HACH LCE009 Assay Microbe
Instructions for Use
Section 1 Trademark, Patents and Copyright Statements
Cepheid®, the Cepheid logo, GeneXpert®, and Xpert® are trademarks of Cepheid,
registered in the U.S. and other countries. All other trademarks are the
property of their respective owners.
THE PURCHASE OF THIS PRODUCT CONVEYS TO THE BUYER THE NON-TRANSFERABLE RIGHT TO USE IT IN ACCORDANCE WITH THESE INSTRUCTIONS FOR USE. NO OTHER RIGHTS ARE CONVEYED EXPRESSLY, BY IMPLICATION OR BY ESTOPPEL. FURTHERMORE, NO RIGHTS FOR RESALE ARE CONFERRED WITH THE PURCHASE OF THIS PRODUCT.
- 2022 Cepheid.
- 2022 Hach.
Section 2 Xpress SARS-CoV-2 plus
Not for Use in Diagnostic Procedures.
Section 3 Proprietary Name
Xpress SARS-CoV-2 plus
Section 4 Common or Usual Name
Xpress SARS-CoV-2 plus
Section 5 Intended Use
The Xpress SARS-CoV-2 plus test is a rapid, real-time RT-PCR test intended for
the qualitative detection of nucleic acid from the SARS-CoV-2 in sample water
and wastewater samples for water-based epidemiological applications. It is not
intended for human or animal testing such as nasopharyngeal, oropharyngeal,
nasal, or mid-turbinate swab and/or nasal wash/ aspirate samples collected
from individuals suspected of COVID-19 by their healthcare provider). Results
are for the detection of SARS-CoV-2 RNA. The SARS-CoV-2 RNA is generally
detectable in sample wastewater samples for communities that have multiple
active positive cases in which the waste is combined and composite samples can
be taken. Negative results do not preclude SARS-CoV-2 in the water samples and
should not be used as the sole basis for any public health decisions. Negative
results should be combined with, case counts epidemiological information, and
other observations when making public health decisions.
Section 6 Summary and Explanation
An outbreak of respiratory illness of unknown etiology in Wuhan City, Hubei
Province, China was initially reported to the World Health Organization (WHO)
on December 31, 2019.1 Chinese authorities identified a novel coronavirus
(2019-nCoV), which has resulted in thousands of confirmed human infections in
multiple provinces throughout China and exported cases in several Southeast
Asian countries and more recently the United States. Cases of severe illness
and some deaths have been reported. The International Committee for Taxonomy
of Viruses (ICTV) renamed the virus SARS-CoV-2.2 Wastewater surveillance, the
measurement of pathogen levels in wastewater, is used to evaluate community-
level infection trends, augment traditional surveillance that leverages
clinical tests and services (e.g., case reporting), and monitor public health
interventions. Approximately 40% of persons infected with SARS-CoV-2, the
virus that causes COVID-19, shed virus RNA in their stool; therefore,
community-level trends in SARS-CoV-2 infections, both symptomatic and
asymptomatic can be tracked through wastewater testing. 3 In nearly 80% of
U.S. households, fecal waste is transported from homes to wastewater treatment
plants within hours. Wastewater represents a pooled community stool sample
that can provide information on infection trends in the community served by
the sewer network (sewer shed), which can range in size from fewer than 2,000
to >3 million persons.
Wastewater surveillance data provide information about symptomatic and asymptomatic SARS-CoV-2 infections. The accuracy of this surveillance approach is not influenced by healthcare access or clinical testing capacity. In addition, SARS-CoV-2 infection trends in a community can be detected in wastewater before other COVID-19 surveillance metrics, such as case reports and hospital admissions. To build sustainable national wastewater surveillance capacity, CDC focused NWSS development in four areas: 1) offering technical assistance to implementing jurisdictions; 2) creating a data portal for centralized data submission and standardized data analysis and visualization; 3) coordinating communities of practice to share best practices among health departments, public health laboratories, and utilities; and 4) building epidemiology and laboratory capacity for wastewater surveillance at health departments.3 The Xpress SARS-CoV-2 plus test is a molecular test that aids in the detection of SARS-CoV-2 in wastewater samples and is based on widely used nucleic acid amplification technology. The Xpress SARS-CoV-2 plus test contains primers probes and internal controls used in RT-PCR for the quantitative detection of SARS-CoV-2 RNA in wastewater samples.
