TiePie engineering HS3 Handy Scope Instruction Manual

May 15, 2024
TiePie engineering

TiePie engineering HS3 Handy Scope

TiePie-engineering-HS3-Handy-Scope-product

Specifications

  • Acquisition system: High-performance data acquisition system
  • Trigger system: Advanced trigger system for precise measurements
  • Arbitrary Waveform Generator: Ability to generate custom waveforms
  • Interface: USB interface for connection to computer
  • Power: Can be powered via USB or external power adapter
  • Physical: Compact and portable design
  • I/O connectors: BNC connectors for signal input/output
  • System requirements: Compatible with Windows 8, Windows 10
  • Environmental conditions: Operating temperature: 0-40°C, Storage temperature: -20-70°C

Product Usage Instructions

Safety

  • When using the Handyscope HS3, always ensure proper safety measures are taken to avoid any electrical hazards.
  • Avoid measuring directly online voltage without proper isolation equipment.

Driver Installation

  • Before using the device, install the necessary drivers based on your computer’s operating system.
  • Follow the instructions provided in the manual for driver installation.

Hardware Installation

  • Power the instrument using either the external power adapter or USB power cable.
  • Connect the instrument to your computer via the USB port.
  • Ensure a stable connection for accurate data transfer.

Front Panel

  • The front panel of the Handyscope HS3 features input connectors for Channel 1 and Channel 2, as well as a GENERATOR output connector.
  • The power indicator shows the status of the device.

Rear Panel

  • The rear panel includes power options such as a USB power cable or a power adapter.
  • Additionally, there is an Extension Connector for expanding connectivity options.

FAQ

Q: Can I measure line voltage directly with the Handyscope HS3?

  • A: It is not recommended to measure line voltage directly with the Handyscope HS3 due to safety concerns.
  • Always use proper isolation equipment when working with high voltages.

Q: What are the system requirements for the Handyscope HS3?

  • A: The Handyscope HS3 is compatible with computers running Windows 8 or Windows 10.
  • Ensure your system meets the minimum requirements for optimal performance.

ATTENTION!

  • Measuring directly on the line voltage can be very dangerous.
  • The outside of the BNC connectors at the Handyscope HS3 are connected to the ground of the computer.
  • Use a good isolation transformer or a differential probe when measuring the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS3 is connected to a positive voltage.
  • This short-circuit current can damage both the Handyscope HS3 and the computer.

Safety

  • When working with electricity, no instrument can guarantee complete safety.
  • It is the responsibility of the person who works with the instrument to operate it safely.
  • Maximum security is achieved by selecting the proper instruments and following safe working procedures.

Safe working tips are given below:

  • Always work according to (local) regulations.
  • Work on installations with voltages higher than 25 VAC or 60 VDC should only be performed by qualified personnel.
  • Avoid working alone.
  • Observe all indications on the Handyscope HS3 before connecting any wiring
  • Check the probes/test leads for damages. Do not use them if they are damaged
  • Take care when measuring voltages higher than 25 VAC or 60 VDC.
  • Do not operate the equipment in an explosive atmosphere or the presence of flammable gases or fumes.
  • Do not use the equipment if it does not operate properly. Has the equipment been inspected by a qualified service personnel? If necessary, return the equipment to TiePie Engineering for service and repair to ensure that safety features are maintained.
  • Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS3 are connected to the ground of the computer. Use a good isolation transformer or a differ-initial probe when measuring the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS3 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS3 and the computer.

Declaration of conformity

  • TiePie engineering Koperslagersstraat 37 8601 WL Sneek
  • The Netherlands

EC Declaration of Conformity

TiePie-engineering-HS3-Handy-Scope-fig-1

Environmental considerations

  • This section provides information about the environmental impact of the Handy-scope HS3.

End-of-life handling

  • Production of the Handyscope HS3 required the extraction and use of natural resources.
  • The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the Handyscope HS3’s end of life.
  • To avoid the release of such substances into the environment and to reduce the use of natural resources, recycle the Handyscope HS3 in an appropriate sys-tem that will ensure that most of the materials are reused or recycled appropriately.
  • The shown symbol indicates that the Handyscope HS3 complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE).

