Information on the use of these instructions

Symbols

FIG 1 Symbols.JPG

You can download the current version of the instructions and the EU declaration of conformity via the following link:

Safety

Read this manual carefully before starting or using the device. Always store the manual in the immediate vicinity of the device or its site of use.

Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may result in electric shock, fire and/or serious injury.
Save all warnings and instructions for future reference.

  • Do not use the device in potentially explosive rooms or areas and do not install it there.
  • Do not use the device in aggressive atmosphere.
  • This appliance is not a toy. Keep away from children and animals.
  • Never immerse the entire device in a liquid. Only the measuring probe is intended to be immersed.
  • Protect the device from permanent direct sunlight.
  • Do not open the device.
  • Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
  • Never charge batteries that cannot be recharged.
  • Different types of batteries and new and used batteries must not be used together.
  • Insert the batteries into the battery compartment according to the correct polarity.
  • Remove discharged batteries from the device. Batteries contain materials hazardous to the environment. Dispose of the batteries according to the national regulations.
  • Remove the batteries from the device if you will not be using the device for a longer period of time.
  • Never short-circuit the supply terminal in the battery compartment!
  • Do not swallow batteries! If a battery is swallowed, it can cause severe internal burns within 2 hours! These burns can lead to death!
  • If you think batteries might have been swallowed or otherwise entered the body, seek medical attention immediately!
  • Keep new and used batteries and an open battery compartment away from children.
  • Only use the device, if sufficient safety precautions were taken at the surveyed location (e.g. when performing measurements along public roads, on building sites etc.). Otherwise do not use the device.
  • Observe the storage and operating conditions (see Technical data).
  • Check accessories and connection parts for possible damage prior to every use of the device. Do not use any defective devices or device parts.

Intended use
Only use the device to determine the pH value and temperature of liquids in aquaria, ponds, swimming pools or in food.
Any use other than the intended use is regarded as misuse.

Reasonably foreseeable misuse
Do not use the device in potentially explosive atmospheres or for measurements at live parts.
The device is not designed to be used under laboratory conditions.
Any unauthorised modifications, alterations or structural changes to the device are forbidden.
Personnel qualification

People who use this device must:

  • be aware of the dangers that occur when handling corrosive liquids such as bases and acids.
  • have read and understood the instructions, especially the Safety chapter.

Residual risks

FIG 3 Residual risks.JPG

FIG 4 Residual risks.JPG

Notice
Do not use abrasive cleaners or solvents to clean the device.

Information about the device

Device description
The pH measuring device BW10 is suited for measuring pH values and temperatures of liquids.
With the automatic temperature compensation and the automatic calibration functions the measuring device permits a quick and precise determination of the pH value between pH 0 and pH 14 in the temperature measuring range between 0 and 50 °C.

The three-point calibration is already carried out before leaving the factory, but it can also be repeated using the supplied pH buffer solution set.

The determined pH value is displayed along with the water temperature.
The device comes equipped with an exchangeable pH electrode.
The electrode and the water temperature probe are protected from external influences by means of a removable protective cap.
The device is further equipped with a Hold function for holding the currently measured value.
The auto-off function switches the device off after 15 minutes and can be deactivated if required.

Device depiction

FIG 5 Device depiction.JPG

Display

FIG 7 Display.JPG

Technical data

FIG 8 Technical data.JPG

Scope of delivery

  • 1 x pH measuring device
  • 4 x 1.5 V LR44 batteries
  • 1 x pH buffer solution set (4.01/7.00/10.01)
  • 1 x Quick guide

Transport and storage

Notice
If you store or transport the device improperly, the device may be damaged.
Note the information regarding transport and storage of the device.

Transport
For transporting the device use a suitable bag to protect it from external influences.
Make sure that the protective cap is attached to the measuring probe.

Storage
When the device is not being used, observe the following storage conditions:

  • dry and protected from frost and heat
  • protected from dust and direct sunlight
  • with a cover to protect it from invasive dust if necessary
  • the storage temperature complies with the values specified in the Technical data
  • Batteries are removed from the device

Operation

Inserting the batteries
Insert the supplied batteries into the device before first use.

