Sinocare C-36300625 2-in-1 Uric Acid Blood Glucose Test Kit
Blood Glucose and Uric Acid Monitoring System User Manual SafeAQ UG
Dear Owner
Sinocare thanks you for choosing the Safe AQ UG blood glucose and uric acid
monitoring system. This user manual contains important information about the
meter and how it works.
Please read this user manual carefully before using your new meter, or the
protection provided by the meter may be impaired.
If you have any questions, please feel free to contact our Customer Service at
+86-731-89935929.
Safe AQ UG Blood Glucose and Uric Acid Monitoring System
Introduction to the Your Blood Glucose and Uric Acid Monitoring System
Intended Use
- The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is designed for the quantitative measurement of glucose and uric acid in fresh capillary whole blood samples and in venous whole blood samples. The Safe AQ UG
- Blood Glucose and Uric Acid Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes and hyperuricemia (HUA).
- The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is intended for in vitro diagnostic use and should not be used for the diagnosis of or screening of diabetes and hyperuricemia (HUA).
- The Safe AQ UG Blood Glucose and Uric Acid Monitoring System includes:
- Safe AQ UG blood glucose and uric acid meter, Safe AQ UG blood glucose test strip, Safe AQ UG uric acid test strip, blood glucose control solution, uric acid control solution
- Unit type: Safe AQ UG Blood Glucose and Uric Acid meter displays blood glucose results in mg/dl or mmol/L and uric acid results in mg/dl or µmol/L.
Test Principle
A blood glucose test or uric acid test is based on measurement of electrical
current caused by the reaction of glucose or uric acid with the reagents
(special chemicals) on the electrode of the strip. The blood or control
solution sample is drawn into the tip of the test strip through capillary
action. Glucose or uric acid in the sample reacts with the special chemicals
and generates electrons, which produce electrical current. The meter measures
the electrical current and calculates the glucose or uric acid result. The
blood glucose results are displayed in mg/dl or mmol/L and uric acid results
are displayed in mg/dl or µmol/L.
Important Safety Information
- For self-testing customer, your Safe AQ UG meter and lancing device are for single person use only. Do not share them with anyone else, even family members.
- For professional use, if the meter is being used on the second person, the meter and lancing device should be disinfected first.
- Safe AQ UG test strip and lancet are for single use only. DO NOT REUSE.
- Do not use other test strips and control solutions with the Safe AQ UG meter.
- Do not use the meter if it is not working properly, or there is any damage.
- Keep the test strip vial away from children. The strip vial, test strips, and control solution bottles can be choking hazards. DO NOT drink the control solution.
- Remove the batteries if the meter is not likely to be used for some time.
DO NOT CHANGE YOUR TREATMENT BASED ON A SINGLE RESULT THAT DOES NOT MATCH HOW YOU FEEL OR IFYOU BELIEVE THAT YOUR TEST RESULT COULD BE FALSE.
Patient Cautions
- Not for neonatal (newborn or infant) use
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on the critical patients.
- For in vitro diagnostic use only.
Important Health-Related Information
The American Diabetes Association (ADA) suggests the following targets for
normal blood glucose range. More or less stringent goals may be appropriate
for each person. If your test result is out of the range or your test result
is not in accordance with your feeling, retest your blood glucose. If still
out of the range, please contact healthcare professional.
Expected Results for people without diabetes:
The Modern clinical laboratory diagnostics-test and clinical (May, 2009, version 2) suggests the reference value range for uric acid are as follow:
- If blood glucose test results greater than 33.3mmol/L (600 mg/dL), the meter will display “HI”. Retest immediately with a new test strip. If your reading is still “HI”, contact healthcare professional immediately.
- If blood glucose test results low than 1.1 mmol/L(20mg/dL), the meter will display “LO”. Retest immediately with a new test strip. If your reading is still “LO”, contact healthcare professional immediately.
- If uric acid test results greater than 1188 µmol/L (20.0mg/dL), the meter will display “HI”. Retest immediately with a new test strip. If your reading is still “HI”, contact healthcare professional immediately.
- If uric acid test results low than 181 µmol/L (3.0mg/dL), the meter will display “LO”. Retest immediately with a new test strip. If your reading is still “LO”, contact healthcare professional immediately.
Note : Do not change your treatment only based on test result and light indications. Please consult healthcare professional for your treatment.
About Your Safe AQ UG Meter
Full-Screen Display
Button Functions
Setting Up Your Safe AQ UG Blood Glucose and Uric Acid meter
Setup steps for date/time
The Safe AQ UG blood glucose and uric acid meter is preset with the date and
time. You may need to adjust it to your local time zone. Check the date and
time each time the batteries are replaced. Reset the date and time if they are
not correct.
STEP 1: Insert the Batteries
Open the Battery Cover on the back side of the Safe AQ UG blood glucose and
uric acid meter. Insert two AAA alkaline batteries as indicated by the”+”
and”-” symbols.
STEP 2: Set Correcting Code of Uric Acid (use code chip)
Code chip matching with test strips shall be used when you use a new vial of
uric acid test strip. Find code chip in test strip package, check if the code
number on code chip is in accordance with code number on test strip package.
If not, you will get inaccurate test results, please call Customer Service at
+86-731-89935929. If there is already a code chip in Safe AQ UG meter, please
take it out and insert a new code chip.
Insert the code chip to code chip port, the meter will make a short beep sound
and perform self-inspection. The meter will display code number and meter will
be off after a short beep sound.
If there is an error, the meter will display E-4. Please pull out the code
chip and insert again. If still failed, please call Customer Service at
+86-731-89935929.
Every code chip only need to be inserted once, the meter will store the code
number until insert new code chip.
STEP 3: How to Set the Date and Time
STEP 4: Setting the Year
STEP 5 Setting the Month
STEP 6 Setting the Day
STEP 7 Setting the Hour
STEP 8 Setting the Minutes
Note
- 24 digital clock-hour time format can be displayed, year can be set from 2000-2099.
- The meter may be preset with time and date. If you need to adjust time and date setting or replace the battery, you need to enter the Time- Date setting mode and reset the time and date.
- Continuous decrease and Increase function: When set up Date and Time, press and hold for more than 1 second, the value will continuous decrease; press and hold for more than 1 second, the value will continuous increase.
- Press and hold ( ) button for more than 3 seconds to exit from Date and Time set -up mode.
Control Solution Testing
Why Perform a Control Solution Test
- Ensures that your meter and test strips are working properly.
- Allows you to practice testing without using blood.
When Should You Perform a Control Solution Test
- When you begin using a new vial of Test Strips.
- When the vial cap is opened for a long time.
- When Test Strips have been exposed to extreme environmental conditions.
- When you want to check whether meter or test strips are work properly.
- When you drop or broken the meter.
- When you want to check whether the test steps are correct.
Important Information
- Use only Sinocare control solution with your Meter.
- Check the expiration date on the control solution bottle. Do not use if expired.
- Use the blood glucose test strips within a period of 6 months from the date that you first open the vial.
- Use the uric acid test strips within a period of 3 months from the date that you first open the vial.
- Record the discard date on the blood glucose control solution bottle, which is 3 months from when you first open it. Discard the bottle after 3 months.
- Record the discard date on the uric acid control solution bottle, which is 2 months from when you first open it. Discard the bottle after 2 months.
- Discard any bottle that appears to be cracked or leaking.
- For in vitro diagnostic use only.
Performing a control solution test
Start with the meter off.
STEP 1 : Wash your Hands
Wash your hands with mild soap and water. Be sure to dry your hands before
performing a test.
STEP 2: Insert a Safe AQ UG Test Strip
Insert a Safe AQ UG Test Strip with printing face up and the contact end into
the test strip port until it cannot go any further. The meter will turn on. If
insert the blood glucose test strip, the meter will display and symbol is
flashing. If insert a Safe AQ UG uric acid test strip, the meter will display
the correcting code and symbol is flashing.
Attention: Only when meter display flashing blood drop symbol ” , “and
control bottle symbol, can you apply control solution. If there is no blood
drop symbol or control bottle symbol, restart the meter again or change the
test mode.
STEP 3: Apply the Control Solution
- Check the expiration date and discard dates on your control solution and Safe AQ UG test strip vials. Do not use control solution or test strip which expired.
- Shake the control solution bottle well, then remove cap. Squeeze the bottle and discard the first drop. Squeeze the bottle again to get a second drop and bring the tip of the test strip to touch the drop of solution until the meter beeps. After 5 seconds’ count-down, meter will display blood glucose control solution test result. After 25 seconds’ count-down, meter will display uric acid control solution test result.
- Control test results must be within range of the control solution level printed on test strip box. If the result is out of range, test again. If still so, do not use test strips or meter for testing. Call customer service at +86-731-89935929 or contact local dealers for help.
- Eject the test strip
Attention :
Discard used test strip and control solution according to local regulations.
Keep control solution away from children and pets.
Questionable Results
Blood Glucose and Uric Acid Testing
Prepare meter, strip, lancing device and lancet before test.
-
STEP 1: Wash your hands or use an alcohol swab to clean the fingertip. MAKE SURE YOUR HAND IS DRY BEFORE TESTING.
-
STEP 2 : Unscrew the lancing device cover.
-
STEP 3 : Insert a lancet into the lancing device until it comes to a full stop.
-
STEP 4 : Twist and pull to remove the lancet protector. Retain the lancet protector for safe lancet disposal.
-
STEP 5 : Screw the lancing device cover back on.
-
STEP 6 : Twist the adjustable comfort tip to adjust the puncture depth.
Tips:- 1-2 (least depth) for soft or thin skin,
- 3-4 (average depth) for average skin,
- 5 (deepest depth) for thick or calloused skin.
-
STEP 7: Pull the spring cover back to arm the lancing The Sinocare lancing device is prepared and ready to lance your finger for a blood sample.
The lancet is for single use only. To prevent the risk of blood borne pathogen transmission, it shall not be shared with other people. -
STEP 8 : Take a strip out of the strip vial. Immediately close the vial lid. Attention: Finish test within 3 minutes after take out strip. Otherwise test result may be inaccurate.
-
STEP 9 : Insert strip into the test strip port, The Safe AQ UG meter will beep and turn on.
The port marked with “G” is for blood glucose strip and the port marked with “U” is for uric acid strip. If blood glucose strip inserted, “GLU” will be displayed on the bottom-left of the screen, if uric acid strip inserted, “UA” will be displayed on the top-left of the screen. After test strip inserted, the meter will display flashing blood drop symbol and you can apply blood sample.
Attention :- If blood is not applied within 3 minutes, the meter will automatically turn off. Reinsert the strip and begin test.
- If the meter display “E-2”, it means the environment temperature is beyond 40 “C or below 5″C, Please put the meter, strip, control solution at appropriate place (10″C-35″C ), wait at least 30 minutes and test again.
- The test strip port must be match the inserted strip, or else the meter will not turn on.
- The meter can test blood glucose and uric acid together.
-
STEP 10 :Take Blood Sample
- Make sure the lancing device is pressed firmly to your finger. Press the button to lance.
- Attention : If you didn’t get enough blood sample, please gently massage your finger. DO NOT SQUEEZE YOUR FINGER. Refer to step 6 to adjust puncture depth and lance again.
-
STEP 11 : Apply blood sample
- When the meter screen displays a flashing blood drop symbol” “, apply the blood sample to the reaction chamber
- Apply blood sample only the blood drop symbol ” ” flashing.
- Before applying blood sample, the disinfection ethanol in the puncture site should completely volatilize.
- Hold the tip of the test strip to the blood drop until the meter beeps. Visually confirm that the blood sample fills the entire reaction chamber at the end of the test strip.
- Do not move meter or strip and press main button when testing.
- If you didn’t apply enough blood sample to test strip, use a new test strip to test again.
- Do not use test strip press finger firmly, otherwise it may cause inaccurate test result. Caution:
- DO NOT smear or scrape the blood onto the Test Strip
- DO NOT apply blood to the Test Strip when the Test Strip is not in the test strip port. DO NOT put blood or foreign objects into the test strip port.
-
STEP 12: Read the result
Blood glucose testing: After a 5-second count down, the meter displays the test result in mg/dL or mmol/L . Uric acid testing: After a 25-second count down, the meter displays the test result in mg/dL or µmol/L. -
STEP 13 : Eject the strip.
-
STEP 14 : Remove the lancing device cover. With the lancet protector on the table, stick the needle of the lancet into the protector. Eject the lancet and screw the lancet cover back on.
Used test strip and lancet may be biohazardous. Please discard them carefully
according to local regulation requirements.
If the meter is used by multiple people, use 75% alcohol to clean meter
surface after testing to avoid pathogen infection.
Viewing test result
After the meter counts down from 5, your blood glucose results appear along
with the unit of measure. After the meter counts down from 25, your uric acid
results appear along with the unit of measure. The result is stored in the
meter memory. Turn the meter off by removing the test strip. Discard the used
test strip carefully to avoid contamination.
If blood glucose test results greater than 33.3mmol/L (600 mg/dL), the meter will display “HI”. Retest immediately with a new test strip. If your reading is still “HI”, contact healthcare professional immediately. If blood glucose test results low than 1.1mmol/L (20 mg/dL), the meter will display “LO”. Retest immediately with a new test strip. If your reading is still “LO”, contact healthcare professional immediately. If uric acid test results greater than 1188 umol/L (20.0 mg/dL), the meter will display “HI”. Retest immediately with a new test strip. If your reading is still “HI”, contact healthcare professional immediately. If uric acid test results low than 181 umol/L (3.0 mg/dL), the meter will display “LO”. Retest immediately with a new test strip. If your reading is still “LO”, contact healthcare professional immediately.
Memory
The Safe AQ UG meter stores the 600 most recent results (500 blood glucose
test results and 100 uric acid test results) with the date and time in its
memory. You can review the individual results by entering the memory mode.
Note: You cannot review your test results when the test strip is in the test strip port.
Step1 : Screening Memory
Step2: Delete Memory
Note: You cannot review your test results when the test strip is in the test strip port.
Press ( ) button 3 seconds to turn the meter off.
Information for Healthcare Professional Use
- Only trained healthcare professional be allowed to use Safe AQ UG blood glucose and uric acid monitoring system on multiple patients.
- Only trained healthcare professional be allowed to collect venous blood samples.
- Any patient who has infectious diseases must use his/her own meter.
- If the same meter be used on multiple patients, there is risk of infection between both patients and healthcare professional.
- Water or other disinfection solution which remains on skin may dilute blood and cause inaccurate test results.
- Please dispose used lancet, test strips and expired control solution as medical waste.
Perform test
- Wear disposable gloves.
- Healthcare professional use an alcohol swab to clean lance site. Then be sure dry stick site before testing.
- Use qualified lancing device and choose suitable penetrate depth for patients.
- Using lancing device to lance.
- Hold the tip of the test strip to the blood sample until the test strip reaction zone is full and the meter begin countdown.
- Wash hands thoroughly with soap and water after handling the meter, lancing device or test strips.
- Cleaning and disinfecting meter and lancing device before use on next patient.
Comparing Meter with Laboratory Results
When comparing results between Safe AQ UG blood glucose and uric acid
monitoring system and a laboratory system, Safe AQ UG blood glucose and uric
acid monitoring system blood tests should be performed within 30 minutes of a
laboratory test.
Cleaning and Disinfecting
Clean and disinfect meter immediately after getting any blood on the meter or
if meter is dirty. If the meter is being used on second person, the meter and
lancing device should be clean and disinfected first. Do not ciean the meter
during the test.
To clean the meter:
- Wash hands thoroughly with soap and water.
- Make sure meter is off and a test strip is not inserted. Use 75% alcohol to rub the entire outside of the meter. Make sure no liquids enter the test strip port or other opening in the meter.
- Let meter air dry thoroughly before using to test.
- Wash hands thoroughly again after handing meter.
- Verify the meter is working properly by performing control solution test. If your meter can’t work properly, please contact customer service at +86-731-89935929 or contact local dealers.
To disinfect the meter:
Please according to clinic institution disinfect method to disinfect the
meter.
Power Saving
If there is no blood applied to the test strip within 3 minutes, the meter
will automatically turn off. After test, the test results will be displayed on
the screen with test strip in strip port, If there no operation on the meter
within 3 minutes, the meter will automatically turn off.
If there is no other operation on the meter within 1 minute, such as enter
into test results review mode, the meter will automatically turn off.
Performance Characteristics
Accuracy: 95%test result of glucose meet the requirement below:
Concen t rationRange | Bias% |
---|---|
< **5 . 5 mmol/ L (100 mg/dl)** | Within ± |
0.83 mmo l /L(± 15mg/d l )
2:5.5 mmol/L(100mg/dl)| Within± 15%
95% test result of uric acid meet the requirements below:
Concentration Range | Bias% |
---|---|
<297 µmol/L(5 mg/dl ) | :S ± 59.4 µmol/ L (1 mg/dl) |
> 297 µmol/L(5 mg/dl ) | :St20% |
Precision: test result of glucose meet the requirement below:
Concentration Range | Requirement |
---|---|
< 5.5 mmol/L(100 mg/dL) | SD< 0.34mmol/L(6 mg/dL) |
2:5 . 5 mmol/ L (100mg/d l ) | CV **< |
**6.0%
Test result of uric acid meet the requirements below
Co n centrationRange | Requiremen t |
---|---|
:S297 µmol/ L (5 mg/d l ) | SD < 22.2 |
µmol/L(0.37mg/dl )
> **297 µmo l / L (5 mg/d l )| CV <
**7 . 5%
Details please refer to blood glucose and uric acid package insert.
Specifications
Blood volume for blood glucose | About 0.6µL |
---|---|
Blood volume for uric acid | About3µL |
Sample type | Capillary whole blood, Venous whole blood |
Calibration | Plasma equivalent |
Measuring time for blood glucose | 5±1s |
Measuringtime for uric acid | 25±1s |
Meter storage/transportation conditions | -20°C- 55°C |
Dimension | 1086622 (mm) |
--- | --- |
Weight | About100g |
Power source | 3V DC, 10mA, 2 AAA alkaline batteries |
Battery life | perform up to 1,000 tests |
Display | LCD |
Memory | 500 blood glucose test results with date and time 100 uric acid test |
results with date and time
Operating condition for blood glucose| Refer to the user manual of blood
glucose test strip.
Operating condition for uric acid| Refer to the user manual of uric acid test
strip.
Construction| Hand-held
Measurement units for blood glucose| mg/dl or mmol/L
Measurement units for uric acid| mg/dl or µmol/L
Measurement range for blood glucose| 20-600 mg/dl or 1.1-33.3 mmol/L
Measurement range for uric acid| 3.0-20.0 mg/dl or 181-1188 µmol/L
Software Version| V01
Additional features
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time count-down
Temperature Warning
Battery power Warning
Used test strips Warning
Maintenance
- Use of this instrument in a dry environment, especially if synthetic materials are present (synthetic clothing, carpets etc.) may cause damaging electrostatic discharges that may cause erroneous results.
- Electromagnetic Compatibility (EMC): The meter complies with the electromagnetic requirements IEC 61326-2-6: 2012 specified in ISO 15197-2013. Electromagnetic emissions are low and unlikely to interfere with other nearby electronic equipment, nor are emissions from nearby electronic equipment likely to interfere with the meter. Immunity to electrostatic is charge meets the requirements of IEC 61326-2-6: 2012. The meter has been tested for radio frequency interference at the frequency range and test levels specified by IEC 61326-2-6: 2012.
- Do not use this instrument in close proximity to sources of strong electromagnetic radiation, as these may interfere with the proper operation.
- Avoid dirt, dust, blood, control solution or liquids into the meter test port.
- Do not keep meter in an area where it may be crushed.
- Store blood glucose monitoring system (meter, test strip, control solution) in a dry place.
- Do not freeze.
- Do not store in the kitchen or bathroom.
Changing the batteries
Make sure your meter is off when you change the batteries.
Caution: Keep batteries away from children. If the battery is swallowed call healthcare professional immediately.
- Step 1 : Slide the battery cover out of the meter.
- Step 2 : Remove the old batteries. Place the new ones in the drawer according to the inner compartment directs of”+” and “-” side.
- Step 3 : Slide the battery cover back into position until it locks into place.
Troubleshooting
Display | What it means | What to do |
---|---|---|
**** Meter does not turn on after inserting a test strip | Battery | |
incorrectly installed | Check that the battery is correctly installed. Pay |
attention to the “+”sign
Battery power is low| Replace battery
Test strip inserted incorrectly| Insert the test strips with the arrow side
facing up until it will go no further
There is no feedback on the display| Remove the battery and wait at least 3
minutes, then reinstall the battery. If the malfunction has not been solved,
please contact the manufacturer or the dealer.
Test strip Inserts into wrong ports| Make sure the right port when insert the
test strips.
The blood glucose test strip insert into port marked with “G”.
The uric acid test strip insert into port marked with “U”
Poor contact with Insert port| Please contact the manufacturer or the dealer.
Display| What it means| What to do
---|---|---
The meter does not begin to test after applying sample| Insufficient sample|
Repeat the test with a new test strip and enough blood sample
Sample is applied to wrong site of the test strip| Apply the sample into test
strip according to the instruction of “Applying sample”
Used test strip| Use new test strip
Meter displays incorrect test result| The test strip is exposed in air too
long after taking out from the vial| Finish the test within 3 minutes after
taking out the strips from the vial
Press the strip too firmly against your finger, sample cannot be smoothly
applied.| Gently touch the blood drop to the top edge of the test strip.
The test strip foil pouch is broken or damaged or left open to air. The test
strips in vial is beyond the expiration or the discard date.| Use new test
strip
Display| What it means| What to do
---|---|---
Meter displays incorrect test result| Expired test strip| Use new test strip
Defective meter or test strip| Contact the manufacturer or the dealer
false test mode| choose the right test mode
without sample applied to, test starts and meter counts down as soon as the
test strip is inserted in| The test strip is exposed in moisture environment
for too long| use a new test strip
The test strip is exposed in moisture environment for too long| use a new test
strip
If you follow the actions recommended but the problem is not resolved, please contact Customer Service at +86-731-89935929 or contact your local dealers for assistance.
Explanation of symbols
Warranty
Changsha Sinocare Inc. provides the following Warranty to the original purchaser of the Safe AQ UG blood glucose and uric acid monitoring system:
- Changsha Sinocare Inc. warrants this meter to be free of defects in materials and workmanship from the date of purchase. If the meter does not work properly, Sinocare will replace it with a new meter or equivalent product free of charge. Failure of the meter due to abuse or use not in accordance with the instructions for use is not covered by this Warranty.
- This warranty does not include the battery supplied with the meter .
- Do not take the meter apart. This action will avoid the warranty and cause the meter to display false results.
- If you meter does not work properly, please fill in warranty card carefully. Use 75% alcohol to clean your meter and send your meter with warranty card to your local dealer.
References
- American Diabetes Association Standards of medical care in diabetes-2016.
- Larsson-Cohn U: Difference between capillary and venous blood glucose during oral glucose tolerance tests. Scand J Clin Lab Invest 36:805-808, 1976
- The Modern clinical laboratory diagnostics-test and clinical (May, 2009, version 2)
Thank you for choosing Safe AQ UG blood glucose meter, a free replacement under normal operation is provided.Please do complete the warranty card and send it back to our agent or our factory.we will inform you with information of the meter.
Thank you for supporting our product and us!
- Purchase date: ____ _
- Meter No.: ———————–
Please reserve this part.and show it when in need)
Appendix 2 Warranty card (returned to producer)
- Name : ____ Tel: ____
- Age : __ Gender:
- Purchase date: _ Meter No.: ___ _
- Zip code : ____
- Address : ___ _ Purchase place (full name of the agent):
Please complete the table correctly, we will provide you with an excellent after-service.