibrations and airflows.
  • The machine is for indoor use only
  • Maintain a 70cm distance from other objects on all sides of the machine.
  • Always power off and unplug the machine when not in use Store it in a safe place away from hazards

Specifications

Type

|

Negative Pressure

---|---

Pumping Speed (L/min)

|

10

Ultimate Pressure (Mpa)

|

0.075

Noise Level (dB)

|

<40

Motor Power (w)

|

20

Diaphragm & Valves

|

NBR

Overload Protection

|

Yes

Over Temperature Protection

|

Yes

Short Circuit Protection

|

Yes

Voltage

|

110-240v

Dimensions (mm)

|

192 X 100 X 135

Product Weight (kg)

|

1.58

Operation

OPERATING THE MACHINE

  • Set the Vacuum Pump on a flat, solid, dry surface away from liquids and hazardous materials. Suction the feet to the surface

  • Connect t he rubber connection tube to the vacuum pump and t hen to t he fil t ration apparatus.

  • Plug the power cord into t he back of the machine.
    Plug the adapter into an outlet with t he correct voltage.

  • Turn on the power switch on the front the machine.

  • To power off, press t he power switch to off Unplug the power cord when not in use and store the machine in a safe place. DO NOT unplug or plug in t he machine when the power is turned on.

Operation

Company Information

WARRANTY

The Fristaden Lab FL-VPl0L has al Year Manufacturer’s Warranty Please contact the Fristaden Lab’s customer service department at support@fristadenllc.com for more information

ABOUT US

Fristaden Lab is a lab and scientific equipment retailer established in 2013 in Chicago, Illinois.
We sell an array of popular products, including analytical balances, centrifuges, glassware, hematology counters, hot plates, magnetic stirrers, pH meters, pipettes and pressure pumps.
We stand behind all of our products, offering a 7-2 year warranty, 30-day returns, and a responsive customer service team committed to solving any questions or issues you may have.

CONTACT

Fristaden Lab .comsupport@fristadenllc.com
Owned and operated by Fristaden & Company, LLC 407 Ryland Street, Suite 200-A Reno NV, 89502 USA

Customer Support

Fristaden Lab com
support@fristadenlle.com
Owned and operated by
Fristaden & Company, LLC
407 Ryland Street, Suite 200-A
Reno NV, 89502 USA

Logo

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals