FRISTADEN LAB B081DCQ9BN Moisture Analyzer User Manual
- June 1, 2024
- FRISTADEN LAB
Table of Contents
Moisture Analyzer
User Manual Model: HCSFYlOS
±0.5% Moisture Content Accuracy
O.OOSg Weighing Accuracy
lOOg Capacity
Summary
IMPORTANT!
Thankyou for purchasing the Fristaden Lab Moisture Analyzer. Please read the
User Manual carefully before use and follow all operating and safety
instructions!
PRODUCT FEATURES
RESULTS YOU CAN TRUST: If you conduct moisture analysis for quality control or
for lab work, a small inaccuracy can compromise your product or ruin your
experiment. That’s why the Fristaden Lab precision moisture analyzer is
accurate to degree of magnitude higher than similar models (0.005g). Every
moisture balance we sell is tested for accuracy and precision before sale.
MOISTURE CONTENT ACCURATE TO ±0.5%: We know how a 1-3% change in moisture
content can make all the difference in your product or sample’s quality.
That’s why this moisture analyzer has a moisture accuracy of ±0.5% (sample
≥5g). This is based on a moisture readability of 0.02% and a repeatability of
0.2%. These results are driven by a balance accurate up to 0.005g and a
halogen light accurate to ± 1 °C.
MANY FEATURES, INTUITIVE CONTROLS: The digital moisture analyzer clearly
displays results on a 5.5” LCD Touch Screen. You can easily switch between
manual, timed and automatic tests. With the automatic test, the measuring
device will alert you once the sample has lost its moisture. (continued)
PRODUCT FEATURES (CONTINUED)
Heat samples, up to lOOg, from 60-160°(. A calibration weight, pan holder,
sample remover and 10 sample pans are included. The moisture scale can be set
up in under 15 minutes and comes with a detailed, user-friendly instruction
manual.
RAPID RESULTS: The Fristaden Lab moisture analyzer can quickly dry samples
with a powerful and accurate (60-160°() halogen lamp. Halogen lamps work
faster than other methods often seen in moisture analyzers, saving you time
and money. The scientific analyzer can be used with food, plants, grain,
minerals, chemicals, soil, wood, plastics and more. All samples can be
measured on a full 0-100% moisture content range.
SPECIFICATIONS
Model | HCSFY105 |
---|---|
Capacity | lOOg |
Readability | 0.005g |
Moisture Readability | 0.02% |
Repeata bi I ity | 0.20% |
Display | 5.5″ LCD Touch Screen |
Pan Diameter | lOOmm |
Heating | High Efficiency Halogen Lamp |
Accuracy for Moisture Measure | ±0.5% (sample ≥5g) |
SPECIFICATIONS
Temp. Increase Step | ±1 °C |
---|---|
Temp. Range | 60-160°C |
Heating Time | 1-99 min with l sec intervals |
Time End Setting | Auto Stop/Timed Stop |
Amount of Storage | 8 |
Moisture Range | 0-100% |
Material | Aluminum alloy base, ABS body |
Modes | Timer Dry/ Auto Dry |
Dimensions | 330mm x 480mm x 140mm |
N.W. | 2.9kg/ G.W. 4.2kg |
Safety
WARNING!
- Ensure everyone operating the machine reads the user manual and understands the operating instructions, cautions and dangers associated with the machine. Only trained staff should operate the machine.
- Do not operate the machine if you are under 18 years of age.
- Do not touch the halogen I ig ht or the top of the instrument when in use! The halogen light can cause severe burns and other injuries if touched.
- Be careful removing samples, as they still will be hot.
IMPORTANT!!
-
Do not cover the instrument while in use. The cover will be hot.
-
Check the machine and accessories prior to each use for damage. Do not use the machine if damaged.
Contact the manufacturer. -
Always supervise the machine while in use! Never leave the machine alone or unattended when it is operating, hot or cooling down.
-
Before plugging the machine into an electrical outlet, make sure the voltage and frequency matches that of the machine.
