ISENO3181 IntelliCal Nitrate Ion Selective Electrode
User Manual DOC022.53.80031
ISENO3181
05/2021, Edition 5 User Manual
Section 1 Product overview
The Intellical ISENO3181 series probes are digital, combination ion-selective
electrodes (ISE) that measure the concentration of nitrate in general water
samples. The probes have a built-in temperature sensor, a non-refillable gel
reference and non-replaceable membranes. Refer to Figure 1.
Figure 1 Probe overview
- Sensing element
- Reference junction
- Sensor protection cap
- Cable
Section 2 Specifications
Specifications are subject to change without notice.
Specifications|
Details
---|---
Probe type| Digital combination probe with a non-refillable reference junction
and a built-in temperature sensor
Measurement range| – 0.1 mg/L (7×10–6 M) to 14,000 mg/L (1 M) NO3 —N
Accuracy| ±0.02 mV or 0.05% (the larger value)
Sample pH range| pH 2 to 11 (pH 3 to 5 after the addition of Nitrate ISA)
Reference type| Ag/AgCl
Reference junction| Double junction (ceramic porous pin and annular porous
PTFE)
Slope| 58 mV/decade (90 to 110% at 25 °C (77 °F) in linear range per Nernstian
theoretical value)
Linear region| – 0.7 mg/L to 14,000 mg/L NO3 —N
Temperature accuracy| ±0.3 °C (±0.54 °F)
Temperature sensor type| 30 kΩ NTC thermistor
Operating temperature| 0 to 50 °C (32 to 122 °F)
Storage temperature| 5 to 35 °C (41 to 95 °F)
Response time in linear region| < 60 seconds (application dependent)
Minimum sample volume| 25 mL
---|---
Minimum immersion depth| 25.4 mm (1 in.)
Body material (standard)| Epoxy
Membrane| Solid-state PVC membrane
Reference electrolyte| Non-refillable Dritek gel reference element
Cable connection| M12 digital output and connector
Dimensions| Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.9 in.) total; 103 mm (4.1 in.) below head
Cable length: ISENO318101: 1 m (3.3 ft); ISENO318103: 3 m (9.8 ft)
Warranty| 1 year on the probe. This warranty covers manufacturing defects, but
not improper use or wear.
Certifications| CE, FCC/ISED
Section 3 Safety information
3.1 Intended use
The Intellical probes are intended for use by individuals who measure water
quality parameters in the laboratory. The Intellical probes do not treat or
alter water.
3.2 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
3.3 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage
to the instrument could occur if not observed. A symbol on the instrument is
referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in
European domestic or public disposal systems. Return old or end-of-life
equipment to the manufacturer for disposal at no charge to the user.
3.4 Product hazards
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
Section 4 Preparation for use
Prepare the probe for calibration and measurement as follows.
-
Remove the sensor protection cap from the probe.
-
Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not wipe the tip of the probe.
-
Condition the probe before use. To condition the probe, soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes. Do not add the ISA to the standard solution.
Note: Condition the probe each day for 15 to 30 minutes or for 30 to 60 minutes after long-term storage. -
Make sure that the meter has the correct date and time settings. The service-life time stamp in the probe comes from the date and time settings in the meter.
Note: Some meters automatically open the date and time settings when the meter starts for the first time, or after battery replacement. -
Connect the probe to the meter.
Section 5 Calibration
The procedure that follows is applicable to meters that can connect to
Intellical ISE probes. Refer to the applicable meter documentation for meter
operation and probe-specific settings.
5.1 Calibration notes
Read the notes that follow before calibration.
- Measure the standard solutions from lowest to highest concentration for best results.
- Keep all of the solutions (standard solutions and samples) at the same temperature (± 2 °C (± 3.6 °F)) for best results.
- Stir the standards and samples at a slow and constant rate to prevent the formation of a vortex.
- Use the default calibration options or change the options in the probe settings menu.
- Use the single display mode for calibration when more than one probe is connected to the meter (if applicable).
- Calibrate the probes and verify the calibration regularly for best results. Use the meter to set calibration reminders.
- The calibration data is stored in the probe. When a calibrated probe is connected to a different meter with the same calibration options, a new calibration is not necessary.
- Air bubbles below the sensor when in solution can cause a slow response or error in the calibration. Make sure to remove air bubbles during calibration.
5.2 Calibration procedure
-
Go to the calibrate menu. Select the probe, if applicable. The display shows the standard solutions to use for calibration.
-
Add 25 mL of each standard solution to different beakers.
-
Add one Nitrate Ionic Strength Adjustor (ISA) Powder Pillow to each 25 mL of standard solution.
-
Add a stir bar to the first standard solution. Put the standard solution on an electromagnetic stirrer. Stir at a moderate rate.
-
Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
-
Put the probe in the standard solution with the sensor fully submerged. Do not put the probe on the bottom or sides of the beaker.
-
Shake the probe from side to side to remove air bubbles.
-
Stir for 30 to 60 seconds, then read the nitrate concentration of the standard solution.
-
Do steps 3 through 8 to read the value of the remaining standard solutions.
-
Save the calibration.
Section 6 Sample measurement
The procedure that follows is applicable to meters that can connect to
Intellical ISE probes. Refer to the applicable meter documentation for meter
operation and probe-specific settings.
6.1 Sample measurement notes
Read the notes that follow before sample measurements.
- Rinse the probe with deionized water and dry with a lint-free cloth between measurements to prevent contamination.
- If the stabilization time is long, try a different stir rate and make sure to condition the probe.
- Keep all of the solutions (standard solutions and samples) at the same temperature (± 2 °C (± 3.6 °F)) for best results.
- If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the meter manual for instructions.
- Stir the standards and samples at a slow and constant rate to prevent the formation of a vortex.
- The meter automatically saves the measurement data when the user manually reads each data point and when the meter is set to read at regular intervals. The user must manually save each data point when the meter is set to read continuously.
- Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure to remove air bubbles before and during measurements.
6.2 Sample measurement procedure
-
Pour 25 mL of fresh sample into a 50mL beaker.
-
Add one Nitrate Ionic Strength Adjustor (ISA) Powder Pillow.
-
Add a stir bar. Put the beaker on an electromagnetic stirrer. Stir at a moderate rate.
-
Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
-
Put the probe in the sample with the sensor fully in the sample. Do not put the probe on the bottom or sides of the beaker.
-
Shake the probe from side to side to refresh the reference junction and remove air bubbles.
-
Stir for 30 to 60 seconds, then read the nitrate concentration of the sample. The display shows the value when the reading is stable.
6.3 Low-level measurements
Use the guidelines that follow for measurements at low concentrations (<1 mg/L
NO3N).
- Fully rinse the probe with deionized water and blot dry between measurements.
- Soak the probe in the lowest concentration standard solution for 1 hour maximum before calibration and measurement.
- Use the meter to set the stability criteria for the probe to a low value.
- Stir the standards and samples at a slow and steady rate to prevent the formation of a vortex.
- Use a dilute ionic strength adjustor (ISA) solution for calibration and measurements:
- Dissolve the contents of one ionic strength adjustor powder pillow in 50 mL of deionized water.
- Add 5 mL of the dilute (ISA) solution to each 25 mL of standard solution or sample.
Note: The ISA is not necessary when all of the conditions that follow apply:
- The sample does not contain interferences.
- The sample pH is within the range given in the specifications.
- Omission of the ISA is accepted by the regulatory reporting agency (if the measurement is for regulatory reporting).
6.4 Interferences
The sensing element will measure some other ions that are known to interfere
with the method. The probe response to other ions usually increases the mV
potential and causes a positive error. The response to other ions can be semi-
quantitatively calculated through the Nikolsky equation, an extended Nernst
equation:
E = Eº + (RT/(zF))ln[aNa + KNax × ax] Where
- ax = the activity of the interfering ion
- KNax = the selectivity coefficient for the interfering ion relative to the primary ion
The primary interferences for nitrate ion-selective electrodes are shown in
Table 1. To remove most of the interferences, add the Nitrate ISA, which
adjusts the pH to between pH 3 and 5. The measurements are usually made at
this pH range to remove carbonate and bicarbonate interference and to decrease
the organic acid interference.
The selectivity coefficient is the approximate apparent increase in the
measured concentration that is caused by one unit of the interfering ion
(e.g., 1 unit of CIO4 increases the nitrate concentration by 0.1). The
approximate selectivity coefficients for some ions with the Intellical Nitrate
ISE are shown in Table 1.
Table 1 Interferences
Interference | K (selectivity coefficient) value |
---|---|
Perchlorate (CIO4 ) | 1200 (interferes more) |
Iodide (I–) | 10 |
Bromide (Br–) | 0.1 |
Chloride (Cl–) | 0.006 |
Nitrite (NO2 ) | 0.001 (interferes less) |
Section 7 Verify the calibration
Measure the value of a fresh standard solution at regular intervals to make
sure the result is accurate. The meter compares the expected standard solution
value to the measured value and accepts or rejects the measurement. The user
can change the standard solution and acceptance criteria for verification in
the probe-specific settings.
Note: Password protection may prevent access to the acceptance criteria.
7.1 Verification procedure
-
Go to the verification menu. The display shows the standard solution to use for verification. Note: Menu name for HQd meters: Run check standard.
-
Pour 25 mL of the standard solution into a 50mL beaker.
-
Add one Nitrate Ionic Strength Adjustor (ISA) Powder Pillow.
-
Add a stir bar. Put the beaker on an electromagnetic stirrer. Stir at a moderate rate.
-
Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
-
Put the probe in the standard solution with the sensor fully in the solution. Do not put the probe on the bottom or sides of the beaker.
-
Shake the probe from side to side to remove air bubbles.
-
Stir for 30 to 60 seconds, then read the value of the standard solution. The meter accepts or rejects the result.
Section 8 Maintenance
Regular maintenance is necessary for the best accuracy, stabilization time and
life of the probe. Fully rinse the probe with deionized water between
measurements to keep the probe clean.
8.1 Clean the probe
Rinse the probe with deionized water regularly to remove mineral or sample
buildup on the sensing element. Symptoms of contamination:
- Incorrect or irregular readings
- Slow stabilization times
- Calibration errors
- Sample material stays on the probe
- Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the tip of the probe.
- Polish the white reference junction on the probe tip with a soft cloth or cotton swab. Refer to Product overview on page 3. Do not polish the PVC sensing element in the center of the probe tip. Do not use cleaning solutions. Make sure to fully rinse the probe with deionized water between sample measurements to prevent mineral or sample buildup on the sensing element.
- Soak the probe for 30 minutes in a 100-mg/L nitrate nitrogen (NO3–N) standard solution.
8.2 Storage
Keep the sensor protection cap on the probe during storage to prevent damage
to the sensing element. Note: Make sure to condition the probe after long-term
storage. Refer to Preparation for use on page 5.
- Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
- Make sure that the probe sensor is dry, then install the sensor protection cap. Refer to Product overview on page 3.
Section 9 Troubleshooting
Refer to Table 2 for general troubleshooting information. To check the probe
performance, refer to Slope check on page 11. To check the accuracy of sample
measurements, refer to Standard additions check on page 11.
Table 2 Troubleshooting information
Problem| Possible cause|
Solution
---|---|---
Decreased probe performance causes slow stabilization and prevents accurate
calibrations or measurements.| The probe has contamination on the sensing
element.| Rinse and condition the probe.
Refer to Clean the probe on page 10.
The reference junction is clogged.| Fully rinse the reference junction with
deionized water. Shake the probe down to remove air bubbles.
The probe is not conditioned to the sample sufficiently.| Condition the probe.
Refer to Preparation for use on page 5.
Sample properties cause slow stabilization or inaccurate measurements.| The
sample pH with ISA is incorrect.| Make sure that the sample pH after the ISA
is added is between 3 and 5. Make sure to add one ISA pillow to each 25 mL of
sample.
---|---|---
The sample temperature is low, or there is a large temperature difference
between samples.| Increase the sample temperature or adjust the temperature of
different samples to be the same (within 2 °C (3.6 °F)).
Procedure problem causes slow stabilization and prevents accurate calibrations
or measurements.| Air bubbles are around or below the probe tip.| Carefully
tap or shake the probe to remove air bubbles.
The ISA was not added.| Add one ISA powder pillow to each 25 mL of sample and
standard solution.
The stir speed is too slow or too fast.| Try a different stir speed.
Magnetic stirrers can become warm and increase the solution temperature.| Put
a piece of insulating material between the stirrer and beaker.
An incorrect standard solution was used or the standard solution has
contamination.| Use the specified standard solution of good quality.
Too much time passed after the ISA was added.| Measure the sample or standard
solutions immediately after the ISA is added.
9.1 Slope check
Use the mV value of two standard solutions to make sure the probe gives the
correct slope.
- Prepare two standard solutions that are ten times apart in concentration (e.g., 10 mg/L and 100 mg/L NO3–N). Select standard solutions with a concentration above and below the typical sample concentration. Use a minimum concentration of 3.1 mg/L.
- Use the measurement procedure to add the ISA and measure the mV value of each standard solution.
- Calculate the difference in the mV value of the two standard solutions to find the slope. If the probe is in good condition, the slope will be 58 mV (within the ± slope limits of the method) at 25 °C (77 °F).
9.2 Standard additions check
To make sure that the sample measurement is accurate, add a small volume of a
standard solution to the sample and calculate the percent recovery. The sample
with the known volume of standard solution is known as a spiked sample.
- Use the measurement procedure to measure the concentration of a 25-mL sample.
- Use a pipet to add the applicable volume of standard solution to the sample. Refer to Table 3.
Table 3 Standard solution volumes and concentrations
Measured sample concentration
| Volume of standard to add|
Concentration of standard solution
---|---|---
1 to 2 mg/L| 0.5 mL| 100 mg/L NO3 —N
3 to 6 mg/L| 1.0 mL| 100 mg/L NO3 —N
7 to 15 mg/L| 0.3 mL| 1000 mg/L NO3 —N
15 to 30 mg/L| 0.5 mL| 1000 mg/L NO3 —N
30 to 60 mg/L| 1.0 mL| 1000 mg/L NO3 —N
3. Measure the concentration of the spiked sample.
4. Calculate the expected (theoretical) concentration of the spiked sample:
CE = (CS x VS/VT) + (CSS x VSS/VT) Where:
- CE = expected (theoretical) concentration of the spiked sample
- CS = concentration of the sample (mg/L) before the standard solution was added
- CSS = concentration of the standard solution (mg/L)
- VS = sample volume (mL) before the standard solution was added
- VSS = volume of the standard solution (mL)
- VT = total volume (standard solution volume (mL) + sample volume)
5. Calculate the percent recovery of the standard addition. A percent
recovery of 100 (±5)% is an indication that the sample measurements are
accurate.
Percent recovery = CM/CE x 100
Where:
- CM = measured concentration of the sample after the addition of the standard solution
- CE = expected (theoretical) concentration of the sample after the addition of the standard solution
Section 10 Consumables
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Description| Quantity|
Item no.
---|---|---
Nitrate Ionic Strength Adjustor Powder Pillows (buffer only)| 100/pkg| 2984799
Nitrate Ionic Strength Adjustor Powder Pillows (includes components to remove
interferences)| 100/pkg| 4456369
Nitrate Ionic Strength Adjustor Buffer Solution (includes components to remove
interferences)| 500 mL| 2488349
Nitrate Standard Solution, 1 mg/L NO3—N| 500 mL| 204649
Nitrate Standard Solution, 10 mg/L NO3—N| 500 mL| 30749
Nitrate Standard Solution, 100 mg/L NO3—N| 500 mL| 194749
Nitrate Standard Solution, 1000 mg/L NO3—N| 500 mL| 1279249
Wash bottle, polyethylene, 500 mL| 1| 62011
Disposable wipes, 11 x 22 cm| 280/pkg| 2097000
Beaker, 30 mL, plastic, colorless| 80/pkg| SM5010
---|---|---
Beaker, 100 mL, polypropylene| 1| 108042
Probe stand for standard Intellical probes| 1| 8508850
10.1 Accessories
Description| Quantity|
Item no.
---|---|---
Beaker, polypropylene, 50 mL, low form| 1| 108041
Disposable wipes, 11 x 22 cm| 280/pkg| 2097000
Wash bottle, polyethylene, 500 mL| 1| 62011
Probe stand for standard Intellical probes| 1| 8508850
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. 970-669-3050
800-227-4224 (U.S.A. only)
Fax 970-669-2932
orders@hach.com
www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
info-de@hach.com
www.de.hach.com
© Hach Company/Hach Lange GmbH,
2010-2011, 2013, 2021.
All rights reserved.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>