SUN23-III Portable Vacuum Autoclave Steam Sterilizer
Specifications
- Inner chamber of the autoclave (Diameter X Depth) 18L: 245mm X 355mm
- Inner chamber of the autoclave (Diameter X Depth) 22L: 245mm X 465mm
- Rated Power: 1800VA, 2000VA (Under Standard Atmospheric pressure)
- Fuse-Melting pipe: T15A, T30A
- Capacity of water tank: 4.5L
Product Usage Instructions
Installation
The autoclave should be installed on a horizontal working base with a proper air-flowing environment. Ensure the clearance around the autoclave is at least 10cm and the top surface is 20cm clear. The air-cooling window must not be clogged.
Preparation Before Operation
Connect the autoclave to the power source by plugging it into the power source hole. Press the green switch on the front side to initialize the autoclave. When the indicator light is on and LD is displayed on the screen, the autoclave is ready for operation.
Display and Keypad Description
The display screen shows real-time sterilizing status like system pressure and temperature. The keypad includes function keys for entering advanced settings, arrow keys for parameter adjustment, and a START/STOP key for program confirmation.
Sterilizing Status Settings
To start a program, press Setting for 6 seconds to enter the program menu, choose the program, confirm it with START/STOP, close the door, and press the Start/stop button to begin working. Date and time adjustments can be made in the program menu.
FAQ
- Q: What is the application scope of the autoclave?
- A: The machine can be used for sterilizing tooth devices, operation devices, sanitary materials, and other heat-resisting articles in departments like Stomatology, Ophthalmology, and labs.
- Q: How should I adjust the height of the autoclave?
- A: The front feet of the autoclave can be adjusted to ensure the front of the machine is slightly higher than the back when placed on a horizontal working base.
- Q: How do I know when the autoclave is ready for operation?
- A: Connect the autoclave to the power source, press the green switch, and wait for the indicator light to turn on while displaying LD on the screen. This indicates that the autoclave is initialized and ready for use.
General Introduction
This Autoclave is a precision instrument specializing in the operation
occasion in need of simple, fast, and efficient sterilization by the doctor or
his/her assistants, and the doctor shall take responsibility for such work.
This Autoclave is specially designed for the occasions in use of frequent
sterilizations. It is equipped with a microprocessor for intelligent control
and a man-machine interface for convenient operation. With the distinctions in
displaying parameters and conditions actively, evaluating malfunction
automatically, and shutting off for self-protection automatically in case of
excessive temperature and pressure, this type of autoclave is a reliable
product for sterilizing.
- Power source socket
- Fuse
- Exit of source water
- Exit of wastewater
Application Scope
This machine can be applied in sterilizing tooth devices, operation devices,
Sanitary Materials and other heat-resisting articles existing in the Dept. of
Stomatology, Ophthalmology, and lab.
Tech. Parameter
- Inner chamber of the autoclave (Diameter X Depth) 18L……245mmX355mm
- Inner chamber of the autoclave (Diameter X Depth) 22L……245mmX465mm
- Rated Voltage………………………………………220V±22V, 50Hz±1Hz ………………………………………110V±11V, 60Hz±1Hz
- Rated Power………………………………………1800VA 2000VA
- Sterilizing (Pressure/temp.)…………………………………90-130KPa/121ºC 200-230KPa/134ºC (Both of the above are Under Standard Atmospheric pressure)
- Fuse-Melting pipe……………………….. ……. T15A T30A
- Capacity of water tank………………………………………………………4.5L
- Water Consumption volume per recycling round…………………….. .0.3-0.5L
- Operating Temp…………………………………………………………0-40ºC
Diagram Specification
- Attention and warning
- I Sort B model device
- Protective Grounding
Installation
- The clearance between the outside of the autoclave and the around articles shall be 10cm as retained, and the top surface shall be 20cm retained.
- The air-cooling window attached to the exterior surface of the autoclave shall be non-clogged by dust or articles, and the autoclave shall work in the proper air-flowing environment.
- The autoclave shall be installed on the horizontal working base.
- The 2 front feet of the autoclave can be adjusted the height, please ensure the front of the machine is slightly higher than the back.
PREPARATION BEFORE OPERATION
Before operation, please connect well to the power source. Connect the power source hole under the back left side of the autoclave and the power socket by the attached plug. Press the green switch right bottom corner of the front side, when the indicator light is on, it means that the autoclave gets through with the electric power, and the process then is in initialization status and shows “LD” on the screen. At this time the autoclave is not heating.
Display and keypad description
display screen
Real-time displaying the sterilizing status like system pressure and
Temperature.
SETTING function keys
Used to enter the advanced settings status
▲UP key
When in the program selection menu or the Advanced Settings menu, the up arrow
is used to indicate movement and parameter values increase.
▼DOWN key
When in the program selection menu or the Advanced Settings menu, the down
arrow is used to indicate movement and parameter values decrease.
START/STOP confirmation key
Identify programs and quite the program settings.
Sterilizing status settings
Choose the program
After turning on the power, the machine Displays the Load, here keep Press
“Setting” for 6 seconds to enter the program menu, use”▲””▼” to choose the
program, and press START/STOP to confirm it, after closing the door, the
machine would display “Ready”, then press “Start/stop” button to start
working!
Adjust the date and time
- In the program menu, choose “Set up” to adjust the time and printer working status.
- Adjust the sterilizing time and drying time, choose the system as the right Photo, and press Start/stop to confirm your choice! Press “Start/stop” to quiet the setting!
Fill & drain water
When you install the autoclave for the first time, after turning on the
switch, then The machine would displaying a lack of source water. Then you
need to use the measuring cup to fill distilled water from the top say around
3000ml for the first time, next time do not fill the water too full, only near
the red mark. When the machine displays the used water full, please drain
waster water out in time from the front connector. Please clean the brass
filter inside the chamber, or it will affect the pressure release during the
sterilization. Please clean the filter together by draining the used water.
Note: Be sure to fill distilled water
- Before filling the distilled water, you must turn on the switch of power!
- Please fill the distilled water only after finish working!
Function description
Vacuum sterilizer is divided into the following processes:
- preheating process;
- the pre-vacuum and pressure-increasing process;
- sterilizing process;
- pressure release and vacuum drying process.
- pressure equilibrium process:
When the drying time is reduced to zero, turn off the exhaust, only the pressure displaying 0 and the machine displaying END, means the machine has finished the sterilizing, please open the door and take the sterilized instrument out by the handle.
Sterilizing data of each program
1 | 2 | 3 | 4 | 5 | 6 | 7 | |
---|---|---|---|---|---|---|---|
134℃ |
UNWRAPPED
| 121℃
UNWRAPPED
| 134℃
WRAPPED
| 121℃
WRAPPED
| 134℃
COTTON
| 121℃
COTTON
| 121℃
PLASTIC
Sterilizing
temperature
| 134| 121| 134| 121| 134| 121| 121
Sterilizing
pressure
| 2.1| 1.1| 2.1| 1.1| 2.1| 1.1| 1.1
Vacuum times| 1| 1| 3| 3| 3| 3| 1
Sterilizing
time
| 4| 20| 4| 20| 20| 20| 20
Drying time| 12| 12| 18| 18| 20| 20| 18
Emergency Condition
- During the working time, the autoclave will automatically alarm, then release pressure, stop heating, and display the error code (The definition of the error code is as the below table: error code and resolving method ). It would ensure the safety of the operator if an emergency occur.
- If any emergency happens, please check the warning code table to resolve the malfunctions. If not possible to resolve, please contact the dealers. We will try our best to provide help to you in a short time.
Error code and resolving measures
NO| Code
display
| Alarm Tone| Malfunction content| Resolving Measures
---|---|---|---|---
1| E1| long “du”| Fault of temperature sensor inside the chamber| Inspect if
anything affected the sensor or wire not connecting well
or the sensor was broken
2| E2| long “du”| Pressure over 240 KPa| Inspect the pressure sensor or vacuum
pump working
3| E3| long “du”| Fault of temperature sensor outside
the chamber
| Inspect the temperature sensor
outside of the chamber
4| E4| long “du”| The super high inner temperature when
pressure rising
| Inspect the inside temperature sensor
or pressure sensor
5| E5| long “du”| Pressure releasing slowly| Inspect the pressure-releasing valve
(the right top one)
6| E6| long “du”| The problem of the door switch or door being opened when working| Inspect door switch or door handle were not revolved to the
right MAX position
7| E7| long “du”| Fault of temperature sensor on steam
generator
| Inspecting the temperature sensor
on steam generator
8| E8| long “du”| Fail on pressure rising| Inspect water flowing road and
steam leaking.
9| EH| long “du”| The steam generator not working| Inspect connecting wire or
resistance of the steam generator
10| EF| long “du”| Pressure not over than 0 KPa| Inspect the water pump/steam
generator
11| EL| long “du”| Pressure not over 20 KPa| Inspect if the electric valve not
working and caused steam leaking
12| Eo| long “du”| Pressure not over 65 KPa| Inspect if the water pump working
weakly.
13| E9| long “du”| Fail to keep pressure and temperature| Inspect the steam
leaking
Maintenance and Service
-
Clean the top water tank with alcohol once a month.
-
Disinfect and clean the interior surface of the autoclave chamber monthly.
-
Replace the fuse.
-
Shut off the power.
-
Revolve the fuse base counter-clockwise by a screwdriver to bring out the melting fuse to be replaced;
-
Replace the new fuse and reset the fuse base on the original position, then use the screwdriver to fasten the base clockwise.
-
Check the correctness of the data of the new fuse replaced.
Clean the sealing ring periodically
The operator shall clean the sealing ring periodically to avoid the influence
of the seal caused by the dust/dirt left due to the long period of use. Use
the smooth cloth with distilled water to wipe the surface of the sealing ring
or sealing cap gently. The user shall discharge the sealing ring for further
cleaning or replacement if the air-leaking problem can not be resolved after
the above process (The user shall discharge the sealing ring periodically for
cleaning and inspection as recommended by the supplier).
The replacement work for sealing ring
Tool: One flat screwdriver (No sharp edge at the head)
- One hand catches the lip of the sealing ring, the other hand inserts the screwdriver into the clearance between the sealing ring and the door to raise the sealing ring.
- After raising part of the ring, you can use your hand to draw the whole ring out. Wash its groove after drawing out the ring and pay attention to see whether it’s spoiled or not in consideration of the necessity of replacement.
- Put back the ring to the original door groove after cleaning. Most Importantly: the inset work must be done equably to the groove During the installation, the four equal points of the ring must be inserted to the groove first, later for the same work for the remaining segments of the ring. After finishing, press the sealing ring equally with the force of your hand.
- Note: The inner circle of the ring may be raised when inserting the ring into the groove. Please press it to the groove by using the screwdriver carefully. Shut off the power and cool the autoclave sufficiently before inserting the ring to avoid scald.
- Shut off the power before service or components’ replacement, and the service or replacement work shall be done by the technician.
Transportation and Storage
-
The preparatory work before transportation and storage: Shut off the power, draw out the plug, and cool the autoclave down.
-
Draining
Empty the water storing tank and the cooling water collecting tank; The terminal side of the attached pipe with no connecting joint shall be inserted into the water exiting joint pipe. As the above picture shows, B is the water exiting the mouth of the cooling water collecting tank, A is the water exiting the mouth of the water storing tank. Revolving the water exit button in an anticlockwise direction to release the water. -
The autoclave shall be transferred and stored with the below conditions:
- Ambiance temperature: -40ºC -+55ºC
- Relative humidity: < 85%
- Atmospheric pressure: 500HPA-1060HPA
Note: Don’t drag or drop during transportation.
Attention
- The autoclave must be placed on the horizontal working base.
- Must use distilled water to last the operation life of the autoclave.
- There shall be no clog to the air cooling window on the outer surface of the autoclave.
- The object apparatus to be sterilized shall be placed on the device plate with enough clearance retained in favor of the air circulation inside the sterilizing room.
- Empty the water inside the cooling water collecting tank frequently. Usually, the cooling water collecting tank shall be emptied once the water storing tank is emptied.
- The door handle must be buttoned up before working.
- Don’t open the door of the autoclave before seeing the “0” bar displayed on the pressure indicator.
- Not be too close to the door of the autoclave in case of scald when opening the door.
- Shut off the power before discharging/installing the sealing ring, and the work shall be carried out after further sufficient cooling in case of scald.
- Don’t drag or drop the autoclave during the relocation period of the autoclave.
- The protecting ground shall be reliable.
- Equipment must be used away from the magnetic field.
- The device’s end of life must be for processing scrapped under the local environmental protection laws and regulations.
- Such as power failure occurs during the working, needs to be disinfected the equipment out, it must be open the door 2 minutes later.
Accessory
1. | Draining pipe | 1 pc | **** |
---|---|---|---|
2. | Sterilizing Tray | 3 pcs | |
3. | Plug | 1 pc | |
4. | Sterilizing shelf | 1 pc | |
5. | Handle | 1 pc | |
6. | Fuse | 2 pcs | |
7. | Operation manual | 1 pc | |
8. | Measuring cup | 1 pc | |
9. | Sealing gasket | 1 pc | |
10. | O ring of tap | 2 pcs |
Appendix
- Steam loop(Work Description)
- Circuit diagram of sterilizer
- BD test method Press the “UP / DOWN” button to choose the BD test system before sterilizing, close the door, and then press the “START/STOP” button to start.