HACH FH950 Handheld Flow Meters
Specifications
Sensor specifications
The sensor specifications for the FH950 are as follows:
- Measurement Range: [Specify measurement range]
- Accuracy: [Specify accuracy]
- Resolution: [Specify resolution]
Portable meter specifications
The portable meter specifications for the FH950 are as follows:
- Display: [Specify display type and size]
- Power Supply: [Specify power supply requirements]
- Operating Temperature: [Specify operating temperature range]
User interface specifications
The user interface specifications for the FH950 are as follows:
- Screen: [Specify screen type and size]
- Buttons: [Specify button functions]
- Menu Navigation: [Specify menu navigation options]
General specifications
The general specifications for the FH950 are as follows:
- Dimensions: [Specify dimensions]
- Weight: [Specify weight]
- Protection Rating: [Specify protection rating]
Product Information
Safety Information
It is important to follow these safety guidelines when using the FH950:
- Always read and understand the user manual before using the product.
- Use the hazard information provided to identify and minimize risks.
- Pay attention to precautionary labels on the product.
- Ensure the product has the necessary certifications for safe operation.
Product Overview
The FH950 is a portable meter designed for [Specify purpose]. It consists
of the following components:
- System Overview: [Describe the overall system and its components]
- Sensor Overview: [Describe the sensor and its features]
- Meter Overview: [Describe the meter and its features]
Product Components
The FH950 comes with the following components:
- [Specify component 1]
- [Specify component 2]
- [Specify component 3]
Installation
[Provide detailed installation instructions for the FH950]
User Interface and Navigation
-
Keypad and Key Functions
[Explain the functions of each keypad button] -
Status Bar
[Describe the information displayed on the status bar] -
Navigation and Main Menu
[Guide the user through navigating the menu and accessing different functions] -
Startup and Self-Test
-
Sleep Mode
[Explain how to activate and exit sleep mode] -
Operation
[Provide instructions on how to operate the FH950] -
Download Data
[Explain how to download data from the FH950] -
Delete Data Files
[Describe the process of deleting data files from the FH950] -
Maintenance
[Provide maintenance instructions for the FH950]
FAQ
Q: How do I calibrate the sensor?
A: To calibrate the sensor, follow these steps:
- Step 1: [Describe the first calibration step]
- Step 2: [Describe the second calibration step]
- Step 3: [Describe the third calibration step]
Q: Can I use the FH950 underwater?
A: No, the FH950 is not designed for underwater use. It is only suitable
for [Specify suitable environments].
Section 1 Specifications
Specifications are subject to change without notice.
Sensor specifications
Specification | Details |
---|
Velocity measurement
Method| Electromagnetic
Range| 0 to 6.09 m/s (0 to 20 ft/s)
Minimum water depth| 3.18 cm (1.25 in.)
Accuracy| ±2% of reading ±0.015 m/s (±0.05 ft/s) 0 to 3.04 m/s (0 to 10 ft/s);
± 4% of reading from
3.04 to 4.87 m/s (10 to 16 ft/s)
Resolution| 0.01 value < 100; 0.1 value < 1000; 1.0 value ≥ 1000
Zero stability| ±0.015 m/s (±0.05 ft/s)
Material| ABS, glass filled
Enclosure rating| IP68
Dimensions (L x W x H)| 11.9 x 4.3 x 6.3 cm (4.7 x 1.7 x 2.5 in.)
Cable material| Polyurethane jacketed
Cable lengths| 1.5, 6.1, 12.2 and 30.5 m (5, 20, 40 and 100 ft)
Depth measurement
Method| Diaphragm type: absolute pressure with single point calibration
Accuracy (static)| The larger of ± 2% of reading or ± 0.015 m (± 0.504
inches). Steady state temperature and static non-flowing water.
Range| 3.05 m (0-10 ft)
Resolution| 0.01 value < 100; 0.1 value < 1000; 1.0 value ≥ 1000
Portable meter specifications
Specification | Details |
---|---|
Pollution degree | 2 |
Protection class | II |
Charging temperature | 0 to 40 ºC (32 to 104 ºF) |
Operating temperature | –20 to 55 ºC (-4 to 131 ºF) |
Storage temperature | –20 to 60 ºC (-4 to 140 ºF ) |
Enclosure rating | IP67 |
Battery life gauge | Five-segment bar graph |
Battery type | Rechargeable lithium ion, 3.6 V, 5.0 Ah, 18 Wh |
Battery life | 18 hours heavy typical day use1; 20 ºC (68 ºF) |
1 Defined as 30 minutes of set up, six 1-hour periods of continuous use with an active sensor and the display at maximum brightness and 30 minutes of sleep mode between use periods, data download and power off.
Specification | Details |
---|---|
Battery charger | External power adapter, energy efficiency VI, 100–240 VAC, |
50–60 Hz, 0.5 A input; 12 VDC, 1.25 A output
Dimensions (L x W x H)| 21.8 x 9.3 x 5.3 cm (8.6 x 3.7 x 2.1 in.)
