Fisher Scientific 88861041 3D Platform Rotator Instruction Manual

June 14, 2024
Fisher Scientific

Fisher Scientific 88861041 3D Platform Rotator

Product Information

Specifications

  • Catalog Number: 88861041 (Nutating Mixer Fixed Speed 120V)
  • Catalog Number: 88861042 (Nutating Mixer Fixed Speed 230V)
  • Catalog Number: 88861043 (Nutating Mixer Variable Speed US)
  • Catalog Number: 88861044 (Nutating Mixer Variable Speed International)
  • Catalog Number: 88861045 (3D Platform Rotator US Plug)
  • Catalog Number: 88861046 (3D Platform Rotator International Plug)

Section 1: Inspection and Installation

Inspection and Installation

  1. Inspect package and contents upon receipt of the instruments. If the package is severely damaged or if there are any missing pieces, please contact the manufacturer immediately.
  2. Unpack the instrument, ensure all parts of the instrument and accessories are not missing or damaged. Make sure to take out all the components before discarding packing. If there are any missing or damaged pieces, please contact the manufacturer immediately.
  3. Place the instrument on a level and firm surface to avoid vibration and noise.
  4. This is the general user manual for three instruments. Please make sure to refer to the corresponding chapter for your instrument.

Packing List

Catalog Number Model Dimpled Rubber Mat General Power Adaptor
88861041 Nutating Mixer Fixed Speed 120V 1 1
88861042 Nutating Mixer Fixed Speed 230V 1 1
88861043 Nutating Mixer Variable Speed US 1 1
88861044 Nutating Mixer Variable Speed International 1 1
88861045 3D Platform Rotator US Plug N/A N/A
88861046 3D Platform Rotator International Plug N/A N/A

Section 2: Overview

Specifications

The specifications for each model are as follows:

  • Catalog Number: 88861041 (Nutating Mixer Fixed Speed 120V)
  • Catalog Number: 88861042 (Nutating Mixer Fixed Speed 230V)
  • Catalog Number: 88861043 (Nutating Mixer Variable Speed US)
  • Catalog Number: 88861044 (Nutating Mixer Variable Speed International)
  • Catalog Number: 88861045 (3D Platform Rotator US Plug)
  • Catalog Number: 88861046 (3D Platform Rotator International Plug)

Environmental Conditions

The product should be used under the following environmental conditions:

  • Temperature: 0°C to 40°C
  • Humidity: 20% to 80% RH (non-condensing)

Safety Instructions

Users should follow all operational guidelines and take necessary safety precautions while using this product to avoid bodily harm and property damage. Failure to follow these guidelines could result in potentially irreparable harm.

Loads and Speed

The product has specific load and speed limits. Please refer to the corresponding chapter for your instrument for more information.

Section 3: Operation

Operation of the Control Panel
The control panel of the product allows users to adjust various settings. Please refer to the corresponding chapter for your instrument for detailed instructions on operating the control panel.

Settings
The product offers various settings for operation. Please refer to the corresponding chapter for your instrument for detailed instructions on adjusting the settings.

Section 4: Safety Tips and Maintenance

Safety Tips
Follow these safety tips to ensure safe usage of the product:

  • Always wear appropriate protective equipment.
  • Keep the product away from water and other liquids.
  • Avoid placing heavy objects on the product.
  • Regularly inspect the product for any signs of damage or malfunction.

Maintenance

Maintenance of the product should be performed by qualified service personnel. Do not attempt to perform internal adjustments or maintenance yourself.

Section 5: Troubleshooting

If you encounter any issues with the product, please refer to the corresponding chapter for your instrument for troubleshooting steps. If the issue persists, contact the manufacturer for further assistance.

Section 6: Optional Accessories/Spare Parts

Optional accessories and spare parts are available for the product. Please contact the manufacturer for more information.

Section 7: Warranty

The product comes with a warranty. Please refer to the warranty information provided by the manufacturer for details on coverage and terms.

FAQ:

Q: Can I use the product under extreme temperatures?

A: The product should be used within the temperature range of 0°C to 40°C.

Q: Can I perform internal adjustments or maintenance myself?

A: No, all internal adjustments and maintenance must be performed by qualified service personnel.

Q: Are there any specific safety precautions I should take while using the product?

A: Yes, users should follow all operational guidelines and take necessary safety precautions while using this product to avoid bodily harm and property damage.

Q: Where can I find spare parts for the product?

A: Optional accessories and spare parts are available. Please contact the manufacturer for more information.

Important before using this product, read this entire operation manual carefully. Users should follow all of the operational guidelines contained in this manual and take all necessary safety precautions while using this product. Failure to follow these guidelines could result in potentially irreparable bodily harm and/or property damage.

  • Caution all internal adjustments and maintenance must be performed by qualified service personnel.
  • Material in this manual is for information purposes only. Fisher Scientific committed to a continuing program of product development and improvement, and reserves the right to change information, such as specifications, appearance, and dimensions, described in this document without notice. Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Fisher Scientific be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
  • No part of this manual may be reproduced or transmitted in any form or by any means, including photocopying, recording, or using information storage and retrieval systems, for any purpose other than the purchaser’s own use, without the express written permission of the manufacture.
  • Any other product names and services identified in this manual are trademarks or registered trademarks of their respective owners. No such use, or the use of any trade name, is intended to convey endorsement or other affiliation with Fisher Scientific.
    @2015 Fisher Scientific. All rights reserved.

Inspection and Installation

  1. Inspect package and contents upon receipt of the instruments. If the package is severely damaged or if there are any missing pieces, please contact the manufacturer immediately.
  2. Unpack the instrument, ensure all parts of the instrument and accessories are not missing or damaged. Make sure to take out all the components before discarding packing. If there are any missing or damaged pieces, please contact the manufacturer immediately.
  3. Place the instrument on a level and firm surface to avoid vibration and noise.
  4. This is the general user manual for three instruments. Please make sure to refer to the corresponding chapter for your instrument.

Packing List

Table -1 Packing List

Fisher-Scientific-88861041-3D-Platform-Rotator-fig-1

Connections

Nutating Mixer Fixed SpeedFisher-Scientific-88861041-3D-Platform-Rotator-
fig-2

Nutating Mixer Variable Speed/3D Platform Rotator

Fisher-Scientific-88861041-3D-Platform-Rotator-fig-3

Structure Diagram

Fisher-Scientific-88861041-3D-Platform-Rotator-fig-4

Overview

Specifications

Rotation Speed

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Variable Speed| 3D Platform Rotator
---|---|---|---
Cat. NO.| 88861041| 88861042| 88861043| 88861044| 88861045| 88861046
Speed Range| 24rpm| 20rpm| 2 to 60rpm| 2 to 80rpm
Display Mode| | LED
Display Accuracy| | 1 rpm
Speed Accuracy| ±1 rpm

Angle

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Variable Speed| 3D Platform Rotator
---|---|---|---
Tilt Angle| 20°| 7°
Tilt Angle Accuracy| ±1°
Rotation Angle| 0 to 360°

Note: Maximum speed/tilt angle may vary with heavy or unbalanced loads.

Load

Maximum Load (Centered on tray)

0.8kg

Time

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Variable Speed| 3D Platform Rotator
---|---|---|---
Timer Range| | 1min to 99h59min
Display Mode| | LED

Power Supply

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Variable Speed| 3D Platform Rotator
---|---|---|---
Cat. NO.| 88861041| 88861042| 88861043| 88861044| 88861045| 88861046
Requirement| 110V, 60z,

0.1A

| 220V, 50Hz,

0.05A

| 100~240V,

50Hz/60Hz,1A

| 100~240V,

50Hz/60Hz, 0.8A

Size

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Variable Speed| 3D Platform Rotator
---|---|---|---
Overall Dimensions| 415×293×273mm (16.3×11.6×10.7in.)| 415×295×170mm (16.3×11.6×6.7in.)
Tray Dimensions| 306×266mm(12.0×10.5in.)
Package Dimensions| 536×501×346mm(21.1×19.7×13.6in.)

Weight

Product Name| Nutating Mixer Fixed Speed| Nutating Mixer Varaioable Speed| 3D Platform Rotator
---|---|---|---
Net Weight| 5.9kg(13lb)| 7.4kg(16.3lb)| 7.3kg(16.1lb)
Gross Weight| 8.6kg(18.9lb)| 9.4kg(20.7lb)| 9.3kg(20.5lb)

Others

Noise Level ﹤55 dB with no load when ambient noise is 45 dB
Certificate ROHS, WEEE, cCSAus, CE Mark

Environmental Conditions
Application Environmental Conditions: indoor use

Temperature 5 to 40 ℃(41 to 104 ℉)
Humidity 20% to 85%,
Altitude ≤2,000 m
Voltage Fluctuation ±10% of the nominal voltage

Storage Environmental Conditions

Temperature 0 to 60 ℃(36 to 124 ℉)
Humidity 20% to 90%, non-condensing

Safety Instructions

Please read the entire instruction manual before operating the Nutating Mixer/3D Platform Rotator.

WARNING DO NOT use the Nutating Mixer/3D Platform Rotator in a hazardous atmosphere or with hazardous materials for which the unit was not designed. Also, the user should be aware that the protection provided by the equipment may be impaired if accessories used are not provided or recommended by the manufacturer, or are used in a manner not specified by the manufacturer.

CAUTION ! To avoid electrical shock, completely cut off power to the unit by disconnecting the power cord from the unit or unplug from the wall outlet.
Disconnect unit from the power supply prior to maintenance and service. Any spills should be removed promptly. Biohazard spills should be cleaned using approved liquid promptly. Solvent spills are a fire hazard. Stop the unit immediately, and DO NOT operate until clean up is complete and vapors have dissipated.
DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.

Position of Loads
Place the loads in the recommended positions below:

  1. Place load at the center of the tray
  2. Place loads symmetrically around the center of the tray
  3. Make sure to use rubber strips to fasten containers onto the tray Centered Placement

Asymmetric loading or unloading may affect the normal operation of the instrument and may even cause unpredictable damage.

Loads and Speed

Nutating Mixer Variable Speed

Cell Culture Flask Specification(cm2)| ≤Capacity(ml)| Maximum Speed

(rpm)

---|---|---
25| 30 / 40| 60
75| 225| 60
150| 450| 60
175| 600| 60
225| 800| 60

3D Platform Rotator

Cell Culture Flask Specification(cm2)| ≤Capacity(ml)| Maximum Speed

(rpm)

---|---|---
25| 30 / 40| 80
75| 225| 80
150| 450| 80
175| 600| 80
225| 800| 80

Warning: The rotation speed is inversely proportional to the load. When the instrument is in use, it is recommended to adjust the rotation speed from low to high, step by step and run the machine at an appropriate speed to avoid accidental spillage of liquids.
For safety, please always using rubber strip to hold samples in place.

Operation

This Chapter covers introduction of the control panel and its operation.

Operations of the Control Panel

Nutating Mixer Variable Speed/3D Platform Rotator

The front panel of the Nutating Mixer Variable Speed/3D Platform Rotator contains all the controls needed to operate the unit.

  1. Start/Stop button: Start or stop the instrument
  2. RUN indicator light: The light is on when the instrument is running and off when the instrument is in standby.
  3. TIME display window: The window shows cumulative time (in continuous mode) or remaining time (in timer mode). The range of time displayed is 0 to 99 hours and 59 minutes. The accuracy is 1 minute.
  4. SPEED display window: The window shows set speed (when the instrument is in standby) or current speed (when the instrument is running).
  5. Set Time Buttons: UP/DOWN Arrow buttons are used to increase/decrease the set time of the instrument.
  6. Set Speed Buttons: UP/DOWN Arrow buttons are used to increase/decrease the set speed of the instrument.

Nutating Mixer Fixed Speed

  1. RUN indicator light: The light is on when the instrument is running.

Preparation

  1. Connect all the components according to the figures shown on page 1-3 of this manual. Use grounded power outlet.
  2. Press power button on the back right side of the instrument and put it to the “|” state and then the instrument is in standby.

Settings
Nutating Mixer Variable Speed/3D Platform Rotator Time Settings

  1. Continuous mode
    Press the arrow button below the TIME display window. When the number shown on the display window starts flashing, press arrow button to decrease the time to 00:00 and then release the button. The time setting is finished after the number shown on the display window has flashed twice.

  2. Timer mode
    Press the arrow button below the TIME display window. When the number shown on the display window starts flashing, press arrow button to increase or decrease the time value. Release the button when the time shown on the display window reaches the set value. The time setting is finished after the number shown on the display window has flashed twice.

Speed Settings
Press the arrow button below the SPEED display window. When the number shown on the display window starts flashing, press arrow button to increase or decrease the speed value. Release the button when the speed shown on the display window reaches the set value. The speed setting is finished after the number shown on the display window has flashed twice.
Note: press the arrow button for a longer time to accelerate the setting.

Run and Stop

  1. Continuous Mode
    Press button and the instrument will start running with the specified settings and the RUN indicator light will be on. The TIME display window will show the cumulative time and the SPEED display window will show the current speed. Press button again and the instrument will slow down until it stops. The instrument will then be in standby and the two display windows will show the set values.

  2. Timer Mode
    Press button the instrument will start running with the specified settings and the RUN indicator light will be on. The TIME display window will show the remaining time and the SPEED display window will show the current speed. Press button again and the instrument will slow down until it stops. The instrument will then be in standby and the two display windows will show the set values.

Nutating Mixer Fixed Speed
Turn the power switch to “|” position. The RUN indicator light will be on and the instrument will start rotating at the fixed speed.

Finish Operation
After the operation is finished, please press the power switch at the back right side the instrument and put it into the “O” state. Unplug the instrument and store the instrument according to the storage guide.

Alarm System
When the normal operation is obstructed or stopped by malfunctioning, the instrument will stop running automatically within 2 minutes and the alarm will sound.

Power Recovery
If the power supply is cut off suddenly while the instrument is in operation, the unit will automatically run at the previously set parameter upon power restoration. The display window will flash. Press any button to stop flashing.

Safety Tips and Maintenance

Safety Tips

  1. Use independent power supply.
  2. Check if the local power supply voltage is suitable for use.
  3. Do not drag the power supply cable when unplugging.
  4. Do not use non-specified power cable nor damage cable.
  5. Service should only be performed by a qualified professional.
  6. The power supply must be unplugged under the following situations:
    • a. when the unit is moved
    • b. when the electrical cabinet or the moving component is opened
    • c. when the equipment is malfunctioning
    • d. when the equipment is not in use

Maintenance

  • a. This instrument uses a high-accuracy motor. It is maintenance free and has a long service time, high quality, and low noise level.
  • b. Surfaces can be cleaned with a mild detergent and water.Liquid is not allowed inside the dust cover.
  • c. Two fuses are installed inside the power socket at the back of the Nutating Mixer Fixed Speed (One is for work and the other is for backup). The installation of the fuses is shown in the figure below:
    Warning: Avoid dripping detergent or water into the instrument during cleaning.

Clean Spill
If accidental spillage of liquids caused by mishandling or contained break -age occurs on the surface of the instrument, please shut down the instrument and clean up the liquid immediately.
If the liquid has already spilled into the unit, please cut off the power supply first and immediately clean up the liquid at the surface of the instrument. Then place the instrument in a ventilated and dry environment for 24 hours before reuse. If the instrument is still not functioning after drying for 24 hours, please contact the manufacturer.

Warning: Disassembling/Assembling without a qualified professional’s guidance may cause malfunctioning of the instrument.

Troubleshooting

Please refer to the following table to troubleshoot if any malfunction occurs. If the problem still exists, contact your local sales representative.

Error Cause Solution

Cannot start machine, LED display window off

| Power disconnected| Connect the power
Power adaptor failure| Replace power adaptor
Switch off| Switch on

No shaking of the tray

| Electrical malfunction| Contact Fisher Scientific
Mechanical malfunction| Contact Fisher Scientific
Over-weighted or unbalanced load| Adjust the weight and position of load, decrease rotation speed

Loud noise

| Load moving| Change Dimpled Rubber Mat or fasten load
Tray loose| Fasten fixing screws on the tray

House loose

| Fasten fixing screws at the House
Other| Keep record for maintenance

Optional Accessories/ Spare Parts

Fisher-Scientific-88861041-3D-Platform-Rotator-fig-14

Warranty

When used in laboratory conditions and according to these operation instructions and maintenance, this product is warranted for 24 months against defective materials or workmanship. The 24 month warranty period begins from the delivery date of this product.
For product quality or performance issues, contact Fisher Scientific Customer Service.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Fisher Scientific User Manuals

Related Manuals