JBC JNA-1UB 120V High Precision Hot Air Control Unit Instruction Manual
- June 16, 2024
- JBC
Table of Contents
JBC JNA-1UB 120V High Precision Hot Air Control Unit
Product Specifications
- Model : JNA
- Available References : JNA-1UB (120V), JNA-2UB (230V), JNA-9UB (100V)
- Packing List
- High-Precision Hot Air Control Unit – 1 unit
- Power Cord – 1 unit (Ref. 0023715 for 120V, Ref. 0023714 for 230V, Ref. 0024092 for 100V)
- Manual – 1 unit (Ref. 0031960)
Product Features and Connections
JNA High-Precision Hot Air Control Unit
- Display
- Pick and Place Activation Switch
- T260 Pick & Place Handle
- J125 Cartridge Range
- USB-A Connector
- Main Switch
- NH High-Precision Heater Hose Set (sold separately)
- NAS Stand for NH High-Precision Heater Hose Set (sold separately)
- Power Socket
- Fuse
- USB Connector
- Equipotential Connector
- Pedal Connector
- Auxiliar Connector FAE / Robot Connector
Product Usage Instructions
JNA Work Screen
JNA offers an intuitive user interface which provides quick access to
station parameters. The default PIN is 0105
- Status Bar
- Instant Power Supplied to Heater
- Current Air Temp.
- Air Temp. Selected
- Hot Air Temp.
- Air Flow Selected
- Power
- Cartridge
- Status Indicator
- Selected Air flow
- Station Information
- Menu Options
Station Update
To update the station, follow these steps
- Download the JBC Update File from www.jbctools.com/software.html.
- Insert the USB flash drive with the downloaded file into the station.
- Press “Update” and then “OK” to start the update process.
Adjustable Stand (Ref. NA-SB)
The adjustable stand allows you to customize the tool holder angle to suit
your work position
Operation
- The tool starts blowing when lifted from the stand and stops when returned to the stand.
- The stand is designed for straight insertion of the tool through its opening, especially when using J325 Cartridges.
Changing Cartridges
To change cartridges, follow these steps
- Removing
- Place the cartridge in the extractor.
- Pull the handle to remove the cartridge.
- Use the ring (J325) or umbrella (J125) around the cartridge as a stop to extract it from the tool.
- Inserting
- Place the handpiece on top of the new cartridge.
- Press down slightly to insert the cartridge.
- Fixing
- Use the holes to fix the cartridge at the right depth.
- The extractor holes also allow correcting the nozzle angle of curved cartridges.
- To adjust the nozzle position, rotate the tool with the cartridge inserted in the extractor until the nozzle is in the desired position.
FAQ
- Where can I find troubleshooting information for the station?
Station troubleshooting is available on the product page at www.jbctools.com.
JNA
High-Precision Hot Air Control Unit
This manual corresponds to the following references
- JNA-1UB (120V)
- JNA-2UB (230V)
- JNA-9UB (100V)
Packing List
The following items are included
High-Precision Hot Air
Control Unit …………………………………………..1 unit
Power Cord ………………………………………… 1 unit
Ref. 0023715 (120V)
0023714 (230V)
0024092 (100V)
Manual …………………………………………………..1 unit
Ref. 0031960
Features and Connections
JNA Work Screen
JNA offers an intuitive user interface which provides quick access to station
parameters. Default PIN : 0105
- Set the station parameters.
- Set the tool parameters, such as cartridge recognition (automatic or manual).
- Display the hours worked in each cycle.
- It is possible to choose the language from a list.
- Allows you to carry out an overall station reset restoring all the parameters to their default values.
Troubleshooting
Station troubleshooting available on the product page at www.jbctools.com
Station Update
- Station Update
- Download the JBC Update File from
- www.jbctools.com/software.html.
- Insert the USB flash drive with the downloaded file into the station.
System Notifications
The following icons will be displayed on the screen’s status
-
USB flash drive is connected.
-
Station is controlled by a PC.
-
Station is controlled by a robot.
-
Station software update.
Press INFO to start the process. -
Warning
Press INFO for failure description. -
Error
Press INFO for failure description, type of error and how to proceed.
Adjustable Stand * (Ref. NA-SB)
Adjust the tool holder angle to suit your work position.
Operation
The tool starts/stops blowing when lifted/returned to the stand
The stand is designed so that the tool is inserted in a straight line through
its opening, rather than being inserted from above, especially when using J325
Cartridges. *not included, sold separately.
Changing Cartridges
Save time and change the cartridge safely without switching the station
off.
-
Removing
Place the cartridge in the ex-tractor and pull the handle to remove it. Use the ring (J325) or umbrella (J125) around the cartridge as a stop to extract it from the tool. -
Inserting
Place the handpiece on top of the new cartridge and press down slightly. -
Fixing
Use the holes to fix the cartridge at the right depth -
The extractor holes also allows correcting the nozzle angle of curved cartridges. To do this, rotate the tool
-
With the cartridge inserted in the extractor until the nozzle is in the right position.
Important
It is essential to insert the cartridge up to the mark for a proper
connection.
Compatible Cartridges
NH works with J125 and J325 Cartridge Ranges. Find the model that best suits your soldering needs at www.jbctools.com
Compatible Accessories
Grips for NH High Precision Heater Hose Set (Grip Ref. 0024473)
-
Inserting Tabs
Insert the slide-on tabs into the new grip. -
Inserting Handle
Push the grip with the tabs onto the handle. -
Removing Tabs
Hold the grip and pull them out. Use a pliers if necessary
Pick & Place
This tool helps you place and remove SMDs of any size easily thanks to the
suction pump.
Set includes the following needles and cups (2 units of each reference)
- Ref. 0026637: 20G, 60º, Ø0.60 mm/0,0236 in
- Ref. 0026638: 25G, 60º, Ø0.26 mm/0.01023 in
- Ref. 0026639: 30G, 60º, Ø0.159 mm/0.00626 in
- Cup Ref. 0940112, used with needle Ref. 0026637
*not included , sold separately
Operation
- Choose the needle that best fits for the component and the suction cup if needed and start as follows
- Insert the needle with the cup for a correct suction process.
- Make sure the needle does not protrude from the cup.
- Once the suction is activated, cover the pen hole with your finger and lift off the component.
- Lift your finger to release the component
Pedal P405
By means of the pedal connection, NH High-Precision Heater Hose Set and the
T260 Pick & Place Tool can be activated and deactivated.
The pedal should be plugged in at the rear of the station into its connector,
which is marked with “Pedal”, to use the Pedal functionality.
Pedal and Tool Settings
Start Mode
Select menu . Then by pressing select “TOOL” and confirm with .
Select “PEDAL FUNCTION” and confirm with OK button. Then choose between “HOT AIR” and “SUCTION”. Confirm again with OK button.
Pedal Activation Mode
In both applications, HOT AIR and SUCTION, there are 2 different pedal
activation modes available.
One Touch Mode
Press the pedal once to activate the function. Press again to stop the
function.
Pressed Mode
Start the function by pressing the pedal continuously and stop the function by
lifting the pedal. The function is active as long as the pedal is kept
pressed.
Select “PEDAL”, confirm with OK button and choose between “PRESS ONCE” and
“PRESS CONT”. Confirm again with OK button.
Recommended
For rework jobs the use of the preheaters* is highly recommended. Find the
preheater that best suits your soldering needs on
www.jbctools.com.
- Applying high temperatures (350 ºC – 450 ºC / 660 ºF – 840 ºF) with a hot air tool for a long time can damage both the printed circuit and the electronic components.
- In addition, significant temperature differences between two points on a printed circuit board can damage the printed circuit and the electronic components due to the stresses generated by the different expansion of the material.
- Preheating is a process that minimizes thermal stress and allows the soldering process to take no longer than it should.
- For most rework processes JBC recommends a preheat temperature between 90 and 130 ºC / 195 and 265 ºF and not exceeding a heating rate of 4 ºC/s.
- With the use of JBC preheaters, a homogeneous and controlled heating is achieved within the recommended parameters.
Maintenance
Before carrying out maintenance, always switch the station off and disconnect it from the mains. Allow the equipment to cool down.
- Clean the station screen with a glass cleaner or a damp cloth.
- Use a damp cloth to clean the casing and the tool. Alcohol can only be used to clean the metal parts.
- Periodically check that the metal parts of the tool and stand are clean so that the station can detect the tool status.
- Periodically check all cables and tubes.
Replace a blown fuse as follows
- Pull off the fuse holder and remove the fuse. If necessary use a tool to lever it off.
- Replace any defective or damaged pieces. Use original JBC spare parts only.
- Repairs should only be performed by a JBC authorized technical service.
- . Insert the new fuse into the fuse holder and return it to the station.
Safety
It is imperative to follow safety guidelines to prevent electric shock, injury, fire or explosion.
- Do not use the units for any purpose other than soldering or rework. Incorrect use may cause fire.
- The power cord must be plugged into approved bases. Make sure that it is properly grounded before use. When unplugging it, hold the plug, not the wire.
- Do not work on electrically live parts.
- The tool should be placed in the stand when not in use to turn off the hot air.
- The soldering tip, the metal part of the tool and the stand may still be hot even when the station is turned off. Handle with care, including during the stand position adjustment.
- Do not leave the appliance unattended when it is on.
- Do not cover the ventilation grills. Heat can cause inflammable products to ignite.
- Avoid the contact of flux with skin or eyes to prevent irritation
- Be careful with the fumes produced when soldering.
- Keep your workplace clean and tidy. Wear appropriate protection glasses and gloves when working to avoid personal harm.
- Utmost care must be taken with liquid tin waste which can cause burns.
- This appliance can be used by children over the age of eight and also persons with reduced physical, sensory or mental capabilities or lack of experience provided that they have been given adequate supervision or instruction concerning use of the appliance and understand the hazards involved. Children must not play with the appliance.
- Maintenance must not be carried out by children unless supervised.
Notes
Specifications
JNA
High-Precision Hot Air Control Unit
Ref.: JNA-1UB 120V 50/60Hz. Main fuse: T2A. Rated current 0.85A
Ref.: JNA-2UB 230V 50/60Hz. Main fuse: T2A. Rated current 0.85A
Ref. : JNA-9UB 100V 50/60Hz. Main fuse: T2A. Rated current 0.85A
- Temperature Selection : 150 – 450 °C / 300 – 840 °F
- Nominal Power : 70W
- Cool Mode : “Temp. off” is used to blow air at room temperature
- Ambient Operating Temp.: 10 – 50 ºC / 50 – 122 ºF
- Air Flow Regulation : 0.15 – 2.5 SLPM
- Vacuum : 53% / 397 mmHg / 15.6 inHg
- Connectors : USB-A / USB-B RJ12 for FAE / Robot Pedal / Stand
- Net Weight : 1.65 kg / 3.64 lb
- Total Package Dimensions/Weight : 330 x 260 x 210 mm / 2.65 kg (L x W x H) 12.99 x 10.24 x 8.27 in / 5.84 lb
Complies with CE standards.
ESD safe.
Warranty
- JBC’s 2 year warranty covers this equipment against all manufacturing defects, including the replacement of defective parts and Labour.
- Warranty does not cover product wear or misuse.
- For the warranty to be valid, equipment must be returned, postage paid, to the dealer where it was purchased.
- Get 1 extra year JBC warranty by registering here: https://www.jbctools.com/productregistration/ within 30 days of purchase.
This product should not be thrown in the garbage.
In accordance with the European directive 2012/19/EU, electronic equipment at
the end of its life must be collected and returned to an authorized recycling
facility.