Sinocare 8187M51626 Blood Pressure Checking Machine Instruction Manual

May 16, 2024
Sinocare

Sinocare 8187M51626 Blood Pressure Checking Machine

IMPORTANT INFORMATION

All physical activity, excitement, stress, eating, drinking, smoking, body posture, and many other activities or factors (including taking a blood pressure measurement) will influence blood pressure value. Because of this, it is mostly unusual to obtain identical multiple blood pressure readings.
Blood pressure fluctuates continually (day and night). The highest value usually appears in the daytime and the lowest one usually at midnight.

Typically, the value begins to increase at around 3:00AM and reaches to highest level in the daytime while most people are awake and active.
Considering the above information, it is recommended that you measure your blood pressure at approximately the same time each day.

Please always relax a minimum of 1 to 1.5 minutes between measurements to allow the blood circulation in your arm to recover. You rarely obtain identical blood pressure readings each time.

  • regarding the effect of blood flow interference and resulting injury to the PATIENT caused by continuous CUFF pressure due to connection tubing kinking;
  • too frequent measurements can cause injury to the patient due to blood flow interference;
  • regarding the application of the CUFF over a wound, as this can cause further injury;
  • regarding the application of the CUFF and its pressurization on any limb where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present because of temporary interference to blood flow and could result in injury to the PATIENT;
  • regarding the information that pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring Me EQUIPMENT on the same limb;
  • regarding the need to check (for example, by observation of the limb concerned) that the operation of the AUTOMATED blood Pressure monitor does not result in prolonged impairment of the circulation of the blood of the PATIENT.

The blood pressure monitor has different features to meet different users.

  • Please kindly read the instruction manual with the relevant Model Number you purchased and notice the difference between them.
  • Please read the Instruction Manual thoroughly before using this unit and please consult your physician for specific information about your blood

Safety Information

To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual.

SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL

  • Warning: Indicates a potentially hazardous situation Which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation Which, if not
  • Caution: avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

OPERATING THE DEVICE

  • Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.

  • DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician.
    Only a physician is qualified to diagnose and treat High Blood Pressure.

  • The monitor is not intended to be a diagnostic device.

  • If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.

  • This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.

  • Those parts of the equipment that shall not be serviced or maintained while in use with a patient

  • Read all of the information in the instruction manual and other literature in the box before operating the unit.

  • This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.

  • Operate the device only as intended. Do not use the device for any other purpose.

  • Dispose of the device, components optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

  • Do not use a cellular phone near the device. It may result in an operational failure.

  • Use only BS-authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.

  • The effects of degraded sensors and electrodes, or loosened electrodes, can degrade performance or cause other problems.

  • The effects are caused by pets, pests, or children.

  • Use only 1.5V dry batteries with this device,

  • Do not use other types of batteries. This may damage the unit.

RISK OF ELECTRICAL SHOCK
A Do not plug or unplug the power cord into the electrical outlet with wet
hands.
A Do not overload power outlets. Plug the device into the appropriate
voltage outlet.

CARE AND MAINTENANCE

  • Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
  • Dore the device and the components that can, said canale.
  • Changes or modifications not approved by BSX Healthcare will void the user warranty. Do not disassemble or attempt to repair the unit or components.
  • No modification of this equipment is allowed.
  • Do not modify this equipment without the authorization of the manufacturer.
  • If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe use of the equipment.
  • Enclosure, screen, button continue contact time is less than 10 min.

BEFORE TAKING A MEASUREMENT

To ensure a reliable reading follow these recommendations:

  1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 min before taking a measurement. Rest for at least 15 min before taking the measurement.
  2. Stress raises blood pressure, avoid taking measurements during stressful times.
  3. Measurements should be taken in a quiet place.
  4. Remove tight-fitting clothing from your arm.
  5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so that the cuff is at the same level as your heart.
  6. Remain still and do not talk during the measurement.
  7. Wait 2-3 minutes between measurements, the wait time allows the arteries to return to the condition before taking the blood pressure measurement. You may need to increase the wait time depending on your physiological characteristics.

PRODUCT INTRODUCTION
Blood pressure is one of the important indicators to reflect the human blood circulatory system. The home blood pressure monitor is easy to use frequently. It is very significant to prevent cardiovascular disease, to improve the initiative of hypertension patients for treatment, and to control high blood pressure.

The automatic blood pressure monitor can measure the upper and lower pressure values simultaneously by fuzzy logic control. Personalized inflation leads to the result of value every time. It is effective in reducing the discomfort by wrongly inflating the CUFF and also in reducing the wrong measurement value, thus, it improves the accuracy of the measurement.

OPERATING PRINCIPLE
The blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading.

INTENDED USE/INDICATIONS FOR USE
The blood pressure monitor is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the arm of which the circumference includes 22 cm-42 cm. It is intended to be used in a hospital environment or at home.

CONTRAINDICATION

  • It is inappropriate for people with serious arrhythmia to use this Electronic Blood Pressure monitor.
  • This device is only intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this unit for any other purpose.
  • The monitor is not intended to be a diagnostic device. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician.

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-1

LCD Display Introductions

  1. Battery level indicator
  2. User ID symbol
  3. Pressure indicator
  4. Pulse signal symbol
  5. Memory Group Number
  6. Memory symbol
  7. Average value symbol
  8. Irregular heartbeat symbol
  9. Alarm symbol
  10. Data/time Symbol
  11. Date/time display
  12. Systolic blood pressure
  13. Blood pressure Unit
  14. Diastolic backpressure
  15. Pulse rate display

SETUP AND OPERATING PROCEDURES

  1. Remove the battery cover. pulling the hook up while pushing it in the direction of the arrow.
  2. Install 4×1.5V “AA” size batteries so the +(positive) and -(negative) polarities match the polarities match the polarities of the battery compartment as indicated.
  3. Replace the battery cover.

NOTES

  • Turn the unit off before replacing the batteries.
  • The supplied batteries may have a shorter life.

Battery level Introduction

  • In shutdown mode, Press the buttons Start/Stop /Memory /Setting, and enter the measuring pattern/recording pattern/setting pattern respectively.
  • When it displays, the system will detect the battery level and will indicate the level on the LCD. Details are as follows:
  • Battery Level Full.
  • Display the battery full icon ( ) or the battery icon ( ), which can take measurements and read the memory.
  • Battery Level Low. Indicate the battery level low symbol ( ), still can take measurements and read the memory. Please charge your battery or replace new batteries.
  • Battery Exhausted.
  • Out of battery symbol (
  • It flashes 5 times and the monitor will shut off immediately.

Use setup

User switching
In the shutdown state, short-press the “SET” button and enter the time mode. Then the Unit icon, user icon, and time icon appeared. Time and date are switched every three seconds. In the time mode, short press the setup button again to switch the user. Long press setup without response. Press the “Start” button or shut down automatically after 60 minutes, and the settings will be saved.

Date, time, and clock setup

  • In the shutdown state, long press the “SET” button to enter the date adjustment, press the “MEM” button to enter the year adjustment, then press the “SET” button the month flickers, and then enter the month adjustment state.
  • Press the “MEM” button to adjust the month, then press the “SET” button the date flickers, and then enter the date setup state.
  • Press the “MEM” button to adjust the date, then press the “SET” button the time icon flickers, and then enter the hour system setup state.
  • Press the “MEM” button to adjust the hour system, then press the “SET” button the hour flickers, and then enter the hour setup state. Afternoon displays PM (if set as a 24-hour system, then afternoon does not display PM).
  • Press the “MEM” button to adjust the hour, then press the “SET” button the minute flickers, and then enter the minute setup state.
  • Press the “MEM” button to adjust the minute, then the minute setup is completed
  • Press the “SET” button and enter the clock setup state, the screen displays AL and user number, and the hour flickers, press the “MEM” button to adjust the hour, press the setup button again, and the minute flickers, press the “MEM” button to adjust the minute, press the “SET” button the clock symbol flickers, press the “MEM” button to adjust whether to enable the clock.
  • Please note that each user can set a clock, and the lock at this time will correspond to that of the user selected previously. When the clock sounds, press any button to exit.
  • Press the setup button again, then shut down and save the setting results.
  • At this time, manual shutdown or automatic shutdown after 60 seconds will save the results

