Menuett 013698 Fever Thermometer Instruction Manual
- June 4, 2024
- Menuett
Table of Contents
Menuett 013698 Fever Thermometer
SAFETY INSTRUCTIONS
Read these instructions carefully before use, and save them for future reference.
WARNING!
-
The product can be used irrespective of the age or gender of the user.
-
The product is only intended for household use.
-
The product cannot replace consultation with a donor.
Consult a do or if you need to interpret the readings or are concerned about your health status. -
Do not allow children to use the product without supervision – it contains small parts that can be swallowed.
-
Product is not watertight – never immerse the product in water or any other liquid.
-
Never modify the product in any way without the express approval of the manufacturer.
-
Do not expose the product to temperatures below -25°C or more than SS°C, or for humidity higher than 95% RH.
-
Store batteries out of the reach of children.
-
Remove the batteries if the product is not going to be used for some time.
ELECTROMAGNETIC COMPATIBILITY
WARNING
- The product must not be used near or on top of other electronic equipment, such as mobile phones, communication radio equipment or radio control equipment. If the product must be used near or on top of other electronic equipment it must be monitored so that normal operation can be guaranteed in the configuration in which it is used.
- The use of spare parts, accessories or cords other than those recommended or sold by the manufacturer can increase the electromagnetic radiation of the product or reduce its electromagnetic immunity.
SYMBOLER
TECHNICAL DATA
Measurement range
- Forehead temperature measuring 34.0 to 43.0°C (93.2 to 109.4°F)
- Object temperature measuring Oto 100°( (32 to 212°F
Accuracy in laboratory
- Forehead temperature measuring: ±0.2°C (0.4°F) in measurement value interval 35.5 to 42.0°C (95.9 to 107.6°F) at ambient temperature 15 to 35°C. ±0.3°C (0.5°F) outside these temperature intervals.
- Object temperature measuring: the higher of ±4% and ±2°C (4°F).
Clinical accuracy
- Clinical systematic deviation: -0.09°C(-0.16°F)
- Clinical repeatability: 0.13°C (0.23 F)
- Compliance limit: 0.87°C (1.57°F)
The product complies with requirements in MDD 93/42/EEC,
approved by notified body no. 0197.
Manufacturer
JOYTECH HEALTHCARE Co., Ltd. No.365, Wuzhou Road, Yuhang
Economic
Development Zone, Hangzhou city,
311100 Zhejiang, China
Telephone no: +86-571-81957767
Fax: +86-571-81957750
European representative
Shanghai International Holding Corp. GmbH (Europe), Eiffestrasse
80, 20537 Hamburg, Germany.
DESCRIPTION
The fever thermometer DET-306 is an IR thermometer intended for measuring body temperature in the forehead through detection of the IR radiation emitted from the forehead. The fever thermometer converts the measured IR radiation intensity to a temperature value, which is shown on the display. The fever thermometer is intended for intermittent body temperature measuring on the skin of the forehead for people of all ages. The correct use of the fever thermometer quickly gives an accurate body temperature value.
PARTS
- Probe
- Start button
- Power switch
- Button to select type of measurement
- Battery cover
FIG. 1
DISPLAY
- Forehead temperature measuring
- Object temperature measuring
- Time and date
- AM/PM indicator for time
- Low battery level indicator
- Memory mode
- Unit: °C or °F
- Fever indicator
- Reading
FIG. 2
FUNCTIONS
- Clock with date reading.
- The time and dates of the temperature measurements in the memory make it easier to follow the body temperature curve.
- Set the clock as per the instructions in the section “Setting the time and date” before using the thermometer.
- Forehead temperature measuring
- The thermometer is only intended to be used as a practical aid. It cannot replace consultation with a doctor. Always compare the readings with your normal body temperature.
- See the instructions for body temperature measuring in the section “How to use”.
- Object temperature measuring
- In object temperature measuring mode there is no correction of the measured temperature as in the case of body temperature measuring on the forehead, the actual measured surface temperature is shown instead. This temperature measuring mode can be used to check that the object, preparation etc. has a suitable temperature, for example that milk has the right temperature for a baby.
- See the instructions for object temperature measuring in the section “How to use”
- Audio alarm.
- If a temperature higher than 37.8°( (100.0 °F) is detected in forehead temperature measuring mode the product gives a fever alarm in the form of an audio prompt with one long and three short beeps.
- Readings memory.
- Up to 10 forehead and object temperature measurements can be saved in the memory. The measurements are saved with time readings.
- °C and °F switch.
- Instructions to switch between displaying in degrees Celsius and degrees Fahrenheit are given in the section “Selection of temperature.
ADVANTAGES OF TEMPERATURE MEASURING ON THE FOREHEAD
Measuring of the forehead temperature with IR technology gives a reliable
value of the temperature of the body’s vital organs, see figure 3. The body
temperature is measured by measuring the skin temperature over the temporal
arteries (temple arteries). The temporal arteries are linked to the heart via
the carotid artery, which in turn is in direct connection with the aorta. The
flow of blood through the temporal arteries is therefore constant. Changes to
the inner body temperature are therefore reflected more quickly in the
forehead than in other accessible parts of the body, for example the oral
cavity, rectum and armpit.
- Vital organ
FIG. 3
USE
SETTING THE TIME ANS DATE
The time and date must be set before using the fever thermometer for the
first time. Check that the thermometer is switched off and then go to settings
mode by pressing the button marked for three seconds.
FIG. 4
-
Choice of 12/24 hour display:
The time can be shown in 12 hour format with AM/PM indicator, or in 24 hour format. Tap the start button to switch between display formats. When the required display format is set, press the button marked . The numbers for hours start flashing.
FIG. 5 -
Setting the hour:
- Step through the hours by repeatedly pressing the start button until the correct hour is shown.
- When the correct hour is set, press the button marked The numbers for minutes start flashing.
FIG. 6
-
Setting the minutes
- Step through the minutes by repeatedly pressing the start button until the correct minutes are shown.
- When the correct minutes are set, press the button marked. The numbers for the year start flashing.
FIG. 7
-
Setting the year
- Step through the years by repeatedly pressing the start button until the correct year is shown.
- When the correct year is set, press the button marked. The numbers for the month start flashing.
FIG. 8
-
Setting the month
- Step through the months by repeatedly pressing the start button until the correct month is shown.
- When the correct month is set, press the button marked. The numbers for the day (today’s date) start flashing.
FIG. 9
-
Setting the day
- Step through the days by repeatedly pressing the start button until the correct day is shown.
- When the correct day is set, press the button marked to finish and exit settings mode.
FIG. 10
MEASURING TIPS
The tips below contribute to accurate and reliable measurements.
-
A measurement of a person’s body temperature does not in itself indicate fever – it is the deviation from the person’s normal body temperature that determines whether fever is present. It is therefore necessary to chart the person’s body temperature profile in good health and in normal conditions. This is done by measuring the body temperature on several occasions when the person is healthy. Verify if possible these measurements with a reference thermometer.
-
The person should have been indoors for at least 30 minutes before the measurements are made.
NOTE:
Both the person and the thermometer should have been at the same ambient temperature for at least 10 minutes before the measurements are made. -
The person should not drink, eat, attend to personal hygiene, use a hair dryer or be physically active before or during the measurements. Remove any headgear and move away any hair hanging over the forehead at least 10 minutes before measuring.
-
Wash the forehead clean from cream, cosmetics and impurities 10 minutes before measuring-the reading can be too low if the skin is not completely clean.
-
Avoid holding a hand against the forehead other than very briefly as this can result in an incorrect reading. Do not measure against scarred tissue, scratches or sores.
-
Do not measure against sweaty skin,
this can result in an incorrect reading. -
Do not measure during or immediately after breast-feeding.
-
Do not measure outdoors.
-
Do not measure near sources of heat, such as fireplaces or stoves.
-
The front lens on the probe is the most sensitive part of the thermometer.
-
Do not touch the front lens on the probe. The accuracy can be reduced if the front lens is damaged or contaminated.
-
If the fever thermometer is stored in a place that is significantly different from where it is to be used it should be left at rest for at least 30 minutes before the first measurement.
-
The fever thermometer is not intended to be used in an atmosphere with high oxygen content or an environment where flammable mixtures of anesthesia gas and air, oxygen or nitrous oxide (laughing gas) can be present.
FOREHEAD TEMPERATURE MEASURING
-
Press the button marked. The display is activated and all the display segments light up briefly as a function check. After the function check the display switches to show the image as per figure 11 and gives audio prompts to indicate that the fever thermometer is ready to start a new measurement.
FIG. 11 -
Point the probe at the center of the forehead, with the tip of the probe two to three cm from the surface of the forehead (see figure 12) and press the start button. Hold the probe still in the measuring position over the forehead until an audio prompt indicates that the measurement is done. If several measurements are to be made in sequence, wait three seconds between the measurements for the most accurate results.
FIG. 12 -
Read off the measured temperature on the display.
-
Switch off the thermometer by pressing the button marked.
SWITCHING BETWEEN FOREHEAD TEMPERATURE
MEASURING & OBJECT TEMPERATUR **MEASURING**
Press the button (I) to switch between forehead temperature measuring and object temperature measuring.
OBJECT TEMPERATURE MEASURING
-
Press the button (I) to start the fever thermometer. The temperature can be measured after two audio prompts (see figure).
FIC. 13 -
Point the probe at the center of the object to be measured, with the tip of the probe one to two cm from the surface of the object.
-
Press the start button and read off the temperature on the display.
-
Switch off the product by pressing the button marked.
AFTER MEASURING
- Switching off: the fever thermometer switches off automatically after 60 seconds of inactivity to save the batteries.
- Clean the probe after use to ensure reliable results and to reduce the risk of contagion (see section “Maintenance”).
FEVER INDICATOR
The fever indicator is active for forehead temperature measuring.
- If the measured temperature is less than 37.8°C (100.0°F) a laughing face symbol is shown G beside the temperature value.
- If the measured temperature is 37.8°( (100.0°F) or higher a sad face symbol is shown G beside the temperature value.
READINGS MEMORY
-
The readings memory is accessible from both forehead temperature measuring mode and object temperature mode.
When the fever thermometer is started and the display has run a self-test and forehead temperature measuring (display as per figure 11) or object temperature measuring (display as per figure 13) has been selected, press the button marked for three seconds. The text Mis shown on the top right of the display (see figure 14).
FIG. 14 -
The fever thermometer will now automatically store the next ten measured temperatures. The measurements are saved with time readings. Press the button marked repeatedly to browse through the stored measurement values in the sequence one to ten. Number 1 is the most recent and number 10 is the oldest of the stored measurements (see figure 15).
FIG. 15 -
In memory mode the symbol iii is always shown for forehead temperature measuring or the symbol ill for object temperature measuring. Press the start button to do new measurements.
SELECTION OF TEMPERATURE UNIT
- The measured temperature can be shown in degrees Celsius (°C) or degrees Fahrenheit (°F).
- Check that the thermometer is switched off and then press the start button for 3 seconds to activate selection of the temperature unit.
- Tap the start button to switch between temperature units.
- When the required temperature unit is selected, press the button to exit temperature selection mode.
MAINTENANCE
REPLACING THE BATTERY
-
The batteries should be replaced when the symbol ex:> is shown in the top right corner of the display.
-
Remove the battery cover by pushing it down.
FIG. 16 -
Take out the used batteries and insert 2 new alkaline batteries, size AAA (LR03).
FIG. 17 -
Replace the battery cover.
CLEANING
- The front lens on the probe must be kept clean, dry and undamaged for accurate and reliable measurement results. The accuracy can be reduced if the front lens is damaged or covered with dust, fingerprints, ear wax or other impurities. A contaminated or damaged sensor can also cause other problems.
- The front lens on the probe is the most sensitive part of the thermometer. Disinfect the fever thermometer and its probe with a soft cloth moistened with a 75% solution of isopropanol. Do not use abrasive detergent. Allow to dry for at least 10 minutes before next temperature measurement.
- Clean the fever thermometer display and casing with a soft, dry cloth.
- Strong electromagnetic fields can interfere with the functionality of the fever thermometer. Special conditions concerning electromagnetic interference must be complied with, see section “Electromagnetic compatibility”.
WARNING!
- No chemicals other than isopropanol must be used to clean the front lens on the probe.
- The fever thermometer is not watertight. Do not immerse the thermometer in water.
STORAGE
- The fever thermometer should be stored dry, clean and free from dust, and protected from direct sunlight.
- Store the fever thermometer in the original packaging when not in use.
CALIBRATION
The fever thermometer is supplied calibrated from the factory.
No regular recalibration is necessary when used in accordance with these
instructions. It is recommended to calibrate the fever thermometer every two
years, or if its clinical accuracy is doubtful. The thermometer should then be
submitted to the manufacturer or retailer for recalibration.
The above recommendation does not replace any legal requirements. Legislation,
rules and regulations, and guidelines for the use, functionality and accuracy
of the fever thermometer should always be complied with.
A clinical summary and procedure to check the calibration of the fever thermometer can be supplied on request (start the thermometer and press the power switch until it switches to calibration mode, whereby the installed programme version is shown).
TROUBLESHOOTING
ELECTROMAGNETIC COMPATIBILITY
The product complies with the EMC requirements in accordance with the international standard IEC 60601-1-2 if the specified conditions in the table below are met. The product is an electrical medical product and as such is subject to special requirements concerning electromagnetic compatibility, which requirements must be indicated in the instructions for the product. Portable and mobile radio communications equipment can affect the product. Do not use the product together with unapproved accessories -this can have a negative effect on the product and change its electromagnetic compatibility. Do not use the product in the immediate vicinity of electrical equipment or between different electrical equipment.
GUIDELINES & MANUFACTURER’S CERTIFICATE CONCERNING ELECTROMAGNETIC
RADIATION
The product is intended to be used in electromagnetic environments in
accordance with the specifications below. It is incumbent on the customer or
the person using the product to ensure that the electromagnetic environment in
which the product is used complies with these specifications.
GUIDELINES & MANUFACTURER’S CERTIFICATE CONCERNING ELECTROMAGNETIC
IMMUNITY
The product is intended to be used in electromagnetic environments in
accordance with the specifications below. It is incumbent on the customer or
the person using the product to ensure that the electromagnetic environment in
which the product is used complies with these specifications.
NOTE: OUT is the mains voltage before the test level is applied.
NOTE:
- The higher frequency range is applicable at 80 MHz and 800 MHz.
- These guidelines do not apply in all situations. Electromagnetic radiation can be affected by absorption in and reflection against building structures and other objects, such as people.
- In the frequency range 0.15-80 MHz there are the following ISM bands (industrial, scientific and medical): 6.765-6.795 MHz, 13.553-13.567 MHz, 26.957-27.283 MHz and 40.66-40.70 MHz. In the frequency range 0.15-80 MHz there are the following communication radio bands: 1.8-2.0 MHz, 3.5-4.0 MHz, 5.3-5.4 MHz, 7-7.3 MHz, 10.1-10.15 MHz, 14-14.2 MHz, 18.07-18.17 MHz, 21.0-21.4 MHz, 24.89-24.99 MHz, 28.0-29.7 MHz and 50.0-54.0 MHz.
- The compliance levels in the ISM bands between 150 kHz and 80 MHz and in the frequency band between 80 MHz and 2.7 GHz are intended to reduce the probability that the product suffers interference from mobile phones and portable communication radio equipment which by mistake is taken into the patient area. For this reason an extra factor of 10/3 has been inserted in the formulas for calculation of the recommended minimum distance to equipment that radiates radio frequency energy in these frequency bands.
- The field strength from permanent transmitters, e.g. base stations for mobile phones and wireless telephony, communication radios, amplitude or frequency modulated radio and television broadcasting, cannot be predicted theoretically with a practical degree of accuracy. In those cases where electromagnetic environmental impact from permanent radio transmitters is important, measurement of electromagnetic field strength in situ should therefore be taken into consideration. If the measured field strength in the place where the product is used exceeds the applicable, above specified compliance levels for radio frequency radiation, the product should be tested and checked to verify that it is working properly. Additional measures may be necessary if incorrect or abnormal functionality is observed, for example the realignment or repositioning of the product.
- In the frequency range 150 kHz to 80 MHz the field strength should be less than 3 V/m.
RECOMMENDED MINIMUM DISTANCE BETWEEN
THE PRODUCT & PORTABLE & MOBILE RADIO COMMUNICATION EQUIPMENT
The product is intended to be used in electromagnetic environments in which radiated radio frequency interference is controlled and limited. The customer or person using the product can reduce the risk of electromagnetic interference by maintaining the recommended minimum distance, given below on the basis of radio transmitted power, between the product and portable and mobile radio communication equipment.
For radio transmitters where the rated power P is not given in the table values the minimum distanced can be calculated with the applicable equation for the transmitter frequency, whereby the transmitting power Pis given in W.
NOTE:
- The minimum distance for the higher frequency range is applicable at so MHz and BOO MHz.
- These guidelines do not apply in all situations. Electromagnetic radiation can be affected by absorption in and reflection against building structures and other objects, such as people.
RECOMMENDED MINIMUM DISTANCE BETWEEN THE PRODUCT & PORTABLE & MOBILE
RADIO **COMMUNICATION EQUIPMENT**
The product is intended to be used in electromagnetic environments in which
radiated radio frequency interference is controlled and limited. The customer
or person using the product can reduce the risk of electromagnetic
interference by maintaining the recommended minimum distance, given below
on the basis of radio transmitted power, between the product and portable and
mobile radio communication equipment.
NOTE:
These guidelines do not apply in all situations. Electromagnetic radiation can
be affected by absorption in and reflection against building structures and
other objects, such as people.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>