MSA 5X Altair Gas Detector Owner’s Manual
- June 14, 2024
- MSA
Table of Contents
MSA 5X Altair Gas Detector
Product Information
Specifications
-
Order No.: 10116951/09
-
Print Spec: 10000005389 (EO)
-
CR: 800000056790
-
Website: www.MSAsafety.com
-
Address: 1000 Cranberry Woods Drive Cranberry Township, PA
16066 USA -
Phone: 1-800-MSA-2222
-
Fax: 1-800-967-0398
Description
Overview
The ALTAIR 5X and ALTAIR 5X IR Multigas Detectors, hereafter also referred
to as device, are designed for use by trained and qualified personnel. They
are used for performing a hazard assessment and interpreting device readings.
Device Hardware Interfaces
The device is equipped with various hardware interfaces for operation and
monitoring. These interfaces include buttons, a display screen, alarms, on-
screen indicators, and additional pages for viewing information.
Alarms
The device is equipped with alarms to alert users of potential hazards.
These alarms include visual and audible signals.
On-Screen Indicators
The device has on-screen indicators that provide information about the gas
levels, battery status, calibration status, and other important parameters.
Monitoring Toxic Gases
The ALTAIR 5X Multigas Detector can detect various toxic gases in gas
mixtures. It is designed for measurement of the oxygen concentration in gas
mixtures for inertisation according to EN 50104 without alarm function.
Monitoring Combustible Gases
The ALTAIR 5X IR Multigas Detector is equipped with an infrared sensor that
can detect specific combustible gases up to 100% Vol.
Product Usage Instructions
Correct Use
The ALTAIR 5X and ALTAIR 5X IR Multigas Detectors should only be used by
trained and qualified personnel. They are intended for performing hazard
assessments and interpreting device readings.
Safety and Precautionary Measures
It is important to follow safety regulations and take precautionary
measures when using the device. Failure to do so can result in serious
personal injury or death. Some of the safety guidelines include:
- Do not remove battery pack, recharge Li Ion battery, or replace alkaline batteries in a hazardous location.
- Do not alter or modify the device.
Device Setup
Before using the device, it is important to perform the initial setup. The
setup process includes:
- Ensure the device is turned off.
- Place the device in a well-ventilated area.
- Check the battery status and replace the battery if necessary.
- Inspect the device for any damage or defects.
- Connect any necessary accessories or sensors.
Measurement Mode [Normal Operation] The device operates in measurement mode for normal operation. In this mode, it continuously monitors the gas levels and displays the readings on the screen. The user can interpret the readings and take necessary actions based on the measured values.
Bluetooth Operation
The device supports Bluetooth connectivity for wireless data transfer and remote monitoring. To use the Bluetooth feature, follow these steps:
- Enable Bluetooth on the device.
- Pair the device with a compatible Bluetooth-enableddevice.
- Install any necessary software or applications for dat a transfer and monitoring.
Calibration
The device requires periodic calibration to ensure accurate measurements. To calibrate the device, follow these steps:
- Ensure the device is turned off.
- Connect the calibration gas cylinder to the device.
- Follow the on-screen instructions to initiate the calibration process.
- Verify the calibration results and adjust if necessary.
FAQ
Q: Can the device detect multiple gases simultaneously?
A: Yes, the device is designed to detect multiple gases simultaneously
based on the installed sensors.
Q: How often should I calibrate the device?
A: The device should be calibrated periodically according to the
manufacturer’s recommendations. The calibration frequency may depend on
factors such as usage, environmental conditions, and regulatory requirements.
Q: Can I use the device in hazardous locations?
A: No, it is not recommended to use the device in hazardous locations.
Follow safety regulations and guidelines to ensure proper usage and avoid
accidents.
Q: How do I replace the battery?
A: To replace the battery, follow these steps:
- Turn off the device.
- Remove the old battery pack.
- Insert the new battery pack into the device.
- Ensure the battery is securely connected.
Operating Manual
Altair® 5X Multigas Detector
Altair® 5XiR Multigas Detector
GB
Order No.: 10116951/09 Print Spec: 10000005389 (EO)
CR: 800000056790
MSAsafety.com
WARNING! Read this manual carefully before using or maintaining the device.
The device will perform as designed only if it is used and maintained in
accordance with the manufacturer’s instructions. Otherwise, it could fail to
perform as designed, and persons who rely on this device could sustain serious
injury or death. The warranties made by MSA with respect to the product are
voided if the product is not installed and used in accordance with the
instructions in this manual. Please protect yourself and your employees by
following the instructions. Please read and observe the WARNINGS and CAUTIONS
inside. For additional information relative to use or repair, call
1-800-MSA-2222 during regular working hours. For countries of Russian
Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector
will be delivered with a passport document that includes valid approval
information. On the CD with manual instruction attached to the gas detector
the user will find the documents “Type Description” and “Test Method” –
appendixes to Pattern Approval Certificate of Measuring instrument, valid in
the countries of use. The Declaration of Conformity can be found under the
following link: https://MSAsafety.com/DoC. MSA is a registered trademark of
MSA Technology, LLC in the US, Europe and other Countries. For all other
trademarks visit https://us.msasafety.com/Trademarks. This product
incorporates Bluetooth® wireless technology. The Bluetooth word mark and logos
are registered trademarks owned by Bluetooth SIG, Inc., and any use of such
marks by MSA is under license. Other trademarks and trade names are those of
their respective owners. Versions of this product manufactured July 2022 and
later may not contain Bluetooth wireless technology. This will be indicated by
the front case of the detector not utilizing a Bluetooth logo. All references
in this manual to Bluetooth will not pertain to these versions of the device.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. You are cautioned
that changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
English: This device complies with RSS-210 of the Industry Canada Rules.
Operation is subject to the following two conditions: (1) This device may not
cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation. French:
Ce dispositif est conforme à la norme CNR-210 d’Industrie Canada applicable
aux appareils radio exempts de licence. Son fonctionnement est sujet aux deux
conditions suivantes: (1) le dispositif ne doit pas produire de brouillage
préjudiciable, et (2) ce dispositif doit accepter tout brouillage reçu, y
compris un brouillage susceptible de provoquer un fonctionnement indésirable.
1000 Cranberry Woods Drive Cranberry Township, PA 16066 USA Phone
1-800-MSA-2222 Fax 1-800-967-0398
For your local MSA contacts, please go to our website www.MSAsafety.com
© MSA 2015, 2017, 2020, 2022. All rights reserved
Contents
1
Calibration Certification and Conformance
5
1.1 Calibration Certification
5
1.2 Conformance Statement
5
2
Safety Regulations
5
2.1 Correct Use
5
2.2 Liability Information
6
2.3 Safety and Precautionary Measures
6
2.4 Warranty
8
3
Description
9
3.1 Overview
9
3.2 Device Hardware Interfaces
9
3.3 Alarms
11
3.4 On-Screen Indicators
12
3.5 Viewing Additional Pages
14
3.6 Sensor Missing Alarm
18
3.7 Monitoring Toxic Gases
18
3.8 Monitoring Oxygen Concentration
19
3.9 Monitoring Combustible Gases
19
3.10 Gas Exposure of 100 % LEL
20
4
Operation
20
4.1 Environmental Factors
20
4.2 Turning ON and Fresh Air Setup
21
4.3 Special Consideration for Oxygen Sensor
23
4.4 Measurement Mode [Normal Operation]
23
4.5 Device Setup
23
4.6 Bluetooth Operation
30
4.7 MSA Link Operation
31
4.8 Function Tests on the Device
31
4.9 Bump Test
31
4.10 Calibration
33
4.11 Time of Day Testing
36
4.12 Device Shutdown
36
5
Maintenance
37
5.1 Troubleshooting
37
5.2 Verifying Pump Operation
38
5.3 Replacing the Battery
38
5.4 Maintenance Procedure – Replacing or Adding a Sensor
39
5.5 Replacing the Pump Filter
41
5.6 Cleaning the Device Exterior
41
5.7 Storage
41
5.8 Shipment
41
6
Technical Specifications
42
6.1 Factory-set Alarm Thresholds and Setpoints
43
6.2 Performance Specifications
43
6.3 Calibration Specifications
45
6.4
Combustible Gas Cross Reference Factors for General-Purpose Calibration Using Calibration Cylinder (P/N 10053022)
46
7
XCell Sensor Patents
47
GB
Altair 5X / Altair 5XiR
3
8
Ordering Information
48
8.1 US
48
8.2 Outside US
49
8.3 Accessories
49
8.4 Replacement Parts
51
9
Flow Charts
53
9.1 Basic Operation
53
9.2 Bump Test/ Informational Pages
54
9.3 Calibrations
56
9.4 Setup
57
9.5 Calibration Options
58
9.6 Alarm Options
59
9.7 Sensor Alarm Setup
60
9.8 Instrument Options
61
9.9 Sensor Setup
63
10 Changeable Feature Summary
64
4
Altair 5X / Altair 5XiR
GB
1 Calibration Certification and Conformance
1 Calibration Certification and Conformance
1.1 Calibration Certification
All applicable inspections, testing, and calibrations were performed using
NIST traceable equipment, where available, in accordance with MSA’s ISO 9001
Certified Quality System. Each material, component, and/or instrument must be
installed, operated and maintained in strict accordance with its labels,
cautions, warnings, instructions, and within the limitations stated in the
supplied instruction manual. Routine calibration checks, equipment
inspections, and applicable preventative maintenance measures must be
performed to verify that the materials, components, and/or instruments are
operating properly. Failure to perform these tasks on a routine basis, or
suggested intervals, with specified equipment or methods, may result in
inaccurate readings.
1.2 Conformance Statement
MSA certifies that the materials, components, and/or instruments delivered in
this shipment conform to all applicable specifications. The items delivered
have been processed through the appropriate approved document controlled
procedures for Receiving, Manufacturing and Inspection. The materials,
components, and/or instruments were inspected, tested, and calibrated, as
applicable, per the associated drawings, standards requirements, and/or
specifications, and were deemed acceptable by appropriate authorized
personnel.
2 Safety Regulations
2.1 Correct Use
The ALTAIR 5X and ALTAIR 5X IR Multigas Detectors, hereafter also referred to
as device, are for use by trained and qualified personnel. They are designed
to be used when performing a hazard assessment to:
· Assess potential worker exposure to combustible and toxic gases and vapors
as well as low level of oxygen. · Determine the appropriate gas and vapor
monitoring needed for a workplace.
The ALTAIR 5X Multigas Detector can be equipped to detect:
· Combustible gases and certain combustible vapors. · Oxygen-deficient or
oxygen-rich atmospheres. · Specific toxic gases for which a sensor is
installed. · US only: While the device can detect up to 30 % oxygen in ambient
air, it is approved for use only up to 21 % oxygen. · Outside the US: Oxygen
for monitoring inertisation applications. The device is suitable and certified
for the
measurement of the oxygen concentration in gas mixtures for inertisation
according to EN 50104 but without alarm function.
The ALTAIR 5X IR Multigas Detector can also contain one infrared sensor to
detect CO2 or specific combustible gases up to 100 % Vol.
WARNING! · Perform a blocked flow test before each day’s use.
· It is recommended that a Bump Test is performed before each day’s use, and
adjust if necessary.
· Perform a Bump Test more frequently if exposed to silicone, silicates, lead-
containing compounds, hydrogen sulfide, or high contaminant levels.
· Recheck calibration if unit is subjected to physical shock.
· Use only to detect gases/vapors for which a sensor is installed.
· Do not use to detect combustible dusts or mists.
· For accurate catalytic combustible readings, make sure adequate oxygen is
present (>10 % O2). · Never block pump inlet, except to perform a sampling
system safety test. Have a trained and qualified person
interpret device readings. Risk of Explosion: Do not remove battery pack,
recharge Li Ion battery, or replace alkaline batteries in a hazardous
location. Do not alter or modify device.
GB
Altair 5X / Altair 5XiR
5
2 Safety Regulations
· Use only MSA-approved sampling lines. · Do not use silicone tubing or
sampling lines. · Wait sufficient time for the reading; response times vary
based on gas and length of sampling line. · Do not use the device for
prolonged periods in an atmosphere containing a concentration of fuel or
solvent vapors
that may be greater than 10 % LEL. Failure to follow these warnings can result
in serious personal injury or death.
US only: This Class A digital apparatus complies with Canadian ICES-003.
It is imperative that this operating manual be read and observed when using
the product. In particular, the safety instructions, as well as the
information for the use and operation of the product, must be carefully read
and observed. Furthermore, the national regulations applicable in the user’s
country must be taken into account for safe use.
Alternative use, or use outside this specification will be considered as non-
compliance. This also applies especially to unauthorized alterations to the
product and to commissioning work that has not been carried out by MSA or
authorized persons.
2.2 Liability Information
MSA accepts no liability in cases where the product has been used
inappropriately or not as intended. The selection and use of this product must
be under the direction of a qualified safety professional who has carefully
evaluated the specific hazards of the jobsite where it will be used and who is
completely familiar with the product and its limitations. The selection and
use of this product and its incorporation into the safety scheme of the
jobsite is the exclusive responsibility of the employer.
Product liability claims, warranties also as guarantees made by MSA with
respect to the product are voided, if it is not used, serviced or maintained
in accordance with the instructions in this manual.
2.3 Safety and Precautionary Measures
WARNING! Carefully review the following safety limitations and precautions
before placing this device in service. Failure to follow this warning can
result in serious personal injury or death.
· Check function ( 4.8 Function Tests on the Device) each day before use. MSA
recommends carrying out a routine inspection prior to each day’s use.
· It is recommended that a Bump Test is performed 4.9 Bump Test) before each
day’s use to verify proper device operation. The device must pass the bump
test. If it fails the test, perform a calibration ( 4.10 Calibration) before
using the device.