Measurement unit switching
In the shutdown state, long press the “Start” button for 15 seconds. If the original unit is mmHg (millimeter of mercury), after a long press, it will be switched to kPa (kilopascal), and if the high-pressure zone of the screen displays 0.0 kPa, it indicates a successful setting. Press the shutdown button to shut down and save the settings.

Mute Setting
Under Off state, long press the “SET” button until the Year flashes, Press the “SET” button to switch to the SP, press the MEM button to switch to OF, press the “Start” button to switch off the monitor, and when switched on again, there will be no voice broadcast.

Read the Memory

  • For the first time, press the “MEM” button, AVG and MR icons are lighted, and display the average value of the first 3 times (when the number of memory is greater than or equal to 3 groups);
  • For the second time, press the “MEM” button, the MR icon is lit, then the measurement records can be viewed, and long press the “MEM” button can quickly view the memory.

Memory deletion

  • Because of the memory, while holding down the “MEM” button, continue to hold down the “Start” button for 2 seconds, and when the screen is fully displayed as -, then the memory is deleted.
  • Memory deletion is a one-time removal of all measurement data under all users.

Entering pressure measurement method

  • a. Short press the “SET” button hold down the “SET” button then hold down the “MEM” button immediately, and after about 1s, the low-pressure display area of the screen shows figure 0, and enter the static pressure measurement mode after releasing the button.
  • b. After entering the static pressure measurement mode, the short press / long press “Start” button can exit the static pressure measurement mode, and return to the time mode.

BLOOD PRESSURE MEASUREMENT METHOD

  1. Applying the cuff on the arm
    • Make sure the air plug is securely inserted into the main unit.
    • Remove tight-fitting clothing from your upper left arm.
    • Push open the cuff into a tubular shape, until it is suitable for the arm to get in.
    • Wrap the cuff around the upper arm, and also make the air duct located on the forearm.
    • The bottom of the cuff shall be located at the inner side of the upper arm elbow joint, going upward for 2cm-3cm.
    • Tightly wrap the cuff, and fix it with a cloth buckle. Leave a finger gap between the arm and the cuff.
    • Align the air duct plug of the cuff with the air duct interface on the left side of the machine body, and insert it to the end.

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-8

Note:

  • A. Please refer to the CUFF circumference range in “SPECIFICATIONS” to make sure that the appropriate CUFF is used.
    • Please keep the CUFF clean. If the CUFF becomes dirty, remove it from the monitor and clear it by hand in a mild detergent, then rinse it thoroughly in cold water.
    • Never dry the CUFF in the clothes dryer or iron it. Clean the CUFF after the usage of every 200 times is recommended.
    • Do not place the CUFF around your arm if the arm has any inflammation, acute diseases, infections skin wounds.
  • An explanation of the operating steps needed to obtain accurate routine resting BLOOD PRESSURE measurements for the condition hypertension including:
    • PATIENT position in NORMAL USE, including
    • comfortably seated
    • legs uncrossed
    • feet flat on the floor
    • back and arm supported
    • middle of the CUFF at the level of the right atrium of the heart.
    • a recommendation that the PATIENT relax as much as possible and not talk during the measurement PROCEDURE;
    • a recommendation that 5 min should elapse before the first reading is taken;
    • Measuring on the same arm each time.
    • OPERATOR position in NORMAL USE;

A Warning:

  • Strangulation due to cables and hoses, particularly due to excessive length;
  • potential allergic reactions to accessible materials used in the ME EQUIPMENT;
  • contact injuries.

Follow the instructions of your physician or licensed healthcare provider.