Introduction

  • Before using the Handyscope HS3 first read Chapter 1 about safety.
  • Many technicians investigate electrical signals. Though the measurement may not be electrical, the physical variable is often converted to an electrical signal, with a special transducer.
  • Common transducers are accelerometers, pressure probes, current clamps, and temperature probes. The advantages of converting the physical parameters to electrical signals are large since many instruments for examining electrical signals are available.
  • The Handyscope HS3 is a portable two-channel measuring instrument with an Arbitrary Waveform Generator.
  • The Handyscope HS3 is available in several models with different maximum sampling rates: 5 MSa/s, 10 MSa/s, 25 MSa/s, 50 MSa/s or 100 MSa/s.
  • The native resolution is 12 bits, but user selectable resolutions of 8, 14 and 16 bits are available too, with adjusted maximum sampling rate:

Resolution| HS3-100| HS3-50| HS3-25| HS3-10| HS3-5
---|---|---|---|---|---
8 bit| 100 MSa/s| 50 MSa/s| 25 MSa/s| 10 MSa/s| 5 MSa/s
12 bit| 50 MSa/s| 50 MSa/s| 25 MSa/s| 10 MSa/s| 5 MSa/s
14 bit| 3.125 MSa/s| 3.125 MSa/s| 3.125 MSa/s| 3.125 MSa/s| 3.125 MSa/s
16 bit| 195 kSa/s| 195 kSa/s| 195 kSa/s| 195 kSa/s| 195 kSa/s

Table 3.1: Maximum sampling rates

  • With the accompanying software, the Handyscope HS3 can be used as an oscilloscope, a spectrum analyzer, a true RMS voltmeter, or a transient recorder.
  • All instruments measure by sampling the input signals, digitizing the values, processing them, saving them, and displaying them.

Sampling

  • When sampling the input signal, samples are taken at fixed intervals. At these intervals, the size of the input signal is converted to a number.
  • The accuracy of this number depends on the resolution of the instrument. The higher the resolution, the smaller the voltage steps in which the input range of the instrument is divided.
  • The acquired numbers can be used for various purposes, e.g. to create a graph.
  • The sine wave in figure 3.1 is sampled at the dot positions. By connecting the adjacent samples, the original signal can be reconstructed from the samples. You can see the result in figure 3.2.

Sampling rate

  • The rate at which the samples are taken is called the sampling rate, the number of samples per second. A higher sampling rate corresponds to a shorter interval between the samples.
  • As is visible in figure 3.3, with a higher sampling rate, the original signal can be reconstructed much better from the measured samples.
  • The sampling rate must be higher than 2 times the highest frequency in the input signal. This is called the Nyquist frequency.
  • Theoretically, it is possible to reconstruct the input signal with more than 2 samples per period.
  • In practice, 10 to 20 samples per period are recommended to be able to examine the signal thoroughly.

Aliasing

  • When sampling an analog signal with a certain sampling rate, signals appear in the output with frequencies equal to the sum and difference of the signal frequency and multiples of the sampling rate.
  • For example, when the sampling rate is 1000 Sa/s and the signal frequency is 1250 Hz, the following signal frequencies will be present in the output data:
Multiple sampling rate 1250 Hz signal -1250 Hz signal
 
-1000 -1000 + 1250 = 250 -1000 – 1250 = -2250
0 0 + 1250 = 1250 0 – 1250 = -1250
1000 1000 + 1250 = 2250 1000 – 1250 = -250
2000 2000 + 1250 = 3250 2000 – 1250 =  750
 

Table 3.2: Aliasing

  • As stated before, when sampling a signal, only frequencies lower than half the sampling rate can be reconstructed. In this case, the sampling rate is 1000 Sa/s, so we can only observe signals with a frequency ranging from 0 to 500 Hz.
  • This means that from the resulting frequencies in the table, we can only see the 250 Hz signal in the sampled data. This signal is called an alias of the original signal.
  • If the sampling rate is lower than twice the frequency of the input signal, aliasing will occur. The following illustration shows what happens.
  • In figure 3.4, the green input signal (top) is a triangular signal with a frequency of 1.25 kHz. The signal is sampled with a rate of 1 kSa/s. The corresponding sampling interval is 1/1000Hz = 1 ms.
  • The positions at which the signal is sampled are depicted with blue dots. The red dotted signal (bottom) is the result of the reconstruction.
  • The period time of this triangular signal appears to be 4 ms, which corresponds to an apparent frequency (alias) of 250 Hz (1.25 kHz – 1 kHz).
  • To avoid aliasing, always start measuring at the highest sampling rate and lower the sampling rate if required.

Digitizing

  • When digitizing the samples, the voltage at each sample time is converted to a number. This is done by comparing the voltage with many levels.
  • The resulting number is the number corresponding to the level that is closest to the voltage.
  • The number of levels is determined by the resolution, according to the following relation: LevelCount = 2Resolution.
  • The higher the resolution, the more levels are available, and the more accurately the input signal can be reconstructed.
  • In figure 3.5, the same signal is digitized, using two different amounts of levels: 16 (4-bit) and 64 (6-bit).
  • The Handyscope HS3 measures at e.g. 12-bit resolution (212=4096 levels). The smallest detectable voltage step depends on the input range.
  • This voltage can be calculated as V oltageStep = F ullInputRange/LevelCount
  • For example, the 200 mV range ranges from -200 mV to +200 mV, therefore the full range is 400 mV.
  • This results in the smallest detectable voltage step of 0.400 V / 4096 = 97.65 µV.

Signal coupling

  • The Handyscope HS3 has two different settings for the signal coupling: AC and DC. In the setting DC, the signal is directly coupled to the input circuit.
  • All signal components available in the input signal will arrive at the input circuit and will be measured.
  • In the setting AC, a capacitor will be placed between the input connector and the input circuit. This capacitor will block all DC components of the input signal and let all AC components pass through.
  • This can be used to remove a large DC component of the input signal, to be able to measure a small AC component at high resolution.
  • When measuring DC signals, make sure to set the signal coupling of the input to DC.

Probe compensation.

  • The Handyscope HS3 is shipped with a probe for each input channel. These are 1x/10x selectable passive probes.
  • This means that the input signal is passed through directly or 10 times attenuated.
  • When using an oscilloscope probe in a 1:1 setting, the bandwidth of the probe is only 6 MHz.
  • The full bandwidth of the probe is only obtained in the 1:10 setting
  • The x10 attenuation is achieved through an attenuation network. This attenuation network has to be adjusted to the oscilloscope input circuitry, to guarantee frequency independency.
  • This is called low-frequency compensation. Each time a probe is used on another channel or another oscilloscope, the probe must be adjusted.
  • Therefore the probe is equipped with a setscrew, with which the parallel capacity of the attenuation network can be altered.
  • To adjust the probe, switch the probe to the x10 and attach the probe to a 1 kHz square wave signal.
  • Then adjust the probe for a square front corner on the square wave displayed. See also the following illustrations.TiePie-engineering-HS3-Handy-Scope-fig-7

Driver installation

  • Before connecting the Handyscope HS3 to the computer, the drivers need to be installed.

Introduction

  • To operate a Handyscope HS3, a driver is required to interface between the measurement software and the instrument. This driver takes care of the low-level communication between the computer and the instrument, through USB.
  • When the driver is not installed, or an old, no longer compatible version of the driver is installed, the software will not be able to operate the Handyscope HS3 properly or even detect it at all.

Computers running Windows 10

  • When the Handyscope HS3 is plugged into a USB port of the computer, Windows will detect the instrument and will download the required driver from Windows Update.
  • When the download is finished, the driver will be installed automatically.

Computers running Windows 8 or older

  • The installation of the USB driver is done in a few steps. Firstly, the driver has to be pre-installed by the driver setup program.
  • This makes sure that all required files are located where Windows can find them.
  • When the instrument is plugged in, Windows will detect new hardware and install the required drivers.

Where to find the driver setup

  • The driver setup program and measurement software can be found in the download section on TiePie Engineering’s website.
  • It is recommended to install the latest version of the software and USB driver from the website. This will guarantee the latest features are included.

Executing the installation utility

  • To start the driver installation, execute the downloaded driver setup program. The driver install utility can be used for a first time installation of a driver on a system and also to update an existing driver.
  • The screenshots in this description may differ from the ones displayed on your computer, depending on the Windows version.
  • When drivers are already installed, the install utility will remove them before installing the new driver. To remove the old driver successfully, the Handyscope HS3 must be disconnected from the computer before starting the driver install utility.
  • When the Handyscope HS3 is used with an external power supply, this must be disconnected too.
  • Clicking ”Install” will remove existing drivers and install the new driver. A remove entry for the new driver is added to the software applet in the Windows control panel.

Hardware installation

  • Drivers have to be installed before the Handyscope HS3 is connected to the computer for the first time. See chapter 4 for more information.

Power the instrument

  • The Handyscope HS3 is powered by the USB, no external power supply is required. Only connect the Handyscope HS3 to a bus-powered USB port, otherwise it may not get enough power to operate properly.

External power

  • In certain cases, the Handyscope HS3 cannot get enough power from the USB port. When a Handyscope HS3 is connected to a USB port, powering the hardware will result in an inrush current higher than the nominal current. After the inrush current, the current will stabilize at the nominal current.
    USB ports have a maximum limit for both the inrush current peak and the nominal current. When either of them is exceeded, the USB port will be switched off. As a result, the connection to the Handyscope HS3 will be lost.

  • Most USB ports can supply enough current for the Handyscope HS3 to work without an external power supply, but this is not always the case. Some (battery-operated) portable computers or (bus-powered) USB hubs do not supply enough current.

  • The exact value at which the power is switched off varies per USB controller, so it is possible that the Handyscope HS3 functions properly on one computer, but does not on another.

  • To power the Handyscope HS3 externally, an external power input is provided. It is located at the rear of the Handyscope HS3. Refer to paragraph 7.1 for specifications of the external power input.

Connect the instrument to the computer

  • After the new driver has been pre-installed (see Chapter 4), the Handyscope HS3 can be connected to the computer. When the Handyscope HS3 is connected to a USB port of the computer, Windows will detect new hardware.
  • Depending on the Windows version, a notification can be shown that new hardware is found and that drivers will be installed. Once ready, Windows will report that the driver is installed.
  • When the driver is installed, the measurement software can be installed and the Handyscope HS3 can be used.

Plug into a different USB port

  • When the Handyscope HS3 is plugged into a different USB port, some Windows versions will treat the Handyscope HS3 as different hardware and will install the drivers again for that port.
  • This is controlled by Microsoft Windows and is not caused by TiePie engineering.

Front panel

TiePie-engineering-HS3-Handy-Scope-fig-11

CH1 and CH2 input connectors

  • The CH1 and CH2 BNC connectors are the main inputs of the acquisition sys-tem. The outside of the BNC connectors is connected to the ground of the Handy-scope HS3. Connecting the outside of the BNC connector to a potential other than ground will result in a short circuit that may damage the device under test, the Handyscope HS3, and the computer.

GENERATOR output connector

  • The OUT BNC connector is the output of the internal Arbitrary Waveform Generator. The outside of this BNC connector is connected to the ground of the Handy-scope HS3.
  • When the generator is switched off in the software, the generator output is switched to a high impedance, floating state, and the output voltage is then undefined.
  • When the generator is switched on in the software and set to pause, the generator output is switched to a low impedance (50 Ω), and the output voltage is at the selected offset level.

Power indicator

  • A power indicator is situated at the top cover of the instrument. It is lit when the Handyscope HS3 is powered.

Rear panelTiePie-engineering-HS3-Handy-Scope-fig-12

Power

  • The Handyscope HS3 is powered through the USB. If the USB cannot deliver enough power, it is possible to power the instrument externally.
  • The Handyscope HS3 has two external power inputs located at the rear of the instrument: the dedicated power input and a pin of the extension connector.
  • Handyscope HS3s with SN# 11832 and lower do not have a dedicated power input at the rear, they only have an external power input on the extension connector.

The specifications of the dedicated power connector are:

Pin Dimension Description

Center pin

Outside bushing

| Ø1.3 mm

Ø3.5 mm

| ground

positive

Figure 7.2: Power connector

  • Besides the external power input, it is also possible to power the instrument through the extension connector, the 25-pin D-sub connector at the rear of the instrument.

  • The power has to be applied to pin 3 of the extension connector. Pin 4 can be used as ground.

The following minimum and maximum voltages apply to both power inputs:**  | Minimum| Maximum**
---|---|---
SN# <12941| 4.5 VDC| 6 VDC
SN# >12941| 4.5 VDC| 12 VDC

Table 7.1: Maximum voltages

  • Note that the externally applied voltage should be higher than the USB voltage to relieve the USB port.

USB power cable

  • The Handyscope HS3 is delivered with a special USB external power cable.
  • One end of this cable can be connected to a second USB port on the computer, the other end can be plugged into the external power input at the rear of the instrument. The power for the instrument will be taken from two USB ports of the computer.
  • The outside of the external power connector is connected to +5 V. To avoid a shortage, first connects the cable to the Handyscope HS3 and then to the USB port.

Power adapter

  • In case a second USB port is not available, or the computer still can’t provide enough power for the instrument, an external power adapter can be used. When using an external power adapter, make sure that:
  • the polarity is set correctly
  • the voltage is set to a valid value for the instrument and higher than the USB voltage
  • the adapter can supply enough current (preferably >1 A)
  • the plug has the correct dimensions for the external power input of the instrument

USB

  • The Handyscope HS3 is equipped with a USB 2.0 High speed (480 Mbit/s) interface with a fixed cable with type A plug. It will also work on a computer with a USB 1.1 interface, but will then operate at 12 Mbit/s.

Extension Connector

To connect to the Handyscope HS3 a 25-pin female D-sub connector is available, containing the following signals:

Pin Description Pin Description
1 Ground 14 Ground
2 Reserved 15 Ground
3 External Power in DC 16 Reserved
4 Ground 17 Ground
5 +5V out, 10 mA max. 18 Reserved
6 Ext. sampling clock in (TTL) 19 Reserved
7 Ground 20 Reserved
8 Ext. trigger in (TTL) 21 Generator Ext Trig in (TTL)
9 Data OK out (TTL) 22 Ground
10 Ground 23 I2 C SDA
11 Trigger out (TTL) 24 I2 C SCL
12 Reserved 25 Ground
13 Ext. sampling clock out (TTL)  

Table 7.2: Pin description Extension connector

  • All TTL signals are 3.3 V TTL signals which are 5 V tolerant, so they can be connected to 5 V TTL systems.
  • For instruments with serial numbers 14266 and higher, pins 9, 11, 12, and 13 are open collector outputs. Connect a pull-up resistor of 1 kOhm to pin 5 when using one of these signals.
  • For older instruments, the outputs are standard TTL outputs and no pull-up is required.

Specifications

Acquisition systemTiePie-engineering-HS3-Handy-Scope-
fig-15

Trigger system

TiePie-engineering-HS3-Handy-Scope-fig-16

Arbitrary Waveform GeneratorTiePie-engineering-HS3-Handy-Scope-
fig-17

InterfaceTiePie-engineering-HS3-Handy-Scope-fig-18

Power

Physical

connectors

System requirements

Environmental conditions

Certifications and Compliances

Probes

TiePie-engineering-HS3-Handy-Scope-fig-25

Package contentsTiePie-engineering-HS3-Handy-Scope-
fig-26

  • If you have any suggestions and/or remarks regarding this manual, please contact: TiePie Engineering Koperslagersstraat 37 8601 WL SNEEK The Netherlands
  • Tel.: +31 515 415 416
  • Fax: +31 515 418 819
  • E-mail:support@tiepie.nl
  • Site:www.tiepie.com.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals