MASTECH MS5911 Safety Pat Plus Tester User Manual
- June 15, 2024
- MASTECH
Table of Contents
- MASTECH MS5911 Safety Pat Plus Tester
- Product Information
- Product Usage Instructions
- Introduction of Instrument **
- Safety instructions
- Description
- Features
- Operation of instrument
- Technical Specifications
- Maintenance of instrument
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
MASTECH MS5911 Safety Pat Plus Tester
Product Information
Specifications
- Measurement Category: CAT II
- Test Voltage: 300V
Description
Introduction of functions of each part
- LCD: Displays the test content and data
- F: Function key that completes combined functions with other keys
- UP: Turns pages when viewing the test results
- A : Starts the CLASS I test
- POWER: Turns on/off the machine
- B: Starts the CLASS II test
- BACKLIGHT: Turns on/off the backlight
- DOWN: Turns pages when viewing the test results
- C: Starts the CORD test
- Output socket: For the output of the test voltage during CLASS I, CLASS II, and CORD tests
- Input socket: For input of the test voltage during CORD and out-phase tests
- Input Earth Bond Lead terminal: A conductor for connecting external earth points on Class I appliances
- 12VDC power connector: A power connector when an external 12VDC power is applied
Display Screen
Indicator | Description |
---|---|
EARTHCONT | Ground Resistance Test / Reading result |
INSULATION | Insulation Resistance Test / Reading result |
LEAKAGE | Leakage Current Test / Reading result |
Low Battery Symbol | Indicates low battery |
Double Insulated / Class II Test | Indicates double insulation or Class II test |
Live – Neutral | Indicates live-neutral connection |
Live – Earth | Indicates live-earth connection |
Neutral – Earth | Indicates neutral-earth connection |
Pass Symbol | Indicates test passed |
Fail Symbol | Indicates test failed |
Product Usage Instructions
Safety Instructions
Read this manual carefully before use and follow the relevant procedures
during the operation.
Warranty
The meter is warranted to be free from defects in material and workmanship
under normal use and service. The warranty period is one year and begins on
the date of shipment. This warranty does not apply to fuses or disposable
batteries.
Safety Symbols
- Note: Important safety information, refer to the instruction manual.
- Caution: Possibility of electric shock
- Equipment protected throughout by double insulation or reinforced insulation.
- Complies with European (EU) safety standards
Operation of Instrument
-
Turn On/Off
Press the POWER button to turn on/off the machine. The instrument will automatically shut down if there is no operation within 5 minutes. Note: When the instrument displays a low battery symbol, it indicates the lack of power and it is necessary to replace the battery or use the 12V DC Power Adapter. -
Calibration
To calibrate the instrument, refer to the instruction manual for detailed procedures. -
Test Voltage Selection
The CLASS II insulation resistance test voltage is defaulted as 500VDC and can be set to 250VDC. Refer to section 6.3 in the instruction manual for specific operation instructions.
FAQ (Frequently Asked Questions)
-
Q: What is the warranty period for this product?
A: The warranty period is one year from the date of shipment. -
Q: Can I use disposable batteries with this meter?
A: No, the warranty does not cover the use of disposable batteries. -
Q: How do I know if a test has passed or failed?
A: The instrument will display a pass or fail symbol depending on the test result.
Introduction of Instrument **
**
The meter is a handheld safety tester which is mainly used for safety performance tests of household appliances and industrial equipment(otherwise know as a Portable Appliance Tester),
Main functions are as follows:
- Safety CLASS I test;
- Safety CLASS II test;
- Cord/Polarity test;
- 230V/110V Sub-Leakage test functionality
- 12V battery-powered and mains powered using 12VDC adaptor, with automatic power-off function so that it will automatically shut down if there is no operation within 5 minutes.
- It has backlight display function and will be automatically shut down within 30s.
Warning
Read this manual carefully before use and follow the relevant procedures
during the operation.
This kit includes
- Safety tester 1pcs
- CORD Main test cord 1pcs
- Alligator clip test cord (including a removable alligator clip) 1pcs
- Probe 1pcs
- AA(LR6) battery 1.5V 8pcs
- Instructions 1pcs
- Rugged padded carrying case 1pcs
Safety instructions
Warnings – Please be aware of the following warnings.