Blood pressure measurement

  1. In the state of shutdown or standby, press the [Start / Stop] for start-up, it will start measurement automatically, and the CUFF will inflate automaticallySinocare-8187M51626-Blood-Pressure-Checking-Machine-FIG-10When the CUFF starts to inflate, the blood Pressure monitor will start to work automatically. In the process of inflation, the blood pressure monitor will also detect the pulse, so please do not move your arm in the whole measurement process, and remain static throughout the process. The arm will feel the CUFF extrusion, and this is a normal phenomenon of inflation.
    If there is any error, please refer to the “Error and fault instructions” in section VIII for troubleshooting.

  2. Stop automatically after full inflation, and start to detect the pulse rate.Sinocare-8187M51626-Blood-Pressure-Checking-Machine-FIG-11
    Note: The pictures in the measurement interface are only for reference. After the measurement is completed, the blood Pressure monitor will display your blood pressure valve and pulse number, and if the arrhythmia is measured, then the arrhythmia symbol will be displayed.

  3. If the monitor has stored results, the LCD will momentarily display the most recent one. If no result has been stored, ‘-” will appear on LCD.

  4. After the measurement is completed, take off the CUFF, press the [Start / Stop) button, and shut down manually. If no button is pressed, then the system will be shut down automatically after 60 seconds.
    Note: Please consult a health care professional for interpretation of pressure measurements. The pictures in the measurement interface are only for reference.

ERROR AND FAULT TROUBLESHOOTING

Troubleshooting tips

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-15

Error indicators

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-16

If the problem is still not resolved, please contact the dealer or our customer service department, to obtain further assistance.

NOTICE

  1. The interval between two measurements shall be at least 2-3 minutes.
    The waiting between two measurements is mainly to get the artery back to the state before the blood pressure is measured.

  2. When in repeated measurement, because the arm blood flow may not be smooth, you may not measure the correct blood pressure. Please conduct measurements after the blood flow is smooth.

  3. When the display screen shows “Err”, the measurement cannot be conducted correctly, please refer to the error and fault instructions in the next page.

  4. When the measured personnel has arrhythmia, the measurement value of blood pressure and pulse rate may not be accurate enough, so for patients with arrhythmia, it is advised to consult a doctor to make a professional diagnosis.

  5. When the CUFF is too loose or too tight, it will affect the accuracy of measurement.

  6. Measurement posture and physical condition etc., will affect the measurement results. Please understand the knowledge of blood pressure, and eliminate all kinds of interference factors.

  7. If the CUFF not provided by the manufacturer is used to replace the original one, it may lead to measurement error. Please do not use the CUFF other than supplied by the manufacturer.

  8. Please do not use it together with the electrical appliances to emit electromagnetic interference, it may affect the accuracy.

  9. The button component list of the machine includes the main control chip, sensor, pump, electromagnetic valve, uniform speed air release valve and CUFF. Parts of the manufacturer’s models and specifications specified in the list must be used; otherwise, it may lead to problems of safety or accuracy.

  10. The CUFF should be placed at the same level as your heart.

  11. During measurement, neither speak nor move your body and arm.

  12. Use on same arm for each measurement.

  13. This Electronic Blood Pressure monitor is designed for adults and should never be used on infants or young children. Consult your physician or other health care professionals before use on older children.

  14. For information regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION.

  15. The monitor might not meet its performance specifications or cause safety hazards if stored or used outside the specified temperature and humidity ranges in the specifications.

  16. Please do not share the CUFF with another person to avoid cross-infection.

  17. Please note that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

MAINTENANCE

REPAIR AND CALIBRATION

  1. When not in use, please keep the blood pressure monitor and its accessories in the packing box. If you do not use the monitor for a long time, please remove the batteries.