-
Operate the machine in a safe environment away from hazards. Do not operate in a flammable, magnetic or dangerous environment. Keep the machine away from magnetic parts (e.g. watches, pacemakers, etc.)
-
Keep the machine away from children, pets and other potentia I hazards.
-
Place the machine on a flat and spacious work surface.
-
Ensure the surface is non-slip, dry, fireproof, stable and away from any potential hazards.
-
Operate the machine in a safe, dry and well ventilated environment. The environment temperature should not exceed 5-35°C and have less than or equal to 50- 75% relative humidity.
-
Do not cover the machine.
IMPORTANT (CONTINUED)
- DAMAGE: Do not use the machine if the machine or any parts are broken. Contact the manufacturer and do not attempt to repair it yourself.
- DAMAGE: Do not replace power cables, plugs or other parts unless they are directly from the manufacturer ..
- Do not exceed the max weight (lOOg) to avoid damaging the sensor.
Display and Controls
DISPLAY CONTROLS
OFF | Turn off the device |
---|---|
CAL | Calibrate the device (or set date/time) |
Time | Timed heating function |
Auto | Automatic heating function |
Latest | The most recent record |
History | Saved records |
Operation
PACKING LIST
- Moisture Analyzer
- Pan Remover
- Pan Holder
- Calibration Weight
- Power Cable
- lOx Sample Pan
- User Manual
- lOOg Calibration Weight
INSTALLATION
CHECK FOR DAMAGE – Open the package and ensure no parts were damaged during
delivery. Make sure all parts shown in the packing list are included.
- Place moisture analyzer on a flat, stable working platform.
- Install sample pan holder and pan.
- Observe the position of the level bubble. If it is offset, adjust the feet until it is level.
- Attach power cord and plug in device.
- The scale accepts voltage of llOV. Please confirm the correct voltage before using the sea le.
- Do not overload the full capacity of the balance.
- If there is an error, calibrate the scale.
GETTTING STARTED
- Set the moisture analyzer on a flat, solid, dry surface away from liquids and hazardous materials.
- Plug in the instrument.
- Turn on the power for at least 30 minutes before use to allow the instrument to adapt to environmental fluctuations. If it is used in cold temperatures, turn on the heating lamp for 30 minutes, then allow the machine to cool to room temperature.
CALIBRATION
- NOTE: The weighing scale should be calibrated under the ta re function.
- NOTE: The calibration should ensure that the internal signal is stable.
- NOTE: Calibrate the scale before use.
- Clean the sample plan and place it on the sample pan holder in the center of the instrument.
- Close the cover and touch the calibration key on the main menu.
- Press the weight calibration key. The display will show the weight value.
- Place the corresponding weight on the pan. Close the upper cover. After 5 seconds, it will display the weight.
- After calibrating, if the display does not show 0 the calibration.
PREPARE THE SAMPLE
- Press Tare on the display. Remove the sample pan.
- Clean the disposable sample plan. Place 5-lOg of the sample into the sample pan. The sample should be a powder, for best results. Larger samples should be crushed in a grinder before moisture analysis.
- Set the heating parameters (time, temperature and ending method) before weighing the sample.Place the sample in the device and close the lid.
DRYING TEMPERATURE SETTINGS
- On the homepage of the display, select the auto heating or ti med heating key.
- Set the temperature in the display frame to the temperature.
- Press confirm to finish.
END MODE SELECTION
- *You can stop the drying with three different modes: Manual, Automatic and Timed.
- Manual: Manually end the drying time.
- Automatic: Once the sample remains steady (less than lmg for a set time), the instrument will end the measurement process. The display will record the heating time.
- Timed: Enter the amount of time you would like to heat the sample for. Once that time is reached, the halogen light will shut off.
AUTOMATIC MEASURING FUNCTION
- Select Auto heating on the display.
- Input the desired temperature.
- Place samples in weighing area.
- Once the balance is stable, begin heating.
TIMED MEASURING FUNCTION
- Select Time heating on the display.
- Input the desired temperature.
- Input the desired measurement time.
- Place samples in weighing area.
- Once the balance is stable, begin heating.
STORED DATA
- *The analyzer automatically stores the 8 previous recordings.
- Press history on the main menu.
- Select the saved sample folder.
- Click on the folder to enter and view the testing para meters.
PRINTING FUNCTION
- Connect the balance to the micro printer using the RS-232 port.
- The test result will show on the display.
- Go to History.
- Press PRINT. The data will be printed.
COMPUTER CONNECTION
- Connect the balance to the computer using the RS- 232 port.
- Download the XCOM V2.2 software. You should receive this in an email with your purchase. If not, contact support@fristadenllc.com.
- When you conduct a test, the test result will show on the display.
- Go to History.
- Press PRINT. The data will be entered in the computer program ..
DATE AND TIME SETTINGS
- Press the calibration key in the main menu.
- Press the ti mer key.
- Set the date and time. Press confirm
Troubleshooting
THE MONITOR IS OFF
REASON | SOLUTION |
---|---|
• The display switch is not open. | Open the switch and touch screen again. |
• Transient Interference | Unplug and restart the machine. |
• Broken Fuse. | Change fuse. |
LOW OR NO LIGHT
REASON | SOLUTION |
---|---|
• Pan holder installed incorrectly. | Reinstall pan holder. |
• Memory calibration is damaged. | Recalibrate the instrument. |
OVERWEIGHT OR ZERO OVERWEIGHT
REASON | SOLUTION |
---|---|
• Heavy objects on the weighing pan. | Remove objects from weighing pan. |
• Over maximum load. | Reduce load. |
• Memory calibration is damaged. | Recalibrate the instrument. |
LARGE INACCURACIES DURING CALIBRATION
REASON | SOLUTION |
---|---|
• Incorrect calibration weight. | Use correct weight. |
• Sample pan contains additional object. | Remove object, tare, and re- |
calibrate.
UNSTABLE WEIGHING RESULT
REASON | SOLUTION |
---|---|
• High airflow. | Avoid, or move away from drafty areas. |
• Unstable platform. | Place balance on a stable platform |
SAMPLE IS TOO HEAVY
REASON | SOLUTION |
---|---|
• The sample weighs more than the maximum load. | Reduce the load. |
INACCURATE WEIGHING RESULT
REASON | SOLUTION |
---|---|
• Not 0 before weighing. | Calibrate the instrument before using. |
• Balance used before weighing, | Touch Okey to tare balance. |
UNRESPONSIVE TOUCHSCREEN
REASON | SOLUTION |
---|---|
• Transient interference. | Reboot the machine. |
• Incorrect power supply voltage. | Use the correct power supply. |
Maintenance
MAINTENANCE AND STORAGE
- IMPORTANT: Do not touch halogen light to avoid damage or injury.
- Keep the moisture analyzer plugged in to reduce warm-up time.
- Power off the moisture analyzer between uses.
- The moisture analyzer should be kept clean. Beware of dust and other items getting into the moisture analyzer.
- Always power off, unplug and wait for the unit to cool completely before cleaning or storing it.
- Do not clean the machine with elements that may corrode the machine. Use a mild detergent and a wet, smooth cloth to clean.
- Make sure no liquid enters the balance.
- Dry the balance after cleaning.
Company Information
WARRANTY
The Moisture Analyzer has al Year Manufacturer’s Warranty. Please contact the
Fristaden Lab customer service department at
support@fristadenllc.com for more
information.
ABOUT US
Fristaden Lab is a lab and scientific equipment retailer established in 2013
in Chicago, Illinois.
We sell an array of popular products, including analytical balances,
centrifuges, glassware, hematology counters, hot plates, magnetic stirrers, pH
meters, pipettes and pressure pumps.
We stand behind all of our products, offering a 1-2year warranty, 30-day
returns, and a responsive customer service team committed to solving any
questions or issues you may have.
CONTACT
FristadenLab.com
support@fristadenllc.com
Owned and operated by
Fristaden & Company, LLC
401 Ryland Street, Suite 200-A
Reno NV, 89502 USA
Made in China