PCE-950 Handheld Material Hardness Tester Metals User Manual
- June 15, 2024
- PCE
Table of Contents
- PCE-950 Handheld Material Hardness Tester Metals
- Product Information
- Product Usage Instructions
- FAQ
- Safety notes
- General description
- Main applications
- Technical specifications
- Delivery contents
- Optional adaptors
- Optional external impact devices
- Device description
- Operation
- Menu structure
- Settings
- Calibration
- Battery recharging
- Automatic Shut Off
- Connection to the computer
- Conversion chart
- Maintenance
- Contact
- Disposal
- PCE Instruments contact information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PCE-950 Handheld Material Hardness Tester Metals
Product Information
Specifications
Hardness measuring range by material:
- Cast steel: HRC 19.8 – 68.5
- Hammered steel: HRB 59.6 – 99.6
- Cold-rolled steel: HRA 59.1 – 85.8
- Stainless steel: HB 80 – 651
- Cast iron: HV 83 – 976
- Ductile iron: HS 32.2 – 115
- Aluminum alloy: HB 143 – 650
- Brass: HRC 20.4 – 67.1
- Bronze: HV 80 – 898
- Forged copper alloy: HRB 45.5 – 101.7
- General specifications:
- Accuracy: Varies by material
- Repeatability: Varies by material
- Total measuring range: Varies by material
- Hardness scales: HRC, HRB, HRA, HB, HV, HS
- Impact device: Varies by model
- Measuring direction: Varies by model
- Display: OLED display
- Data storage: Yes
- Power supply: Battery or power supply
- Battery life: Varies by usage
- Data interface: USB
- Dimensions: Varies by model
- Weight: Varies by model
Product Usage Instructions
Safety Precautions
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
General Description
The PCE-950 Hardness Tester is a device used for measuring the hardness of various materials. It is designed for qualified personnel and comes with a range of features to ensure accurate measurements.
Main Applications
The PCE-950 Hardness Tester is commonly used in industrial settings for quality control and material testing. It can be used to measure the hardness of materials such as cast steel, hammered steel, cold-rolled steel, stainless steel, cast iron, ductile iron, aluminum alloy, brass, bronze, and forged copper alloy.
Technical Specifications
Refer to the specifications section above for detailed technical specifications of the PCE-950 Hardness Tester.
Delivery Contents
The package includes:
- 1 x durometer PCE-950
- 1 x brush
- 1 x miniature adapter
- 1 x USB cable
- 1 x power supply
- 1 x test block
- 1 x user manual
- 1 x carrying case
The software can be downloaded here.
Optional Adaptors
The PCE-950 Hardness Tester supports optional adaptors for specific measurement tasks. The following adaptors are available:
Order Code | Description |
---|---|
HK16.530 | Adaptor to measure spherical surfaces as for example in mould |
making. Enables centering and prevents slipping and resulting
measurement errors. Radius: 16.5 – 30 mm
HK11-13| Adaptor to measure spherical surfaces as for example in mould
making. Enables centering and prevents slipping and resulting
measurement errors. Radius: 11 – 13 mm
Z25-50| Adaptor to measure outer radii, e.g., on round steel and pipes.
Enables centering and prevents slipping and resulting measurement
errors. Radius: 25 – 50 mm
Z10-15| Adaptor to measure outer radii, e.g., on round steel and pipes.
Enables centering and prevents slipping and resulting measurement
errors. Radius: 10 – 15 mm
HZ12.517| Adaptor to measure inner radii, e.g., in pipes and tanks.
Enables centering and prevents slipping and resulting measurement
errors. Radius: 12.5 – 17 mm
Optional External Impact Devices
The PCE-950 Hardness Tester supports optional external impact devices for specific measurement requirements. The following external impact devices are available:
Order Code | Type |
---|---|
SG-D | D, 1800 HV |
SG-G | G, 1800 HV |
SG-DL | DL, 1800 HV |
SG-C | C, 1800 HV |
SG-D+15 | D + 15, 1800 HV |
SG-DC | DC, 1800 HV |
Device Description
Device
- On/Off key: Used to turn the device on or off.
- Release button: Used to release the impact device.
- Charging LED: Indicates the charging status of the device.
- OLED display: Displays the measurement results.
