VIQUA Arros 9 Process Water Purification Unit Owner’s Manual
- June 17, 2024
- VIQUA
Table of Contents
Owner’s Manual Models:
Arros 9, Arros 15, Arros 22
NSF Standard 55 Class B
Validated Models:
Arros 9-V, Arros 15-V, Arros 22-V Arros 9-V, Arros 15-V, Arros 22-V
Tested and Certified by
NSF International against CSA
B483.1 and NSF/ANSI 55 for
Disinfection Performance, Class B
425 Clair Rd. W, Guelph, Ontario, Canada N1L 1R1
t. (+1) 519.763.1032 • tf. (+1) 800.265.7246 (US and Canada only)
t. (+31) 73 747 0144 (Europe only) • f. (+1) 519.763.5069
e-mail: info@viqua.com
www.viqua.com
Safety Information
These are the original instructions. Please read this entire manual before
operating this equipment. Pay attention to all danger, warning, and caution
statements in this manual. Failure to do so could result in serious personal
injury or damage to the equipment. Make sure that the protection provided by
this equipment is not impaired. DO NOT use or install this equipment in any
manner other than that specified in the installation manual.
1.1 Potential Hazards:
Read all labels and tags attached to the system. Personal injury or damage to
the system could occur if not observed.
| Waste electrical and electronic equipment (WEEE). This symbol indicates
that you should not discard wasted electrical or electronic equipment (WEEE)
in the trash. For proper disposal, contact your local recycling/reuse or
hazardous waste center.
---|---
| This symbol indicates there is Mercury present.
| This is the safety alert symbol. Obey all safety messages that follow this
symbol to avoid potential injury. When on the equipment, refer to the
Operational and Maintenance manual for additional safety information.
| This symbol indicates a risk of electrical shock and/or electrocution
exists.
| This symbol indicates the marked equipment may contain a component that can
eject forcibly. Obey all procedures to safely depressurize.
| This symbol indicates the system is under pressure.
| This symbol indicates the marked item could inject fluid into the skin.
| This symbol indicates there is a potential UV hazard. Proper protection must
be worn.
| This symbol indicates the marked item could be hot and should not be touched
without care.
| This symbol indicates there is a potential for VERY hot water when flow is
started.
| This symbol indicates not to store any combustible or flammable material
close to the system.
| This symbol indicates that the contents of the transport package are fragile
and the package should be handled with care.
| This symbol indicates safety glasses with side protection is required for
protection against UV exposure.
| This symbol indicates gloves must be worn.
| This symbol indicates safety boots must be worn.
| This symbol indicates the operator must read all available documentation to
perform required procedures.
****| This symbol indicates that the system should only be connected to a
properly grounded, grounding-type controller receptacle that is protected by a
Ground Fault Circuit Interrupter (GFCI).
1.2 Safety Precautions:
DANGER
****| Failure to follow these instructions will result in serious injury or
death.
• Electric Shock: To avoid possible electric shock, special care should be
taken since water is present near the electrical equipment. Unless a situation
is encountered that is explicitly addressed by the provided maintenance and
troubleshooting sections, DO NOT attempt repairs yourself, refer to an
authorized service facility.
• GROUNDING: This product must be grounded. If it should malfunction or
breakdown, grounding provides a path of least resistance for electric current
to reduce the risk of electrical shock. This system is equipped with a cord
having an equipment-grounding conductor and a grounding plug. The plug must be
plugged into an appropriate outlet that is properly installed and grounded in
accordance with all local codes and ordinances. Improper connection of the
equipment-grounding conductor can result in a risk of electrocution. Check
with a qualified electrician or service personnel if you are in doubt as to
whether the outlet is properly grounded. DO NOT modify the plug provided with
this system – if it does not fit in the outlet, have a proper outlet installed
by a qualified electrician. DO NOT use any type of adapter with this system.
• GROUND FAULT CIRCUIT INTERRUPTER PROTECTION: To comply with the National
Electrical Code (NFPA 70) and to provide additional protection from the risk
of electric shock, this system should only be connected to a properly
grounded, grounding-type controller receptacle that is protected by a Ground
Fault Circuit Interrupter (GFCI) or to a residual current device (RCD) having
a rated residual operating current not exceeding 30 mA. Inspect operation of
GFCI as per manufacturer’s suggested maintenance schedule.
• DO NOT operate the UV system if it has a damaged cord or plug, if it is
malfunctioning or if it has been dropped or damaged in any manner.