  2. To prevent the blood Pressure monitor from being damaged, please comply with the following items, and improper use will result in the loss of warranty services:

    • Please do not place the blood pressure monitor and the CUFF in an environment of high temperature, dampness, full of water vapor, or direct sunlight. Please refer to the specification table for details. If stored and used outside the specified range of temperature and humidity, the system may not be able to achieve the claimed specifications.
    • Please do not firmly fold the CUFF or the air duct.
    • Please do not disassemble the Blood Pressure monitor.
    • Please do not allow the blood pressure meter to be subjected to strong impact or vibration (for example, dropping the blood pressure monitor on the ground).
    • Please do not use volatile liquid to clean the blood Pressure monitor. Please use a soft and dry cloth to clean the outer shell and battery.
    • Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile. Do not attempt to clean the cuff.
    • Please do not use water to clean or moisten the CUFF, and when necessary, absorbent cotton dipped with 75% medical alcohol can be used to wipe it gently, to conduct disinfection.
    • Drop, collision bending, etc. may reduce the performance of the battery; if the battery is found abnormal, such as packaging damage and battery pack deformation etc, do not use these batteries.
    • Please do not place the blood pressure monitor and the CUFF in an environment of high temperature, dampness, full of water vapor, or direct sunlight. Please refer to the specification table for details. If stored and used outside the specified range of temperature and humidity, the system may not be able to achieve the claimed specifications.
    • Please do not firmly fold the CUFF or the air duct.
    • Please do not disassemble the Blood Pressure monitor.
    • Please do not allow the blood pressure meter to be subjected to strong impact or vibration (for example, drop the blood pressure monitor on the ground).
    • Please do not use volatile liquid to clean the blood Pressure monitor. Please use a soft and dry cloth to clean the outer shell and battery.
    • Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile. Do not attempt to clean the cuff.
    • Please do not use water to clean or moisten the CUFF, and when necessary, absorbent cotton dipped with 75% medical alcohol can be used to wipe it gently, to conduct disinfection.
    • Drop, collision bending, etc. may reduce the performance of the battery; if the battery is found abnormal, such as packaging damage and battery pack deformation etc, do not use these batteries.
    • The manufacturer will make available on request circuit diagrams, part lists, and descriptions.
    • The patient can perform changing batteries.
  3. Calibration instructions, or other information that will assist service personnel to repair those parts of my equipment that are designated by the manufacturer as repairable by service personnel.

  4. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements.

  5. For customer service: Shenzhen BS Technology Electronics Co., Ltd.
    For warranty service, please contact BSX or European
    The representative listed on the instructions

  6. Calibration
    It is recommended the performance should be checked every 1 year or after repair. Please contact the service center.

CLEANING AND DISINFECTION

Cleaning:
Recommended cleaning agent: Household-neutral cleaning agent.

Cleaning steps:

  1. Remove the batteries before cleaning.
  2. The body of the Blood Pressure Monitor can be cleaned with a clean soft cloth.
  3. The body can be wiped with a clean soft cloth stained with a small amount of neutral detergent or water.
    • Do not use any corrosive cleaning agent. When cleaning, be careful not to immerse any part of the Electronic Blood Pressure Monitor to avoid liquid flow into the instrument.
    • It is suggested to clean once every week. Complete the cleaning in 3min each time. The number of repeated cleanings each time shall not exceed 3 times

Disinfection:
Recommended disinfecting agent: Isopropanol solution with 70% concentration

Disinfection steps

  1. Wipe the machine body with a clean soft cloth stained with a small amount of the above disinfectant, and dry it immediately.
  2. The body of the Electronic Blood Pressure Monitor can be wiped with a clean cloth stained with a small amount of 75% medical alcohol for disinfection.
    • Do not disinfect by high-temperature steam or ultraviolet irradiation, which might damage the instrument or accelerate its aging!
    • It is suggested to disinfect the Electronic Blood Pressure Monitor before and after use each time. Each time disinfection shall be completed within 1min. The number of repeated disinfection each time shall not exceed 2 times.

STATEMENT

  1. The electronic blood Pressure monitor products manufactured by our company have passed clinical verification, and their safety and effectiveness are in line with the relevant national regulatory requirements.
  2. The arm-type electronic blood Pressure monitor products manufactured by our company have been conducted contrast test with mercury column blood Pressure monitor, the blood pressure value measured by this equipment is equivalent to the measurement value of the auscultatory method and its error meets the specified requirements of IEC 80601-2-30, which is within the acceptable range.

