**TROLEX TX6377 Sentry Transportable Sensor Station
**
TO REPLACE AN EMODULE
Checkpoint
Sentra modules can be replaced underground even when the Sentry is powered up and in use.
-
On the front half of the Sentry use a cross head screwdriver and remove the six screws securing the front panel of the Sentry. There are three screws at the top and three at the bottom.
-
Swing the front panel out of the way. Identify the module(s) to be replaced.
-
The module locations are shown in the illustration below.
-
Using a flat blade screwdriver turn the module retaining screw anti-clockwise to release the defective module. Repeat for all modules being replaced.
-
Remove all defective eModule(s) from the module bay.
-
Insert the replacement eModule(s) into the module bay.
-
Ensure the correct module is fitted in the correct location, see illustration above. The coding stops prevent the wrong module being fitted in the wrong module bay.
-
Ensure the connector is fully engaged and the seal on the top of the module is intact and firmly attached. Repeat for all affected modules.
-
Using a flat blade screwdriver, turn the module retaining screw clockwise to secure the module. Repeat for all replaced modules.
-
After fitting the modules close the Sentry front panel and secure using the six captive screws and a cross head screwdriver.
Checkpoint
Tighten the six screws evenly to avoid the possibility of distorting the Sentry front panel. DO NOT OVER-TIGHTEN the six screws. -
Ensure the Sentry is switched on.
-
Wait two minutes for the background gas concentration readings to stabilise.
-
Continue normal operation of the Sentry including testing the outside gas concentration when required.