• DO NOT use this UV system for other than intended use (potable water
applications). The use of attachments not recommended or sold by the
manufacturer / distributor may cause an unsafe condition.
• DO NOT install this UV system where it will be exposed to the weather or to
temperatures below freezing.
• DO NOT store this UV system where it will be exposed to the weather.
• DO NOT store this UV system where it will be exposed to temperatures below
freezing unless all water has been drained from it and the water supply has
been disconnected.
---|---
WARNING
| • During extended periods of no water flow, the water in your chamber can
become very hot (Approx. 60 °C) and potentially lead to scalding. It is
recommended to run your water until this hot water has been purged from your
chamber. Do not allow water to contact your skin during this time. To
eliminate this condition, a temperature management valve can be installed at
the outlet of your UV system.
• Do not pass water through the UV system for a minimum of 5 minutes after
applying power (including after power interruptions) to avoid passing
under-treated water that may, in rare instances, pose health hazards.
• This system contains a UV lamp unit. Do not operate the UV lamp unit when it
is removed from the chamber. Unintended use or damage of the
system may result in the exposure of dangerous UV radiation. UV radiation may,
even in little doses, cause harm to the eyes and skin.
• Changes or modifications made to this system without the consent of the
manufacturer could render the system unsafe for operation and may void
the manufacturer’s warranty.
---|---
| WARNING: This product can expose you to chemicals including phthalates,
which is known to the state of California to cause cancer, and mercury,
which is known to the State of California to cause birth defects or other
reproductive harm. For more information go to www.P65Warnings.ca.gov.
CAUTION
| Failure to follow these instructions could result in minor or moderate
injury.
• Carefully examine the UV system after installation. It should not be plugged
in if there is water on parts not intended to be wet such as, the controller
or lamp connector.
• Hg EXPOSURE: The UV lamp unit contains mercury. If the lamp breaks, then
avoid inhalation or ingestion of the debris and avoid exposure to eyes and
skin. Never use a vacuum cleaner to clean up a broken lamp as this may scatter
the spilled mercury. Obey local regulations and guidelines for the removal and
disposal of mercury waste.
---|---
NOTICE
| • The Arros 9, Arros 15 and Arros 22 UV systems have been validated through
microbial testing. Through this testing, performance data has been generated
for UV dose delivery to inactivate Cryptosporidium, Giardia, Escherichia coli
(E. coli) and Fecal Coliform.
• The Arros 9-V, Arros 15-V and Arros 22-V UV systems inactivate heterotrophs,
Escherichia coli (E. coli) and Fecal Coliform.
---|---
| • The UV lamp unit inside the UV system is rated at an effective life of
approximately 9000 hours. To ensure continuous protection, replace the UV lamp
unit annually.
• The UV system is not to be used or played with by children. Persons with
reduced physical, sensory or mental capabilities, or lack of experience and
knowledge, are also not to handle the UV system unless they have been given
supervision or instruction.
• This system is intended to be permanently connected to the water lines.
• This system is not intended to be used in or above water or outdoors or used
in swimming pools when persons are in the pool.
• EXTENSION CORDS: If an extension cord is necessary, use only 3-wire
extension cords that have 3-prong grounding-type plugs and 3-pole cord
connectors that accept the plug from this system. Use only extension cords
that are intended for outdoor use. Use only extension cords having an
electrical rating not less than the rating of the system. A cord rated for
less amperes or watts than this system rating may overheat. Exercise caution
when arranging the cord so that it will not be tripped over or pulled. DO NOT
use damaged extension cords. Examine extension cord before using and replace
if damaged. DO NOT abuse extension cord. Keep extension cord away from heat
and sharp edges. Always disconnect the extension cord from the receptacle
before disconnecting this system from the extension cord. Never yank cord to
pull plug from outlet. Always grasp the plug and pull to disconnect.
• If the supply cord is damaged, it must be replaced by a special cord or
assembly available from the manufacturer or its service agent.
• SYSTEM PROTECTION: To protect your Controller, a UL1449 certified (or
equivalent) transient voltage surge suppressor is strongly recommended.
• The UV lamp unit in this system conforms to the applicable provisions of the
Code of Federal Regulations (CFR) requirements including, Title 21, Chapter 1,
Subchapter J, Radiological Health.
• Only the following Class B systems Arros 9-V, Arros 15-V and Arros 22-V or
component conforms to NSF/ANSI 55 for the supplemental bactericidal treatment
disinfected public drinking water or other drinking water that has been tested
and deemed acceptable for human consumption by the state or local health
agency having jurisdiction. The system is only designed to reduce normally
occurring non-pathogenic nuisance microorganisms. Class B systems are not
intended for treatment of contaminated water.
• Read and understand the Owner’s Manual before operating and performing any
maintenance on this equipment.
| • This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
• This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/ TV technician for help.
• Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
1.3 Water Chemistry
Water quality is extremely important for the optimum performance of your UV
system. The following levels are recommended for installation:
Water Quality and Minerals | Level |
---|---|
Iron | < 0.3 ppm (0.3 mg/L) |
Hardness* | < 7 gpg (120 mg/L) |
Turbidity | < 1 NTU |
Manganese | < 0.05 ppm (0.05 mg/L) |
Tannins | < 0.1 ppm (0.1 mg/L) |
UV Transmittance | > 75% (call factory for recommendations on applications |
where UVT < 75%)
- Where total hardness is less than 7 gpg, the UV unit should operate efficiently provided the quartz envelope is cleaned periodically. If total hardness exceeds 7 gpg, the water should be softened. If your water chemistry contains levels in excess of those mentioned above, proper pre-treatment is recommended to correct these water problems prior to the installation of your UV system. These water quality parameters can be tested by your local dealer, or by most private analytical laboratories. Proper pre-treatment is essential for the UV system to operate as intended.
General Information
Item | Description | Part Number | UV Systems |
---|---|---|---|
1 | Flow Restrictor (Only for certified systems) | FRSS-A9 | Arros 9-V |
FRSS-A15 | Arros 15-V | ||
FRSS-A22 | Arros 22-V | ||
2 | UV Lamp Unit- Hard glass coated UV lamp for long consistent life (9000 | ||
hours) sealed within a GE 214 fused quartz sleeve | VA09LU | Arros 9, Arros 9-V | |
VA15LU | Arros 15, Arros 15-V | ||
VA22LU | Arros 22, Arros 22-V | ||
3 | Arros Controller 100-240VAC/50/60Hz | VA-CTRL | All Models |
4 | Power Cords for Arros Controller | 260251-R | Arros 9, Arros 15, Arros 22 |
260252-R | Arros 9/2, Arros 15/2, Arros 22/2 | ||
260253-R | Arros 9/2B, Arros 15/2B, Arros 22/2B | ||
260254-R | Arros 9/2A, Arros 15/2A, Arros 22/2A | ||
5 | Drain Plug | 411111-R | All Models |
6 | Drain Plug O-ring | 411138-R | All Models |
7 | Mounting Bracket Arros | 420618-R | Arros 9 & Arros 15 |
Mounting Bracket Arros | 420624-R | Arros 22 | |
8 | Mounting Bracket Clamp | 420619-R | All Models |
9 | Lamp Head O-ring | 411173-R | All Models |
Installation of UV System
CAUTION
Electronic controller must be connected to a Ground Fault Protected Circuit
(GFCI) receptacle. Ensure green ground wire ring terminal is securely fastened
to ground stud on UV chamber.
3.1 Installing UV System
The UV system is designed to be mounted either horizontally or vertically at
the point-of-use or point-of-entry depending on the specific flow rate of the
unit.
If installing the chamber in horizontal position the outlet port must be
pointing upwards to ensure all air is fully purged from the chamber.
Note: The preferred installation is vertical with the lamp connector on top.
This is to prevent water damage from occurring on the lamp pins and lamp
connector. If horizontal install is required ensure inlet port is positioned
facing downward.
Note: To reverse inlet port to suit installation plumbing see Section 3.1
procedure 4.
Prerequisites:
- Determine appropriate indoor location for the UV system. Refer to Figure 3.
- Ensure adequate clearance above chamber to allow for removal of the UV lamp unit.
- Make sure to turn off the main water supply and depressurize the line.
- Mount the system to the wall with appropriate lag bolts through the two mounting holes located on the metal bracket.
- The supplied lag bolts are suitable for attachment to wood. (Attachments to other materials will require purchasing of alternative hardware).
- Make all necessary plumbing connections. Refer to Figure 3.
Procedure:
- Figure 3 shows the installation of a typical UV system and the related components that may be used for the installation.
The use of a by-pass assembly is recommended in case the system requires “off- line” maintenance. In this case, note the system requires a supplementary disinfection for the distribution system if any water is used during by-pass condition. In addition, during by-pass, the water will NOT be disinfected and a “DO NOT CONSUME THE WATER” tag should be physically installed on the by-pass assembly until such time as the system is sanitized and returned to service. For more information, refer to Section 3.2. If the water is to be consumed while the system is off-line, the water must be boiled for twenty minutes prior to consumption. Item| Description| Item| Description
---|---|---|---
1| Main water shut-off| 5| UV System
2| Filter Cartridge| 6| Optional flow restrictor
Note: Mandatory for NSF credited systems
3| Optional by-pass assembly
4| Drain Plug| 7| Customer supplied GFCI circuit to match voltage
|
A| 6” Note: By-pass pipe should be at least 6” above the Arros UV unit and as close to the wall as possible to ensure lamp can be removed. - Select a suitable location for the UV system and its related components. As it is recommended to install a GFCI, make sure that this is taken into consideration prior to any installation. The system can either be installed vertically as shown in Figure 4 A, or horizontally as shown in Figure 4 B. However, the vertical installation is the most preferred method. When selecting a mounting location, leave enough space to allow the removal of the lamp unit (typically leave a space equal to the size of the chamber itself).
- Mount the system to the wall using the supplied bracket. Various connection methods can be used to connect the water source to the system, however union type connectors are recommended. For validated systems, the use of a flow restrictor device will help to maintain the manufacturers rated flow. The flow restrictor should be installed on the outlet port and is designed to be installed in one direction only. Ensure that the flow of the water matches the flow direction as indicated on the flow restrictor. Refer to Figure 5.
Note: DO NOT solder connections while attached to the system as this could damage the O-ring seals.![VIQUA Arros 9 Process Water Purification Unit
-
Flow Restrictor](https://manuals.plus/wp-content/uploads/2024/02/VIQUA- Arros-9-Process-Water-Purification-Unit-Flow-Restrictor.png)
-
Reversing the inlet and outlet ports.
The factory configuration of the UV system is with the inlet port on the left side of the unit. If required due to plumbing installations then this can be reversed as follows:
1. Remove the locknut found on the back side of the UV system.2. Remove the two Philips screws that attach the controller unit to the chamber.
3. Pull the controller off of the chamber and rotate it 180 degrees. 4. Align the mounting holes on the controller with the welded10 stud mounts on the chamber and insert the ground wire back onto the
threaded stud.
5. Push the controller firmly down onto the chamber.6. Reassemble the mounting screws to secure the controller to the chamber
7. Tighten down the locking nut to secure the grounding wire. -
Install the UV lamp unit into the chamber as shown in Figure 6. Twist lamp unit until fully seated then swing lamp connector around and position over top of the lamp and press down until the lamp pins are fully inserted into the lamp connector.
-
When all plumbing connections are complete, slowly turn on the water supply and check for leaks. The most likely cause of leaks is from an O-ring seal. In case of a leak, shut water off, drain chamber, remove the lamp unit or drain plug, wipe the O-rings and threads. Clean and re-install.Once it is determined that there are no leaks, plug the system into the ground fault interrupter and check controller to ensure the system is operating properly. The controller should illuminate without any alarms.
-
Allow the water to run for a few minutes to clear any air or dust that may be in the chamber.
Note: When there is no flow, the water in the chamber will become warm. To remedy this, run a cold water tap anywhere in the house for a minute to flush out the warm water. -
Once it is determined that there are no leaks, plug the system into the ground fault interrupter and check controller to ensure the system is operating properly. The controller should illuminate without any alarms.
Note: DO NOT look directly at the glowing UV lamp unit.
-
3.2 Cleaning Procedure
As UV does not provide a disinfection residual, it is imperative that the
entire distribution system located after the UV be chemically disinfected to
ensure that the plumbing system is free from any bacteriological contaminants.
The disinfection process must be performed immediately after the UV unit is
installed and repeated thereafter whenever the UV is shut down for service
without power, or inoperative for any reason. The procedure for sanitizing the
plumbing system is readily accomplished as follows:
-
Ensure that the unit remains powered on for the entire cleaning process.
-
Shut off the water supply.
• Close each faucet. -
Press the pressure button to release the pressure from the cartridges.
-
Remove sump housing(s) using sump wrench.
-
Remove cartridge(s) and pour 2 cups of household bleach solution into the filter housing(s).
Note: DO NOT use Hydrogen Peroxide. -
Reinstall sump only to filter head.
-
Turn on water supply.
• Open a tap down stream and allow water to fill the chamber until you smell bleach. -
Turn on the cold water supply followed by hot water (if available) until you smell the bleach.
• Close all faucets and allow bleach to settle in the water lines for 30 minutes. -
Shut off the water supply.
• Press the pressure button to release the pressure. -
With all faucets closed, remove filter housing(s) using sump wrench.
-
Reinstall the cartridge(s) into filter housing(s) and connect to the unit.
• Turn on water supply and flush all water outlets until bleach can no longer be smelled (at least 5 minutes). -
Press the pressure button to purge air and to complete the cleaning procedure.
Note:
- The addition of chlorine (bleach) to a hot water tank that has in the past been fed with untreated raw water with high levels of other contaminants (iron, manganese, hydrogen sulphide, organics, etc.) will result in oxidation of these contaminants and may require repeated flushing of the hot water tank. This contingency must be dealt with independently under the start-up procedure for any other conditioners that may form a part of the pre-treatment for the UV unit.
- The above cleaning procedure will result in a massive chlorine residual far in excess of the 0.5 to 1.0 mg/L typically present in municipally chlorinated water and of a magnitude consistent with the minimum 50 mg/L chlorine solution recommended for the cleaning of distribution systems known to be contaminated. Do not consume water until complete system has been flushed.
Maintenance
WARNING
| • Always disconnect power before performing any work on the UV system.
• Always shut-off water flow and release water pressure before servicing.
• Regularly inspect your UV system to ensure that the power indicators are on
and no alarms are present.
• Replace the UV lamp unit annually upon controller alert, typically after 1
year or 9000 hours of operation.
• Always drain the chamber when closing a seasonal home or leaving the unit in
an area subject to freezing temperatures.
---|---
4.1 Replacing UV Lamp Unit
NOTICE
- Refer to Section 5.1. Refer to www.lamprecycle.org for lamp disposal.
- Do not use water during replacement of lamp unit.
Lamp unit replacement is a quick and simple procedure requiring no special
tools. The UV lamp unit must be replaced after 9000 hours of continuous
operation (approximately one year) in order to ensure adequate performance.
Procedure:
-
Shut off the water line to UV chamber and release system pressure before servicing.
-
Disconnect main power source and allow the unit to cool for 10 minutes.
-
To remove lamp connector from the lamp unit pull lamp connector straight up until the pin sockets disengage from the lamp unit pins.
• Rotate lamp connector clockwise until a stop is hit. Then press down to lock in the service position. -
To remove lamp unit rotate counter clockwise until the lamp unit becomes free from the bayonet mounting taps.
• Carefully pull lamp unit straight out from chamber ensuring the lamp unit does not contact the chamber.
• Always hold the lamp unit by the plastic lamp head. -
Insert the new lamp unit fully into the chamber and rotate clockwise until lamp is fully inserted and aligns with the lock position shown on the controller top.
-
Rotate lamp connector counter clockwise from the service position so it is over top of the lamp unit connector pins.
• Press down until the lamp connector fully engages the lamp unit pins.
4.2 Cleaning UV Lamp Envelope
Minerals in the water may slowly coat the quartz envelope. It may periodically
become necessary to clean the quartz envelope.
Operations
WARNING
The advanced warning system has been installed to provide the optimum
protection against microbiological contamination in water. DO NOT disregard
the warning signals. The best way to ensure optimum UV performance is to have
the water microbiologically tested by a recognized testing agency on a regular
basis.
5.1 UV System Operation
Note: Firmware updates may modify the UI screens and processes. If
necessary, refer to the online manual for most up-todate information.
5.1.1 Start Up Sequence
- During the startup sequence the Controller will cycle thru the following screens: (Use the arrows to scroll and press
when the appropriate value is displayed)
a. Language Selection: English, French, Spanish with English being the default option.b. Model Select: Pressing the ? Icon will direct you to where the model number label can be found.c. Screen Orientation Setup: When mounting the UV system in a horizontal installation the Display Screen can be configured to the most preferred orientation.
Click on the arrows to change the LCD screen orientation. Once desired orientation is shown press done.d. Warranty Registration (Recommended): Follow the QR code to www.trojantechnologies.com/register to complete the online registration. Once completed (or if skipping this step), press Done to complete the Startup Sequence.
www.trojantechnologics.com/register
5.2 Home Screen
- The Home Screen displays the lamp life remaining and counts down from 365 to 0 days as the system is ON.
- The Home Screen also displays Lamp Status ON/OFF.
- A Minor Warning Alarm is initiated once the lamp life reaches 30 days remaining and will also alarm when the lamp life reaches 14 days remaining.
- A Major Lamp Alarm will be initiated once the lamp life reaches 0 days.
Note: Use of a non-VIQUA lamp unit will result in an ongoing fault mode (major lamp alarm) and will impair accurate lamp-life tracking.
5.3 Menu Screen
- From the Home Screen, press the top left icon to access the main menu screen.
- The Menu Screen provides access to four main user screens:
a. A spare parts menu screen for finding replacement part numbers.
b. A Setting menu for setting or changing system settings such as System model number, Display and Halo brightness, orientation, user language, screen saver time, and Reset that returns all settings back to default settings.
c. Information menu which provides system information such as firmware version, Model number and serial number.
d. Contact dealer information menu provides information on finding a dealer in your area.
5.4 Lamp Head Status Lights
The white top cap (Halo) of the lamp assembly will illuminate when the lamp is
operating. For normal operation (no faults, lamp on) the caps is blue. When
there is a minor alarm condition (lamp life less than 30 days left), the top
cap is amber. When there is a major alarm condition, the cap will be red.
5.5 Firmware Updates
- When firmware updates are developed to enhance the performance or add features to the current controller this new firmware can be uploaded via a smart phone and associated downloadable application program.
- For details on this feature go to www.trojantechnologies.com for details of when upgrades are available and instructions on how to upload the new firmware.
Troubleshooting
Symptom | Possible Cause | Possible Solution |
---|---|---|
No Power | GFCI and or circuit breaker has tripped | Reset GFCI and or circuit |
breaker
Transient voltage surge suppressor damaged| Replace TVSS
Controller damaged| Replace controller and use a TVSS
GFCI or circuit breaker repeatedly trips| Connection between lamp and lamp
plug is wet| Clean and dry area around the lamp and lamp connector, check unit
for leaks or condensation
Short circuit in electrical assembly| Replace controller
Leak at inlet or outlet| Threaded pipe fittings are leaking| Clean threads,
reseal with Teflon tape and re tighten
Leak detected from area of UV chamber| Condensation of warm moist air on cold
chamber (slow accumulation)| Control humidity or relocate unit
O-rings damaged, deteriorated or incorrectly installed| Inspect o-rings and
replace if deteriorated.
Lamp unit not properly installed (not fully inserted into controller lamp
locking assembly)| Tighten and ensure lamp assembly is fully inserted into
controller locking assembly
Alarm| Refer to Alarm Menu Sec 5| Follow alarm menu help guidance
System is operating but water tests reveal bacterial contamination| Equipment
downstream of UV system is acting as a breeding ground for pathogens| Ensure
UV is the last piece of treatment equipment
Pathogens are residing in the distribution lines post UV| Ensure all
distribution lines have been disinfected with chlorine. Refer to Section 3.2.
Recontamination from pipe dead-ends| Remove any pipe dead-ends and flush with
chlorine. Refer to Section 3.2.
Pressure Drop| Sediment Pre-Filter clogged| Replace filter cartridge with
appropriate 5 micron cartridge.
Note: Check source water supply as fluctuations may occur in source
pressure
Flow regulator| Flow regulator will result in pressure drop when approaching
full flow.
High Bacteria Counts| Lamp unit quartz envelope is stained or dirty| Clean
quartz envelope of lamp unit with scale cleaner and eliminate the source of
staining problem (i.e soften hard water) Refer to Section 5
Change in feed water quality| It is imperative that the effluent water stream
be shocked with Chlorine (bleach) before water leaves UV system-disinfection
system must have a bacterial free distribution system to work effectively
Refer to Section 3.2.
Contamination in water lines after UV system| Have source water tested to
ensure that the water quality is still within allowable limits for this system
Possible break-through of sediment through pre-filter| Have source water
tested for turbidity- may need stepped filtration in order to catch all
sediment entering water system (20 micron filter followed by a 5 micron filter
followed by the UV system)
Heated Product Water| Common problem caused by infrequent use of water| Run
water until it returns to ambient temperature
Water Appears Milky| Caused by air in the water lines| Run water until air is
purged
Unit Leaking Water| Problem with O-ring seal on Lamp Assembly or Drain Plug
Assembly| Ensure O-ring is in place, check for any cuts or abrasions, clean
O-ring, moisten with water/lubricant and re-install, replace if necessary
Condensation on chamber caused by excessive humidity and cold water| Check
location of UV system and control humidity
Inadequate inlet/outlet port connections| Check thread connections, reseal
with Teflon tape and re- tighten connection
System shutting down intermittently| Interrupted Controller| Ensure system has
been installed on its own circuit as other equipment may be drawing power away
from UV (ie pump or fridge)
UV system should not be installed on a circuit which is incorporated into a
light switch
Lamp Failure Alarm on New Lamp| Loose connection between lamp and lamp
connector| Disconnect lamp from connector and reconnect, ensuring that a tight
fit is accomplished
Moisture build up in connector may keep lamp and connector from making a solid
connection| Eliminate chance of any moisture getting to the connector and/or
lamp pins
Manufacturer’s Dose Flow Chart
Note: Not performance tested or certified by NSF
NSF Standard 55 Class B Certified Systems
Model | Flow Rate | Min Dose @ 70% UVT |
---|---|---|
Arros 9-V | 6.4 gpm | 16 mJ/cm² |
Arros 15-V | 9.4 gpm | 16 mJ/cm² |
Arros 22-V | 12.9 gpm | 16 mJ/cm² |
Note: Performance tested or certified by NSF
Specifications
Model| ARROS 9 ARROS 9-V| ARROS 15 ARROS 15-V| ARROS 22
ARROS 22-V
---|---|---|---
Flow Rate1| | *NSF class B certified 16mJ/cm2 @70% UVT| 6.4 gpm (24.2 lpm)
(1.45 m3/hr)| 9.4 gpm (35.58 lpm) (2.13 m3/hr)| 12.9 gpm (48.8 lpm) (2.93
m3/hr)
Not performance tested or certified by NSF| US Public Health 16mJ/cm2 @
95%UVT| 17 gpm (64.3 lpm) (3.86 m3/hr)| 29 gpm (109.8 lpm) (6.59 m3/hr)| 42
gpm (159 lpm) (9.54 m3/hr)
VIQUA Standard 30mJ/cm2 @ 95%UVT| 9 gpm (34.1 lpm) (2.0 m3/hr)| 15 gpm (56.8
lpm) (3.41 m3/hr)| 22 gpm (83.3 lpm) (5.0 m3/hr)
NSF/EPA 40mJ/cm2 @ 95%UVT| 7 gpm (26.5 lpm) (1.59 m3/hr)| 12 gpm (45.4 lpm)
(2.73 m3/hr)| 16 gpm (60.6 lpm) (3.63 m3/hr)
Dimensions| Chamber| 28.4 cm x 10.16 cm (11.2″ x 4.0″)| 42.0 cm x 10.16 cm
(16.5” x 4.0”)| 57.2 cm x 10.16 cm (22.52” x4.0”)
Controller| 26.67cm x 12.7 cm x 16.51cm (10.5” x5.0” x 6.5”)| 26.67cm x 12.7
cm x 16.51cm (10.5” x5.0” x 6.5”)| 26.67cm x 12.7 cm x 16.51cm (10.5” x5.0” x
6.5”)
Inlet/Outlet Port Size| 3/4” MNPT| 3/4” MNPT| 3/4” MNPT
Shipping Weight| 12 lbs| 15 lbs| 18 lbs
Electrical| Voltage2| 100-240 V / 50/60 Hz| 100-240 V / 50/60 Hz| 100-240 V /
50/60 Hz
Max. Current| 1.5 Amp| 1.5 Amp| 1.5 Amp
Power consumption| 35 W| 50 W| 65 W
Lamp Watts| 24 W| 38 W| 52 W
Maximum Operating Pressure| 125 psi| 125 psi| 125 psi
Ambient Water Temperature| 2-40 °C (36-104 0F)| 2-40 °C(36-104 0F)| 2-40 °C
(36-104 0F)
Lamp Type| High Output| High Output| High Output
Display| LCD Touch Screen| LCD Touch Screen| LCD Touch Screen
Lamp Replacement Reminder| Yes| Yes| Yes
Visual Lamp Life| Yes| Yes| Yes
True Lamp Running time| Yes| Yes| Yes
System Total Running Time| Yes| Yes| Yes
Chamber Material| 304 SS| 304 SS| 304 SS
1Flow rates based on End of Lamp Life; 20°C.
2Units ending in “/2” are for 230V applications.
Manufacturer’s Warranty
Our Commitment
VIQUA is committed to ensuring your experience with our products and
organization exceeds your expectations. We have manufactured your UV system to
the highest quality standards and value you as our customer. Should you need
any support, or have questions about your system, please contact our Technical
Support team at 1.800.265.7246 or
technicalsupport@viqua.com and we
will be happy to assist you.
How to Make a Warranty Claim
Note: To maximize the performance and reliability of your VIQUA product,
the system must be properly sized, installed and maintained. Guidance on the
necessary water quality parameters and maintenance requirements can be found
in your Owner’s Manual.
In the event that repair, or replacement of parts covered by this warranty are
required, the process will be handled by your dealer. If you are unsure
whether an equipment problem or failure is covered by warranty, contact our
Technical Support team at 1.800.265.7246 or e-mail
technicalsupport@viqua.com. Our fully
trained technicians will help you troubleshoot the problem and identify a
solution. Please have available the model number (system type), the date of
purchase, the name of the dealer from whom you purchased your VIQUA product
(“the source dealer”), as well as a description of the problem you are
experiencing. To establish proof of purchase when making a warranty claim, you
will either need your original invoice, or have previously completed your
product registration online.
Specific Warranty Coverage
Warranty coverage is specific to the VIQUA range of products. Warranty
coverage is subject to the conditions and limitations outlined under “General
Conditions and Limitations”.
Ten-Year Limited Warranty for VIQUA UV Chamber
VIQUA warrants the UV chamber on the VIQUA product to be free from defects in
material and workmanship for a period of ten (10) years from the date of
purchase. During this time, VIQUA will repair or replace, at its option, any
defective VIQUA UV chamber. Please return the defective part to your dealer
who will process your claim.
Five-Year Limited Warranty for Electrical and Hardware Components
VIQUA warrants the electrical (controller) and hardware components to be free
from defects in material and workmanship for a period of five (5) years from
the date of purchase. During this time, VIQUA will repair or replace, at its
option, any defective parts covered by the warranty. Please return the
defective part to your dealer who will process your claim.
One-Year Limited Warranty for UV lamp Units
VIQUA warrants UV lamp unit to be free from defects in material and
workmanship for a period of one (1) year from the date of purchase. During
this time, VIQUA will repair or replace, at its option, any defective parts
covered by the warranty. Your dealer will process your claim and advise
whether the defective item needs to be returned for failure analysis.
Note: Use only genuine VIQUA replacement lamp units in your system.
Failure to do so may seriously compromise performance and affect warranty
coverage.
General Conditions and Limitations
None of the above warranties cover damage caused by improper use or
maintenance, accidents, acts of God or minor scratches or imperfections that
do not materially impair the operation of the product. The warranties also do
not cover products that are not installed as outlined in the applicable
Owner’s Manual. Parts repaired or replaced under these warranties will be
covered under warranty up to the end of the warranty period applicable to the
original part. The above warranties do not include the cost of shipping and
handling of returned items.
The limited warranties described above are the only warranties applicable to
the VIQUA range of products. These limited warranties outline the exclusive
remedy for all claims based on a failure of or defect in any of these
products, whether the claim is based on contract, tort (including negligence),
strict liability or otherwise. These warranties are in lieu of all other
warranties whether written, oral, implied or statutory. Without limitation, no
warranty of merchantability or of fitness for a particular purpose shall apply
to any of these products. VIQUA does not assume any liability for personal
injury or property damage caused by the use or misuse of any of the above
products. VIQUA shall not in any event be liable for special, incidental,
indirect or consequential damages. VIQUA’s liability shall, in all instances,
be limited to repair or replacement of the defective product or part and this
liability will terminate upon expiration of the applicable warranty period.
425 Clair Rd. W, Guelph, Ontario, Canada N1L 1R1
t. (+1) 519.763.1032 • tf. (+1) 800.265.7246 (US and Canada only)
t. (+31) 73 747 0144 (Europe only) • f. (+1) 519.763.5069
e-mail:info@viqua.com
www.viqua.com
Documents / Resources
|
VIQUA Arros 9 Process Water Purification
Unit
[pdf] Owner's Manual
Arros 9, Arros 15, Arros 22, Arros 9 Process Water Purification Unit, Process
Water Purification Unit, Water Purification Unit, Purification Unit, Unit
---|---
References
- Lamp Recycle – Environmental Responsiblility Starts Here
- P65Warnings.ca.gov
- Trojan Technologies | Trojan Technologies
- Product Registration - VIQUA
- VIQUA Home Page - VIQUA
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>