· The ALTAIR 5X MultiGas Detectors are designed to detect gases and vapors in
air only. · If your device contains Bluetooth functionality, Bluetooth
operation is dependent upon signal availability of the
wireless service(s) necessary to maintain the communication link. Loss of
wireless signal will prevent communication of alarms and other information to
linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs. · Perform a Bump Test more frequently if the device is
subjected to physical shock or high levels of contaminants. Also, check
calibration more frequently if the tested atmosphere contains the following
materials, which may desensitize the combustible gas sensor and reduce its
readings:
Organic silicones Silicates Lead-containing compounds Sulfur compound
exposures over 200 ppm or exposures over 50 ppm for one minute.
6
Altair 5X / Altair 5XiR
GB
2 Safety Regulations
· The minimum concentration of a combustible gas in air that can ignite is
defined as the Lower Explosive Limit (LEL). A combustible gas reading of XXX
indicates the atmosphere is above 100 % LEL, and an explosion hazard exists.
Move away from hazardous area immediately.
· Do not use the device to test for combustible or toxic gases in the
following atmospheres as this may result in erroneous readings:
Oxygen-deficient or oxygen-rich atmospheres
Reducing atmospheres
Furnace stacks
Inert environments (only IR sensors acceptable for use)
Atmospheres containing combustible airborne mists/dusts.
· Do not use the ALTAIR 5X and ALTAIR 5X IR Multigas Detectors to test for
combustible gases in atmospheres containing vapors from liquids with a high
flash point (above 38 °C, 100 °F) as this may result in erroneously low
readings.
· Allow sufficient time for device to display accurate reading. Response times
vary based on the type of sensor being utilized ( 6.2 Performance
Specifications). Allow a minimum of 1 second per foot (3 seconds per meter) of
sample line to allow the sample to be drawn through the sensors.
· Sampling lines made from 0.062 inch (1.57 mm) inner diameter tubing provide
fast transport times to the device; however, they must be limited to 50 feet
(15 m) in length.
· Sampling of reactive toxic gases (Cl2, ClO2, NH3) must only be done with the
reactive gas sample line and probe kits listed in 8 Ordering Information.
· All device readings and information must be interpreted by someone trained
and qualified in interpreting device readings in relation to the specific
environment, industrial practice and exposure limitations.
· Outside the US: The device is suitable and certified for the measurement of
the oxygen concentration in gas mixtures for inertisation according to EN
50104 without alarm function.
Observe Proper Battery Maintenance
Use only battery chargers made available by MSA for use with this device;
other chargers may damage the battery pack and the device. Dispose of in
accordance with local health and safety regulations.
Be Aware of Environmental Conditions
A number of environmental factors may affect the sensor readings, including
changes in pressure, humidity and temperature. Pressure and humidity changes
also affect the amount of oxygen actually present in the atmosphere.
Be Aware of the Procedures for Handling Electrostatically Sensitive
Electronics
The device contains electrostatically sensitive components. Do not open or
repair the device without using appropriate electrostatic discharge (ESD)
protection. The warranty does not cover damage caused by electrostatic
discharges.
NOTICE This equipment has been tested and found to comply with the limits for
a Class A digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
To comply with FCC RF exposure requirements MSA ensures that an approved
antenna is installed per FCC ID: 7V1316.
PAN1326 is licensed to meet the regulatory requirements of Industry Canada
(IC), license: IC: 216Q-1316 PAN1326
This device complies with part 15 of the FCC Rules. Operation is subject to
the following conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including any
interference that may cause undesired operation.
GB
Altair 5X / Altair 5XiR
7
2 Safety Regulations
WARNING!
This is a class A product in accordance with CISPR 22. In a domestic
environment, this product may cause radio interference, in which case the user
may be required to take adequate measures. Failure to follow this warning can
result in serious personal injury or death.
This Class A digital apparatus complies with Canadian ICES-003.
Be Aware of the Warranty Regulations
The warranties made by MSA The Safety Company with respect to the product are voided if the product is not used and maintained in accordance with the instructions in this manual. Please protect yourself and others by following them. We encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to use or service.
Be Aware of the Product Regulations
Follow all relevant national regulations applicable in the country of use.
2.4 Warranty
Item Chassis and electronics XCell COMB EX,O2, H2S, CO, SO2, NO2, and MSA IR
sensors XCell Cl2, NH3 sensors Series 20 ClO2, HCN, NO, NO2, PH3 sensors,
XCell COMB EX-H In-Box Accessories including Replacements
*Period begins date of received shipment.
Warranty Period* Three years Three years Two years One year Two years
This warranty does not cover filters, fuses, etc. As the battery pack ages, there will be a reduction in usable device run time. Certain other accessories not specifically listed here may have different warranty periods. This warranty is valid only if the product is maintained and used in accordance with Seller’s instructions and/or recommendations.
The Seller shall be released from all obligations under this warranty in the event repairs or modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the Seller to any affirmation, representation or warranty concerning this product. Seller makes no warranty concerning components or accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser’s sole and exclusive remedy for breach of the above warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the replacement at Seller’s option, of any equipment or parts thereof, which after examination by Seller is proven to be defective.
Replacement equipment and/or parts will be provided at no cost to Purchaser, F.O.B. Seller’s Plant. Failure of Seller to successfully replace any nonconforming equipment or parts shall not cause the remedy established hereby to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic, special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by reason of non-operation of the goods. This exclusion is applicable to claims for breach of warranty, tortious conduct or any other cause of action against seller.
8
Altair 5X / Altair 5XiR
GB
3 Description
3.1 Overview
3 Description
Figure 1 Device view
1
LEDs 2 red “Alarm”, 1 green “Safe” and 1 yellow “Fault”
8
IRDA Communication port
2 Horn
9 Pump inlet
3 Display
10 RFID tag
4 p Button
11 Filter
5 q Button
12 Belt Clip (ALTAIR 5X only)
6 q Button
13 Charging port
7 Bluetooth Status LED
14 Charge Status LED
The device monitors gases in ambient air and in the workplace.
The ALTAIR 5X is available with a maximum of four sensors, which can display
readings for five separate gases (one TwoTox Sensor provides both CO and H2S
or CO and NO2 sensing capabilities in a single package).
The ALTAIR 5X IR is available with a maximum of five sensors, which can
display readings for six separate gases (one Two-Tox Sensor provides both CO
and H2S or CO and NO2 sensing capabilities in a single package).
The ALTAIR 5X and ALTAIR 5X IR Multigas Detectors are available with either a
monochrome or color display.
The alarm levels for the individual gases are factory-set and can be changed
through the Instrument Setup Menu. These changes can also be made through MSA
Link Software. Ensure that the latest version of the MSA Link software has
been downloaded from MSA’s website www.msasafety.com.
It is recommended that after making changes using MSA Link software, the
device should be turned OFF and ON.
US only: While the device can detect up to 30 % oxygen in ambient air, it is
approved for use only up to 21 % oxygen.
3.2 Device Hardware Interfaces
Device operation is dialog driven from the display with the aid of the three
function buttons ( Figure 1 ).
The device has three buttons for user operation. Each button can function as a
“soft key”, as defined on the display directly above the button.
GB
Altair 5X / Altair 5XiR
9
3 Description
Button Definitions
Button q q
p
Description
The q button is used to turn the device ON or OFF and to confirm user action
selections. The q button is used to page down through data screens or to
decrease the values in setup mode. This button is also used to initiate a Bump
Test for the installed sensors, directly from the MEASURING page.If the user
is granted access to the MotionAlert setting feature, this button can be used
to activate the InstantAlertTM alarm.
See 4.5 Device Setup for the means to allow/disallow user access. The p button
is used to reset Peak, STEL, TWA and alarms (where possible) or perform
calibration in measuring mode. It is also used as page up or to increase the
values in setup mode.
When the p button and the q button are pressed simultaneously while in normal
measure mode, the Setup mode can be entered after the password is confirmed.
LED Definitions
LED RED (Alarm) GREEN (Safe)
YELLOW (Fault)
BLUE (Bluetooth Status)
Description The red alarm LEDs are visual indications of an alarm condition or
any type of error in the device. The Safe LED flashes once every 15 seconds to
notify the user that the device is ON and operating under the conditions
defined below:
· The green SAFE LED is enabled · Combustible reading is 0 % LEL or 0 % Vol ·
Oxygen (O2) reading is 20.8 % · Carbon Dioxide (CO2) reading is 0.03 % · All
other sensor readings are 0 ppm · No gas alarms are present (low or high) ·
Device is not in Low Battery warning or alarm · STEL and TWA readings are 0
ppm
This option can be turned OFF through the MSA Link software. The Fault LED
activates if any of several fault conditions are detected during device
operation. This includes:
· A device memory error · A sensor determined to be missing or inoperative · A
pump fault
These faults are also indicated by activation of device alarm LEDs, horn, and
vibrating alarm. If your instrument supports Bluetooth, the blue LED is a
visual indication of the Bluetooth connection status.
· Off = Bluetooth board OFF or Undiscoverable · Fast Flash = Discoverable Mode
· Slow Flash = Connected
10
Altair 5X / Altair 5XiR
GB
3 Description
3.3 Alarms The device is equipped with multiple alarms for increased user safety:
Icon
Alarm Vibrating Alarm
The device vibrates when any alarm condition is active. This can be turned OFF through the SETUP- ALARM OPTIONS menu ( 4.5 Device Setup).
Horn
The device is equipped with an audible alarm.
The horn can be turned OFF through the SETUP- ALARM OPTIONS menu (4.5 Device Setup).
InstantAlertTM Alarm
The InstantAlert exclusive feature allows the user to manually activate an
audible alarm to alert those nearby to potentially dangerous situations.
Holding the q button for approximately 5 seconds while in Normal Measure Mode
activates the
InstantAlert alarm. Access to this feature may be restricted. See 4.5 Device
Setup for
means to allow/disallow user access.
MotionAlertTM Alarm
If MotionAlert is turned ON ( 4.5 Device Setup), the device activates a “Man Down” alarm if motion is not detected within 30 seconds. The Alarm LEDs flash, and the horn activates with an increasing audible frequency. MotionAlert is always turned OFF when the device is turned OFF.
Stealth Mode Sensor Life Alarm
Access to this feature may be restricted by user settings. See 4.10
Calibration for means to allow/disallow user access.
Stealth Mode disables the visual, audible and vibrating alarms. MSA recommends
that this feature be left in its default “OFF” state. Stealth mode can be
turned ON through the SETUP – INSTRUMENT OPTIONS menu (4.5 Device Setup). The
message “Alarms OFF” flashes on the monochrome display when Stealth mode is
ON. On the color display, all three alarm icons are shown as OFF.
The device evaluates the condition of the sensors during Calibration.
As the end of a sensor’s life approaches, a warning is provided. While the sensor is still fully functional, the warning gives the user time to plan for a replacement sensor to minimize downtime. The Sensor Life indicator displays during ongoing operations as a reminder of a sensor’s pending end of life.
When a sensor’s end-of-life is reached, sensor calibration will not be successful, and the user is then alerted by a Sensor Life Alarm. A flashing Sensor Life indicator displays during ongoing operations until the sensor is replaced and/or successfully calibrated.
On the monochrome display, the Sensor Life indicator appears on the display at the same position as the MotionAlert indicator. If MotionAlert is enabled (the
- indicator displays) and a Sensor Life warning or alarm occurs, the Sensor Life indicator takes priority and is shown instead.
Backlight
On the color display, each displayed gas will have its own Sensor Life
indicator. If a sensor is in end-of-life warning, its indicator will be an
orange . If a sensor has reached end-of-life, it is in alarm and its Sensor
Life indicator will be a continuous blinking red .
See4.10 Calibration for additional details on Sensor Life determination and
indication.
The backlight automatically activates when any front panel button is pressed
and remains ON for the duration of user-selected timeout.
This duration can be changed using the SETUP – INSTRUMENT SETUP( 4.5 Device Setup) or through MSA Link software.
GB
Altair 5X / Altair 5XiR
11
3 Description
Icon
Alarm Operating Beep
3.4 On-Screen Indicators Monochrome Display
This operating beep activates every 30 seconds by momentarily beeping the horn
and flashing the alarm LEDs under the following conditions:
· Operating beep is enabled · Device is on normal Measure Gases page · Device
is not in battery warning · Device is not in gas alarm.
The Operating Beep can be disabled using the SETUP – INSTRUMENT OPTIONS ( 4.5
Device Setup) or through MSA Link software.
Figure 2 Monochrome Display
1 Gas type
5
2 Current time
6
3 “Soft Key” q Indicator
7
4 Gas reading
8
Battery condition Successful Bump Test/Calibration Indicator “Soft Key” p
indicator MotionAlert (+ = ON)
Sensor Life Indicator
Bluetooth On/Off
On a monochrome display, a message appears every 30 seconds if the Vibration, Horn or LED alarms are turned OFF. Color Display
Figure 3 Color display
1
“Soft Key” q Indicator
8
Vibration alarm OFF
12
Altair 5X / Altair 5XiR
GB
3 Description
2
Gas reading
9
Horn OFF or successful Bump Test/Calibration Indicator
3
Gas Concentration Units 10
LED OFF
4
Current Time
11
Battery charge level
5
Motion Alert symbol ON 12
Gas Type
6
or
Wireless USB or Bluetooth ON
13
“Soft Key” p Indicator
7
Combustible Gas/VOC type
14
Sensor Life Indicator
Battery Charge Level Indicator
The battery condition icon continuously displays in the upper right-hand
corner of the color display and in the lower righthand corner of the
monochrome display. A bar represents the charging level of the battery.
The nominal run-time of the device (COMB, O2, CO, H2S, with pump and
monochrome display) at room temperature is 20 hours. Actual run-time will vary
depending on ambient temperature, battery and alarm conditions.
Low Battery Warning
WARNING!
If battery warning alarm activates while using the device, leave the area
immediately as the end of battery life is approaching. Failure to follow this
warning can result in serious personal injury or death.
Figure 4 Battery Warning
The duration of remaining device operation during a Low Battery Warning
depends on ambient temperatures, battery condition alarm status. Nominal
battery life is 30-60 minutes after the Battery Warning activates.
When the device goes into battery warning the:
· Battery life indicator continuously blinks · Alarm sounds and alarm LEDs
flash every 30 seconds · Safe LED no longer flashes · Device continues to
operate until it is turned OFF or battery shutdown occurs.
Battery Shut Down
WARNING!
If battery alarm displays, stop using the device as it no longer has enough
power to indicate potential hazards, and persons relying on this product for
their safety could sustain serious personal injury or death.
The device goes into battery shutdown mode 60 seconds before final shutdown (when the batteries can no longer operate the device):
· “BATTERY ALARM” flashes on the display · Alarm sounds
GB
Altair 5X / Altair 5XiR
13
3 Description
· Alarm LEDs flash · Fault LED is on · No other pages can be viewed; after
approximately one minute, the device automatically turns OFF.
Figure 5 Battery Shut Down When battery shutdown condition occurs ( Figure 4
):
1. Leave the area immediately. 2. Recharge or replace the battery pack.
Battery Charging
WARNING! · Risk of explosion. Do not recharge device in hazardous area. · Use
of any charger, other than the charger supplied with the device, may damage or
improperly charge the batteries. Failure to follow these warnings can result
in serious personal injury or death.
For users in Australia/ New Zealand: The charge cradle is a Class A product.
In a domestic environment, this product may cause radio interference, in which
case, the user may be required to take adequate measures.
The charger is capable of charging a completely depleted pack in less than six
hours in normal, room-temperature environments. NOTE: Allow very hot or cold
devices to stabilize for one hour at room temperature before attempting to
charge
· Minimum and maximum ambient temperature to charge the device is 10 °C (50
°F) and 35 °C (95 °F), respectively. · For best results, charge the device at
room temperature 23 °C (73 °F). To Charge the Device · Firmly insert the
charger connector into the charge port on the back of the device. · An LED in
the battery pack is used to indicate on the charge status.
Red = charging, Green = charged, yellow = fault · If a problem is detected
during charging (LED turns yellow):
Disconnect the charger momentarily to reset the charge cycle. · The battery
pack may be charged separately from the device. · During periods of non-use,
the charger may remain connected to the device/battery pack. NOTE: The charger
must be disconnected for the device to operate.
3.5 Viewing Additional Pages The Main Screen appears at device turn-ON.
Additional displays can be viewed by pressing the q button to move to the
screen as indicated by the “soft key”.
14
Altair 5X / Altair 5XiR
GB
3 Description
(For the monochromatic display the name of the page is displayed, for the
color version it is represented by an icon.) The sequence of pages are as
follows and are described Figure 1 :
Advance on
MAIN MEASURE
PAGE
BUMP TEST ?
PEAK *
MIN * 1
Advance on
WUSB *
Advance on
MOTION ALERT STATUS
- IF ENABLED ** IF WUSB IS INSTALLED 1 NOT VALID FOR ALL SENSORS
STEL1 TWA 1 DATE
DISCOVERY MODE*
VOC*
CAL DUE DATE *
Figure 6 Sequence of Pages Bump Test (BUMP page)
LAST CAL DATE *
This page allows the user to perform an automated Bump Test on the device. To
perform the test, the (YES) button is pressed. See 4.9 Bump Test for details
on performing the Bump Test. If the q button is pressed, the Bump Test is not
performed, and the display shows the next page in the sequence (PEAK). If the
p button is pressed, the Bump Test is not performed, and the display reverts
back to the normal MEASURE page.
Peak Readings (PEAK page)
Monochromatic display
Color display
PEAK
This page shows the highest levels of gas recorded by the device since turn-ON
or since peak readings were reset. To reset the peak readings:
1. Access the PEAK page. 2. Press the p button.
This page can be de-activated through MSA Link software.
Minimum Readings (MIN page)
Monochromatic display
Color display
MIN
GB
Altair 5X / Altair 5XiR
15
3 Description
This page shows the lowest level of oxygen recorded by the device since turn-
ON or since the MIN reading was reset. It is only shown if an oxygen sensor is
installed and enabled. To reset the MIN reading:
1. Access the MIN page. 2. Press the p button. Short Term Exposure Limits
(STEL page)
WARNING! If the STEL alarm activates, leave the contaminated area immediately;
the ambient gas concentration has reached the preset STEL alarm level. Failure
to follow this warning will cause over-exposure to toxic gases and persons
relying on this product for their safety could sustain serious personal injury
or death.
Monochromatic display
STEL
Color display
This page shows the average exposure over a 15-minute period. When the amount
of gas detected by device is greater than the STEL limit:
· Alarm sounds, alarm lights flash. · Alarm LEDs flash · “STEL ALARM” message
flashes. To reset the STEL: 1. Access the STEL page. 2. Press the p button.
The STEL alarm is calculated over a 15-minute exposure. STEL calculation
examples: Assume the device has been running for at least 15 minutes: 15
minute exposure of 35 ppm:
(15 minutes x 35 ppm) 15 minutes
= 35 ppm
10 minute exposure of 35 ppm and 5 minutes exposure of 15 ppm:
(10 minutes x 35 ppm) + (5 minutes x 5 ppm) 15 minutes
= 25 ppm
This page can be de-activated through MSA Link software.
16
Altair 5X / Altair 5XiR
GB
Time Weighted Average (TWA page)
3 Description
WARNING!
If the TWA alarm activates, leave the contaminated area immediately; the
ambient gas concentration has reached the preset TWA alarm level. Failure to
follow this warning will cause over-exposure to toxic gases and persons
relying on this product for their safety could sustain serious personal injury
or death.
This page shows the average exposure over 8 hours since the device was turned ON or since the TWA reading was reset. When the amount of gas detected is greater than the eight-hour TWA limit:
Monochromatic display
TWA
Color display
· Alarm sounds · Alarm LEDs flash · “TWA ALARM” message flashes. To reset the TWA Readings: 1. Access the TWA page. 2. Press the p button. The TWA alarm is calculated over an eight-hour exposure. TWA calculation examples: 1 hour exposure of 50 ppm:
(1 hour x 50 ppm) + (7 hours x 0 ppm) 8 hours
= 6.25 ppm
4 hour exposure of 50 ppm and 4 hour exposure of 100 ppm:
(4 hours x 50 ppm) + (4 hours x 100 ppm) = 75 ppm 8 hours
12 hour exposure of 100 ppm:
(12 hours x 100 ppm) 8 hours
= 150 ppm
This page can be de-activated through MSA Link software.
Date Display
Current date appears on the display in the format: MM-DD-YY.
Last cal page
Displays the device last successful calibration date in the format: MM-DD-YY.
This page can be de-activated through MSA Link software or the SETUP – CAL
OPTIONS page.
Cal due page
Displays the days until the device’s next calibration is due (user
selectable). This page can be de-activated through MSA Link software or the
SETUP – CAL OPTIONS page.
GB
Altair 5X / Altair 5XiR
17
3 Description
Discoverable Mode page Allows the user to put the device into Bluetooth
discoverable mode in order to pair with another device. This page can be
deactivated through the SETUP – INSTRUMENT OPTIONS page.
Motion Alert Activation Page When the MotionAlert feature is active, the +
symbol appears. The device enters pre-alarm when no motion is detected for 20
seconds. This condition can be cleared by moving the device. MotionAlert is
turned OFF each time the device is powered OFF. After 30 seconds of no motion,
the full MotionAlert alarm is triggered. This alarm can only be cleared by
pressing the p button. This page displays if it was selected in Setup Mode. To
activate or deactivate the MotionAlert feature, press the p button while the
MOTIONALERT ACTIVATION page is displayed.
3.6 Sensor Missing Alarm Enabled IR and XCell sensors are continuously
monitored for proper function. If, during operation, the IR or an XCell sensor
is detected as failed or disconnected, this alarm message appears.
· “SENSOR MISSING” flashes on the display. · The problematic sensor is
indicated. · The alarm sounds and the Fault and Alarm LEDs flash. · The alarm
can be silenced by pressing the p button; no other pages can be viewed.
WARNING! When this alarm occurs, the device is inoperative for measuring
gases. The user must exit the hazardous area, the device must be powered down,
and the sensor situation must be corrected. Failure to follow this warning can
result in serious personal injury or death.
3.7 Monitoring Toxic Gases The device can monitor the concentration of a
variety of toxic gases in ambient air. Which toxic gases are monitored depends
on the installed sensors. The device displays the gas concentration in parts
per million (ppm), µmol/mol or mg/m3 on the Measuring page. Gas units are
selected in the SETUP – INSTRUMENT OPTIONS page.
WARNING! If an alarm is triggered while using the device, leave the area
immediately. Remaining in the area under such circumstances can cause serious
personal injury or death.
The device has four gas alarms:
· HIGH Alarm · LOW Alarm · STEL Alarm · TWA Alarm
Figure 7 Alarm Conditions (here High Alarm)
18
Altair 5X / Altair 5XiR
GB
3 Description
If the gas concentration reaches or exceeds the alarm set point or the STEL or
TWA limits, the: · Alarm message displays and flashes in combination with the
corresponding gas concentration · Backlight turns on · Alarm sounds (if
active) · Alarm LEDs flash (if active) · Vibrating alarm triggers (if active)
3.8 Monitoring Oxygen Concentration The device monitors the oxygen
concentration in ambient air. The alarm set points can be set to activate on
two different conditions:
· Enriched – oxygen concentration > 20.8% or · Deficient – oxygen
concentration < 19.5%.
WARNING!
If an alarm is triggered while using the device, leave the area immediately.
Remaining in the area under such circumstances can cause serious personal
injury or death.
When the alarm set point is reached for either of the above conditions the:
· Alarm message displays and flashes in combination with the corresponding gas
concentration · Backlight turns on · Alarm sounds (if active) · Alarm LEDs
flash (if active) · Vibrating alarm triggers (if active)
The LOW alarm (oxygen deficient) is latching and will not automatically reset
when the O2 concentration rises above the LOW set point. To reset the alarm
press the p button. If the alarm is latching, the p button silences the alarm
for five seconds. Alarms can be made latching or unlatching via MSA Link
software.
False oxygen alarms can occur due to changes in barometric pressure
(altitude), humidity or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at the temperature
and pressure of use. Be sure that the device is in known fresh air before
performing a calibration.
3.9 Monitoring Combustible Gases
The device can be equipped with a catalytic combustible sensor that detects a
variety of combustible gases up to 100 % LEL and displays the reading as
either % LEL or % CH4. The ALTAIR 5X IR can also contain an IR combustible
sensor. The IR sensor displays the reading in % Vol or %LEL.
WARNING!
If an alarm is triggered while using the device, leave the area
immediately.Remaining in the area under such circumstances can cause serious
personal injury or death.
The catalytic combustible, 25 % Vol Butane IR, and 100% LEL Propane IR sensors
have two alarm setpoints: · HIGH Alarm · LOW Alarm
If the gas concentration reaches or exceeds the alarm set point, the device: ·
Alarm message displays and flashes in combination with the corresponding gas
concentration:
GB
Altair 5X / Altair 5XiR
19
4 Operation
· Backlight turns on · Alarm sounds (if active) · Alarm LEDs flash (if active)
The 100 % Vol IR sensors have no alarm setpoints.
3.10 Gas Exposure of 100 % LEL
When gas reading exceeds 100 % of the lower explosive limit (LEL), the device
enters a Lock Alarm state and displays “XXX” in place of the actual reading.
WARNING! A catalytic combustible gas reading of “XXX” indicates the atmosphere
could be above 100 % LEL or 5.00 % Vol CH4 and an explosion hazard exists.
Move away from contaminated area immediately. Failure to follow this warning
can result in serious personal injury or death.
For ALTAIR 5X IR devices with an enabled 100 % Vol methane IR sensor, the
LockAlarm will clear, and the catalytic combustible again displays combustible
concentrations when the gas sample drops to a lower level. For devices without
an enabled 100 % Vol methane IR sensor, the user can clear the LockAlarm state
only by turning the device OFF, and then ON again in a fresh air environment.
When catalytic combustible gas reading digits appear, the device is available
for measuring gases once again.
LockAlarm of the catalytic combustible sensor occurs during Bump Testing and
calibration of a % Vol combustible IR sensor. After the IR sensor Bump Test,
the LockAlarm must be cleared (as described above) before the catalytic
combustible sensor is again able to measure and provide readings.
NOTE: Check your national standard values for 100 % LEL.
4 Operation
Device operation is dialog driven from the display with the aid of the three
function buttons ( 3.2 Device Hardware Interfaces).
For more information, see the flow charts in 9 Flow Charts.
4.1 Environmental Factors
A number of environmental factors may affect the gas sensor readings,
including changes in pressure, humidity and temperature. Pressure and humidity
changes affect the amount of oxygen actually present in the atmosphere.
Pressure Changes
If pressure changes rapidly (e.g., stepping through airlock), the oxygen
sensor reading may temporarily shift and possibly cause the device to go into
alarm. While the percentage of oxygen may remain at or near 20.8 Vol %, the
total amount of oxygen present in the atmosphere available for respiration may
become a hazard if the overall pressure is reduced by a significant degree.
Humidity Changes
If humidity changes by any significant degree (e.g., going from a dry, air
conditioned environment to outdoor, moisture laden air), oxygen readings can
be reduced by up to 0.5 %, due to water vapor in the air displacing oxygen.
The oxygen sensor has a special filter to reduce the effects of humidity
changes on oxygen readings. This effect will not be noticed immediately, but
slowly impacts oxygen readings over several hours.
Temperature Changes
The sensors have built-in temperature compensation. However, if temperature
shifts dramatically, the sensor reading could shift.
20
Altair 5X / Altair 5XiR
GB
4 Operation
4.2 Turning ON and Fresh Air Setup
Device operation is dialog driven from the display with the aid of the three function buttons ( 3.2 Device Hardware Interfaces).
For more information, see the flow charts in 9 Flow Charts.
Turn the device ON with the button.
The device performs a self test:
During the self test, the device checks alarm LEDs, audible alarm, vibrating alarm and installed sensors.
The device displays:
· Startup logo · Software version, device serial number, company name, department and user names · IC / FCC ID Identifier · Sampling system safety test
During the turn-ON sequence, if a sensor was changed since the previous device
operation, the current listing of the installed sensors displays and user
interaction is required.
NOTE: The user must accept the new configuration by pressing the p button. If
the current sensor configuration is not accepted, the device alarms and is not
usable.
· Combustible gas type, and installed sensor indication · Combustible gas type
and sensor units (monochrome display only) · Alarm setpoints Low Alarm · Alarm
setpoints High Alarm · Alarm setpoints STEL Alarm (if enabled) · Alarm
setpoints TWA Alarm (if enabled) · Settings for calibration cylinder · Current
date · Last calibration date (if enabled) · CAL due date. If the calibration
due date is enabled, the message “CAL DUE; X DAYS” appears on the device
display. X = the number of days until a calibration is due, user selectable
for 1 to 180 days.
NOTE: If the number of days until calibration is due reaches 0, an alert
occurs and “CAL DUE, NOW” displays. Press the p button to clear the alert
· Sensor warm-up period · Fresh Air Setup option (if enabled).
The Main Measure Page will appear.
The presence of a indicator on the display means a sensor is approaching or has reached its end-of-life. See 3.3 Alarms for details on the Sensor Life Alarm situation.
Refer to flowchart in 9.1 Basic Operation.
Sampling System Safety Test
Upon startup, an alarm (visual, audible and vibrating) is triggered and the customer is prompted to block the pumps/sampling system of the device within 30 seconds.
When the device detects a pump flow block, it will display a PASS message. The startup sequence will resume.
If the device does not detect a pump flow block, it will display an error message.
GB
Altair 5X / Altair 5XiR
21
4 Operation
The device will shut OFF after the customer acknowledges this message by
pressing the p button.
Check your sampling system if this occurs and contact MSA as needed.
Users can check the operation of the sampling system anytime during operation
by blocking the sampling system to generate a pump alarm.
WARNING! · Do not use the pump, sample line, or probe unless the pump alarm
activates when the flow is blocked. Lack of an
alarm is an indication that a sample may not be drawn to the sensors, which
could cause inaccurate readings. Failure to follow the above can result in
serious personal injury or death. · Never let the end of the sampling line
touch or go under any liquid surface. If liquid is drawn into the device,
readings will be inaccurate and device could be damaged. We recommend the use
of an MSA sample probe containing a special membrane filter, permeable to gas
but impermeable to water, to prevent such an occurrence. Failure to follow
this warning can result in serious personal injury or death.
4.2.1 Fresh Air Setup (FAS) at Device Turn-ON
The Fresh Air Setup (FAS) is for automatic ZERO adjustment of the device.
The FAS has limits. If a hazardous level of gas is present, the device ignores
the FAS command and the device alarm activates.
The ability to perform an FAS at device turn-ON can be disabled by using MSA
Link software.
NOTE: The Fresh Air Setup is not available for the CO2 sensor.
WARNING! · Do not perform the Fresh Air Setup unless you are certain you are
in fresh, uncontaminated air; otherwise,
inaccurate readings can occur which can falsely indicate that a hazardous
atmosphere is safe. If you have any doubts as to the quality of the
surrounding air, do not use the Fresh Air Setup feature. · Do not use the
Fresh Air Setup as a substitute for daily calibration checks. The calibration
check is required to verify span accuracy. Failure to follow this warning can
result in serious personal injury or death.
Figure 8 Fresh Air Setup
The device displays a blinking “FRESH AIR SETUP?”, prompting the user to perform a Fresh Air Setup:
1. Press the p button to bypass the Fresh Air Setup. The Fresh Air Setup is
skipped and the device goes to the Measuring page (Main page).
2. Press the q button to perform the Fresh Air Setup. a. The device starts
the FAS sequence and the FAS screen displays. b. A progress bar shows the user
how much of the FAS has been completed. c. At the end of the FAS, the device
displays either “FRESH AIR SETUP PASS” or “FRESH AIR SETUP FAIL”.
NOTE: If the FAS fails, perform a zero calibration ( 4.10 Calibration).
22
Altair 5X / Altair 5XiR
GB
4 Operation
4.3 Special Consideration for Oxygen Sensor Under the following situations,
the oxygen sensor display reading may be suppressed for up to 30 minutes at
device turnON as a sensor ‘cook down’ is performed. This could occur if the:
· Oxygen sensor was just installed · Battery pack was allowed to be deep-
discharged · Battery pack was removed from the device. During this time, the
oxygen sensor numeric position on the display indicates “PLEASE WAIT”. While
this message displays, the device cannot respond to a: · Fresh Air Setup ·
Calibration · Bump Test procedure. When the numeric oxygen reading appears,
the FAS, calibration, or Bump Test procedures may be performed.
4.4 Measurement Mode [Normal Operation] The following options pages can be
executed from the Measurement screen:
BUMP page Peak Page*
This page allows user to perform a Bump Test on installed sensors This page shows the peak readings for all sensors.
Min Page
This page shows the minimum readings for the oxygen sensor.
STEL Page TWA Page Date Page
This page shows the calculated STEL readings of the device. This page shows the calculated TWA readings of the device. This page shows actual date settings of the device.
Last Cal Date
This page shows the date of the last calibration.
Cal Due*
This page shows the set date for the next calibration.
Discoverable Mode
Motion Alert Wireless USB
This page allows the user to put the instrument into Bluetooth discoverable
mode in order to pair with another device. This page allows the Motion Alert
Feature to be activated or deactivated.
This page allows the Wireless USB communication to be -activated or
deactivated.
- The display of these pages can be de-activated through MSA Link software. For further information see 10 Changeable Feature Summary.
4.5 Device Setup The device has provisions to access and modify the following parameters through direct button interface:
· Calibration Options · Alarm Options · Instrument Options
GB
Altair 5X / Altair 5XiR
23
4 Operation
These menus can be accessed only from the measure page by pressing and holding
the q and p buttons simultaneously until you are prompted for a password. The
operation is as follows:
1. Turn the device ON and wait until the measure page appears. 2.
Simultaneously press and hold the q and p buttons for approximately five
seconds.
a. The default password is “672”.
PASSWORD
000 3. Enter the first digit by pressing the q or p button and confirm with
the q button. The cursor jumps to the second digit. 4. Enter the second as
well as the third digits. Incorrect password: device returns to the Main Page.
Correct password: user can enter the Setup mode.
The password can be changed with a PC through the MSA Link software. If the
password is forgotten, it can be reset by using MSA Link software. Contact MSA
Customer Service for assistance. The following Options are available by
pressing the q and p buttons:
· Calibration Options – see 9.5 Calibration Options · Alarm Options – see 9.6
Alarm Options · Instrument Options – see 9.8 Instrument Options
4.5.1 Calibration Setup CALIBRATION OPTIONS
The Calibration Options menu has provisions to:
· modify the calibration cylinder settings (CYLINDER SETUP) · enable/disable calibration due and to set the number of days (CAL DUE OPTIONS) · enable/disable the option to show the last cal date at turn on and (LAST CAL DATE)
When enabled, the date of the last device calibration displays during the turn-ON process.
· enable/disable the option for password protected calibration (CAL PASSWORD)
When enabled, the device setup password must be entered prior to calibration.
Press:
· the button go to next page
the button to go previous page
the button to enter setup.
Setting Calibration Cylinder
24
Altair 5X / Altair 5XiR
GB
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
1. Press the button to enter setup.
NOTE: The screen for the first calibration cylinder displays.
2. Press the q or p button to change the value. the button to confirm the
setup. NOTE: With this confirmation the device automatically moves to the next
cylinder setting.
3. Repeat the sequence for changing the required settings for all necessary
gas values. NOTE: After the last setting is performed, the device returns to
the Calibration Options menu.
Setting Cal Due Options
1. Press the button to enter setup. 2. Press the q or p button to
enable/disable this option. 3. Press the button to confirm. 4. After
confirmation the device prompts the user to enter the number of days for the
reminder. 5. Change number of days by pressing the q or p button. 6. Press the
button to go to the next menu.
Setting Last Cal Date
1. Press the button to enable/disable this option. 2. Press the q button to
go to the next page. 3. Press the p button to go to the previous page.
Setting Calibration Password
1. Press the button to enable/disable this option. 2. Press the q button to
go to the next page. 3. Press the p button to go to the previous page.
Back To Main Menu
1. Press the button to go to Device Setup Menu NOTE: The Cal Options screen
displays
2. Press the q button to go to the next (Alarm options) or the p button to
exit the Setup menu.
4 Operation
GB
Altair 5X / Altair 5XiR
25
4 Operation 4.5.2 Alarm Setup
ALARM OPTIONS
The Alarm Options Menu allows the user to: · Enable/disable the vibrating
alarm · Enable/disable the audible alarm (horn) · Enable/disable the Alarm
LEDs · Enable/disable the MOTIONALERT SELECTION page. NOTE: If disabled, the
user cannot change the device MotionAlert setting. · Set Sensor Alarms.
Press · the q button go to next page · the p button to go previous page · the
button to enter setup.
Setting Vibrating Alarm Press the button to enable/disable this option.
Setting Horn Alarm Press q button to enable/disable this option. Setting LED
Alarm Press button to enable/disable this option. Setting MotionAlert Access
Setting this parameter allows the user to access the MOTIONALERT page from the
MEASURE page. If access is denied here:
· the user cannot access the MOTIONALERT page to enable or disable that
feature · the InstantAlert feature (3.3 Alarms) cannot be activated. 1. To
grant or deny user access to the MOTIONALERT page, use the button to change
the indicated selection.
User access is: permitted when the setting indicates ON. denied when the
setting indicates OFF.
2. The selection is confirmed by pressing either the q or p button. Setting
Sensor Alarms This page allows modifying the preset alarm values of:
· LOW Alarm · HIGH Alarm · STEL Alarm · TWA Alarm. NOTE: Factory set alarm
levels are shown in 6.1 Factory-set Alarm Thresholds and Setpoints.
26
Altair 5X / Altair 5XiR
GB
1. Press the button to enter Sensor Alarm setup. LOW Alarm Setup screen displays.
Figure 9 Sensor Alarm Setup 2. Press · the q button to abort the operation or · the p button to go to next alarm setup or · the button to change the alarm setpoints. NOTE: Alarm Value for the first Sensor displays.
Figure 10 Sensor Alarm Setup 3. Set values for Sensor Alarm by pressing the q
or p button. 4. Press the q button to confirm set value. 5. Repeat setting for
all other sensors. 6. Press the p button to return to the Alarm Options menu.
7. Repeat setting for all other alarm types.
4.5.3 Instrument Options
SETTINGS
The Instrument Options menu allows modification of different device options:
· Sensor Setup (enable/disable the channel) · Language Setup · Time Date Setup
· Datalog Intervals · Stealth Mode · Operating Beep · Display Contrast
(monochrome only) · Backlight Options
GB
Altair 5X / Altair 5XiR
4 Operation 27
4 Operation
· Bluetooth Press
· the q button go to next page the p button to go previous page the button to
enter setup.
Setting Sensor Options 1. Press the button to enter setup. NOTE: Following
screen displays:
Figure 11 Sensor Options Setup 2. Press the button to select sensor, press the
button to make changes. NOTE: The sensor information is displayed and the
sensor can be enabled or disabled.
Other operations such as changing the gas type (Methane, Butane, Propane etc.
for the combustible sensor) and units (ppm to mg/m3) are only possible using
the MSA Link software.
3. Change status by pressing the q or p button. 4. Press the button to
confirm and advance to next screen (next sensor). 5. Perform the sequence for
all other sensors.
NOTE: After setting up the last sensor the device goes to the next Setup Page.
Language Setup This option is for setting the language of the device.
1. Press the button to enter setup. 2. Change language by pressing the q or p
button. 3. Confirm with the button.
NOTE: The device goes to the next Setup Page. Time and Date Setup This option
is for setting the device time and date. The device first prompts to set the
time and then it prompts for the date.
The time can be set up for either regular AM/PM or military time (through MSA Link software). AM/PM time is the default setting.
1. Press the button to enter setup. 2. Change hours by pressing the q or p button. 3. Confirm with the button. 4. Change minutes by pressing the q or p button. 5. Confirm with the button.
28
Altair 5X / Altair 5XiR
GB
NOTE: The device goes to the Set Date Page. 6. Change month, date and year by
pressing the q or p button and confirming with the button.
NOTE: The device goes to the next Setup Page. 7. Confirm with the button.
NOTE: The device goes to the next Setup Page.
Setting Datalog Intervals
This option is for setting the intervals at which all the readings will be logged.
1. Press the button to enter setup. 2. Change interval by pressing the q or p
button. 3. Confirm with the button.
NOTE: The device goes to the next Setup Page.
Setting Stealth Mode
Stealth mode disables the visual, audible and vibrating alarms.
1. Press the q button to change mode (ON/OFF). 2. Press the q button to go to the next page or the p button to return to previous page.
Setting Operating Beep
1. Press the q button to change mode (ON/OFF). 2. Press the q button to go to the next page or the p button to return to previous page.
Setting Display Contrast (monochrome display) 1. Press q or p button to adjust the contrast level. 2. Press q button to confirm the contrast level.
Setting Backlight
1. Press the q button to enter setup. Change option by pressing the q or p
button.
2. Press the q button to enter. 3. Change timeout by pressing the q or p
button. 4. Press q button to confirm timeout.
Setting Bluetooth
1. Press the q button to change mode (ON/OFF). 2. Press the q button to go to the next page or the p button to return to previous page.
Back To Main Menu There are three options at this point:
the q button the p button the q button
Sensor Options menu
Previous Setup page in the Instrument Options menu
Instrument Options menu
4 Operation
GB
Altair 5X / Altair 5XiR
29
4 Operation
4.6 Bluetooth Operation
NOTE: Versions of this product manufactured July 2022 and later may not
contain Bluetooth wireless technology. This will be indicated by the front
case of the detector not utilizing a Bluetooth logo. All references in this
manual to Bluetooth will not pertain to these version of the device.
The Bluetooth communication device must be enabled for any Bluetooth functions
to operate. See 4.5 Device Setup. Compatible Bluetooth host with appropriate
software is required for proper operation.
Bluetooth Security
The Bluetooth connection is encrypted and secured with a unique six digit pin
that must be double confirmed on both device and Bluetooth host at the time of
pairing.
Discovery Mode
This device mode is used to enable a Bluetooth host to pair with the device
for the first time or if a different Bluetooth host was connected with the
device previously.
Note that the device will automatically enter discovery mode for five minutes
at device turn on if Bluetooth has been enabled. Discovery mode will also be
entered for 5 minutes following a disconnection.
To manually enter discovery mode: 1. Page down through the menu pages in
Measurement Mode using the q button until the Discovery Mode page is
displayed. 2. Press the q button to enter discovery mode. The blue led will
blink rapidly indicating that the device is in Discovery Mode.
Connecting the device to a Bluetooth host for the first time
1. Ensure that the device is on and in Discovery Mode 2. On the Bluetooth
host, locate the Bluetooth device list. Select “A5X-xxxxxxxx” from the list.
Both the device and Bluetooth host will display a unique six digit security
code to ensure that the correct devices are being paired. 3. After confirming
that the six digit codes match, confirm the pairing request on the device by
pressing the q button. 4. Confirm on the Bluetooth host as well.
Connecting the Device to a Bluetooth Host
If this was the last device connected to the Bluetooth host, the Bluetooth
host can connect to the device whether or not the device is in discovery mode
as long as Bluetooth is enabled. The six digit code confirmation will not be
displayed.
The device will only recall the last Bluetooth host it was paired with. If
connecting to another Bluetooth host, the device must be placed into discovery
mode to be detected.
Pairing the device to a Bluetooth host
This device has an integrated RFID chip to facilitate a faster Bluetooth
pairing process with a Bluetooth host that supports a RFID or NFC reader with
appropriate software. Simply align the RFID or NFC reader of the Bluetooth
host directly over the MSA logo on the front of the device.The device and
Bluetooth host should become paired and connected.
Disconnecting the Device from a Bluetooth Host
The device does not have a disconnect feature as this would be initiated by
the Bluetooth host. Use the Bluetooth host functions to purposefully
disconnect the device from the Bluetooth host.
Device Configuration over Bluetooth Connection
30
Altair 5X / Altair 5XiR
GB
4 Operation
The device has the ability to receive updates to device settings over the
Bluetooth connection. The user must successfully pair the device and Bluetooth
host confirming that the six digit security code matches both on the device
and the Bluetooth host. After a configuration change has been initiated, the
user must confirm the request on the device by pressing the q button.
Evacuation alert over Bluetooth connection
The device has the ability to receive an evacuate message over the Bluetooth
connection. The user must successfully pair the device and Bluetooth host
confirming that the six digit security code matches both on the device and the
Bluetooth host. Once connected, an evacuate message sent to the device will
send the device into alarm while displaying EVACUATE on the display. Press the
p button to silence the Evacuation alert and confirm the alert was received.
Press the p button a second time to reset the Evacuation alert once in a safe
area.
4.7 MSA Link Operation
Connecting device to PC
1. Switch ON the device and align the Datalink Communication port on the
device to the IR interface of the PC. 2. Start the MSA Link software on the PC
and start the connection by clicking the connect icon.
4.8 Function Tests on the Device
Alarm Test
1. Turn ON the device.
The user should verify that:
· alarm LEDs flash · horn sounds briefly · vibrating alarm triggers briefly.
4.9 Bump Test
WARNING!
Perform a Bump Test before each day’s use to verify proper device operation.
Failure to perform this test can result in serious personal injury or death.
Bump test frequency is often stipulated by national or corporate regulations; however, bump testing before each day’s use is generally the accepted best safety practice and is therefore MSA’s recommendation.
This test quickly confirms that the gas sensors are functioning. Perform a
full calibration periodically to ensure accuracy and immediately if the device
fails the Bump Test. The Bump Test can be performed using the procedure below
or automatically using the GALAXY GX2 Test Stand.
CSA requires (per 22.2 NO. 152) that combustible sensor sensitivity be tested
before each day’s use on a known concentration of methane equivalent to 25 to
50 % of full scale concentration. ACCURACY MUST BE WITHIN 0 to +20 % OF
ACTUAL. Correct accuracy by performing the calibration procedure described in
4.10 Calibration.
NOTE: The Automated Test Stands cannot test the following sensors:
GALAXY Chlorine Dioxide % Vol Butane % Vol Propane % Vol Methane
GALAXY GX2 Chlorine Dioxide % Vol Butane % Vol Propane
For these sensors, use this Bump Test procedure.
GB
Altair 5X / Altair 5XiR
31
4 Operation
Equipment
See 0.1 Accessories for ordering information for these components.
· Calibration Check Gas Cylinder(s)
See 6.3 Calibration Specifications for calibration gas target values and
appropriate MSA calibration gas cylinders.
· Demand Flow Regulator(s) · Tubing appropriate for the gases to be tested ·
Kits containing tubing and regulators suitable for reactive and non-reactive
gases are available from MSA.
Performing a Bump Test
For ALTAIR 5X IR devices with combustible % Vol IR sensors, the following gas
levels should not be exceeded when used for daily Bump Tests:
· IR Butane 25 % Vol – 8 % Vol Butane Cal Check gas · IR Propane 100 % Vol –
50 % Propane Cal Check gas · IR Methane 100 % Vol – 20 % Methane Cal Check
gas. · IR Propane 100% LEL 1. While the device is turned ON in a clean, fresh
air environment, verify that readings indicate no gas is present. 2. From the
normal measure screen press the q button to display “BUMP TEST?”. 3. Verify
the gas concentrations displayed match the Calibration Check Gas Cylinder. If
they do not, adjust the values
through the Calibration Setup menu. NOTE: Depending on the sensors installed,
there could be one to five separate Bump Tests performed, each with a
different cylinder, regulator, and tubing used. 4. Attach the demand regulator
(supplied in the calibration kit) to the cylinder providing the indicated
gases. 5. Connect tubing (supplied in the calibration kit) to the regulator.
6. Attach the other end of tubing to the device pump inlet. 7. Press the
button to start the bump test:
the progress bar advances the sensors respond to the gas.
The message BUMP TEST PASS indicates a successful Bump Test of the sensors.
If any sensor fails the Bump Test the:
· Message BUMP TEST FAIL appears · Failed sensor is indicated.
If there are more sensors to be Bump Tested, the next sensor displays and the
process repeats from Step 4.
If there are no more sensors to be Bump Tested, the tubing can be removed from
the device pump inlet.
In the ALTAIR 5X IR, Bump Testing of a combustible IR sensor causes the
catalytic combustible sensor to enter the LockAlarm condition. While a device
with a % Vol methane IR sensor automatically recovers from LockAlarm, propane
or butane IR devices do not. For these devices, clear the LockAlarm state by
turning the device OFF, then ON while in a fresh air environment. See 3.9
Monitoring Combustible Gases for additional details.
After the Bump Test
After all installed sensors pass the Bump Test, the symbol displays on the
-MEASURE page. This symbol appears on the:
· color display in the upper feature bar · monochrome display in the lower-
right corner.
32
Altair 5X / Altair 5XiR
GB
4 Operation
If any sensor was not bump tested, or fails the Bump Test, the symbol does not
display.
The color display:
· temporarily shows the symbol at each gas reading for successfully bump
tested sensors · symbol is then replaced by the present gas reading.
The monochrome display does not show symbols for individual gas readings. The
symbol shows for 24 hours after the Bump Test. If a sensor fails the Bump
Test, calibrate the device as described in 4.10 Calibration.
4.10 Calibration
The ALTAIR 5X can be calibrated either manually using this procedure or
automatically using the GALAXY or GALAXY GX2 test stand. Refer to4.10
Calibration .
The use of the demand regulators listed in 8 Ordering Information is
recommended. If a new sensor has been installed, the battery pack has been
depleted or a new battery pack has been installed allow sensors to stabilize
for 30 minutes before calibration is performed.
WARNING! · Special conditions with toxic gases! If the device is to be checked
or calibrated for reactive gases, prerequisites are
required; otherwise, incorrect calibration would result in incorrect device
operation. · Reactive toxic gases (e.g., chlorine, ammonia, chlorine dioxide)
have the property of diffusing into the rubber and
plastic tubes so that the volume of test gas available in the device would no
longer be sufficient to correctly perform device calibration. · When
calibrating the device with toxic gases, certain prerequisites are required,
otherwise incorrect calibration could result:
A special pressure regulator Shortest possible connection tubes between the
pressure regulator and the device Connection tubes made from a material that
does not absorb the test gases (e.g., PTFE). Failure to follow these warnings
can result in serious personal injury or death.
NOTE: If using normal tubes and pressure regulators, expose them to the
required test gas for an extended time period. Keep these materials dedicated
for use with that test gas only; do not use them for other gases.
For example, for chlorine, allow the entire contents of a test gas cylinder to
flow through the pressure regulator and tubes before using to calibrate the
device. Mark these materials for use with chlorine only.
4.10.1 Zero Calibration Procedure
1. Press the p button for five seconds in Normal Measurement page. ZERO
screen displays.
To skip the ZERO procedure and move directly to the span calibration
procedure, push the p button. If no button is pushed for 30 seconds, the
device prompts user to perform a SPAN calibration before device returns to the
Normal Measurement page.
To perform ONLY a Fresh Air Setup at this time, press the button. The device
then performs a Fresh Air Setup as described in 4.2 Turning ON and Fresh Air
Setup. When the Fresh Air Setup is complete, the device returns to the normal
Measure screen.
GB
Altair 5X / Altair 5XiR
33
4 Operation
2. Press the q button to confirm the ZERO screen, i. e. to execute zero
calibration. The message “SENSOR REFRESH” displays, followed by the message
“ZERO CALIBRATION”. The “REFRESH” message does not appear if a catalytic
combustible sensor is not installed. ZERO calibration starts. A progress bar
shows the user how much of the calibration has been completed.
During the first moments of a ZERO calibration, the combustible sensor reading
may be replaced by a moving display of “PLEASE WAIT”. This is normal. NOTE:
After the ZERO calibration is completed the device displays either
· “ZERO CALIBRATION PASS”
or
· “ZERO CALIBRATION FAIL”.
NOTE: Only if the device passes the zero calibration the SPAN screen displays.
4.10.2 Span Calibration
To skip the Span calibration procedure, push the p button.
NOTE: If the SPAN calibration of the combustible sensor is skipped after a
successful ZERO calibration, the combustible sensor reading may be replaced
with a moving display of “PLEASE WAIT” for a few moments. This is normal, and
the device is fully operational once a combustible gas reading reappears.
If no button is pushed for 30 seconds, span calibration is skipped.
Because of the different possible combinations of gases that are possible,
skipping a Span calibration could advance the user to the Span calibration of
another installed sensor, or back to Measuring mode.
When calibrating with combustible gases > 100 % LEL, select the “Yes” option
to prompt “Span Calibration?” BEFORE applying gas to the device.
1. Connect one end of tubing to the cylinder regulator (supplied in the
calibration kit).
2. Connect the other end of the tubing to the pump inlet.
3. Press the q button to calibrate (span) the device. “SPAN CALIBRATION”
flashes SPAN calibration starts. A progress bar shows the user how much of the
calibration has already been completed.
34
Altair 5X / Altair 5XiR
GB
NOTE: After the SPAN calibration is completed, the device displays either ·
“SPAN CALIBRATION PASS” or · “SPAN CALIBRATION FAIL”
NOTE: The device returns to Measuring mode.
4 Operation
If a sensor is nearing its end-of-life, the “PASS” display is followed by the
Sensor Life indicator display.
· While the sensor is still fully functional, this warning gives the user time
to plan for a replacement sensor to minimize downtime.
· The indicator blinks as the device returns to Measure mode. · After 15
seconds, the blinking stops, but the indicator continues to display during
ongoing operations as a reminder
of a sensor’s pending end-of-life.
If a span calibration fails:
· The Sensor Life Indicator blinks to show a sensor has reached its end-of-
life and should be replaced. · The device remains in the Sensor Life alarm
condition until the p button is pressed. · After the alarm is cleared, the
device enters Measure mode and the Sensor Life indicator blinks during ongoing
operations until the sensor is replaced and/or successfully calibrated.
Span calibration can fail for reasons other than a sensor at the end of its
life. If a span calibration failure occurs, verify items such as:
· sufficient gas remaining in the calibration cylinder · gas expiration date ·
integrity of calibration tubing/fittings, etc. 4. Reattempt the span
calibration before replacing the sensor.
4.10.3 Finishing Successful Calibration 1. Remove the calibration tube from pump inlet.
The calibration procedure adjusts the span value for any sensor that passes the calibration test. Sensors that fail calibration are left unchanged.
In the ALTAIR 5X IR, a combustible IR sensor calibration causes the catalytic combustible sensor to enter the LockAlarm condition.
· While a device with a % Vol methane IR sensor automatically recovers from LockAlarm, propane and butane IR devices do not; for these devices, the LockAlarm state is cleared by turning the device OFF, then ON while in a fresh air environment ( 4.2 Turning ON and Fresh Air Setup for details)
On the color display, each successfully calibrated sensor temporarily shows a symbol at its gas reading.
These symbols remain visible for a few moments and are then replaced by the present gas reading.
The monochrome display does not show symbols for individual gas readings.
Since residual gas may be present, the device may briefly go into an exposure alarm after the calibration sequence is completed.
2. Press the p button to reset the alarm as necessary.
GB
Altair 5X / Altair 5XiR
35
4 Operation
A symbol displays on the MEASURE page. This symbol appears on the: · color
display in the upper feature bar · monochrome display in the lower-right
corner.
The symbol displays for 24 hours after the calibration and then turns off.
If the horn alarm is turned OFF, the calibration symbol does not appear on the
color display.
Calibration with an Automated Test System The device can be calibrated using
the GALAXY or GALAXY GX2 Automated Test System – contact MSA for a list of
compatible gases and concentrations. Similar to the successful (manual)
calibration described in 4.10 Calibration, a symbol displays on the MEASURE
page after successful GALAXY or GALAXY GX2 calibration. This symbol appears on
the:
· color display in the upper feature bar · monochrome display in the lower-
right corner. The symbol displays for 24 hours after the calibration and then
turns off.
If the horn alarm is turned OFF, the calibration symbol does not appear on the
color display.
4.11 Time of Day Testing This feature permits the device to be automatically
calibrated on a user-defined interval. The most common use of this feature
allows the user to configure the ALTAIR 5X and the GALAXY GX2 System to
automatically calibrate a device prior to the start of work-shift. See the
GALAXY GX2 Operating Manual (“Automated Testing Features” section) for a
complete description of how to configure the GALAXY GX2 for this mode. On
ALTAIR 5X devices with firmware version 1.30 or higher, the following settings
must be configured using either MSA Link or the GALAXY GX2 Instrument Setup
page:
· Calibration Due must be enabled and a non-zero Calibration Interval must be
entered for automatic calibration testing
· Bump Due must be enabled and a non-zero Bump Interval must be entered for
automatic bump testing The firmware version is displayed on instrument
startup. Carefully follow all GALAXY GX2 set-up directions as described in the
GALAXY GX2 Operating Manual for proper setup.
4.12 Device Shutdown For device shutdown press and hold the q button.
Figure 12 Shutdown
The device displays a blinking “HOLD BUTTON FOR SHUTDOWN” and a progress bar
shows the user how much longer to hold the button to complete the shutdown.
36
Altair 5X / Altair 5XiR
GB
5 Maintenance
5 Maintenance
If irregularities occur during operation, use the displayed error codes and
messages to determine appropriate next steps.
WARNING! Repair or alteration of the device beyond the procedures described in
this manual or by anyone other than a person authorized by MSA, could cause
the unit to fail to perform properly. Use only genuine MSA replacement parts
when performing any maintenance procedures described in this manual.
Substitution or incorrect installation of components can seriously impair
performance of the unit, alter intrinsic safety characteristics or void agency
approvals. Failure to follow this warning can result in serious personal
injury or death.
Refer to EN 60079-29-2 (Guide for the selection, installation, use and maintenance of apparatus for the detection and measurement of combustible gases or oxygen) and EN 45544-4 (Guide for the selection, installation, use and maintenance of electrical apparatus used for the direct detection and direct concentration measurement of toxic gases and vapors).
5.1 Troubleshooting
Error State Alternating display
ADC ERROR MEM ERROR PROG ERROR RAM ERROR BT ERROR LOW BATTERY
Details
Analogue measurement error Memory error Program error RAM error Bluetooth
error Battery Warning repeats every 30 seconds
Recommended Action
Contact MSA Contact MSA Contact MSA Contact MSA Contact MSA Remove from
service as soon as possible and recharge or replace battery
(flashing) BATTERY ALARM
Battery is completely discharged
Device does not turn ON
Battery fully discharged
SENSOR MISSING NO SENSORS
Sensor damaged or missing No sensors are enabled
Sensor warning Sensor alarm
(flashing)
PUMP ERROR
Pump malfunction or flow path blockage
INVALID CONFIGURATION Sensor(s) installed in incorrect location.
Device is no longer sensing gas; Remove from service and recharge or replace
battery. Remove from service as soon as possible and recharge or replace
battery pack. Replace sensor Device must have at least one sensor enabled at
all times Sensor is near the end of its life
Sensor has reached the end of its life and cannot be calibrated. Replace
sensor and recalibrate.
Check flowpath for blockage. If error persists, remove from service. Install
sensors as shown in Figure 14.
GB
Altair 5X / Altair 5XiR
37
5 Maintenance
5.2 Verifying Pump Operation Users can check operation of the sampling system
any time during operation by blocking the sampling system to generate a pump
alarm. When the pump inlet, sample line or probe is blocked, the pump alarm
must activate. Once gas readings are displayed, plug the free end of the
sampling line or probe.
The pump motor shuts down and an alarm sounds. PUMP ERROR will flash on the
display. 1. Press the p button to reset the alarm and restart the pump. If the
alarm does not activate: Check the sample line and probe for leaks. Once leak
is fixed, recheck pump alarm by blocking the flow. 2. Press the p button to
reset the alarm and restart the pump.
WARNING! · Do not use the device, sample line, or probe unless the pump alarm
activates when the flow is blocked. · Lack of an alarm is an indication that a
sample may not be drawn to the sensors, which could cause inaccurate
readings. If a sample line or probe is installed and the pump alarm does not
activate, remove the line or probe and repeat the test. This will provide
information on where the blockage is located. Failure to follow these warnings
can result in serious personal injury or death.
Never let the end of the sampling line touch or go under any liquid surface.
If liquid is drawn into the device, readings will be inaccurate and device
could be damaged. We recommend the use of an MSA sample probe containing a
special membrane filter, permeable to gas but impermeable to water, to prevent
such an occurrence. During operation, a pump alarm may occur when the:
· Flow system is blocked · Pump is inoperative · Sample lines are attached or
removed. To Clear Pump Alarm 1. Correct any flow blockage. 2. Press the p
button.
The Pump will now restart.
5.3 Replacing the Battery
WARNING! Never replace the battery in a hazardous area. This could result in
an explosion. Failure to follow this warning can result in serious personal
injury or death.
38
Altair 5X / Altair 5XiR
GB
5 Maintenance
Figure 13 Battery Replacement 1 Captive screw
2 Battery pack
1. Unscrew the two captive screws on the rear of the device. 2. Pull the battery pack out of the device by gripping the sides and lifting it up and away from the device.
Figure 14 Battery Replacement
1 Battery holder
3. For alkaline battery packs (ALTAIR 5X only): a. Remove the battery holder
circuit board from the pack door. b. Replace the 3 cells, using only those
listed on the label. c. Be sure to observe proper polarity on the cells. d.
Place the battery holder circuit board back in the device and reinstall the
door.
4. When replacing the battery, be sure to accurately align screws and battery
with housing. 5. Screws should be tightened and torqued to 5.5 in lb (0,62
Nm).
5.4 Maintenance Procedure – Replacing or Adding a Sensor
Any factory-installed Series 20 sensor may be removed or replaced with a like
type. Any XCell sensor may be removed or replaced according to the positions
allowed in table below Figure 15 .
If the type of any sensor (including the IR sensor) is to be changed, the
device must be returned to an authorized service center.
NOTICE Before handling the PC board, ensure you are properly grounded;
otherwise, static charges from your body could damage the electronics. Such
damage is not covered by the warranty. Grounding straps and kits are available
from electronics suppliers.
GB
Altair 5X / Altair 5XiR
39
5 Maintenance
WARNING! Remove and reinstall sensors carefully, ensuring that the components
are not damaged; otherwise device intrinsic safety may be adversely affected,
wrong readings could occur, and persons relying on this product for their
safety could sustain serious personal injury or death.
While device case is open, do not touch any internal components with
metallic/conductive objects or tools. Damage to the device can occur.
Figure 15 Possible positions for sensor replacement
1
Combustible sensor
3
NH3, CL2, H2S, SO2 sensor (position with adapter) or Series 20 sensor
2
O2 sensor, TwoTox sensor
4
NH3, CL2, SO2, CO-HC sensor, Two-Tox sensor
SENSOR
XCell combustible sensor XCell O2 sensor XCell Two-Tox sensor XCell SO2, Cl2,
NH3 XCell CO-HC Series 20 sensor
OPERATIONAL ONLY IN POSITION 1
2 2 or 4 3 or 4 4 3
1. Verify that the device is turned OFF. 2. Remove the battery pack. 3.
Remove the two remaining case screws, and remove the case front. 4. Gently
remove the sensor to be replaced. 5. Carefully align the new sensor contact
pins with the sockets on the printed circuit board. 6. Press the new sensor
into place. 7. Note the position restrictions in the table above.
Adapter (P/N 10110183) is required for XCell usage in position 3.
If a sensor is removed and will not be replaced, be sure to install a sensor plug in its place in order to maintain correct device function.
The plug for XCell positions is P/N 10105650.The Series 20 plug is P/N 10088192. 8. Reinstall the sensor gasket in the case front. 9. Attach front case and tighten two case screws using 5.5 in-lbs of torque. 10. Attach the battery pack and tighten the two battery pack screws using 5.5 in-lbs of torque.
If a change in XCell Sensor configuration is detected during the device turn- ON process:
40
Altair 5X / Altair 5XiR
GB
5 Maintenance
· The “ACCEPT?” prompt appears on the display · The q button accepts the
sensor configuration · The p button rejects the sensor configuration; the
device is not operational. When an XCell sensor is replaced, the device
automatically enables the sensor after the change has been accepted. If a
Series 20 sensor is replaced, it must be manually enabled ( 4.5 Device Setup
SETTING SENSOR OPTIONS). If the oxygen sensor was replaced, see 4.2 Turning ON
and Fresh Air Setup regarding the oxygen reading display.
11. Allow sensors to stabilize at least 30 minutes before calibration. 12.
Calibrate device before use.
WARNING! Calibration is required after a sensor is installed; otherwise, the
device will not perform as expected and persons relying on this product for
their safety could sustain serious personal injury or death.
5.5 Replacing the Pump Filter 1. Turn OFF the device. 2. Unscrew the two
captive screws from the clear filter cover on the back of the device to access
the filter. 3. Carefully lift out the O-ring and the filter disk(s). 4. Use
both the paper-like filter and the fibrous dust filter (the thicker disk) as
supplied in Maintenance Kit if the device is NOT configured to use a reactive
toxic gas sensor (does not have a Cl2, ClO2, or NH3 sensor). NOTE: Use ONLY
the paper filter supplied in the Reactive Gas Maintenance Kit if the device IS
configured to use a reactive toxic gas sensor (Cl2, ClO2, or NH3). 5. Place
the new paper-like filter into the recess in the back of the device. If it is
to be used, place the fibrous dust filter into the clear filter cover.
WARNING! Use of the fibrous dust filter or the incorrect paper filter for the
measurement of reactive gases could cause erroneous readings. Failure to
follow this warning can result in serious personal injury or death.
6. Replace the O-ring in the recess. 7. Re-install the clear filter cover on
the back of the device.
5.6 Cleaning the Device Exterior Clean the exterior of the device regularly
using only a damp cloth. Do not use cleaning agents as many contain silicones
which will damage the combustible sensor.
5.7 Storage When not in use, store the device in a safe, dry place between 18
°C (65 °F) and 30 °C (86 °F). After storage, always recheck device calibration
before use. If not to be used in 30 days remove battery pack or connect it to
a charger.
5.8 Shipment Pack the device in its original shipping container with suitable
padding. If the original container is unavailable, an equivalent container may
be substituted.
GB
Altair 5X / Altair 5XiR
41
6 Technical Specifications
6 Technical Specifications
Weight Weight (with IR Sensor) Dimensions (cm)
Dimensions (cm) (with IR Sensor) Alarms Volume of audible alarm Displays
Battery types
0.45 kg (1 lb.) – device with battery and clip (ALTAIR 5X unit) 0.52 kg (1.15 lb.) 17 x 8.87 x 4.55 (6.69″ H x 3.49″ W x 1.79″ D) pumped, without belt clip (ALTAIR 5X unit) 17 x 8.94 x 4.88 (6.68″ H x 3.52″ W x 1.92″ D) LEDs, audible alarm, vibrating alarm 95 dB typical Monochrome / Color Rechargeable Li ION battery
Replaceable AA alkaline (ALTAIR 5X only)
Charging time
For battery types see the Altair 5X / XIR User Instruction Addendum (P/N
10223097).
6 hours The maximum safe area charging voltage
Normal Temperature range Extended Temperature range
Um = 6.7 Volts DC -10 °C to 40 °C (14 °F to 104 °F) -20 °C to 50 °C (-4 °F to 122 °F) Monochrome display
-10 °C to 50 °C (14 °F to 122 °F) color display
Short Term Operations (15 minutes) temperature range
Humidity range
-20 °C to 40 °C (-4 °F to 104 °F) for devices with ClO2 sensors -40 °C to 50 °C (-40 °F to 122 °F) for devices without PID sensors
15 – 90 % relative humidity, non-condensing,
Atmospheric pressure range Ingress protection Measuring methods
5 – 95 % RH intermittent 80 kPa to 120 kPa (11.6 to 17.4 PSIA) IP 65 Combustible gases – Catalytic or Infrared sensor
Warranty
Oxygen and Toxic gases – Electrochemical or Infrared sensor See 2.4 Warranty
Measuring Range CIO2 Cl2 CO CO – HC Combustible
H2S H2S – LC HCN
0-1.00 ppm 0-10 ppm 0-2000 ppm 0-10,000 ppm 0-100 % LEL 0-5.00 % CH4 0-200 ppm
0-30 ppm
NH3 NO NO2 (S20) NO2 (XCell) O2
PH3 PID SO2
0-100 ppm 0-200 ppm 0-20.0 ppm 0-50.0 ppm 0-30 % Vol.
0-5.00 ppm 0-2000 ppm 0-20.0 ppm
42
Altair 5X / Altair 5XiR
GB
6.1 Factory-set Alarm Thresholds and Setpoints
6 Technical Specifications
Check the monitor or calibration certificate for exact alarm levels as they vary depending on national or corporate regulations.
Sensor
LOW alarm HIGH alarm SETPOINT min SETPOINT max STEL
TWA
CL2 ClO2 CO
0.5 ppm 0.1 ppm 25 ppm
1.0 ppm 0.3 ppm 100 ppm
0.3 ppm 0.1 ppm 10 ppm
7.5 ppm 0.9 ppm 1700 ppm
1.0 ppm 0.3 ppm 100 ppm
0.5 ppm 0.1 ppm 25 ppm
CO-HC COMB
25 ppm 10 % LEL
100 ppm 20 % LEL
10 ppm 5 % LEL
8500 ppm 60 % LEL
100 ppm –1
25 ppm –1
H2S H2S-LC HCN
10 ppm 5 ppm 4.5 ppm
15 ppm 10 ppm 10.0 ppm
5 ppm 1 ppm 2.0 ppm
175 ppm 70 ppm 20.0 ppm
15 ppm 10 ppm 10 ppm
10 ppm 1 ppm 4.5 ppm
HCN
4.5 ppm
10.0 ppm
2.0 ppm
20.0 ppm
10.0 ppm
4.5 ppm
NH3
25 ppm
50 ppm
10 ppm
75 ppm
35 ppm
25 ppm
NO
25 ppm
75 ppm
15 ppm
100 ppm
25 ppm
25 ppm
NO2 (S 20) NO2 (XCell) O2
PH3 SO2 IR Butane (25 % Vol)
2.0 ppm 2.5 ppm 19.5 % 0.3 ppm 2.0 ppm 8 % Vol
5.0 ppm 5.0 ppm 23.0 % 1.0 ppm 5.0 ppm 15 % Vol
1.0 ppm 1.0 ppm 5.0 % 0.3 ppm 2.0 ppm 5 % Vol
17.5 ppm 47.5 ppm 24.0 % 3.75 ppm 17.5 ppm 25 % Vol
5.0 ppm 5.0 ppm –1
1.0 ppm 5.0 ppm –1
2.0 ppm 2.5 ppm –1
0.3 ppm 2.0 ppm –1
IR CO2 (10 % Vol)
0.5 % Vol
1.5 % Vol
0,2 % Vol
8 % Vol
0.5 % Vol
1.5 % Vol
IR Methane
– -2
– -2
– -2
– -2
–1
–1
(100 % Vol)
IR Propane
15% LEL
29% LEL
9% LEL
100% LEL
–1
–1
(100 % LEL)
IR Propane
– -2
– -2
– -2
– -2
–1
–1
(100 % Vol)
1STEL and TWA not applicable for combustible and oxygen gases. 2No alarm thresholds are possible for the 0-100 % Vol Methane and Propane IR sensors. In environments with >100 % LEL combustible gas present, devices with a catalytic combustible LEL sensor will be in a latching over-range alarm, and the 100 % Vol IR sensors will display the % Vol gas reading.
6.2 Performance Specifications
Sensor
Range
Combustible Gas 0 to 100 % LEL or
0 to 5 % CH4
Resolution 1 % LEL or 0.05 Vol % CH4
Reproducibility Normal temp. range:
<50 % LEL: 3 % LEL
50-100 % LEL: 5 % LEL <2.5 % CH4: 0.15 % CH4 2.5-5.00 % CH4: 0.25 % CH4
Response time
t(90)< 15 sec (Pentane) (normal temp.)
t(90)< 10 sec (Methane) (normal temp.)
GB
Altair 5X / Altair 5XiR
43
6 Technical Specifications
Sensor
Range
Oxygen
0 30% O2*
Carbon -Monoxide 0-2000 ppm CO
Hydrogen Sulfide 0-200 ppm H2S
Resolution
Reproducibility
Response time
Extended temp. range: <50 % LEL: 5 % LEL 50-100% LEL: 8% LEL
<2.5 % CH4: 0.25 % CH4
0.1% O2
2.5-5.00 % CH4: 0.40 % CH4 0.7 % O2 for 0 30 % O2
1 ppm CO
normal temperature range: ±5 ppm CO or 10 % of reading, whichever is greater
extended temperature range: ±10 ppm CO or 20 % of reading, whichever is greater
1 ppm H2S, for 3 to normal temperature range:
200 ppm H2S
±2 ppm H2S or 10 % of reading,
whichever is greater
extended temperature range:
±20 ppm H2S or 20 % of reading, whichever is greater
t(90)< 10 sec (normal temp.) t(90)< 15 sec (normal temp.)
t(90)< 15 sec (normal temp.)
6.2.1 IR Sensors Sensor
Resolution
Carbon Dioxide Methane Propane Propane Butane
0-10 % Vol
0.01 % Vol
0-100 % Vol 0-100 % Vol 0-100%LEL 0-25 % Vol
1 % Vol 1 % Vol 1% LEL 0.1 % Vol
Response time at 20 °C
t90 35 s
Reproducibility of the zero point
± 0.01 % Vol
34 s 36 s 32 s 35 s
± 5 % Vol ± 3 % Vol ± 3% LEL ± 0.5 % Vol
Reproducibility of the measured value1)
±4 %
±10 % ±8 % ±8% ±4 %
6.2.2 Additional Toxic Sensors
Sensor
Normal temperature
range: Cl2 Chlorine
Range (ppm)
Resolution (ppm)
Extended temp. range:
0 – 10
0.05
ClO2 Chlorine
0-1
0.01
dioxide
CO-HC Carbon – 0 – 10000
5
Monoxide
Reproducibility
±0.2 ppm or 10 % of reading, whichever is greater ±0.1 ppm or 10 % of reading,
whichever is greater ±5 ppm or 10 % of reading, whichever is
Nominal response*
±0.5 ppm or 20 % of reading, whichever is greater
±0.2 ppm or 20 % of reading, whichever is greater
±10 ppm or 20 % of reading, whichever is
t(90)< 30 s t(90)< 2 min t(90)< 15 s
44
Altair 5X / Altair 5XiR
GB
6 Technical Specifications
Sensor
Range (ppm)
Resolution (ppm)
Reproducibility
Normal temperature
range:
Extended temp. range:
greater
H2S-LC Hydrogen 0 – 100
0.1
Sulfide
±0.2 ppm or 10 % of reading, whichever is greater
HCN Hydrogen 0 – 30
0.5
cyanide
±1 ppm or 10 % of reading, whichever is greater
NH3 Ammonia
0 – 100
1
±2 ppm or 10 % of reading, whichever is greater
NO2 Nitrogen
0 – 20
0.1
dioxide (S 20)
±2 ppm or 10 % of reading, whichever is greater
NO2 Nitrogen
0 – 50
0.1
dioxide (XCell)
±1 ppm or 10 % of reading, whichever is greater
NO Nitric oxide 0 – 200
1
±5 ppm or 10 % of reading, whichever is greater
PH3 Phosphine 0 – 5
0.05
±0.2 ppm or 10 % of reading, whichever is greater
SO2 Sulfur
0 – 20
0.1
dioxide
±2 ppm or 10 % of reading, whichever is greater
- Response time is for normal temperature range with sensor in position #3
Nominal response*
greater
±0.5 ppm or 20 % of reading, whichever is greater
±2 ppm or 20 % of reading, whichever is greater
±5 ppm or 20 % of reading, whichever is greater
±3 ppm or 20 % of reading, whichever is greater
±2 ppm or 20 % of reading, whichever is greater
±10 ppm or 20 % of reading, whichever is greater
±0.25 ppm or 20 % of reading, whichever is greater
±3 ppm or 20 % of reading, whichever is greater
t(90)< 15 s t(90)< 30 s t(90)< 40 s t(90)< 40 s t(90)< 15 s t(90)< 40 s t(90)< 30 s t(90)< 20 s
6.3 Calibration Specifications
Sensor
Zero Gas
Zero Cal Value**
COMB Pentane COMB Methane (0 – 5 % Vol) COMB Methane (4.4 % Vol) COMB Propane (2.1 % Vol) COMB Propane (1.7 % Vol) COMB Butane (1.4 % Vol) COMB Methane (5 % Vol) COMB Hydrogen (4,0 % Vol) O2
Fresh Air Fresh Air Fresh Air Fresh Air Fresh Air Fresh Air Fresh Air Fresh Air Fresh Air
0 0 0 0 0 0 0 0 20.8 %
Span Cal Gas
Span Cal
Value
1.45 % Vol Methane 2.5 % Vol Methane 1.45 % Vol Methane 1.45 % Vol Methane 1.45 % Vol Methane 1.45 % Vol Methane 1.45 % Vol Methane 1.45 % Vol Methane 15 % O2
58 % LEL 2,5 % 33 % LEL 46 % LEL 37 % LEL 46 % LEL 29 % LEL 33 % LEL 15 %
Time (min) 1 1 1 1 1 1 1 1 1
GB
Altair 5X / Altair 5XiR
45
6 Technical Specifications
Sensor
Zero Gas
Zero Cal Value**
Span Cal Gas
Span Cal
Value
Time (min)
CO
Fresh Air
0
60 ppm CO
60 ppm
1
H2S
Fresh Air
0
20 ppm H2S
20 ppm
1
SO2
Fresh Air
0
10 ppm SO2
10 ppm
1
Cl2
Fresh Air
0
10 ppm Cl2
10 ppm
2
NO
Fresh Air
0
50 ppm NO
50 ppm
4
NO2 NH3 PH3 HCN
Fresh Air
0
Fresh Air
0
Fresh Air
0
Fresh Air
0
10 ppm NO2 25 ppm NH3 0.5 ppm PH3 10 ppm HCN
10 ppm
2
25 ppm
2
0.5 ppm
1
10 ppm
4
*ClO2 IR CO2 (10 % Vol) IR Butane (25 % Vol)
Fresh Air Fresh Air Fresh Air
0 0.03 % 0
2 ppm Cl2 2.5 % CO2 8 % Vol
0.8 ppm
6
2.5 %
2
8 % Vol
2
IR Propane (100 % Vol)
Fresh Air
0
50 % Vol Propane
50 % Vol
2
IR Propane (100 % LEL)
Fresh Air
0
0.6 % Vol Propane
29% LEL
2
IR Methane (100 % Vol)
Fresh Air
0
50 % Vol Methane
50 % Vol
2
Span values can be changed if using different gas cylinders than those listed. Changes can be made using MSA Link software and through calibration cylinder setup.
*For most accurate results, calibration with ClO2 is recommended. **Zero cal time is one minute if a catalytic combustible sensor is installed – 30 seconds if not.
LEL values, if not listed here, are according to EN 60079-20-1. Local
regulations may differ.
6.4 Combustible Gas Cross Reference Factors for General-Purpose Calibration
Using Calibration Cylinder (P/N 10053022)
See the applicable addendum on the Product CD.
46
Altair 5X / Altair 5XiR
GB
7 XCell Sensor Patents
SENSOR
Combustible Oxygen Carbon Monoxide / Hydrogen Sulfide Ammonia Chlorine Sulphur
Dioxide
PART NUMBER 10106722 10106729 10106725
US8826721 US8790501 US8790501, US8702935
10106726 10106728 10106727
US8790501, US8623189 US8790501, US8623189 US8790501, US8623189
7 XCell Sensor Patents
GB
Altair 5X / Altair 5XiR
47
8 Ordering Information
8 Ordering Information
8.1 US Gas Cylinder Parts List
Gases
Gases Mix
1
10 % CO2 in N2
1
8 % Butane in N2 (6L)
1
50 % Vol Methane in N2 (103L)
1
100 % Vol Methane
1
10 ppm NO2 in Air
1
10 ppm SO2 in Air
1
25 ppm NH3 in N2
1
10 ppm Cl2 in N2
1
2 ppm Cl2 in N2
1
10 ppm HCN in N2
1
0.5 ppm PH3 in N2
3
1.45 % CH4, 15.0 % O2, 20 ppm H2S
3
2.50 % CH4, 15.0 % O2, 20 ppm H2S
3
1.45 % CH4, 15.0 % O2, 60 ppm CO
3
2.50 % CH4, 15.0 % O2, 60 ppm CO
4
1.45 % CH4, 15.0 % O2, 60 ppm CO,
10 ppm NO2
4
1.45 % CH4, 15.0 % O2, 60 ppm CO,
20 ppm H2S
4
2.50 % CH4, 15.0 % O2, 60 ppm CO,
20 ppm H2S
4
2.50 % CH4, 15.0 % O2, 60 ppm CO,
10 ppm NO2
5
1.45 % CH4, 15.0 % O2, 60 ppm CO,
20 ppm H2S, 2.5% CO2
5
1.45 % CH4, 15.0 % O2, 60 ppm CO,
20 ppm H2S, 10 ppm SO2
MSA P/N ECONO-CAL (34 L)
10075802
711068 711070 711078 711066 711082 711072 711088 10048790 10048888 10048789
10049056
10058036
RP (58 L)
10081603
10075804
711014 808977 808978 814866 806740 10028080 809351 710533 10048788 10048889
478191 (100L) 813718 (100L) 10058034
Recommended CAL Gas for:
25 % Vol Butane IR 100 % Vol Methane IR
NO2 sensor SO2 sensor NH3 sensor Cl2 sensor ClO2 sensor HCN sensor PH3 sensor
10048280
10045035
10048981
10048890
10058172
10103262 10 % CO2 IR
10098855
10117738 SO2 sensor
48
Altair 5X / Altair 5XiR
GB
8 Ordering Information
8.2 Outside US
Description Gas Cylinder 34L, 60 ppm CO Cylinder 34L, 40 ppm H2S Cylinder 34L, 25 ppm NH3 Cylinder 34L, 10 ppm Cl2 Cylinder 34L, 10 ppm SO2 Cylinder 34L, 10 ppm NO2 Cylinder 34L, 0.5 ppm PH3 Cylinder 34L, 2 ppm Cl2 (To calibrate ClO2 sensor) Cylinder 34L, 10 ppm HCN Calibration Cylinder 58L (1.45 % CH4, 15.0 % O2, 60 ppm CO, 20 ppm H2S) Calibration Cylinder 58L (1.45 % CH4, 15.0 % O2, 60 ppm CO, 20 ppm H2S) (US) Cylinder 34L, 50 ppm NO Cylinder 58L (0,4 % Propane, 15 % O2, 60 ppm CO, 20 ppm H2S) Cylinder 58L (1.45 % CH4, 2.5 % CO2,, 60 ppm CO, 20 ppm H2S, 15 % O2) Cylinder 34L (1.45 % CH4, 15 % O2, 60 ppm CO, 20 ppm H2,S, 10 ppm SO2,) Cylinder 58L (1.45 % CH4, 15 % O2, 60 ppm CO, 20 ppm H2,S, 10 ppm SO2,) For IR Sensors Cylinder 34L, 2.5 % Vol CO2 Cylinder 34L, 8 % Vol Butane Cylinder 34L, 20 % Vol Methane Cylinder 34L, 50 % Vol Methane
Part Number
10073231 10011727 10079807 10011939 10079806 10029521 10029522 711082 711072
10053022 10045035 10126429 10086549 10102853 10122425 10122426
10069618 10078012 10022595 10029500
8.3 Accessories
Description Universal Demand Regulator kit MSA Link USB dongle MSA Link Datalogging Software Shoulder Strap Retractable Line with Belt Clip Holster, leather Sampling Probe, flexible 30 cm, conductive Sampling Line, 1,5 m, conductive Sampling Line, 3 m, conductive Sampling Line, 5 m, conductive Sampling System w/floating probe, 5 m, PU conductive Sampling Line, 20 m, conductive Sampling Line, 30 m, conductive
Part Number 10034391 10082834 10088099 474555 10050976 10099648 10103191 10103188 10103189 10103190 10082307 10159430 10159431
GB
Altair 5X / Altair 5XiR
49
8 Ordering Information
Description Probe, 1 ft. straight PEEK Probe, 3 ft. straight PEEK Polyurethane Sample Line, 10 ft. Polyurethane Sample Line, 25 ft. Polyurethane Sample Line, 3 ft. Coiled (Cl2, ClO2, NH3) 5 ft.PU Coiled Sample line & probe, kit (Cl2, ClO2, NH3) 5 ft. PU Sample line & probe, kit (Cl2, ClO2, NH3) 10 ft. Teflon Sample line & probe, kit Replacement Filters for probe, 10 pack Charger only (North America) Charger only (Global version) Charging Cradle with Barrier- (North America) Charging Cradle – (North America) Charging Cradle (Europe) Charging Cradle (Australia) Vehicle Charger Cradle Cradle Only – (no charger) ALTAIR 5/5X Multi-Unit Charger, 4 Unit (North American) ALTAIR 5/5X Multi-Unit Charger, 4 Unit (Europe) ALTAIR 5/5X Multi-Unit Charger, 4 Unit (UK) ALTAIR 5/5X Multi-Unit Charger, 4 Unit (Australia) ALTAIR 5/5X Multi-Unit Charger, 4 Unit No Power Cord Carrying Case
Part Number
10042622 10040665 10040664 10040667 10105210 10105251 10105839 801582 10087913
10092936 10093055 10093054 10093057 10093056 10099397 10093053 10127427
10127428 10127429 10127430 10128704 10152079
Please note that not all accessories are available in every local market. Check availability with the local MSA representative .
50
Altair 5X / Altair 5XiR
GB
8.4 Replacement Parts
1 3 5 9
6 8
7
8 Ordering Information
2 4
No.
Description
Part Number
1
Case assembly, upper, with label
10114853
Case assembly, upper, (phosphorescent), with label
10114854
2
Battery pack, rechargeable, North America, ALTAIR 5X
10114835
Battery pack, rechargeable, EU/Aus, ALTAIR 5X
10114836
Battery pack, rechargeable, North America, ALTAIR 5X IR
10114839
Battery pack, rechargeable, EU/Aus, ALTAIR 5X IR
10114851
Battery Pack, rechargeable, North America ALTAIR 5X IR (phosphorescent)
10114840
Battery Pack, rechargeable, European/Australian ALTAIR 5X IR (phosphorescent) 10114852
3
Kit, belt clip replacement, (ALTAIR 5X rechargeable)
10094830
Kit, maintenance (includes filters, o-ring, screws)
10114949
Kit, maintenance, reactive gas (Cl2, ClO2, NH3) (includes filters, o-ring, screws)
10114950
GB
Altair 5X / Altair 5XiR
51
8 Ordering Information
No.
Description
Part Number
4
Filter cover assembly
10083591
5
Display assembly, monochrome
10111389
Display assembly, color
10148366
6
Sensor bracket assembly with pump, ALTAIR 5X (includes vibrator motor)
10114804*
Sensor bracket assembly with pump, ALTAIR 5X IR (includes vibrator motor)
10114805*
7
Kit, pump cap replacement
10114855
8
Sensor, HCN (Series 20)
10106375
XCell Sensor, Cl2 Sensor, ClO2 (Series 20) XCell Sensor, SO2 Sensor, NO2 (Series 20) XCell Sensor, NH3 Sensor, PH3 (Series 20) XCell Sensor, COMB
10106728 10080222 10106727 10080224 10106726 10116638 10106722
XCell Sensor, O2 XCell Sensor, CO
10106729 10106724
XCell Sensor, H2S XCell Sensor, CO-H2S, Two-Tox XCell Sensor, CO/ NO2 XCell Sensor, CO-HC
10106723 10106725 10121217 10121216
XCell Sensor, H2S-LC/CO XCell Sensor, CO-H2 Res/H2S Sensor, NO (Series 20)
10121213 10121214 10114750
XCell Sensor plug
10105650
20 mm sensor plug
10088192
9
XCell adapter socket
10110183
*In order to replace this item on a detector manufactured prior to July 2022, additional service to the main PCBA will be required by an MSA Service Center. Contact MSA for additional help. Detectors manufactured July 2022 and afterward may use this item with no further action required.
**The date is located under the serial number in the XYY format. X = month, A through L. (A = January, B = February, … L = December) YY = final 2 digits of year. For example, G22 = July 2022
Description
Part Number
IR sensors – Must be repaired or replaced at a Certified Repair Center
IR sensor HC 0-25 % Vol Butane
10145739-SP
IR sensor HC 0-100 % Vol Methane
10145752-SP
IR sensor HC 0-100 % Vol Propane
10145740-SP
IR sensor 0-10 % Vol CO2 IR sensor 0-100% LEL Propane
10145738-SP 10145751-SP
52
Altair 5X / Altair 5XiR
GB
9 Flow Charts
9.1 Basic Operation
TURN-ON OR BATTERY ATTACH
INSTRUMENT INFORMATION
PUMP TEST
SENSOR DISCOVERY
BLUETOOTH INFORMATION
SENSOR AND CALIBRATION INFORMATION
FRESH AIR SETUP ?
BUMP TEST /
INFORMATIONAL PAGES
Press
MAIN MEASURE
PAGE
Hold
For 5 seconds
CALIBRATE
InstantAlertTM *
Hold For 5
seconds
Hold For 5 seconds
Hold and
Simultaneously for 5 seconds
SETUP
TURN OFF
- IF ENABLED
9 Flow Charts
GB
Altair 5X / Altair 5XiR
53
9 Flow Charts
9.2 Bump Test/ Informational Pages
BUMP TEST / INFORMATIONAL
PAGES
PERFORM SESOR
BUMP TEST
Press
BUMP TEST?
Press
Press
Press
PEAK *
Press
Press
Press
MIN
Press
Press
Press
STEL *
Press
Press
MAIN MEASURE
PAGE
Press
TWA *
Press
MAIN MEASURE
PAGE
RESET PEAK
RESET MIN
RESET STEL
RESET TWA
- IF ENABLED (NOT VALID FOR ALL SENSORS)
54
Altair 5X / Altair 5XiR
GB
Continued
Press
DATE
Press
Press
LAST CAL DATE *
Press
Press
CAL DUE DATE *
Press
Press
DISCOVERY MODE*
Press
Press
MOTION ALERT STATUS * Press
Press
TOGGLE ON /OFF
Press
WIRELESS **
Press
TOGGLE ON /OFF
Press
MAIN MEASURE
PAGE
- IF ENABLED
** IF WIRELESS IS INSTALLED
9 Flow Charts
GB
Altair 5X / Altair 5XiR
55
9 Flow Charts 9.3 Calibrations
CALIBRATE
PASSWORD ?
(IF REQUIRED)
PASSWORD FAIL
ZERO CALIBRATION
Press
ZERO CAL ?
Press
Press
SPAN CALIBRATION
Press
SPAN CAL ?
(REPEATS AS NECESSARY)
Press
MAIN MEASURE
PAGE
PERFORM FRESH AIR
SETUP
56
Altair 5X / Altair 5XiR
GB
9.4 Setup
SETUP
Hold and
Simultaneously for 5 seconds
PASSWORD FAIL
ENTER PASSWORD
CALIBRATION OPTIONS ?
Press
Press
ALARM OPTIONS ?
Press
Press
INSTRUMENT OPTIONS ?
Press
Press
EXIT ?
Press
MAIN MEASURE
PAGE
CALIBRATION OPTIONS
(See 9.5)
ALARM OPTIONS
(See 9.6)
INSTRUMENT OPTIONS
(See 9.7)
9 Flow Charts
GB
Altair 5X / Altair 5XiR
57
9 Flow Charts
9.5 Calibration Options
CALIBRATION OPTIONS
CYLINDER SETUP
Press
Press
CALIBRATION DUE OPTIONS
Press
Press
SHOW LAST CAL DATE AT STARTUP
Press
Press
PASSWORD PROTECTED CALIBRATION
Press
Press
BACK TO SETUP ?
Press
BACK TO MAIN MENU
SET SPAN GAS FOR
EACH SENSOR
ON / OFF, SET # OF
DAYS
TOGGLE ON / OFF
TOGGLE ON / OFF
58
Altair 5X / Altair 5XiR
GB
9.6 Alarm Options
ALARM OPTIONS
VIBRATE
Press
Press
HORN
Press
Press
ALARM LEDS
Press
Press
USER ACCESS TO
MOTION ALERT
Press
Press
SENSOR ALARMS ?
Press
Press
BACK TO SETUP ?
Press
BACK TO MAIN MENU
TOGGLE ON/OFF
TOGGLE ON/OFF
TOGGLE ON/OFF
TOGGLE ON/OFF
SENSOR ALARM SETUP
9 Flow Charts
GB
Altair 5X / Altair 5XiR
59
9 Flow Charts
9.7 Sensor Alarm Setup
SENSOR ALARM SETUP
SETUP
LOW
Press
ALARMS ? Press
Press
SETUP
HIGH
Press
ALARMS ? Press
Press
SETUP
STEL
Press
ALARMS ? Press
Press
SETUP
TWA
Press
ALARMS ? Press
Press
MODIFY LOW ALARMS FOR
ALL SENSORS
MODIFY HIGH ALARMS FOR
ALL SENSORS
MODIFY STEL
ALARMS
MODIFY TWA
ALARMS
RETURN TO ALARM OPTIONS
NOTE: STEL AND TWA ARE NOT VALID FOR ALL SENSORS
60
Altair 5X / Altair 5XiR
GB
9.8 Instrument Options
INSTRUMENT OPTIONS
SENSOR SETUP
Press
Press
LANGUAGE
Press
Press
TIME / DATE
Press
DATALOG INTERVAL
Press
Press Press
STEALTH MODE
Press
Press
SENSOR SETUP
(See 9.9)
SELECT
SET HH:MM MM:DD:YY
SET MM:SS
TOGGLE ON/OFF
9 Flow Charts
GB
Altair 5X / Altair 5XiR
61
9 Flow Charts Continued
OPERATING BEEP
Press
Press
TOGGLE ON/OFF
CONTRAST (MONO ONLY)
Press
Press
ADJUST
BACKLIGHT
Press
ON / OFF
TIMEOUT
Press
PERIOD
BLUETOOTH
Press
Press
ON / OFF
BACK TO SETUP ?
Press
BACK TO MAIN MENU
62
Altair 5X / Altair 5XiR
GB
9.9 Sensor Setup
SENSOR SETUP
COMB SENSOR **
Press
Press
SHOW STATUS
Press
Press
TOGGLE SENSOR ON/OFF
TOXIC SENSOR **
(REPEATS AS Press NECESSARY)
SHOW STATUS
Press
Press
Press
TOGGLE SENSOR ON/OFF
IR SENSOR **
Press
Press
SHOW STATUS
Press
Press
TOGGLE SENSOR ON/OFF
RETURN TO INSTRUMENT
OPTIONS
** IF INSTALLED
9 Flow Charts
GB
Altair 5X / Altair 5XiR
63
10 Changeable Feature Summary
10 Changeable Feature Summary
Feature
Initial Setting
Setup Password Vibrating Alarm Horn Alarm LED Alarm Safe LED (green) Operating Beep (alarm LEDs & horn) Stealth MotionAlert – Access MotionAlert
672 ON ON ON ON OFF OFF Allowed OFF
Sensor Alarm Levels
Enable / Disable High & Low Alarms Enabled
Turn Sensors ON / OFF
ON
Show Peak Show STEL, TWA Cal Cylinder Setup Show Last Cal Date Show Cal Due Cal Password Required Backlight Backlight Duration Display Contrast Language Date, Time Datalog Interval Custom Logo Screen Device S/N Company Name Dept./User Name VOC RF ON/OFF Bump Due ON/OFF Bump Interval Cal Due Interval
ON ON
ON ON OFF Enabled 10 s Factory-set User-set User-set 3 min Factory-set
Factory-set Blank Blank ON OFF 1 30
Setup Path to Change this Setting
ALARM OPTIONS ALARM OPTIONS ALARM OPTIONS INSTRUMENT OPTIONS INSTRUMENT
OPTIONS ALARM OPTIONS Use q button from MEASURE page ALARM OPTIONS / SENSOR
ALARM SETUP INSTRUMENT OPTIONS / SENSOR SETUP CAL OPTIONS CAL OPTIONS CAL
OPTIONS CAL OPTIONS INSTRUMENT OPTIONS INSTRUMENT OPTIONS INSTRUMENT OPTIONS
INSTRUMENT OPTIONS INSTRUMENT OPTIONS Certified service center INSTRUMENT
OPTIONS INSTRUMENT OPTIONS
Change with MSA link? Yes Yes Yes Yes Yes Yes No No No
Yes
Yes Yes
Yes Yes Yes No Yes No No Yes No No Yes Yes Yes No Yes Yes Yes Yes Yes Yes
Change via Bluetooth?
No Yes Yes Yes No No No Yes Yes
Yes
Yes No
No No Yes No No No Yes Yes No Yes Yes No No No Yes Yes No Yes Yes Yes
64
Altair 5X / Altair 5XiR
GB
References
- MSA Safety | Global
- MSA Safety | Global
- MSA Safety | Global
- Cookie Consent - Mine Safety Appliances DAM
- Trademark Information | MSA Safety | United States
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>