DISPOSAL TREATMENT STATEMENT
The product manufacturing date can be seen in the qualification certificate. and the product’s estimated life is 5 years.

If electronic products and batteries are directly thrown into the trash can, they can cause harm.

CAUTION: Use of the device, components, and optional accessories according to applicable local regulations. Illegal disposal may cause environmental pollution.
If electronic products and batteries are directly thrown into the trash can, they can cause harm to the environment. At the end of the blood pressure monitor, please do not throw it into the trash can, and please dispose of it according to the local law

ELECTROMAGNETIC COMPATIBILITY WARNING AND STATEMENT

Notes
The equipment complies with the relevant electromagnetic compatibility requirements of EN 60601-1-2; Users shall conduct installation and use according to the electromagnetic compatibility information provided by random files; Portable and mobile RF communication equipment may affect the performance of the device, so strong electromagnetic interference shall be avoided when in use, such as close to mobile phone and microwave oven etc.

A The equipment shall not be close to or stacked with other equipment when in use, and if it is necessary to close to or stack with other equipment when in use, it shall be observed and verified whether it can run normally under its configuration;

  • Except for the cable sold by the equipment manufacturer as the spare parts of internal components, the use of unspecified accessories and cables may lead to emission increases or anti-interference reduction of the equipment.
  • potential allergic reactions to accessible materials used in the me equipment.

Guidance and manufacturer’s declaration – electromagnetic emissions
The model BSX516 is intended for use in the electromagnetic environment specified below. The customer or the user of the model BSX516 should ensure that it is used in such an environment.

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-17

The model BSX516 is intended for use in the electromagnetic environment specified below. The customer or the user of the model BSX516 should ensure that they are used in such anervironment

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-18

The model BSX516 is intended for use in the electromagnetic environment specified below. The customer or the user of the model BSX516 should assure that it is used in such an environment

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-19Sinocare-8187M51626-Blood-
Pressure-Checking-Machine-FIG-20

NOTE 1 At 80 MHz and 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered If the measured field strength in the location in which the model BSX516 is used exceeds the applicable RF compliance level above, the model BSX516 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model BSX516.

  • Over the frequency range of 150 kHz to 80 MHz. field strength should be less than 3Vim.

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-21

  • a) For some services, only the uplink frequencies are included.
  • b) The carrier shall be modulated using a 50% duty cycle square wave signal.
  • c) As an alternative to FM modulation, 50% pulse modulation at 18Hz may be used because while it does not represent actual modulation, it would be the worst case.

Recommended separation distances between portable and mobile RF communications equipment and the model BSX516.
The model BSX516 is intended for use in an electromagnetic environment in which radiated RF disturbances is controlled. The customer or the user of the model BSX516 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model BSX516 as recommended below, according to the maximum output power of the communications equipment.

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer.

  • NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.
  • NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

DISPLAY SYMBOLS

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-14

PRODUCT SPECIFICATION

Sinocare-8187M51626-Blood-Pressure-Checking-Machine-
FIG-22

If there is any performance improvement, the above specifications are subject to change without notice.

  • Manufacturer: Shenzhen BSX Technology Electronics Co., Ltd.
  • Address: Rm101&2/F-4/F, Building No. 13, Ailian Industrial Park, Wulian Community, Longgang District, Shenzhen
  • Telephone: +86-755-28719103
  • Email:[email protected]
  • FAX: +86-755-28882567

Authorized European Representative:
Company Name: MedNet EC-REP GmbH
Address: Borkstrasse 10,48163 Münster, Germany
Phone: +49 251 32266-61
Fax: +49 251 32266-22
Email: [email protected]

Please thoroughly read the instruction Manual before using the unit.
Please save this instruction Manual for further reference

  • Version: A.O
  • REF No.: 8187M51601
  • Revision Date: 2017.11.01

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals