hygiena BAX System Q7 Instruction Manual

June 4, 2024
Hygiena

hygiena logo BAX® System Q7 Software Upgrade
Version 4.22.0.10335
Features and Instructions

BAX System Q7

BAX System Q7 software release v4.22 contains several user interface features, new target options, and error logging improvements.
We strongly recommended that you always install the most recent version as soon as possible for optimal performance. Please read all instructions thoroughly and ensure all required prerequisites are confirmed prior to prior to beginning installation of the software update.
If you currently have a WIN7 computer system, please contact diagnostics.support@hygiena.com or your local Hygiena® representative for upgrade support. This release is NOT applicable to WIN7 systems.

Software Updates and Features for Release v4.22.0.10335 for WIN10 ONLY

„ New Assays with Result Interpretations:

  • foodproof® Salmonella Genus plus Enteritidis & Typhimurium Detection LyoKit
  • foodproof® Salmonella plus Cronobacter Detection LyoKit
  • foodproof® Enterobacteriaceae plus Salmonella Detection LyoKit
  • foodproof® Vibrio Detection LyoKit
  • foodproof® Yeast and Mold Quantification LyoKit
  • foodproof® Aspergillus Detection LyoKit
  • microproof® Legionella Quantification LyoKit

„ SDS Equipment Driver Improvements
„ Messages and Confidence Levels:

  • All foolproof and micro proof assays

Software Upgrade Instructions:

  1. Required Prerequisites:
    „ Regional Settings are in English
    „ Read/Write Permissions for the below file locations for all users
    • C:\Program Files\MATLAB
    • C:\Program Files (x86)\BAX
    • C:\ProgramData\BAX Q7 System
    • All Power and Sleep Settings are set to Never
    „ Obtain HEX Calibration Kit from Hygiena
    • This additional calibration plate is a standalone product and is not included with other calibration kits
    • Please visit the contact us page for ordering and technical support

  2. Calibration storage and removal of current software:
    „ Back up the system’s calibration file
    • Select Operations > BAX Maintenance > Export Calibration from the menu bar
    • Navigate to a new folder in which to store the backup file and click Save
    • Name the calibration file (example: Q7Cal_date.bqc)
    „ Uninstall the previous version of the BAX Q7 application
    • Navigate to Windows Settings and select Apps
    hygiena BAX System Q7 - Figure 1 „ Click on BAX System Q7, select Uninstall, and agree (click Yes) on the next window to initiate removal
    hygiena BAX System Q7 - Figure 2
    „ If updating from 4.21, removal of MATLAB is not required. If updating from any previous versions, click on the MATLAB application, select Uninstall, and click Agree on the next window (this may take several minutes)}
    hygiena BAX System Q7 - Figure 3 **Restart the computer BEFORE continuing to step 3
    Software Installation Instructions for Steps 3 and 4


  3. Download 4.22.0.10335 software by clicking this link or the button from the BAX Q7 Page
    „ Transfer download to a memory/USB stick if the computer doesn’t have internet access
    hygiena BAX System Q7 - Figure 4 „ Windows protect screen may pop up. Select More Info, then select Run Anyway.
    • ****
    On the next pop-up, select Yes to allow install to begin
    hygiena BAX System Q7 - Figure 5

  4. MATLAB updates will install first. Follow the prompts below and select the default responses:
    hygiena BAX System Q7 - Figure 6****
    In case this install error occurs: VCRUNTIME140.dll was not found. (shown below)
    „ Please download and install the patch: vc_redist.x64.exe from
    https://www.microsoft.com/en-us/download/details.aspx?id=52685.
    The installation will end when you click OK.
    After installing this patch, MATLAB Installation is continued

  5. hygiena BAX System Q7 - Figure 8 | hygiena BAX System Q7 - Figure 9****
    ---|---
    hygiena BAX System Q7 - Figure 10| hygiena BAX
System Q7 - Figure 11
    hygiena BAX System Q7 - Figure 12| hygiena BAX
System Q7 - Figure 13
    hygiena BAX System Q7 - Figure 14

  6. Following MATLAB installation, the BAX Q7 application install will initiate. Read the prompts carefully. Choose the default responses.
    • Welcome Window – click Nexthygiena BAX System Q7 - Figure
15 „ Calibration File Warning – ENSURE YOU HAVE BACKED UP THE CALIBRATION FILE
    • Check the box and click Next
    hygiena BAX System Q7 - Figure 16 „ License Agreement
    • Check Accept Terms box and click Next
    hygiena BAX System Q7 - Figure 17 „ Custom Setup
    • Click Next
    hygiena BAX System Q7 - Figure 18 „ Password Protection
    • BAX default is disabled password protection. Choose an option for your lab, then click Next
    hygiena BAX System Q7 - Figure 19 „ Ready to Install Program
    • Click Install
    hygiena BAX System Q7 - Figure 20 „ Installing BAX System Q7 Software
    • During installation, an Error Log Reporting window will pop up. We recommend NOT checking the box so the system can collect software and/or hardware errors in the Log Files to assist Diagnostics Support in troubleshooting. However, please review and decide what is appropriate for your lab.
    • Select Next to complete the install
    hygiena BAX System Q7 - Figure 21 „ InstallShield Wizard Complete
    • Select Finish
    hygiena BAX System Q7 - Figure 22• Click on the desktop icon to initiate the program. The new open screen will have software version 4.22.0.10335
    • Enter Instrument Name and Serial Number
    • This will appear on reports and should be utilized
    hygiena BAX System Q7 - Figure 23

  7. CHECK system’s calibration (NEW FEATURE)
    „ From the menu bar, select Diagnostics > Calibration to open the Q7 Calibration Diagnostics window
    • Click the Check Calibration button to determine the status:
    • If calibration is available, no further actions are needed
    • If calibration is NOT optimal or NOT found:
    Import the backed-up calibration file by selecting Operations > BAX Maintenance > Import Calibration from the menu bar and selecting the backed-up calibration file from step 2
    • To utilize any food-proof  assay, additional calibration using the HEX calibration kit must be performed. Please follow the onboard instructions in the BAX Software to perform calibration.
    hygiena BAX System Q7 - Figure 24****

  8. Ready to perform BAX System Analysis
    „ Define proper targets that are used in your lab by selecting View > Select Targets > then in the window, highlight targets being used, and then use arrow buttons to move targets from Define in System to Available for Selection window
    hygiena BAX System Q7 - Figure 25****

hygiena logo****

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals