Sensirion’s relative humidity and temperature sensors SHTxx (where xx serves as placeholder) offer best in class performance in a broad range of applications. However, to make the most of their excellent performance, it is important to take some precautions during storage, assembly, and packaging. We recommend to carefully review these handling instructions, ideally during the design-in phase and before production release. Special attention should be paid to avoiding exposure to volatile organic compounds in high concentrations and/or for long periods of time, particularly during manufacturing and storage. Therefore, proper handling to mitigate any risks and correct material selection is crucial to ensure highest performance. Application of SHTxx in the field in ambient environment is not critical.
Key Instructions
- Protection against ESD is mandatory.
- Do not use polyethylene antistatic bags.
- Do not apply board wash.
- Do not apply spray to unprotected sensor.
- Be careful exposing the sensor to VOC.
- Prevent sensor from exposure to cleaning agents.
- Cover the sensing element during coating.
This document is applicable to all Sensirion SHTxx humidity and temperature sensors.
https://feedback.echonovum.com/OeoAIs
General
1.1 ESD
To ensure proper functioning of the sensor, it must be protected from
Electrostatic Discharge (ESD) at all times. All handling of the sensor should
take place exclusively in Electrostatic Discharge Protected Areas (EPAs) that
have been properly set up to minimize the risk of ESD. This includes grounding
personnel with wrist-straps or similar measures, grounding all conductive
objects and excluding insulating materials from the EPA.
Additionally, all operations should be conducted on a grounded conductive
floor. To further protect the sensor, it should be packaged using ESD
protective materials when not being handled within an EPA. 1.2 Storage
To ensure optimal performance of the sensors, we strongly recommend storing
them in the original sealed ESD bag prior to assembly or use. The ideal
storage conditions for the sensors are as follows:
- Temperature: 10 °C…50 °C (0…125 °C for a limited time)
- Humidity: 20…60 %RH (sensors that are not stored in ESD bags).
By following these guidelines, you can ensure that the sensors are in the best
possible condition for assembly and use. Instruction 2. Store sensors in
original, unopened ESD bag. Place additional stickers only on the outside of
the ESD bag.
To ensure the longevity and accuracy of your sensors, we recommend storing
them in metal-in antistatic shielded ESD bags once they have been removed from
their original packaging. This will help protect the sensors from
electrostatic discharge and other external influences. 1
Additionally, we recommend to not use any adhesive or adhesive tapes to reseal
the sensor bag after opening.
This can help avoid contamination and maintain the integrity of the sensor.
The ESD bags listed in Table 1. ESD Bag compatibilitycan be recommended for
this purpose, as they do not have a polluting effect on the humidity sensor.
Manufacturer | Product | Compatibility |
---|---|---|
Stroebel | Topshield Bags | Compatible |
Table 1. ESD Bag compatibility
When packaging sensors, it is important to ensure that the materials used do
not have the potential to cause sensor pollution. To maintain the accuracy and
longevity of your sensors, only use packaging materials that do not outgas.
Metal-in antistatic shielded ESD bags, paper or cardboards based packaging,
and deep-drawn plastic trays such as PE, PET, and PP may be considered. Do not
use antistatic polyethylene bags, bubble foils, and foams, as they may
contaminate the sensor. Additionally, be cautious when using stickers inside
the packaging, as they can also cause contamination if not used properly.
Sticker size should be kept to a minimum, and the sticky side should fully
adhere to a surface.
It is important to note that many packaging materials may contain additives
(plasticizers) that can have a polluting effect on the sensor. As a general
rule, if a material emits a strong odour, it should not be used. Additionally,
even materials that are recommended for use may contain additives, so it is
important to be aware of this when selecting packaging materials.
For high safety and to ensure that the sensor is not affected, it is
recommended to conduct a qualification test on the final device housing and
shipment packaging2 . Such a test may include exposing the final device with
the sensor in its shipment packaging to a temperature of ≥65 °C for at least
168 hours. If shipping or storage conditions are expected to be harsh, the
qualification test conditions for the packaging material should be customized.
The sensor reading should show no deviation from a reference compared to the
same measurements taken before the exposure. Instruction 3. Do not use
polyethylene antistatic bags (light blue, pink or rose color). Do not use
adhesive tapes inside packaging.
If you need guidance on testing your humidity sensors, we recommend consulting
the “Testing Guide for
Humidity Sensors” [1]. This guide will provide you with simple and
straightforward instructions to help you
¹This recommendation also applies to devices with assembled sensors.
²Especially for harsh shipping conditions.
ensure your sensors are working correctly. Furthermore, in case of doubt,
kindly contact the Sensirion support line.
1.3 Exposure to chemicals
Humidity and temperature sensors are highly accurate environmental sensors
that require special care to ensure optimal performance. The sensors are
susceptible to pollutants and must be protected from exposure to volatile
chemicals, acids, bases, and cleaning agents. Especially Ketenes, Acetone,
Ethanol, Isopropyl Alcohol, Toluene, etc.3 , might cause drift, in some cases
even irreversible. To ensure the sensors maintain their outstanding
performance, it is important to follow these guidelines:
- Avoid exposing the sensor to pollutants, especially volatile chemicals such as solvents or organic compounds. High concentrations and long exposures should be avoided, as they can cause drift in the humidity reading or even irreversibly damage the sensor.
- Be aware that certain chemicals are often found in epoxies, glues, adhesives, and plastics, and may outgas during baking and curing, potentially affecting the sensor.
- Avoid exposure to acids and bases, as they can irreversibly damage the sensor. These include HCl H2SO4 , HNO3 , and NH3 . High concentrations of ozone or H2O2 should also be avoided. • Avoid contact with cleaning agents, such as PCB board wash after soldering, or strong air blasts from an air-pistol 4 , as they can cause drift in the reading or complete breakdown of the sensor.
- Ensure good ventilation to avoid high concentrations of volatile chemicals, such as solvents, cleaning
solutions, and detergents, e.g. ethanol, isopropanol, methanol, acetone, etc..
Please note that this is not a complete list of harmful substances, and it is important to carefully read the application notes and follow all the guidelines provided to ensure optimal performance of the sensor.
Assembly
The sensors can be soldered using standard reflow soldering ovens. They are designed to withstand the soldering process, including the profile according to IPC/JEDEC J-STD-020 with peak temperatures at 260 °C for up to 30 seconds during Pb-free assembly in IR/Convection reflow ovens (see Figure 1).5 Figure
- Soldering profile according to JEDEC standard. TP ≤ 260 °C and tP ≤ 30 sec
above 255 °C for Pbfree assembly. TL < 220 °C and tL < 150 sec. Ramp-up rate
<3 °C and ramp-down rate <6 °C/sec for temperatures >TL .
3 Such chemicals are integral part of epoxies, glues, adhesives, etc. and outgas during baking and curing. These chemicals are also added as plasticisers into plastics, used for packaging materials, and do out-gas for some period.
4 Oil free air does not cause any harm to the sensor.
5Vapor phase soldering may not be recommended for the assembly of this specific product. If no alternative options are available, iadvised that the end-user or a qualified party conducts a compatibility qualification. Please note that the compositions of heat transliquids can vary among manufacturers, and therefore Sensirion cannot provide specific information regarding their use in vapor phassoldering.
When soldering humidity sensors, it is important to ensure that the maximum temperatures and exposure times are respected to prevent damage to the sensors. If the PCB goes through multiple solder cycles, we recommend assembling in the last solder cycle to reduce the risk of sensor pollution6 . The use of “no clean” type ≥3 solder paste7 is strongly recommended as it eliminates the need for a board wash, which can be harmful to the sensors.
It is important to use the appropriate amount of solder paste to achieve a
stand-off height (clearance between the package body and the substrate) of
50-75 μm. Please consult the corresponding sensor datasheet for more
information on the metal land pattern and recommendations on solder paste
printing stencils. Standard pick & place equipment and vacuum nozzles for
standard QFN packages can be used for assembly of SHTxx sensors.
Manual soldering is not recommended, and rework soldering should be limited to
five seconds at up to 350 °C. Additionally please note that after exposure to
high temperatures, such as during reflow soldering, humidity sensors may
temporarily read a negative humidity offset, typically between -1…-2 %RH. This
offset will disappear by itself when the sensor is exposed to ambient
conditions, typically within one to three days. If humidity testing is
performed immediately after reflow soldering, this offset should be considered
when defining the test limits.
It is important to note that the diced edge or side faces of the I/O pads may
oxidize over time, so a solder fillet may or may not form. Therefore, there is
no guarantee for solder joint fillet heights of any kind.
Sensors in SMT packages are classified as Moisture Sensitivity Level 1
(IPC/JEDEC J-STD-020), meaning that it is recommended to process the sensors
within 1 year of the date of delivery to ensure optimal performance.
Before beginning the assembly process, it is important to carefully read the
User’s Guide and Datasheets to ensure proper handling and usage of the
sensors.
It is important to note that no mechanical force should be applied to any part
of the sensor during assembly or usage to prevent damage.
To protect electronic assemblies, including soldered contacts of sensors, in
corrosive environments such as condensed water or corrosive gases, a
passivation should be applied. This can be done through conformal coating or
an adhesive. However, it is important to use high viscous conformal coatings
or potting materials to avoid flowing into the sensor opening and covering the
sensing element, rendering the humidity sensor inoperable. Alternatively use
the Sensirion solution with protective cover. Additionally, the sensor opening
should not be covered with any coating, so that the top surface of the sensor
remains free.
6 Max. 3 solder cycles with sensor.
7 Solder types are related to the solder particle size in the paste: Type 3
covers the size range of 25 – 45 µm as specified in IPC J-STD005A Instruction
5. If conformal coating is applied, the top surface of the sensor must remain
free of coating.
When applying conformal coatings to electronic assemblies, it’s important to
keep in mind that solvent vapors are produced during the curing and processing
of the coating. Proper ventilation should be established throughout the
application, staging, and curing process to prevent pollution of the sensor.
This can be achieved by using ovens with fresh air supply, to ensure that the
concentration of these gaseous substances remains low.
It is important to follow the manufacturer’s guidelines for coating thickness
and curing times, and to apply the upper time limit if provided. This is
especially important if the PCB design features large components with
undercuts, where high layer thicknesses on/under components or drain-off edges
can form. Tack-free state of the coating is not an indication of full curing,
so be sure to refer to the application notes of the manufacturer for full
curing. It is also important to ensure process stability for coating thickness
and apply corresponding safety margins for curing times.
Table 2 lists conformal coatings that have been tested for pollution of the
humidity sensor and are known to be suitable if applied and fully cured under
good ventilation and according to the respective datasheet. Some coatings are
specifically marked as non-compatible. Please note that this recommendation is
valid from January 2023, but the suitability of a coating might change without
prior notification if the manufacturer changes the formulation of the product.
Manufacturer | Product | Compatibility |
---|---|---|
Peters | SL 1307 FLZ 23 | Compatible |
Peters | Elpeguard SL 1301 ECO-BA-FLZ | Not compatible |
Peters | Elpeguard SL 1307 | Not compatible |
Peters | Elpeguard DSL 1706FLZ | Not compatible |
Dow Corning | 1-2620 | Compatible |
Humiseal | 1R32A-2 | Compatible |
Humiseal | 1B51NSLU | Compatible |
Humiseal | 1B12LU | Not Compatible |
Humiseal | 1A27 | Not Compatible |
Humiseal | 1B73EPA | Not Compatible |
Table 2. Conformal coating compatibility.When using spray coating techniques, such as applying
varnish or conformal coating, it is recommended to use the SHT4x sensor with
the protective cover option.
If adhesive tapes need to be used, we only recommend tapes listed in Table 3.
Manufacturer | Product | Compatibility |
---|---|---|
3M | Kapton® Electrical Tape 92 | Compatible |
Table 3. Adhesive Tape compatibility Instruction 7. In conjunction with the
sensor, use only recommended Kapton® tape Electrical Tape 92 by 3M™. Do not
apply any other adhesive tapes such as Scotch Tape, Sello Tape, Tesa Film etc.
Please use the materials listed in the following tables for adhesives and
encapsulants. They have been tested and are recommended for use, as long as
they are applied and fully cured in a well-ventilated environment with fresh
air supply. Be cautious when using materials and brands that are not listed in
the tables, as they may be harmful. Always test them carefully before use.
Please note that this recommendation is valid from January 2023, but the
compatibility might change without prior notification if the manufacturer
changes the formulation of the product.
Manufacturer | Product | Compatibility |
---|---|---|
EPO-TEK | H70E/S | Compatible |
EPO-TEK | T6067 | Compatible |
EPO-TEK | 93-86 | Compatible |
Lord | MD-130 | Compatible |
Eccobond | E3503-1 | Compatible |
Ablebond | 84-3 | Compatible |
Table 4. Epoxy Adhesive compatibility.
Manufacturer | Product | Compatibility |
---|---|---|
Stycast | 50300 / 50400 series | Compatible |
Hysol | EO1061/EO1062 | Compatible |
Table 5. Epoxy Glob Top compatibility.
Manufacturer | Product | Compatibility |
---|---|---|
RTV | 6424 (Momentive) | Compatible |
Dow Corning | 732 | Compatible |
Table 6. Silicone Adhesives compatibility.
Manufacturer | Product | Compatibility |
---|---|---|
Stycast | S 7503 | Compatible |
Table 7. Silicone Glob Top Encapsulants compatibility.
Manufacturer | Product | Compatibility |
---|---|---|
Loctite | 401 | Compatible |
Table 8. Instant Adhesives compatibility.
Manufacturer | Product | Compatibility |
---|---|---|
Peters | VU 4457-61 | Compatible |
Table 9. PUR (Polyutherane) Casting Compounds compatibility.
Hot melts may absorb moisture, consequently influencing the response time of
the sensor found in the device.
Therefore, hot melt should be used judiciously.
Firstly, it is recommended to place the sensor in the device only when all the
materials used in the assembly process have completely cured or dried up, and
in case they have not, to ensure proper ventilation is available in the curing
ovens and assembly lines. Secondly, care should be taken to prevent any dust
or particles from entering the sensor entrance, as this can impact its
performance. It is advisable to wear clean gloves or finger cots while
handling the sensor and avoid touching the sensor entrance.
During repair and rework of assemblies containing sensors, it is recommended
to cover the sensor with Kapton tape, as per the specific recommendation in
Table 3. Additionally, it is important to be cautious when using potent
cleaning agents, such as detergents, alcohols, brominated or fluorinated
solvents, as cleaning any part of the product can result in a high
concentration of cleaning agents on the sensor. Therefore, it is necessary to
remove any sensors or devices containing sensors before cleaning the
production area and tools. Finally, it is important to ensure good ventilation
and that any solvents have evaporated before resuming production to prevent
any damage to the sensors.
Operation in the Field
When operating a capacitive humidity sensor in the field, there are generally
no impairments to its performance if the design is good, see Design-In Guide
[2]. However, wrongful exposure to chemicals and other contaminants can lead
to degradation of the sensor’s accuracy over time. Please refer to section 1.3
to see a list of known chemicals that can have an impact on the sensors
performance if not handled with care. This is because contaminates can induce
a change in the dielectric constant of the sensing material like water
molecules. As the sensor cannot distinguish if the induced change comes from
contaminants or water molecules, they can falsify the measurement. While water
desorbs once the humidity of the environment decreases, some contaminants are
bound too strongly and will not leave the system. This leads to an
irreversible drift in the sensors reading not to be confused with the
reversible creep of sensors used at extreme conditions (high temperature and
relative humidity above 90%) [3].
To combat this induced drift, the newest sensors are equipped with a heater
that help to free the sensor from different types of contaminants. Please
follow the protocol in the document Creep_Mitigation_SHT4x [3] to ensure
continuous drift mitigation. It is important to note that some contaminants
may have a more severe impact on sensor accuracy than others, so it is crucial
to consider the environment in which the sensor will be operating when
selecting the appropriate design.
Extreme Conditions and Reconditioning
Certain applications require the exposure of humidity and temperature sensors
to extreme conditions, which SHT sensors withstand in most cases. However,
certain prophylactic measures must be adopted.
For exposure to extreme humidity and temperature, please refer to the
datasheet of the respective product.
Make sure that the exposure time of the sensor to the maximum operating
conditions is limited as mentioned in the datasheet. Prolonged and high
concentration exposure to volatile organic compounds is not only critical
during assembly but also in the field. Such applications need to be tested and
validated thoroughly. As described above, exposure to acids or bases may also
be critical. For it to be critical, their concentration must be such that the
polymers are affected. As a rule of thumb, bases with a pH below 9 are not
critical. Etching substances such as H2O2 , NH3 , etc. in high concentrations
are critical to the sensor.
Corrosive substances in low concentrations are not detrimental to the sensor
itself. However, they may corrode the soldered contacts. Hence, the contacts
must be well-protected (passivated) in potentially corrosive environments –
see also Section 2.
The application of Sensirion humidity and temperature sensors to severe
conditions must be cautiously tested and validated. Sensirion qualifies its
humidity and temperature sensors to function appropriately within ambient
clean air – the qualification for use in extreme environments is the
responsibility of the user.
As stated above extreme conditions or exposure to solvent vapors may offset
the sensor. The following reconditioning procedure may bring the sensor back
to calibration state:
- Baking: 100 – 105 °C at <5 %RH for 10 h
- Re-Hydration: 20 – 30 °C at ~75 %RH for 12 h
The 75 %RH can conveniently be generated with a saturated NaCl solution. For further information please refer to the cited article on Humidity Fixed Points of Binary Saturated Aqueous Solutions from the NIST Journal of Research [4].
Removable Protective Cover
There are some visual anomalies which can occur when handling the removable
protective cover. When removing the protective cover some adhesive residues
might be left behind on the top surface of the sensor.
It shall be noted that this is a purely optical phenomenon which does not
affect the sensor functionality.
It also might happen that the anti-adhesion layer, highlighted in Figure 2
contracts slightly during processes where high temperatures are reached, such
as reflow soldering. This change in shape occurs only to the antiadhesion
layer while the protective cover
retains its full shape and functionality.
Coating residues might be found on the edges of the sensor packaging after
removal of the protective cover, particularly when depositing thick coating
layers.
Bibliography
-
Sensirion, “Sensors Specification Statement and Testing Guide,” 19 April 2021. [Online]. Available:
https://sensirion.com/products/downloads/. -
Sensirion, “SHTxx Design Guide,” [Online]. Available: www.sensirion.com.
-
Sensirion, “Creep Mitigation SHT4x,” 04 2022. [Online]. Available:
https://sensirion.com/media/documents/A88858C9/629626D4/Application_Note_Creep_Mitigation_SHT4x.pdf. -
Greenspan, “Humidity Fixed Points of Binary Saturated Aqueous Solutions,” J Res Natl Bur Stand A Phys Chem, vol. 81A(1), no. doi: 10.6028/jres.081A.011. Epub 1977 Feb 1. PMCID: PMC5295834., pp. 89-96, 1977.
-
IEEE, “IEEE Reference Guide,” 2018. [Online]. Available: https://ieeeauthorcenter.ieee.org/wpcontent/uploads/IEEE-Reference-Guide.pdf. [Accessed 20 1 2023].
Revision History
Date | Revision | Pages | Changes |
---|---|---|---|
June 2007 | 0.1 | Initial revision (Preliminary) | |
January 2009 | 1.0 | Complete rework. | |
March 2010 | 1.1 | New format, implementation of pictograms. | |
June 2012 | 2.0 | Moisture Sensitivity Level adapted. | |
November 2013 | 3 | Document partially restructured and clarifications added. | |
May 2014 | 4 | ESD bag recommendation changed, reconditioning procedure added | |
March 2018 | 5 | ESD bag recommendation changed, Pictograms updated, conformal |
coating added
May 2020| 6| | Conformal coating recommendation changed; coating information
added
June 2021| 7| | Soldering information added; new suitable conformal coatings
added
February 2023| 8| All
6
7| Reformatting and reformulation
Updated compatible & Incompatible conformal coatings.
Updated recommendation from SHT3x to SHT4x
June 2023| 8.1| 4| Added Disclaimer for Vapor Phase Soldering in section 2
October 2023| 8.2| Multiple 9
9
10| Corrected typos Image format changed
Changed “absorbs” to “desorbs”
Added section about protective cover visual anomalies
February 2023| 8.3| 10| Added disclaimer about conformal coating residues
Important Notices
Warning, Personal Injury
Do not use this product as safety or emergency stop devices or in any other
application where failure of the product could result in personal injury. Do
not use this product for applications other than its intended and authorized
use. Before installing, handling, using or servicing this product, please
consult the data sheet and application notes. Failure to comply with these
instructions could result in death or serious injury.
If the Buyer shall purchase or use SENSIRION products for any unintended or
unauthorized application, Buyer shall defend, indemnify and hold harmless
SENSIRION and its officers, employees, subsidiaries, affiliates and
distributors against all claims, costs, damages and expenses, and reasonable
attorney fees arising out of, directly or indirectly, any claim of personal
injury or death associated with such unintended or unauthorized use, even if
SENSIRION shall be allegedly negligent with respect to the design or the
manufacture of the product.
ESD Precautions
The inherent design of this component causes it to be sensitive to
electrostatic discharge (ESD). To prevent ESD-induced damage and/or
degradation, take customary and statutory ESD precautions when handling this
product. See application note “ESD, Latchup
and EMC” for more information.
Warranty
SENSIRION warrants solely to the original purchaser of this product for a
period of 12 months (one year) from the date of delivery that this product
shall be of the quality, material and workmanship defined in SENSIRION’s
published specifications of the product.
Within such period, if proven to be defective, SENSIRION shall repair and/or
replace this product, in SENSIRION’s discretion, free of charge to the Buyer,
provided that:
- notice in writing describing the defects shall be given to SENSIRION within fourteen (14) days after their appearance;
- such defects shall be found, to SENSIRION’s reasonable satisfaction, to have arisen from SENSIRION’s faulty design, material, or workmanship;
- the defective product shall be returned to SENSIRION’s factory at the Buyer’s expense; and
- the warranty period for any repaired or replaced product shall be limited to the unexpired portion of the original period.
This warranty does not apply to any equipment which has not been installed and
used within the specifications recommended by SENSIRION for the intended and
proper use of the equipment. EXCEPT FOR THE WARRANTIES EXPRESSLY SET FORTH
HEREIN, SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH RESPECT
TO THE PRODUCT. ANY AND ALL WARRANTIES, INCLUDING WITHOUT LIMITATION,
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSLY EXCLUDED AND DECLINED.
SENSIRION is only liable for defects of this product arising under the
conditions of operation provided for in the data sheet and proper use of the
goods. SENSIRION explicitly disclaims all warranties, express or implied, for
any period during which the goods
are operated or stored not in accordance with the technical specifications.
SENSIRION does not assume any liability arising out of any application or use
of any product or circuit and specifically disclaims any and all liability,
including without limitation consequential or incidental damages. All
operating parameters, including without
limitation recommended parameters, must be validated for each customer’s
applications by customer’s technical experts.
Recommended parameters can and do vary in different applications.
SENSIRION reserves the right, without further notice, (i) to change the
product specifications and/or the information in this document and (ii) to
improve reliability, functions and design of this product.
Headquarters and Subsidiaries
Sensirion AG
Laubisruetistr. 50
CH-8712 Staefa ZH
Switzerland| Sensirion Inc., USA
phone: +1 312 690 5858
info-us@sensirion.com
www.sensirion.com| Sensirion Korea Co. Ltd.
phone: +82 31 337 7700~3
info-kr@sensirion.com
www.sensirion.com/kr
---|---|---
phone: +41 44 306 40 00
fax: +41 44 306 40 30
info@sensirion.com
www.sensirion.com| Sensirion Japan Co. Ltd.
phone: +81 45 270 4506
info-jp@sensirion.com
www.sensirion.com/jp| Sensirion China Co. Ltd.
phone: +86 755 8252 1501
info-cn@sensirion.com
www.sensirion.com/cn
Sensirion Taiwan Co. Ltd
phone: +886 2 2218-6779
info@sensirion.com
www.sensirion.com| To find your local
representative, please visit
www.sensirion.com/distributors
Copyright© 2024, by SENSIRION.
CMOSens® is a trademark of Sensirion.
All rights reserved
www.sensirion.com / D1
Version 8.3 – February 2024
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>