MASTECH MY60 Digital Multimeter User Guide
- June 5, 2024
- MASTECH
Table of Contents
MASTECH MY60 Digital Multimeter
Contents
Specifications
| Range| Resolution| Accuracy
---|---|---|---
| 200mV
2V
20V/200V
750V
| 100µV
1mV
10mV/100mV
1V
| ±(1.2%+3)
±(0.8%+3)
±(0.8%+3)
±(1.2%+3)
| 200mV/2V/20V/200V
1000V
| 0.1mV/1mV/10mV/100mV
1V
| ±(0.5%+2)
±(0.8%+2)
| 20µA
200µA
2mA
20mA
200mA
10A
| 0.01µA
0.1µA
1µA
10µA
0.1mA
10mA
| ±(2.0%+5)
±(1.0%+5)
±(1.0%+5)
±(1.0%+5)
±(1.8%+5)
±(3.0%+7)
| 20µA
200µA/2mA/20mA
200mA
10A
| 0.01µA
0.1µA/1µA/10µA
0.1mA
10mA
| ±(2.0%+5)
±(0.8%+1)
±(1.5%+1)
±(2.0%+5)
R| 200Ω
2kΩ/20kΩ/200kΩ/2MΩ
20MΩ
200MΩ
| 0.1Ω
1Ω/10Ω/100Ω/1kΩ
10kΩ
100kΩ
| ±(0.8%+3)
±(0.8%+2)
±(1.0%+2)
±(6.0%+10)
-
-
-
-
-
-
-
-
- hFE
-
-
-
-
-
-
-
Important Note:
The special attention should be paid when using the test and measurement instrument because the improper usage may cause electric shock and damage the equipment.
- Use proper measurement category (CAT), voltage, and amperage rated probes, test leads, and adapters for the measurement.
- Do not use or store the device around explosive gas, vapor, or under high temperature and humidity.
- When handling the test probes and crocodile clips, keep your fingers behind the physical guard.
- Examine the measuring test leads (if present) before use. Any element of which the insulation is deteriorated (even partially), replace them with proper functional test leads.
- Connect the common test lead before the live test lead and remove the live test lead before the common test lead.
- Before measuring current, make sure the instrument fuses available and disconnect the power supply to the circuit test
- Remove the batteries if the device is not used for an extended period of time, or if stored in temperatures above 45 °C. If the batteries are not removed, battery leakage can damage the device.
Read User Manual Online (PDF format)