Spoilage Yeast Detection 1 LyoKit
Ready Reference Guide
Revision A, November 2023

Product No. KIT230121 (LP), KIT230122 (RP)
PCR kit for the qualitative or quantitative detection of Dekkera/Brettanomyces spp., Zygosaccharomyces spp. and Saccharomyces spp. using real-time PCR instruments.
Before starting, it is strongly recommended to read the entire product manual available on our website.

PROGRAM SETUP

Program your real-time PCR instrument before setting up the PCR reactions. Select the following channels:

  • FAM (Dekkera/Brettanomyces spp.), HEX (Zygosaccharomyces spp.), ROX (Saccharomyces spp.) and Cy5 (Internal Control).

hygiena KIT230121 LP Spoilage Yeast Detection 1 LyoKit - Internal
Control

Pre-incubation: 1 cycle
Step 1: 37 °C for 4 min
Step 2: 95 °C for 5 min
Amplifi cation: 50 cycles
Step 1 : 95 °C for 5 sec
Step 2*: 60 °C for 60 sec

  • Fluorescence detection

For some real-time PCR instruments the probe quencher as well as the use of a passive reference dye must be specified. This kit contains probes with a non- fluorescent “dark” quencher and no passive reference dye.
A Color Compensation is necessary for users of the LightCycler® 480 System: Color Compensation Set 5 (Product No. KIT230011).

DATA INTERPRETATION

Verify results of positive (Control Template) and negative (H₂O) controls, before interpreting the sample results.
Always compare samples to positive and negative controls. Review data from each channel and interpret results as described in the table.

FAM + HEX ROX Cy5 Result Interpretation
+ + or – + or – + or – Positive for Dekkera/Brettanomyces spp.
+ or – + + or – + or – Positive for Zygosaccharomyces spp.
+ or – + or – + + or – Positive for Saccharomyces spp.
+ Negative for targeted spoilage yeasts
Invalid

foodproof® is a registered trademark of Hygiena. The above mentioned real-time PCR instruments are registered trademarks of their respective holders.

PREPARATION OF THE PCR MIX

Take appropriate precautions to prevent contamination, e.g., by using fi lter tips and wearing gloves.

  1. PLACE STRIPS IN RACK
    Take needed number of PCR tube strips out of aluminum bag.
    Important: close bag tightly afterwards. Place strips in a suitable PCR tube rack.
    If needed, gently tap the tubes to move the lyophilized pellets to the bottom of all tubes.

  2. DECAP
    Immediately before fi lling, carefully open strips and discard caps.
    Do not leave open longer than necessary.

  3. ADD SAMPLES AND CONTROLS
    Pipette 25 µL of samples, Negative Control (colorless cap) or Control Template (purple cap) into respective wells.
    If using less volume, add PCR-grade H₂O to reach 25 µL.

  4. SEAL
    Carefully seal the tubes with the provided 8-cap strips.

  5. MIX
    Resuspend pellet after sealing by mixing thoroughly.
    Alternatively, resuspend pellet by pipetting up and down multiple times in Step 3.

  6. CENTRIFUGE
    Briefl y spin strips, e.g., 5 seconds at 500 – 1,000 x g, in a suitable centrifuge.

  7. START REAL-TIME PCR RUN
    Cycle samples as described above.
    Place tubes in a vertical, balanced order into the cycler, e.g., two strips can be placed in the fi rst and last column.

foodproof® Spoilage
Yeast Detection 1 LyoKit| KIT230121 / 22
Kit for 48 reactions
Store kit at 2 to 8 °C| For food testing purposes
FOR IN VITRO USE ONLY
Made in Germany
---|---|---

hygiena LOGO Hygiena®
Camarillo, CA 93012 USA
[email protected]
www.hygiena.com

Documents / Resources

| hygiena KIT230121 LP Spoilage Yeast Detection 1 LyoKit [pdf] User Guide
KIT230121 LP, KIT230121 LP Spoilage Yeast Detection 1 LyoKit, Spoilage Yeast Detection 1 LyoKit, Yeast Detection 1 LyoKit, Detection 1 LyoKit, 1 LyoKit
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Hygiena User Manuals

Related Manuals