owon FDS1102A Series Channel 14 Bits Android Based Digital Storage
Oscilloscope
Product Information
Specifications
- Product Name: FDS Dual-Channel Series
- Edition: Aug. 2023 edition V1.0.4
- Trademark: LILLIPUT
- Manufacturer: Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
- Address: No. 19, Heming Road Lantian Industrial Zone, Zhangzhou 363005 P.R. China
- Contact: Tel: +86-596-2130430, Fax: +86-596-2109272
- Warranty: 3 years for the product, 12 months for accessoriessuch as probes
General Safety Requirements
To ensure safe usage of the product, please follow these general safety requirements:
- Read the user manual carefully before using the product.
- Do not expose the product to extreme temperatures or moisture.
- Avoid using the product in environments with high electromagnetic interference.
- Keep the product away from children and pets.
- Do not disassemble or modify the product without proper authorization.
Safety Terms and Symbols
The product uses safety terms and symbols to indicate potential hazards and precautions. Familiarize yourself with the following safety terms and symbols:
- DANGER: Indicates an imminent hazard that will result in severe injury or death if not avoided.
- WARNING: Indicates a potential hazard that could result in serious injury or death if not avoided.
- CAUTION: Indicates a potential hazard that could result in minor or moderate injury if not avoided.
- NOTICE: Provides important information or instructions regarding the product.
Primary User Guide
A general knowledge of the structure of the instrument
The instrument consists of a front panel and a rear panel:
- Front Panel: This panel contains various controls and indicators for operating the instrument.
- Rear Panel: This panel provides input/output ports and power connections.
A general knowledge of the user interface of the instrument
The user interface of the instrument includes various menus and settings
for configuring and controlling the device.
Power Output Inspection
To inspect the power output of the instrument, follow these steps:
- Connect the appropriate probes to the output ports.
- Select the power output inspection function from the menu.
- Adjust the settings as needed.
- Observe and analyze the power output waveform.
How to Implement the Probe Compensation
To implement probe compensation, follow these steps:
- Connect the probe to the designated input port.
- Select the probe compensation function from the menu.
- Follow the on-screen instructions to perform the compensation process.
How to Set the Probe Attenuation Coefficient
To set the probe attenuation coefficient, follow these steps:
- Select the probe attenuation coefficient setting from the menu.
- Choose the appropriate attenuation value based on the probe specifications.
- Apply the selected attenuation coefficient to the probe.
How to Use the Probe Safely
To use the probe safely, follow these guidelines:
- Ensure the probe is properly connected to the instrument.
- Avoid applying excessive voltage or current to the probe.
- Handle the probe with care to prevent damage.
- Follow proper safety procedures when working with high voltages.
How to conduct self-calibration
To conduct self-calibration of the instrument, follow these steps:
- Select the self-calibration function from the menu.
- Follow the on-screen instructions to perform the calibration process.
Use the Android System
The instrument features an Android system for additional functionality. The
Android system area is located on the front panel.
Use the Oscilloscope
A general Knowledge of Oscilloscope
An oscilloscope is a key feature of the instrument that allows users to
visualize and analyze electrical waveforms.
A general knowledge of vertical system
The vertical system controls the vertical position, scale, and triggering
of the waveform displayed on the oscilloscope.
A general Knowledge of Horizontal System
The horizontal system controls the timebase and triggering options for the
oscilloscope.
A general knowledge of Trigger System
The trigger system allows users to synchronize the oscilloscope’s waveform
display with specific trigger events.
How to use touch screen control
The instrument features a touch screen for intuitive control. Follow these
instructions to use touch screen control:
- Familiarize yourself with the touch screen gestures, such as tapping, swiping, and pinching.
- Use your finger to interact with the on-screen menus and controls.
Use the touch screen to operate the menu
To operate the menu using the touch screen, follow these steps:
- Tap on the desired menu option to select it.
- Swipe up or down to scroll through the menu options.
- Tap on sub-menus or settings to access further options.
Operate the touch screen
To operate the touch screen in waveform amplification mode, follow these steps:
- Select the waveform amplification mode from the menu.
- Pinch or spread your fingers on the touch screen to zoom in or out of the waveform.
- Drag your finger across the screen to navigate through the waveform.
Other touch screen operations
The touch screen supports additional operations, such as:
- Changing display settings
- Adjusting waveform parameters
- Enabling measurement functions
How to conduct automatic measurement
To conduct automatic measurements using the oscilloscope, follow these
steps:
- Select the automatic measurement function from the menu.
- Choose the desired measurement parameters.
- Initiate the automatic measurement process.
- Review the measurement results on the screen.
FAQ (Frequently Asked Questions)
Q: How long is the warranty period for the product?
A: The warranty period for the product is 3 years from the date of purchase
by the original purchaser.
Q: What is the warranty period for accessories such as probes?
A: The warranty period for accessories such as probes is 12 months.
Q: How do I obtain service under the warranty?
A: To obtain service under the warranty, you must notify our company of the
defect before the expiration of the warranty period. You will be responsible
for packaging and shipping the defective product to our designated service
centre, along with a copy of your proof of purchase.
Q: Are there any guarantees for maintenance beyond the warranty?
A: Except for the after-sales services provided in this summary or the
applicable warranty statements, we do not offer any guarantee for maintenance
definitely declared or hinted, including but not limited to the implied
guarantee for marketability and special-purpose acceptability. We should not
take any responsibilities for any indirect, special, or consequential damages.
FDS Dual-Channel Series QUICK_GUIDE
Aug. 2023 edition V1.0.4
Copyright © LILLIPUT Company. All rights reserved.
The LILLIPUT’s products are under the protection of the patent rights,
including ones which have already obtained the patent rights and those which
are applying for. The information in this manual will replace all that in the
materials published originally.
The information in this manual was correct at the time of printing. However,
LILLIPUT will continue to improve products and reserves the rights to change
specification at any time without notice.
is the registered trademark of the LILLIPUT Company.
Fujian LILLIPUT Optoelectronics Technology Co., Ltd. No. 19, Heming Road Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Fax: +86-596-2109272
General Warranty
We warrant that the product will be free from defects in materials and
workmanship for a period of 3 years from the date of purchase of the product
by the original purchaser from our company. The warranty period for
accessories such as probes is 12 months. This warranty only applies to the
original purchaser and is not transferable to a third party.
If the product proves defective during the warranty period, we will either
repair the defective product without charge for parts and labour, or will
provide a replacement in exchange for the defective product. Parts, modules
and replacement products used by our company for warranty work may be new or
reconditioned like new. All replaced parts, modules and products become the
property of our company.
To obtain service under this warranty, the customer must notify our company of
the defect before the expiration of the warranty period. Customer shall be
responsible for packaging and shipping the defective product to our designated
service centre, a copy of the customer’s proof of purchase is also required.
This warranty shall not apply to any defect, failure or damage caused by
improper use or improper or inadequate maintenance and care. We shall not be
obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than our company representatives to install,
repair or service the product; b) to repair damage resulting from improper use
or connection to incompatible equipment; c) to repair any damage or
malfunction caused by the use of not our supplies; or d) to service a product
that has been modified or integrated with other products when the effect of
such modification or integration increases the time or difficulty of servicing
the product.
Please contact the nearest sales and service offices for services.
Excepting the after-sales services provided in this summary or the applicable
warranty statements, we will not offer any guarantee for maintenance
definitely declared or hinted, including but not limited to the implied
guarantee for marketability and special-purpose acceptability. We should not
take any responsibilities for any indirect, special or consequent damages.
Table of Contents
1. General Safety Requirements . 1
2. Safety Terms and Symbols .3
How to conduct a general inspection .5 How to conduct function inspection . 6
3. Primary User Guide .7
A general knowledge of the structure of the instrument . 8 Front Panel . 8
Rear Panel . 9 A general knowledge of the user interface of the instrument .
10 Power Output Inspection . 11 How to Implement the Probe Compensation . 12
How to Set the Probe Attenuation Coefficient .13 How to Use the Probe Safely
.14 How to conduct self-calibration . 14
4. Use the Android System . 15
Front Panel Android System Area .15
5. Use the Oscilloscope .17
A general Knowledge of Oscilloscope . 17 Front Panel Oscilloscope Area . 17 A
general knowledge of vertical system .18 A general Knowledge of Horizontal
System . 19 A general knowledge of Trigger System . 20 How to use touch screen
control .20 Use the touch screen to operate the menu . 20 Operate the touch
screen .22 Operate the touch screen in waveform amplification mode . 24 Other
touch screen operations . 25 How to conduct automatic measurement . 28
6. Use Arbitrary Waveform/Function Generator . 29
Arbitrary waveform/function generator area of front panel . 29 Display window
of arbitrary waveform/function generator . 29 Setting Window of arbitrary
waveform/function generator . 30 Connect the output end . 30 Set the channel .
31 Set the waveform . 32 Output Built-in Waveform . 33
i
Output Modulation Waveform .34 Output Sweep Waveform . 34 Generate Burst . 35
7. Use the Power Supply .36
Power Supply Area on the Front Panel .36 Power Display Window . 36 Power
Setting Window . 37
8. Use the Multimeter . 38
Multimeter Area on the Front Panel . 38 Multimeter Zoom Box . 38 Multimeter
Setting Window .39
9. Appendix . 41
Appendix A: Enclosure . 41 Appendix B: General Care and Cleaning .41
ii
1.General Safety Requirements
1.General Safety Requirements
Before use, please read the following safety precautions to avoid any possible
bodily injury and to prevent this product or any other connected products from
damage. In order to avoid any contingent danger, ensure this product is only
used within the range specified.
Only the qualified technicians can implement the maintenance.
To avoid Fire or Personal Injury:
Connect the probe correctly. The grounding end of the probe corresponds to the
grounding phase. Please don’t connect the grounding end to the positive phase.
Use Proper Power Cord. Use only the power cord supplied with the product and
certified to use in your country.
Connect or Disconnect Correctly. When the probe or test lead is connected to a
voltage source, please do not connect and disconnect the probe or test lead at
random.
Product Grounded. This instrument is grounded through the power cord grounding
conductor. To avoid electric shock, the grounding conductor must be grounded.
The product must be grounded properly before any connection with its input or
output terminal.
When powered by AC power, it is not allowed to measure AC power source
directly, because the testing ground and power cord ground conductor are
connected together, otherwise, it will cause short circuit.
When powered by battery, the product must ground connection. To avoid electric
shock, there must be a ground wire connect between ground and the ground port
(on the back of product panel).
Check all Terminal Ratings. To avoid fire or shock hazard, check all ratings
and markers of this product. Refer to the user’s manual for more information
about ratings before connecting to the instrument.
Do not operate without covers. Do not operate the instrument with covers or
panels removed.
Use Proper Fuse. Use only the specified type and rating fuse for this
instrument.
Avoid exposed circuit. Do not touch exposed junctions and components when the
instrument is powered.
Do not operate if in any doubt. If you suspect damage occurs to the
instrument, have it inspected by qualified service personnel before further
operations.
Use your Oscilloscope in a well-ventilated area. Make sure the instrument
installed with proper ventilation, refer to the user manual for more details.
1
1.General Safety Requirements Do not operate in wet conditions. Do not operate
in an explosive atmosphere. Keep product surfaces clean and dry. When
measuring current, turn off the circuit power before connecting the
multimeter in the circuit. Remember to place the multimeter in series with the
circuit. When servicing the multimeter, use only the specified replacement
parts. Use caution when working above 60 V DC, 30 V AC RMS, or 42.4 V peak.
Such voltages pose a shock hazard. When using the test leads, keep your
fingers behind the finger guards on the test leads. Disconnect circuit power
and discharge all high-voltage capacitors before testing resistance,
continuity, diodes, or capacitance. Use the proper terminals, function, and
range for your measurements. When the range of the value to be measured is
unknownchoose the auto ranging mode. To avoid damages to the multimeter, do
not exceed the maximum limits of the input values shown in the technical
specification tables. Connect the common test lead before you connect the live
test lead. When you disconnect the leads, disconnect the live test lead first.
Before changing functions, disconnect the test leads from the circuit under
test.
2
2.Safety Terms and Symbols
2.Safety Terms and Symbols
Safety Terms
Terms in this manual. The following terms may appear in this manual:
Warning: Warning indicates the conditions or practices that could result in
injury or loss of life.
Caution: Caution indicates the conditions or practices that could result in
damage to this product or other property.
Terms on the product. The following terms may appear on this product: Danger:
It indicates an injury or hazard may immediately happen. Warning: It indicates
an injury or hazard may be accessible potentially. Caution: It indicates a
potential damage to the instrument or other property might occur.
Safety Symbols
Symbols on the product. The following symbol may appear on the product:
Hazardous Voltage
Protective Earth Terminal
Chassis Ground
Direct current (DC)
Alternating (AC)
current
Both direct and alternating current
Conforms
to
European
Union
directives
Equipment protected
throughout by double
insulation
or
reinforced insulation
3
Test Ground
Fuse
Caution, risk of danger (refer to this manual for specific Warning or Caution
information)
Category II overvoltage protection
Category III protection
overvoltage
Category IV overvoltage protection
2.Safety Terms and Symbols To avoid body damage and prevent product and
connected equipment damage, carefully read the following safety information
before using the test tool. This product can only be used in the specified
applications.
Warning:
The two channels of the oscilloscope are not electrically isolated. The
channels should adopt a common ground during measuring. To prevent short
circuits, the 2 probe grounds must not be connected to 2 different non-
isolated DC levels.
Warning:
Note when measuring the channel to the public base, otherwise may cause short
circuit because of the oscilloscope probe ground wire. The diagram of the
oscilloscope ground wire connection:
The diagram of the ground wire connection when the battery-powered
oscilloscope is connected to the AC-powered PC through the ports:
It is not allowed to measure AC power when the oscilloscope is AC powered, or
when the battery-powered oscilloscope is connected to the AC-powered PC
through the ports.
4
2.Safety Terms and Symbols
Warning:
To avoid fire or electrical shock, when the oscilloscope input signal
connected is more than 42V peak (30Vrms) or on circuits of more than 4800VA,
please take note of below items: Only use accessory insulated voltage probes
and test lead. Check the accessories such as probe before use and replace it
if there are any damages. Remove probes, test leads and other accessories
immediately
after use. Remove USB cable which connects oscilloscope and
computer. Do not apply input voltages above the rating of the instrument
because the probe tip voltage will directly transmit to the oscilloscope. Use
with caution when the probe is set as 1:1. Do not use exposed metal BNC or
banana plug connectors. Do not insert metal objects into connectors.
How to conduct a general inspection
After you get a new device, it is recommended that you should make a check on
the instrument according to the following steps:
1. Check whether there is any damage caused by transportation. If it is found
that the packaging carton or the foamed plastic protection cushion has
suffered serious damage, do not throw it away first till the complete device
and its accessories succeed in the electrical and mechanical property tests.
2. Check the Accessories The supplied accessories have been already described
in the “Appendix A: Enclosure” of this Manual. You can check whether there is
any loss of accessories with reference to this description. If it is found
that there is any accessory lost or damaged, please get in touch with our
distributor responsible for this service or our local offices.
3. Check the Complete Instrument If it is found that there is damage to the
appearance of the instrument, or the instrument can not work normally, or
fails in the performance test, please get in touch with our distributor
responsible for this business or our
5
2.Safety Terms and Symbols
local offices. If there is damage to the instrument caused by the
transportation, please keep the package. With the transportation department or
our distributor responsible for this business informed about it, a repairing
or replacement of the instrument will be arranged by us.
How to conduct function inspection
Conduct a quick function inspection to confirm if the instrument is running in normal state.
Power-on Inspection
To power on the instrument, long press host.
in the lower left of the
The instrument screen shows the startup screen, will display Android screen and wait a few seconds, the relay in the host will make a slight click sound. After performing all self-inspection items on the instrument, the instrument will directly enter instrument system.
6
3.Primary User Guide
3.Primary User Guide
This chapter elaborates the following topics: A general knowledge of the
structure of the instrument A general knowledge of the user interface of the
instrument How to conduct probe compensation How to set the probe attenuation
coefficient How to use the probe safely How to conduct self-calibration
7
3.Primary User Guide
A general knowledge of the structure of the instrument
This chapter gives a brief description and introduction to the operations and
functions of the front panel of the instrument, so as to facilitate your
operations of the instrument in the shortest time.
Front Panel
On the instrument panel, knobs and function buttons are used to set different
options for current menu, and other buttons are used to enter different
function menus or directly obtain specific function application.
Figure 3- 1 Front panel 1. Display area. 2. Touch key. 3. Android function
keys. 4. General knob and arrow keys. 5. Oscilloscope key and knob control
area. 6. Probe compensation: 5V/1kHz signal output. 7. Multimeter keys and
port control area. 8. Signal generator key and port control area. 9. USB Host
Interface: When the oscilloscope is connected to an external
USB device as a “master device”, the USB Host interface is used to transmit
the data. 10.Power key and output port control area. 11.Headphone socket.
12.Instrument switch with memory (self-lock) switch and auto memory of last
shutdown; if the instrument is shut down by powering the supply off, there is
no need to press the switch to start it up after next power-on; if the
instrument is shut down by pressing the switch key, it is required to press
8
3.Primary User Guide the switch key again to power it on.
Rear Panel
Figure 3- 2Rear Panel 1. Foldable Handle. 2. Heat Emission Hole. 3. Power
Outlet. 4. Fuse . 5. Foot Rest: To adjust the inclined angle of the
oscilloscope. 6. HDMI Interface: To connect HDMI output to the external
monitor or
projector. 7. USB Host Interface: When the oscilloscope is connected to an
external
USB device as a “master device”, the USB Host interface is used to transmit
the data. 8. USB Device Interface: When the oscilloscope is connected to an
external USB device as a “slave device”, the USB Device interface is used to
transmit the data. For example, use the interface to connect a PC or printer.
9. LAN Interface: The network interface to connect a PC or router. 10.External
Trigger Input Interface (EXT Trig In). 11.Trig Out (P/F) Interface: Trigger
output or pass/fail output port.
9
3.Primary User Guide
A general knowledge of the user interface of the instrument
Figure 3- 3Illustrative Drawing of Display Interfaces
1. Shortcut soft keys of oscilloscope functions.
2. It indicates that a USB Device cable is inserted currently.
3. It indicates that a network cable is connected, that is, the LAN port is
accessed(If icon for the
it indicates the Wi-Fi is enabled and
connected currently).
4. It indicates that a USB flash drive is inserted currently.
5. Display the system set time.
6. Display window of signal generator.
7. Power display window.
8. Multimeter display window.
9. Main menu, click to pop up the oscilloscope and general function menu window.
10. Information display bar of oscilloscope horizontal, acquire and trigger functions.
11. Information display bar of oscilloscope channel.
12. Waveform display area.
10
3.Primary User Guide 13. Display current operating status.
Oscilloscope Inspection
1. Set the Switch in the Oscilloscope Probe as 10X and Connect the
Oscilloscope with CH1 Channel. Align the slot in the probe with the plug in
the CH1 connector BNC, and then tighten the probe with rotating it to the
right side. Connect the probe tip and the ground clamp to the connector of the
probe compensator.
2. Push the Autoset Button at the top left of the screen or on the front
panel. The square wave of 1 kHz frequency and 5V peak-peak value will be
displayed in several seconds (see Figure 3- 4).
Figure 3- 4Auto set Check CH2 by repeating Step 2 and Step 3.
Power Output Inspection
Output inspection is to ensure that the instrument can achieve its rated
outputs and properly respond to operation from the front panel. 1. Voltage
Output Inspection
11
3.Primary User Guide
(1) When the instrument is under no load, select a channel and ensure the
output current setting for this channel is not at zero.
(2) Turn on the channel output, then ensure the channel is in Constant Voltage
output mode.
(3) Set some different voltage values on this channel; check if the actual
voltage value displayed is close to the set voltage value, and also that the
actual current value displayed is nearly to zero.
(4) Check CH1 and CH2 channel if the output voltage can be adjusted from 0.1V
to the maximum rating.
2. Current Output Inspection The following steps check basic current
functions with a short across the power supply’s output: (1) Power on. (2)
Connect a short across (+) and (-) output terminals with an insulated test
lead on this channel. Use a wire size sufficient to handle the maximum
current. (3) Set the output voltage to the maximum rating on this channel. (4)
Turn on the channel output. Ensure the channel you used is in Constant Current
output mode. (5) Set some different current values on this channel; check if
the actual current value displayed is close to the set current value, and to
check if the actual voltage value displayed is nearly zero. (6) Check CH1 and
CH2 channel if the output current can be adjusted from zero to the maximum
rating, When it is set to the maximum or minimum, a beep is heard, indicating
that the limit has been reached.. (7) Turn off the channel output and remove
the short circuit from the output terminals.
How to Implement the Probe Compensation
When connect the probe with any input channel for the first time, make this
adjustment to match the probe with the input channel. The probe which is not
compensated or presents a compensation deviation will result in the measuring
error or mistake. For adjusting the probe compensation, please carry out the
following steps:
1. Set the attenuation coefficient of the probe in the menu as 10X and that
of the switch in the probe as 10X (see “How to Set the Probe Attenuation
Coefficient” on P13), and connect the probe with the CH1 channel. If a probe
hook tip is used, ensure that it keeps in close touch with the probe.
12
3.Primary User Guide
Connect the probe tip with the signal connector of the probe compensator and
connect the reference wire clamp with the ground wire connector of the probe
connector, and then push the Autoset button on the front panel. 2. Check the
displayed waveforms and regulate the probe till a correct compensation is
achieved (see Figure 3- 5 and Figure 3- 6).
Overcompensated
Compensated correctly
Under
compensated
Figure 3- 5 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if needed.
Figure 3- 6 Adjust Probe
How to Set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the
vertical scale factor of the oscilloscope. To change or check the probe
attenuation coefficient in the menu of oscilloscope: (1) Push the function
menu button of the used channels(CH1 channel or CH2
channel). (2) Select Probe Attenu (1X , 10X or other custom probe
magnifications) in
the displayed channel setting window. The setting will remain in effect until
changed again after selection This setting will be valid all the time before
it is changed again.
13
3.Primary User Guide
Caution: The default attenuation coefficient of the probe on the instrument is
preset to 10X. Make sure that the set value of the attenuation switch in the
probe is the same as the menu selection of the probe attenuation coefficient
in the oscilloscope. The set values of the probe switch are 1X and 10X (see
Figure 3- 7).
Figure 3- 7 Attenuation Switch Caution: When the attenuation switch is set to
1X, the probe will limit the bandwidth of the oscilloscope in 5MHz. To use the
full bandwidth of the oscilloscope, the switch must be set to 10X.
How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against any
electric shock, shown as Figure 3- 8.
Figure 3- 8 Finger Guard
Warning: To avoid electric shock, always keep your finger behind the safety
guard ring of the probe during the operation. To protect you from suffering
from the electric shock, do not touch any metal part of the probe tip when it
is connected to the power supply. Before making any measurements, always
connect the probe to the instrument and connect the ground terminal to the
earth.
How to conduct self-calibration
Self-calibration program is used to quickly make the oscilloscope be in the
optimum condition to obtain the most accurate measurement. This program can be
performed at any time. It is especially necessary when the ambient
14
4.Use the Android System temperature reaches or exceeds 5. To conduct self-
calibration, disconnect all probes and wires from the input connector. Then,
press Settings key at the oscilloscope area of the front panel or click ” ” in
the lower right of the screen, select Oscilloscope from the right menu and
Self-calibration from the menu displayed, and click Start in the self-
calibration display box.
4.Use the Android System Front Panel Android System Area
Four keys are contained at the front panel area of the instrument. They are
Back, Home, Task and Touch. Note: The Touch Lock is only available in EduInstr
system and does not work on the main homepage and other application interface.
Android System Homepage Window
15
4.Use the Android System 1. Application shortcut key 2. Task key 3. App Drawer
(Click to see all apps) 4. Home key 5. Back key
System built-in application list
Open the application drawer of the main page. There are 17 built-in
applications in the system, including: App Center, Calculator, Calendar,
Chrome, Browser, Clock, Email, File Explorer, Gallery, Instruments, Music,
Pydroid3, Screen Recorder, Settings, Sound Recorder, USB Camera, Web Service
and WPS Office, as illustrated in the figure below:
16
5.Use the Oscilloscope
5.Use the Oscilloscope A general Knowledge of Oscilloscope
Front Panel Oscilloscope Area
Instructions:
1. Trigger state.
2. Oscilloscope function shortcut keys.
3. General knob and arrow keys.
General knob: When
appears in the screen menu, it indicates that
the General knob can be turned to select current menu or set the value.
Arrow keys: Move to select the parameter.
4. Function key area: There are 9 keys in total.
5. Shortcut key: Run/Stop,Signal.
6. Trigger control area:
Contain two keys and one knob.
“Trigger Level” knob is to adjust the trigger level.
Menu button correspond to the settings of the trigger system;
Force button is the forced trigger shortcut button.
7. Vertical control area:
Contain two keys and four knobs.
In the oscilloscope state, the “CH1” key and “CH2” key correspond to the channel switch of Channel 1 and Channel 2, respectively. Two “Vertical
Displacement” knobs are to control the vertical displacement of Channel 1 and Channel 2, respectively. And two “Gear” knobs are used to control the voltage of Channel 1 and Channel 2, respectively
17
5.Use the Oscilloscope
8. Horizontal control area: Contain one key and two knobs. In the
oscilloscope state, the “Horizontal Menu” key corresponds to the setup menu of
the horizontal system, the “Horizontal Displacement” knob is to control the
horizontal displacement triggered, when pressing the knob horizontal
displacement can be zero and the “Gear” knob is used to control the time base
gear.
9. Main menu button: Click to display/hide the main menu, including the
oscilloscope and general function menu list.
10. Oscilloscope function information box: Display horizontal, acquisition,
and trigger information content separately. Click to display/hide
corresponding function setting window.
11. Vertical channel information box: Display the configuration information
content of channel 1 and channel 2 respectively. The information content
includes:channel switch, voltage gear, zero position, bandwidth limit, etc.In
the display content, BW indicates that the bandwidth limit is 20MHz. The icon
indicates the coupling mode of the channel: ” ” indicates DC coupling ” ”
indicates AC coupling ” ” indicates ground coupling
A general knowledge of vertical system
As shown in Figure 5-1, there are two buttons and four knobs in the Vertical
Control Area. The following exercises are to guide you through the use of the
vertical settings.
Figure 5- 1:Vertical Control Area 1. Press the Vertical Displacement knob to
center display the signal in the
waveform window. It is used to control the vertical display position of the
signal. When turning the Vertical Displacement knob, the pointer
18
5.Use the Oscilloscope
indicating the Grounding Reference Point of the channel moves up and down
following the waveform. When press the Vertical Displacement button, the
vertical displacement can be zero. Measuring Skills If the DC coupling mode is
adopted for the channel, observe the gap between the waveform and the signal
ground to quickly measure the DC component of the signal. If the AC coupling
mode is adopted for the channel, the DC component of the signal is filtered
out, facilitating you to display the AC component of the signal with higher
sensitivity. 2. Change the vertical settings and observe the resulting status
change. You can determine the change of the vertical gear factor of any
channel through the information displayed in the status bar at the lower part
of the waveform window. Turn the Vertical Gear knob to change the Vertical
Gear Factor
(Voltage Gear), and the gear factor of corresponding channel in the status bar
changes accordingly. 3. Press the CH1 and CH2 key to enable or disenable
corresponding channel.
A general Knowledge of Horizontal System
As shown in Figure 5-2,there is a key and two knobs in the Horizontal Contral
Area. The following exercises are to guide you through the use of the
horizontal time base settings.
Figure 5- 2:Horizontal Control Area 1. Turn the Horizontal Gear knob to change
the settings of the horizontal
time base and thus observe the resulting change of the status information,
then the corresponding Horizontal Time Base in the status bar changes
accordingly. Press the Horizontal Gear knob enter or exit waveform
amplification mode. 2. Turn the Horizontal Displacement knob to adjust the
horizontal displacement of the signal in the waveform window. 3. The
Horizontal Displacement knob is to control the triggered horizontal
displacement of the signal; when turning the knob, the waveform moves
horizontally with the knob.When press the Horizontal Displacement button, the
horizontal displacement can be zero. 4. Press the HOR key to pop up the
horizontal Settings window.
19
5.Use the Oscilloscope
A general knowledge of Trigger System
As shown in Figure 5-3, there is a knob and two keys in the Trigger Control
Area. The following exercises are to guide you through the settings of the
trigger system.
Figure 5- 3: Trigger Control Area 1. Press the Trigger Menu key to call the
trigger menu, and operate the
menu displayed to change the trigger settings. 2. Use Trigger Level knob to
change the settings of the trigger level.
Turn the Trigger Level knob, the trigger pointer on the screen moves up and
down as the knob turns. While moving the trigger pointer, the value of the
trigger level on the screen changes accordingly. Note: Turning the Trigger
Level knob can not only change the trigger level value, but also set the
shortcut key of the trigger level at the vertical midpoint of the trigger
signal amplitude. 3. Press Forced Trigger key to generate a trigger signal
forcibly, mainly used in the “Normal” and “Single” trigger mode.
How to use touch screen control
The touch screen can be used to control the oscilloscope by various gestures.
Operate the touch screen when the light of the touch lock in the upper left of
the front panel is off; click the key to light up the indicator; the touch
function is disabled when the touch lock is in locked state. The instructions
of the touch screen operations are as follows, the contents in the parentheses
indicate the key or knob that plays the same role.
Use the touch screen to operate the menu
Open the Settings window function: Directly click menu items in the lower menu
bar or the right menu bar of the display area to pop up the setting window of
the corresponding function.
Set Menu Item: In the Settings window, you can change the configuration of the
relevant menu item by touching it. The types of operable parts include:
switch, button, radio, gear hobbing (scrolling list), etc. The
20
5.Use the Oscilloscope following box selects the radio type, click directly to switch the options.
Scroll List: When the scroll bar appears in the menu or file system window,
swipe the screen up and down with the finger to scroll the list, as shown in
the figure below.
Open Main Menu: Click the
icon in the lower right of the display area,
the main menu window pops up, as shown below, with each item
corresponding to the keys with the same name on the panel or the shortcut
keys at the upper part of the screen. Clicking the item is equivalent to
pressing the corresponding key on the panel or the shortcut key at the
upper part of the screen.
21
5.Use the Oscilloscope
Operate the touch screen
Select a Channel (CH1 channel or CH2 channel): Click the channel pointer at
the left side or the channel waveform to select the channel point.
Set the vertical position of the selected channel waveform (Vertical
Displacement knob): Waveform area on the right side of two for trigger level
touch mobile area, slide up and down in the area can change the trigger level,
as shown in the figure below.
Set the trigger level of the signal source in the Trigger Menu (Trigger Level
knob): Swipe the screen up and down with the finger near the trigger area two
squares to the right side, as shown in the figure below.
22
5.Use the Oscilloscope
Set Horizontal Position (Horizontal Displacement knob): Swipe the screen left
and right with the finger in the waveform display area, as shown in the figure
below.
Control the voltage gear and time base by pinch action: Pinch the thumb and
the index finger up and down/left and right in the waveform display area, as
shown in the figure below. 23
5.Use the Oscilloscope
Operate the touch screen in waveform amplification mode
Press Horizontal Gear Knob to enter the waveform amplification mode, the main
window is displayed at the top half of the screen and the amplified window is
displayed at the bottom half of the screen. The amplified window is the
amplified part of the main window that is selected.
24
5.Use the Oscilloscope
Other touch screen operations
Click and drag the open menu item to move itself to the appropriate location.
Control the horizontal or vertical cursor lines under cursor measurement: 25
5.Use the Oscilloscope
Run/Stop: Click or in the upper left of the display area to switch Run/Stop.
Parameter Setting Keyboard in Menu Item: There are digital input mode and gear
input mode. Digital input mode: Click to input.
Gear input mode, as shown in the figure below: 26
5.Use the Oscilloscope 27
5.Use the Oscilloscope
How to conduct automatic measurement
Push the Measure button to display the menu for the settings of the Automatic
Measurements. At most 8 types of measurements could be displayed on the bottom
left of the screen.
The oscilloscopes provide 43 parameters for auto measurement, including
Period, Period, Frequency, + Width, – Width, Rise Time, Fall Time, Positive
Duty Ratio, Negative Duty Ratio , Screen Pulse Width Ratio, Average Value,
Peak-to-Peak Value, Root-mean-square Value, Maximum Value, Minimum Value, Top
Value, Bottom Value, Amplitude, Pre-shoot, Over-shoot , Cycle Root-mean-square
Value,Standard deviationNumber of Positive Pulse, Number of Negative Pulse,
Number of Rising Edge, Number of Falling Edge, Area , Cycle Area,Delayed (1 -2
)Delayed (1 -2 )Delayed (1 -2 ) Delayed (1 -2 ) Phase Position (1 -2 ) Phase
Position (1 -2 ) Phase Position (1 -2 ) Phase Position (1 -2 )FRR (1 -2 )FRF
(1 -2 )FFR (1 -2 )FFF (1 -2 )LRR (1 -2 )LRF (1 -2 )LFR (1 -2 )LFF (1 -2 ).
The waveform channel must be opened for measurement. The automatic measurement
can not be performed in the following situation: 1) On the saved waveform. 2)
On the Dual Wfm Math waveform.
For example: To measure the cycle and frequency of CH1 channel signal, operate
according to the following steps:
1. Press Measure key and the automatic measurement menu is displayed on the
screen.
2. Click the Switch to On state and the circle is highlighted. 3. Click
Signal Source CH1 to highlight it. 4. Click Cycle and Frequency in the
horizontal measurement, and the
circle is checked and highlighted.
The measured values will be automatically displayed at the lower left of the
screen.
6.Use Arbitrary Waveform/Function Generator
6.Use Arbitrary Waveform/Function Generator
Arbitrary waveform/function generator area of front panel
Display window of arbitrary waveform/function generator
The display window of arbitrary waveform/function generator is located at the
upper right of the screen. Refer to Channel 1 for the descriptions of Channel
2.
Descriptions: 1. Display the channel name and the channel output switch
status.
6.Use Arbitrary Waveform/Function Generator 2. Display the frequency/cycle. 3.
Display the amplitude/high level. 4. Display the offset/low level. 5. Initial
phase . 6. Figure A: Display the pulsewidth/duty ratio. 7. Figure A: Display
the current waveform. 8.Figure B:Display the current modulation mode.
Setting Window of arbitrary waveform/function generator
The display window of arbitrary waveform/function generator is located at the
upper right of the screen. Refer to Channel 1 for the descriptions of Channel
2.
Descriptions: 1. Select the channel. 2. Enable or disable the channel output.
3. Select the load: high resistance or Custom load(Range is a to 10k,default
is 50). 4. Waveform selection area. 5. Output parameter setting area.
Connect the output end
Connect BNC cable to the output end of Channel 1 or Channel 2 signal generator
of the oscilloscope’s front panel. Press the On/Off keys of two channels. To
view the output of the signal
6.Use Arbitrary Waveform/Function Generator
generator, connect the other end of BNC cable to the signal input channel of
the oscilloscope’s front panel.
Set the channel
How to switch the channels displayed in the menu Open the signal source
setting window and click CH1, CH2 or CH1/2 at the upper of the window to
switch among Channel 1 menu, Channel 2 menu and Channel Copy menu.
How to enable/disable the channel output Press On/Off key of two channels to
enable/disable the output of corresponding channel. The key light of
corresponding channel turns on when the output is enabled. Channel copy menu
Press CH1/CH2 softkey to switch to the Channel Copy menu.
Align Phase Click Align Phase softkey in the menu below to align the initial
phases of two channel signals. Copy channel Select From CH2CH1 in the menu
below to copy the parameters of Channel 2 to Channel 1. Select From CH1CH2 in
the menu below to copy the parameters of Channel 1 to Channel 2. Frequency
synchronization Select Frequency Sync as On in the menu below. When adjusting
the frequency of one channel, that of the other changes simultaneously.
Amplitude synchronization Select Amplitude Sync as On in the menu below. When
adjusting the amplitude of one channel, that of the other changes
simultaneously.
6.Use Arbitrary Waveform/Function Generator
Set the waveform
(1) Click CH 1 or CH 2 display window to display the setting window of the
signal generator.
(2) Select the waveform required and corresponding waveform setting menu from
the menu items at the lower part of the window.
(3) Operate the setting menu to set the parameters of the waveform required.
Example: Press Frequency in the menu below (if there is only Period other than
Frequency, click the title to switch it to Frequency), and set the required
value, the specific methods are as follows. There are 3 methods to change the
parameter values selected: Use digital soft keyboard input mode: Click Numeric
Display Box and
the soft keyboard appears, and input the required value directly.
6.Use Arbitrary Waveform/Function Generator Use gear input mode:
Use General Knob: Turn General Knob to increase or decrease the value
at the cursor. Press
/
to move the cursor left and right, see
the gear input mode diagram for details. Parameters that can be set for each waveform
Nane Sine
Square
Ramp
Pulse Noise Arbitrary
Menu Items Frequency/Period, Amplitude/High Level, Offset/Low Level, Start Phase Frequency/Period, Amplitude/High Level, Offset/Low Level, Start Phase Frequency/Period, Amplitude/High Level, Offset/Low Level, Start Phase, Symmetry Frequency/Period, Amplitude/High Level, Offset/Low Level, Start Phase, Pulse width/Duty Cycle Amplitude/High Level, Offset/Low Level Frequency/Period, Amplitude/High Level, Offset/Low Level, Start Phase, Built- in/External
Output Built-in Waveform
There are 28 kinds of built-in waveforms. To select the built-in waveform, follow the steps below:
6.Use Arbitrary Waveform/Function Generator
(1) Click CH 1 or CH 2 display window to display the setting window of the
signal generator.
(2) Select Arbitrary Wave in the menu below and click Built-in External. (3)
Select the classification of the built-in wave in the menu: Engineering,
Maths, Medical, Trigonometric or Others. For example, select Others. (4) Click
DC to output the DC voltage waveform.
Output Modulation Waveform
Supported modulation types include: Amplitude Modulation (AM), Frequency Modulation (FM), Phase Modulation (PM) and Frequency-shift Keying (FSK).
Press Setting key in the signal source area of the panel or directly click the signal source display window at the right side of the screen to pop up the
signal source setting window, click Composite and select Modulation; click Type menu to select the modulation type. To disable the modulation, pop up the signal source setting window again, select Continue in the Composite. Parameters that can be set for various modulation types
Type AM FM PM FSK
Settings Mod.WaveMod.FreqMod.Depth Mod.WaveMod.FreqFrequency deviation Mod.WaveMod.FreqPhase.Dev Mod.FreqHop Freq
Output Sweep Waveform
In the sweep mode, the output varies from the start frequency to stop frequency within the specific scan time. Only sine wave, square wave, sawtooth wave or arbitrary wave (except DC) can be used to product sweep waveforms. When the output signal is sine wave, square wave, ramp wave or arbitrary wave, press Setting key at the signal source area of the front panel or directly press the signal source display window at the right side of the screen to pop up the signal source setting window, then click Sweep in the Composite to enter the sweep mode. Parameters that can be set include sweep time, Linear/Log, Start Freq/Ctr Freq, Stop Freq/Freq Span, and trigger source. When the trigger source is switched to manual, the soft key is displayed at the bottom. Click the soft key to perform a frequency sweep trigger. To turn off the sweep, call up the signal
6.Use Arbitrary Waveform/Function Generator
source setting window again and select continuous in the composite.
Generate Burst
Press Setting function key at the signal source area of the front panel or
directly click the signal source display window at the right side of the
screen to pop up the signal source setting window, click Burst in the
Composite to generate pulse train waveform output of various waveform
functions. The pulse trains can last a specific number of waveform cycles (N
cycle pulse trains). Sine wave, square wave, sawtooth wave, pulse wave or
arbitrary wave functions can be used (this function is unavailable for noise
wave).
7.Use the Power Supply
7.Use the Power Supply
Power Supply Area on the Front Panel
Power Display Window
The power display window is located in the middle right of the screen. Please
refer to Channel 1 for the descriptions of Channel 2.
7.Use the Power Supply Descriptions: 1. Display channel name and channel
output switch status. 2. Actual Output Value of Voltage. 3. Actual Output
Value of Current. 4. Actual Output Power. 5. Set Values of Output
Voltage/Current. 6. Display Constant Voltage/Constant Current.
CV: Constant Voltage Output; CC: Constant Current Output.
Power Setting Window
Click Screen Power Display window or press Setting key at the power supply
area on the screen to display/hide the power measurement setting window.
Descriptions: 1. Set current function mode. 2. Select CH1 or CH2 as the
channel output parameter. 3. Click to enable/disable measured value output. 4.
Voltage/Current setting area. 5. Overvoltage/Overcurrent protection Settings.
Use the Multimeter
Multimeter Area on the Front Panel
In the Multimeter area on the front panel of the instrument, there is 1 multimeter setting key and 4 multimeter input ends. The input end of the multimeter is located at the lower right of the front panel,
with 4 sockets marked
, COM, A mA and 20A, respectively.
Figure 8- 1: Multimeter Area
Multimeter Zoom Box
The multimeter Zoom Box is located at the lower right of the screen.
Figure 8- 2: Multimeter Zoom Box Descriptions: 1. Display window drag icon. 2.
Measuring type indication:
AC —— AC Voltage Measurement
8.Use the Multimeter DC —— AC Voltage Measurement ACI —— AC Current
Measurement DCI —— DC Current Measurement
—— Capacitance Measurement —— Resistance Measurement
—— On-off Measurement —— Diode Measurement 3. Measured display value; if the
measured value is out of range, “OL” is displayed. 4. Current Range. 5. Manual
Range/Automatic Range Indication: Manual indicates manual range; Auto
indicates automatic range. 6. Zoom Frame Off key.
Multimeter Setting Window
Click Multimeter display window on the screen or press Setting key at the
multimeter area of the front panel to display/hide the multimeter measurement
setting window.
Figure 8- 3Multimeter setting Window
8.Use the Multimeter
Function Settings Zoom box
Soft Key
Descriptions
mV, V Measure the voltage.
Measure the capacitance.
uA, mA, A
AC/DC Hold REL Auto
Measure the current.
Measure
the
resistance/on-off
testing/measure the diode.
Switch AC/DC upon voltage or current measurement.
Keep the current reading on the display.
Reading is the difference between the stored reference value and the input signal.
Select auto range mode.
Manual
Select manual range mode and press to switch the range.
Open/Close the parameter display zoom box.
Appendix
9.Appendix
Appendix A: Enclosure
(The accessories subject to final delivery.)
Standard Accessories:
Power Cord
Quick Guide
USB Cable
Probe
Probe Adjust
Multimeter Lead BNC to alligator banana plug to clip Q9 cable alligator clip
cable
Appendix B: General Care and Cleaning
General Care Do not store or leave the instrument where the liquid crystal
display will be exposed to direct sunlight for long periods of time.
Caution: To avoid any damage to the instrument or probe, do not exposed it to
any sprays, liquids, or solvents.
Cleaning Inspect the instrument and probes as often as operating conditions
require. To clean the instrument exterior, perform the following steps:
1. Wipe the dust from the instrument and probe surface with a soft cloth. Do
not make any scuffing on the transparent LCD protection screen when clean the
LCD screen.
2. Disconnect power before cleaning your Oscilloscope. Clean the instrument
with a wet soft cloth not dripping water. It is recommended to scrub with soft
detergent or fresh water. To avoid damage to the instrument or probe, do not
use any corrosive chemical cleaning agent.
9.Appendix
Warning: Before power on again for operation, it is required to confirm that
the instrument has already been dried completely, avoiding any electrical
short circuit or bodily injury resulting form the moisture.