TiePie engineering Handyscope HS5 USB Oscilloscope Instruction Manual

June 15, 2024
TiePie engineering

unnamed

TiePie engineering Handyscope HS5 USB Oscilloscope

Specifications

  • Acquisition system: 12-bit ADC
  • Sampling rate: up to 100 MS/s
  • Input channels: 2
  • Input voltage range: ±20 V
  • Trigger system: analog, digital, external
  • Arbitrary Waveform Generator: 14-bit DAC, 1 MS/s
  • Power: USB-powered or external power adapter (12-30 V DC)

Product Usage Instructions

Safety
Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS5 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS5 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS5 and the computer.

Driver Installation
Before using the Handyscope HS5, you need to install the driver. Follow the instructions below based on your computer’s operating system:

Computers running Windows 10

  1. Download the driver setup file from the manufacturer’s website.
  2. Run the installation utility and follow the on-screen instructions to complete the installation.

Computers running Windows 8 or older

  1. Download the driver setup file from the manufacturer’s website.
  2. Locate the driver setup file on your computer.
  3. Run the installation utility and follow the on-screen instructions to complete the installation.

Hardware Installation
Once the driver is installed, you can proceed with the hardware installation:

Power the instrument

  • Connect the Handyscope HS5 to a power source using either the external power adapter or the USB power cable.
  • Connect the instrument to the computer
  • Connect one end of the USB cable to the Handyscope HS5 and the other end to an available USB port on your computer.

Plug into a different USB port
If you encounter any connection issues, try plugging the USB cable into a different USB port on your computer.

Operating conditions
Ensure that the Handyscope HS5 is operated within the specified operating conditions mentioned in the user manual.

Combining and Synchronizing Instruments
If you have multiple Handyscope HS5 devices, you can combine and synchronize them for multi-channel measurements. Refer to the user manual for detailed instructions on combining and synchronizing instruments.

Front Panel
The front panel of the Handyscope HS5 includes the following:

  • CH1 and CH2 input connectors
  • AWG output connector
  • Power indicator

Rear Panel
The rear panel of the Handyscope HS5 includes the following:

  • Power connection (power adapter or USB power cable)
  • USB port
  • Extension Connector
  • AUX I/O

FAQ

Can I measure line voltage directly with the Handyscope HS5?
No, measuring directly on the line voltage can be dangerous. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies.

How many input channels does the Handyscope HS5 have?
The Handyscope HS5 has 2 input channels.

What is the maximum sampling rate of the Handyscope HS5?
The Handyscope HS5 has a maximum sampling rate of up to 100 MS/s.

How is the Handyscope HS5 powered?
The Handyscope HS5 can be powered either through an external power adapter (12-30 V DC) or via USB power.

ATTENTION!
Measuring directly on the line voltage can be very dangerous.
The outside of the BNC connectors at the Handyscope HS5 are connected with the ground of the computer. Use a good isolation transformer or a dif-ferential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS5 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS5 and the computer.

Copyright ©2023 TiePie engineering. All rights reserved. Revision 2.44, October 2023 This information is subject to change without notice. Despite the care taken for the compilation of this user manual, TiePie engineering can not be held responsible for any damage resulting from errors that may appear in this manual.

Safety

When working with electricity, no instrument can guarantee complete safety. It is the responsibility of the person who works with the instrument to operate it in a safe way. Maximum security is achieved by selecting the proper instruments and following safe working procedures. Safe working tips are given below:

  • Always work according (local) regulations.
  • Work on installations with voltages higher than 25 VAC or 60 VDC should only be performed by qualified personnel.
  • Avoid working alone.
  • Observe all indications on the Handyscope HS5 before connecting any wiring
  • Check the probes/test leads for damages. Do not use them if they are damaged
  • Take care when measuring at voltages higher than 25 VAC or 60 VDC.
  • Do not operate the equipment in an explosive atmosphere or in the presence of flammable gases or fumes.
  • Do not use the equipment if it does not operate properly. Have the equipment inspected by qualified service personal. If necessary, return the equipment to TiePie engineering for service and repair to ensure that safety features are maintained.
  • Measuring directly on the line voltage can be very dangerous. The outside of the BNC connectors at the Handyscope HS5 are connected with the ground of the computer. Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power supplies! A short-circuit current will flow if the ground of the Handyscope HS5 is connected to a positive voltage. This short-circuit current can damage both the Handyscope HS5 and the computer.

Environmental considerations
This section provides information about the environmental impact of the Handyscope HS5.

End-of-life handling
Production of the Handyscope HS5 required the extraction and use of natural resources. The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the Handyscope HS5’s end of life.

In order to avoid release of such substances into the environment and to reduce the use of natural resources, recycle the Handyscope HS5 in an appropriate system that will ensure that most of the materials are reused or recycled appropriately. The shown symbol indicates that the Handyscope HS5 complies with the European Union’s requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE).

Introduction

Before using the Handyscope HS5 first read chapter 1 about safety.

Many technicians investigate electrical signals. Though the measurement may not be electrical, the physical variable is often converted to an electrical signal, with a special transducer. Common transducers are accelerometers, pressure probes, current clamps and temperature probes. The advantages of converting the physical parameters to electrical signals are large, since many instruments for examining electrical signals are available.

The Handyscope HS5 is a portable two channel measuring instrument with Arbitrary Waveform Generator. The Handyscope HS5 is available in several models with different maximum sampling rates: 50 MSa/s, 100 MSa/s, 200 MSa/s or 500 MSa/s. The native resolutions are 8, 12 and 14 bits and a user selectable resolution of 16 bits is available too, with adjusted maximum sampling rates:

Handyscope HS5 Channels Resolution
8 / 12 bit 14 bit 16 bit
Model 540 CH1 500 MSa/s
CH1+CH2 200 MSa/s
Model 530 CH1 500 MSa/s
CH1+CH2 200 MSa/s
Model 220 CH1 200 MSa/s
CH1+CH2 100 MSa/s
Model 110 CH1 100 MSa/s
CH1+CH2 50 MSa/s
Model 055 CH1 50 MSa/s
CH1+CH2 20 MSa/s

Table 3.1: Maximum sampling rates

The Handyscope HS5 supports high speed continuous streaming measurements. The maximum streaming rates are:

Handyscope HS5 Channels Resolution
8 bit 12/14 bit 16 bit
Model 540 CH1 40 MSa/s

10 MSa/s

| 6.25 MSa/s
CH1+CH2| 20 MSa/s
Model 530| CH1| 40 MSa/s| 20 MSa/s

10 MSa/s

| 6.25 MSa/s
CH1+CH2| 20 MSa/s
Model 220| CH1| 20 MSa/s| 10 MSa/s

5 MSa/s

| 3.125 MSa/s
CH1+CH2| 10 MSa/s
Model 110| CH1| 10 MSa/s| 5 MSa/s

2 MSa/s

| 1.25 MSa/s
CH1+CH2| 4 MSa/s
Model 055| CH1| 4 MSa/s| 2 MSa/s

1 MSa/s

| 625 kSa/s
CH1+CH2| 2 MSa/s

Table 3.2 : Maximum streaming rates

The Handyscope HS5 is available with two memory configurations, these are:

Memory Model 540 Model 530 Model 220 Model 110 Model 055
Standard model 128 KiSa 128 KiSa 128 KiSa 128 KiSa 128 KiSa
Option XM 32 MiSa 32 MiSa 32 MiSa 32 MiSa 32 MiSa

Table 3.3 : Maximum record lengths per channel

Optionally available for the Handyscope HS5 are SureConnect connection test and resistance measurement. SureConnect connection test tells you immediately whether your test probe or clip actually makes electrical contact or not. No more doubt whether your probe doesn’t make contact or there really is no signal. This is useful when surfaces are oxidized and your probe cannot get a good electrical contact. Simply activate the SureConnect and you know whether there is contact or not. Also when back probing connectors in confined places, SureConnect immediately shows whether the probes make contact or not.

Models of the Handyscope HS5 with SureConnect come with resistance measurement on all channels. Resistances up to 2 MOhm can be measured directly. Resistance can be shown in meter displays and can also be plotted versus time in a graph, creating an Ohm scope.

With the accompanying software the Handyscope HS5 can be used as an oscilloscope, a spectrum analyzer, a true RMS voltmeter or a transient recorder. All instruments measure by sampling the input signals, digitizing the values, process them, save them and display them.

Sampling
When sampling the input signal, samples are taken at fixed intervals. At these intervals, the size of the input signal is converted to a number. The accuracy of this  number depends on the resolution of the instrument. The higher the resolution, the smaller the voltage steps in which the input range of the instrument is divided. The acquired numbers can be used for various purposes, e.g. to create a graph.

The sine wave in figure 3.1 is sampled at the dot positions. By connecting the adjacent samples, the original signal can be reconstructed from the samples. You can see the result in figure 3.2

Sampling rate

The rate at which the samples are taken is called the sampling rate, the number of samples per second. A higher sampling rate corresponds to a shorter interval between the samples. As is visible in figure 3.3, with a higher sampling rate, the original signal can be reconstructed much better from the measured samples.

The sampling rate must be higher than 2 times the highest frequency in the input signal. This is called the Nyquist frequency. Theoretically it is possible to recon- struct the input signal with more than 2 samples per period. In practice, 10 to 20 samples per period are recommended to be able to examine the signal thor- oughly.

Aliasing
When sampling an analog signal with a certain sampling rate, signals appear in the output with frequencies equal to the sum and difference of the signal frequency and multiples of the sampling rate. For example, when the sampling rate is 1000 Sa/s and the signal frequency is 1250 Hz, the following signal frequencies will be present in the output data:

Multiple of sampling rate 1250 Hz signal -1250 Hz signal
 
-1000 -1000 + 1250 = 250 -1000 – 1250 = -2250
0 0 + 1250 = 1250 0 – 1250 = -1250
1000 1000 + 1250 = 2250 1000 – 1250 = -250
2000 2000 + 1250 = 3250 2000 – 1250 = 750
 

Table 3.4: Aliasing

As stated before, when sampling a signal, only frequencies lower than half the sampling rate can be reconstructed. In this case the sampling rate is 1000 Sa/s, so we can we only observe signals with a frequency ranging from 0 to 500 Hz. This means that from the resulting frequencies in the table, we can only see the 250 Hz signal in the sampled data. This signal is called an alias of the original signal. If the sampling rate is lower than twice the frequency of the input signal, aliasing will occur. The following illustration shows what happens

In figure 3.4, the green input signal (top) is a triangular signal with a frequency of 1.25 kHz. The signal is sampled with a rate of 1 kSa/s. The corresponding sam- pling interval is 1/1000Hz = 1ms. The positions at which the signal is sampled are depicted with the blue dots. The red dotted signal (bottom) is the result of the re- construction. The period time of this triangular signal appears to be 4 ms, which corresponds to an apparent frequency (alias) of 250 Hz (1.25 kHz – 1 kHz).

WARNING : To avoid aliasing, always start measuring at the highest sampling rate and lower the sampling rate if required.

Digitizing

When digitizing the samples, the voltage at each sample time is converted to a number. This is done by comparing the voltage with a number of levels. The re- sulting number is the number corresponding to the level that is closest to the voltage.

The number of levels is determined by the resolution, according to the following relation: LevelCount = 2Resolution.

The higher the resolution, the more levels are available and the more accurate the input signal can be reconstructed. In figure3.5, the same signal is digitized, using two different amounts of levels: 16 (4-bit) and 64 (6-bit). The Handyscope HS5 measures at e.g. 14 bit resolution (214=16384 levels). The smallest detectable voltage step depends on the input range.

This voltage can be calculated as: V oltageStep = FullInputRange/LevelCount For example, the 200 mV range ranges from -200 mV to +200 mV, therefore the full range is 400 mV. This results in a smallest detectable voltage step of 0.400 V / 16384 = 24.41 µ V.

Signal coupling
The Handyscope HS5 has two different settings for the signal coupling: AC and DC. In the setting DC, the signal is directly coupled to the input circuit. All signal components available in the input signal will arrive at the input circuit and will be measured.

In the setting AC, a capacitor will be placed between the input connector and the input circuit. This capacitor will block all DC components of the input signal and let all AC components pass through. This can be used to remove a large DC com- ponent of the input signal, to be able to measure a small AC component at high resolution.

WARNING: When measuring DC signals, make sure to set the signal coupling of the input to DC

Probe compensation
The Handyscope HS5 is shipped with a probe for each input channel. These are 1x/10x selectable passive probes. This means that the input signal is passed through directly or 10 times attenuated.

WARNING: When using an oscilloscope probe in 1:1 the setting, the bandwidth of the probe is only 6 MHz. The full bandwidth of the probe is only obtained in the 1:10 setting

The x10 attenuation is achieved by means of an attenuation network. This atten- uation network has to be adjusted to the oscilloscope input circuitry, to guaran- tee frequency independency. This is called the low frequency compensation. Each time a probe is used on an other channel or an other oscilloscope, the probe must be adjusted.

Therefore the probe is equiped with a setscrew, with which the parallel capacity of the attenuation network can be altered. To adjust the probe, switch the probe to the x10 and attach the probe to a 1 kHz square wave signal.  Then adjust    the probe for a square front corner on the square wave displayed. See also the following illustrations

TiePie-engineering-Handyscope-HS5-USB-Oscilloscope-fig-
\(7\)

Driver installation

Before connecting the Handyscope HS5 to the computer, the drivers need to be installed

Introduction
To operate a Handyscope HS5, a driver is required to interface between the mea- surement software and the instrument. This driver takes care of the low level communication between the computer and the instrument, through USB. When the driver is not installed, or an old, no longer compatible version of the driver is installed, the software will not be able to operate the Handyscope HS5 properly or even detect it at all.

Computers running Windows 10
When the Handyscope HS5 is plugged in into a USB port of the computer, Win- dows will detect the instrument and will download the required driver from Win- dows Update. When the download is finished, the driver will be installed automat- ically.

Computers running Windows 8 or older
The installation of the USB driver is done in a few steps. Firstly, the driver has to be pre-installed by the driver setup program. This makes sure that all required files are located where Windows can find them. When the instrument is plugged in, Windows will detect new hardware and install the required drivers.

Where to find the driver setup
The driver setup program and measurement software can be found in the down- load section on TiePie engineering’s website. It is recommended to install the latest version of the software and USB driver from the website. This will guarantee the latest features are included.

Executing the installation utility
To start the driver installation, execute the downloaded driver setup program. The driver install utility can be used for a first time installation of a driver on a system and also to update an existing driver. The screen shots in this description may differ from the ones displayed on your computer, depending on the Windows version.

When drivers were already installed, the install utility will remove them before in- stalling the new driver. To remove the old driver successfully, it is essential that the Handyscope HS5 is disconnected from the computer prior to starting the driver install utility. When the Handyscope HS5 is used with an external power supply, this must be disconnected too.

Clicking Install will remove existing drivers and install the new driver. A remove entry for the new driver is added to the software applet in the Windows control panel.

Hardware installation

Drivers have to be installed before the Handyscope HS5 is connected to the computer for the first time. See chapter 4 for more information.

Power the instrument
The Handyscope HS5 is powered by the USB, no external power supply is required. Only connect the Handyscope HS5 to a bus powered USB port, otherwise it may not get enough power to operate properly.

External power
In certain cases, the Handyscope HS5 cannot get enough power from the USB port. When a Handyscope HS5 is connected to a USB port, powering the hardware will result in an inrush current higher than the nominal current. After the inrush current, the current will stabilize at the nominal current.

USB ports have a maximum limit for both the inrush current peak and the nominal current. When either of them is exceeded, the USB port will be switched off. As a result, the connection to the Handyscope HS5 will be lost.

Most USB ports can supply enough current for the Handyscope HS5 to work with- out an external power supply, but this is not always the case. Some (battery oper- ated) portable computers or (bus powered) USB hubs do not supply enough cur- rent. The exact value at which the power is switched off, varies per USB controller, so it is possible that the Handyscope HS5 functions properly on one computer, but does not on another.

The Handyscope HS5 comes with a universal power supply, that can be connected to a power outlet using the appropriate adapter. The 3.5 mm connector attached to the power supply must be plugged into the power connector at the rear of the Handyscope HS5. Refer to paragraph 8.1 for specifications of the external power intput.

When the Arbitrary Waveform Generator is used, the power that the Handyscope HS5 requires may strongly increase. It is recommended to use the external power supply when the Handyscope HS5 Arbitrary Waveform Generator is used.

Connect the instrument to the computer
After the new driver has been pre-installed (see chapter 4), the Handyscope HS5 can be connected to the computer. When the Handyscope HS5 is connected to a USB port of the computer, Windows will detect new hardware. Depending on the Windows version, a notification can be shown that new hard- ware is found and that drivers will be installed. Once ready, Windows will report that the driver is installed. When the driver is installed, the measurement software can be installed and the Handyscope HS5 can be used.

Plug into a different USB port
When the Handyscope HS5 is plugged into a different USB port, some Windows versions will treat the Handyscope HS5 as different hardware and will install the drivers again for that port. This is controlled by Microsoft Windows and is not caused by TiePie engineering.

Operating conditions
The Handyscope HS5 is ready for use as soon as the software is started. However, to achieve rated accuracy, allow the instrument to settle for 20 minutes. If the in- strument has been subjected to extreme temperatures, allow additional time for internal temperatures to stabilize. Because of temperature compensated calibra- tion, the Handyscope HS5 will settle within specified accuracy regardless of the surrounding temperature.

Combining and synchronizing instruments
When more channels are required than one instrument can offer, multiple instruments can be combined into a larger combined instrument. To combine two or more instruments, the instruments need to be connected to each other using special cables.

The CMI (Combine Multiple Instruments) interface that is available by default on the Handyscope HS5 provides an easy way to couple multiple oscilloscopes to one combined oscilloscope.

TiePie-engineering-Handyscope-HS5-USB-Oscilloscope-fig-
\(11\)

The CMI interface supports automatic recognition of the instrument. The high speed trigger bus is automatically terminated with the correct impedance and the high speed sampling bus is automatically configured and terminated at the beginning and end of the bus. The high speed sampling bus takes care that each Handyscope is fully synchronized to ensure that even at the highest sampling rate the instruments operate at the same sample clock (0 ppm clock error!). The connection order when combining multiple instruments is not important. The CMI interfacehas built-in intelligence to detect the connections and terminate all buses properly at both ends of the bus. So instruments can be connected to each other in random order. Placing terminators and determining the proper connection order is not required!

Advantages of the CMI (Combine Multiple Instruments) interface are:

  • automatic instrument recognition,
  • automatic creation and termination of the high speed trigger bus,
  • automatic creation and termination of the high speed sampling bus,
  • automatic master/slave setting of the sampling clock bus.

Connecting is done by daisy chaining the CMI connectors of the instruments prior to starting the software, using special coupling cables (order number TP-C50H). The software will detect how the instruments are connected to each other and will automatically terminate the connection bus. The software will combine the connected instruments to one large instrument. The combined instruments will sample using the same clock, with a deviation of 0 ppm.

A six channel instrument is easily made by connecting three Handyscope HS5s to each other.

When combining one or more Handyscope HS5s with other instruments including Handyscope HS6 DIFFs and/or WiFiScope WS6s and/or WiFiScope WS6 DIFFs and/or Automotive Test Scope ATS610004D-XMSGs and/or Automotive Test Scope ATS605004D-XMSs and/or Automotive Test Scope ATS610004DW-XMSGs and/or Automotive Test Scope ATS605004DW-XMS s, the daisy chained CMI bus mustbegin or end with a Handyscope HS6, Handyscope HS6 DIFF, WiFiScope WS6,WiFiScope WS6 DIFF, Automotive Test Scope ATS610004D-XMSG, Automotive Test Scope ATS605004D- XMS, Automotive Test Scope ATS610004DW-XMSG or AutomotiveTest Scope ATS605004DW-XMS. Additionally, the maximum sampling rate is limited to 100 MSa/s at 14 bit resolution.

Front panel

CH1 and CH2 input connectors
The CH1 and CH2 BNC connectors are the main inputs of the acquisition system. The outside of the BNC connectors is connected to the ground of the Handyscope HS5. Connecting the outside of the BNC connector to a potential other than ground will result in a short circuit that may damage the device under test, the Handyscope HS5 and the computer.

AWG output connector
The AWG BNC connector is the output of the internal Arbitrary Waveform Generator. The outside of this BNC connector is connected to the ground of the Handyscope HS5. When the generator is switched off in the software, the generator output is switched to a high impedance, floating state, the output voltage is then undefined. When the generator is switched on in the software and set to pause, the generator output is switched to a low impedance (50 Ω), the output voltage is at the selected offset level.

Power indicator
A power indicator is situated at the top cover of the instrument. It is lit when the Handyscope HS5 is powered.

Rear panel

Power
The Handyscope HS5 is powered through the USB. If the USB cannot supply enough power, it is possible to power the instrument externally. The Handyscope HS5 has two external power inputs located at the rear of the instrument: the dedicated power connector and a pin of the 9 pin D-sub extension connector. The specifications of the dedicated power connector are:

To power the instrument through the extension connector, the power has to be applied to pin 7 of the extension connector. Pin 6 can be used as ground. The following minimum and maximum voltages apply to the power inputs:

Minimum 4.5 VDC / 2 A max.
Maximum 15 VDC / 1 A max.

Note that the externally applied voltage should be higher than the USB voltage to relieve the USB port.

Power adapter
The Handyscope HS5 comes with an external power adapter that can be connected to any mains power net that supplies 100 – 240 VAC, 50 – 60 Hz. The external power adapter can be connected to the dedicated power connector.

USB power cable
A special USB external power cable is supplied with the Handyscope HS5 that can be used instead of a power adapter. One end of this cable can be connected to a second USB port on the computer, the other end can be plugged in the dedicated power connector at the rear of the instrument. The power for the instrument will then be taken from two USB ports.

USB
The Handyscope HS5 is equipped with a USB 2.0 High speed (480 Mbit/s) interface with a fixed cable with type A plug. It will also work on a computer with a USB 1.1 interface, but will then operate at 12 Mbit/s.

Extension Connector

A 9 pin female D-sub connector is available at the back of the Handyscope HS5 containing the following signals:

Pin Description
1 EXT 1 (LVTTL)
2 EXT 2 (LVTTL)
3 EXT 3 (LVTTL)
4 I2 C SDA
--- ---
5 I2 C SCL
6 GND
7 Power IN
--- ---
8 Power OUT
9 reserved

Pins EXT 1, EXT 2 and EXT 3 have internal 1 kOhm pull-up resistors to 2.5 V. These pins can simultaneously be used as inputs and outputs. Each pin can be configured as external digital trigger input for the acquisition system and/or the generator of the Handyscope HS5. Also, each pin can be configured to output one of the following function generator outputs:

  • Generator start
  • Generator stop
  • Generator new period

The I2C pins have internal 2.2 kOhm pull-up resistors connected to 3 V. Pin 8, Power OUT, has the same potential as the Handyscope HS5 power supply. When USB powered, it is at USB power level. When externally powered, it is at the same level as the external power input.

AUX I/O
The Handyscope HS5 has two Auxiliary I/O connectors at the rear of the instrument, connected to the CMI bus. These are used to combine multiple instruments to a single combined instrument to perform synchronized measurements. They can also be used to provide or get an external sampling clock.

Pin Description
1 GND
2 EXT CLK IN/OUT P (LVDS)
3 EXT CLK IN/OUT N (LVDS)

When using the the LVDS external clock (pins 2 and 3) as clock input, the signal must be 10 MHz, 1%.

The Auxiliary I/O connectors use HDMI type C sockets, but are not HDMI compliant. They can not be used to connect an HDMI device to the Handyscope HS5.

Specifications

To achieve rated accuracy, allow the instrument to settle for 20 minutes. When subjected to extreme temperatures, allow extra time for internal temperatures to stabilize. Because of temperature compensated calibration, the Handyscope HS5 will settle within specified accuracy regardless of the surrounding temperature.

Acquisition system

TiePie-engineering-Handyscope-HS5-USB-Oscilloscope-fig-
22

Trigger system

TiePie-engineering-Handyscope-HS5-USB-Oscilloscope-fig-
24

Arbitrary Waveform Generator

TiePie-engineering-Handyscope-HS5-USB-Oscilloscope-fig-
25TiePie-engineering-Handyscope-HS5
-USB-Oscilloscope-fig- 26

Power

Multi-instrument synchronization

Physical

Interface

I/O connectors

System requirements

Environmental conditions

Certifications and Compliances

Probes

Package contents

Declaration of conformity
TiePie engineering Koperslagersstraat 37 8601 WL Sneek The Netherlands

EC Declaration of conformity

We declare, on our own responsibility, that the product

  • Handyscope HS5-540(XM/S/XMS)
  • Handyscope HS5-530(XM/S/XMS)
  • Handyscope HS5-220(XM/S/XMS)
  • Handyscope HS5-110(XM/S/XMS)
  • Handyscope HS5-055(XM/S/XMS)

for which this declaration is valid, is in compliance with EC directive 2011/65/EU (the RoHS directive) including up to amendment 2021/1980, EC regulation 1907/2006 (REACH) including up to amendment 2021/2045, and with

  • EN 55011:2016/A1:2017
  • IEC 61000-6-1:2019 EN
  • EN 55022:2011/C1:2011
  • IEC 61000-6-3:2007/A1:2011/C11:2012
  • EN according the conditions of the EMC standard 2004/108/EC,

also with

  • Canada: ICES-001:2004
  • Australia/New Zealand: AS/NZS
  • CISPR 11:2011 and IEC 61010-1:2010/A1:2019
  • USA: UL 61010-1, Edition 3 and is categorized as 30 Vrms, 42 Vpk, 60 Vdc Sneek, 1-9-2022 ir. A.P.W.M. Poelsma

If you have any suggestions and/or remarks regarding this manual, please contact: TiePie engineering Koperslagersstraat 37 8601 WL SNEEK The Netherlands

Tel.: +31 515 415 416
Fax: +31 515 418 819
E-mail: support@tiepie.nl
Site: www.tiepie.com

TiePie engineering Handyscope HS5 instrument manual revision 2.44, October 2023