- Before using the unit and test leads inspect both items. Do not use the unit and test leads if they are damaged or the case is broken or damaged as this could result in electrical shock.
- Replace the battery as soon as the battery indicator ” ”appears. Low battery levels may produce inaccurate readings.
- Do not use or store the unit in high temperature, humid, flammable or electromagnetic environments as damage and incorrect readings may result.
- If the tester requires repair or replacement please contact your local service center.
Declaration of Conformity
This product and declares that this product conforms to the following
standards:
-
BS EN 61326
Electrical equipment for measurement, control and laboratory use -
BS EN 61010-1
Safety Requirements for electrical equipment for measurement, control, and laboratory use.
Warranty
The meter is warranted to be free from defects in material and workmanship
under normat use and service. The warranty period one year and begins on the
date of shipment Part. This warranty does not apply to fuses, disposable
batterise.
Safety Symbols
CAT II: MEASUREMENT CATEGORY II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
Description
Introduction of functions of each part
Display Screen
Features
-
Auto Switch Off
The unit will automatically switch off after approximately 5 minutes if no buttons are pressed. -
It has backlight display function and will be automatically shut down within 30s.
-
Power: AA 1.5V*8 or 12VDC adapter
-
Low voltage detection: ≤ 9.5V
-
Data: Logging 300 groups
-
AbPO: Auto Power Off
-
LBD: Low Battery Display
-
Product Size: 204mm×155mm×76mm
-
Product Weight: 1270g
Operation of instrument
Turn on / off
- Press POWER to turn on/off the machine;
- The instrument will be automatically shut down if there is no operation within 5 minutes;
- Note: When the instrument displays , this indicates the lack of power and it is necessary to replace the battery. Or to use the 12V DC Power Adapter
Calibration
When using the test leads for the 1st time or replacing the test leads,
including the IEC lead (used when testing extension cables / C Type
functionality) or Earth Bond leads these functions require recalibration of
the meter before use. (This calibration will compensate for the lead
resistance during testing)
- Calibrate the CLASS I test grounding resistance.
- Insert the Earth bond lead main into the output socket (12) with the probe supplied, insert the probe into the earth thermal and the input socket (10). Press and hold the F (2) and A (4) function key together until the calibration interface is enabled.
- Press the A (4) key this will then perform the lead calibration/ adjustment, Press the F (2) to exit the calibration process.
- Calibrate the CORD/IEC test grounding resistance.
Insert the CORD/IEC main test cord into the output socket (10) and the other end into input socket (11). Press the POWER key to turn on the instrument, and press and hold the combination key F + A until entering the calibration interface. Press the C key to calibrate the CORD test grounding resistance; after the calibration is completed, the resistance of the current test cord (line) will be displayed. Press the F key to exit.
Test Voltage selection
In addition to selecting the output voltage (250V/500V) of the test insulation
resistance in the CLASS I/II test, you can select the working voltage
(110V/230V) of the equipment to be tested according to the working voltage of
the equipment under test. The setting of the working voltage (110V/230V) of
the equipment to be tested will affect the leakage current test results and
the out-phase detection function.
-
Setting of the output voltage of the insulation resistance test voltage.
Press the POWER key to turn on the instrument, press and hold the combination key F + B until entering the voltage setting interface. Set the insulation resistance test voltage by pressing ▲ or ▼ key. Switch to INSULATION, press the POWER key to select the voltage, and switch 250V/500V through ▲ and ▼ keys. Press the POWER key to confirm and exit. -
Setting of the working voltage of the equipment to be tested.
Press the POWER key to turn on the instrument, and press and hold the combination key F + B until entering the voltage setting interface. Set the insulation resistance test voltage or the working voltage of the equipment to be tested by pressing ▲ or ▼ key. Switch to LEAKAGE, press the POWER key to select the voltage, and switch 110V/230V through ▲ and ▼ keys. Press the POWER key to confirm and exit.
CLASS I test
- Connect the equipment to be tested to the meter with the mains test cord supplied with the appliance to the meter output socket and the other end to the power connector of the equipment to be tested.
- Connect the Earth bond lead to the meter (Terminal 12) then connect the other end to a metal conductive part of the CLASS1 appliance (connected to the protective conductor of the equipment).
After correct connection, press the A key to start the CLASS I test. Ensure that the equipment to be tested is powered on during the test. After the test is started, the meter will light the symbol at the right hand side of the LCD screen to indicate that the CLASS I test is being performed currently. The test will be done automatically according to the following steps:
- Grounding Test
- Test the resistance between the conductive part of the equipment to be tested and the protected area.
- If the measured resistance is greater than 25Ω, the buzzer and the grounding icon flash to indicate the grounding fault. After 5 seconds, exit and return to the main interface;
- If the measured resistance is less than 25Ω, compare with the set resistance, display the resistance and “√” or “×”, and then conduct the next test.
- Insulation Resistance Test
- If the power switch of the equipment to be tested is off, the instrument may display LO LOAD to indicate that the equipment to be tested has an insufficient load capacity. Press the A key or turn on the power switch of the equipment to be tested to continue the insulation resistance test. Press any other key to exit and return to the main interface.
- After the test is completed, the insulation resistance is displayed.
- Compare it with the set value and display the comparison result “√” or “×”.
- Leakage current test
- After the test is completed, the leakage current value is displayed.
- Compare it with the set value and display the comparison result “√” or “×”.
- Final Result
- Make a final judgment based on the test results obtained in the three steps above, display PASS or FAIL, and save the test results.
- Press any key to or automatically exit and return to the main interface after about 10 seconds.
CLASS II test
- Connect the equipment to be tested to the meter with the mains test cord: connect one end of the mains test cord to the meter output socket(10) and the other end to the power interface of the equipment to be tested.
After correct connection, press the B key to start the CLASS II test. Ensure that the equipment to be tested is powered on during the test. After the test is started, the meter will light the marking at the right side of the LCD screen to indicate that the CLASS II test is being performed currently. If the insulation resistance test voltage used is 500V, the marking will also be lit. The test will be done automatically according to the following steps:
-
Insulation Resistance Test.
- If the power switch of the equipment to be tested is off, the instrument may display LO LOAD to indicate that the equipment to be tested has an insufficient load capacity. Press the A key or turn on the power switch of the equipment to be tested to continue the insulation resistance test.
- Press any other key to exit and return to the main interface.
- After the test is completed, the insulation resistance is displayed. Compare it with the set value and display the comparison result “√” or “×”.
-
Leakage current test
After the test is completed, the leakage current value is displayed. Compare it with the set value and display the comparison result “√” or “×”. -
Final Result
Make a final judgment based on the test results obtained in the two steps above, display PASS or FAIL, and save the test results. Press any key to or automatically exit and return to the main interface after about 10 seconds.
Note: The CLASS II insulation resistance test voltage is defaulted as 500VDC and can be set to 250VDC. The specific operation is shown in 6.3.
CORD test
- Connect the CORD cable to be tested to the meter: Connect one end of the CORD cable to the meter output socket(10) and the other end to the meter IEC input socket(11).
- After the correct connection, press the C key to start the CORD test.
- After the test is started, the meter will lighten the marking at the right hand side of the LCD screen to indicate that the CORD test is being performed currently.
Grounding Test
Detect the connectivity of the Earth terminal of the CORD cable. If the
connectivity is poor or it is not connected, the buzzer and the grounding icon
flash to indicate the grounding fault, provide an initial error prompt after
the polarity analysis, and exit and return to the main interface after 5
seconds;
If the connectivity is good, compare it with the set resistance, display the
resistance and “√” or “×”, and conduct the next test.
Insulation Resistance Test
After the test is completed, the insulation resistance is displayed. Compare
it with the set value and display the comparison result “√” or “×”.
Polarity Test
Test whether there is an open circuit, short circuit and reverse connection,
display test results with PASS/OPEN/SHOR/CORS/FAIL to respectively indicate
the correct connection, open circuit, short circuit, reverse connection and
concurrence of many errors.
Final Result
Make a final judgment based on the test results obtained in the three steps
above, display GOOD or ERROR, and save the test results. Press any key to or
automatically exit and return to the main interface after about 10 seconds.
Power socket test
- Connect the socket to be tested to the meter with the CORD main test cord: connect one end of the IEC test cord to the meter input socket(11) and the other end to the socket to be tested.
- After correct connection, if the socket to be tested is powered on, the instrument will automatically perform the out-phase test. Other operations will be invalid during this period until the socket to be tested is powered off or the power cord is unplugged from the socket.
- After entering the test interface, the meter will lighten the marking at the right hand side of the LCD screen to indicate that the instrument is under the instrument is performing the out-phase test.
- Note: This test result is affected by the setting of the working voltage of the equipment under test. The setting operation is shown in 6.3.
Test result:
- If it is normal, display LN “√”, LE “√”, NE “√”;
- If the Zero line is out of phase, display LN “×”, LE“√” and NE“×”;
- If the ground line is out of phaset , display LN “√”, LE“×” and NE“×”;
Reading test date
- When the instrument is in the standby mode, press the ▲ key or the ▼ key to view the test results, where a maximum of last 300 test results can be viewed. the meter only saves the results of the CLASS I test, the CLASS II test and the CORD test, the results will be saved only when the test is completed successfully.
- For the test results, , and respectively indicate the test type corresponding to the data displayed currently as CLASS I, CLASS II and CORD.
- The test results viewed indicate data groups with 001 – 300 to distinguish which data belongs to which test result. The last test result is classified as 001; the greater the number, the earlier the test time. Press the POWER key to view the determined final test result of the current test result and display it in the group position.
- Press the F key to exit the view page and return to the main interface.
Technical Specifications
Test Conditions:
Function | Range | Accuracy | Technical parameters |
---|---|---|---|
Grounding performance test | 0Ω~20Ω | ±5%+3 | Test current>200mA (R < 2Ω) |
If 0.01~0.1 passed, display” √” If 0.1~0.5Ω passed, display”√ “and flash
If 0.51~19.9 failed disolav”x”
Test voltage 5VDC, test time 5S
Insulation resistance test| Test voltage 500VDC| | Test current> 1mA (R =
SOOKΩ) Test time 5S
0.2MΩ~19.99MΩ| +5%+3|
Test voltage 250VDC| | Test current> 2mA (R = 1 KΩ), test time 5S
0.2MΩ~19.99MΩ| +5%+3|
CLASS I| | If 1 M passed, display” √”
CLASS 11| | lf 2M □passed, display” √”
CORD| | If 1 M □passed, display “”I/”
Leakage current test| 0.10mA~10mA| ±5%+3| Test voltage 50V rms, frequency 50Hz
Test time 5S
---|---|---|---
CLASS I| | If 0.1~0.75mA passed, display” √”
If 0.75~3.SmA passed, display” √”and flash If 3.5~10.50mA failed disolav “x”
CLASS II| | If 0.1~0.25mA passed, display” √”
If 0.25~10.50mA failed, display “x”
Polarity test| Correct| Disolav ” PASS”
Lopen| “LN x, LE x, NE ,OPEN”
N open| “LN x, LE , NEx, OPEN”
Gopen| “LN , LE x, NE x, OPEN”
L N short| “LN x, LE , NE , SHOR”
L G short| “LN ** LE x NE SHOR”
N G short| “LN ** LE NE x SHOR”
L N reverse| “LN x LE NE CORS”
L E reverse| “LN ** LE x NE CORS”**
N E reverse| “LN,LE ,NE x, CORS”
Multiple errors| “LN X LEX NE X FAIL”
Power socket test| Correct| Display “LN , LE , NE “| 230VAC±15%
110VAC±15%
Ooen Neutral| Display “LN x, LE , NEx”
Open ground| Display “LN , LEx, NEx”
CLASS I TEST:
CLASS II TEST: CORD TEST:
Maintenance of instrument
Maintenance Battery Replacement
When the low battery symbol ” ”appears, replace the batteries by:
- Disconnecting all test leads before battery replacement.
- Power off the unit.
- Unscrew the battery cover.
- AA(LR6) battery 1.5V.
- Ensure the polarity of the batteries is correct.
- Re-attach the battery cover.
Cleaning
- Clean only with a dry cloth; do not use solvents.
- Before use, ensure unit is clean and dry; visually inspect all leads, connectors, and case. Any damage or wear must be rectified to preserve user safety.
Maintenance & Service
- This unit should be calibrated and repaired by an authorized service center. To ensure the accuracy of the device it must be calibrated annually.
- There are no user serviceable parts.
Environmental rating
- Operating temperature range 0°C to 40°C,
- Do not expose to moisture or condensation as incorrect or non-genuine readings may result.
- Storage temperature range -25° C to 65°C.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>