Best price-performance ratio Mass Flow Controller, Flow Meter SENSIRION
SFC6000D Mass Flow Controller Mass Flow Meter-

Highlights

  • Disruptive design with integrated electronics
  • High accuracy/repeatability (2% / 0.2% set point)
  • Wide control range (500:1)
  • RS485 and I2 C digital interface
  • No drift and no re-calibration required
  • Ultra-fast settling time (100 ms)

The heart of SFC6000D and SFM6000D is the unsurpassed Sensirion CMOSens® technology. It combines a high precision sensor element with state-of-the-art signal processing on a single chip and thereby providing an accurately calibrated and temperature  compensated signal. In fact, the full functionality of the device is integrated directly on the dedicated ASIC chip, which drastically limits the number of electrical components used. Thanks to CMOSens® sensor technology, Sensirion’s flow meters and controllers  achieve unmatched ratings for speed, accuracy and repeatability at very attractive system cost. Due to the excellent long-term stability of CMOSens® chips, no recalibration is required.

Device Overview (Page 14: Full product list)

Product Variant Description
SFC6000D- 5 slm High performance, low cost

mass flow controller with
multi-gas calibration.
20 slm
50 slm
SFM6000D-| 5 slm| High performance, low cost
mass flow meter with
multi-gas calibration.
20 slm
50 slm

High performance and low cost of SFC6000D and SFM6000D make them the best choice for a wide range of applications, such as gas flushing, FOUP, analytical instrumentation, coating and medical equipment, process automation and gas mixing to name a  few. CMOSens® SFC6000D and SFM6000D measure gas mass flow by the calorimetric principle based on heat transfer. A heater element on a thermally insulated membrane and two thermopiles up-stream and downstream are integrated on a single silicon  chip. In the presence of gas flow, the temperature distribution up- and downstream is disturbed. This asymmetry is then measured. The measurement area as well as the A/D-converter and further signal processing are based on a single chip using CMOS  standard processes (see Figure 1). Moreover, the same chip also takes over the function of a microprocessor in a standard mass flow controller, directly regulating the valve. This drastically reduces the complexity and the number of electronic components in  SFC6000D.
MEMS-based CMOSens® technology enables a significantly larger cross section of the gas channel than bypass capillary diameter in conventional mass flow controllers. This channel design makes the device more robust against particles, clogging and  humidified gases.
The minimal thermal mass of the membrane results in an ultra-fast sensor response time of 3-4 ms. Since the whole design of the amplification, A/D conversion, digital linearization and temperature compensation is matched to the sensor speed, a fully  compensated flow measurement value can be delivered every millisecond.
Combined with advanced control algorithms, SFC6000D offers greatly reduced settling times compared to conventional mass flow controllers.
Furthermore, a special arrangement of the two temperature sensors, on-chip temperature compensation and the minimization of noise sources lead to unbeatable performance with regards to repeatability and accuracy over a large dynamic range. Thanks to the  unique CMOSens® technology, the SFC6000D mass flow controllers and SFM6000D mass flow meters show zero-drift performance and control true mass flow independently of the ambient temperature and pressure changes.

SENSIRION SFC6000D Mass Flow Controller Mass Flow Meter- gas
channel

Scan me to provide feedback

https://feedback.echonovum.com/UXJ7Zk

Quick Start Guide

SFC6000D and SFM6000D can be evaluated easily using the EK-F5x evaluation kit. EK-F5x is compatible with all Sensirion mass flow meters and controllers of 5000 and 6000 families. A quick start guide is included in EK-F5x and can also be found on  Sensirion website for the flow meter and mass flow controller. SENSIRION SFC6000D Mass Flow Controller Mass Flow Meter- Sensirion flow
meters

Figure 2. EK-F5x is the fastest way to evaluate Sensirion flow meters and controllers of 5000 and 6000 families.

Sensor Specifications

Specification Air & N2* 2 CO2* N20 Ar
Calibrated gas number 1 (default) 0 2 3 4
Full scale flow [slm] SFX6000D-Sslm 5 5 2 2
SFX6000D-20slm 20 20 10 10 10
SFX6000D-SOslm 50 50 20 20 20

Accuracy’
whichever is greater| [% set point]| 2.0| 3.| 2.0| 3.0| 3.0
[% full scale]| 0.20| 0.25| 0.20| 0.30| 0.30
Repeatability whichever is greater| [% set point]| 0.20|
[% full scale]| 0.02|
Specification| Flow range| Unit
5 slm| I 20 slm| I 50 slm
Typical settling time2 (SFC only)| 100| ms
Measurement frequency| 1000| Hz
Control / measurement range (SFC / SFM)| 0.2% – 100% full scale (500:1)|
Operating temp. (ambient & gas)| 5 — 50 / 41 — 122| °C / °F
Warming up time 3| 1 (during the first 5 minutes the accuracy may
be below specifications)| s
Maximum inlet overpressure| 10 / 145| bar/ sig
Maximum differential pressure’| SFC6000D| 7 / 102| 5/5/80| 4.0 / 58| bar/sig
SFM6000D| non applicable
Pressure drop at full flow N2 / Air| SFC6000D| 2.0 / 29| 2.0 / 29| 3/5/51| bar/sig
SFM6000D| 0.01 / 0.15| 0.05 / 0.75| 0.15 / 2.2
Supplied fittings| Push-in: tube outer diameter 6 mm (mounted)
Downmount (exchangeable with Festo QSP 18 mm interlace)|
Interlace| Digital: RS 485 and 12C|
Flow units| slm|

Table 1. Overview of CMOSens® SFC6000D Mass Flow Controller and SFM6000D Mass Flow Meter
Specifications. X stands for C=controller, or M=Meter. All data, unless otherwise noted, apply for the following calibration conditions:
Temperature 20°C, Air and horizontal mounting position. SFC is calibrated at 3.0 bar overpressure (inlet: 4.0 bar absolute) against atmosphere (outlet: 1.0 bar absolute). SFM – at atmospheric pressure.
*Air/N2 and CO2 are used for physical calibration. The other gasses are modelled. The accuracy stated for the modelled gasses is “typical” and cannot be guaranteed.

  1. Including offset, non-linearity and hysteresis. Measured against NIST traceable reference.
  2. Step answer from 10% to 100% of full scale within 5% of set point error band.
  3. to within ± 2 % of setpoint
  4. Pressure between flow inlet and flow outlet. For availability of higher differential pressure option, please contact Sensirion

2.1 Gas calibration
SFC6000D and SFM6000D are factory-calibrated for multiple gasses. Table 1 lists the available calibrations and the calibration gas numbers saved in internal memory of the device. The desired calibration can be selected by the user. Please see the application  notes describing the communication interfaces or quick start guide for instructions on how to activate a desired calibration. These can be found on http://www.sensirion.com/sfc6000.
Please note that the maximum flow rate achievable with each MFC is strongly dependent on the gas measured.
Table 1 lists the maximum flow rates for all gasses for which the products have been calibrated.
The calibration for Air and N2 is identical. It can be used with both gasses with the same result.

2.2 Accuracy
Accuracy describes how precisely the mass flow controller / meter is able to control / measure flow rate with respect to the absolute flow of a given gas. Accuracy is mostly determined by the quality of the calibration and can be different for each calibration gas.
For example: with mass flow controller set point of 10 slm and a real flow rate measured by an external reference of 10.2 slm, the set point accuracy would be calculated as:

Accuracy 10.2 − 10 = 2.0% set point

10.2

2.3 Repeatability
Unlike accuracy, repeatability is not influenced by calibration quality and is directly related to the build quality of the mass flow controller / meter. It describes how reliably mass flow controller is able to reach a given setpoint (applied repeatably). For mass flow  meter, it describes, how reliably the instrument measures the same flow repeatably.
For example: if a mass flow meter measures the absolute flow of 10 slm multiple times and the resulting flow measurement results follow a Gaussian distribution centered around 9.97 slm with 0.02 standard deviation, then the repeatability is calculated as:

Repeatability 0.02 0.2% set point

9.97

Generally, for mass flow controllers/meters repeatability is better than accuracy. For mass flow controllers, in applications where an additional calibration or feedback loop exists, it is possible to relay on repeatability rather than accuracy of mass flow  controller. One example would be an optimized process, where the setpoint value of mass flow controller is fine- tuned to give the desired outcome. In such case, it is not important that the absolute flow rate is close to the set point – instead, it is important that the optimized set point can be achieved repeatably to give the same process outcome.
Accuracy and repeatability of SFC6000D and SFM600D devices are dominated by set point error at high flows and by the full-scale error at low flows. Figure 3 demonstrates this.

SENSIRION SFC6000D Mass Flow Controller Mass Flow Meter- set
point

2.4 Settling time
The CMOSens® SFC6000D mass flow controller has an ultra-fast settling time. Figure 4 shows the typical response time of the SFC6000D in comparison to a mass flow controller using conventional capillary technology. SENSIRION
SFC6000D Mass Flow Controller Mass Flow Meter- mass flow
controller

2.5 Wide control and measurement range
The wide control range of the SFC6000D brings a decisive benefit in applications with a wide dynamic range of gas flows. Instead of two devices used for high flow and low flow ranges, a single SFC6000D device can efficiently cover a flow range of three orders of magnitude.
Control and measurement range is defined as 0.2% – 100% full scale. This means that a mass flow controller with e.g. 50 slm full scale flow, can control flows lower than 0.1 slm. When relying on repeatability rather than absolute accuracy, even lower flows can  be controlled.
2.6 Pressure drop
Mass flow controllers need pressurized gas sourced to operate. Pressure drop is generated, when gas passes through a mass flow controller. When evaluating a mass flow controller, it is important to verify that at maximum required flow rate, for a given gas, the  pressure drop will be smaller than the inlet pressure – otherwise the desired maximum flow rate will not be possible to achieve.
Mass flow meters have generally much lower pressure drop due to the absence of the valve and hence this is rarely a practical problem.
Table 1 lists pressure drop at full scale flow of Air or N2. At the same setpoint, for gasses heavier than air, the pressure drop would generally be higher. For gasses lighter than air – it would be lower. Pressure drop scales approximately linearly with the density of  the gas.

Electrical specifications

Parameter Min Typ Max Unit
Supply voltage, VDD 23. 24.0 25. VDC
Max current supply SFC6000D 210 mA
Max current supply SFM6000D 20 mA
Max standby current < 15 mA
Electrical Connector M8 (4 pin)

Table 2. Electrical specifications

Sensor Operation

SFC6000D and SFM6000D are configured with RS485 and I2 C interfaces. Please see the Technical download (sensirion.com/sfc6000) section for instruction on how to use these.
For OEM projects with MOQ of 100 pcs / yr Analog Voltage interface is available on sister platform SFC6000 and SFM6000. Profibus, Modbus and other protocols are available on request.
4.1 RS485 interface
The pinout of the M8 connector available on SFC6000D and SFM6000D is shown below.SENSIRION SFC6000D Mass Flow Controller Mass Flow Meter-
interface

The maximum data readout rate available with RS485 interface is 200 Hz. It is recommended that the cable length between the master and the sensor does not exceed 3m.
EK-F5x evaluation kit is designed to work with RS485 interface. When using the evaluation kit with Windows PC, the highest achievable data readout rate is 20 Hz.
4.2 I2C interface
SFC6000D and SFM6000D can be operated by standard I2C interface. To access I2C interface, please remove the green cap from the sensor and unplug the 6-pin cable inside the sensor. Please make sure that the sensor is not powered, when doing so.  Unplugging the internal 6-pin cable, while the device is powered will damage it.
An example of a matching cable assembly is Molex Micro-Lock Plus 0451110606. The sensor is equipped with a Molex Micro-Lock Plus connector with 6 circuits and a 1.25 mm pitch (UPC: 889056511957), which is shown in Figure 6. SENSIRION SFC6000D Mass Flow Controller Mass Flow Meter- interface
2

Pin # Name Description
1 VDD Supply Voltage
2 GND Ground
3 NC Not Connected
4 SCL Serial Clock
5 ADDR I2 C Address Selection
6 SDA Serial Data

Figure 6. Pinout of 6-pin connector on the sensor PCB used for I2C communication.
Using I2 C interface offers a benefit of faster data readout rates – up to 1 kHz, as compared to 200 Hz with RS485.

4.3 InitStep and Controller gain
The regulation parameters (InitStep and Controller gain) allow fine tuning of the PID controller. They let the user achieve a desired compromise between speed, overshoot, and control stability. The configured gain and initstep are reset upon unplugging the  device.

4.3.1 InitStep
InitStep refers to a normalized initial valve voltage. This voltage value will be added to the valve control value if the regulator is in the normal regulation mode and a setpoint value unequal zero is specified. This normalized InitStep value reflects the fact that for  instance a 24 V valve starts to open only if the valve voltage is higher than several volts. The customer has the possibility to change this value. Setting a higher InitStep will result in a faster controller but may lead to overshoots. A lower value will result in a  smoother opening behavior but may lead to a slower response. The ideal InitStep depends on input pressure and temperature. Please note, InitStep value is stored in volatile memory and will be reset to the factory value, after the device is unplugged or reset.

4.3.2 Controller gain
The Controller gain refers to a normalized gain parameter, which is set to 1 by default. This gain factor will be multiplied to the control deviation, which is the difference between desired setpoint and measured value, and a correction of the valve control voltage  will be applied depending on this control deviation value. If the customer gain factor is increased, this results in a larger change in the output for a given change in the control deviation.
A too low Controller gain value results in a smaller output response to a large input control deviation and therefore to a less responsive / slower controller. On the other hand, if the Controller gain is too high, the system can become unstable: show oscillations  and overshoots. As the valve characteristics depend for instance on pressure, the control action may be too strong when responding to system disturbances at high pressures. Therefore, a smaller gain factor may be advisable at higher pressures and the customer  has the possibility to tune this parameter according to his conditions and requirements. Please note, that it is also possible to set Controller gain values smaller than 1.

Physical Specification

5.1 Fittings
SFC6000D and SFM6000D are factory-fitted with 6 mm push-in fittings (QSPK18-6). These fittings can be safely removed to reveal Festo QSP-type mounting slot. Any compatible QSP fittings can be installed without impacting the performance of the device.
Please note that the supplied QSP push-in fittings are not compatible with O2-rich gasses due to the grease used. To use your device with O2-rich gasses, please replace factory-installed fittings on SFC6000D and SFM6000D with the downmount flange  provided. When using third-party fittings, please make sure the substitute is compatible with the gas you want to use.
For an example list of compatible fittings and the instructions on how to replace them, please see the application note at Technical download (sensirion.com/sfc6000). Other custom-made fittings are available from Sensirion for OEM projects.

5.2 Wetted materials & compatibility
Table 3 gives an overview of the materials wetted by the gas. For high volume OEM applications different specialized materials for the body, valve and sealing can be used with sister products, SFC6000 and SFM6000.

Parameter|
---|---
Wetted materials| Flow body and
flow sensor| PAl2
Si, Si3N4, SiOx
Epoxy
Glass-filled epoxy Stainless steel
FKM
Valve| Stainless steel FKM
PTFE
Push-in fittings| PA
NBR
Grease (not 02-compatible)
Downmount fittings| PAl2 FKM
Flow channel is free of silicone and copper
REACH, RoHS| REACH and RoHS compliant

Table 3. Overview of wetted materials
5.3 Safety instructions
5.3.1 Oxygen-rich gases
Most push-in fittings are not compatible with O2-rich gasses (more than 21% O2) due to the grease used. This is also the case with the factory-provided QSP push-in fittings on SFC6000D and SFM6000D. At elevated pressures and temperatures, there is a risk  of self-ignition of the grease. To use your device with O2-rich gasses, please use the downmount flanges included or install other O2-compatible fittings.
5.3.2 Toxic gases
Due to relatively high leakage rate of the push-in connectors, the sensors are not suitable for operation with toxic gasses. If the connectors are replaced, the whole gas assembly must be checked for leakage before applying toxic gas to the device.
5.3.3 Aggressive or corrosive gases
Please make sure that the gases used are compatible with the wetted materials listed in this chapter. In case of doubt, please contact Sensirion for further advice. Corrosive gasses can damage the sealing or the CMOSens® chip.
5.3.4 Flammable gases
SFC6000D and SFM6000D are not designed to be operated with flammable gasses. Please check SFC5xxx and SFM5xxx series to use such gasses.
5.4 Physical dimensions and mounting information
Physical dimensions and mounting information are shown below. The flexible cable between the valve and the green housing can add around 4-6 mm to the total height of SFC6000D.
It is recommended to mechanically fix the sensor during measurement. This will avoid movement of the device and flow disturbances.
The sensor can be mounted from below with PT screws: size K30, max. depth inside housing: 6mm. Alternatively, it can be mounted from the side, using the through-holes. Finally, SFC6000D has two M2.5 mounting slots below the valve – these can be used to  fix the sensor from above but require a (temporary) removal of the green cap. Downmount fittings available with SFC6000D and SFM6000D offer additional mounting holes.SENSIRION SFC6000D Mass Flow
Controller Mass Flow Meter- dimensions

Figure 7. Physical dimensions and mounting information for SFC6000D with push- in fittings. All units are in [mm]. SENSIRION SFC6000D Mass Flow Controller
Mass Flow Meter- dimensions 2

Figure 8. Physical dimensions and mounting information for SFM6000D with push- in fittings. All units are in [mm]. SENSIRION SFC6000D Mass Flow Controller
Mass Flow Meter- dimensions 3

Figure 9. Physical dimensions and mounting information for SFC6000D with downmount fittings. All units are in [mm]. SENSIRION SFC6000D Mass Flow
Controller Mass Flow Meter- dimensions 4

Figure 10. Physical dimensions and mounting information for SFM6000D with downmount fittings. All units are in [mm].

Flow units

SFC6000D and SFM6000D are calibrated in “standard units” – namely slm (standard liter per minute). Table 4 lists other common units used for measuring gas flow and a relationship between them. If your device appears to show c.a. 7% error vs your  reference, it is possible the reference is displaying flow in “norm” units.

Flow unit Reference condition
Gas Temperature Gas Pressure
slm (standard liter per minute) 20 °C / 68° F
In/min (norm liter per minute) 0 °C / 32° F

Table 4. Common units used to express gas flow rates.

OEM options

SFC6000D and SFM6000D mass flow controllers/meters are all special versions of Sensirion’s SFC6000 and SFM6000 platform. SFC6000D and SFM6000D were designed as distribution products to meet a wide range of needs. In case SFC6000D or SFM6000D does not meet exactly the requirements for a given application, Sensirion recommends looking at SFC6000 and SFM6000 mass flow controllers / meters, which are available with a variety of configurations (fittings, flow ranges, calibrations,  interfaces) and are generally built on order.
These devices are available directly from Sensirion with a minimum order quantity (MOQ) of 100 pcs. SFC6000 and SFM6000 also allow for OEM product development in the context of high-volume projects. Possibilities involve different communication  interfaces, wetted materials and gas calibrations.

Ordering Information

Flow unit Reference condition
Gas Temperature Gas Pressure
slm (standard liter per minute) 20 °C / 68° F
In/min (norm liter per minute) 0 °C / 32° F

Table 5. Products in SFC6000D and SFM6000D series and a compatible evaluation kit

Revision History

Date Version Pages Changes
Mar-23 1.0 all Initial version
May-23 1. 6 Corrected Figure 3

Important Notices

Warning, Personal Injury
Do not use this product as safety or emergency stop devices or in any other application where failure of the product could result in personal injury. Do not use this product for applications other than its intended and authorized use. Before installing, handling,  using or servicing this product, please consult the data sheet and application notes. Failure to comply with these instructions could result in death or serious injury.
If the Buyer shall purchase or use SENSIRION products for any unintended or unauthorized application, Buyer shall defend, indemnify and hold harmless SENSIRION and its officers, employees, subsidiaries, affiliates and distributors against all claims, costs, damages and expenses, and reasonable attorney fees arising out of, directly or indirectly, any claim of personal injury or death associated with such unintended or unauthorized use, even if SENSIRION shall be allegedly negligent with respect to the design or the manufacture of the product.
ESD Precautions
The inherent design of this component causes it to be sensitive to electrostatic discharge (ESD). To prevent ESD-induced damage and/or degradation, take customary and statutory ESD precautions when handling this product. See application note “ESD,  Latchup and EMC” for more information.
Warranty
SENSIRION warrants solely to the original purchaser of this product for a period of 12 months (one year) from the date of delivery that this product shall be of the quality, material and workmanship defined in SENSIRION’s published specifications of the product.
Within such period, if proven to be defective, SENSIRION shall repair and/or replace this product, in SENSIRION’s discretion, free of charge to the Buyer, provided that:

  • notice in writing describing the defects shall be given to SENSIRION within fourteen (14) days after their appearance;
  • such defects shall be found, to SENSIRION’s reasonable satisfaction, to have arisen from SENSIRION’s faulty design, material, or workmanship;
  • the defective product shall be returned to SENSIRION’s factory at the Buyer’s expense; and the warranty period for any repaired or replaced product shall be limited to the unexpired portion of the original period.
  • This warranty does not apply to any equipment which has not been installed and used within the specifications recommended by SENSIRION for the intended and proper use of the equipment. EXCEPT FOR THE WARRANTIES EXPRESSLY SET FORTH  HEREIN, SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH RESPECT TO THE PRODUCT. ANY AND ALL WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES OF MERCHANTABILITY OR FITNESS  FOR A PARTICULAR PURPOSE, ARE EXPRESSLY EXCLUDED AND DECLINED.
    SENSIRION is only liable for defects of this product arising under the conditions of operation provided for in the data sheet and proper use of the goods. SENSIRION explicitly disclaims all warranties, express or implied, for any period during which the  goods are operated or stored not in accordance with the technical specifications.
    SENSIRION does not assume any liability arising out of any application or use of any product or circuit and specifically disclaims any and all liability, including without limitation consequential or incidental damages. All operating parameters, including  without limitation recommended parameters, must be validated for each customer’s applications by customer’s technical experts.
    Recommended parameters can and do vary in different applications.
    SENSIRION reserves the right, without further notice, (i) to change the product specifications and/or the information in this document and (ii) to improve reliability, functions and design of this product.

Headquarters and Subsidiaries

Sensirion AG
Laubisruetistr. 50
CH-8712 Staefa ZH
Switzerland
phone: +41 44 306 40 00
fax: +41 44 306 40 30
info@sensirion.com
www.sensirion.com
Sensirion Taiwan Co. Ltd
phone: +886 2 2218-6779
info@sensirion.com
www.sensirion.com| Sensirion Inc., USA
phone: +1 312 690 5858
info-us@sensirion.com
www.sensirion.com
Sensirion Japan Co. Ltd.
phone: +81 45 270 4506
info-jp@sensirion.com
www.sensirion.com/jp
To find your local representative, please visit www.sensirion.com/distributors| Sensirion Korea Co. Ltd.
phone: +82 31 337 7700~3
info-kr@sensirion.com
www.sensirion.com/kr
Sensirion China Co. Ltd.
phone: +86 755 8252 1501
info-cn@sensirion.com
www.sensirion.com/cn
---|---|---

Copyright© 2022, by SENSIRION. CMOSens® is a trademark of Sensirion. All rights reserved