HACH DOC023 Inductive Conductivity Sensors User Manual
- June 14, 2024
- HACH
Table of Contents
DOC023.97.80079
Inductive Conductivity Sensors
03/2023, Edition 9 Basic User Manual
Section 1 Online user manual
This Basic User Manual contains less information than the User Manual, which is available on the manufacturer’s website.
Section 2 Specifications
Specifications are subject to change without notice.
Specification | Details |
---|---|
Dimensions | Refer to Figure 1 on page 5. |
Pollution degree | 2 |
Overvoltage category | I |
Protection class | III |
Altitude | 2000 m (6562 ft) maximum |
Operating temperature | -20 to 60 °C (-4 to 140 °F) |
Storage temperature | -20 to 70 °C (-4 to 158 °F) |
Weight | Approximately 1 kg (2.2 lbs) |
Wetted materials | Polypropylene, PVDF, PEEK or PFA |
Sensor cable | 5-conductor (plus two isolated shields), 6 m (20 ft); rated at |
150 °C (302 °F)— polypropylene
Conductivity range| 0.0 to 200.0 μS/cm; 0 to 2,000,000 μS/cm
Accuracy| 0.01 % of reading, all ranges
Repeatability/precision| > 500 μS/cm: ±0.5% of reading; < 500 μS/cm: ±5 μS/cm
Maximum flow rate| 0–3 m/s (0–10 ft/s)
Temperature/pressure limit| Polypropylene: 100 °C at 6.9 bar (212 °F at 100
psi); PVDF: 120 °C at 6.9 bar (248 °F at 100 psi); PEEK and PFA: 200 °C at
13.8 bar (392 °F at 200 psi)
Transmission distance| 200 to 2000 μS/cm: 61 m (200 ft); 2000 to 2,000,000
μS/cm: 91 m (300 ft)
Temperature measurement range| -10 to 135 °C (14 to 275 °F) limited by sensor
body material
Temperature sensor| Pt 1000 RTD
Calibration methods| Zero calibration, 1-point conductivity calibration,
1-point temperature calibration
Sensor interface| Modbus
Certifications| Listed by ETL (US/Canada) for use in Class 1, Division 2,
Groups A, B, C, D, Temperature Code T4 – Hazardous Locations with Hach SC
Controller.
Conforms to: CE, UKCA, FCC, ISED, ACMA, KC, CMIM. Sanitary sensors certified
by 3A.
Warranty| 1 year; 2 years (EU)
Section 3 General information
In no event will the manufacturer be liable for damages resulting from any
improper use of product or failure to comply with the instructions in the
manual. The manufacturer reserves the right to make changes in this manual and
the products it describes at any time, without notice or obligation. Revised
editions are found on the manufacturer’s website.
3.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is soley responsible to identify critical
application risks and install appropriate mechanisms to protect processes
during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this
equipment. Pay attention to all danger and caution statements. Failure to do
so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do
not use or install this equipment in any manner other than that specified in
this manual.
3.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
3.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage
to the instrument could occur if not observed. A symbol on the instrument is
referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for
operation and/or safety information.
** Electrical equipment marked with this symbol may not be disposed of in
European domestic or public disposal systems. Return old or end-of-life
equipment to the manufacturer for disposal at no charge to the user.
3.2 Product overview
DANGER
Chemical or biological hazards. If this instrument is used to monitor a
treatment process and/or chemical feed system for which there are regulatory
limits and monitoring requirements related to public health, public safety,
food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to
have sufficient and appropriate mechanisms in place for compliance with
applicable regulations in the event of malfunction of the instrument.
NOTICE
Use of this sensor may lead to fissures of the coating, exposing the
underlying substrate to the environment in which the sensor is immersed.
Therefore, this sensor has not been developed for, and is not meant for use in
applications where the liquid is expected to conform to certain purity or
cleanliness parameters and in which contamination could result in substantial
damages. These applications typically include semiconductor manufacturing
applications and may include other applications in which the user must assess
risk of contamination and subsequent impact on product quality. The
manufacturer advises against the use of the sensor in these applications and
assumes no responsibility for any claims or damages arising as a result of the
sensor being used in or in relation to these applications.
This sensor is designed to work with a controller for data collection and
operation. Different controllers can be used with this sensor. This document
assumes sensor installation and use with an SC4500 Controller. To use the
sensor with other controllers, refer to the user manual for the controller
that is used.
Refer to Figure 1 for the sensor dimensions.
3.3 Product components
Make sure that all components have been received. Refer to Figure 2 and Figure
3. If any items are missing or damaged, contact the manufacturer or a sales
representative immediately.
Note:** The sensor can be ordered without the digital gateway that is shown
in Figure 3.
- Convertible-style sensor—for installation in a pipe tee or an open vessel with applicable mounting hardware
- Sanitary-style sensor—for installation in a 2-inch sanitary tee
- Ferrite
- Digital gateway
- Mounting bracket
- Nut with lock washer, #8-32
- Flat washer, #8 (2x)
- Screw, cross-slot, #8-32 x 1.25 in.
- Screwdriver (for the terminal block)
Section 4 Installation
WARNING
Multiple hazards. Only qualified personnel must conduct the tasks described in
this section of the document.
4.1 Install the sensor in the sample stream
WARNING
Explosion hazard. For installation in hazardous (classified) locations, refer
to the instructions and control drawings in the controller Class 1, Division 2
documentation. Install the sensor according to local, regional and national
codes. Do not connect or disconnect the instrument unless the environment is
known to be non-hazardous.
WARNING
Explosion hazard. Make sure that the mounting hardware for the sensor has a
temperature and pressure rating sufficient for the mounting location.
Refer to Figure 4 for installation of the sensor in different applications.
The sensor must be calibrated before use. Refer to Calibrate the sensor on
page 12.
Make sure that the routing of the sensor cable prevents exposure to high
electromagnetic fields (e.g., transmitters, motors and switching equipment).
Exposure to these fields can cause inaccurate results.
- Sanitary (CIP) flange mount
- Union T-mount
- End of pipe immersion
- Ball valve insertion
4.2 Electrical installation
4.2.1 Prepare the sensor wires
If the sensor cable length is changed, prepare the wires as shown in Figure 5.
1 Outer shield wire1 | 5 Red wire |
---|---|
2 Inner shield wire2 | 6 Clear heat-shrink tubing3 |
3 Twisted pair, yellow wire and green wire | 7 Black heat-shrink |
tubing3
4 Twisted pair, white wire and blue wire| 8 Inner conductor shields4
4.2.2 Electrostatic discharge (ESD) considerations
NOTICE
Potential Instrument Damage. Delicate internal electronic components can be
damaged by static electricity, resulting in degraded performance or eventual
failure.
Refer to the steps in this procedure to prevent ESD damage to the instrument:
- Touch an earth-grounded metal surface such as the chassis of an instrument, a metal conduit or pipe to discharge static electricity from the body.
- Avoid excessive movement. Transport static-sensitive components in anti-static containers or packages.
- Wear a wrist strap connected by a wire to earth ground.
- Work in a static-safe area with anti-static floor pads and work bench pads.
4.2.3 Connect the sensor to an SC Controller
Use one of the options that follows to connect the sensor to an SC Controller:
- Install a sensor module in the SC Controller. Then, connect the bare wires of the sensor to the sensor module. The sensor module converts the analog signal from the sensor to a digital signal.
- Connect the bare wires of the sensor to an sc digital gateway, then connect the sc digital gateway to the SC Controller. The digital gateway converts the analog signal from the sensor to a digital signal.
1. The shield wire for the sensor cable
2. The shield wire for the green and yellow twisted pair
3. User supplied
4. The inner conductor shields are foil tubes with a conductive interior side and a nonconductive exterior side. Make sure to keep the electrical isolation between the interior side of the inner conductor shields. Make sure that the interior side of the inner conductive shields is not exposed.
Refer to the instructions supplied with the sensor module or sc digital
gateway.
4.2.4 PRO-series Model E3 electrodeless conductivity transmitter
To connect the sensor to a PRO-series Model E3 electrodeless conductivity
transmitter, remove power to the transmitter and refer to Figure 6 and Table
Table 1 Sensor wiring information
Terminal (TB2) | Wire | Terminal (TB2) | Wire |
---|---|---|---|
1 | White | 4 | Red |
2 | Blue | 5 | Yellow |
3 | Clear (inner shield)5 | 6 | — |
3 | Black (outer shield)5 | 7 | Green |
Section 5 Operation
WARNING
Fire hazard. This product is not designed for use with flammable liquids.