Main Screen
- MiniUSB connection: Used to connect the device to a computer or power supply.
- Loading sheath: Used to load the impact device.
- Connection external impact device: Used to connect optional external impact devices.
- Operating keys: Used to navigate through the device’s menu and settings.
FAQ
Q: Can the PCE-950 Hardness Tester measure the hardness of other materials not listed in the manual?
A: The PCE-950 Hardness Tester is designed to measure the hardness of a wide range of materials. However, for accurate results, it is recommended to refer to the manual for the specified measuring ranges and hardness scales for each material.
Q: How long does the battery last on the PCE-950 Hardness Tester?
A: The battery life of the PCE-950 Hardness Tester varies depending on usage. It is recommended to fully charge the device before each use and recharge it as needed.
Q: Can I connect the PCE-950 Hardness Tester to a computer?
A: Yes, the PCE-950 Hardness Tester has a USB data interface that allows you to connect it to a computer for data transfer and analysis using compatible software.
PCE-950 Hardness Tester
User manuals in various languages (français, italiano, español, português,
nederlands, türk, polski, русский, 中文) can be found by using our
product search on: www.pce-instruments.com
Last change: 25 August 2022 v1.0
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this
manual.
We expressly point to our general guarantee terms which can be found in our
general terms of business.
General description
- Customised curve for a specific testing material
- OLED display of 128×32 matrix
- Converts to all common hardness scales (HV ,HB,HRC,HRB,HRA,HS)
- Easy to use menu system With miniUSB interface, multiple communication modes are adopted to meet customized requirements of various users
- Max. 600 groups of data can be stored in memory
- Upper and lower limit of hardness can be preset; with alarm
- Test values software calibration function
- Working voltage:3.7 V rechargeable Li-Ion battery
- Continuous working period: approx. 50 h
- Can be used with impact device Type C, DC, G, D+15, DL, E
- Smart and skid-proof design
- Charging indicator
Main applications
- Direct measurement of heavy workpieces and permanently installed parts
- Measurements during production process, especially in mass production
- Material identification of metal materials
- Failure analysis of pressure vessels
- Bearings and other parts
- Testing of parts in confined or hard-to-access areas
Technical specifications
Hardness measuring range by material|
---|---
Cast steel| HRC: 19.8 … 68.5
HRB: 59.6 … 99.6
HRA: 59.1 … 85.8
HB: 80 … 651
HV: 83 … 976
HS: 32.2 … 115
Hammered steel| HB: 143 … 650
Cold-rolled steel| HRC: 20.4 … 67.1
HV: 80 … 898
Stainless steel| HRB: 45.5 … 101.7
HB: 85 … 655
HV: 85 … 802
Cast iron| HB: 93 … 334
Ductile iron| HB: 131 … 387
Aluminum alloy| HRB: 23.8 … 84.6
HB: 19 … 164
Brass| HRB: 13.5 … 95.3
HB: 40 … 173
Bronze| HB: 60 … 290
Forged copper alloy| HB: 45 … 315
General specifications|
Accuracy| ± 6 HL at 730 … 790 HL
± 10 HL at 490 … 570 HL
Repeatability| 6 HL at 730 … 790 HL
10 HL at 490 … 570 HL
Total measuring range| 170 … 960 HL
Hardness scales| HL (HLD): Leeb (Type D impact device) HB: Brinell
HRC: Rockwell C HRB: Rockwell B HRA: Rockwell A HV: Vickers
HS: Shore
Impact device| Type D
Measuring direction| 360°
Display| 128 x 32 OLED
Data storage| Internal memory saves up to 600 readings
Power supply| Built-in rechargeable Li-ion battery
Battery life| Approx. 50 hours
Data interface| Mini USB
Dimensions| Approx. 153 x 54 x 24 mm / 6.02 x 2.13 x 0.95 in
Weight| Approx. 250 g / < 1 lb
Delivery contents
1 x durometer PCE-950, 1 x brush, 1 x miniature adapter, 1 x USB cable, 1 x power supply, 1 x test block, 1 x user manual, 1 x carrying case The software can be downloaded here: https://www.pce-instruments.com/english/download- win_4.htm
Optional adaptors
Optional external impact devices
The external impact devices just have to be connected to the device. The device will recognize them automatically (for connection, see device description).
No. | Order code | Type |
---|---|---|
1 | SG-D | D, 1800 HV |
2 | SG-G | G, 1800 HV |
3 | SG-DL | DL, 1800 HV |
4 | SG-C | C, 1800 HV |
5 | SG-D+15 | D + 15, 1800 HV |
6 | SG-DC | DC, 1800 HV |
Device description
Device
- On/Off key
- MiniUSB connection
- Release button
- Loading sheath
- Charging LED
- Connection external impact device
- OLED display
- Operating keys:
Arrow keys – move cursor
MENU/OK – Main menu/confirm
Main screen
- Alarm upper limit
- Reading
- No. of measurements
- Battery level indication
- Alarm lower limit
- Hardness scale
- Impact direction
-
Upper / lower limit alarm
When measured value is over/below the limit, the icon will be full. -
Measured value
Current measured value (without “AVE”) or current average value (with “AVE”) will be displayed. The average value can only be displayed if the desired no. of impacts has been set in advance. If ↑ is displayed, this means that the measured value exceeds the measurement range / the convertible range. If ↓ is displayed, this means that the measured value is below the measurement range / the convertible range. -
No. of measurements
The number of measurements is displayed here. The no. of measurements after which the average value is shown can be set previously in the menu. -
Battery level indication
The battery icon shows the current battery level. -
Hardness scale
The hardness scale currently used is shown here. -
Impact direction
The current impact direction is displayed here.
No.
|
Type ofimpactdevice
| hardness value of standard Leeb hardness
block
|
Error ofdisplayedvalue
|
Repeatabilityofdisplayedvalue
---|---|---|---|---
1
|
D
| 760±30HLD
530±40HLD
| ±6 HLD
±10 HLD
| 6 HLD
10 HLD
Operation
Preparation prior to testing
- If the surface to be tested is too rough, measuring errors can occur. The surface of the sample must be smooth, flat, and free from oil or grease.
- Curved surfaces: Make sure that the testing surface of workpiece is plane. When the radius of curvature R of the curved surface to be tested is less than 30 mm,a small support ring or should be used.
- When the sample is magnetic, this can falsify the result.
- When the sample weighs over 5 kg, no support is needed.
- Samples made of profile material or with thin wall thicknesses and parts weighing between 2 and 5 kg need a support of more than 5 kg which must be tightly and evenly coupled with the sample to avoid deformations caused by the impact of the impact device or by vibrations.
- Samples with a weight below 2 kg also have to be coupled tightly and evenly to a support of more than 5 kg. The surface between the sample and the support should be even and smooth. Apply an adequate but not an excessive amount of coupling agent to these surfaces.
- The minimum weight of a sample should always be 100 g.
- The minimum thickness of the sample is 3 mm.
- The hardness penetration depth of hardened parts must be at least 0.8 mm.
- To avoid errors caused by rough surfaces, the surface should be polished and have a roughness Ra <2 μm or Rz <l0 μm
Testing the accuracy
A standard Leeb hardness block should be used to verify the accuracy of the
portable hardness tester and the mesasurement error and repeatability should
not exceed the values in the following chart.
No.| Type of impact
device
| Hardness value of standard Leeb
hardnesss test block
| Accuracy| Repeatability
---|---|---|---|---
1
|
D
| 760±30HLD
530±40HLD
| ±6 HLD
±10 HLD
| 6 HLD
10 HLD
Note: It is recommended that you take a minimum of 5 tests and obtain your average. That value should be within the tolerance shown in the chart above. If the value exceeds this tolerance, you must send the device to PCE Instruments for calibration.
Measurement
Press the On/Off key to turn on the device. The following will be shown on the
screen:
PCE950
Hardness Tester
The device will automatically enter measurement mode.
Push down the loading sheath to lock the impact body. Now place the device
with the support ring on the surface of test sample. The direction of impact
should be vertical with the testing surface. Make sure that the meter is
placed firmly on the test surface and push down the release button on the top
of the device to make a measurement. Five measurements should be carried out
per measuring point of the test sample. The deviation of the mean value should
not exceed the values in the chart.
The distance between any two indentations, or the distance between any indentation center and the edge of the test sample should be in accordance with the specifications in the following chart:
Type of impact device| The distance of two indentation centers|
The distance between indentation center and edge of test piece
---|---|---
No less than| No less than
D| 3| 5
How the results are displayed
- The hardness value will be displayed ahead of the unit HL (Leeb hardness), followed by the type of impact device. For example, 700 HLD expresses that the Leeb hardness is 700 and was measured by means a D-type impact device.
- For values converted into other hardness units, the corresponding hardness scale should be added ahead of Leeb hardness abbreviation. For example, 400HVHLD expresses that the Vickers hardness value is 400, which was converted from the Leeb hardness value, measured by a D-type impact device.
Information on hardness scales
Not all materials can be converted into every hardness scale. The hardness scale will return to Leeb hardness (HL) automatically after the material has been changed. Therefore, the material must be selected first when setting the measurement parameters, and the hardness scale should be set next.
Menu structure
When you are in measurement mode, press the MENU/OK key to enter the main menu. Here you can set all parameters.
Settings
Function overview
Press the MENU/OK key to enter the main menu.
You can move the cursor by using the arrow keys. With the MENU/OK key, you can select a function.
Function | Description |
---|---|
Save Average | Saves the measured average. |
Delete Single | **Deletes individual data measured to calculate the |
average.
Cal. Average| Calculates the average of individual measurement
results.
MemoryManager**|
View From No.1| All measured values (averages)
View From End| All measured values from new to old
View From No.| Here you can set what measurements you want to view.
Print| Function not yet available
Delete By No.| Here you can delete certain measurements
Delete All| All measured data are deleted from the memory
Exit|
Save Average
This function is used to save the current average value. This function is only
available if you have made as many measurements as previously set under Test
Set – Average or via the “Cal.Average” function that stops the measurement
immediately.
Delete Single
If you wish to delete the displayed measured value or average value, go to
“Delete Single”, use the arrow keys to select “Yes” or “No”. The latter
applies if you decide not to delete the value.
Cal. Average
This function is to calculate the average value and to terminate the test.
Memory Manager
Use the MENU/OK key to go to the main menu.
Navigation
Navigate through the data by using the arrow keys. To return to the last page, press the MENU/OK key.
Deletion
Use the arrow keys to select the file to be deleted and confirm with the
MENU/OK key.
Select „Yes“ by using the arrow keys and confirm with MENU/OK.
Test Set
Press the MENU/OK key to enter the main menu. Press the arrow keys to move the cursor and press MENU/OK to select “Test Set”. Move the cursor to the parameter you wish to set and press MENU/OK.
Impact Direction
Press the arrow keys to select the impact direction and confirm with MENU/OK.
Average
Here you can set the no. of impacts / measurements to calculate the average. You can select a no. between 1 and 10. The number can be changed using the arrow keys, At the end, confirm with the MENU/OK key.
Material
Press the arrow keys to select a material and confirm with MENU/OK.
Note: When a new material is measured, the value will automatically be
displayed in HL: Therefore, it is important to first select the material and
then the hardness scale.
Hardness Scale
Press the arrow keys to select a hardness scale. Press MENU/OK to confirm.
Note: Only provide hardness scales that are available for the selected
material. When a new material is measured, the value will automatically be
displayed in HL: Therefore, it is important to first select the material and
then the hardness scale.
Tolerance Limit
Change the lower and upper limit with the arrow keys and use the MENU/OK key
to get to the next digit. When the last digit has been set, you will enter
measurement mode.
Note: When the value set exceeds the measurement range, you will be asked to
set it again. When the Min value exceeds the Max value, Min and Max will be
swapped.
System Setting
-
Warn Sound
Here you can enable or disable the alarm tones. Use the MENU/OK key to enable or disable the function. -
Key Sound
Here you can enable or disable key sounds. Use the MENU/OK key to enable or disable the function. -
Auto Save
This function allows automatic saving of the averages. Use the MENU/OK key to enable or disable the function. -
Auto Delete
If you enable this function, the system will automatically delete those test results that are out of range. The deleted values need to be replaced by making the same number of measurements as values deleted. Use the MENU/OK key to enable or disable the function. -
Auto Transfer
After the average is calculated, the data can be transferred in text format to a computer automatically via the USB cable. Use the MENU/OK key to enable or disable the function. -
LCD Brightness
Use the arrow keys to set the brightness of the display. Confirm with MENU/OK:
Time Date Set
Press the arrow keys to select the menu item Time Date Set. Press MENU/OK to
confirm.
Press MENU/OK to move the cursor to the next digit and change it by using the
arrow keys. At the end, press MENU/OK to confirm.
About Software
Press MENU/OK to enter the main menu. Use the arrow keys to select the menu
item “About Software” and confirm with MENU/OK.
In this menu item, you will find some information about the device, including
the serial number and the software version. To navigate, use the arrow keys.
By pressing the MENU/OK key, you return to the measurement mode.
Exit
Via this menu item, you can leave the menu.
The device needs to be calibrated before first use or after a long period of
non-use.
Press and hold the On/Off key and the arrow keys at the same time to enter the
calibration screen. Set the impact direction【↓】and make 5 measurements of the
test block that comes with the meter.
The meter will then display the average value.
Use the arrow keys to enter the standard value which you can find on the test
block. Press MENU/OK to finish the calibration. The accuracy is ±6HL.
Calibration
Battery recharging
This icon (flashing) indicates low battery. Recharge the battery to avoid the device switching off automatically.
Automatic Shut Off
The auto shut off function is designed to save battery power.
- The battery will automatically shut off after 5 minutes of inactivity.
- The screen will flash for 20 seconds before automatic shut off. Press any button other than the On/Off key to terminate the shutting off.
- When the battery is too low, the system will be shut off automatically and the battery icon will flash.
Connection to the computer
- Switch on the device.
- Install the driver and the software on your computer. Both can be downloaded here: https://www.pce-instruments.com/english/download-win_4.htm.
- Connect the device to the computer, using the USB cable.
- Open the software and click “CONN” in the lower part of the software screen so that the software recognises the device.
- Now you can load your measurements and view them as graphics by clicking “READ” or “LOAD.”
Conversion chart
Material
| ****
HL
| ****
HRC
| ****
HRB
| HB| ****
HS
| ****
HV
---|---|---|---|---|---|---
30D2| 10D2
Cast steel| 300 ~
900
| 20.0 ~ 68.0| 38.4 ~ 99.5| 80 ~ 647| | 32.5 ~ 99.5| 80 ~ 940
CWT steel| 300 ~| 20.4 ~| | | | | 80 ~ 898
Stainless steel| 300 ~| 19.6 ~| 46.5 ~| 85 ~ 655| | | 80 ~ 802
Grey cast iron| 360 ~| | | 93 ~ 334| | |
Nodular cast| 400 ~| | | 131 ~| | |
Cast aluminium| 174 ~| | | | 20 ~| |
Copper zinc| 200 ~| | 13.5 ~ 95.3| | 40 ~| |
Copper alum.| 300 ~| | | | 60 ~| |
Wrought| 200 ~| | | | 45 ~| |
Maintenance
Troubleshooting
Failure | Cause | Solution |
---|---|---|
Failure in starting | Battery empty | Recharge battery |
No measuring value | Probe wire open circuit | Replace probe wire |
Value is inaccurate | Calibration data loss | Repeat calibration |
Other maintenance
- For every 1000-2000 impact, use nylon brush to clean the pipe/tube and the impact body of the impact device. Take the supporting ring and the impact body off, insert and twist nylon brush counterclockwise to the bottom and pull back. Repeat for 5 times and re-install the supporting ring and the impact body.
- Keep the impact body released after use.
- Using of any kind of lubricant is strictly prohibited in the impact device.
- When calibrating error is more than 2 HRC, it might be necessary to replace the test tip or the impact body. Contact PCE Instruments.
- The impact body, the supporting ring and the loading sheath are excluded from the warranty.
Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
Disposal
For the disposal of batteries in the EU, the 2006/66/ECdirective of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste. They must be given to collection points
designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back.
We either re-use them or give them to a recycling company which disposes of
the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in
accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Instruments contact information
United Kingdom
PCE Instruments UK Ltd
Unit 11 Southpoint Business Park Ensign Way, Southampton Hampshire
United Kingdom, SO31 4RF
Tel: +44 (0) 2380 98703 0
Fax: +44 (0) 2380 98703 9
info@pce-instruments.co.uk
www.pce-instruments.com/english
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
- Download | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>