WHP3000 1200W Weller Hot Plate Instruction Manual

June 15, 2024
Weller

WHP3000 1200W Weller Hot Plate

Product Specifications

Power Heating Zone Display Control Accessories
200 / 600 W 600 / 1200 W 3-digit 7-segment display Remote control, High

Power button, UP button, DOWN button,
Mains switch, Optical indication of the state of regulation| External sensor

Product Usage Instructions

Placement in Operation
  1. Connect the product to a power source.
  2. Turn on the hands switch located on the product.
Operation with External Sensor

To use the product with an external sensor, follow these steps:

  1. Connect the external sensor to the product.
  2. Make sure the sensor is properly attached and secured.
  3. Turn on the product using the hand switch.
  4. The product will now display the temperature readings from the external sensor.

Error Indications on the Display
If there are any errors or issues with the product, they will be indicated on the display. Refer to the user manual for a list of error codes and their meanings.

Other Power On Routines

The product has additional power on routines that can be accessed by following these steps:

  1. Press and hold the HIGH POWER button for 5 seconds.
  2. The product will enter a different power on mode.
  3. Refer to the user manual for more information on the different power on modes and their functions.

Accessories

The product comes with the following accessories:

  • External sensor

Items Supplied

The product is supplied with the following items:

  • Product unit
  • External sensor

Frequently Asked Questions

Q: How much power does the product consume?
A: The product has two power settings: 200 W and 600 W.

Q: What are the heating zone options?
A: The product has two heating zones: 600 W and 1200 W.

Q: Can I use an external sensor with the product?
A: Yes, the product is compatible with an external sensor. Refer to the user manual for instructions on how to connect and use the external sensor.

Q: How do I access the different power-on modes?
A: To access the different power modes, press and hold the HIGH POWER button for 5 seconds. Refer to the user manual for more information on the different power-on modes and their functions.

Overview

  1. Remote LED (external control via RS232)
  2. High Power LED (large 1200 W / small 600 W heating zone)
  3. Indication, external sensor regulation
  4. Display (3-digit 7-segment display)
  5. “UP” button
  6. Mains switch
  7. “DOWN” button
  8. Optical indication of the state of regulation
  9. “HIGH POWER” button (switchover large 1200 W small 600 W heating zone)

WHP 3000

Quick Reference WHP 3000

  • The UP button increases the value
  • The DOWN button decreases the value
  • Change over small/large heating zone
  • Manual heating shut down (Off)
  • Automatic heating shut down (Auto Off)

Key function via “Power-On Routine”

  • Standby function ON / OFF
  • Change over °C / °F
  • Resetting to the factory setting (FSE)

We thank for the confidence you have shown by purchasing the Weller Heating Plate WHP 3000. During manufacture the strictest quality requirements are applied; these assure the correct function of the device and make it possible to obtain optimal soldering results.

Attention!
Prior to placing the device in operation, please carefully read these operating instructions and the safety instructions enclosed. If the safety instructions are not observed, there is a risk of injury. The manufacturer accepts no liability for usage other than that described in the operating instructions or for unauthorised modifications The WELLER heating plate WHP 3000 complies with the EU declaration of conformity as per the essential safety require-ments in the directives 2004/108/EU, 2006/95/EU and 2011/65/EU (RoHS).

Description

The WHP 3000 heating plate is equipped with 3 infrared high-temperature lamps and enables electronic assemblies to be pre-heated in numerous ways. The high- temperature lamps emit radiation primarily in the wavelength range of 2 – 10 μm and heat modern materials rapidly and efficiently. Digital regulation electronics ensure precise temperature behaviour and support various special functions such as “AUTO OFF” or standby temperature. Setpoints and actual values are indicated digitally. Two different size heating zones are available. Using an optional external sensor, the temperature can be regulated at defined measuring points. An integrated RS232 interface enables the device to be controlled externally from the Weller WHA 3000P / WHA 3000V hot air station. Here the WHP 3000 heating plate is used as a bottom heater in a 3-step temperature-time profile.

Technical data

  • Dimensions:
    • 254 x 395 x 70 mm
  • (W x L x H)
    • 10 x 15,55 x 2,75 inch
  • Mains voltage:
    • 230 V (120 V); 50 Hz (60 Hz)
  • Power:
    • T0053338699
    • small heating zone 600 W
    • T0053364699
    • large heating zone 1200 W
  • Temperature range:
    • 50°C – 400°C (150°F – 750°F)
  • Protection class:
    • 1

Placing in operation

Remove all temperature-sensitive and flammable objects from the vicinity of the heating plate. Ensure that the heating plate is switched off. Ensure that the mains voltage is correct. Connect the device to the mains (12). Switch on the device at the main switch (6). When the device is switched on, a self-test is performed during which all display elements (4) are operated. The temperature set (setpoint) and the tem-perature scale (°C / °F) are then displayed briefly. The display then switches automatically to the indication of the actual value. The red dot on the display illuminates (8). This dot is a visual indication of the state of the regulation. Continuous illumination indicates the system is warming up. Flashing indicates that the operating temperature has been reached.

Adjusting temperature

The digital display (4) normally indicates the actual tempe-rature. The digital display (4) switches to the current setpoint when the ”UP” or “DOWN” button (5)(7) is pressed. The setpoint (flashing indication) can now be changed as required by pressing or pressing and holding the “UP” or “DOWN” button (5)(7). If the button is pressed and held down, the set point changes quickly. Approx. 2 sec. after the button is released, the digital display (4) automatically switches back to the actual value.

Switching over between large/small heating zones

Two different power settings and sizes of the active heating surface are available. Switchover is performed by pressing the HIGH POWER button (9).

Small heating zone:

Large heating zone:

Note:
The heating zone should be selected prior to use (when the device is cold). Switching over from the small to the large heating zone when the device has reached a steady-state temperature will result in long warm-up times for the additional lamps.

Manual heating shut down (OFF)

The device heating is shut down by simultaneously pressing the “UP” and “DOWN” buttons. “Off” appears on the display (4). If the standby function is also active, the temperature is reduced to 100°C (212°F). “Stb” appears on the display (4).

Automatic heating shut down (AUTO OFF function)

The auto-off time for the heating shut down is displayed flashing by pressing and holding (approx. 3 sec.) the HIGH POWER button (2). The shut-down time can be adjusted in 5-minute steps in the range of 5 – 600 min by pressing the “UP” or “DOWN” button (5)(7). A setting of less than 5 min switches off the automatic heating shut-down and “OFF” appears on the display (4). If the standby function is also active, the temperature is reduced to 100°C (212°F). “Stb” appears on the display (4).

Operation using RS232 serial interface

When operated together with the WHA 3000P / WHA 3000V hot air station, the WHP 3000 heating plate is controlled via the RS232 serial interface (11). Here the heating plate is used as a bottom heater for electronic assemblies and is integrated into a 3-step temperature-time profile. When used in automatic mode (Remote LED (1) illuminated) it is not possible to make any entries directly at the heating plate. Only the switchover between the heating zones (9) remains acti-ve. The temperature setting is made via the WHA 3000P /WHA 3000V hot air station. When the program is not active, the heating plate is switched off. “OFF” appears on the display(4). If the standby function is also active, the temperature is reduced to 100°C (212°F). “Stb” appears on the display (4).

STANDBY function

In standby mode, the temperature is reduced to 100°C (212°F) if a heating shut-down occurs (using OFF, AUTO OFF, RS232). The standby mode is activated via a “Power-On Routine”. For this purpose, the device is first switched off at the mains switch (6). Press the HIGH POWER button (9) and switch on the device. Keep the button pressed until the — 11 –appears on the display (4). When the “HIGH POWER” button is released the setting is saved. The standby function is switched on. Use the same procedure for switching off. — 00 –appears on the display (4) (factory setting).

Operation with external Sensor

It is also possible to measure the control variable for the tem-perature regulation using an external sensor. If an external sensor is connected, the current temperature from the external sensor is displayed and controlled instead of the temperature of the heating lamps. The external sensor, an insulated thermocouple type K, is connected to a socket (10). LED (3) on the display (4) illuminates.

Note:
The sensor must be adequately in contact with the assembly or component for correct operation. When working with the external sensor, the temperature setting (setpoint) must be adjusted to suit the measuring point.

Error indications on the display (4)

  • – – -: No temperature sensor detected
  • E10: Maximum housing temperature exceeded

Other power-on routines

°C / °F change over
For this purpose the device is first switched off at the mains switch (6). Press the DOWN button (7) and switch on the device. Keep the button pressed until the “°F” appears on the display (4). When the “DOWN” button is released the setting is saved. Use the same procedure for the change over to “°C”.

Resetting to the factory setting (FSE)
For this purpose, the device is first switched off at the main switch (6). Press the “UP” (5) + “DOWN” (7) + “HIGH POWER” (9) button simultaneously and switch on the device. Keep buttons (5, 7, 9) pressed until “FSE” appears on the display (4). When the buttons (5, 7, 9) are released the setting is saved.

Accessories

  • T005 31 190 99: External sensor type K (0.5mm)
  • T005 31 191 99: Interface cable
  • T005 33 162 99: WBH 3000 Circuit board holder
  • T005 33 163 99: WBH 3000S Circuit board holder with stand WHA 3000
  • T005 33 346 99: WHA 3000P Hot air station
  • T005 33 366 99: WHA 3000V Hot air station

Items supplied

WHP 3000 heating plate
Mains cable
Operating instructions
Safety information

Subject to technical change without notice!
See the updated operating instructions at www.weller-tools.com.

Symbols

CE mark of conformity
British Conformity Mark

Disposal
Waste light sources have to be removed from equipment. Check with your local authority or retailer for recycling advice and collection points. According to local regulations, retailers may have an obligation to take back waste electrical and electronic equipment free of charge. Your contribution to the reuse and recycling of waste electrical and electronic equipment helps to reduce the demand for raw materials. Waste electrical and electronic equipment contain valuable, recyclable materials, which can adversely impact the environment and human health, if not disposed of in an environmentally compatible manner. Delete personal data from waste equipment, if any. Contaminated filters must be treated as special waste. Dispose of replaced equipment parts, filters or old devices in ac-cordance with the rules and regulations applicable in your country.

EC declaration of conformity

We hereby declare that the products described herein comply with the following guidelines:
SISI 20162016 No.1No.1101,101, SISI 20162016 SI 2008 No.1597, SI 2012 No.3032, SI 2016 No.1091No.1091,No.1091, SISI 20082012 No. No.15973032

Besigheim, 2023-08-28

Philippe Buidin
Managing director
Authorised to compile technical documentation.
Apex tool Group (UKAuthorised to compile technical document Operations) Limited citation. Piccadilly, Tamworth StafApex Tfordshire B78 2ER pool Group UK Ltd Registered in England, Company  Number 14127816 Registered Offi ce: 20 Farringdon Street, 8th Floor London, EC4A 4AB, United Kingdom

Weller Tools GmbH
Carl-Benz-Straße 2, 74354 Besigheim, Germany

Back View WHP 3000

  • External sensor
  • RS232 interface (external control)
  • Mains connection ****

Circuit Diagram

Product Registration

www.weller-tools.com/registration

GERMANY
Weller Tools GmbH
Carl-Benz-Straße 2
74354 Besigheim
Tel: +49 (0)7143 580-0

CHINA
Apex Tool Group
Room 302A,
NO 177 Bibo Road
Shanghai, 201202
Tel: +86 (21) 60880288
info@weller-tools.com
www.weller-tools.com

THE NETHERLANDS
Apex Tool Group B. V. Phileas Foggstraat 16 7821 AK Emmen
Tel: +31 (0) 591 66 75 00

GREAT BRITAIN
Apex Tool Group UK Ltd Registered in England,
Company Number 14127816 Registered Office:
C/O TMF Group 13th Floor, One Angel Court, London, EC2R 7HJ, United Kingdom

USA
Apex Tool Group, LLC.
Weller Professional Tools Division 1000 Lufkin Road
Apex, NC 27539
+866-498-0484

www.weller-tools.com