5.1 User navigation
Refer to the controller documentation for the touchscreen description and
navigation information.
5.2 Configure the sensor
Use the Settings menu to enter identification information for the sensor and
to change options for data handling and storage.
-
Select the main menu icon, then select Devices. A list of all of the available devices shows.
-
Select the sensor and select Device menu > Settings.
5 For the best immunity to electrical noise, connect the inner shield wire and the outer shield wire together with solder before they are put in the terminal block. -
Select an option.
• For sensors connected to a conductivity module, refer to Table 2.
• For sensors connected to an sc digital gateway, refer to Table 3.
Table 2 Sensors connected to conductivity module
Option | Description |
---|---|
Name | Changes the name that corresponds to the sensor on the top of the |
measurement screen. The name is limited to 16 characters in any combination of
letters, numbers, spaces or punctuation.
Sensor S/N| Lets the user enter the serial number of the sensor. The
serial number is limited to 16 characters in any combination of letters,
numbers, spaces or punctuation.
Measurement type| Changes the measured parameter to Conductivity
(default), Concentration, TDS (total dissolved solids) or Salinity. When the
parameter is changed, all other configured settings are reset to the default
values.
Format| Changes the number of decimal places that are shown on the
measurement screen to Auto, X.XXX, XX.XX or XXX.X. When Auto is selected, the
decimal places automatically change.
Note: The Auto option is only available when the Measurement type setting
is set to Conductivity.
Conductivity unit| Note: The Conductivity unit setting is only
available when the Measurement type setting is set to Conductivity or
Concentration.
Changes the conductivity units—Auto, µS/cm, mS/cm or S/cm.
Temperature| Sets the temperature units to °C (default) or °F.
T-compensation| Adds a temperature-dependent correction to the measured
value—None, Linear (default: 2.0%/°C, 25 °C), Natural water or Temperature
compensation table.
When Temperature compensation table is selected, the user can enter x, y (°C,
%/°C) points in ascending order.
Note: The Natural water option is not available when the Measurement type
setting is set to TDS or Concentration.
Concentration measurement| Note: The Concentration measurement
setting is only available when the Measurement type setting is set to
Concentration.
Sets the type of concentration table to use—Built-in (default) or User
compensation table.
When Built-in is selected, the user can select the chemical that is measured—
H3PO4: 0–40%; HCl: 0–18% or 22–36%; Na OH: 0–16%; CaCl2 0–22%; HNO3: 0–28% or
36–96%; H2SO4: 0–30%, 40–80% or 93–99%; HF: 0–30%; Na Cl: 0–25%; HBr, KOH,
Seawater
When User compensation table is selected, the user can enter x, y
(conductivity, %) points in ascending order.
TDS (total dissolved solids)| Note: The TDS (total dissolved solids)
setting is only available when the Measurement type setting is set to TDS.
Sets the factor that is used to convert conductivity to TDS—Na Cl (default) or
Custom (enter a factor between 0.01 and 99.99 ppm/µS, default: 0.49 ppm/µS).
Temperature element| Sets the temperature element for automatic
temperature compensation to PT100, PT1000 (default) or Manual. If no element
is used, set to Manual and set a value for temperature compensation (default:
25 °C).
When Temperature element is set to PT100 or PT1000, refer to Adjust the
T-factor for non-standard cable lengths on page 12 to set the T factor
setting.
Note: If Temperature element is set to Manual and the sensor is replaced
or the sensor days are reset, Temperature element automatically changes back
to the default setting (PT1000).
Cell constant parameters| Changes the cell constant to the actual
certified K value from the label on the sensor cable. When the certified K
value is entered, the calibration curve is defined. Default: 4.70
Option| Description
---|---
Filter| Sets a time constant to increase signal stability. The time
constant calculates the average value during a specified time—0 (no effect,
default) to 200 seconds (average of signal value for 200 seconds). The filter
increases the time for the sensor signal to respond to actual changes in the
process.
Data logger interval| Sets the time interval for sensor and temperature
measurement storage in the data log— 5, 30 seconds or 1, 2, 5, 10, 15
(default), 30, 60 minutes
Reset settings to default values| Sets the Settings menu to the factory
default settings and resets the counters. All sensor information is lost.
Table 3 Sensors connected to SC digital gateway
Option | Description |
---|---|
Name | Changes the name that corresponds to the sensor on the top of the |
measurement screen. The name is limited to 16 characters in any combination of
letters, numbers, spaces or punctuation.
Measurement type| Changes the measured parameter to Conductivity
(default), Concentration, TDS (total dissolved solids) or Salinity. When the
parameter is changed, all other configured settings are reset to the default
values.
Conductivity unit| Note: The Conductivity unit setting is only
available when the Measurement type setting is set to Conductivity,
Concentration or Salinity.
Changes the conductivity units—µS/cm (default), ms/cm or S/cm.
Cell constant parameters| Note: The Cell constant parameters setting
is only available when the Measurement type setting is set to Conductivity or
Salinity.
Changes the cell constant to the actual certified K value from the label on
the sensor cable. When the certified K value is entered, the calibration curve
is defined. Default: 4.70
Concentration measurement| Note: The Concentration measurement
setting is only available when the Measurement type setting is set to
Concentration.
Sets the type of concentration table to use—Built-in (default) or User
defined.
When Built-in is selected, the user can select the chemical that is measured—
H3PO4: 0–40%; HCl: 0–18% or 22–36%; Na OH: 0–16%; CaCl2 0–22%; HNO3: 0–28% or
36–96%; H2SO4: 0–30%, 40–80% or 93–99%; HF: 0–30%
When User defined is selected, the user can enter x, y (conductivity, %)
points in ascending order.
TDS (total dissolved solids)| Note: The TDS (total dissolved solids)
setting is only available when the Measurement type setting is set to TDS.
Sets the factor that is used to convert conductivity to TDS—Na Cl (default) or
User defined (enter a factor between 0.01 and 99.99 ppm/µS, default: 0.49
ppm/µS).
Temperature| Sets the temperature units to °C (default) or °F.
T-compensation| Adds a temperature-dependent correction to the measured
value—None, Linear (default: 2.0%/°C, 25 °C), Natural water or Temperature
compensation table.
When Temperature compensation table is selected, the user can enter x, y (°C,
%/°C) points in ascending order.
Note: The Natural water option is not available when the Measurement type
setting is set to TDS.
Note: The T-compensation setting is set to None when the Measurement type
setting is set to Concentration.
Data logger interval| Sets the time interval for sensor and
temperature measurement storage in the data log— Disabled (default), 5, 10,
15, 30 seconds, 1, 5, 10, 15, 30 minutes or 1, 2, 6, 12 hours
Alternating current frequency| Selects the power line frequency to
get the best noise rejection. Options: 50 or 60 Hz (default).
Option| Description
---|---
Filter| Sets a time constant to increase signal stability. The time
constant calculates the average value during a specified time—0 (no effect,
default) to 60 seconds (average of signal value for 60 seconds). The filter
increases the time for the sensor signal to respond to actual changes in the
process.
Temperature element| Sets the temperature element for automatic
temperature compensation to PT1000 (default) or Manual. If no element is used,
set to Manual and set a value for temperature compensation (default: 25 °C).
When Temperature element is set to PT1000, refer to Adjust the T-factor for
non-standard cable lengths on page 12 to set the Factor setting.
Note: If Temperature element is set to Manual and the sensor is replaced
or the sensor days are reset, Temperature element automatically changes back
to the default setting (PT1000).
Last calibration| Sets a reminder for the next calibration (default: 60
days). A reminder to calibrate the sensor shows on the display after the
selected interval from the date of the last calibration. For example, if the
date of the last calibration was June 15 and Last calibration is set to 60
days, a calibration reminder shows on the display on August 14. If the sensor
is calibrated before August 14, on July 15, a calibration reminder shows on
the display on September 13.
Sensor days| Sets a reminder for sensor replacement (default: 365 days).
A reminder to replace the sensor shows on the display after the selected
interval.
The Sensor days counter shows on the Diagnostics/Test > Counter menu.
When the sensor is replaced, reset the Sensor days counter on the
Diagnostics/Test > Counter menu.
Reset setup| Sets the Settings menu to the factory default settings and
resets the counters. All sensor information is lost.
5.3 Adjust the T-factor for non-standard cable lengths
When the sensor cable is extended or shortened from the standard 6 m (20 ft),
the resistance of the
cable changes. This change reduces the accuracy of temperature measurements.
To correct for this
difference, calculate a new T-factor.
-
Measure the temperature of a solution with the sensor and with an independent, reliable instrument such as a thermometer.
-
Record the difference between the temperature measured from the sensor and from the independent source (actual).
For example, if the actual temperature is 50 °C and the sensor reading is 53 °C, the difference is 3 °C. -
Multiply this difference by 3.85 to get the adjustment value.
Example: 3 x 3.85 = 11.55. -
Calculate a new T-factor:
• Sensor temperature > actual—add the adjustment value to the T-factor that is on the label on the sensor cable
• Sensor temperature < actual—subtract the adjustment value from the T-factor that is on the label on the sensor cable -
Select Settings > Temperature element > T factor (or Factor) and enter the new T-factor.
5.4 Calibrate the sensor
WARNING
Fluid pressure hazard. Removal of a sensor from a pressurized vessel can be
dangerous. Reduce the process pressure to below 7.25 psi (50 kPa) before
removal. If this is not possible, use extreme caution. Refer to the
documentation supplied with the mounting hardware for more information.;
WARNING
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
5.4.1 About sensor calibration
The wet cal method should be used to calibrate the conductivity sensor:
- Wet cal—use air (Zero Cal) and a reference solution or process sample of known value to define a calibration curve. A reference solution calibration is recommended for best accuracy. When the process sample is used, the reference value must be determined with a secondary verification instrument. Be sure to enter the T-factor in Temperature element of the Settings menu for accurate temperature compensation.
During calibration, data is not sent to the Datalog. Thus, the Datalog can
have areas where the data is intermittent.
5.4.2 Change calibration options
For sensors connected to a conductivity module, the user can set a reminder or
include an operator ID with calibration data from the Calibration options
menu.
Note: This procedure is not applicable to sensors connected to an sc
digital gateway.
- Select the main menu icon, then select Devices. A list of all of the available devices shows.
- Select the sensor and select Device menu > Calibration.
- Select Calibration options.
- Select an option.
Option | Description |
---|---|
Calibration reminder | Sets a reminder for the next calibration (default: Off). |
A reminder to calibrate the sensor shows on the display after the selected
interval from the date of the last calibration. For example, if the date of
the last calibration was June 15 and Last calibration is set to 60 days, a
calibration reminder shows on the display on August 14. If the sensor is
calibrated before August 14, on July 15, a calibration reminder shows on the
display on September 13.
Operator ID for calibration| Includes an operator ID with calibration data—Yes
or No (default). The ID is entered during the calibration.
5.4.3 Zero calibration procedure
Use the zero calibration procedure to define the unique zero point of the
conductivity sensor. The zero point must be defined before the sensor is
calibrated for the first time with a reference solution or process sample.
-
Remove the sensor from the process. Wipe the sensor with a clean towel or use compressed air to make sure the sensor is clean and dry.
-
Select the main menu icon, then select Devices. A list of all of the available devices shows.
-
Select the sensor and select Device menu > Calibration.
-
Select Zero calibration (or 0-point calibration).;
-
Select the option for the output signal during calibration:
Option| Description
---|---
Active| The instrument sends the current measured output value during the calibration procedure.
Hold| The sensor output value is held at the current measured value during the calibration procedure.
Transfer| A preset output value is sent during calibration. Refer to the controller user manual to change the preset value. -
Hold the dry sensor in the air and push OK.
-
Do not push OK until the calibration result shows on the display.
-
Review the calibration result:
• “The calibration was successfully completed.”—The sensor is calibrated and ready to measure samples. The slope and/or offset values are shown.
• “The calibration failed.” —The calibration slope or offset is outside of accepted limits. Repeat the calibration with fresh reference solutions. Clean the sensor if necessary. -
Push OK.
-
Proceed to the calibration with a reference solution or process sample.
5.4.4 Calibration with a reference solution
Calibration adjusts the sensor reading to match the value of a reference
solution. Use a reference solution that is at the same value or higher than
the expected measurement readings.
Note: If the sensor is being calibrated for the first time, be sure to
complete the zero calibration first.
-
Thoroughly rinse the clean sensor in deionized water.
-
Put the sensor in the reference solution. Support the sensor so that it does not touch the container. Make sure that the sensing area is fully immersed in the solution (Figure 7). Stir the sensor to remove bubbles.
-
Wait for the sensor and solution temperature to equalize. This can take 30 minutes or more if the temperature difference between the process and reference solution is significant.
-
Select the main menu icon, then select Devices. A list of all of the available devices shows.
-
Select the sensor and select Device menu > Calibration.
-
Select Conductivity solution ( or Conductivity calibration if the sensor is connected to an sc digital gateway).
-
Select the option for the output signal during calibration:
Option| Description
---|---
Active| The instrument sends the current measured output value during the calibration procedure.
Hold| The sensor output value is held at the current measured value during the calibration procedure.
Transfer| A preset output value is sent during calibration. Refer to the controller user manual to change the preset value. -
Enter the reference temperature of the reference solution and push OK.
-
Enter the slope of the reference solution and push OK.
-
With the sensor in the reference solution, push OK.
-
Wait for the value to stabilize and push OK.
Note: The screen may advance to the next step automatically. -
Enter the value of the reference solution and push OK.
-
Review the calibration result:
• “The calibration was successfully completed.”—The sensor is calibrated and ready to measure samples. The slope and/or offset values are shown.
• “The calibration failed.” —The calibration slope or offset is outside of accepted limits. Repeat the calibration with fresh reference solutions. Clean the sensor if necessary. -
Push OK to continue.
-
Return the sensor to the process and push OK.
The output signal returns to the active state and the measured sample value is shown on the measurement screen.
5.4.5 Calibration with the process sample
The sensor can remain in the process sample, or a portion of the process
sample can be removed for calibration. The reference value must be determined
with a secondary verification instrument.
Note: If the sensor is calibrated for the first time, be sure to complete
the zero calibration first.
-
Select the main menu icon, then select Devices. A list of all of the available devices shows.
-
Select the sensor and select Device menu > Calibration.
-
Select Conductivity calibration, TDS calibration or Concentration calibration (or Calibration).
Note: Use the Measurement type setting to change the parameter that is calibrated. -
Select the option for the output signal during calibration:
Option| Description
---|---
Active| The instrument sends the current measured output value during the calibration procedure.
Hold| The sensor output value is held at the current measured value during the calibration procedure.
Transfer| A preset output value is sent during calibration. Refer to the controller user manual to change the preset value. -
With the sensor in the process sample, push OK. The measured value is shown.
-
Wait for the value to stabilize and push OK.
Note: The screen may advance to the next step automatically. -
Measure the conductivity (or other parameter) value with a secondary verification instrument. Use the arrow keys to enter the measured value and push OK.
-
Review the calibration result:
• “The calibration was successfully completed.”—The sensor is calibrated and ready to measure samples. The slope and/or offset values are shown.
• “The calibration failed.” —The calibration slope or offset is outside of accepted limits. Repeat the calibration with fresh reference solutions. Clean the sensor if necessary. -
Push OK to continue.
-
Return the sensor to the process and push OK.
The output signal returns to the active state and the measured sample value is shown on the measurement screen.
5.4.6 Temperature calibration
The instrument is calibrated at the factory for accurate temperature
measurement. The temperature can be calibrated to increase accuracy.
- Put the sensor in a container of water.
- Measure the temperature of the water with an accurate thermometer or independent instrument.
- Select the main menu icon, then select Devices. A list of all of the available devices shows.
- Select the sensor and select Device menu > Calibration.
- Select 1-point temperature calibration (or Temperature adjustment).
- Enter the exact temperature value and push OK.
- Return the sensor to the process.
5.4.7 Exit calibration procedure
- To exit a calibration, push the back icon.
- Select an option, then push OK.
Option | Description |
---|---|
Quit calibration (or Cancel) | Stop the calibration. A new calibration must |
start from the beginning.
Return to calibration| Return to the calibration.
Leave calibration (or Exit)| Exit the calibration temporarily. Access to other
menus is allowed. A calibration for a second sensor (if present) can be
started.
5.4.8 Reset the calibration
The calibration can be reset to the factory default settings. All sensor
information is lost.
- Select the main menu icon, then select Devices. A list of all of the available devices shows.
- Select the sensor and select Device menu > Calibration.
- Select Reset to default calibration values or Reset to calibration defaults (or Reset setup), then push OK.
- Push OK again.
5.5 Modbus registers
A list of Modbus registers is available for network communication. Refer to
the manufacturer’s website for more information.
Section 6 Maintenance
WARNING
Multiple hazards. Only qualified personnel must conduct the tasks described in
this section of the document.
WARNING
Explosion hazard. Do not connect or disconnect the instrument unless the
environment is known to be non-hazardous. Refer to the controller Class 1,
Division 2 documentation for hazardous location instructions.
WARNING
Fluid pressure hazard. Removal of a sensor from a pressurized vessel can be
dangerous. Reduce the process pressure to below 7.25 psi (50 kPa) before
removal. If this is not possible, use extreme caution. Refer to the
documentation supplied with the mounting hardware for more information.
WARNING
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
6.1 Clean the sensor
Pre-requisite: Prepare a mild soap solution with warm water and dishwashing
detergent, Borax hand soap or a similar soap.
Examine the sensor periodically for debris and deposits. Clean the sensor when
there is a buildup of deposits or when performance has degraded.
- Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with clean, warm water.
- Soak the sensor for 2 to 3 minutes in the soap solution.
- Use a soft bristle brush to scrub the entire measuring end of the sensor. Scrub the inside of the toroid.
- If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5% HCl for a maximum of 5 minutes.
- Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
- Rinse the sensor with clean water.
Always calibrate the sensor after maintenance procedures are done.
Section 7 Troubleshooting
7.1 Intermittent data
During calibration, data is not sent to the Datalog. Thus, the Datalog can
have areas where the data is intermittent.
7.2 Test the conductivity sensor
If a calibration fails, first complete the maintenance procedures in
Maintenance on page 16.
- Disconnect the sensor wires.
- Use an ohmmeter to test the resistance between the sensor wires as shown in Table 4.
Note: Be sure that the ohmmeter is set to its highest range for all
infinite (open circuit) resistance readings.
Table 4 Conductivity resistance measurements
Measurement points | Resistance |
---|---|
Between red and yellow wires | 1090–1105 ohms at 23–27 °C6 |
Between blue and white wires | Less than 5 ohms |
Between green and yellow wires | Less than 5 ohms |
Between white and shield wires | Infinite (open circuit) |
If one or more of the measurements is incorrect, call technical support.
Supply technical support with the serial number of the sensor and the
resistance values measured.
7.3 Diagnostics/Test menu
The Diagnostics/Test menu shows current and historical information about the
sensor. Refer to Table 5. Push the main menu icon, then select Devices. Select
the device and select Device menu > Diagnostics/Test.
Table 5 Diagnostics/Test menu
Option | Description |
---|---|
Module information | For sensors connected to a conductivity module |
only—Shows the version and the serial number of the conductivity module.
Sensor information| For sensors connected to a conductivity module—Shows
the sensor name and the serial number entered by the user.
For sensors connected to an sc digital gateway—Shows the sensor model number
and the sensor serial number. Shows the software version and driver version
installed.
Last calibration| For sensors connected to a conductivity module
only—Shows the number of days since the last calibration was done.
Calibration history| For sensors connected to a conductivity
module—Shows the calibration slope and date of the previous calibrations.
For sensors connected to an sc digital gateway—Shows the cell constant
parameters, offset correction and date of the last calibration.
Reset calibration history| For sensors connected to a conductivity
module only—For service use only
Sensor signals (or Signals )| For sensors connected to a conductivity
module only—Shows the current conductivity and temperature reading.
For sensors connected to an sc digital gateway—Shows the current temperature
analog to digital converter counter. Select Sensor signal to show current
measurement analog to digital converter counter or set the sensor range
(default: 6). Select Sensor measurement to show the sensor reading.
Sensor days (or Counter )| Shows the number of days that the sensor
has been in operation.
For sensors connected to an sc digital gateway—Select Counter to show the
number of days that the sensor has been in operation. To reset the counter to
zero, select Reset.
Reset the Sensor days counter when the sensor is replaced.
Reset| For sensors connected to a conductivity module only—Sets the
Sensor days counter to zero.
Reset the Sensor days counter when the sensor is replaced.
Factory calibration| For sensors connected to a conductivity module
only—For service use only
6 An infinite value (open circuit) or 0 ohms (short circuit) identifies a
failure.
7.4 Error list
When an error occurs, the reading on the measurement screen flashes and all
outputs are held when specified in the Controller > Outputs menu. The screen
changes to red. The diagnostics bar shows the error. Push on the diagnostic
bar to show the errors and warnings. As an alternative, push the main menu
icon, then select Notifications > Errors.
A list of possible errors is shown in Table 6.
Table 6 Error list
Error | Description | Resolution |
---|---|---|
Conductivity is too high. | The measured value is > 2 S/cm, 1,000,000 | |
ppm, 200% or 20,000 ppt. | Make sure that the Conductivity unit setting is set |
for the correct measurement range.
Conductivity is too low.| The measured value is < 0 µS/cm, 0 ppm, 0% or
0 ppt or the sensor cell constant is not correct.| Make sure that the sensor
is configured for the correct cell constant.
Zero is too high.| The zero calibration value is > 500,000 counts.| Make
sure that the sensor is held in air during zero calibration and is not located
near radio frequency or electromagnetic interference. Make sure that the cable
is shielded by metal conduit.
Zero is too low.| The zero calibration value is < –500,000 counts.
Temperature is too high.| The measured temperature is > 130 °C.| Make
sure that the correct temperature element is selected. Refer to Configure the
sensor on page 9.
Temperature is too low.| The measured temperature is < –10 °C.
ADC failure| The analog to digital conversion failed.| Power off and
power on the controller. Contact technical support.
Sensor is missing.| The sensor is missing or disconnected.| Examine the
wiring and connections for the sensor and for the module (or digital gateway).
Make sure that the terminal block is fully inserted into the module, if
applicable.
Measurement value is out of range.| The sensor signal is outside of
the accepted limits (2 S/cm).| Make sure that the Conductivity unit setting is
set for the correct measurement range.
7.5 Warning list
A warning does not affect the operation of menus, relays and outputs. The
screen changes to an amber color. The diagnostics bar shows the warning. Push
on the diagnostic bar to show the errors and warnings. As an alternative, push
the main menu icon, then select Notifications > Warnings. A list of possible
warnings is shown in Table 7.
Table 7 Warning list
Warning | Description | Resolution |
---|---|---|
Zero is too high. | The zero calibration value is >300,000 counts. | Make |
sure that the sensor is held in air during zero calibration and is not located
near radio frequency or electromagnetic interference. Make sure that the cable
is shielded by metal conduit.
Zero is too low.| The zero calibration value is < –300,000 counts.
Temperature is too high.| The measured temperature is > 100 °C.| Make
sure that the sensor is configured for the correct temperature element.
Temperature is too low.| The measured temperature is < 0 °C.
Calibration is overdue.| The Cal Reminder time has expired.| Calibrate
the sensor.
Warning| Description| Resolution
---|---|---
The device is not calibrated.| The sensor has not been calibrated.|
Calibrate the sensor.
Replace a sensor.| The Sensor days counter is more than the interval
selected for sensor replacement. Refer to Configure the sensor on page 9.|
Replace the sensor. Reset the Sensor days counter on the Diagnostics/Test >
Reset menu (or Diagnostics/Test > Counter menu.
Calibration is in progress…| A calibration was started but not
completed.| Return to calibration.
Outputs on hold| During calibration, the outputs were set to hold for a
selected time.| The outputs will become active after the selected time period.
As an alternative, remove and then supply power to the controller.
Linear temperature compensation is out of range.| The user-defined
linear temperature compensation is out of range.| The value must be between 0
and 4%/°C; 0 to 200 °C.
Temperature compensation table is out of range.| The user-defined
temperature compensation table is out of range.| The temperature is above or
below the temperature range defined by the table.
Incorrect user concentration table.| The concentration measurement is
outside of the range of the user table.| Make sure that the user table is set
for the correct measurement range.
Incorrect built-in temperature table.| The measured temperature is
outside of the range of the built-in temperature compensation table.| Make
sure that the temperature compensation is configured correctly.
Incorrect built-in concentration table.| The concentration measurement
is outside of the range of the built-in concentration table.| Make sure that
the concentration measurement is configured for the correct chemical and
range.
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. 970-669-3050
800-227-4224 (U.S.A. only)
Fax 970-669-2932
orders@hach.com
www.hach.com| HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
info-de@hach.com
www.de.hach.com| HACH LANGE Sarl
6, route de Compose
1222 Veenas
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499
---|---|---
© Hach Company/Hach Lange GmbH, 2010, 2014, 2018–2023.
All rights reserved. Printed in Germany.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>