USB connector| Type Mini-B, 5-pin, rated to IP67 when capped
Material| Polycarbonate with a thermoplastic elastomer (TPE) overmold
User interface specifications
Specification | Details |
---|---|
Graphics display | Color, LCD 3.5″ QVGA transflective (readable in direct |
sunlight)
Measurement resolution| 0.01 value < 100; 0.1 value < 1000; 1.0 value ≥ 1000
Keypad| Alpha-numeric
Operating modes| Real time, profiling
Profile types| Stream, conduit
Conduit shapes| Circular, rectangular, trapezoidal, 2/3 egg, inverted 2/3 egg
Stream entries| Fixed, non-fixed stations
Noise rejection| User-selectable, 50 Hz or 60 Hz
Units of measure| Velocity: ft/s, m/s, cm/s, mm/s
Flow: ft3/sec, million gal/day, gal/day, gal/min, m3/s, m3/min, m3/hour,
m3/day, liters/s, liters/min
Depth: in., ft, m, cm, mm
Stream flow calculation| Mean-section or mid-section method
Diagnostics| Self test, keypad, display, event log
Conduit profiling methods| 0.9 x Vmax, 0.2/0.4/0.8, velocity and level
integrator, 2D
Stream profiling methods| 1, 2, 3, 4, 5 and 6 point (velocity method – USGS
and ISO), 2 point KREPS method, 1 point surface method, 1 and 2 point ice
method
File types| Real-time, profiling, event log
Languages| English, French, Spanish, German, Italian, Dutch, Danish, Swedish,
Chinese, Polish, Japanese, Korean, Portuguese, Slovak, Russian, Hungarian,
Bulgarian, Romanian, Czech, Turkish, Finnish, Greek
General specifications
Specification | Details |
---|---|
Maximum number of real-time files | 50 each with up to 75 readings captured by |
the user.
Firmware| The sensor and portable meter are field upgradeable via USB
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical
application risks and install appropriate mechanisms to protect processes
during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this
equipment. Pay attention to all danger and caution statements. Failure to do
so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do
not use or install this equipment in any manner other than that specified in
this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage
to the instrument could occur if not observed. A symbol on the instrument is
referenced in the manual with a precautionary statement.
Certification
C A U T I O N
This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
- Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class A: Supporting test records reside with the manufacturer.
- This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
- Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class “A” Limits
Supporting test records reside with the manufacturer. The device complies with
Part 15 of the FCC Rules. Operation is subject to the following conditions:
- The equipment may not cause harmful interference.
- The equipment must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
- Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
- If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
- Move the equipment away from the device receiving the interference.
- Reposition the receiving antenna for the device receiving the interference.
- Try combinations of the above.
Product overview
The portable velocity system is used in the field, laboratory and
municipalities. Turbulent, noisy and low flows can be measured with this
system.
The meter and sensor get velocity information in conduits and streams. These
measurements are important for calibration in municipal wastewater industries,
as well as for environmental-impact evaluations.
Two types of sensor are available: velocity-only and velocity plus depth. This
manual covers both types of sensors. If information applies to a specified
type of sensor, this fact is noted in the text.
System overview
An overview of an assembled system is shown in Figure 1. Refer to the
documentation supplied with the individual components or accessories for more
information.
Figure 1 Assembled components
1 | Portable meter | 5 | Adjustable mount for portable meter |
---|---|---|---|
2 | Sensor height lock/release device | 6 | Sensor assembly |
3 | Top setting wading rod (accessory) | 7 | Flow direction |
4 | Sensor cable |
Sensor overview
Figure 2 shows the main sensor components. Instructions for how to attach the
sensor on a standard or top-setting wading rod are supplied with the
accessory.
Figure 2 Sensor components
1 | Sensor electrodes | 5 | Pressure cell (sensors with depth option) |
---|---|---|---|
2 | Sensor body | 6 | Sensor with depth option |
3 | Sensor connection plug | 7 | Sensor without depth option |
4 | Sensor attachment thumb screw |
Meter overview
Figure 3 shows the features of the meter.
1 USB connection port2 | 6 Threaded hole for adjustable meter mount |
---|---|
2 Keypad | 7 Slots for velcro or strap attachment |
3 Meter display | 8 Slot for neck strap attachment threads (2x) |
4 Expansion port (not used) | 9 Sensor connection port |
5 Battery compartment cover | 10 Battery charger connector |
Product components
When purchasing a complete system, refer to Figure 4 to make sure that all
components have been received. If any of these items are missing or damaged,
contact the manufacturer or a sales representative immediately.
1 Carrying case (with slot for sensor cable)| 8 Universal sensor
mount
---|---
2 USB cable with mini-USB connector, 1 m (3 ft)| 9 Velcro strap
3 Adjustable portable meter mount| 10 Lanyard
4 Battery charger, 100–240 VAC| 11 Portable meter
5 Sensor| 12 Lithium battery pack
6 Extra thumb screws (4x)| 13 Sensor as connected to meter inside
case
7 Cloth to dry the sensor
Section 3 Installation
Installation with optional accessories
Mount the meter on an optional wading rod for use in low-stage stream
environments where the stream can be waded. Optional accessories let the user
take measurements from a bridge or cable over a stream. A torpedo shaped
weight attached below the sensor keeps the sensor in place when under water.
For more information, refer to the documentation supplied with the accessory.
Install the sensor on the universal sensor mount
Use the universal sensor mount to attach the sensor to poles 1 inch or less in
diameter. For correct operation and accurate readings, the front of the sensor
must be pointed upstream with the electrodes in full contact with the flow.
Note : Keep the sensor electrodes free from nonconductive substances such as oil and grease. To remove sensor contamination, refer to Clean the sensor on page 28.
- The front part of the sensor is round and contains three electrodes. The sensor has a mounting hole in back and a thumbscrew on top. Put the mounting shaft of the universal mount in the mounting hole at the back of the sensor. Make sure that the mounting shaft is completely engaged with the mounting hole and the thumbscrew is engaged with the groove.
- Hand tighten the thumbscrew.
- Move a pole 1 inch or less in diameter through the clamp of the universal sensor mount. Tighten the clamp.
Note : Instructions for how to mount the sensor on a standard or top setting wading rod are supplied with the accessory.
Lithium battery safety
Fire and explosion hazard. Lithium batteries may get hot, explode or ignite
and cause serious injury if exposed to abuse conditions.
- Do not use the battery if there is visible damage.
- Do not use the battery after strong shock or vibration occurs.
- Do not expose the battery to fire.
- Keep the battery at temperatures less than 60 ºC (140 ºF).
- Keep the battery dry and away from water.
- Prevent contact between the positive and negative battery terminals.
- Do not let unauthorized persons touch the battery.
- Discard the battery in accordance with local, regional and national regulations.
- Do not use or store the instrument in direct sunlight, near a heat source or in high temperature environments such as a closed vehicle in direct sunlight.
Install the battery
W A R N I N G
Fire and explosion hazard. This equipment contains a high energy lithium
battery which can ignite and cause fire or explosion, even without power. To
maintain the safety provided by the instrument enclosure, the instrument
enclosure covers must be installed and secured with the supplied hardware.
W A R N I N G
Explosion and fire hazard. Battery substitution is not permitted. Use only
batteries that are supplied by the instrument manufacturer.
Install the supplied lithium battery as shown in Figure 5.
Charge the battery
Note : Battery charger substitution is not permitted. Use only the
charger specified in the list of parts and accessories for the instrument.
Refer to Replacement parts and accessories on page 30.
Before initial use, fully charge the instrument battery with the supplied
battery charger. A discharged battery is fully charged in approximately 8
hours.
A full battery charge will supply power to the system for approximately 10–11
hours with constant use. When the level of battery charge drops to 3.4 V or
less, the display shows a warning and the meter automatically powers off. The
battery must be charged before the unit becomes functional again.
- Make sure that the correct type of plug for the geographic location is installed on the battery charger.
- Connect the battery charger to an electrical outlet with earth ground.
- Connect the battery charger to the battery charger connector on the instrument. Refer to Meter overview on page 8. The battery level shows on the display.
A blue light shows around the charge port while the battery charges. When the charge process is complete, the blue light goes off.
Note : The meter is not operational while the battery charges. The battery does not charge through the USB cable connection.
Connect the sensor to the meter
Attach the lanyard
Attach the lanyard to wear the meter safely around the neck.
Attach the velcro strap
Use the velcro strap to hold the extra cable. Refer to Figure 6.
Section 4 User interface and navigation
Keypad and key functions
Figure 7 shows the meter keypad. Table 1 gives the functions of each key or
key type.
Figure 7 Keypad
1 Power On/Off | 6 Main Menu |
---|---|
2 OK | 7 Underscore or decimal |
3 Up and Down arrows | 8 Backspace |
4 Quick Jump | 9 Alpha-numeric |
5 Right and Left arrows | 10 Previous menu |
Table 1 Key description
Key | Description |
---|---|
Power On/Off | Energizes and de-energizes the meter. |
OK | Confirms an entry or highlighted menu option. |
Up and Down arrows | Moves up or down in the display. If the cursor is at the |
top or bottom of the display, the cursor wraps to the bottom or top when the
UP or DOWN arrow is pushed.
Quick Jump| In normal operation, this key jumps to the Select conduit shape
screen. In profiler operation, the Quick Jump key is disabled.
If the auto-zero feature is disabled, hold this key for five seconds to do a manual zero of the depth sensor.
In Real-Time mode, the Quick Jump key toggles between the digital and graph views.
Right and Left arrows| Moves to the right or left in the display.
Main Menu| Moves to the Main Menu from any submenu or screen.
Underscore or decimal| Puts in an underscore or decimal character. In numeric-
only fields, this key automatically puts a decimal point in the cursor
position.
Backspace| Moves the cursor back one space.
Alpha-numeric| Puts in the key alpha or numeric value. Values are put in the
order shown on the key. After 2 seconds, the value shown in the display is
stored and the cursor advances.
Previous menu| Moves to the previous screen.
Status bar
A status bar is shown in the top of the display. Descriptions of the
information in the status bar are given in Table 2.
Table 2 Status bar indicators
Indicator | Description |
---|---|
Time and Date | Shows the current time and date. |
USB | Shows when a USB cable is connected. If a USB cable is connected and this |
indicator does not show in the status bar, the USB cable is not fully engaged.
Make sure that the USB cable is pushed in completely and makes full contact
with the connection port.
Conductivity| If the sensor is out of the water and non-conductive, a blue
ring appears next to the battery icon. If the sensor is in the water and
conductive, the indicator is a solid blue circle.
Battery| A five-bar graph shows the level of charge in the battery.
File access| Shows while the meter gets access to a file.
Auto zero depth indicator| If the depth sensor was zeroed in the last 30
minutes, a solid green circle shows next to the Conductivity indicator. If the
depth sensor was not zeroed in the last 30 minutes, this indicator flashes
red.
Navigation and Main Menu
Push OK to confirm a selected menu option or a value shown in the display.
Select More and push OK to see additional screens and options if available.
Push the Main Menu button to go to the Main Menu from a submenu.
Note : Some operations cannot be completed unless a sensor is connected to the meter. If these operations are tried when there is no sensor connected, the display shows an error message. Connect a sensor and try the operation again.
- Real time—Select this option to get real-time velocity and depth information. (A sensor with depth capability is necessary to read depth). An example of a Real Time screen for sensors with velocity only is shown in Figure 8. Real time screens for sensors with both velocity and depth is shown in Figure 9. The format of the information and options shown depends on the type of sensor used. In Real Time mode, the Quick Jump key toggles between digital and graphic views of Real Time information. The velocity is updated in FPA filter mode according to fixed period averaging time. In RC filter mode, the velocity is updated continuously on the screen every 250 ms.
Profiler— Select this option to do stream or conduit velocity measurements. The meter shows prompts when user input is necessary. A percentage of the remaining memory is given in 1%resolution. Refer to the Appendix on page 32 for more information about profiles.
Set up—Select this option to change general system settings and preferences.
Diagnostics — Select this option to troubleshoot problems with the meter or an attached sensor. For more information about the Diagnostics options, refer to Diagnostics on page 29.
Section 5 Startup and self-test
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a
treatment process and/or chemical feed system for which there are regulatory
limits and monitoring requirements related to public health, public safety,
food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to
have sufficient and appropriate mechanisms in place for compliance with
applicable regulations in the event of malfunction of the instrument.
W A R N I N G
Fire and explosion hazards. Do not use or store the instrument in direct
sunlight, near a heat source or in high temperature environments such as a
closed vehicle in direct sunlight. Failure to take this precaution can make
the battery overheat and cause a fire or explosion.
The battery must be installed in the meter and charged before use. For more information about battery installation and replacement, refer to Install the battery on page 10. For information on how to charge the battery, refer to Charge the battery on page 11.
Note : The meter is not operational while the battery charges.
-
Push the meter power button until an audible beep is heard.
The meter does a self test and the display shows the results. If the meter fails the self-test, the display shows FAIL next to the failed parameter. If the sensor fails, attach a different sensor if available. -
When the self test is complete, push OK to go to the Main Menu.
-
To de-energize the meter, push the power button again. In the Confirmation screen, select Yes and push OK.
If the portable meter becomes unresponsive, push and hold the power button for more than 3 seconds to force the power off. Do not force off the power in normal operation or when the file access icon is visible.
Sleep mode
The meter backlight goes dim after 30 seconds of no activity and goes into
sleep mode after
60 seconds of no activity. These actions do not occur if the meter is in real-
time mode or while the meter is measuring. After 30 minutes in sleep mode, the
meter power goes off.
To cancel the sleep mode, push any key. The display brightness goes back to
the normal level and all keys go back to their normal functions.
Section 6 Operation
Stream profiles
Stations and station spacing
For a well-chosen cross-section, division into 25 to 30 partial sections is
typically sufficient. If the cross-section is very smooth and the velocity
distribution very consistent, it is possible to decrease the number of
stations.
Make the distance between the partial stations so that no individual station
contains more than 10%of the discharge. The ideal measurement is one in which
each partial station contains 5% or less (≤ 5%) of the total discharge, but
this is rarely possible when 25 stations are used. Partial stations should not
have equal widths across the entire cross-section unless the discharge is
well-distributed.
Distances between stations are generally smaller where water depth and flow
velocities change significantly. Places where depth and velocities frequently
change significantly include bank areas, vertical or steep slopes, ledges in
divided cross-sections and transitions from the main stream bed to the
foreland. Stations should also be located at points of significant changes in
the stream bed profile.
The measurement cross-section must be set at right angles to the direction of
flow. Cross-sections must not contain still areas, counter currents or eddies.
Do not put the sensor in deep pools, below large inflows, or near ship
moorings, ferries or sluices.
Use Table 3 as a guide for the number of stations necessary for an acceptable
measurement. The information is based on EN – ISO 748 standards.
Table 3 Number of stations in relation to the waterway width
Feet | Meters | Number of stations |
---|---|---|
< 1.6 | < 0.5 | 5 to 6 |
> 1.6 and < 3.3 | > 0.5 and < 1 | 6 to 7 |
> 3.3 and < 9.8 | > 1 and < 3 | 7 to 12 |
> 9.8 and < 16.4 | > 3 and < 5 | 13 to 16 |
> 16.4 | ≥ 5 | ≥ 22 |
Measure velocity
Measurement quality is dependent on the correct selection of a measurement
cross-section. Select a section of stream with the following characteristics:
- The flow directions at each measurement point across the stream are parallel to the bank and perpendicular to the cross-section.
- The streambed is stable and free of large rocks, weeds and protruding obstructions such as piers that cause turbulence.
It is often not possible to completely satisfy all of these conditions. Use
the criteria to select the best possible section and then select a cross-
section.
The general procedure to take velocity measurements in river and stream
profiles is described below. Make the first measurement in a stream profile at
the top or bottom. Make each subsequent measurement below the last one.
- In fixed mode, divide the channel into stations of equal width.
- Conduct a velocity measurement at each station. The portable meter shows and stores the depth and measured velocity information.
- When the stream profile is completed, the meter automatically calculates the total flow.
- For accurate measurement results, stand to the side of the instrument. Refer to Figure 10.
Measure
velocities in a cross-section
A typical stream cross-section is shown in Figure 11.
To measure velocities in a cross-section:
-
In the Main Menu, select Profiler.
-
A list of options will show.
Option Description -
Select Setup > Maximum depth. Select Manual or Automatic. If Automatic is selected, enter the distance from the bottom of the channel to the bottom of the sensor mount (Offset). The Offset distance value can be different based on the used wading rod and the properties of the bottom. Refer to Figure 12.
-
If necessary, change or update the filter parameters in Profiler Setup.
-
In the Profiler menu, select Stream.
-
Enter the operator name. Make operator names alpha-numeric with a maximum of 23 characters. Push OK to save the operator name.
-
Enter the name of the new stream profile. Make profile names alpha-numeric with a maximum of 23 characters. Push OK to save the profile name or select Clear to delete all current stream data.
-
In the Station menu, select Edge/Obstruction. Select one of the options.
Option Description -
Select Distance to Vertical and enter the information.
-
Select Set Depth and enter the information. If at an edge, the meter automatically sets this value to 0.00.
- If Manual mode was selected in the Profiler setup, enter the total depth of water at this vertical position.
- If Automatic mode was selected in the Profiler setup, put the probe at the bottom and push OK to set the maximum flow depth at the value shown.
Note : The maximum depth of a vertical probe is the measured inmersion plus the offset.
-
If Left or Right was selected in Step 8, enter an edge factor for the vertical. Select a factor from the list or User-defined. For User-defined values, enter a roughness factor between 0.50 (very rough) and 1.00 (smooth). The roughness factor is relevant only for right angled cross sections. It is used as a factor in the calculation of the discharge proportion of edge areas. For example:
- Smooth edge with no vegetation (e.g., concrete, steel, cement)— 0.8 to 0.9
- Brick sides with vegetation— 0.7
- Rough walls with heavy vegetation—0.6 to 0.5
-
Enter the stage reference. The stage reference is usually an elevation value from an object that cannot be moved (e.g., a survey marker, staff gauge, bridge).
-
Select Next to move to the next station.
-
Select Distance to vertical.
-
Enter the distance of the vertical to the reference point at the bank (e.g., tagline zero).
-
Select Set Depth.
-
Enter the maximum depth (manual mode) or measure the maximum depth (automatic mode).
-
Enter the stage reference.
-
Select Measure Velocity. Select an applicable velocity measurement method.
-
Select a measurement point from the list. Obey the instrument prompts and adjust the sensor to the correct depth. If the sensor has a depth option, adjust the sensor depth until the depth box is green. This means the sensor is in at the correct position.
Note : Red indicates more adjustment is necessary. Yellow indicates the depth is close to the correct depth. -
Select Capture to start the measurement process.
-
If necessary, the setup can be changed and the measurement can be repeated. When the measurement is complete, push OK to store the data.
-
Repeat steps 20–22 for the other measurement points on the vertical.
-
When all measurements for the station are complete, select Main or Verify results. Push OK to return to the list of measurement points.
Option Description
Main Returns to the station menu.
Verify Shows the average velocity reading for the station based on the measurement method. -
Select Next to go to the next station.
-
Repeat steps 14–25 for all of the remaining stations selected for velocity measurements.
-
If the stop bank shows, select Edge/Obstruction. Repeat steps 8–12. To examine the measurement before exit, select Channel Summary.
-
Select Save data and Exit. Enter a name for the file and push OK.
Note : A warning flag will show if the discharge in one or more segments is more than 5% of the total discharge.
Insert or delete a station
Prev, Next, Ins and Del options show at the bottom of the display in the
Station screen. Prev and Next are used to navigate to a previous or subsequent
station. Ins and Del are used to insert or delete a station.
For example, after measurements have been done at 10 stations, a user may wish to insert a new station between stations 3 and 4. The steps below describe how to do this. These steps can be applied in similar situations.
-
Select Prev and push OK until the display shows the information for Station 3.
-
Select Ins and push OK.
The instrument adds a new station named Station 4. Subsequent stations are automatically given new sequential numbers. -
To delete the current station (when in non-fixed mode), select Del and push OK.
Conduit profiles
It is possible to use all of the methods for conduit profiles in this section
in sites with a typical profile shape and sufficient depth to measure 3-point
velocities. The 0.9 x Vmax method can also be used when the depth is not
sufficient for multi-point profiles.
Note: In typical conduit profiles, the first measurement is made at the bottom. Subsequent measurements are made above the one made before. A different procedure may be necessary for some profiles.
-
In the Main Menu, select Profiler.
-
Enter the operator name.
-
In the list of options, select Conduit.
-
Enter a name for the new Conduit profile.
-
Select the conduit shape.
Note : The input screens that show next depend on the shape selected. -
Enter values at the screen prompts.
When the necessary values have been entered, the display shows the Select Method menu. -
Select a profile method and do the steps for the method.
0.9 x Vmax
measurement method
The meter uses the maximum velocity measurement in the conduit and multiplies
this value by 0.9 to calculate the total flow.
Note : The RC filter mode with a value of 2 to 4 seconds is recommended for this method.
-
In the Select Method menu, select 0.9 x Vmax.
-
With the sensor in the flow, select Measure Velocity to get a velocity measurement. The measured values are shown on the graph.
-
Move the sensor until a point of maximum velocity is found, then push OK.
The meter calculates and shows the flow, maximum and average velocity values. -
Select Save.
The information is saved to a data file.
0.2/0.4/0.8 method
Do measurements at one, two or three points to calculate an average velocity.
Each point represents a percentage of the maximum depth as measured on the
center line as shown in Figure 13.
Figure 13 2-4-8 profile
-
In the Select Method menu, select 0.2/0.4/0.8.
-
Select one of the options.
-
For the selected option, select a measurement point. The meter shows the sensor adjustment information.
-
If necessary, adjust the sensor as necessary.
-
Select Capture.
The meter gets information from the sensor and shows the velocity value in numerical and graphical form. -
If necessary, select Setup to change the Y-axis range or the data filter parameters.
-
Push OK.
-
Do steps 3–7 for all of the other measurement points then push OK to return to the list of measurement points.
-
Select Flow.
-
Select Save to save the information to a data file.
Velocity/Level Integration measurement method
Measurements are done at 10 different depths. The results from all segments
are integrated to calculate the flow value.
- Select Prev or Next to go to another measurement.
- Select Main to return to the Select Method menu.
-
In the Select method menu, select Vel./Lev. Integ. The display shows the first measurement screen.
-
Select Measure Velocity.
The sensor depth information is shown. -
If necessary, adjust the sensor depth as shown.
-
Select Capture.
The handheld unit gets information from the sensor and shows the average velocity value in numeric and graphical form. -
If necessary, select Setup to change the Y-axis range in FPA filtering mode, the X and Y-axis range when in RC filtering mode or the data filter parameters.
-
Push OK to confirm the information.
-
Select Next. The next measurement screen in the series appears.
-
Do steps 2–7 for the other measurement depths.
-
Select one of the options at the bottom of the screen.
2D
measurement method
Velocity is measured while the sensor is moved through the flow as shown in
Figure 14. Select Cancel at any time to cancel the measurement and return to
the Select Method menu. Note: The RC filter mode with a value of 2 to 4
seconds is recommended for this method.
-
In the Select method menu, select 2D.
The sensor depth information is shown. -
If necessary, adjust the sensor depth.
-
Select Capture. While the sensor collects data, move the sensor through the entire cross-section in the pattern shown in Figure 14.
-
If necessary, select Setup to change the Y-axis range in FPA filtering mode, the X and Y-axis range when in RC filtering mode or the data filter parameters.
-
Push OK.
-
Select one of the options shown at the bottom of the screen.
Figure 14 Path of the sensor in the flow
1 Start | 2 Finish |
---|
Download data
The meter directory is Read Only. In Windows, the meter operates as a mass
storage device or removable hard drive.
- Set the meter to USB Mass Storage mode.
- To edit the data in a file, drag and drop the file to a laptop or PC. File names are limited to eight characters.
- Data files are kept in the tab separated variable (.TSV) format. To see files in Microsoft® Excel, double or right-click a file and open the file with Excel.
Real time files are stored in a directory called RT. Stream and conduit profile files are stored in a directory called P.
Delete data files
- To delete all files from USB memory:
- Go to Main Menu > Diagnostics > Delete Files.
- In the confirmation window, select Yes.
- Push OK once to delete the files, then one more time to return to the previous screen.
- To delete Real Time files:
- Go to Main Menu > Real-Time > Files.
- Select Delete All or use the UP or DOWN arrow to select a file in the list.
- Push OK once to delete the files, then one more time to return to the previous screen.
- To delete Profiler files:
- Go to Main Menu > Profiler > Files.
- Select Delete All or use the UP or DOWN arrow to select a file in the list.
- Push OK once to delete the files, then one more time to return to the previous screen.
Section 7 Maintenance
Download the PVM utility
The PVM utility is used to update the firmware in the portable meter. The PVM
Utility is available at http://www.hach.com or http://www.hachflow.com. Do
the steps listed for the selected URL.
- Go to http://www.hach.com.
- Search for FH950.
- Select the product.
- Click the Downloads tab.
- Click the download link.
- Click PVMSetup.msi.
- Select Save or Run.
- Go to http://www.hachflow.com.
- Click Data Management.
- Click Software Download Center.
- Click the FH950 link.
- Click PVMSetup.msi.
- Select Save or Run.
Update the firmware
Note : All data files in mass storage are lost when the firmware is updated. To download data, refer to Download data on page 27.
The PVM Utility is necessary for this procedure. To download the PVM Utility,
refer to Download the PVM utility on page 27.
This is the general procedure to update firmware for the meter and sensor. To
update the sensor firmware, the sensor must be connected to the meter.
-
Double-click the PVM Utility desktop icon.
-
Push the power button on the meter. When the meter self-test is complete, push OK.
-
In the Main Menu, select Setup, then push OK.
-
Select USB>CDC, then push OK.
-
Connect the USB cable to the portable meter and the PC.
-
In the PVM Utility window, select Connect.
-
In the drop-down menu, select the PVM (COM X) port, where COM X is the virtual port number assigned to the PVM by Windows. Push OK.
-
In the left-side panel, select Firmware Update, then select the Meter or Sensor tab.
-
Select the correct firmware version, then click Start.
The firmware download starts. A “Firmware update successful” message shows when the download is complete. For meter updates, the instrument display turns off until the instrument completes the update. Then, the instrument automatically resets and powers up again after a few minutes. Do not try to make the instrument power on or off before the update process is complete. -
In the Main Menu, select Diagnostics > About. Make sure that the firmware versions for both the Handheld Boot and the Handheld Application are correct.
Clean the sensor
Clean the
sensor when unexpected increases or decreases in flow or level trends occur
and after use in sandy or muddy waterways.
For heavy contamination, soak the sensor in clear water for a few minutes to
help make the contamination easy to remove.
Disconnect the sensor from the meter before it is cleaned. Use only solutions
listed as acceptable in Table 4 to clean the sensor. For sensors with a
pressure cell (i.e., velocity plus depth sensors), make sure the holes for the
pressure cell chambers are washed out and clear of contamination. Rinse the
sensor with clean water before re-attaching the sensor to the assembly.
Table 4 Acceptable and unacceptable cleaning solutions
Acceptable | Do not use |
---|---|
Dish detergent and water | Concentrated bleach |
Window cleaner | Kerosene |
Isopropyl alcohol | Gasoline |
Aromatic hydrocarbons
Clean the meter
- Push the power button to de-energize the meter.
- Use a clean, moist cloth to clean the meter exterior. Mix the water with a mild detergent if necessary.
- Dry the meter exterior with a clean cloth. Let the meter dry in air completely before it is energized again.
Note : Do not use paper-based cloths to clean the display. This type of cloth can cause damage to the display screen.
Replace the battery
Replace the battery when the battery does not hold power when charged.
Note: If the instrument must be returned to the factory for repair or
maintenance, remove the battery and put the battery in a protective cover
before shipment.
-
Set the power to off.
-
Remove the battery. Refer to Install the battery on page 10.
-
Discard the battery in accordance with local, regional and national regulations.
-
Install a new battery in the same location and with the same orientation. Refer to Install the battery on page 10.
Use only batteries that are supplied by the instrument manufacturer. -
Install the battery cover. Make sure that the cover is secure to keep the enclosure rating.
-
Charge the battery if necessary. Refer to Charge the battery on page 11.
Section 8 Troubleshooting
Diagnostics
In the Main Menu, select Diagnostics to see information about the meter and do
the diagnostic tests in Table 5.
Table 5 Meter diagnostics
Option | Description |
---|---|
About | Shows information about the meter and the sensor. Includes the serial |
number and the firmware version.
Delete files| Deletes all files from memory to make space for new
measurements. Make sure that the data is downloaded to a PC before this option
is selected. The system automatically reformats the memory after file
deletion.
Sensor| Shows diagnostic information about the sensor.
Option| Description
---|---
Self test| Makes the meter do a diagnostic self test.
Key pad test| Does a test of any button to make sure that the button is
functional.
Display test| Does a test on the display to make sure that the display is
functional.
Event log| Lets the user see, delete or export the event log. Export the event
log to make the contents available as an accessible file through USB mass
storage. This option is used primarily by factory service.
Troubleshoot errors
The meter and sensor contain no user-serviceable parts. For the errors and
messages listed, try the corrective action.
If the problem does not go away or a problem occurs that is not in the list,
contact the manufacturer.
Message or problem | Solution |
---|---|
Sensor is not connected | Connect a sensor and try the action again. |
Value is out of range | Change the measurement parameters or put in a different |
value, then try the action again.
Sensor data is known to be not correct or not accurate| Clean the sensor and
test.
Sensor is not recognized| Check the sensor connection. Make sure that the lock
nut on the connection port is tight (finger-tighten only).
Display is dim or is not visible| Push a key on the keypad.
Data is not available or access to the data is not possible| Make sure that
the USB option (Main Menu) is set to Mass Storage.
Meter is unresponsive| Push and hold the power button for at least 3 seconds.
This de- energizes the meter. Energize the meter again.
Note: Do not use this method to power off while in normal operation or if the file access icon is visible in the display.
Section 9 Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury,
damage to the instrument or equipment malfunction. The replacement parts in
this section are approved by the manufacturer.
Note : Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Table 6 Velocity only sensor
Description | Item no. |
---|---|
Sensor with 1.5 m (5 ft) cable | EM950.0005 |
Sensor with 6.1 m (20 ft) cable | EM950.0020 |
Sensor with 12.2 m (40 ft) cable | EM950.0040 |
Sensor with 30.5 m (100 ft) cable | EM950.0100 |
Table 7 Velocity with depth sensor
Description | Item no. |
---|---|
Sensor with 1.5 m (5 ft) cable | EM950.1005 |
Sensor with 6.1 m (20 ft) cable | EM950.1020 |
Sensor with 12.2 m (40 ft) cable | EM950.1040 |
Sensor with 30.5 m (100 ft) cable | EM950.1100 |
Table 8 Handheld meter
Description Item no.
English/Metric FH950.1
Accessories
Table 9 Wading rods
Description | Item no. |
---|---|
Standard wading rod kit, English | 75002 |
Standard wading rod kit, Metric (includes the bullet items below) | 75002M |
Bottom section for wading rod, English/Metric | 43000M/43001M |
Intermediate section for wading rod, English/Metric | 43010M/43011M |
Base plate for wading rod, English/Metric | 43015M |
Double-end hanger for wading rod, English/Metric | 43020M |
Top set wading rod kit, English | 75013 |
Top set wading rod kit, Metric | 75013M |
Table 10 General accessories
Description | Item no. |
---|---|
Absorbent wipe | 9073500 |
Adjustable meter mount | 9071700 |
Power supply, 90–264 VAC, 47–63 Hz, 12 V | 002782 |
AC Adapters, 4-plug wall clip set | 002781 |
Carrying case | 9073400 |
Lanyard, double ended loop | 9072700 |
Lithium battery pack | 9073600 |
Suspension cable kit (includes bulleted items below)3 | 75003 |
• Sensor mount | 42033 |
Description | Item no. |
--- | --- |
• Link connector | 43025 |
• Weight hanger | 43030XX01 |
• Weight pin | 43035M |
• 6.8 kg (15 lb) brass weight | 43040 |
Thumb screw kit (includes four thumb screws) | 9073200 |
Universal sensor mount | 75015 |
USB cable | 9070800 |
Appendix A Appendix
A.1 Mean-section and Mid-section methods
The user can select the Mean-section or the Mid-section method for flow
calculations. The Mean-section method divides the cross-section into
individual flow segments. Pairs of adjacent verticals are the limits of the
segments. The two edges of the cross-section are given values of 0 for the
velocity and depth. The total flow is the sum of the partial flows of all
segments. Figure 15 shows the definitions and the equation for the Mean-
section method.
Where:
- V = velocity at vertical
- b = distance to vertical from bank
- h = depth at vertical
- q = flow at vertical
The Mid-section method also divides the cross-section into individual flow
segments. With the Mid-section method, the segments are not between verticals
but are defined by half of the distance to neighbor verticals in each case.
For this reason, the first and last verticals should be as near to the edges
as possible (i.e., left edge of water (LEW) and right edge of water (REW)).
Boundary conditions dictate the proximity of the first and last vertical to
the edge of water.
Experience shows that the Mid-section method gives more exact results compared
to the Mean-section method so it is the default setting. Figure 16 shows the
definitions and equation for the Mid-section method.
Where:
- m = station number
- n = total number of stations
- V = velocity at vertical
- b = distance to vertical from bank
- h = depth at vertical
- q = flow at vertical
A.2 Profiles and measurements
Profiles can be set up for streams or conduits. Figure 17 shows an example of
a typical profile shape in a conduit. In a typical profile, velocity is less
near a wall or edge than at the center and decreases near the surface.
Multiple velocity measurements in the profile are averaged to calculate the
total flow. Measurements for conduit profiles are made from the bottom up.
Measurements for stream profiles are made from the top down or bottom up.
In the Main Menu, select Profiler. The meter prompts for the operator name,
the type of profile
(stream or conduit) and the profile name. Do the instructions for the selected
profile type.
Figure 17 Typical profile
1 Depth | 2 Velocity |
---|
A.3 Site selection
A site with the typical profile shape gives the most accurate results. Visual
inspection is typically sufficient to identify problem sites. Use the
information in these guidelines to help select the best site.
These guidelines apply to conduit and stream profiles.
- The channel should have as much straight run as possible. If the length of the straight run is limited, the length upstream from the profile should be two times the downstream length.
- The channel should be free of flow disturbances. The site must not have protruding pipe joints, sudden changes in diameter, contributing side-streams, outgoing side-streams or obstructions. Remove all rocks, sediment or other debris from the bottom of the pipe.
- The flow should not have visible swirls, eddies, vortices, back-flow or dead zones.
- Do not select areas immediately downstream from sharp bends or obstructions.
- Do not select areas with converging or diverging flow (approaches to a flume) or vertical drops.
- Do not select areas immediately downstream from sluice gates or places where the channel spills into a body of stationary water.
A.4 Do a velocity calibration
Use this feature to remove a velocity offset if necessary. The velocity offset
stays active until the meter power is switched off.
- Collect a bucket of water from the water in the profile area. The bucket must be non-metallic and at least 20.32 cm (8 in.). The water depth must be at least 15.24 cm (6 in.).
- Put the sensor in the center of the bucket so that it does not touch the sidewall or the bottom of the bucket.
- Let the water become still.
- Let the velocity reading stabilize.
- Select Zero Velocity.
A.5 Wet/Dry threshold
The wet/dry threshold is the trigger point for the meter to know when the
sensor is in or out of the water. This information is important because if the
meter does not know that the sensor is under the surface of the water, the
meter sets the velocity value to zero. For a profile or real-time reading, the
meter prompts the user to submerge the sensor in the water.
If the specific conductivity of the water being measured is very low, adjust
the threshold value for the best performance. Go to Setup Menu>Wet/Dry
Threshold.
For troubleshooting, the present reading is used as a guide to set a custom
threshold. Get a reading in the water (wet) and then out of the water (dry).
The threshold value must be between the wet and dry value. For example, if the
actual wet value is 17 and the actual dry value is 2, put in a threshold value
half way between 2 and 17. The default value is 20.
HACH COMPANY World Headquarters
- P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. 970-669-3050
- 800-368-2723 (U.S.A. only)
- U.S.A. – orders@hach.com
- International – intlsupport@hach.com flowtechsupport@hach.com
- www.hach.com/flow
© Hach Company/Hach Lange GmbH, 2012, 2014, 2018, 2020. All rights reserved.