HACH B3500ul Bio Tector Online TOC Analyzer Instruction Manual
- June 15, 2024
- HACH
Table of Contents
DOC023.52.90722
BioTector B3500ul Online TOC Analyzer
Installation and Operation
08/2022, Edition 1
Specifications
Specifications are subject to change without notice.
This product does not comply with, and is not intended to be put into,
regulated bodies of water or fluid, which includes drinking water or food
contact materials in food and beverage.
Table 1 General specification
Specification | Details |
---|---|
Dimensions (H x W x D) | 1000 x 500 x 320 mm (39.37 x 19.69 x 12.60 in.) |
Enclosure | Rating: IP44 with the doors closed and latched; optional IP54 with |
air purge or vortex cooler
Material: Fiberglass reinforced polyester (FRP)
Weight| < 50 kg (110 lb)
Mounting| Wall mount, indoor installation
Protection class| Class 1 (PE connected)
Pollution degree| 2
Installation category| II
Electrical requirements| 110–120 VAC, 50/60 Hz, 300 W (2.6 A), or
200–230 VAC, 50/60 Hz, 300 W (1.3 A)
Refer to the product rating label for the electrical requirements. Use a
permanent field wiring connection.
Cable entry| Typically, four cable glands (strain relief fittings) are
supplied with the analyzer.
Note: PG13.5 cable glands have a clamping range of 6-12 mm. PG11 cable
glands have a clamping range of 5-10 mm.
Mains power wire| 2 Core +PE1 +Screened; 1.5 mm2 (16 AWG) rated 300 VAC, 60
°C, VW-1;
The cable type is to be SJT, SVT, SOOW or
The power cable installed in accordance with local and regional codes,
suitable for end application. Connected to a dedicated and isolated branch
circuit protected supply rated 10 A.
Signal wire| 4 wires (twisted pair, shielded cable) and more 2 wires for each
additional signal, 0.22 mm2 (24 AWG) minimum and rated 1 A; depending on the
configuration and options installed on the analyzer
Modbus RTU wire| 2 wires (twisted pair, shielded cable), 0.22 mm2 (24 AWG)
minimum UL AWM Style 2919 or equivalent for application
Fuses| Refer to the fuse location diagram in the enclosure. In addition, refer
to the Maintenance and Troubleshooting manual for the specifications.
Operating temperature| 5 to 45 °C (41 to 113 °F)
For the best accuracy, keep the ambient temperature stable, ±3 °C (±5 °F)
maximum
Cooling options are available for the analyzer.
Operating humidity| 5 to 85% non-condensing relative humidity
Storage temperature| 2 to 60 °C (35 to 140 °F); ETL approved devices 5 to 40
°C (41 to 104 °F)
Altitude| 2000 m (6562 ft) maximum
Display| High contrast, 40 character x 16 line backlit LCD with LED backlight
Sound| < 60 dBa
Sample streams| One or two sample streams and one manual grab sample. Refer to
Table 2 for sample requirements.
Data storage| 6000 measurements and 99 error entries in analyzer memory
Table 1 General specification (continued)
Specification | Details |
---|---|
Data send | MMC/SD card to save data, software updates and configuration |
updates
Analog outputs| Four 4–20 mA output signal, programmable (direct or multiplex
mode), optically isolated, self powered, 500 Ω impedance maximum
Analog inputs| Two 4–20 mA input signals, programmable
Relays| Six configurable relays; volt free contacts, 1 A at 30 VDC maximum
Remote control| Digital inputs for remote standby, remote stream selection and
remote grab sample measurement
In addition, the analyzer can be controlled remotely with Modbus.
Communications (optional)| Modbus RTU, Modbus TCP/IP or Profibus. The software
requirement for Modbus RTU and TCP/IP is version 2.12 or later.
Note: When the Profibus option is selected, the analyzer sends the
digital output signals through the Profibus converter with the specific
communication protocol of Profibus.
Reagents| 1.2 N sodium hydroxide (NaOH)
1.8 N sulfuric acid (H2SO4) that contains 80-mg/L manganese sulfate
monohydrate For the reagent usage rate, refer to Plumb the reagents on page
42.
Instrument air| Dry, oil and dust free, > -20 °C (-4 °F) dew point, 22 L/h at
6 bar (87 psi), 5 to 40 °C (41 to 104 °F).
Setpoint:
• 1.5 bar (21.7 psi)
• 1.5 and 0.9 bar (21.7 and 13 psi) when the oxygen concentrator is on.
• 1.2 bar (17.4 psi) when the BioTector air compressor is used.
Note: A filter pack is recommended if the instrument air is not within
specifications.
Purge air| 4 to 7 bar (58 to 101.5 psi), -20 °C (-4 °F) dew point (free of
water, oil and dust) Initial purge air consumption is less than 15 m3/hour.
Normal operation air consumption is less than 6 m3/hour.
Calibration standard| Zero calibration: None or Ultra pure water. For the best
accuracy, use ultra pure water (18.2 MΩ·cm, < 5 ug/L TOC).
Span calibration: TIC (total inorganic carbon) and TOC (total organic carbon)
concentration in the calibration standard is based on the operation range
selected for span calibrations.
Certifications| CE, cETLus
Optional: Class 1 Division 2 and ATEX Zone 2 hazardous area certifications
Warranty| 1 year
Table 2 Sample requirements
Specification | Details |
---|---|
Sample types | Samples cannot contain fats, greases, oils or calcium. Samples |
can contain 0.1% chlorides (salts) maximum. Refer to Table 5 for sodium
chloride interference.
Note: Chlorides (salts) do not interfere with measurements, but can cause
corrosion.
pH| 1 to 12
Sample particle size| 100 µm diameter maximum, soft particulates
Note: Hard particulates (e.g., sand) will cause damage to the analyzer.
Sample pressure| Ambient at sample and manual (grab sample) inlets
Note: For pressurized sample streams, use the optional Sample Overflow
Chamber to supply sample at ambient pressure to the analyzer.
Drain pressure| Ambient
Note: For pressurized drains, use the optional systems available.
Table 2 Sample requirements (continued)
Specification | Details |
---|---|
Sample temperature | 2 to 60 °C (36 to 140 °F) |
Sample flow rate | 100 mL minimum for each sample stream |
Sample volume (usage) | 12.0 mL maximum |
Table 3 Performance specifications
Specification | Details |
---|---|
Operation range 2 | 0 to 5000 µgC/L |
Cycle time | 5 minutes to measure TIC and TOC (minimum) |
Note: The cycle time is based on the operation range and application.
Repeatability| TOC 0 to 5000 µgC/L — ±2% of reading or ±6 µgC/L (the
larger value)
Accuracy| 2% or ±15 µg/L (the larger value)
Signal drift (1 year)| < 5%
Limit of quantitation| 80 µg/L
TIC interference| At 500 µg/L TIC (as bicarbonate), 2% carryover into TOC may
occur.
Table 4 Analysis specifications
Specification | Details |
---|---|
Oxidation method | Two-stage advanced oxidation process (TSAO) with hydroxyl |
radicals
TOC measurement| NDIR (non-dispersive infrared sensor) measurement of CO2
after oxidation
VOC, COD and BOD| Calculated with correlation algorithm that includes TOC
measurement results
Table 5 Sodium chloride interference
Parameter | Interference level |
---|---|
TOC | None |
Note: Chlorides (salts) do not interfere with measurements, but can cause corrosion .
2 There is one operation range maximum for each parameter (e.g., TOC) and each sample stream (e.g.,STREAM 1).
Specifications—IECEx/ATEX analyzer
Specifications are subject to change without notice.
This product does not comply with, and is not intended to be put into,
regulated bodies of water or fluid, which includes drinking water or food
contact materials in food and beverage.
Table 6 General specification
Specification | Details |
---|---|
Dimensions (H x W x D) | 1000 x 500 x 320 mm (29.53 x 19.69 x 12.60 in.) |
Enclosure | Rating: IP44 with the doors closed and latched; optional IP54 with |
air purge or vortex cooler
Material: Fiberglass reinforced polyester (FRP)
Weight| < 50 kg (110 lb)
Mounting| Wall mount, indoor installation
Protection class| Class 1 (PE connected)
Pollution degree| 2
Installation category| II
Electrical requirements| 110–120 VAC, 50/60 Hz, 300 W (2.6 A), or
200–230 VAC, 50/60 Hz, 300 W (1.3 A)
Refer to the product rating label for the electrical requirements. Use a
permanent field wiring connection.
Cable entry| Typically, four cable glands (strain relief fittings) are
supplied with the analyzer.
Note: PG13.5 cable glands have a clamping range of 6-12 mm. PG11 cable
glands have a clamping range of 5-10 mm .
Mains power wire| 2 Core +PE3 +Screened; 1.5 mm2 (16 AWG) rated 300 VAC, 60
°C, VW-1;
The cable type is to be SJT, SVT, SOOW or
The power cable installed in accordance with local and regional codes,
suitable for end application. Connected to a dedicated and isolated branch
circuit protected supply rated 10 A.
Signal wire| 4 wires (twisted pair, shielded cable) and more 2 wires for each
additional signal, 0.22 mm2 (24 AWG) minimum and rated 1 A; depending on the
configuration and options installed on the analyzer
Modbus RTU wire| 2 wires (twisted pair, shielded cable), 0.22 mm2 (24 AWG)
minimum UL AWM Style 2919 or equivalent for application
Fuses| Refer to the fuse location diagram in the enclosure. In addition, refer
to the Maintenance and Troubleshooting manual for the specifications.
Operating temperature| 5 to 35 °C (41 to 95 °F)
For the best accuracy, keep the ambient temperature stable, ±3 °C (±5 °F)
maximum
Cooling options are available for the analyzer.
Operating humidity| 5 to 85% non-condensing relative humidity
Storage temperature| 2 to 60 °C (35 to 140 °F); ETL approved devices 5–40 °C
(41–104 °F)
Altitude| 2000 m (6562 ft) maximum
Display| High contrast, 40 character x 16 line backlit LCD with LED backlight
Sound| < 60 dBa
Sample streams| One or two sample streams and one manual grab sample. Refer to
Table 7 for sample requirements.
Data storage| 6000 measurements and 99 error entries in analyzer memory
3 Protective earth
Table 6 General specification (continued)
Specification | Details |
---|---|
Data send | MMC/SD card to save data, software updates and configuration |
updates
Analog outputs| Three 4–20 mA output signal, programmable (direct or multiplex
mode), optically isolated, self powered, 500 Ω impedance maximum
Analog inputs| Two 4–20 mA input signals, programmable
Relays| Three configurable relays; volt free contacts, 1 A at 30 VDC maximum
Remote control| One digital input for remote standby or remote grab sample
measurement In addition, the analyzer can be controlled remotely with Modbus.
Communications (optional)| Modbus RTU, Modbus TCP/IP or Profibus. The software
requirement for Modbus RTU and TCP/IP is version 2.12 or later.
Note: When the Profibus option is selected, the analyzer sends the
digital output signals through the Profibus converter with the specific
communication protocol of Profibus.
Reagents| 1.2 N sodium hydroxide (NaOH)
1.8 N sulfuric acid (H2SO4) that contains 80-mg/L manganese sulfate
monohydrate For the reagent usage rate, refer to Plumb the reagents on page
42.
Instrument air| Dry, oil and dust free, > -20 °C (-4 °F) dew point, 22 L/h at
6 bar (87 psi), 5 to 40 °C (41 to 104 °F).
Setpoint:• 1.5 bar (21.7 psi)
• 1.5 and 0.9 bar (21.7 and 13 psi) when the oxygen concentrator is on.
• 1.2 bar (17.4 psi) when the BioTector air compressor is used.
Note: A filter pack is recommended if the instrument air is not within
specifications .
Purge air| 4 to 7 bar (58 to 101.5 psi), -20 °C (-4 °F) dew point (free of
water, oil and dust) Initial purge air consumption is less than 15 m3/hour.
Normal operation air consumption is less than 6 m3/hour.
Calibration standard| Zero calibration: None or Ultra pure water. For the best
accuracy, use ultra pure water (18.2 MΩ·cm, < 5 ug/L TOC).
Span calibration: TIC (total inorganic carbon) and TOC (total organic carbon)
concentration in the calibration standard is based on the operation range
selected for span calibrations.
Certifications| CE, cETLus
Class 1 Division 2 and ATEX Zone 2 hazardous area IECEx Zone 1—II 2 G, Ex px
IIC T4 Gb Ta ≤ 40 °C
ATEX Zone 1—II 2 G, Ex px IIC T4 Ta ≤ 40 °C
Warranty| 1 year
Table 7 Sample requirements
Specification | Details |
---|---|
Sample types | Samples cannot contain fats, greases, oils or calcium. Samples |
can contain 0.1% chlorides (salts) maximum. Refer to Table 10 for sodium
chloride interference.
Note: Chlorides (salts) do not interfere with measurements, but can
cause corrosion .
pH| 1 to 12
Sample particle size| 100 µm diameter maximum, soft particulates
Note: Hard particulates (e.g., sand) will cause damage to the analyzer
.
Sample pressure| Ambient at sample and manual (grab sample) inlets
Note: For pressurized sample streams, use the optional Sample Overflow
Chamber to supply sample at ambient pressure to the analyzer .
Table 7 Sample requirements (continued)
Specification | Details |
---|---|
Drain pressure | Ambient |
Note: For pressurized drains, use the optional systems available .
Sample temperature| 2 to 60 °C (36 to 140 °F)
Sample flow rate| 100 mL minimum for each sample stream
Sample volume (usage)| 12.0 mL maximum
Table 8 Performance specifications
Specification | Details |
---|---|
Operation range4 | 0 to 5000 µgC/L |
Cycle time | 5 minutes to measure TIC and TOC (minimum) |
Note: The cycle time is based on the operation range and application.
Repeatability| TOC 0 to 5000 µgC/L — ±2% of reading or ±6 µgC/L (the
larger value)
Accuracy| 2% or ±15 µg/L (the larger value)
Signal drift (1 year)| < 5%
Limit of quantitation| 80 µg/L
TIC interference| At 500 µg/L TIC (as bicarbonate), 2% carryover into TOC may
occur.
Table 9 Analysis specifications
Specification | Details |
---|---|
Oxidation method | Two-stage advanced oxidation process (TSAO) with hydroxyl |
radicals
TOC measurement| NDIR (non-dispersive infrared sensor) measurement of CO2
after oxidation
VOC, COD and BOD| Calculated with correlation algorithm that includes TOC
measurement results
Table 10 Sodium chloride interference
Parameter | Interference level |
---|---|
TOC | None |
Note: Chlorides (salts) do not interfere with measurements, but can cause corrosion .
4 There is one operation range maximum for each parameter (e.g., TOC) and each sample stream (e.g., STREAM 1).
General information
In no event will the manufacturer be liable for damages resulting from any
improper use of product or failure to comply with the instructions in the
manual. The manufacturer reserves the right to make changes in this manual and
the products it describes at any
time, without notice or obligation. Revised editions are found on the
manufacturer’s website.
3.1 Safety information
The manufacturer is not responsible for any damages due to misapplication
or misuse of this product including, without limitation, direct, incidental
and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is soley responsible to identify
critical application risks and install appropriate mechanisms to protect
processes during a possible equipment malfunction. Please read this entire
manual before unpacking, setting up or operating this equipment.
Pay attention to all danger and caution statements. Failure to do so could
result in serious injury to the operator or damage to the equipment. Make sure
that the protection provided by this equipment is not impaired. Do not use or
install this equipment in any manner other than that specified in this manual.
3.1.1 Safety symbols and markings
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
The safety symbols and marking that follow are used on the equipment and in
the product documentation. The definitions are in the table that follows.
| Caution/Warning. This symbol identifies that an appropriate safety
instruction should be followed or a potential hazard exists.
---|---
| Hazardous voltage. This symbol indicates that hazardous voltages are present
where a risk of electrical shock exists.
| Hot surface. This symbol indicates that the marked item can be hot and
should not be touched without care.
| Corrosive substance. This symbol identifies the presence of a strong
corrosive or other hazardous substance and a risk of chemical harm. Only
individuals qualified and trained to work with chemicals should handle
chemicals or perform maintenance on chemical delivery systems associated with
the equipment.
| Toxic. This symbol indicates a toxic/poisonous substance hazard.
| This symbol indicates the presence of devices sensitive to Electro-static
Discharge (ESD) and indicates that care must be taken to prevent damage with
the equipment.
| This symbol indicates a flying debris hazard.
| Protective earth. This symbol indicates a terminal which is intended for
connection to an external conductor for protection against electric shock in
case of a fault (or the terminal of a protective earth (ground) electrode).
| Noiseless (clean) earth. This symbol indicates a functional earthing
(grounding) terminal (e.g., a specially designed earthing (grounding) system)
to avoid a malfunction of the equipment.
| This symbol indicates an inhalation hazard.
---|---
| This symbol indicates there is a lifting hazard because the object is heavy.
| This symbol indicates a fire hazard.
| Electrical equipment marked with this symbol may not be disposed of in
European domestic or public disposal systems. Return old or end-of-life
equipment to the manufacturer for disposal at no charge to the user.
3.1.2 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
3.1.3 Ozone precautions
CAUTION
| Ozone inhalation hazard. This instrument produces ozone that is contained
within the equipment, specifically within the internal plumbing. The ozone
could be released under fault conditions.
---|---
It is recommended to plumb the exhaust gas port to a fume hood or to the
building exterior in accordance with local, regional and national
requirements.
Exposure to even low concentrations of ozone can damage delicate nasal,
bronchial and pulmonary membrane. In sufficient concentration, ozone can cause
headaches, coughing, eye, nose and throat irritation. Immediately, move the
victim to uncontaminated air and seek first aid.
The type and harshness of symptoms are based on the concentration and exposure
time (n). Ozone poisoning includes one or more of the symptoms that follow.
- Irritation or burning of the eyes, nose or throat
- Lassitude
- Frontal headache
- Sensation of sub-sternal pressure
- Constriction or oppression
- Acid taste in mouth
- Asthma
In case of more severe ozone poisoning, the symptoms can include dyspnea,
cough, choking sensation, tachycardia, vertigo, lowering of blood pressure,
cramping, chest pain, and generalized body pain. Ozone can cause a pulmonary
oedema one or more hours after exposure.
3.2 Compliance and certification
CAUTION
This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Canadian Radio Interference-Causing Equipment Regulation, ICES-003, Class
A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
InterferenceCausing Equipment Regulations.
FCC Part 15, Class “A” Limits
Supporting test records reside with the manufacturer. The device complies
with Part 15 of the FCC Rules. Operation is subject to the following
conditions:
- The equipment may not cause harmful interference.
- The equipment must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
- Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
- If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
- Move the equipment away from the device receiving the interference.
- Reposition the receiving antenna for the device receiving the interference.
- Try combinations of the above.
3.3 Compliance and certification marks
****| The CE (European Conformity “Conformité Européene”) mark on the
instrument indicates that “The instrument complies with the European product
directives, health, safety and environmental protection legislations”.
---|---
| The ETL (Electrical Testing Laboratories) Listed mark on the instrument
indicates that “This product has been tested to Safety Requirements of
Electrical Equipment for Measurements, Control and Laboratory use; Part 1:
General Requirements of ANSI/UL 61010-1 and CAN/CSA-C22.2 No 61010-1”.
The Intertek ETL listed mark on the instrument identifies that the product has
been tested by Intertek, found in compliance with accepted national standards,
and the instrument meets the minimal requirements required for sale or
distribution.
3.4 EMC compliance statement (Korea)
Type of equipment | Additional information |
---|---|
Class A equipment (Industrial Broadcasting and Communication Equipment) | This |
equipment meets Industrial (Class A) EMC requirements. This equipment is for use in industrial environments only.
3.5 Product overview
NOTICE
Perchlorate Material—Special handling may apply. Refer to
www.dtsc.ca.gov/hazardouswaste/perchlorate.
This perchlorate warning applies only to primary batteries (provided singly or
installed on this equipment) when sold or distributed in California, USA.The
B3500ul TOC analyzer is intended for the measurement of total organic carbon
and has an internal oxygen concentrator.
The analyzer can measure the parameters that follow in wastewater, process
water, surface water and seawater:
- TIC—Total in-organic carbon in µgC/L
- TOC (NPOC)—Total organic carbon in µgC/L, includes NPOC (non-purgeable organic carbon)
- TOC (NPOC + POC)—Total organic carbon in µgC/L, includes NPOC and POC (purgeable organic carbon)
- TC—TIC + TOC
- COD5 —Chemical oxygen demand
- BOD5—Biochemical oxygen demand
- VOC (POC)5 —Volatile organic carbon, includes POC
The analyzer uses the analysis methods in Table 4 on page 5.
For theory of operation information, refer to the BioTector B3500 videos on
youtube.com and Hach Support Online
(https://support.hach.com).
5 Calculated with a correlation algorithm that includes TOC. To show the
calculated results on the display, set the DISPLAY setting on the COD and/or
BOD PROGRAM menu to YES.
The analyzer is factory configured as one of the systems that follow:
- TIC + TOC system 6 —Measures the total inorganic carbon (TIC) and total organic carbon (TOC) content of a sample. The TOC result is the non-purgeable organic carbon (NPOC). The TIC + TOC system is used to measure samples that does not contain volatile organic material or contain a very small concentration of volatile organic material.
- TC system—Measures the total carbon (TC) content of a sample. The TC result is the sum of TIC, NPOC and purgeable organic carbon (POC) content of a sample.
- VOC system—Measures the TIC, TOC, TC and volatile organic carbon (VOC) contents of a sample with two analysis reactions in a single reactor configuration. The VOC result is the purgeable organic carbon (POC). The TOC result is calculated from the TC and TIC measurements as a TC – TIC result. Thus, the TOC result includes the VOC (POC) content of the sample. The TOC result is the sum of NPOC and POC content.
Figure 1 shows an overview of the analyzer exterior.
Refer to Analysis enclosure on page 54 for interior views of the analyzer.
NOTICE
The accessories for the analyzer (e.g., compressor) have separate user
manuals.
For installation in hazardous (classified) locations, refer to the
instructions in the ATEX Category 3 Zone 2 manual and Series 5 Z-purge manual
for the analyzer.
6 The standard analyzer is a TIC + TOC system.
Figure 1 Product overview with side view
1| Reagent, sample and drain fittings (Figure 2)| 5| Cable strain-
relief fittings for electrical connections
---|---|---|---
2| Single-stream analyzer| 6| Display and keypad
3| Dual-stream analyzer| 7| Door lock
4| Fan| 8| Door key
Figure 2 Reagent, sample and drain fittings
1 Single-stream analyzer | 2 Dual-stream analyzer |
---|
3.6 Product components
Make sure that all components have been received. Refer to the supplied
documentation.
If any items are missing or damaged, contact the manufacturer or a sales
representative immediately.
Installation and startup checklist
Use the checklist that follows to complete installation and startup. Do the
tasks in the order given.
If the analyzer is certified for hazardous areas, read the hazardous area
documentation supplied with the analyzer. The hazardous area documentation
contains important information for compliance with explosion protection
regulations.
Task | Initial |
---|---|
Wall mounting: |
Identify the correct installation location. Refer to Installation guidelines
on page 23.
Install the mounting brackets. Attach the analyzer to a wall. Refer to Wall
mount on page 23.|
Electrical connections:|
Connect an earth cable to the M8 earth bolt that is above the cable glands on
the left side of the analyzer.
Connect the analyzer to power. Then, install the supplied ferrite on the power
cable. Refer to Connect power on page 26.
The analyzer is a permanently wired device and configured for 120V or 240V as indicated on the product type label on the left side of the upper enclosure.
Do not set the power to on.
(Optional) Connect the relays (e.g., fault relay) to external devices. Refer
to Connect the relays on page 28.|
(Optional) Connect the 4–20 mA outputs to external devices. Refer to Connect
the analog outputs on page 29.|
(Optional) Connect the digital inputs to external devices to remotely control
the analyzer. Refer to Power, analog output and relay terminals on page 29.|
Connect the Modbus TCP/IP option if installed. Refer to Connect Modbus TCP/IP
(Ethernet) on page 34.|
Connect the Modbus RTU option if installed. Refer to Connect Modbus RTU
(RS485) on page 31.|
Make sure that there are no loose electrical connections in the analyzer.|
Plumbing:|
The orientation of the ferrules used to connect the tubing is important. Refer
to Tube connections on page 36.
Plumb a sample stream(s) to the SAMPLE fitting(s) on the analyzer. Connect a 2
to 2.5 m (79 to 98 inch) length of tubing to the MANUAL fitting. Refer to
Plumb the sample stream(s) and manual stream on page 37.|
Plumb the drain lines. Refer to Plumb the drain lines on page 40.|
Plumb instrument air to the INSTRUMENT AIR fitting on the left side of the
analyzer. Refer to Connect instrument air on page 41. Make sure that the set
point of the instrument air pressure supply is 1.5 bar (21.7 psi) (or 1.2 bar
(17.4 psi) for the BioTector air compressor).
Note: The minimum air supply flow rate is 8.4 m 3 /hour. The average air consumption is less than 5.4 m 3 /hour, and typically
3.6 m 3 /hour during online operation.
|
Plumb the EXHAUST fitting to a ventilated area. Refer to Plumb the exhaust on
page 42.|
Plumb the reagent containers to the fittings on the left side of the analyzer.
Refer to Plumb the reagents on page 42.|
Install the tubing on the sample pump. Refer to Install the sample pump tubing
on page 46.|
Connect the tubes that were disconnected for shipping. Refer to Connect the
internal tubing on page 47.|
Make sure that there are no loose plumbing connections in the analyzer.|
If the analyzer is supplied as an “air purge ready” system (no fan) or there
are corrosive gasses in the area, connect the air purge to the analyzer. Refer
to Connect the air purge on page 48.|
Look at all of the tubes and connections for possible leaks. Repair the leaks
found.|
Task| Initial
---|---
Startup:|
Set the circuit breaker for the analyzer to on, then set the main power switch
to on. Refer to Set the power to on on page 51.
Set the language that shows on the display (default: English). Refer to Set
the language on page 51.|
Set the time and date on the analyzer. Refer to Set the time and date on page
51.|
Adjust the display brightness and contrast as necessary. Refer to Adjust the
display brightness and contrast on page 52.|
Make sure that the set point of the instrument air pressure supply is 1.5 bar
(21.7 psi) (or 1.2 bar (17.4 psi) for the BioTector air compressor). The
instrument air pressure goes between 1.5 and 0.9 bar (21.7 and 13 psi) when
the oxygen concentrator is on.|
Select MAINTENANCE > DIAGNOSTICS > O2-CTRL STATUS. Make sure that the pressure
reading that shows on the display is between 390 and 400 mbar when the MFC is
off.|
Select MAINTENANCE > DIAGNOSTICS > SIMULATE. Select MFC. Set the flow to 60
L/h. Push to start the mass flow controller (MFC). Select O2-CTRL STATUS.
Make sure that the pressure reading is not less than 320 mbar.|
Identify if there is CO2 contamination in the oxygen supply. Refer to Examine
the oxygen supply on page 52.|
Examine the sample, acid and base pump operation. Refer to Examine the pumps
on page 52.|
Make sure that the valves open and close correctly. Refer to Examine the
valves on page 53.|
Set the reagent volumes on the analyzer and start a new reagent cycle. Refer
to Set the reagent volumes on page 54.
Note: The new reagent cycle includes a zero calibration .|
Push to go to the main menu, then select OPERATION > START,STOP > START to
start the analyzer. Do 10 measurements until the measurements are stable.|
Do another zero calibration. Select CALIBRATION > ZERO CALIBRATION > RUN ZERO
CALIBRATION.|
Measure ultra pure water five times at operation range 1 to make sure that the
zero calibration is correct. Connect ultra pure water to the MANUAL fitting.
Refer to Measure ultra pure water on page 54.
If the CO2 peak values on the display are not almost zero, do a pH test. Refer
to the instruction in the Maintenance manual.|
When the startup tests are complete, make sure that the top-left corner of the
Reaction Data screen does not show “SYSTEM FAULT” or “SYSTEM WARNING”.
Note: If “SYSTEM FAULT” or “SYSTEM WARNING” show, select OPERATION >
FAULT ARCHIVE. Faults and warnings preceeded by an “*” are active. Refer to
Troubleshooting in the Maintenance and Troubleshooting Manual for more
information.|
Configuration:|
Set the INTERVAL setting to set the time between reactions. Refer to Set the
measurement interval on page 59.
Set the sample pump forward times for each sample stream. Refer to Set the
sample pump times on page 59.|
Set the stream sequence, the number of reactions to do at each stream and the
operation range for each stream. Refer to Set the stream sequence and
operation range on page 60.
Note: If Modbus RTU or TCP/IP is installed, the Modbus master controls
the stream sequence and operation ranges (default) .|
(Optional) Set the analyzer to show the calculated COD and/or BOD information
on the display. Refer to Configure the COD and BOD settings on page 61.|
Configure the install new reagents settings. Refer to Configure the install
new reagents settings on page 62.|
Configure the alarm settings for low reagents and no reagents. Refer to Set
reagent monitoring on page 62.|
Task| Initial
---|---
Configure the analog outputs that are connected to an external device. Refer
to Configure the analog outputs on page 63.|
Configure the relays that are connected to an external device. Refer to
Configure the relays on page 65.|
Make sure that the operation of the analog outputs and relays is correct.
Refer to the instructions in the Maintenance manual.|
If the optional Modbus TCP/IP module is installed in the analyzer, configure
the Modbus settings. Refer to Configure the Modbus TCP/IP settings on page
68.|
Set the PRINT MODE setting to select the type of reaction data saved to the
MMC/SD card (STANDARD or ENGINEERING) and the type of decimal point (POINT (.)
or COMMA (,). Refer to Configure the communication settings on page 67.
Note: The manufacturer recommends that PRINT MODE is set to
ENGINEERING so that troubleshooting data is saved .|
Calibration:|
Let the analyzer operate for 24 hours for measurements to become stable.
Set the operation range and the calibration standard for span calibrations.
Refer to Start a span calibration or span check on page 72.|
Plumb the calibration standard to the MANUAL fitting. Refer to Plumb the
calibration standard on page 74.|
Start a span calibration. Select CALIBRATION > SPAN CALIBRATION > RUN SPAN
CALIBRATION.|
When the span calibration is complete, examine two or three reactions
(measurements). Make sure that the CO2 peak values are correct. Refer to
Reaction Graph screen on page 79.|
(Optional) Set the days and time when the analyzer does a span calibration,
span check, zero calibration and/or zero check. Refer to the instructions in
the Advanced Configuration manual.|
Save the changes:|
Put the supplied MMC/SD card in the MMC/SD card slot if not already installed.
Refer to Figure 24 on page 84.
Push to go to the main menu, then select MAINTENANCE > DIAGNOSTICS > DATA
OUTPUT > SEND ALL DATA to save the reaction archive, fault archive, analyzer
settings and diagnostics data to the MMC/SD card.|
Installation
DANGER
| Multiple hazards. Only qualified personnel must conduct the tasks described
in this section of the document.
---|---
5.1 Installation guidelines
- Install the analyzer near an open drain. The analyzer waste usually has a low pH (acidic) and can be hazardous. Refer to local regulatory agency instructions for disposal.
- Install the analyzer as near to the sampling point as possible to decrease analysis delay.
- Install the analyzer indoors in a clean, dry, well-ventilated and temperature-controlled location. Refer to the operating temperature and humidity specifications in Specifications on page 3.
- Mount the analyzer upright and level on a flat, vertical surface.
- Do not install the analyzer in direct sunlight or near a heat source.
- Install the analyzer so that the power disconnect device is visible and easily accessible.
- If the analyzer has a Class 1 Division 2 or ATEX Zone 2 hazardous area certification, read the hazardous area documentation supplied with the analyzer. The documentation contains important compliance information and explosion protection regulations.
5.2 Wall mount
WARNING
| Personal injury hazard. Make sure that the wall mounting is able to hold 4
times the weight of the equipment.
---|---
WARNING
| Personal injury hazard. Instruments or components are heavy. Use assistance
to install or move.
---|---
NOTICE
To prevent instrument damage, make sure that there is at least 300 mm (12 in.)
of clearance on the sides and 1500 mm (59 in.) in the front of the analyzer.
Refer to Figure 3 for dimensions.
- Attach the wall mounting brackets to the back of the analyzer. Refer to the documentation supplied with the wall mounting brackets.
- Install mounting hardware on a wall that can hold 4 times the weight of the analyzer (size M8 bolts minimum). Refer to Figure 3 for the mounting hole dimensions. Refer to Specifications on page 3 for the weight of the analyzer. Mounting hardware is supplied by the user.
- Lift the analyzer with a forklift to attach the analyzer to the wall with the wall mounting brackets.
- Make sure that analyzer is level.
Figure 3 Mounting hole dimensions
5.3 Electrical installation
DANGER
| Electrocution hazard. Always remove power to the instrument before making
electrical connections.
---|---
CAUTION
| Multiple hazards. This instrument must be installed by Hach trained
installation engineer in accordance with local and regional electrical codes.
---|---
The analyzer is a permanently wired device and configured for 120V or 240V as
indicated on the product type label on the left side of the upper enclosure.
5.3.1 Electrostatic discharge (ESD) considerations
NOTICE
| Potential Instrument Damage. Delicate internal electronic components can
be damaged
by static electricity, resulting in degraded performance or eventual failure.
---|---
Refer to the steps in this procedure to prevent ESD damage to the instrument:
- Touch an earth-grounded metal surface such as the chassis of an instrument, a metal conduit or pipe to discharge static electricity from the body.
- Avoid excessive movement. Transport static-sensitive components in anti-static containers or packages.
- Wear a wrist strap connected by a wire to earth ground.
- Work in a static-safe area with anti-static floor pads and work bench pads.
5.3.2 Open the door
5.3.3 Connect power
DANGER
Electrocution hazard. Protective Earth Ground (PE) connection is required.
DANGER
Electrical shock and fire hazards. Make sure to identify the local
disconnect clearly for the installation.
WARNING
Potential electrocution hazard. If this equipment is used in potentially wet
locations, a Ground Fault Interrupt device must be used for connecting the
equipment to its mains power source.
NOTICE
Install the device in a location and position that gives easy access to the
disconnect device and its operation.
Do not use a power cord to supply power. Refer to the illustrated steps that
follow and Power, analog output and relay terminals on page 29 to connect
power. Make sure to install the supplied ferrite on the power cable
(illustrated step 3).
The analyzer is a permanently wired device and configured for 120V or 240V as
indicated on the product type label on the left side of the upper enclosure.
The analyzer requires a dedicated branch circuit protected power source and an
isolator within 1 m (3.3 ft).
-
Install a 2-pole, 10 A maximum local disconnect switch for the analyzer within 2 m (6.5 ft) of the analyzer. Put a label on the disconnect that identifies it as the main disconnect device for the analyzer.
-
Make sure that the mains wire power and safety ground service drops for the analyzer are a 2-wire and protective earth cable, 1.5 mm² (16 AWG), 10 A minimum and the wire insulation is rated 300 VAC minimum, 60 °C (140 °F) minimum and VW-1 for fire. Use a shielded mains power cable connected to a shielded earth to comply with the Electromagnetic Compatibility Directive (2004/108/EC).
Use SJT, SVT SOOW orequivalent cable as applicable to the application. -
Connect the disconnect switch to a branch circuit/miniature circuit breaker (MCB) protection rated 10 A/ Type D. Install an earth leakage circuit breaker in accordance with local and regional regulations if applicable.
-
Connect equipment in accordance with local, regional or national electrical codes.
-
Typically, four cable glands (strain relief fittings) are supplied with the analyzer.
PG13.5 cable glands have a clamping range of 6–12 mm. PG11 cable glands have a clamping range of 5–10 mm.
5.3.4 Connect the relays
DANGER
Electrocution hazard. Do not mix high and low voltage. Make sure that the
relay connections are all high voltage AC or all low voltage DC.
WARNING
Potential Electrocution Hazard. Power and relay terminals are designed for
only single wire termination. Do not use more than one wire in each terminal.
WARNING
Potential fire hazard. Do not daisy-chain the common relay connections or
jumper wire from the mains power connection inside the instrument.
CAUTION
Fire hazard. Relay loads must be resistive. Always limit current to the
relays with an external fuse or breaker. Obey the relay ratings in the
Specifications section.
The analyzer has a maximum of six non-powered relays. The relays are
programmable.
The relays are rated at 1 A, 30 VDC maximum.
Use the relay connections to start or stop an external device such as an
alarm. Each relay changes state when the selected condition for the relay
occurs.
Refer to Power, analog output and relay terminals on page 29 and Table 11 to
connect an external device to a relay. Refer to Configure the relays on page
65 to select the condition that sets each relay to on.
(18 to 16 AWG) wire (as determined by load application)7
The relay terminals accept 1.0 to 1.29 mm2 . Wire gauge less than 18 AWG is
not recommended. Use wire with an insulation rating of 300 VAC or higher. Make
sure that the field wiring insulation is rated 80 °C (176 °F) minimum.
Make sure to have a second switch available to remove power from the relays
locally if there is an emergency or for maintenance.
Table 11 Wiring information—relays
NO | COM | NC |
---|---|---|
Normally open | Common | Normally closed |
5.3.5 Connect the analog outputs
The analyzer has a maximum of four 4–20 mA analog outputs. Use the analog
outputs for analog signaling or to control external devices.
Refer to Power, analog output and relay terminals on page 29 to connect an
external device to an analog output.
Wrap the 4-20 mA cables one time around the supplied ferrite to form a single
loop.
Depending on the configuration and options installed on the analyzer, the
minimum specifications for signal and communications cable is 4 wires (twisted
pair, shielded cable) and more 2 wires for each additional signal, 0.22 mm²
(24 AWG) minimum and rated 1 A. Select the full scale value shown as 20 mA on
each analog output. Select the analysis result each analog output shows. Refer
to Configure the analog outputs on page 63.
Notes:
- The analog outputs are isolated from the other electronics, but are not isolated from each other.
- The analog outputs are self-powered. Do not connect to a load with voltage that is independently applied.
- The analog outputs cannot be used to supply power to a 2-wire (loop-powered) transmitter.
5.3.6 Power, analog output and relay terminals
DANGER
Electrocution hazard. Always remove power to the instrument before making
electrical connections.
Refer to Figure 4 for the location of the power, analog output, relay
terminals and digital inputs. Table 12 gives the terminal descriptions.
The four digital inputs in Table 12 are used for remote standby, remote stream
selection and remote grab sample measurement. To identify the digital input
functions, select MAINTENANCE > DIAGNOSTICS > INPUT/OUTPUT STATUS > DIGITAL
INPUT.
There are two digital inputs. Contact technical support for information about
the digital input functions and configuration.
Make electrical connections through the cable strain-relief fittings on the
side of the analyzer.
Always remove power to the instrument before making electrical connections.
Refer to Figure 5 on page 32.
To keep the environmental rating:
- Do not put more than one cable (or two wires) through a strain-relief fitting.
- Make sure that the strain-relief fittings that are unused have rubber cable plugs in them.
Figure 4 Power and Input/Output Board
Table 12 Terminal descriptions
Terminal | Description | Terminal | Description |
---|---|---|---|
E | Protective earth for mains power and screened earth cable | 41 | Power supply |
2 output, 24 VDC
N| Neutral (or L2 for US and Canada)| 42| Power supply 2 output, 0 VDC
P| 110–120 VAC or 200–230 VAC
1 phase
| 43| Power supply 2 output, 0 VDC
11| Relay 1, NO| 44| Power supply 1 output, 24 VDC
12| Relay 1, COM| 45| Power supply 1 output, 0 VDC
13| Relay 1, NC| 46| RS232 output: TX (not used)
14| Relay 2, NO| 47| RS232 output: RX (not used)
Table 12 Terminal descriptions (continued)
Terminal | Description | Terminal | Description |
---|---|---|---|
15 | Relay 2, COM | 48 | RS232 output: GND (not used) |
16 | Relay 2, NC | 51 | 4–20 mA signal in 1+ |
17 | Relay 3, NO | 52 | 4–20 mA signal in 1- |
18 | Relay 3, COM | 53 | 4–20 mA signal in 2+ |
19 | Relay 4, NO | 54 | 4–20 mA signal in 2- |
20 | Relay 4, COM | 71 | Modbus D + * |
21 | Relay 5, NO | 72 | Modbus D – * |
22 | Relay 5, COM | 73 | Modbus ground * |
23 | Relay 6, NO | 74 | Modbus power * |
24 | Relay 6, COM | 61 | 4–20 mA signal out 1+ |
31 | Digital input 1, 24 VDC | 62 | 4–20 mA signal out 1– |
32 | Digital input 1, 0 VDC | 63 | 4–20 mA signal out 2+ |
33 | Digital input 2, 24 VDC | 64 | 4–20 mA signal out 2– |
34 | Digital input 2, 0 VDC | 65 | 4–20 mA signal out 3+ |
35 | Digital input 3, 24 VDC | 66 | 4–20 mA signal out 3– |
36 | Digital input 3, 0 VDC | 67 | 4–20 mA signal out 4+ |
37 | Digital input 4, 24 VDC | 68 | 4–20 mA signal out 4+ |
38 | Digital input 4, 0 VDC |
-
= optional
5.3.7 Connect Modbus RTU (RS485)
For Modbus RTU data transmission, connect the Modbus RTU terminals in the analyzer to a Modbus master device as follows:-
Remove power to the analyzer. Refer to the illustrated steps in Figure 5.
-
Put a 4-wire, twisted pair, shielded cable through a cable strain-relief fitting on the left side of the analyzer. Use wire gauge of 0.2 mm² (24 AWG) minimum.
-
Connect three of the wires to the Modbus RTU terminals in the analyzer. Refer to Figure 6 and Table 13 for wiring information.
Refer to Figure 7 for the location of the Modbus RTU terminals in the analyzer. -
Connect the shield wire of the cable to terminal 73 of the analyzer. Refer to Figure 6 and Table 13 for wiring information.
Note: As an alternative, connect the shield wire to the ground terminal of the Modbus master device. -
Tighten the cable strain-relief fitting.
-
Connect the other end of the cable to a Modbus master device. Refer to Figure 6.
-
Make sure that the wire connected to terminal 71 (D+) is positively biased compared to terminal 72 (D–) when the bus is in an idle condition.
-
To terminate the bus, install a jumper on J18 of the motherboard. Refer to Figure 7.
The motherboard is in the electronic enclosure on the door behind the cover.
-
1 | Modbus master | 3 | To other RS485 devices |
---|---|---|---|
2 | Analyzer |
Table 13 Wiring information
Terminal | Signal |
---|---|
71 | D+ |
72 | D– |
Table 13 Wiring information (continued)
Terminal | Signal |
---|---|
73 | Modbus ground |
74 | Modbus power |
1 Modbus RTU terminals | 3 Bus-termination jumper (J18) |
---|
2 Motherboard
5.3.8 Connect Modbus TCP/IP (Ethernet)
If the optional Modbus TCP/IP module is installed in the analyzer, configure
the Modbus module and connect the module to a Modbus master device. Refer to
the sections that follow.
The Modbus TCP/IP module is marked “MODBUS” and is below the terminals for the
mains power, analog output and relays.
5.3.8.1 Configure the Modbus TCP/IP module
-
Set the analyzer power to on.
-
Use an Ethernet cable to connect a laptop to the Modbus TCP/IP (RJ45) connector in the analyzer. Refer to Figure 8 on page 35.
-
On the laptop, click the Start icon and select Control Panel.
-
Select Network and Internet.
-
Select Network and Sharing Center.
-
On the right side of the window, select Change adapter settings.
-
Right-click Local Area Connection and select Properties.
-
Select Internet Protocol Version 4 (TCP/IPv4) from the list, then click Properties.
-
Record the properties to go back to the properties in the future as necessary.
-
Select Use the following IP address.
-
Enter the IP address and subnet mask that follow:
• IP address: 192.168.254.100
• Subnet mask: 255.255.255.0 -
Click OK.
-
Close the open windows.
-
Open a web browser.
-
In the address bar of the web browser, enter the default IP address (192.168.254.254).
The web-interface of the Modbus TCP module shows. -
Enter the user name and password:
• User name: Admin
• Password: admin -
Use a web-interface at port 80 to change the configuration of the Modbus TCP module, such as the IP address (192.168.254.254) or the TCP/IP port (502).
5.3.8.2 Connect the Modbus TCP/IP module
For Modbus TCP data transmission, connect the Modbus TCP/IP connector in
the analyzer to a Modbus master device as follows:
- Put an Ethernet cable through a cable strain-relief fitting on the left side of the analyzer.
- Connect the Ethernet cable to the Modbus TCP/IP connector in the analyzer. Refer to Figure 8.
- Tighten the cable strain-relief fitting.
- Connect the other end of the Ethernet cable to a Modbus master device. Refer to Figure 9.
If the analyzer has two Modbus TCP/IP connectors, fully redundant data transmission is possible. To connect an analyzer to two Modbus master devices, refer to Figure 10.
Figure 8 Modbus TCP/IP connector
1 Modbus TCP/IP connector
Figure 9 Normal Modbus TCP wiring
1 Analyzer | 2 Modbus master |
---|
Figure 10 Redundant Modbus TCP wiring
1 Modbus Master 1 | 3 Analyzer |
---|
2 Modus Master 2
5.4 Plumbing
5.4.1 Tube connections
The orientation of the ferrules used to connect the tubing is important.
Incorrect ferrule orientation can cause leaks and/or air bubbles in the
analyzer tubing. Refer to Figure 11 for the correct ferrule orientation.
- Cut tubing with a tube cutting tool. Do not use a blade or scissors or leaks can occur.
- Put the tube fully into the fitting.
- Tighten the nut by hand. If the fittings are tighten too much, damage to the fittings and leaks will occur.
• Stainless Steel fittings—Tighten another 1¼ turn with an adjustable wrench.
Stainless steel fittings used on 1/8-in. ID PFA tubing must be tightened only another ¾ of a turn.
• PFA fittings—Tighten another ½ turn with an adjustable wrench.
• Swagelok PFA tee fittings with blue nuts—Tighten until the nut stops (or cannot be tightened more) with an adjustable wrench. The tee fittings with blue nuts have only one ferrule and no back cutting ring or back ferrule.
• Acid and base pump fittings—Tighten the ferrule by hand. Do not use an adjustable wrench.
To tightening a fitting that was tightened before, tighten the number of turns the fitting was tightened before plus a little more with an adjustable wrench.
Figure 11 Ferrule orientation
1 PFA and PVDF fittings | 5 Back ferrule | 9 CTFE ferrule |
---|---|---|
2 Stainless steel fittings (SS-316) | 6 Nut | 10 PFA ferrule |
3 Front ferrule | 7 Acid and base pump fittings | |
4 Back cutting ring | 8 Swagelok PFA tee fittings with blue nuts |
5.4.2 Plumb the sample stream(s) and manual stream
Refer to Specifications on page 3 for the sample specifications. The sample
pressure at the sample inlet must be at ambient pressure.
For pressurized sample streams, install the optional Sample Overflow Chamber
in the sample line to supply sample at ambient pressure. Refer to Install a
sample overflow chamber (optional) on page 40.
-
Use 1/4-in. OD x 1/8-in. ID PFA tubing to plumb the SAMPLE fitting to a sample stream. Make the sample line as short as possible.
Refer to Sample line guidelines on page 37 for instructions. -
Plumb the SAMPLE 2 fitting to a sample stream, if available.
-
Connect 1/4-in. OD x 1/8-in. ID PFA tubing to the MANUAL fitting as necessary.
Make sure that the tubing length is 2 to 2.5 (6.5 to 8.2 ft).
Use the MANUAL fitting to measure grab samples or ultra pure water as necessary and to measure the calibration standard during span calibrations. -
When all of the tubes are connected, look for possible leaks. Repair the leaks found.
5.4.3 Sample line guidelines
Select a good, representative sampling point for the best instrument
performance. The sample must be representative of the entire system.
To prevent erratic readings:
- Collect samples from locations that are sufficiently distant from points of chemical additions to the process stream.
- Make sure that the samples are sufficiently mixed.
- Make sure that all chemical reactions are complete.
Install the sample tubing in an open channel or a pipe as shown in Figure 12
or Figure 13.
To connect the sample tube to a metal pipe, use a Swagelok reducer (e.g., SS-
400R-12).
The length of the sample tubing, between the water surface and the SAMPLE
fitting, must be 2.5 m (8.2 ft).
Figure 12 Sample line in an open channel
1| Sleeve for sample tubing| 4| Depth mark on tube| 7|
Sample tubing goes past end of sleeve (20 mm)
---|---|---|---|---|---
2| Sleeve bracket| 5 Sample tubing, 1/4-in. OD x 1/8-in. ID PFA|
8| Silt
3 Compression gland to hold sample tubing| 6| Clamps| 9|
Sleeve opening8
8 Sleeve must be below the low water level but more than 50 mm above the silt.
Figure 13 Sample line in a pipe
1 Bottom of analyzer| 4 Stainless steel tube, 1/4-in. OD x 1/8-in.
ID| 7 Dirt moves under the sample pipe
---|---|---
2 Sample tubing, 1/4-in. OD x 1/8-in. ID, PFA| 5 Distance between
analyzer and pipe9| 8 Larger pipe (1.5 to 2 times larger diameter) so
pressure does not increase
3 Connection between PFA tubing and stainless steel tube| 6 Air
bubbles move over the sample pipe| 9 Open drain as near to this location
as possible
9 A 30 cm (12 in.) difference in height gives a pressure of 30 mbar (04. psi)
if the flow rate is low.
5.4.4 Install a sample overflow chamber (optional)
For pressurized sample streams, install the optional Sample Overflow Chamber
(19BAS-031) in the sample line to supply sample at ambient pressure.
Figure 14 Installation of the sample overflow chamber
1| Sample inlet tube (flow rate: 0.7 to
1.7 L/min)
| 4| Vent tube| 7| Drain tube
---|---|---|---|---|---
2| Cap| 5| Sample overflow tube| 8| Manual drain valve
3 Sample tube to analyzer| 6| Open drain
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
NOTICE
Incorrect installation of the drain lines can cause liquid to go back into
the instrument and cause damage.
Make sure that the open drain used for the analyzer is in a ventilated area.
Oxygen and very small quantities of carbon dioxide, ozone and volatile gases
can be present in the waste liquids plumbed to the drain.
- Make the drain lines as short as possible.
- Make sure that the drain lines have a constant slope down.
- Make sure that the drain lines do not have sharp bends and are not pinched.
- Make sure that the drain lines are open to air and are at zero pressure.
Use the supplied1/4 -inch OD X 1/8-inch ID tubing to plumb the SAMPLE OUT
fitting to an open drain. Refer to Figure 15. The maximum distance between the
fitting and the drain is 2 m (6.5 ft).
Figure 15 Plumb the drain
1 Open drain| 2 1/4-inch OD x 1/8-inch ID tubing| 3 SAMPLE OUT
fitting
---|---|---
5.4.6 Connect instrument air
Use 3/8-in. OD tubing to connect instrument air (or the BioTector air
compressor and optional air filter pack) to the INSTRUMENT AIR fitting on the
left side of the analyzer.
Refer to the instrument air specifications in Specifications on page 3.
The air plumbed to the oxygen concentrator must be -20 °C dew point, 5 to 40
°C (41 to 104 °F) and contain no water, oil or dust. The optional air filter
pack is recommended. Oxygen quality: The oxygen supplied by the oxygen
concentrator is 93% oxygen minimum and the remaining gas is argon.
Safety precautions for compressed air:
- Use the same precautions necessary for high-pressure or compressed gas systems.
- Obey all local and national regulations and/or the manufacturer’s recommendations and guidelines.
5.4.7 Plumb the exhaust
Use ¼-in. OD PFA tubing to plumb the EXHAUST fitting to a ventilated area.
The maximum tubing length is 10 m (33 ft). If longer tubing is necessary, use
a larger ID tubing or pipe.
Make sure that the tubing has a constant downward slope from the analyzer so
condensation or liquid at the outlet of the tubing cannot freeze.
5.4.8 Plumb the reagents
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance
with local, regional and national regulations.
NOTICE
Keep dust and particles out of the reagent tubing or damage to the acid and
base pumps can occur. The acid and base pumps have a 50 µm particle tolerance.
Plumb the reagents to the analyzer. The top of the reagent containers must be
at the same level and no more than 0.6 m (2 ft) below the analyzer. Refer to
Figure 16.
Items supplied by the user:
- Personal protective equipment (refer to MSDS/SDS)
- Base reagent, 19 to 25 L—1.2 N sodium hydroxide (NaOH), low in carbonate
- Acid reagent, 19 to 25 L—1.8 N sulfuric acid (H2SO4) that contains 80-mg/L manganese sulfate monohydrate
Use ultra pure water that contains less than 5 µg/L (ppb) of organics to
prepare reagents.
For the reagent usage at 100% online time, refer to Table 14.
-
Put on the personal protective equipment identified in the safety data sheets (MSDS/SDS).
-
Put a reagent spill tray (bund) under the reagent containers to keep spills contained.
-
Assemble the supplied caps for the reagent containers. Refer to the documentation supplied with the caps. Only one of the two acid reagent cap assemblies (19PCS-021) is used.
Note: If a supplied cap is not the correct size for the reagent container, use the cap that came with the reagent container. For the acid reagent, make an opening in the cap and install the supplied tube fitting in the cap. For the base reagent, refer to Use a stainless steel fitting for the base reagent (optional) on page 45. -
Attach the weight supplied with each cap (item 4 in Figure 16) to the end of the reagent tube that goes in the reagent container.
Note: The weights contain a 70 µm filter. -
Install the caps on the reagent containers.
• Base reagent container—Install the cap that has a port on the side of the fitting.
The port is used to connect the supplied CO2 filter. As an alternative to the tube fitting supplied, use a stainless steel fitting. Refer to Use a stainless steel fitting for the base reagent (optional) on page 45.
• Acid reagent container—Install the cap that has 1/8-inch OD PFA tubing and a weight. -
Tighten the tube fittings on the caps so the weight stay at the bottom of the reagent containers.
-
Remove the tube that connects the ACID and BASE fittings on the left side of the analyzer. Discard the deionized water in the tube. Refer to illustrated steps 1–2 in Figure 16.
-
Plumb the reagent containers to the reagent fittings on the left side of the analyzer with 1/8-inch OD tubing. Refer to illustrated step 3 in Figure 16. Make the reagent lines as short as possible (2 m (6.5 ft) maximum).
-
Remove the tape from the CO₂ filter (item 3 in Figure 16).
-
Connect the supplied CO₂ filter to the cap of the base reagent container. Refer to illustrated step 4 in Figure 16. Make sure that the connection is air tight.
Note: If atmospheric CO₂ enters the base reagent container, the analyzer TIC and TOC readings will increase.
Figure 16 Reagent installation
1 Base reagent | 3 CO2 filter |
---|---|
2 Acid reagent | 4 Weight |
Table 14 Reagent usage
Reagent | Container size | 0–5000 µgC/L |
---|---|---|
Acid | 19 L | 122 days |
Base | 19 L | 122 days |
5.4.8.1 Use a stainless steel fitting for the base reagent (optional)
As an alternative to the plastic tube fitting supplied for the base reagent
container, use a stainless steel fitting. Refer to Figure 17. The tee fitting
must supply an air tight seal with the cap. If atmospheric CO2 enters the base
reagent container, the analyzer TIC and
TOC readings will increase.
Figure 17 Base reagent container
1 Air inlet| 3 Swagelok SS-400-3TST tee fitting, drilled to 7.0 mm
(0.28 in.)
---|---
2 CO2 filter| 4 Swagelok SS-45ST-N nut
5.4.9 Install the sample pump tubing
5.4.10 Connect the internal tubing
Refer to the illustrated steps that follow.
Note: The reagent pump tubing contains deionized water, which should be
discarded.
5.4.11 Connect the air purge
Connect the air purge to supply positive air pressure in the analyzer if
one or more of the statements that follow are true:
- There are corrosive gasses in the area.
- The analyzer is supplied as a “purge ready” system
A “purge ready” system has a Purge Air Inlet (¼-inch Swagelok fitting) on the
left side of the analyzer and no fan.
If the analyzer is not a “purge ready” system, contact technical support to
connect the air purge.
-
From the inner side of the electrical enclosure, remove the fitting blank (plug) from the Purge Air inlet.
-
Supply clean, dry instrument-quality air at 60 L/min to the Purge Air Inlet on the left side of the analyzer.
Clean, dry instrument-quality air is -20 °C dew point air that does not contain oil, water vapor, contaminants, dust or flammable vapor or gas. -
Install a 40 micron (or smaller) air filter in the air purge line.
Additional requirements:
• Make sure that all of the purge gas supplies are made to prevent contamination.
• Make sure that the purge gas pipe has protection from mechanical damage.
• Make sure that the air compressor intake for the purge gas is in an unclassified location.
• If the compressor intake line goes through a classified location, make sure that the compressor intake line is made of noncombustible material and made to prevent leakage of flammable gases, vapors or dusts into the purge gas. Make sure that the compressor intake line has protection from mechanical damage and corrosion.
Startup
6.1 Set the power to on
6.2 Set the language
Set the language that shows on the display.
- Push to go to the main menu, then select MAINTENANCE > SYSTEM CONFIGURATION > LANGUAGE.
- Select the language, then push . An asterisk (*) identifies the language selected.
6.3 Set the time and date
Set the time and date on the analyzer.
Note: When the time is changed, the analyzer may automatically start
tasks that are scheduled to start before the new time setting.
- Push to go to the main menu, then select OPERATION > TIME & DATE.
- Select an option. Use the UP and DOWN arrow keys to change the setting.
Option | Description |
---|---|
CHANGE TIME | Sets the time. |
CHANGE DATE | Sets the date. |
DATE FORMAT | Sets the date format (e.g., DD-MM-YY). |
6.4 Adjust the display brightness and contrast
- Push to go to the main menu, then select OPERATION > LCD ADJUST.
- Select an option. Use the UP and DOWN arrow keys to change the setting.
Option| Description
---|---
CONTRAST| Sets the display contrast (default: 50%).
BACKLIGHT| Sets the display brightness (default: 50%). - Push to save.
6.5 Examine the oxygen supply
Identify if there is CO2 contamination in the oxygen supply as follows:
- Let the oxygen concentrator operate for a minimum of 10 minutes.
- Select MAINTENANCE > DIAGNOSTICS > SIMULATE.
- Select MFC. Set the flow to 10 L/h.
- Push to start the mass flow controller (MFC).
- Operate the MFC for 10 minutes. The measured CO2 in the oxygen supply shows at the top of the display.
- If the reading is not ± 0.5% (B3500e, B3500c/s and B3500dw) or ± 2.5% (B3500ul) of the CO2 analyzer range, do the steps that follow:
a. Remove the CO2 filter from the base reagent container.
b. Install the CO2 filter between the cooler and the CO2 analyzer inlet port.
Note: Temporary connections can be made with EMPP tube.
c. Do steps 3 to 5 again.
If the reading is less than before, there is CO2 contamination in the oxygen supply. Identify if the CO2 analyzer has dirty lenses. Identify if the CO2 filters on the CO2 analyzer have contamination. Identify if the CO2 analyzer operation is correct.
If the reading is not less than before, there is no CO2 contamination in the oxygen supply.
d. Remove the CO2 filter from between the cooler and CO2 analyzer inlet port.
e. Connect the CO2 filter to the base reagent container.
6.6 Examine the pumps
Make sure that the acid pump, base pump and sample pump operate correctly as follows:
-
Select CALIBRATION > ZERO CALIBRATION > RUN REAGENTS PURGE to start a reagent purge cycle. A reagent purge cycle primes the reagents in the analyzer.
-
If the reagent tubing does not fill with reagents during the reagents purge cycle, prime the pumps manually as follows:
a. Remove the dip tubes from the reagent containers.
b. Close the reagent containers.
c. Put the dip tubes in a small container of deionized water or tap water.
d. Hold the container of water higher than the analyzer.
e. Do step 1 again.
f. Install the dip tubes in the reagent containers.
g. Do step 1 again. -
Make sure that there are no leaks at the sample pump, acid pump or base pump.
-
Put the outlet tubing of the acid pump into an empty 10-mL graduated cylinder.
-
Select MAINTENANCE > DIAGNOSTICS > SIMULATE.
-
Select ACID PUMP.
-
Select ON, then enter the number of pulses identified in Table 15.
Note: 400 pulses must be done as 2 x 200 pulses due to an internal system interlock. -
Push to start the acid pump.
-
Wait the number of pulses identified in Table 15.
400 pulses = 24 seconds, 16 pulses = 8 seconds -
Compare the volume of water in the graduated cylinder to Table 15. Record the volume.
-
Discard the reagent from the graduated cylinder. Make sure that the graduated cylinder is fully empty.
-
Do steps 4 and 6 to 10 again for the base pump.
Make sure that the difference in the measured volumes for the acid pump and base pump is 5% (0.2 mL) or less. -
Use a 1.5 mm Allen wrench to adjust the acid or base pump volume as necessary.
The adjustment dial is on the back of the pump. Only turn the adjustment dial a few degrees.
Note: Turn the adjustment dial clockwise to decrease the pump volume. Each notch on the adjustment dial is approximately 1 µL. -
Do steps 4 and 6 to 10 again for the sample pump.
-
Connect the tubing that was disconnected.
Table 15 Pump volumes
Pump | Pulses | Volume |
---|---|---|
ACID PUMP | 400 | 3.8 to 4.2 mL |
BASE PUMP | 400 | 3.8 to 4.2 mL |
SAMPLE PUMP | 16 | 5.5 to 7.5 mL |
6.7 Examine the valves
Make sure that the valves open and close correctly and there are no leaks. Refer to Analysis enclosure on page 54 for the valve locations.
-
Make sure there are no leaks at the valves.
-
Push to go to the SIMULATE menu.
-
Select REACTOR VALVE (MV3) on the display to open the reactor valve. The LED is on when the valve is open.
-
Do step 3 again for the valves that follow:
• SAMPLE VALVE (MV4)
• EXHAUST VALVE (MV1) 10
• STREAM VALVE (MV6)
• MANUAL VALVE (MV5) -
If the sample (out) valve (MV4), manual valve (MV5) or stream valve (MV6) does not open, disassemble the valve and clean the membrane seal.
6.8 Set the reagent volumes
- Select OPERATION > REAGENTS SETUP > INSTALL NEW REAGENTS.
- Change the reagent levels that show on the display as necessary.
- If the ZERO WATER setting in the SYSTEM CONFIGURATION menu is set to YES, plumb ultra pure water to the MANUAL fitting for the zero calibrations. The default setting for ZERO WATER is NO (no sample). For the best accuracy, set ZERO WATER to YES.
- If the SPAN CALIBRATION or SPAN CHECK setting is set to YES on the MAINTENANCE > COMMISSIONING > NEW REAGENTS PROGRAM menu, install the calibration standard before a span calibration is started. Refer to Plumb the calibration standard on page 74.
- Scroll down to START NEW REAGENT CYCLE, then push .
The analyzer fills all of the reagent lines with the new reagents and does a zero calibration.
In addition, if the SPAN CALIBRATION or SPAN CHECK setting is set to YES on the MAINTENANCE > COMMISSIONING > NEW REAGENTS PROGRAM menu, the analyzer does a span calibration or span check after the zero calibration.
If the CO2 LEVEL setting is set to AUTO, the analyzer sets the reaction check levels for TOC.
6.9 Measure ultra pure water
Measure ultra pure water five times to make sure that the zero calibration
is correct as follows:
-
Connect ultra pure water to the MANUAL fitting.
-
Set the analyzer to do five reactions at operation range 1. Refer to Measure a grab sample on page 81.
If the results of the measurements are almost 0 mgC/L CO2, the zero calibration is correct. -
If the results of the measurements are not almost 0 mgC/L CO2, do the steps that follow:
a. Do a pH test. Use ultra pure water for the sample. Refer to Do a pH test in the Maintenance and Troubleshooting Manual.
b. Measure the TIC pH. Make sure that the TIC pH is less than 2.
c. Measure the BASE pH. Make sure that the BASE pH is more than 12.
d. Measure the TOC pH. Make sure that the TOC pH is less than 2.
e. Measure the ultra pure water two more times. Refer to step 2.
f. Do the steps in Set the reagent volumes on page 54 again.
6.10 Analysis enclosure
Figure 18 shows the interior view of the analyzer.
Figure 19 shows how to open the inner door.
Approximately September 2022, the oxygen concentrator parts changed.
Figure 20 shows the internal view with the inner door open after the change.
Figure 21 shows the internal view with the inner door open before the change.
Figure 18 Internal view
1 Exhaust valve, MV1 | 9 Acid pump, P3 |
---|---|
2 Sample (out) valve, MV4 | 10 NDIR CO2 analyzer |
3 Sample and Manual valve, MV511 | 11 Ozone generator |
4 Sample 1 and Sample 2 valve, MV612 | 12 Cooler |
5 Sample sensor | 13 Ozone line filter |
6 Manual valve, MV512 | 14 Ozone destructor |
7 Sample pump, P1 | 15 Reactor valve, MV3 |
8 Base pump, P4 | 16 Mixer reactor |
11 Single-stream analyzers (one exterior sample fitting)
12 Dual-stream analyzers (two exterior sample fittings)
Figure 19 Open the inner door
Figure 20 Internal view—Inner door open
1 Mixer reactor motor, P2 | 8 Cable tie (2x) |
---|---|
2 Filter Board | 9 Oxygen concentrator |
3 Cooler fan | 10 Mass flow controller (MFC) |
4 Oxygen Control Board | 11 Exhaust filter |
5 Termination Board | 12 Pressure relief valve |
6 HEPA filter | 13 Air isolation valve, OV1 |
7 Oxygen pressure regulator
Figure 21 Internal view—Inner door open (before September 2022)
1 Mixer reactor motor, P2 | 8 Oxygen pressure regulator |
---|---|
2 Filter Board | 9 HEPA filter |
3 Cooler fan | 10 Mass flow controller (MFC) |
4 Oxygen Control Board | 11 Oxygen concentrator rotary valve, OV2 |
5 Molecular sieve beds for oxygen concentrator | 12 Exhaust filter |
6 Termination Board | 13 Pressure relief valve |
7 Oxygen tank | 14 Air isolation valve, OV1 |
Configuration
7.1 Set the measurement interval
Set the time between reactions to set the measurement interval.
- Select MAINTENANCE > COMMISSIONING > REACTION TIME.
- Select an option.
Option | Description |
---|
REACTION
TIME| Shows the total reaction time (minutes and seconds) for operation range
1 (default: 5m00s). The analyzer calculates the total reaction time with the
OXIDATION PROGRAM 1 settings in the SYSTEM PROGRAM menu.
INTERVAL| Sets the time between reactions. Options: 0 (default) to 1440
minutes (1 day).
Note: When the analyzer automatically increases the reaction time because
of a high level of TIC and/or TOC in the sample, the analyzer
subtracts the added reaction time from the interval time.
TOTAL| Shows the total reaction time plus the interval time.
7.2 Set the sample pump times
Set the forward and reverse times for the sample pumps.
Note: If the forward or reverse times are more than the maximum time, the
analyzer adjusts the measurement interval setting. The maximum times are based
on the SYSTEM PROGRAM 1 settings.
-
Do a sample pump test for each sample stream to identify the correct forward and reverse times. Refer to Do a sample pump test on page 59.
-
Select MAINTENANCE > COMMISSIONING > SAMPLE PUMP.
The default sample pump times show for each stream (default: 45s forward, 60s reverse). -
Enter the FORWARD time from the sample pump test.
The analyzer sets the REVERSE time as the FORWARD time plus 15 seconds.
Note: The REVERSE time for a Manual stream is only set if an optional manual bypass valve is installed. The manual bypass valve sends the previous grab sample (or calibration standard) out the drain line.
7.2.1 Do a sample pump test
Do a sample pump test to identify the correct forward and reverse times for
the sample pump for each sample stream.
- Select MAINTENANCE > DIAGNOSTICS > PROCESS TEST > SAMPLE PUMP TEST.
- Select an option.
Option | Description |
---|---|
VALVE | Sets the SAMPLE or MANUAL valve used for the test. For example, to |
select the SAMPLE valve, select STREAM 1.
PUMP
FORWARD
TEST| Starts the sample pump in the forward direction.
Note: First select PUMP REVERSE TEST to empty the sample lines, then select
PUMP FORWARD TEST.
1. Push to stop the timer when the sample comes out the SAMPLE OUT fitting on
the left side of the analyzer.
2. Record the time on the display. Add 10 seconds to the time. The time is
the correct forward time for the selected stream.
PUMP
REVERSE TEST| Starts the sample pump in the reverse direction.
SAMPLE PUMP| Goes to the MAINTENANCE > COMMISSIONING > SAMPLE PUMP menu to set
the forward and reverse times for each sample stream.
7.3 Set the stream sequence and operation range
Set the sample stream sequence, the number of reactions to do at each sample
stream and the operation range for each sample stream.
- Select MAINTENANCE > COMMISSIONING > STREAM PROGRAM.
- Select an option.
Option | Description |
---|---|
SAMPLER | Note: The B3500 analyzers cannot be operated with a sampler |
(venturi or vacuum samplers).
Set to YES if a sampler is used with the analyzer (default: NO). When SAMPLER
is set to YES (default), the sampler time shows on the SAMPLE PUMP screen.
CONTROL| Set to BIOTECTOR (default) to control the stream sequence and
operation ranges with the analyzer.
Set to EXTERNAL to control the stream sequence and operation ranges with an
external device (e.g., Modbus master).
STREAM INPUT is not used with single-range analyzers.
START-UP
RANGE| Note: The START-UP RANGE setting is available when CONTROL is set
to BIOTECTOR and the first operation range setting for a stream is set to
AUTO.
Sets the operation range used for the first reaction when the analyzer starts
(default: 1).
RANGE LOCKED| Note: The RANGE LOCKED setting is available if one or more
of the
RANGE settings for the stream sequence is set to AUTO.
Sets the operation range to change automatically (NO) or to stay at the START-
UP RANGE setting (YES, default).
PROGRAMMED
STREAMS| Shows the number of streams installed and configured.
VOC CYCLE| Note: VOC CYCLE is only available on analyzers configured as a
VOC system.
Sets the frequency of TC analysis cycles, and TC TIC and NPOC analysis cycles
(default: 1). When VOC CYCLE is set to 1, the analyzer does TC TIC and NPOC
analysis cycles continuously.
When VOC CYCLE is set to 10, the analyzer does one TC TIC and NPOC analysis
cycle and then nine TC analysis cycles.
TC CHECK STM
1 to 3| Note: TC CHECK STM is only available on analyzers configured as a
VOC system.
Sets the % and mgC/L ranges for TC analysis.
When VOC CYCLE is more than 1, the analyzer does a TC TIC and NPOC analysis
cycle for the stream if the TC result is not within the % and mgC/L ranges
(default: 10%, 5.000 µgC/L). If the TC results is within the % and mgC/L
ranges, the analyzer does the scheduled TC analysis cycles.
STREAM x, x
RANGE x| Note: If CONTROL is set to EXTERNAL, an external device (e.g.,
Modbus master) controls the stream sequence and operation ranges. Sets the
number of reactions and the operation range for each stream. STREAM—The first
setting is the stream valve number. The second setting is the number of
reactions done at the sample stream before the analyzer does reactions with
the next sample stream. When STREAM is set to “- , -” and RANGE is set to “-“,
the stream is not measured. RANGE —Sets the operation range for each sample
stream. Options: 1, 2, 3 (default) or AUTO. Select OPERATION > SYSTEM RANGE
DATA to see the operation ranges. Note: The AUTO (automatic) range option is
disabled in analyzers with more than one stream.
7.4 Configure the COD and BOD settings
Set the analyzer to show COD and/or BOD information on the Reaction Data
screen as necessary. Set the values used to calculate the COD and/or BOD
results.
- Select MAINTENANCE > COMMISSIONING > COD/BOD PROGRAM.
- Select COD PROGRAM or BOD PROGRAM.
- Select an option.
Option | Description |
---|---|
DISPLAY | Sets the analyzer to show COD and/or BOD information on the Reaction |
Data screen and show the COD and/or BOD (mgO/L) results on a 4–20 mA output if
configured (default: NO).
STREAM 1–3| The first setting is the overall factor (default: 1.000). Refer to
the equation that follows. The second setting is the offset factor (default:
0.000). The stream factors for each stream come from the procedures in the
information sheet I030. TOC to COD or BOD Correlation Method. STREAM 1 factors
are used for manual samples and calibration standards.
COD (and/or BOD) = Overall factor × { (TOC FACTOR × TOC) } + Offset factor
TOC| Sets the TOC FACTOR (default: 1.000).
FACTOR| Note: In TC analysis mode, TC FACTOR shows on the display and is
used in the equation as an alternative to the TOC FACTOR.
7.5 Configure the DW PROGRAM settings
DW PROGRAM settings are not used with B3500e or B3500ul analyzers. Do not
change the default settings.
7.6 Configure CF PROGRAM settings
CF PROGRAM settings are not used with the B3500 analyzers, because the samples
cannot contain oils or grease. Do not change the default settings.
7.7 Configure the install new reagents settings
Configure the analyzer options for the OPERATION > REAGENTS SETUP > INSTALL
NEW REAGENTS function.
- Select MAINTENANCE > COMMISSIONING > NEW REAGENTS PROGRAM.
- Select an option.
Option | Description |
---|---|
SPAN | Sets the analyzer to do a span calibration during the INSTALL NEW |
CALIBRATION | REAGENTS cycle (default: NO). Refer to Start a span calibration |
or span check on page 72 for the span calibration function.
If set to YES, make sure to install the calibration standard before a span
calibration is started. Refer to Plumb the calibration standard
on page 74.
SPAN CHECK| Note: It is not possible to set SPAN CALIBRATION and SPAN
CHECK to YES.
Sets the analyzer to do a span check during the INSTALL NEW REAGENTS cycle
(default: NO). Refer to Start a span calibration or
span check on page 72 for the span check function.
If set to YES, make sure to install the calibration standard before a span
check is started. Refer to Plumb the calibration standard on page 74.
AUTOMATIC RESTART| Set the analyzer to go to back to operation when the
INSTALL NEW REAGENTS cycle is completed (default: YES).
7.8 Set reagent monitoring
Configure the alarm settings for low reagents and no reagents. Set the reagent volumes.
- Select MAINTENANCE > COMMISSIONING > REAGENTS MONITOR.
- Select an option.
Option | Description |
---|
REAGENTS
MONITOR| Sets the Reagent Status screen to show on the display (default: YES).
LOW REAGENTS| Sets the low reagents alarm as a notification or a warning.
Options:
NOTE (default) or WARNING
LOW REAGENTS AT| Sets the number of days before the reagent containers are
empty when a 85_LOW REAGENTS alarm should occur (default: 20 days).
Note: The analyzer calculates the number of days before the reagent containers
are empty.
NO REAGENTS| Sets the no reagents alarm as a notification, warning or a fault.
NOTE—A relay for notifications is set to on when a no reagents alarm
occurs if configured. WARNING (default)—A relay for warning events is set to
on and a 20_NO REAGENTS warning occurs if configured. FAULT
—The fault relay is set to on, measurements stop and a 20_NO REAGENTS fault
occurs.
ACID VOLUME| Sets the volume (liters) of the acid reagent in the reagent
container.
BASE VOLUME| Sets the volume (liters) of the base reagent in the reagent
container.
7.9 Configure the analog outputs
Set what is shown on each 4–20 mA output, the full scale range of each 4–20 mA
output and when each 4–20 mA output changes. Set the fault level for the 4–20
mA outputs.
After the analog outputs are configured, do a 4–20 mA output test to make sure
that the correct signals are received by the external device. Refer to the
instructions in the Maintenance and Troubleshooting manual.
- Select MAINTENANCE > COMMISSIONING > 4-20mA PROGRAM.
- Select OUTPUT MODE.
- Select an option.
• DIRECT (default)—Refer to Table 16 to configure the settings. Configure each channel (4–20 mA output) to show a specified stream (STREAM 1) and result type (e.g., TOC).
• BASIC—The 4–20 mA outputs (Channels 1–4) that are set to show STREAM 1 also show the zero and span calibration/check results. Refer to Table 16 to configure the settings.
• STREAM MUX—Refer to Table 17 to configure the settings. The CHANNEL 1 setting cannot be changed. Configure Channels 2 to 4 (4–20 mA Outputs 2 to 4 ) to each show one result type (e.g., TOC). The 4–20 mA outputs can show a maximum of 35 results. Refer to 4–20 mA output modes in the Advanced Configuration Manual for more information.
• FULL MUX—Refer to Table 18 to configure the settings. The CHANNEL 1–4 settings cannot be changed. No other channels are used. The 4–20 mA outputs can show a maximum of 35 results. Refer to 4-20 mA output modes in the Advanced Configuration Manual for more information.
Table 16 Direct mode settings
Option | Description |
---|---|
CHANNEL 1–4 | Sets what is shown on the 4–20 mA Outputs 1–4 (Channel 1–4), the |
full scale range of each 4–20 mA output and when each 4–20 mA output changes.
First setting—Sets what the 4–20 mA output shows.
• STREAM # (default)—Shows the selected sample stream (e.g., STREAM
1).
• MANUAL # —Shows the selected manual grab sample (e.g., MANUAL 1).
• CAL —Shows the zero and span calibration results.
• CAL ZERO —Shows the zero calibration results.
• CAL SPAN —Shows the span calibration results.
Second setting—Sets the result type. Options: TOC, TIC, TC, VOC, COD or BOD.
In TIC + TOC analysis mode, TC is the sum of TIC and TOC.
Third setting—Sets the result that the output shows as 20 mA (e.g., 25µgC/L).
The output shows 4 mA for 0 µgC/L.
Fourth setting—Sets when the outputs change.
• INST —The output changes at the end of each reaction.
• AVRG —The output (average result of the last 24 hours) changes at
the AVERAGE UPDATE time selected in SYSTEM CONFIGURATION > SEQUENCE PROGRAM >
AVERAGE PROGRAM.
Note: The 4–20 mA outputs that show calibration results change when
the system completes the number of calibration reactions set in MAINTENANCE >
SYSTEM CONFIGURATION > SEQUENCE PROGRAM > ZERO PROGRAM or SPAN PROGRAM .
SIGNAL FAULT| Sets all of the 4–20 mA outputs to change to the FAULT LEVEL
setting when a fault occurs.
YES (default)—All of the 4–20 mA outputs change to the FAULT LEVEL
setting when a fault occurs.
NO —The 4–20 mA outputs continue to show the results when a fault occurs.
FAULT LEVEL| Sets the fault level (default: 1.0 mA).
Table 16 Direct mode settings (continued)
Option | Description |
---|
OUTPUT <
4mA
| Sets the percentage applied to the result shown at the output if the output
value is less than 4 mA, which is a negative result (default: 0%).
For example, if the OUTPUT setting is 100%, the analyzer sends 100% of the
negative result as the 4–20 mA signal. If the OUTPUT setting is 50%, the
analyzer sends 50% of the negative result as the 4–20 mA signal. When the
OUTPUT setting is 0%, the analyzer does not send a negative result. The
analyzer shows a negative result as 4 mA (0 µgC/L).
EXCLUDE RESULTS| Note: The EXCLUDE RESULTS setting is only available when
the OUTPUT MODE setting in the 4-20mA PROGRAM menu is set to DIRECT .
Sets the number of sample reactions that the analyzer ignores after a zero
calibration, span calibration, zero check or span check (default: 3). The
analyzer keeps the calibration output relay active until the analyzer
completes the selected number of sample reactions.
Table 17 Stream multiplex mode settings
Option | Description |
---|---|
CHANNEL 1–4 | Sets the result type that shows on the 4–20 mA outputs (Channels |
1–4). Options: TOC, TIC, TC, VOC, COD or BOD. The Channel 1 setting cannot be
changed. In TIC + TOC analysis mode, TC is the sum of TIC and TOC.
Note: The CHANNEL # and OUTPUT # settings identify what Channels 2 to 4
show. Refer to the OUTPUT option description for more information .
OUTPUT PERIOD| Sets the time to show a full set of reaction results (results
sequence) on the 4–20 mA outputs plus the idle time before the next results
sequence starts (default: 600s).
If a new result is available during the idle period, the results sequence
starts. The idle period is not completed.
If a new result is available before a results sequence is completed, the
analyzer shows the new result then continues the result sequence.
Make sure that OUTPUT PERIOD is sufficient to complete a results sequence. Use
the formulas that follow to calculate the minimum OUTPUT PERIOD:
• Stream multiplex mode —OUTPUT PERIOD = [2 x (SIGNAL HOLD TIME) + 1
second] x [number of streams] • Full multiplex mode —OUTPUT PERIOD =
{[2 x (SIGNAL HOLD TIME) + 1 second] x (number of result types)]} x [number of
streams]
SIGNAL HOLD TIME| Sets the amount of time that Channel 1 holds a signal before
Channel 1 goes to 4 mA (change level) or to the next stream identification
level (e.g., 6 mA = STREAM 2). Default: 10s
When the SIGNAL HOLD TIME setting is 10 seconds, Channels 2 to 4 hold their
signal for 20 seconds (2 x SIGNAL HOLD TIME).
SIGNAL FAULT| Refer to SIGNAL FAULT in Table 16.
FAULT LEVEL| Refer to FAULT LEVEL in Table 16.
OUTPUT < 4mA| Refer to OUTPUT < 4mA in Table 16.
OUTPUT 1–35| Sets what is shown on the 4–20 mA outputs (Channels 2 to 4 ), the
full scale value of each 4–20 mA output and when each 4–20 mA output changes.
The result type in the OUTPUT setting (e.g., TOC) identifies the channel
(Channel 2 to 4 ) on which the result shows. For example, if CHANNEL 3 is set
to TOC and the OUTPUT 1 setting has a result type of TOC, the result
identified in the OUTPUT 1 setting shows on Channel 3. If OUTPUT 1 is set to
STREAM 1, TOC, 5000 µgC/L and INST, when the Channel 1 signal identifies
STREAM 1, Channel 3 shows the TOC result where 5000 µgC/L is shown as 20 mA.
Refer to CHANNEL in Table 16 for descriptions of the four settings for each
OUTPUT setting.
Table 18 Full multiplex mode settings
Option | Description |
---|---|
CHANNEL 1–4 | The CHANNEL 1–4 settings cannot be changed. |
Note: The OUTPUT # settings identify what Channels 3 and 4 show .
OUTPUT PERIOD| Refer to OUTPUT PERIOD in Table 17.
SIGNAL HOLD TIME| Sets how the long Channel 1 and 2 hold their signal before
the channels go to 4 mA (change level or not defined level) or to the next
stream identification level or result type level. Default: 10s
When the SIGNAL HOLD TIME setting is 10 seconds, Channel 3 holds the signal
for 20 seconds (2 x SIGNAL HOLD TIME).
SIGNAL FAULT| Refer to SIGNAL FAULT in Table 16.
FAULT LEVEL| Refer to FAULT LEVEL in Table 16.
OUTPUT < 4mA| Refer to OUTPUT < 4mA in Table 16.
OUTPUT 1–35| Sets what is shown on the 4–20 mA outputs (Channels 3 and 4), the
full scale value of each 4–20 mA output and when each 4–20 mA output changes.
The result type in the OUTPUT setting (e.g., TOC) identifies the channel on
which the result shows. For example, if CHANNEL 3 is set to TOC and the OUTPUT
1 setting has a result type of TOC, the result identified in the OUTPUT 1
setting shows on Channel 3. If OUTPUT 1 is set to STREAM 1, TOC, 5000 µgC/L
and INST, when the Channel 1 signal identifies STREAM 1, Channel 3 shows the
TOC result where 5000 µgC/L is shown as 20 mA.
Refer to CHANNEL in Table 16 for descriptions of the four settings for each
OUTPUT setting.
7.10 Configure the relays
Configure the relay idle conditions and the conditions that set the relays to on. After the relays are configured, do a relay test to make sure that the relays operate correctly. Refer to the instructions in the Maintenance and Troubleshooting manual.
- Configure the relays as follows:
a. Select MAINTENANCE > COMMISSIONING > PROGRAMMABLE OUTPUTS >PWR BRD OUT 1.
Note: PWR BRD OUT 1 is Relay 1 on the Power and Input/Output Board. Refer to Power, analog output and relay terminals on page 29.
b. Select DEFAULT STATE.
c. Set the relay to N/E (normally energized) or N/D (normally de-energized).
d. Select the conditions that trigger the relay. Refer to Table 19.
Note: One or more conditions can be selected. For example, when CAL SIGNAL and MAINT SIGNAL are selected, the relay is set to on when a zero or span calibration starts or when the maintenance switch is set to on. The selected conditions are marked with an asterisk “*”.
Table 19 RELAY settings
Setting | Description | Setting | Description |
---|---|---|---|
– – – | No setting | ZERO CAL | The relay is set to on when the |
manual valve opens during a zero calibration or zero check.
STOP| The relay is set to on when the analyzer is stopped.
Note: Remote standby does not set the relay to on .| CAL SIGNAL|
The relay is set to on when a zero or span calibration, or a zero or span
check starts.
FAULT| The relay is set to on when a system fault occurs (normally
energized relay).| STREAM 1–3| The relay is set to on when the sample
inlet valve opens.
Table 19 RELAY settings (continued)
Setting | Description | Setting | Description |
---|---|---|---|
WARNING | The relay is set to on when a warning occurs (normally | ||
energized relay). | MANUAL 1–3 | The relay is set to on when the manual |
valve opens.
NOTE| The relay is set to on when a notification is saved to the fault
archive.| SAMPLE STATUS 1–3| The relay is set to on when there is no
sample or the sample quality is less than 75% (default) for stream 1 or 2. For
example, when there are a lot of air bubbles in the stream/manual grab sample
lines.
SAMPLER FILL| The relay is set to on when from the start of the sampler
fill time to the completion of the sample injection. The relay controls the
sampler.| STM ALARM 1–3| The relay is set to on when a selected alarm
condition occurs. The alarm conditions are set on the ALARM PROGRAM screen.
Refer to step 2 that follows.
SAMPLER EMPTY| The relay is set to on for 5 seconds after the sample
pump reverse operation is complete. The relay controls the sampler.| CO2
ALARM 1–3| The relay is set to on when a CO2 ALARM occurs for stream 1 or
2. Refer to step 2 that follows.
SAMPLER ERROR| The relay is set to on when a BioTector sampler error
occurs.| 4-20mA CHNG| The relay is set to a 4–20 mA change flag relay.
The relay is set to on for a period of 10 seconds when a new result on any
sample stream causes an analog output value to change.
SYNC| The relay is set to a synchronization relay. A synchronization
relay is used to synchronize the analyzer with external control devices.|
4-20mA CHNG 1–3| The relay is set to a 4–20 mA change flag relay for a
specific sample stream (1–2). The relay is set to on for a period of 10
seconds when a new result on stream 1 or 2 causes an analog output value to
change.
REMOTE STANDBY| The relay is set to on when the remote standby switch
(digital input) is set to on.| 4-20mA READ| The relay is set to on when
the 4–20 mA outputs are set to stream multiplex or full multiplex mode and
there are valid/stable values on the 4–20 mA outputs.
MAN MODE TRIG| The relay is set to on when manual reactions (grab sample
measurements) are started at the keyboard or with the Manual-AT Line option.
Note: The Manual-AT Line option is a small box with only a green button.
The Manual- AT Line cable is connected to the analyzer.| SAMPLE FAULT
1–3| The relay is set to on when the external SAMPLE FAULT 1 input signal
is activated.
MAINT SIGNAL| The relay is set to on when the maintenance switch
(digital input) is set to on.| COMPRESSOR| The relay is set to on when
the compressor (Valve 1–J7 on the Oxygen Control Board) is on.
TEMP SWITCH| The relay is set to on when the temperature switch of the
analyzer sets the fan to on (default: 25 °C).| EQUALIZATION| The relay
is set to on when the sample pump operates in forward direction until the
sample valve opens to let sample leave the analyzer.
CAL| The relay is set to on when the manual valve opens during a span
calibration or span check.| SAMPLE PUMP REVERSE| The relay is set to on
when the sample pump operates in the reverse direction.
2. Select MAINTENANCE > COMMISSIONING > ALARM PROGRAM.
3. Configure the relays set to ALARM and CO2 ALARM if configured.
Option | Description |
---|---|
ALARM 1–3 | Sets the idle condition of the ALARM relay and the condition that |
sets an ALARM relay to on.
First setting—Sets the parameter (TOC, TC, VOC, COD, BOD or TIC).
Second setting—Selects stream 1 or 2.
Third setting—Sets the minimum concentration (default: 10 µg/L) that sets the
alarm relay to on at the end of a reaction for any of the sample streams.
Note: For the TIC + TOC and VOC analysis types, the TOC results of the last
completed reaction control the alarm relays. For the TC analysis type, the TC
results control the alarm relays.
CO2 ALARM 1–3| Note: Only use the CO2 ALARM settings with multi-stream systems
that operate on fixed operation ranges, or systems that operate on a single
operation range. Do not use the CO2 ALARM setting with an analyzer that uses
automatic range change.
Sets the CO2 peak value for stream 1 or 2 that sets the CO2 ALARM relay to on
(e.g., 500 ppm). The default is 500 ppm. Carefully select the CO2 peak value.
Think about the temperature effect, which could have a important effect on the
CO2 peaks. To disable the alarm relay, select 0.0 ppm.
The CO2 alarm identifies a possible high TOC (COD and/or BOD if programmed)
level. The CO2 alarm supplies a warning of an unusually high TOC result from
the rising slope of the CO2 peak during a reaction.
Note: In TIC + TOC and VOC analysis types, the CO2 peak used for the CO2
alarm is the TOC CO2 peak. In the TC analysis type, the CO2 peak used for the
CO alarm is the TC CO peak.
4. Select MAINTENANCE > SYSTEM CONFIGURATION > OUTPUT DEVICES > SYSTEM
OUTPUTS.
5. Select an option.
Option | Description |
---|---|
POWERED ALLTIME | Sets if the relays are powered all of the time, even if the |
analyzer is stopped or paused (YES) or powered only when necessary (NO).
VALVE
ACTIVATION| Sets when the multi-stream valve changes to the next sample
stream.
SPF (sample pump forward, default)—Sets the valve to the next stream when the
sample pump starts forward to pull sample from the next stream.
SPR (sample pump reverse)—Sets the valve to the next stream when the reverse
operation of the sample pump is completed or when power is first supplied to
the analyzer after power was removed.
Note: If SAMPLER is set YES on the STREAM PROGRAM screen, VALVE
ACTIVATION is set to SPF/SAMPLER.
OUTPUT 1–6| Note: OUTPUT 4-6 menus are reserved for future use.
Sets the internal output settings that are associated with system operation.
The relay outputs are on the motherboard and are optional features.
STREAM 1 is always functional as default in the system software.
One or more conditions can be selected. For example, when CAL SIGNAL and MAINT
SIGNAL are selected, the relay is set to on when a zero or span calibration
starts or when the maintenance switch is set to on. The selected conditions
are marked with an asterisk “*”. Refer to Table 19.
7.11 Configure the communication settings
Configure the communication settings for the output devices: MMC/SD card
and/or Modbus.
Configuration
Note: Analyzer communication with a printer or Windows PC is no longer available.
- Select MAINTENANCE > COMMISSIONING > DATA PROGRAM.
- Select MMC/SD CARD.
- Select an option.
Option | Description |
---|---|
PRINT MODE | Sets the type of data sent to the MMC/SD card. Options: STANDARD |
or ENGINEERING (default).
Refer to Table 24 on page 84 and Table 25 on page 85 for descriptions of the
reaction data sent when STANDARD or ENGINEERING is
selected.
Note: The manufacturer recommends that PRINT MODE is set to
ENGINEERING so that troubleshooting data is saved.
REACTION ONLINE| No longer used. Sends the reaction data to the printer at the
end of each reaction (default: NO).
FAULT ON-LINE| No longer used. Sends the faults and warnings to the printer
when a fault or warning occurs (default: NO).
CONTROL
CHARS| Sends the control characters with the Modbus RS232 data (default: NO).
DECIMAL| Sets the type of decimal point included in the reaction data sent to
the MMC/SD card (default: POINT). Options: POINT (.) or COMMA (,)
7.12 Configure the Modbus TCP/IP settings
If the optional Modbus TCP/IP module is installed in the analyzer, configure
the Modbus settings.
Note: The Modbus register maps are supplied in the Advanced Configuration
Manual.
- Select MAINTENANCE > COMMISSIONING > MODBUS PROGRAM.
- Select an option.
Option | Description |
---|---|
MODE | Shows the Modbus operating mode: BIOTECTOR. The MODE setting cannot be |
changed.
BAUDRATE| Sets the Modbus baudrate for the instrument and the Modbus master
device (1200 to 115200 bps, default: 57600).
Note: For Modbus TCP/IP, do not change the BAUDRATE setting.
The RTU-to-TCP converter uses the default BAUDRATE setting.
PARITY| Sets the parity to NONE (default), EVEN, ODD, MARK or SPACE.
Note: For Modbus TCP/IP, do not change the PARITY setting. The
RTU-to-TCP converter uses the default PARITY setting.
DEVICE BUS
ADDRESS| Sets the Modbus address of the instrument (0 to 247, default: 7).
Enter a fixed address that a Modbus protocol message cannot change.
If DEVICE BUS ADDRESS is set to 0, the analyzer will not communicate with the
Modbus Master.
MANUFACTURE ID| Sets the manufacturer ID of the instrument (default: 1 for
Hach).
DEVICE ID| (Optional) Sets the class or family of the instrument (default:
2816).
SERIAL NUMBER| Sets the serial number of the instrument. Enter the serial
number that is on the instrument.
LOCATION TAG| Sets the location of the instrument. Enter the country where the
instrument is installed.
FIRMWARE REV| Shows the firmware revision installed on the instrument.
REGISTERS MAP REV| Shows the Modbus register map version used by the
instrument.
Refer to the Modbus register maps in the Advanced Configuration Manual.
7.13 Save the settings to memory
Save the analyzer settings to internal memory or an MMC/SD card. Then install
the saved settings on the analyzer as necessary (e.g., after a software update
or to go back to the previous settings).
- Select MAINTENANCE > SYSTEM CONFIGURATION > SOFTWARE UPDATE.
- Select an option.
Option
LOAD FACTORY
CONFIG
SAVE FACTORY
CONFIG
LOAD CONFIG FROM
MMC/SD CARD
SAVE CONFIG TO
MMC/SD CARD
UPDATE SYSTEM
SOFTWARE| Description
Installs the analyzer settings saved to internal memory with the SAVE FACTORY
CONFIG option.
Saves the analyzer settings to internal memory.
Installs the analyzer settings from the MMC/SD card with the SAVE CONFIG TO
MMC/SD CARD option.
Note: Use this option to go back to the previous settings or install the
settings after a software update.
Saves the analyzer settings to the syscnfg.bin file on the MMC/SD card.
Note: The MMC/SD card supplied with the analyzer contains the factory default
settings in the syscnfg.bin file.
Installs a software update. Contact the manufacturer or distributor for the
software update procedure.
---|---
7.14 Set security passwords for menus
Set a four-digit password (0001 to 9999) to restrict access to a menu level as
necessary.
Set a password for one or more of the menu levels that follows:
- OPERATION
- CALIBRATION
- DIAGNOSTICS
- COMMISSIONING
- SYSTEM CONFIGURATION
1. Select MAINTENANCE > SYSTEM CONFIGURATION > PASSWORD.
2. Select a menu level, then enter a 4-digit password.
Note: When a password is set to 0000 (default), the password is disabled.
7.15 Show the software version and serial number
Show the contact information for technical support, the software version or
the serial number of the analyzer.
- Select MAINTENANCE > COMMISSIONING > INFORMATION.
- Select an option.
Option | Description |
---|---|
CONTACT INFORMATION | Shows the contact information for technical support. |
SOFTWARE | Shows the software version installed on the analyzer. Shows the date |
the software version was released.
IDENTIFICATION| Shows the serial number of the analyzer.
Calibration
8.1 Start a zero calibration or zero check
Start a zero calibration after a maintenance task or after reagent replacement
or addition.
After maintenance, measure water ten times before a zero calibration is done
to remove contamination from the analyzer.
A zero calibration sets the zero offset values. Start a zero check to see if
the zero offset values set by the analyzer are correct as necessary.
The zero adjust values remove the effect that the items that follow can have
on measurement results:
- Contamination in the analyzer
- Organic carbon in the acid reagent and base reagent
- Absorbed CO2 in the base reagent
1. Select CALIBRATION > ZERO CALIBRATION.
2. Select an option.
Option | Description |
---|---|
ZERO ADJUST | (Optional) Sets the zero adjust values for zero calibrations |
manually for each range (1, 2 and 3) and each parameter. When the zero adjust
values
are entered manually, the analyzer records the information in the reaction
archive with the prefix “ZM” (zero manual).
Note: The TOC zero adjust values are the zero offset values in mgC/L measured
by the CO2 analyzer.
RUN
REAGENTS
PURGE| Starts a reagent purge cycle, which primes the reagents in the
analyzer.
Note: To change the pump operation time for the reagent purge cycle, select
MAINTENANCE > SYSTEM CONFIGURATION > SEQUENCE
PROGRAM > REAGENTS PURGE.
RUN ZERO
CALIBRATION| Starts a zero calibration, which sets the zero adjust values
automatically for each range (1, 2 and 3) for each parameter. Zero calibration
reactions have the prefix “ZC”. Stop measurements before a zero calibration is
started.
Note: Based on the ZERO WATER setting in the SYSTEM CONFIGURATION menu, a zero
calibration reaction is a reaction with no sample or with ultra pure water,
and the sample pump does not operate in the reverse direction.
If the ZERO WATER setting in the SYSTEM CONFIGURATION menu is set to YES, make
sure to plumb ultra pure water (< 5 ppb TOC) to the
MANUAL fitting before a zero calibration is started. The default setting for
ZERO WATER is NO (no sample). For the best accuracy, set ZERO WATER to YES.
Note: Approximately 500 to 800 mL of ultra pure water is used for a zero
calibration or zero check.
At the end of a zero calibration, the analyzer does the actions that follow:
• TOC zero adjust value—The analyzer uses the uncalibrated TOC measurement
(not the results that show on the display) to calculate and set new zero
adjust values.
• CO2 LEVEL setting—The analyzer sets the CO2 LEVEL setting to AUTO
(automatic) on the REACTION CHECK screen. Then a new reaction check CO2 level
is saved.
• CO2 level—The analyzer compares the CO2 level to the BASE CO2 ALARM setting
in the FAULT SETUP menu. If the measured CO2 level is more than the BASE CO2
ALARM value, a 52_HIGH CO2 IN BASE warning occurs.
RUN ZERO CHECK
Starts a zero check. A zero check is the same as a zero calibration, but
the analyzer does not change of the zero adjust values or the CO2 LEVEL
settings. Zero check reactions have the prefix “ZK”. Stop measurements before
a zero check is started.
If the ZERO WATER setting in the SYSTEM CONFIGURATION menu is set to YES, make
sure to plumb ultra pure water to the ZERO WATER or MANUAL fitting before a
zero check is started. If those fittings are not available, plumb ultra pure
water to the SAMPLE 1 fitting. The default setting for ZERO WATER is NO (no
sample). For the best accuracy, set ZERO WATER to YES. At the end of a zero
check, the analyzer does the actions that follow:
• The analyzer identifies the zero response at each range and shows the
suggested zero adjust values in brackets “[ ]” near the zero adjust values set
by the analyzer.
Note: Manually change the zero adjust value settings on the RUN ZERO CHECK
screen if necessary.
• The analyzer compares the CO2 level to the BASE CO2 ALARM setting in the
FAULT SETUP menu. If the measured CO2 level is more than the BASE CO2 ALARM
value, a 52_HIGH CO2 IN BASE warning occurs.
ZERO PROGRAM
Note: Do not change the default setting unless necessary. Changes can
have a negative effect on the zero adjust values.
Sets the number of zero reactions done during a zero calibration or zero check
for each operation range (R1, R2 and R3).
Note: The analyzer does not do a zero reaction for the operation ranges
set to 0. The analyzer calculates the zero adjust values for the operation
ranges set to 0.
ZERO AVERAGE
Note: Do not change the default setting unless necessary. Changes can have a
negative effect on the zero adjust values.
Sets the number of zero reactions averaged for each operation range at the end
of the zero cycles for all of the measured parameters.
8.2 Start a span calibration or span check
Set the operation range and the calibration standards for span calibrations. Start a span calibration to set the span adjust values, which adjust the measurement results. Start a span check to identify if the span adjust values saved to the analyzer are correct.
- Select CALIBRATION > SPAN CALIBRATION.
- Select an option.
Option | Description |
---|
TIC SPAN ADJUST TOC SPAN
ADJUST| (Optional) Sets the TIC and TOC span adjust values for span
calibrations manually for each range.
RUN SPAN
CALIBRATION| Starts a span calibration, which sets the span adjust values
automatically. Span calibration reactions have the prefix of “SC”. Make sure
that measurements are stopped before a span calibration is started.
Make sure to install the calibration standard before a span calibration is
started. Refer to Plumb the calibration standard on page 74.
Note: The analyzer uses the same span adjust value calculated for the selected
RANGE for the other ranges, unless the span adjust values are manually
changed.
A span calibration reaction is the same as a normal reaction, but the prepared
calibration standard is measured and the sample pump does not operate in
reverse.
RUN SPAN
CHECK| Starts a span check. A span check is the same as a span calibration,
but the analyzer does not change the span adjust values. Span check
reactions have the prefix “SK”. Stop measurements before a span check is
started.
Make sure to install the calibration standard before a span check is started.
Refer to Plumb the calibration standard on page 74.
At the end of a span check, the analyzer identifies the span response at each
range and shows the suggested span adjust values in brackets “[ ]” near the
span adjust values set by the analyzer.
Note: Manually change the span adjust value settin gs on the RUN SPAN CHECK
screen if necessary.
SPAN PROGRAM| Note: Do not change the default setting unless necessary.
Changes can have a negative effect on the span adjust values.
Sets the number of span reactions done during a span calibration and a span
check (default: 5).
SPAN AVERAGE| Note: Do not change the default setting unless necessary.
Changes can have a negative effect on the span adjust values.
Sets the number of reactions the analyzer uses to calculate the average value
used for the span adjust values (default: 3).
RANGE| Sets the operation range for span calibration reactions and span check
reactions (default: 1). Select the operation range that agrees with normal
measurements for the sample stream(s).
Refer to the System Range Data screen to see the operation ranges. Select
OPERATION > SYSTEM RANGE DATA.
Note: If the RANGE setting is not applicable to the TIC CAL STD and TOC CAL
STD setting, the analyzer shows “CAUTION! REACTION
RANGE OR STANDARD IS INCORRECT”.
TIC CAL STD
TOC CAL STD| Sets the concentrations of the TIC and TOC calibration standards
for span calibrations.
Enter concentrations that are more than 50% of the full scale value for the
operation range selected in the RANGE setting. For example, if the operation
range for TIC or TOC is 0 to 5000 µgC/L, 50% of the full scale value is 2500
µgC/L.
If a calibration standard selected is 0.0 µgC/L, the analyzer does not change
the span adjust value for that parameter.
Note: In VOC systems, do not do the TIC and TOC calibrations together.
Do the TIC and TOC calibrations with different calibration standards.
TC CAL STD| Note: The TC CAL STD menu only shows in VOC systems.
Shows the TC CAL STD value, which is the sum of the TIC CAL STD and TOC CAL
STD.
If the TOC CAL STD or TIC CAL STD setting is 0.0 µgC/L, the TC CAL STD is set
to 0.0 µgC/L so that the analyzer does not change the span adjust value for
TC. In addition, the warning set with the TC BAND setting does not occur.
TIC CHECK STD
TOC CHECK STD| Sets the concentrations of the TIC and TOC calibration
standards for span checks (defaults: TIC = 0 µgC/L and TOC = 2500 µgC/L).
If the calibration standard selected is 0.0 µgC/L, the analyzer ignores the
span check results. In addition, the warning set with the TIC BAND or TOC BAND
settings do not occur.
TC CHEK STD| Note: The TC CHEK STD menu only shows in VOC systems.
Shows the TC CHEK STD value, which is the sum of the TIC CHECK STD and TOC
CHECK STD.
If the TOC CHECK STD or TIC CHECK STD setting is 0.0 µgC/L, the TC CHEK STD
is set to 0.0 µgC/L so that the analyzer ignores the span check results for
TC. In addition, the warning set with the TC BAND setting does not occur.
8.3 Plumb the calibration standard
Plumb the calibration standard container to the MANUAL fitting.
- Prepare the calibration standard. Refer to Prepare the calibration standard on page 74.
- Connect 1/4-in. OD x 1/8-in. ID PFA tubing to the MANUAL fitting. Make sure that the tubing length is 2 to 2.5 (6.5 to 8.2 ft).
- Put the tubing in the calibration standard container. Make sure that the calibration standard container is 100 to 500 mm (4 to 20 inches) below the analyzer.
8.4 Prepare the calibration standard
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance
with local, regional and national regulations.
Items to collect:
- Ultra pure water, 5 L
- Volumetric flask, 1 L (5x)
- Personal protective equipment (refer to MSDS/SDS)
Before starting:
- Put all hygroscopic chemicals in crystal form in a 105 °C oven for 3 hours to remove all water.
- Mix the prepared solutions with a magnetic stirrer or invert the solutions until all of the crystals are fully dissolved.
- If the purity of the chemical to be used is different than the purity given for the chemical in the steps that follow, adjust the quantity of chemical used. Refer to Table 20 for an example.
Shelf life and storage of calibration standards:
- TOC standards prepared from potassium hydrogen phthalate (KHP) are normally stable for 1 month when kept in a closed glass container at 4 °C.
- All the other standards (e.g., TOC prepared from acetic acid and TIC standards) should be used within 48 hours.
Prepare the calibration standard for TIC/TOC span calibrations and span checks as follows.
Note: The concentration of the calibration standard and the operation range for span calibrations and span checks are set on the SPAN CALIBRATION screen. Refer to Start a span calibration or span check on page 72.
Procedure:
-
Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
-
For the TOC standard, use an off-the-shelf TOC standard. Refer to Replacement parts and accessories in the Maintenance Manual for ordering information.
-
Prepare a 1000-mgC/L TIC standard solution as follows:
a. Add one of the chemicals that follow to a clean 1-L volumetric flask.
• Sodium carbonate (Na2CO3)—8.84 g (99.9% purity)
• Sodium hydrogen carbonate (NaHCO3)—7.04 g (99.5% purity)
• Potassium carbonate (K2CO3)—11.62 g (99.0% purity)
b. Fill the flask to the 1-L mark with ultra pure water. -
To prepare a TOC only standard with a concentration less than 1000-mgC/L, dilute the prepare standards with ultra pure water.
For example, to prepare a 50 mg/L standard solution, put 50 g of the 1000-mg/L prepared standard in a clean 1-L volumetric flask. Fill the flask to the 1-L mark with ultra pure water. -
To prepare a standard with a concentration less than 5-mg/L, prepare the standard with two or more dilution steps.
For example, to prepare a 1-mgC/L (ppm) standard, first prepare a 100-mgC/L standard. Then use the 100-mgC/L standard to prepare the 1-mgC/L standard. Put 10 g of the 100-mgC/L standard in a clean 1-L volumetric flask. Fill the flask to the 1-L mark with ultra pure water. -
To prepare a standard with a concentration at µg/l (ppb) levels, use multiple dilution steps.
Table 20 Quantity of KHP at different purities to prepare a 1000-mgC/L standard
Purity of KHP | Quantity of KHP |
---|---|
100% | 2.127 g |
99.9% | 2.129 g |
99.5% | 2.138 g |
99.0% | 2.149 g |
Table 21 Quantity of KHP to prepare different concentrations of TOC standard
Concentration of TOC standard | Quantity of 99.9% KHP |
---|---|
1000 mgC/L | 2.129 g |
1250 mgC/L | 2.661 g |
1500 mgC/L | 3.194 g |
2000 mgC/L | 4.258 g |
5000 mgC/L | 10.645 g |
10000 mgC/L | 21.290 g |
User interface and navigation
9.1 Keypad description
1 Back key—Push to go back to the previous screen or cancel changes. Push
for 1 second to go to the main menu.| 3 Display
---|---
2 Arrow keys— Push to select menu options or to enter
numbers and letters.| 4 Enter key—Push to confirm and go to the next screen.
9.2 Reaction Data screen
The Reaction Data screen is the default (home) screen. The Reaction Data
screen shows the current reaction information and the results of the last 25
reactions. Refer to Figure 22.
Note: If no key is pushed for 15 minutes, the display goes back to the
Reaction Data screen.
Push to see the Reagent Status screen and then the main menu.
Note: To see more than the last 25 reactions, push the enter key to go to
the main menu, then select OPERATION > REACTION ARCHIVE. Enter the reaction
date for the first reaction to show on the display.
Figure 22 Reaction Data screen
1 Status message (refer to Status messages on page 78)| 5 Operation
range (1, 2 or 3)
---|---
2 Reaction start time and date| 6 Reaction time since start
(seconds)
3 Reaction type| 7 Total reaction time (seconds)
4 Reaction phase| 8 Results of the last 25 reactions: start time,
date, record type13 and results. Refer to Table 22 for the record types.
Table 22 Record types
Symbol | Description | Symbol | Description |
---|---|---|---|
S1 … S2 | Sample Stream 1 to 2 | ZK | Zero check |
M1 | Manual Stream 1 | ZM | Zero adjust value set manually |
√ | There is sample or the quantity of air bubbles in the sample stream and | ||
manual stream is small. | SC | Span calibration | |
x | There is no sample or the quantity of air bubbles in the sample stream and | ||
manual stream is large. | SK | Span check | |
RS | Remote standby reaction | SM | Span adjust value set manually |
ZC | Zero calibration | A1 … A2 | 24-hour average result, Sample Stream 1 to 2 |
9.3 Status messages
A status message shows in the top-left corner of the Reaction Data screen and
the Reagent Status screen. The sequence of the status messages in Table 23
show the priority from highest to lowest.
13 TIC, TOC, TC and VOC. In addition, the calculated results (COD and BOD)
show on the display when the DISPLAY setting on the COD PROGRAM and/or BOD
PROGRAM menu is set to YES (default: OFF).
Table 23 Status messages
Message | Description |
---|---|
SYSTEM MAINTENANCE | The instrument is in maintenance mode. The maintenance |
switch is set to on.
SYSTEM FAULT| The instrument needs immediate attention. Measurements have
stopped. The 4–20 mA outputs are set to the FAULT LEVEL setting (default: 1
mA). The fault relay is on.
To identify the system fault, push to go to the main menu, then select
OPERATION > FAULT ARCHIVE. Faults and warnings preceeded by an “ ” are active.
To start the analyzer again, complete the troubleshooting steps in the
Maintenance and Troubleshooting Manual.
Note: “FAULT LOGGED” shows intermittently on the top-right corner of the
screen where the date and time show .
SYSTEM WARNING| The instrument needs attention to prevent a failure in the
future. Measurements continue. The fault relay is on.
To identify the warning, push to go to the main menu, then select
OPERATION > FAULT ARCHIVE. Faults and warnings preceeded by an “ ” are active.
Complete the troubleshooting steps in the Maintenance and Troubleshooting
Manual.
Note: “FAULT LOGGED” shows intermittently on the top-right corner of the
screen where the date and time show .
SYSTEM NOTE| There is a notification. The notification shows on the display
(e.g., 86POWER UP.
Note: “FAULT LOGGED” shows intermittently on the top-right corner of the
screen where the date and time show .
SYSTEM CALIBRATION| The instrument is in calibration mode (span calibration,
span check, zero calibration or zero check).
SYSTEM RUNNING| Normal operation
SYSTEM STOPPED| The instrument was stopped with the keypad or a fault
occurred.
REMOTE STANDBY| The instrument was put in remote standby with the optional
digital input for remote standby. The analog outputs and relays do not change.
Refer to REMOTE STANDBY in Start or stop measurements on page 81.
Note: A grab sample measurement can be done when the instrument is in
remote standby ._
9.4 Reaction Graph screen
Push to go to the Reaction Graph screen. The Reaction Graph screen shows the
reaction in progress. Refer to Figure 23.
Note: To go to back to the Reaction Data screen, push the enter key.
Figure 23 Reaction Graph screen
1 | Atmospheric pressure | 6 | CO2 instantaneous (i) measured value |
---|---|---|---|
2 | TIC µgC/L un-calibrated (µgu), no compensation for atmospheric | ||
pressure | 7 | CO2 zero (z) value at start of reaction | |
3 | CO2 peak value | 8 | Reaction time since start (seconds) |
4 | Oxygen flow (L/hour) | 9 | Total reaction time |
5 | Temperature of the analyzer (°C) |
Operation
10.1 Start or stop measurements
- Push to go to the main menu, then select OPERATION > START,STOP.
- Select an option.
Option | Description |
---|
REMOTE
STANDBY| An optional digital input is used to put the analyzer in remote
standby (e.g., from a flow switch). When the analyzer is in remote standby:
• “REMOTE STANDBY” shows in the top-left corner of the Reaction Data screen
and the Reagent Status screen.
• Measurements stop and the analog outputs and relays do not change.
• The analyzer does one remote standby (RS) reaction at 24-hour intervals at
the time set in the PRESSURE/FLOW TEST menu (default: 08:15 AM) in the SYSTEM
CONFIGURATION > SEQUENCE PROGRAM menu.
• Sample is not used during the remote standby reaction, only acid reagent and
base reagent is used.
• A grab sample measurement can be done.
When REMOTE STANDBY is unselected, the analyzer starts measurements unless the
analyzer was stopped with the keypad or a fault occurred.
START| Starts the analyzer. The analyzer does an ozone purge, pressure test,
flow test, reactor purge and analyzer purge, then starts the analysis of the
first stream in the programmed stream sequence. If a fault has occurred, the
analyzer cannot be started until the fault is removed.
Note: To start the analyzer without the pressure test or flow test (quick
startup), select START and push the RIGHT arrow key at the same time.
When a quick startup is done, a 28_NO PRESSURE TEST warning occurs.
The warning stays active until a pressure test is passed.
• Ozone purge—Pushes residual ozone through the ozone destructor.
• Pressure test—Identifies if there is a gas leak in the analyzer.
• Flow test—Identifies if there is a blockage in the gas exhaust or the sample
out lines.
• Reactor purge—Removes liquid from the reactor through the SAMPLE OUT
fitting.
• Analyzer purge—Removes CO2 gas from the CO2 analyzer through the EXHAUST
fitting.
Note: If the analyzer is started while the remote standby signal is active,
the analyzer goes to remote standby.
FINISH &
STOP| Stop the analyzer after the last reaction is completed. The analyzer
does an ozone purge, reactor purge and analyzer purge and then stops.
EMERGENCY
STOP| Stops the analyzer before the last reaction is completed. The analyzer
does an ozone purge, reactor purge and analyzer purge and then stops.
Note: If EMERGENCY STOP is selected soon after FINISH & STOP is selected, an
EMERGENCY STOP is done.
10.2 Measure a grab sample
The grab sample settings can be changed while the analyzer is in operation, unless:
- A manual mode (grab sample) sequence is scheduled to start when the last reaction is completed.
- A manual mode sequence has started.
Plumb and configure the analyzer to do a grab sample measurement as follows:
-
Use 1/4-in. OD x 1/8-in. ID PFA tubing to plumb the grab sample container to the
MANUAL fitting. Make sure that the tubing length is 2 to 2.5 (6.5 to 8.2 ft).
Refer to Table 2 on page 4 for the sample specifications. -
Put the tubing in the grab sample. Make sure that the grab sample is 100 to 500 mm (4 to 20 inches) below the analyzer.
-
Do a sample pump test for the manual stream(s) to identify the correct forward and reverse times. Refer to Do a sample pump test on page 59.
-
Set the sample pump times for the manual stream(s). Refer to Set the sample pump times on page 59.
-
Select OPERATION > MANUAL PROGRAM.
-
Select an option.
Option | Description |
---|
RUN AFTER
NEXT
REACTION| Starts the manual mode (grab sample) sequence after the next
reaction. If the analyzer is stopped, the manual mode sequence will start
immediately.
Note: If the analyzer has the Manual-AT Line option, push the green button to
select RUN AFTER NEXT REACTION. The Manual-AT Line option is a
small box with only a green button. The Manual-AT Line cable is connected to
the analyzer.
Note: When a manual mode sequence starts, all of the cleaning cycle,
pressure/flow tests, zero or span cycles stop temporarily. In addition, the
reverse operation of the sample pump is disabled (default).
RUN AFTER| Starts the manual mode (grab sample) sequence at a selected time
(default: 00.00).
RETURN TO
ON-LINE
SAMPLING| Sets the analyzer to stop or go back to online operation when the
manual mode sequence is completed. YES—The analyzer goes back to online
operation. NO (default)—The analyzer stops.
RESET
MANUAL
PROGRAM| Sets the MANUAL PROGRAM settings back to the factory default
settings.
MANUAL x, x
RANGE x| Sets the number of reactions and the operation range for each manual
(grab sample) stream.
MANUAL—The first setting is the manual valve number (e.g., MANUAL VALVE 1 is
connected to the MANUAL 1 fitting on the side of the analyzer).
The second setting is the number of reactions done at the manual stream before
the analyzer does reactions at the next manual stream.
RANGE —Sets the operation range for each manual stream. Options: 1, 2 or 3
(default). Refer to the SYSTEM RANGE DATA screen to see the operation ranges.
Select OPERATION > SYSTEM RANGE DATA. If the concentration of the grab sample
is not known, select AUTO.
Note: If RANGE is set to AUTO, enter 5 for the number of reactions so
theanalyzer can find the best operation range. It may be necessary to discard
the first two or three analysis results.
Note: When a MANUAL is set to “- , -” and RANGE is set to “-“, the manual
stream is not measured.
10.3 Save data to an MMC/SD card
Save the reaction archive, fault archive, configuration settings and/or
diagnostic data to an MMC/SD card.
- Put the supplied MMC/SD card in the MMC/SD card slot. Refer to Figure 24.
- Select MAINTENANCE > DIAGNOSTICS > DATA OUTPUT.
- Select an option.
Option | Description |
---|---|
OUTPUT DEVICE | Sets where the analyzer sends the data (MMC/SD CARD). This |
setting cannot be changed.
To configure the settings for the MMC/SD card, select MAINTENANCE >
COMMISSIONING > DATA PROGRAM. Refer to Configure the communication settings on
page 67.
Make sure that the MMC/SD card is configured with FAT, FAT12/16 or FAT32 file
systems. As an alternative, use an SDHC card. Data is
saved to an MMC/SD card in text format. The binary files on the card are
system firmware (sysfrmw.hex) and system configuration
(syscnfg.bin).
SEND REACTION
ARCHIVE| Sends the contents of the reaction archive to the output device. Set
the start date and number of entries to send, then select START
SENDING. OUTPUT ITEMS shows the number of entries sent. The analyzer sends the
data in the display language.
If PAUSE SENDING is selected, entries are not sent for 60 seconds or until
PAUSE SENDING is selected again.
If the output device is an MMC/SD card, the reaction archive is saved to the
file RARCH.txt.
Note: To see the reaction archive, go to the main menu, then select OPERATION
REACTION ARCHIVE.
Refer to Table 24 and Table 25 for descriptions of the data sent. To select standard or engineering data, select DATA PROGRAM >
PRINT MODE.
SEND FAULT
ARCHIVE| Sends the contents of the fault archive to the output device. Select START SENDING. OUTPUT ITEMS shows the number of entries
sent. The data is sent in the display language.
If PAUSE SENDING is selected, entries are not sent for 60 seconds or until PAUSE SENDING is selected again.
If the output device is an MMC/SD card, the fault archive is saved to the file FARCH.txt.
Note: To see the fault archive, go to the main menu, then select OPERATION > FAULT ARCHIVE. The fault archive contains the last
99 faults and warnings.
SEND
CONFIGURATION| Sends the analyzer settings to the output device. Select START SENDING. OUTPUT ITEMS shows the number of entries sent. The
data is sent in the display language.
If PAUSE SENDING is selected, entries are not sent for 60 seconds or until PAUSE SENDING is selected again.
If the output device is an MMC/SD card, the analyzer settings are saved to the file CNFG.txt.
SEND ALL DATA| Sends the reaction archive, fault archive, analyzer settings and diagnostics data to the output device. Select START SENDING. The
data is sent in English.
If PAUSE SENDING is selected, entries are not sent for 60 seconds or until PAUSE SENDING is selected again.
If the output device is an MMC/SD card, the analyzer settings are saved to the file ALLDAT.txt.
DATA PROGRAM| Goes to the MAINTENANCE > COMMISSIONING > DATA PROGRAM menu to set the communication settings for the output
devices: MMC/SD card and Modbus.
Figure 24 Insert MMC/SD card
Table 24 Reaction archive data—Standard mode
Item | Description |
---|---|
TIME | Time the reaction started |
DATE | Date the reaction started |
S1:2 | Reaction type (e.g., Stream 1) and operation range (e.g., 2) |
TCµgC/L | Calibrated TC value in µgC/L (TC is TIC + NPOC + POC) |
TICµgC/L | Calibrated TIC value in µgC/L |
TOCµgC/L | TIC + TOC analysis —Calibrated TOC value in µgC/L (TOC is NPOC) |
VOC analysis —Calculated TOC value in µgC/L (TOC is calculated as TC –
TIC)
COD/BODµgO/L| Calculated COD and/or BOD value in µgO/L (if set to on in the
COD PROGRAM and/or BOD PROGRAM menu)
VOCµgC/L| Calculated VOC value in µgC/L (VOC is calculated as TC – TIC – NPOC)
Table 25 Reaction archive data—Engineering mode (TIC + TOC analysis)
Item | Description |
---|---|
TIME | Time the reaction started |
DATE | Date the reaction started |
S1:2 | Reaction type (e.g., Stream 1) and operation range (e.g., 2) |
CO2z | Zero adjust value for the CO2 analyzer for the last reaction |
CO2p | Maximum height of the CO2 peak |
µgu | Un-calibrated value in µgC/L |
µgc | Calibrated value in µgC/L |
COD/BODµgO/L | Calculated COD and/or BOD value in µgO/L (if set to on in the |
COD PROGRAM and/or BOD PROGRAM menu)
BT_DegC| Analyzer temperature (°C)
MB_DegC| Motherboard temperature (°C)
Atm| Atmospheric pressure (kPa)
SAMPLE| Sample quality (%) from the Sample Sensor signal used to activate the
SAMPLE STATUS output
SMPL PUMP| The five items, which are number coded or number data, give
information on the sample pump as follows:
- Operation mode (0 = time mode or 1 = pulse mode)
- Number of pulses during operation (e.g., injection)
- Total time (milliseconds) for the total number of pulses
- The time (milliseconds) for the last pulse
- Error counter (0 to 6). When a pulse is not done or identified, the pump
goes to time mode for that specific operation (e.g., injection or
synchronization). A pump warning only occurs if there are six consecutive
failures.
ACID PUMP| Error counter for the acid pump. Refer to SMPL PUMP description.
BASE PUMP| Error counter for the base pump. Refer to SMPL PUMP description.
COOLER| The status of the cooler (e.g., OFF).
O3 HEATER| The status of the ozone destructor heater (e.g., OFF).
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. 970-669-3050
800-227-4224 (U.S.A. only)
Fax 970-669-2932
orders@hach.com
www.hach.com| HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
info-de@hach.com
www.de.hach.com| HACH LANGE Sàrl
6, route de Compois
1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499
---|---|---
©Hach Company/Hach Lange GmbH, 2022. All rights reserved. Printed in Ireland.
References
- Geräte und Reagenzien für die Wasserqualitätsanalyse | Hach
- DTSC.CA.GOV/HAZARDOUSWASTE/PERCHLORATE
- Hach | Hach
- MyHach - Customer Service
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>