SCILOGEX SK-R330-Pro LCD Digital Rocking Shaker
Specifications:
- Model: SK-R330-PRO / SK-D3307-PRO / SK-D3309-PRO
- Type: LCD digital Rocking Shaker / LCD digital 3D Shaker / LCD digital 3D Shaker
- Designed for: Mixing sticky substances in schools, laboratories, or factories
Product Usage Instructions:
1. Safety Instructions:
Before using the instrument, ensure the following safety precautions:
- Make sure to have protective ground contact before use.
- Avoid splashing liquids and using broken glass containers.
- Operate the instrument on a clean, non-slip, dry, and fireproof surface.
- Do not use the instrument in explosive atmospheres, with hazardous substances, or underwater.
2. Proper Use:
The instrument is designed for mixing sticky substances in educational institutions, laboratories, or industrial settings. It can be used with a variety of impellers based on the medium’s viscosity. Avoid using the instrument in residential areas or other restricted environments outlined in Chapter 1. Always follow the manufacturer’s recommendations for accessories and usage instructions to prevent unsafe situations.
3. Inspection:
3.1 Receiving Inspection:
Upon receiving the instrument, carefully unpack and check for any damages incurred during transport. If any damage is found, contact the manufacturer or supplier for technical support.
3.2 Listing of Items:
Items | Qty |
---|---|
Main unit | 1 |
Power cable | 1 |
User manual | 1 |
4. Control and Display:
4.1 Control:
The instrument features various controls for operation:
- Speed control knob
- Timer control knob
- Speed display
- Time display
- Power switch
4.2 Display:
The display provides information on set values, faults, and running values of the instrument.
5. Operation:
Ensure that the operating voltage matches the power supply voltage and that the socket is properly earthed. You can adjust the speed value during operation.
6. Faults:
In case of any faults or errors during operation, refer to the user manual for troubleshooting steps or contact the manufacturer for technical support.
FAQ
-
Q: What should I do if the instrument displays an error?
- A: If an error occurs, stop using the instrument immediately and refer to the user manual for guidance on troubleshooting steps. If the issue persists, contact the manufacturer’s service department for assistance.
-
SK-R330-Pro LCD digital Rocking Shaker
-
SK-D3307-Pro LCD digital 3D Shaker
-
SK-D3309-Pro LCD digital 3D Shaker
Preface
Welcome to the “Rocking Shaker & 3D Shaker User Manual”. Users should read this manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
Service
When help needed, you can always contact the service department of manufacturer for technical support in the following ways:
SCILOGEX, LLC
- 500 Four Rod Road
- Suite 122
- Berlin, CT 06037 USA
- Tel: 1-860-828-5289
- Fax: 1-860-828-5389
- E-mail: info@scilogex.com
- http://www.scilogex.com
Please provide the customer care representative with the following information:
- Serial number ( on the rear panel )
- Certification
- Description of problem ( i.e., hardware or software )
- Methods and procedures adopted to resolve the problems
- Your contact information
Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation. For claims under the warranty please contact your local supplier. You may also send the instrument directly to the manufacturer, enclosing the invoice copy and by giving reasons for the claim.
Safety Instructions
| Warning!
• Read the operating instructions carefully before use.
• Ensure that only trained staff works with the
instrument.
---|---
| Protective ground contact!
• Ensure the socket must be grounded (protective ground contact) before use.
- When working wear personal safety guards to avoid the risk of:
- Splashing liquids
- Broken glass containers
- Follow the safety instructions, guidelines and accident prevention regulations.
- Do not touch the running parts, moving instruments care not rolling your fingers.
- Set up the instrument in a spacious area on a stable, clean, non-slip, dry and fireproof surface. Do not operate the instrument in explosive atmospheres, with hazardous substances or under water.
- If the instrument does not run smoothly, please decrease the motor speed.
- Firmly secure the accessories and vessels in place to avoid damage or risk.
- The preparation of samples may lead to dangerous flammable. Only process samples that will not react dangerous.
- Use the standard accessories listed in the “accessories” section, and follow the instructions to use accessories to ensure safety. Please switch off the power before assembly of accessories, confirm the instrument and accessories are intact before switch on each time.
- The instrument only be opened by expert, please switch off before use.
- The voltage stated on the nameplate must correspond to the mains voltage.
- Do not cover the instrument during running. Prevent the collision and extrusion to instrument and accessories.
- Keep away from high magnetic field.
Proper Use
The instrument is designed for mixing sticky substance in schools, laboratories or factories. It can be installed on a variety of impeller, for different viscosity of the medium. This instrument is not suitable for using in residential areas or other constraints mentioned in Chapter 1. Do not use the accessories recommended by the manufacturer, or failure to use the instructions may be caused unsafe situation.
Inspection
Receiving Inspection
Unpack the instrument carefully and check for any damages which may have arisen during transport. If it happens, please contact manufacturer/supplier for technical support
Note:
If there is any apparent damage to the system, please do not connect to the power line
Listing of Items
Items | Qty |
---|---|
Main unit | 1 |
Power cable | 1 |
User manual | 1 |
Control and Display
Items | Descriptions |
---|
Speed control
knob
| Set target speed. Push on/off the knob to
start/stop the shaking function.
Timer control
knob
| Set target time. Push on/off the knob to
start/stop the shaking function.
Speed display| Display the set and real speed values.
Time display| Display the set and remain time values.
Power switch| Switch ON/OFF instrument.
Display
Display | Descriptions |
---|---|
Set | Display when set target value. |
Err | Display in case of error happening. |
Display area | When Set display, this area shows setting value; When Set |
disappears, this
area shows running value.
Operation
- Place the instrument in safe and stable surface, ensure the required operating voltage and power supply voltage matched, and the socket must be earthed reliably.
- Connect the power cable.
- Switch ON instrument
- The instrument begins self-checking.
- When initialization is over, LCD displays the last running values of speed.
- Rotate the speed control knob to set target speed.
- a. Press the speed control knob to start the shaking function. Press the speed control knob again to stop the shaking function.
- b. Rotate the timer control knob to set the target time. Press the timer control button to start shaking function at the set speed and time. During operation, motion can be stopped at any time by pressing the speed control knob or timer control knob. If the timer control knob is pressed again, the shaking function will start again and the timer will restart countdown. When the timer reaches zero, the unit will be automatically halted and an alert will sound. If these operations above are normal, the instrument is ready to operate. If not, the instrument may be damaged during transportation, please contact technical support of the manufacturer/supplier.
Note:
The speed value can be altered during shaking.
Faults
- When switched ON, the instrument does not work.
- Check whether the power cable is connected.
- Check whether the fuse is broken.
- Shaking function suddenly stop.
- LCD displays “Er 03”, indicate the current failure is “overload protection”.
- Pressing the speed control knob or timer control knob first to close shaking function, and then restart shaking function after lowering load. If shaking function cannot start, then repeat the process and gradually reduce load.
Maintenance and Cleaning
- Proper maintenance can keep the instrument working properly and lengthen its lifetime.
- Do not spray cleanser into the instrument when cleaning.
- Do not remove the power line when cleaning. Only use recommended cleansers:
Dyes | Isopropyl alcohol |
---|
Construction
materials
| Water containing tenside /
Isopropyl alcohol
Cosmetics| Water containing tenside /
Isopropyl alcohol
Foodstuffs| Water containing tenside
Fuels| Water containing tenside
- Before using other method for cleaning or decontamination, the user must ascertain with the manufacturer that this method will not damage the instrument. Wear the proper protective gloves during cleaning of the instrument.
Note:
- Electronic device cannot clean with cleanser.
- If you require maintenance service, must be cleaned the instrument in advance to avoid pollution of hazardous substances, and to send back into original packing.
- If the instrument will not use for a long time, please switch off and place in a dry, clean, room temperature and stable location
Associated standards and regulations
Construction in accordance with the following safety
standards:
- EN 61010-1
- UL 3101-1
- CAN/CSA C22.2(1010-1)
- EN 61010-2-10
- Construction in accordance with the following EMC
standards:
- EN 61326-1
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense
Specifications
Items
| Specifications
---|---
Rocking shaker| 3D shaker
Overall Dimensions [mm]| ****
360×420×155
| ****
430×360×106
Weight [kg]| 12| 9
Permissible ambient temperature [°C]| ****
5-40
| ****
5-40
Permissible relative
humidity
| ****
80%
| ****
80%
Protection class acc. to
DIN EN60529
| ****
IP21
| ****
IP21
Items
| Specifications
---|---
Rocking shaker| 3D shaker
Voltage [VAC]| 100-240| 100-240
Frequency [Hz]| 50/60| 50/60
Power [W]| 40| 50
Tile Angel [o]| ±9| ±7, ±9
Max. Shaking
Weight [kg]
| 7| 5
Motor Type| DC motor| DC motor
Speed range [rpm]| 10-70| 10-70
Speed display| LCD| LCD
Timer range [min]| 1-1199| 1-1199
Timer display| LCD| LCD
Run type| Timer/Continuous
operation
| Timer/Continuous
operation
Ordering Informations