Section 7 Principle of the Procedure
The Xpress SARS-CoV-2 plus test provides quantitative results using an
automated extraction and analysis of the SARS-CoV-2 virus. The Xpert SARS-
CoV-2 test is performed on GeneXpert Instrument Systems.The GeneXpert
Instrument Systems automate and integrate sample preparation, nucleic acid
extraction and amplification, and detection of the target sequences in simple
or complex samples using real-time PCR assays. The systems consist of an
instrument, computer, and preloaded software for running tests and viewing the
results. The systems require the use of single-use disposable cartridges that
hold the RT-PCR reagents and host the RT-PCR process. Because the cartridges
are self-contained, cross-contamination between samples is minimized. For a
full description of the systems, see the GeneXpert Dx System Operator Manual.
The Xpress SARS-CoV-2 plus test includes reagents for the detection of RNA
from SARS-CoV-2 in wastewater samples. A Sample Processing Control (SPC) and a
Probe Check Control (PCC) are also included in the cartridge utilized by the
GeneXpert instrument. The SPC is present to control for adequate processing of
the sample and to monitor for the presence of potential inhibitor(s) in the
RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions
(temperature and time) are appropriate for the amplification reaction and that
the RT-PCR reagents are functional. The PCC verifies reagent rehydration, and
PCR tube filling, and confirms that all reaction components are present in the
cartridge including monitoring for probe integrity and dye stability.
The wastewater sample is collected and homogenized and 300 uL is placed into
the sample chamber of the Xpress SARS-CoV-2 plus cartridge. The GeneXpert
cartridge is loaded onto the GeneXpert Instrument System platform, which
performs hands-off, automated sample processing, and real-time RT-PCR for
detection of viral RNA.
Section 8 Reagents and Instruments
Materials Provided
The Xpress SARS-CoV-2 plus kit contains sufficient reagents to process 10
samples or quality control samples. The kit contains the following:
-
Xpress SARS-CoV-2 plus Cartridges with Integrated Reaction Tubes
- 10
-
Bead 1, Bead 2, and Bead 3 (freeze-dried)
- 1 of each per cartridge
-
Lysis Reagent
- 1.5 mL per cartridge
-
Binding Reagent
- 1.5 mL per cartridge
-
Wash buffer
- 0.5 mL per cartridge
-
Elution Reagent
- 3.0 mL per cartridge
-
Disposable Transfer Pipettes
- 12 per kit
-
Flyer
- 1 per kit
Directions to locate the Product Insert on https://www.hach.com. To download the Instructions for Use (IFU) or ADF, visit https://www.hach.com, search LCE009, and click on Downloads.
Note: Safety Data Sheets (SDS) are available at https://www.hach.com.
Note: The bovine serum albumin (BSA) in the beads within this product was produced and manufactured exclusively from bovine plasma sourced in the United States. No ruminant protein or other animal protein was fed to the animals; the animals passed ante- and post-mortem testing. During processing, there was no mixing of the material with other animal materials.
Section 9 Storage and Handling
- Store the Xpress SARS-CoV-2 plus cartridges at 2-28 °C.
- Do not open a cartridge lid until you are ready to perform testing
Section 10 Warnings and Precautions
General
- For wastewater analysis use.
- Performance characteristics of this test have been established with water and wastewater samples only. This system is not intended for use for human or animal testing.
- Treat all wastewater samples and use cartridges as biological specimens.
- Follow safety procedures set by your institution for working with chemicals and handling wastewater samples.
- Consult your institution’s environmental waste personnel on the proper disposal of used cartridges, which may contain amplified material. This material may exhibit characteristics of federal EPA Resource Conservation and Recovery Act (RCRA) hazardous waste requiring specific disposal requirements. Check state and local regulations as they may differ from federal disposal regulations. Institutions should check the hazardous waste disposal requirements within their respective countries.
- Refer to the GeneXpert Dx System Operator Manual for additional errors and warnings while operating the instrument.
Samples
- Maintain proper storage conditions during sample transport to ensure the integrity of the sample (see Sample Collection, Transport, and Storage on page 6). Sample stability under shipping conditions other than those recommended has not been evaluated. It is recommended to follow the CDC guidelines for sample collection and storage.
Assay/Reagent
- Do not open the Xpress SARS-CoV-2 plus cartridge lid except when adding the wastewater sample.
- Do not use a cartridge that has been dropped after removing it from the packaging.
- Do not shake the cartridge. Shaking or dropping the cartridge after opening the cartridge lid may yield non-determinate results.
- Do not place the sample ID label on the cartridge lid or on the barcode label on the cartridge.
- Do not use a cartridge with a damaged barcode label.
- Do not use a cartridge that has a damaged reaction tube.
- Each single-use Xpress SARS-CoV-2 plus cartridge is used to process one test. Do not reuse processed cartridges.
- Each single-use disposable pipette is used to transfer one sample. Do not reuse disposable pipettes.
- Do not use a cartridge if it appears wet or if the lid seal appears to have been broken.
- Wear clean lab coats and gloves. Change gloves between the handling of each sample.
- In the event of a spill of the wastewater sample or controls, wear gloves and absorb the spill with paper towels. Then, thoroughly clean the contaminated area with 10% freshly prepared household chlorine bleach. Allow a minimum of two minutes of contact time. Ensure the work area is dry before using 70% denatured ethanol to remove bleach residue. Allow the surface to dry completely before proceeding. Or, follow your institution’s standard procedures for a contamination or spill event. For equipment, follow the manufacturer’s recommendations for decontamination of equipment.
- Wastewater samples, transfer devices, and used cartridges should be considered capable of transmitting infectious agents requiring standard precautions. Follow your institution’s environmental waste procedures for proper disposal of used cartridges and unused reagents. These materials may exhibit characteristics of chemical hazardous waste requiring specific disposal. Dispose of in accordance with local regulations. Dispose of waste in accordance with environmental legislation.
Section 11 Chemical Hazards
Consult the Safety Data Sheet (SDS).
Section 12 Sample Collection, Transport, and Storage
Proper sample collection, storage, and transport are critical to the
performance of this test. Inadequate sample collection, improper sample
handling and/or transport may yield a false result. It is important to follow
the CDC guidelines for proper sample collection and storage:
https://www.cdc.gov/healthywater/surveillance/wastewater-
surveillance/developing-a-wastewater-surveillance-sampling-strategy.html
Sampling Safety
Standard practices associated with wastewater treatment plant operations
should be followed to protect wastewater workers from SARS-CoV-2. These
standard practices can include engineering and administrative controls,
handwashing, specific safe work practices, and personal protective equipment4
normally required when handling untreated wastewater. Beyond CDC
recommendations for how to protect against COVID-19, no additional
COVID-19–specific protections are recommended for workers managing wastewater,
including those at wastewater treatment facilities.4
Storage
Never store samples at temperatures higher than refrigeration (4 °C).
Refrigerate samples during the collection process. If possible, process
samples within 24 hours of collection, as effective actionable wastewater
surveillance relies on rapid data collection. The remaining samples can be
frozen at -70 °C for archiving. Avoid more than one freeze-thaw cycle.
Preliminary data have shown potential loss of signal following freezing.
Shipping
When sending samples to laboratories, CDC recommends packing samples with cold
packs (4 °C) and using same-day or overnight shipping. Package and ship
samples as Category B infectious substance (UN 3373), in accordance with the
U.S. Department of Transportation’s Hazardous Materials Regulations and the
International Air Transport Association Dangerous Goods Regulations.