Notice
Make sure that the surface of the device is dry and the device is switched off.

1. Open the battery compartment by unscrewing the lid (1).

2. Insert the batteries (4 x LR44 button cells) into the battery compartment with correct polarity.

3. Screw the lid back on the battery compartment. In doing so, make sure that the rubber seal is properly seated.

Removing the protective cap
The device has already been calibrated at the factory using a highly concentrated saline solution. This often leads to the formation of salt crystals at the diaphragm electrode (fibre bundle) during later usage. These visible residues are completely normal for pH measuring electrodes of this type. Such residues are harmless for the measuring electrode and can easily be washed off with water; thus they do not represent any quality defect!

The measuring electrode is protected from external influences by means of a protective cap (9).

1. Prior to every measurement pull the protective cap off the measuring electrode (8).

2. After every measurement put the protective cap back on the electrode.

Switching the device on
1. Press the Power (5) button to switch the device on.

Performing a calibration

Info
Before calibration immerse the measuring probe (7) in distilled water for approx. 10 to 15 min.

To perform a calibration of the pH measuring device, you first have to prepare the buffer solutions contained in the supplied set.

The buffer solutions correspond to the pH values 4 (red), 7 (green) and 10 (blue) with an accuracy of ±0.01 pH (at 25 °C). Please proceed as follows to prepare a buffer solution:

Caution
The powder for the buffer solutions as well as the buffer solutions made thereof are unfit for human consumption.
Caution
Sodium carbonate (Na2CO3, contained in buffer solution 10.01) may cause eye irritation. Upon contact with eyes, thoroughly rinse with water and seek medical advice.
Keep sodium carbonate out of the reach of children.
Do not breathe dust containing sodium carbonate.

1. Fill the contents of a sachet (e.g. pH 7 = green) into a beaker glass or suitable glass vessel with a capacity of at least 250 ml.
2. Add 250 ml of distilled water.
3. Stir the solution with a glass rob until the powder is completely dissolved.

Please note that the pH value of the buffer solution may vary depending on the temperature. The table below indicates the pH value as a function of the temperature (see also imprint on plastic sachets):

FIG 13.JPG

To perform the calibration, please proceed as follows:
Info

Always use fresh buffer solutions for calibration.

  1. Press the Power (5) button to switch the device on.
    The currently measured pH and temperature values are indicated on the display (2).

  2. With light circular motions, immerse the measuring probe (7) in the buffer solution for the pH value 7 first.
    The measuring electrode (8) should be completely surrounded by the buffer solution.

  3. Wait until a stable measured value is shown on the measurement value display (10).

  4. Press the CAL button (3) for approx. 3 s.
    CAL is displayed on the measurement value display.

  5. Let go of the CAL button.
    After approx. 2 s, SA is indicated on the measurement value display (10).
    Afterwards, End is displayed.
    Calibration for pH value 7 is now completed and the current measurement result is displayed.

  6. Repeat steps 2 to 5 for the buffer solution which is closest to the expected pH value.

Info
If calibration is aborted, this is either due to a wrong buffer solution or to a defective measuring electrode (8).
First check whether the correct buffer solution was used by means of a second measuring device if necessary. If the buffer solution corresponds to the pH value to be calibrated, the measuring electrode is defective and must be replaced, see chapter Replacing the measuring electrode.

Carrying out a measurement

  1. Immerse the measuring probe (7) in the liquid to be measured.
    The measuring electrode (8) should be completely surrounded by the liquid.

  2. Press the Power (5) button to switch the device on.
    The currently measured pH and temperature values are indicated on the display (2).

Info
If the measured value is outside the measuring range, this is indicated on the display (2).
The measurement value display (10) will show — for the pH value and either L or H for too low or too high temperature values.

Changing the unit °C / °F
In default setting, the measuring device is set to °C.
Please proceed as follows to change between the units °C and °F:
The measuring device is switched off.

  1. Simultaneously press the CAL (3) and Power button (5) until the unit °C (11) or °F (13) is indicated on the display (2).
  2. Press the CAL button (3) to switch between °C and °F.
  3. Press the HOLD button (4) to save the setting.
    SA appears on the display.
    The setting is saved and the current measurement result is displayed.

Holding a measured value (HOLD)

  1. Press the HOLD button (4) to freeze the currently measured value.
  2. Press the HOLD button again to show the currently measured values again.

Switching the device off
The device comes equipped with an automatic switch-off function and switches off automatically if no button was pressed for approx. 15 min. Please proceed as follows to deactivate the automatic switch-off function if required.

The device is switched on.

  1. Press the CAL button (3) once to deactivate the automatic switch-off function.
    Off appears briefly on the display (2).
    The cursor of the measured value display (10) does not flash.
    Automatic switch-off is deactivated.

  2. Press the CAL button (3) again to reactivate the automatic switch-off function.
    On appears briefly on the display (2).
    The cursor of the measured value display (10) flashes.
    Automatic switch-off is activated.

Please proceed as follows to switch off the device manually:
1. Press the Power (5) button to switch the device off.

Maintenance and repair

Battery change
A battery change is required when the device can no longer be switched on (see chapter Inserting the battery).

Replacing the measuring electrode
The measuring electrode must be replaced if it is damaged or can no longer be properly calibrated. This manifests itself in the fact that calibration is aborted despite using the correct buffer solution.

1. Pull the protective cap (9) off the measuring electrode (8).

2. Loosen the screwing at the measuring electrode and slide it downwards.

3. Pull the measuring electrode and the seal ring off the device.

4. Place the new seal ring onto the new measuring electrode.
5. Put the new measuring electrode on the device. In doing so, pay attention to the guide rails and connections.

FIG 18 Replacing the measuring electrode.JPG

6. Put the screwing back on and tighten it firmly.

7. Calibrate the device with the new measuring electrode, see chapter Operation.

Cleaning
Clean the device with a soft, damp and lint-free cloth. Make sure that no moisture enters the housing. Do not use any sprays, solvents, alcohol-based cleaning agents or abrasive cleaners, but only clean water to moisten the cloth.

Cleaning the measuring probe
When cleaning the measuring probe, proceed with great caution:

  • Rinse the electrode with distilled water.
  • Avoid unnecessary friction / contact with the glass ball electrode as it may be damaged or age faster.

Repair
Do not modify the device or install any spare parts. For repairs or device testing, contact the manufacturer.

Disposal

Always dispose of packing materials in an environmentally friendly manner and in accordance with the applicable local disposal regulations.

The icon with the crossed-out waste bin on waste electrical or electronic equipment is taken from Directive 2012/19/EU. It states that this device must not be disposed of with the household waste at the end of its life. You will find collection points for free return of waste electrical and electronic equipment in your vicinity. The addresses can be obtained from your municipality or local administration. You can also find out about other return options that apply for many EU countries on the website https://hub.trotec.com/?id=45090. Otherwise, please contact an official recycling centre for electronic and electrical equipment authorised for your country.

The separate collection of waste electrical and electronic equipment aims to enable the re-use, recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment.

In the European Union, batteries and accumulators must not be treated as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of the European Parliament and of the Council of 6 September 2006 on batteries and accumulators. Please dispose of batteries and accumulators according to the relevant legal requirements.

Only for United Kingdom
According to Waste Electrical and Electronic Equipment Regulations 2013 (SI 2013/3113) (as amended) and the Waste Batteries and Accumulators Regulations 2009 (SI 2009/890) (as amended), devices that are no longer usable must be collected separately and disposed of in an environmentally friendly manner.

Trotec GmbH
Grebbener Str. 7
D-52525 Heinsberg
+49 2452 962-400
+49 2452 962-200
info@trotec.com
www.trotec.com

Read More About This Manual & Download PDF:

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals