EKO LBL215 DUO Digital Stethoscope User Manual
- June 17, 2024
- EKO
Table of Contents
- EKO LBL215 DUO Digital Stethoscope
- Product Usage Instructions
- FAQ
- Indications for Use
- Setup
- Turning On and Off, Changing Filters, and Taking a Recording
- Operating Conditions
- Eko App Use
- Electrical Safety
- Electrical Safety
- Electrical Safety
- Manufacturing and Regulatory Information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
EKO LBL215 DUO Digital Stethoscope
Specifications
- Product: Eko DUO System
- Intended Use: Analyzing cardiac and other internal organ sounds
- Components: DUO device, smartphone application (App), web application (Dashboard)
- Restrictions: Not for use on infants weighing less than 10kg
Product Usage Instructions
Indications for Use
The device data should be used in conjunction with physician oversight and other relevant patient data. Not for use on infants weighing less than 10kg.
Introduction, Warnings, and Safety
The Eko DUO System is designed for healthcare professionals to analyze cardiac and internal organ sounds. Includes DUO device, smartphone App, and web Dashboard.
CAUTION: Only for sale to licensed medical professionals.
For Help and Assistance
Contact Eko for general inquiries, support, and warranty information.
- Website: www.ekohealth.com
- Email: support@ekohealth.com
- Phone: 1.844.356.3384
Safety Related Labels & Symbols
Refer to instructions for use. Contains no natural rubber latex. Non-sterile product. Do not attempt to re-sterilize.
Caution
Follow safety instructions provided in the manual for optimal usage.
Skin Preparation
Ensure patient’s skin is clean and free from excess hair for quality ECG tracing. Do not use over wounds or broken skin. Avoid alcohol-based cleaners.
FAQ
Q: Can the DUO device be used on infants?
A: The device should not be used on infants weighing less than 10kg.
Q: What is the warranty information for the DUO system?
A: Eko provides a limited warranty for the DUO system. For full warranty details, visit www.ekohealth.com/warranty.
Indications for Use
- The Eko DUO System is intended to be used by healthcare professionals to electronically amplify, filter, and transfer body sounds and single-channel electrocardiogram (ECG) waveforms. The Eko
- DUO System also displays ECG waveforms and phonocardiogram waveforms on the accompanying mobile application for storage and sharing (when prescribed or used under the care of a physician).
- It can be used to record heart sounds and cardiac murmurs, bruits, respiratory sounds, and abdominal sounds during physical examination in normal patients or those with suspected diseases of the cardiac, vascular, pulmonary, or abdominal organ systems. The device can be used on adults and pediatrics.
- The data offered by the device is only significant when used in conjunction with physician over read as well as consideration of other relevant patient data.
- The device should not be used on infants weighing less than 10kg.
Introduction, Warnings, and Safety
- The Eko DUO System is designed to support healthcare professionals in analyzing cardiac and other internal organ sounds. The Eko DUO System includes a device capturing heart sounds and ECG readings (herein referred to as DUO), a smartphone application (App), and a web application (Dashboard).
- DUO features audio and ECG data transmission via Bluetooth that allows the user to play sounds and visualize phonocardiograms (PCG) and electrocardiograms (ECG) in a mobile application on compatible iOS and Android smartphones and tablets. The app provides the ability to save audio recordings and waveforms (PCG and ECG), share patient recordings with other practitioners, and annotate notes on recorded audio. DUO is intended for use on pediatric and adult patients. The device is not intended for infants weighing less than 10 kg.
- CAUTION: Federal (USA) law restricts this device to sale to or on the order of a licensed medical professional.
For Help and Assistance
For general and product related comments, questions, or concerns, please contact Eko directly.
Serious Incident Reporting
If a serious incident has occurred in relation to the device, it should be
reported to the manufacturer and the competent authority of the Member State
in which the user and/or patient is established. A serious incident means any
incident that directly or indirectly led, might have led or, in case of
recurrence, could lead to any of the following: the death of a patient, user,
or other person, the temporary or permanent serious deterioration of a
patient’s, user’s, fetus or other person’s state of health, or a serious
public health threat.
Manufacturer’s Address:
Eko Health, Inc. 2100 Powell St., Suite 300 Emeryville, CA 94608 USA
General Assistance and FAQs
ekohealth.com/getstarted
Direct Contact
support@ekohealth.com
Phone Support
1.844.356.3384
Warranty Information
Eko provides a limited warranty for DUO. Please visit ekohealth.com/warranty
for a full description of the warranty.
Product Reference and Information
www.ekohealth.com
Safety Related Labels & Symbols
Caution
-
To reduce the risk of device interference, keep DUO at least 2 meters away from all RF emitters including Wifi routers and radios when operating or charging.
-
To reduce the risks associated with infection follow all cleaning and disinfecting instructions included in this manual. Establish and follow a cleaning and disinfecting schedule after each use.
-
To reduce the risks associated with inaccurate data acquisition store and operate this
device only as instructed in this manual. It is recommended that the battery be recharged within thirty minutes of the LED indicator turning red. Recharge the battery using only the wireless charging pad provided with the device. -
DO NOT immerse the device in a liquid or subject it to any sterilization processes other than those described in this manual. The device is non-sterile.
-
To reduce the risks associated with very strong electromagnetic fields avoid using the device near strong radio frequency (RF) signals or portable and/or mobile RF devices. If sudden
or unexpected sounds are heard, move away from any radio transmitting antennas. Using accessories, transducers, and cables not produced by Eko may result in increased RF emissions or decreased immunity. -
Please read, understand, and follow all safety information contained in these instructions prior to using DUO. It is recommended that these instructions be retained for future reference.
-
DUO contains a Bluetooth Class 2 wireless data link. The maximum radio frequency field strength generated by the device is below three volts per meter, a level that is considered safe to use with other medical devices. However, audio, video, and other similar equipment may cause electromagnetic interference. If such devices are encountered and cause interference, immediately move DUO away from that device and/or turn the Bluetooth feature of the interfering device OFF.
-
To reduce the risks associated with environmental contamination follow applicable regulations when disposing of this device. DUO contains a rechargeable battery; please properly dispose of the device as mandated by local directives.
-
No modification of this equipment is allowed. There are no repairable parts inside DUO.
-
Disperse any static electricity before using
the unit. -
Warning: MR-unsafe! Do not expose the device to a magnetic resonance (MR) environment. The device may present a risk of projectile injury due to the presence of ferromagnetic materials that can be attracted by the MR magnet core. Thermal injury and burns may occur due to the metal components of the device that can heat during MR scanning. The device may generate artifacts in the MR image. The device may not function properly due to the strong magnetic and radio frequency fields generated by the MR scanner.
- This device does not detect or measure all heart rate, heart rhythm and heart waveform changes.
- The device has not been tested for use on infants weighing less than 10kgs.
- Conductive parts of electrodes and associated connectors for Type BF Applied Parts, including the neutral electrode, should not contact other conductive parts including earth.
- Do not use DUO while it is charging.
- Do not use on patients with cardiac pacemakers or other electronic implanted devices.
- If used on a portion of the body with significant body fat, body hair, or very dry skin, a successful recording may not be possible.
- Do not use DUO while charging your mobile device, this will cause electrical interference with the ECG signal.
- Do not store DUO in extremely hot, cold, humid, or wet conditions.
- Do not expose to strong electromagnetic fields.
- Do not use as sole basis for medication or treatment decisions.
- Do not take a recording if the electrodes are dirty. Clean them first, according to the cleaning instructions in this User Manual.
- Do not use DUO over broken skin or wound areas.
Skin Preparation
Excessive hair, dirty skin, dry skin, or oily skin can impact the quality of
the ECG tracing. Clean the patient’s skin with water and soap and remove
excess hair as needed to achieve the highest quality tracing. Do not use DUO
over wound areas or areas of broken skin. Do not use an alcohol based skin
cleaner as this dries out the patient’s skin and increases resistance. Rub the
skin vigorously to increase capillary blood flow to the tissues.
ECG gels or saline solutions can be used on the electrodes to improve signal
quality.
EMC Compliance
FCC Intentional Radiator Certification
Eko DUO:
FCC ID: 2ANB3-DUO
IC: 23063-DUO
MIC ID: 020-210076
Wireless Touch Charging Pad:
FCC ID: 2ANB3-W1
This equipment is approved by the FCC under the FCC ID numbers shown above.
This device complies with Part 15 of the FCC rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesirable operation.
NO MODIFICATION: Modifications to this device shall not be made without the written consent of Eko. Unauthorized modifications may void the authority granted under Federal Communications Commission rules permitting the operation of this device.
EMC Compliance Europe
This equipment complies with the EMC requirements of the IEC 60601-1-2.
FCC and Industry Canada Compliance Statement
This device complies with FCC Rules Part 15 and with Industry Canada license- exempt RSS standard(s). Operation is subject to two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference that may be received or that may cause undesired operation.
Cleaning
To ensure a proper ECG and PCG recording, clean the electrodes with an alcohol-based cleaning solution prior to use.
Cleaning and Disinfecting Procedure
The device and earpieces should be cleaned between each patient use. All
cleaning instructions pertaining to stethoscopes and ECG units in general
apply.
Under normal conditions it is unnecessary to separate the device from the
earpieces for cleaning. All external parts of the hardware can be cleaned with
70% isopropyl alcohol wipes.
NOTE: DO NOT immerse the device in any liquid or subject it to any high- pressure/autoclave sterilization processes.
Precautions
- The device is intended to be prescribed by licensed medical professionals. The device may be used on patients during a physical assessment in a clinical setting or by patients with a prescription and under the supervision of a clinician.
- The system provides one source of data that is significant only when used in conjunction with clinician oversight and consideration of other relevant patient information.
- DUO should be used only by qualified clinicians or prescribed to patients with an adequate understanding of the device. DUO is intended for use on patients that can be auscultated normally with an acoustic stethoscope.
- This manual provides instructions for the use of DUO and Eko App. It is assumed that the user is familiar with basic mobile application use on iOS™ and Android™ devices.
- Standard procedures for auscultation should be followed including background noise reduction and optimal patient positioning. The quality of the ECG is dependent on proper preparation practices including, but not limited to, cleaning the contact area and using ECG gel.
- The device uses a Bluetooth Class 2 wireless data link. The Bluetooth range will be reduced when objects (walls, furniture, people, etc) are between DUO and a paired mobile device. To improve Bluetooth connection, reduce the distance and/or allow a line of sight between DUO and the mobile device.
- In order to transmit sounds and record electrical activity to Eko App, DUO and the mobile device must be connected via Bluetooth, and in order to fully use certain functions, the mobile device must be connected to the Internet.
- In addition to security features embedded in the system, it is highly recommended that users of the mobile app and web dashboard use networking security features to protect patient data created and stored using this software.
- Common examples include, strong passwords, biometric authorization, two-factor authentication, and VPN encryption when available.
- DUO is not intended to be used in conjunction with OXYGEN RICH ENVIRONMENTS.
- DUO is not intended for use with flammable anaesthetic.
- DUO is not intended for use with flammable agents.
- NOTICE: Some of the features of Eko App require a minimum Internet connection speed. The minimum recommended upload speed for the mobile app is 4000 Kbps. 4G cellular data service or similar is recommended for the app.
- The app can be used to visualize waveforms and tracings without an Internet connection, however an Internet connection is necessary to save the data.
Patient Privacy
- The privacy of patient health information may be protected by state, federal, or international/foreign laws that regulate how such information can be used, stored, transmitted, and disclosed. The Eko system employs security features that are compliant with HIPAA policies. Third party access may be prohibited to such information without obtaining written authorization from the patient.
- The user is fully responsible for understanding and following all laws that regulate storage, transmission, and disclosure of any electronic patient data through the use of software. If the user becomes unable to comply with a law or restriction that applies to use and disclosure of such data, the user should not proceed to collect or save such information.
- This application may require entry of individually identifiable health information in order to function. Records are stored and recalled through the use of patient name, date of birth, and/or patient ID number. By entering this information, the user assumes any and all risks of and liabilities incurred with using or transmitting such information.
Contents and Operation
This package includes:
- 1 Eko DUO
- 1 Wireless Touch Charging Pad
- 1 5W / 2A USB Power Adapter
- 1 Micro USB to USB Cable
- 1 Eko EARPIECE
- 6 Silicone Rubber Ear Tips
- 1 Getting Started Instruction Card
For full functionality, the system requires users to connect their DUO with an Internet-enabled smart mobile device using Eko App (not included). The app supports many Apple® and Android™ mobile devices.
NOTE: DUO can be used with other audio equipment or headphones through the 3.5mm audio jack. However, no performance guarantees are claimed using other audio products.
System Requirements
For full functionality, the system requires users to connect their Eko DUO
with an internet-enabled smart mobile device using the Eko App. The app
supports Apple® mobile devices. Make sure your system and mobile device meets
or exceeds the minimum performance specifications. Additional information on
the most up-to-date system requirements can be found at
support.ekohealth.com.
NOTICE: Some of the features of the Eko App require a minimum internet
connection speed. The minimum recommended upload speed for the mobile app is
4000 Kbps. A minimum of 4G cellular data service or similar is recommended for
the app. The app can be used to visualize waveforms and tracings without an
internet connection, however an internet connection is necessary to save the
data.
DUO uses Bluetooth® LE; mobile devices used must be compatible with Bluetooth®
LE.
Apple®, iPhone®, iPad®, iPad Air®, and iPad Mini™ are registered trademarks of
Apple, Inc.
Android is a trademark of Google LLC.
Bluetooth® is a registered trademark of Bluetooth SIG, Inc.
Setup
Install included Earpiece by plugging the audio jack into the 3.5mm audio port at the base of DUO and tightening the threads.
Charging
To charge DUO, place the device on the Wireless Touch Charging Pad. Connect
the charging pad to a power source using the included USB charging cable and
power adapter. The lights circling the center button will turn on, indicating
that the device is charging and properly placed. An orange light will also
appear on the changing pad. The lights on DUO will turn off once fully
charged. DUO should be periodically recharged even when in storage. Lithium
ion batteries slowly lose charge when in storage and may fall to an
unacceptably low level, damaging the battery.
NOTE: DUO will not operate or connect to Eko App while charging.
Turning On and Off, Changing Filters, and Taking a Recording
DUO is controlled using the center/action button located on the top face. An LED ring illuminates when the device is powered on.
Power Save Mode
DUO will automatically power off after 5 minutes of continuous inactivity. To
resume DUO, press and hold the center/action button for two seconds.
Volume Control
The device’s sound level can be amplified in 12 increments. Change the volume
level by pressing the top (+) and bottom (-) of the volume button on the side
of the device. Increasing the volume by one level can be confirmed by the
increasing number of LEDs lit in the LED ring.
LED Indicator Lights
Capturing Sounds and ECGs
To capture sounds and ECGs, DUO can be used on various locations and
orientations of the chest. Each position will produce a unique heart sound and
ECG tracing.
Audio
Capture sounds by placing DUO anywhere on the body. For best audio, press the
device firmly against the skin to ensure good contact.
ECG
Capture the ECG signal by placing DUO on the chest. One position that works
well is over the left pectoral muscle, centered below the clavicle and angled
at 45°, as shown in the illustration below. If the patient has particularly
dry skin, significant body fat or chest hair, then the conductive gel used
with other ECG systems may be applied to DUO electrodes to improve the quality
of the ECG signal.
Placement
DUO streams a single lead ECG. Best placement can vary.
ECG electrodes must be placed on the skin.
DUO can also be used to auscultate at all anatomical positions.
Placement
The dots above indicate generally accepted DUO placement positions for
listening to the heart (front) and lungs (back). DUO is sensitive to vibration
and hand noise. Remember to apply firm and constant pressure against the body
to ensure good contact.
Operating Conditions
Environmental
The operating range of DUO is 10° to 45°C (50 to 113°F), and 10% to 95%
relative humidity. The storage and transport range is -10° to 60° C and 10% to
95% relative humidity. Acceptable pressure is 1 atm.
It is recommended to avoid exposure to extreme heat, cold, solvents and oils.
Extreme heat and cold will negatively affect the lithium ion battery in the
device, and may affect battery life.
Operating Warnings
Failure to follow care and maintenance recommendations could result in damage
to the internal components of the device. Internal damage to the product could
cause malfunction of the product, possibly leading to complete loss of
function. If problems are encountered with the device, do not attempt to
repair it. Please notify our support team for assistance.
Eko App Use
Eko App helps you easily connect to DUO for secure transmission and analysis
of your recordings.
Downloading Eko App allows you to:
- Pair DUO to your phone
- Listen wirelessly
- View PCG and ECG visualizations
- Save and share recordings
- Enable AI analysis*
- Take advantage of additional usage guides
- Download Eko App at one of the app stores below:
*available with Pro Plan
Electrical Safety
Guidance and Manufacturer’s Declaration – Electromagnetic Emission
DUO is intended for use in the electromagnetic environment specified below.
The user of DUO should assure that it is used in such an environment.
Warning: The use of accessories other than those specified, with the
exception of accessories sold by Eko as replacement parts, may result in
increased emissions or decreased immunity of DUO.
Warning: DUO should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, DUO should be observed to
verify normal operation in the configuration in which it will be used.
Electrical Safety
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
DUO is intended for use in the electromagnetic environment specified below.
The user of the Eko DUO should assure that it is used in such an environment.
NOTE 1 – At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 – These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and is affected by absorption and
reflection from structures, objects and people
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To address the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which DUO is used exceeds the
applicable RF compliance level above, DUO should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating DUO.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Electrical Safety
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Eko Electronic Stethoscope System DUO is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The user of DUO can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and DUO as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d is meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Manufacturing and Regulatory Information
Eko Health, Inc.
2100 Powell St., Suite 300 Emeryville, CA 94608 USA
ekohealth.com
Emergo Europe Westervoortsedijk 60
6827 AT, Arnhem
The Netherlands
Australia Sponser:
Emergo Australia
201 Sussex Street.
Darling Park, Tower II, Level 20 Sydney NSW 2000,
Australia
MedEnvoy Switzerland Gotthardstrasse 28 6302 Zug Switzerland
©2023 Eko Health, Inc. All rights reserved
Eko, the Eko logo, and DUO are trademarks of Eko. Other company and product
names may be trademarks of their respective owners.
LBL215 Rev 2.0 – Date of Issue: June 2023
References