Community Wastewater Surveillance
Sampling wastewater for SARS-CoV-2 as it enters a treatment plant (referred to
as untreated influent) is used to evaluate trends in infection within the
community contributing water to the sewer system. Select the number of
treatment plants for community-level wastewater disease surveillance based on
the public health data needs in the region and the availability of resources.
COVID-19 rates and trends in the community, distribution of the population,
and characteristics of the sewer system may also influence your selection.
Wastewater treatment plants may be selected for community wastewater
surveillance to:
- Cover a certain percentage of the population.
- Provide data on communities at higher risk for COVID-19 or at increased risk for severe illness from COVID-19.
- Provide data on communities where timely COVID-19 clinical testing is underutilized or unavailable.
- Represent several sewer sheds that serve a larger interconnected population, such as in dense urban areas.
Prior to selecting a wastewater treatment plant for community wastewater surveillance, it is critical to consult with wastewater engineers and utility managers to understand:
- Geographic area and population served by the utility
- The relative contribution of the types of waste inputs (industrial, commercial, residential)
- Operating factors that could influence the detection of SARS-CoV-2 (e.g., pre-treatment of incoming wastewater or diversion of wastewater to adjust flow upstream of the sampling site)
- Available sampling locations at the treatment plant
- Utility capacity for sample collection, documentation, and shipping
- Availability of utility meta-data needed for public health interpretation (e.g., influent flow measurements, chemical/physical water quality measurements, service area shapefile)
Targeted Wastewater Surveillance
Targeted wastewater surveillance5 entails sampling wastewater from upstream in
the wastewater network (e.g., lift stations, interceptors, manholes). Targeted
wastewater surveillance may provide a better understanding of how SARS-CoV-2
infections are distributed within a sewer shed. However, there is currently
little data demonstrating the application of this approach. When deciding
whether targeted wastewater surveillance would be useful for public health
action, it is important to consider the following:
- SARS-CoV-2 RNA concentrations are more variable upstream from the wastewater treatment plant than at the plant intake because upstream wastewater has had less time to mix and contains feces from fewer people.
- Access to sewer lines serving only the intended target population may require infrastructure alterations or may not be possible.
- Depending on the size of the target population, conducting effective targeted wastewater surveillance may be more costly and logistically challenging than case surveillance.
Wastewater Sampling Frequency
Wastewater sampling frequency depends on how the data will be used for public
health and the prevalence of COVID-19 in the community. With sufficient
testing frequency, wastewater testing may be used to track trends over time.
Single samples or very infrequent (e.g., monthly) sampling will likely not be
informative for establishing trends, but could be used for establishing the
presence of COVID-19 in a community. If the goal of wastewater surveillance is
to screen for the presence of SARS-CoV-2 in wastewater, sampling once per week
may be adequate. If the goal is early indication of infection trends, at least
three sampling points are needed within a trend period of interest for
surveillance. There are little data available describing how rapidly
wastewater concentrations may change under various epidemic scenarios.
Consider the following when determining sample frequency at a specific
location:
- A minimum of three samples is required to detect wastewater trends over time. The time between consecutive wastewater samples determines the minimum length of time over which a trend may be detected. For example, if samples are collected twice per week, 8 days is the minimum timespan over which a trend can be confirmed.
- Laboratory testing capacity and supply chain shortages may limit the maximum sampling frequency.
- One-time sampling will not provide actionable data beyond the presence of SARS-CoV-2 infection within the sewer shed.
Wastewater Sample Types
Sample type is an important consideration for collecting representative
samples and will depend on the sample collection location and factors specific
to the wastewater treatment plant. Closely consult with treatment plant staff
to determine appropriate sample types that will best represent the target
population. Samples should be collected at locations that precede the addition
of chemicals or mixing of waste streams at the wastewater treatment plant. The
two wastewater surveillance sample types are untreated wastewater and primary
sludge.
- Untreated wastewater: Untreated wastewater includes waste from household or building use (e.g., toilets, showers, sinks), which contains human fecal waste, as well as waste from non-household sources (e.g., rainwater, industrial use). Untreated wastewater may be sampled from wastewater treatment plant influent (prior to primary treatment) or upstream in the wastewater collection network. Changes in SARS-CoV-2 RNA concentrations in wastewater samples collected from wastewater treatment plant influent have been shown to correlate with trends in reported cases. In most cases, untreated wastewater will likely require concentration6 prior to RNA extraction. The number of infections needed to detect the virus in wastewater without concentration is difficult to determine because it depends on both the method detection limit and the amount of virus in feces, for which there are few data.
- Primary sludge: Primary sludge comprises suspended solids that settle out of wastewater during the first solids removal (“sedimentation”) process at a wastewater treatment plant. Primary sludge is distinct from secondary sludge following primary treatment. Do not use secondary sludge for wastewater surveillance. Changes in SARS-CoV-2 RNA concentrations in primary sludge samples have been shown to correlate with trends in reported cases. An advantage of primary sludge samples compared to untreated wastewater is that SARS-CoV-2 concentrates in sludge, which reduces the sample volume required to detect the virus and may eliminate the need to concentrate the sample prior to quantification. However, the extent of SARS-CoV-2 RNA concentration in sludge is not well characterized. Sludge samples may also present challenges that must be evaluated for each wastewater treatment plant, such as chemicals added at the treatment plant, increased concentrations of compounds that can interfere with laboratory methods, or the addition of recycled waste streams from other parts of the treatment plant.
Selecting a Sample Type
Untreated wastewater and primary sludge are both acceptable community
wastewater surveillance sample types. For upstream targeted wastewater
surveillance, only untreated wastewater samples are available. If laboratory
methods are available, sludge sampling is recommended to evaluate infection
presence within a sewer shed with few known case patients because the virus
will be more concentrated in sludge. Untreated wastewater samples are
recommended when wastewater treatment plants apply disinfectant before sludge
can be sampled, sludge testing demonstrates high assay inhibition or poor
virus recovery, or solids residence time within the primary clarifier is
unknown.
Wastewater Sample Collection
Collection Methods
- Grab: Grab samples can be collected rapidly and do not require automated equipment. However, grab samples may be less representative of community fecal contributions than composite samples. For untreated wastewater and sludge, grab samples represent a single moment in time and are highly influenced by daily fluctuations in wastewater flow and composition. At the treatment plant level, grab samples may provide similar concentrations to composite samples if the proportion of the community that is infected is sufficiently high. However, at this time, the minimum proportion of the community that needs to be infected for grab and composite samples to be similar is unknown.
- Composite: Composite samples are collected by pooling multiple grab samples at a specified frequency over a set time period – typically 24 hours for wastewater surveillance. You can collect composite samples of untreated wastewater manually or using automated samplers with refrigeration capacity that collect flow-weighted samples (e.g., one sub-sample per 200,000 gallons of flow). Continuous composite samplers (versus flow-weighted) may improve how representative the sample is of the community contributing to the sewer. Composite samples are considered more representative of community fecal contributions than grab samples.
Selecting a Sample Volume
The volume of sample to collect will depend on the sample type (wastewater or
sludge). A 1 liter (L) composite wastewater sample or 100 milliliter (ml) grab
sludge sample volume should be adequate for testing. The maximum amount of
sludge solids that may be directly extracted is typically around 2 grams. The
remaining sample volume (if any) can be used for repeat measurement or to
assess biological variability. The volume of sample that is concentrated and
quantified will determine the lowest amount of SARSCoV- 2 RNA that can be
detected. Concentrating more than 1 L of wastewater may result in poor
recovery or viral signal inhibition. If using grab samples, consult with
wastewater treatment plant staff to collect representative samples that
capture peak times of human fecal loading and to understand the solids
residence time for sludge.
Section 13 Procedure
Preparing the Cartridge
Important: Start the test within 30 minutes of adding the sample to the cartridge.
Note: The preferred sampling strategy is to use a composite sample, allow the solids to settle, and pipette the sample from the top 1/3 of the sample volume. See Wastewater Sample Collection on page 8 for information on composite samples.
- Remove a cartridge from the package.
- Check the sample container is closed.
- Mix the sample by inverting the sample container 5 times. Let the sample sit for at least two minutes. Open the cap on the sample container.
- Open the cartridge lid.
- Remove the transfer pipette from the wrapper.
- Squeeze the top bulb of the transfer pipette completely until the top bulb is completely flat. While continuing to hold the bulb, place the pipette tip in the top 1/3 of the sample within the sample container (see Figure 1).
Figure 1 Transfer Pipette
- Squeeze here
- Pipette
- Overflow Reservoir Bulb
- Sample
- Release the top bulb of the pipette to fill the pipette before removing it from the tube. After filling the pipette, the excess sample will be seen in the overflow reservoir bulb of the pipette (see Figure 1). Check that the pipette does not contain bubbles.
- To transfer the sample to the cartridge, squeeze the top bulb of the transfer pipette completely again to empty the contents of the pipette (300 µL) into the large opening (Sample Chamber) in the cartridge shown in Figure 2. Dispose of the used pipette.
Figure 2 Xpress SARS-CoV-2 plus Cartridge (Top View)
- Close the cartridge lid.
Starting the Test
Note: Before you start the test, make sure that the system contains modules with GeneXpert Dx software version 4.7b or higher and that the Xpress SARS-CoV-2 plus Assay Definition File is imported into the software. This section lists the default steps to operate the GeneXpert Instrument System. For detailed instructions, see the GeneXpert Dx System Operator Manual.
Note: The steps you follow may be different if the system administrator has changed the default workflow of the system.
-
Turn on the GeneXpert Instrument System:
- First turn on the instrument, and then turn on the computer. Log into the Windows operating system. The GeneXpert software may launch automatically or may require double-clicking on the GeneXpert Dx shortcut icon on the Windows® desktop.
-
Log on to the System software. The login screen appears. Type your username and password.
-
In the GeneXpert System window, click Create Test.
-
Scan or type in the Sample ID (optional). If typing the Sample ID, make sure the Sample ID is typed correctly. The Sample ID is shown on the left side of the View Results window and is associated with the test result.
-
Scan or type in the Sample ID. If typing the Sample ID, make sure the Sample ID is typed correctly. The Sample ID is shown on the left side of the View Results window and is associated with the test result.
-
Scan the barcode on the Xpress SARS-CoV-2 plus cartridge. Using the barcode information, the software automatically fills the boxes for the following fields: Reagent Lot ID, Cartridge SN, Expiration Date and Selected Assay.
Note: If the barcode on the Xpress SARS-CoV-2 plus cartridge does not scan, then repeat the test with a new cartridge. -
Click Start Test if Auto-Submit is not enabled. In the dialogue box that appears, type your password, if required.
- Locate the module with the blinking green light, open the instrument module door and load the cartridge.
- Close the door. The test starts and the green light stops blinking. When the test is finished, the light turns off and the door will unlock. Remove the cartridge.
- Dispose of used cartridges in the appropriate sample waste containers according to your institution’s standard practices.
Note: Do not turn off or unplug the instruments while a test is in progress. Turning off or unplugging the GeneXpert instrument or computer will stop the test.
Section 14 Quality Control
Internal Controls
Each cartridge includes a Sample Processing Control (SPC) and Probe Check
Control (PCC).
Sample Processing Control (SPC) – Ensures that the sample is processed correctly. The SPC verifies that sample processing is adequate. Additionally, this control detects sample-associated inhibition of the real-time PCR assay, ensures that the PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The SPC should be positive in a negative sample and can be negative or positive in a positive sample. The SPC passes if it meets the validated acceptance criteria.
Probe Check Control (PCC) – Before the start of the PCR reaction, the GeneXpert System measures the fluorescence signal from the probes to monitor bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC passes if it meets the validated acceptance criteria.
Section 15 Retests
Reasons to Repeat the Assay
If any of the test results mentioned below occur, repeat the test once
according to the instructions in the Retest Procedure on page 12.
- An INVALID result indicates that the control SPC failed. The sample was not properly processed, PCR was inhibited, or the sample was not properly collected.
- An ERROR result could be due to, but not limited to, Probe Check Control failure, system component failure, no sample added, or the maximum pressure limits being exceeded.
- A NO RESULT indicates that insufficient data were collected. For example, a cartridge failed an integrity test, the operator stopped a test that was in progress, or a power failure occurred.
If an External Control fails to perform as expected, repeat the external control test.
Retest Procedure
To retest a non-determinate result (INVALID, NO RESULT, or ERROR), or a
PRESUMPTIVE POS result, use a new cartridge. Use the leftover sample from the
original specimen transport medium tube or new external control tube.
- Put on a clean pair of gloves. Obtain a new Xpress SARS-CoV-2 plus cartridge and a new transfer pipette.
- Check the specimen transport tube or external control tube is closed.
- Mix the sample by rapidly inverting the specimen transport medium tube or external control tube. Open the cap on the specimen transport tube or external control tube.
- Open the cartridge lid.
- Using a clean transfer pipette (supplied), transfer sample (one draw) to the sample chamber with the large opening in the cartridge.
- Close the cartridge lid.
Section 16 References
-
Centres for Disease Control and Prevention. https://www.cdc.gov/coronavirus/2019-ncov/index.html. Accessed February 9, 2020.
-
bioRxiv. (https://www.biorxiv.org/content/10.1101/2020.02.07.937862v1). Accessed March 3, 2020.
-
Kirby AE, Walters MS, Jennings WC, et al. Using Wastewater Surveillance Data to Support the COVID-19 Response — United States, 2020–2021. MMWR Morb Mortal Wkly Rep
2021;70:1242–1244. DOI: http://dx.doi.org/10.15585/mmwr.mm7036a2externalicon -
Guidance for Reducing Health Risks to Workers Handling Human Waste or Sewage. Centres for Disease Control and Prevention. December 9, 2021.
https://www.cdc.gov/healthywater/global/sanitation/workers_handlingwaste.html. -
Targeted Wastewater Surveillance at Facilities and Institutions. Centers for Disease Control and Prevention. January 26, 2022. https://www.cdc.gov/healthywater/surveillance/wastewater-surveillance/targeted-use-case.html.
-
Wastewater Surveillance Testing Methods. Centres for Disease Control and Prevention. January 26, 2022. https://www.cdc.gov/healthywater/surveillance/wastewater-surveillance/testing-methods.html.
Section 17 Technical Assistance
Hach Company should be the first point of contact for Technical Support.
Before contacting Technical Support, collect the following information:
- Product name
- Lot number of kit
- Serial number of instrument
- Error messages (if any)
- Software version and, if applicable, Computer Service Tag Number
Region | Telephone | |
---|---|---|
US | + 1 800 227 4224 | techhelp@hach.com |
EU | + 33 563 825 319 | [techsupport- |
eu@hach.com](mailto:support@cepheideurope.com)
Section 18 Table of Symbols
HACH COMPANY World Headquarters
- P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
- Tel. 970-669-3050
- 800-227-4224 (U.S.A. only)
- Fax 970-669-2932
- orders@hach.com
- www.hach.com
Hach Company/Hach Lange GmbH, 2023. All rights reserved. Printed in U.S.A.
References
- Geräte und Reagenzien für die Wasserqualitätsanalyse | Hach
- Hach | Hach
- Severe acute respiratory syndrome-related coronavirus: The species and its viruses – a statement of the Coronavirus Study Group | bioRxiv
- Coronavirus Disease 2019 (COVID-19) | CDC
- Guidance for Reducing Health Risks to Workers Handling Human Waste or Sewage | Global Water, Sanitation and Hygiene | Healthy Water | CDC
- cdc.gov/healthywater/surveillance/wastewater-surveillance/developing-a-wastewater-surveillance-sampling-strategy.html
- cdc.gov/healthywater/surveillance/wastewater-surveillance/testing-methods.html
- Hach | Hach