heidolph Hei-SHAKE Multi Reax Instruction Manual
- June 13, 2024
- heidolph
Table of Contents
Hei-SHAKE Multi Reax
Instruction Manual
Introduction
About this document
This operating manual describes the features and operation of vortexers, Hei-
SHAKE Multi Reax type. The operating instructions are an integral part of the
described device!
Typographic conventions
Standardized symbols, signal words and highlighting elements are used in this
document to warn of hazards and to identify important information or rather
special text contents.
Symbol | Signal word / explanation |
---|---|
Warning symbols in combination with a signal word indicate dangers: |
DANGER
Indicates an immediate dangerous situation. Failure to respect the indications
will exult in death or serious
injury.
WARNING indicates a potentials danger. Failure to respect the
indications will result in serious injuries. CAUTION Indicates a potential
hazard which, If not voided, damage to property and minor to moderate juries
can occur.
| Mandatory signs are used to indicate important and useful information on
handling a product. This information is used to ensure operational safety and
to maintain the value of the product.
[GUI]| Parameter designations, display texts, and device labels are
typographically highlighted in the text and in tables to facilitate their
assignment on the device.
➜| The arrow indicates specific instructions to be followed to ensure
operational safety when handling the product.
Copyright protection
This document is protected by copyright and is generally intended for use by
the purchaser of the product.
No transfer to third parties, reproduction in any form, including excerpts,
and by any means, as well as utilization and/or disclosure of the contents is
permitted without the prior written consent of Heidolph Instruments GmbH & Co.
KG. Any violation is subject to compensation for damage.
Safety instructions in the official languages of the European
Union
A summary of all safety instructions in the official languages of the European
Union can be found in our Safety Guide for the product group shakers & mixers
(Ref. 01-005-006-99). This document is available for download on our homepage
in the most up-to-date version.
General product information
Directives applied, product certification
CE marking
The device meets all requirements of the following directives:
- European Machinery Directive, 2006/42/EC
- EMC Directive, 2004/108/EC
California Residents
Important information for California residents regarding Prop 65. Please visit
www.P65Warnings.ca.gov for more information.
Residual risk
The device has been designed and manufactured in accordance with the state-of-
the-art standards at the time of development and the recognized safety
regulations. During mounting and use, as well as during maintenance, repair
and cleaning work, there are nevertheless certain residual risksassociated
with the described device.
These are identified and described at the appropriate points in this document.
Intended use
The described vortexer in this document has been specially developed for the
following tasks:
- Shaking
- Mixing
- Emulsifying
- Suspending
- Dissolving
The areas of application of the described device include chemical, biological
and environmental-analytical laboratory and research applications,
applications of basic research and comparable facilities.
Due to its design, the device in its delivery condition may only be used in
analytical processes or in laboratory-like conditions in the food, cosmetics,
and pharmaceutical industries as well as other comparable industries that
manufacture products intended for consumption by humans or animals, or for use
on humans or animals. Any other use of this device is not considered as
intended!
Compliant use
The user is generally responsible for evaluating the conformity of his
application and, if necessary, for taking additional measures.
Reasonably foreseeable misuse
Additional measures may be necessary for use under conditions or for purposes
deviating from the intended use, and/or specific guidelines and safety
regulations must be observed.
Corresponding requirements must be evaluated and implemented by the operator
in each individual case.
Compliance with and implementation of all relevant directives and safety
measures for the respective field of application is generally the
responsibility of the operator.
All risks resulting from improper use are borne generally by the operator.
The device may generally be operated by authorized and instructed personnel.
Training and qualification of the operating personnel as well as ensuring that
the device is handled responsibly are the general responsibility of the
operator!
Transportation
During transportation, avoid severe shocks and mechanical stresses that can
cause damage to the device. Keep the original packaging in a dry and protected
place for later use. Storage
Always store the device in its original packaging. To protect against damage
and excessive material aging, store the device in an environment that is as
dry, temperature-stable and dustfree as possible.
Acclimatization
After each transportation and after storage under critical climatic conditions
(e.g. high temperature difference between inside and outside), allow the
device to acclimatize at room temperature for at least two hours to prevent
possible damage due to condensation before commissioning it in the place of
use. If necessary, extend the acclimatization phase if the temperature
differences are very high.
Make all supply connections (power supply, tubing) only after the device has
been acclimatized!
Permissible ambient conditions
The device is designed for indoor use only. The device is NOT suitable for
outdoor use! The device is NOT suitable for use in potentially explosive
areas!
When used in corrosive atmospheres, the service life of the device may be
reduced depending on the concentration, duration and frequency of exposure.
Safety
General safety instructions
- Before commissioning and using the device, familiarize yourself with all the safety regulations and occupational safety guidelines applicable at the place of use and observe them at all times.
- Only operate the device if it is in faultless technical condition. In particular, ensure that there is no visible damage on the device itself and, where applicable, on connected devices or the supply connections.
- If there is missing or misleading information on the device or regarding occupational safety, contact the responsible safety specialist or our technical service.
- Only use the device in accordance with the regulations for intended use.
Safety
Electrical safety
- Before connecting the device to the power supply, ensure that the voltage indicated on the rating plate matches the specifications of the local power utility company.
- Ensure that the power supply circuit provided is protected by means of a residual-current device (RCD).
- Always use the three-pole power supply cord supplied with the device.
- Prior to use, check that the device and the power supply cord are free of visible damage.
- Always have repairs and/or maintenance work on the device carried out by an authorized electrician or by the technical service department of Heidolph Instruments.
- Always switch off and disconnect the device from the power supply, preventing reconnection, before carrying out maintenance work, cleaning, or repairs.
Operational safety
- Operate the device under a closed ventilated fume hood when working with potentially hazardous substances (see EN 14175 and DIN 12924).
- Do not make any unauthorized changes or modifications to the device!
- Only use genuine spare parts and accessories, or those expressly approved by the manufacturer!
- Rectify malfunctions or faults on the device immediately.
- Switch off and disconnect the device from the power supply, preventing reconnection, if it is not possible to eliminate the malfunction or rectify the fault immediately.
- Observe all other applicable regulations such as laboratory and workplace guidelines, recognized safety technology rules and special local regulations.
Occupational safety
-
Always use the prescribed personal protective equipment (PPE) such as protective clothing, safety goggles, protective gloves, safety shoes, etc.
-
Do not operate any other devices in the immediate vicinity of the device … – which can generate electromagnetic fields in the frequency range between 9 × 10 3 Hz to 3 × 10 11 Hz,
– which generate emission or radiation sources in the frequency range 3 × 10 11 Hz to 3 × 10 15 Hz (in the optical spectral range wavelengths from 1,000 μm to 0,1 μm),
– which generate ultrasonic or ionizing waves. -
Do not process any substances that could release energy in an uncontrolled manner (e.g. self-ignition).
-
Do not process substances in which the energy input through mixing poses a danger.
-
Wipe off any liquid that may have spilled on the device immediately.
Personal protective equipment (PPE)
The operating company must determine and provide the necessary PPE, depending
on the respective application and the media and chemicals used.
The corresponding instruction of the personnel is generally within the
operating company’s responsibility.
Environmental protection
When processing environmentally hazardous substances, take appropriate
measures to avoid hazards to the environment.
The evaluation of corresponding measures such as the marking of a hazardous
area, their implementation, and the training of the relevant personnel is the
general responsibility of the operator!
Biohazard
When processing biohazardous substances, take appropriate measures to prevent
hazards to persons and the environment, including:
- Instruction of the personnel regarding the necessary safety measures.
- Provision of personal protective equipment (PPE) and instruction of the personnel in its use.
- Marking of the device with the biohazard warning symbol.
The evaluation of corresponding measures such as the marking of a hazardous
area, their implementation, and the training of the relevant personnel is the
general responsibility of the operator!
Other regulations
In addition to the notes and instructions in this document, observe all other
applicable regulations such as laboratory and workplace guidelines, hazardous
substances ordinances, recognized rules of safety engineering and occupational
medicine as well as particular local regulations!
Noncompliance will invalidate any warranty claims against Heidolph
Instruments. The operator is generally liable for all damage resulting from
unauthorized changes or modifications to the device, from the use of
unapproved or non-genuine spare parts and accessories, or from disregarding
the safety instructions and hazard warnings or the manufacturer’s
instructions!
Device description
Mechanical design
| Legend
---|---
1| Device base
2| On/off switch (on the side of the housing)
3| Device display
4| Start/Stop button with LED
5| Shift button [Select] to switch between the shaking frequency display and
the time, with LED
6| Arrow keys [up]/[down] to increase/reduce the displayed value
7| Control for the shaking frequency (motor speed)
8| Adapter for twelve vessels, illustration by way of example
Commissioning
Set up the device
CAUTION
To use the device, place it on a clean, stable, level and horizontal surface.
Make sure that the power plug can be reached directly at any time, in order to
disconnect the device from the power supply at any time pitchout delay.
Before switching on, make sure that the device is sufficiently stable. Before
switching on, always make sure that all setups and vessels are fixed on the
base unit with sufficient stability. Keep all support and contact surfaces
clean and dry.
The operator is generally liable for all damage resulting from unauthorized
changes or modifications to the device, from the use of unapproved or non-
genuine spare parts and accessories, or from disregarding the safety
instructions and hazard warnings or the manufacturer’s instructions!
Noncompliance will invalidate any warranty against Heidolph Instruments.
Power supply
DANGER
Please note the safety instructions in section „Electrical safety“ on page 25.
Connect the power supply cord
➜ Before connecting the power supply cord, make sure that the main switch of
the device is in position [0] (off).
➜ Connect the cable coupling of the supplied cord to the IEC appliance inlet
on the back of the device.
➜ Connect the power supply cord to a properly secured power socket.
Disconnecting the power supply cord
➜ Before disconnecting the power supply cord, switch off the device.
➜ Disconnect the power supply cord from the power socket.
➜ Disconnect the power supply cord from the IEC appliance inlet on the back of
the device.
Fixing laboratory vessels
The device can be operated with an adapter for 26 vessels (diameter of up
to 16 mm) or an adapter for twelve vessels (diameter of 16 – 32 mm).
Before replacing the assembled adapter, remove all vessels, pull the adapter
carefully up and place the new adapter on the attachment. Please consider the
positions of the two guides!
Further information can be found on our website at
www.heidolph.com.
WARNING
Whenever possible, use closed vessels for processing corrosive, toxic or
biohazardous substances and seal them safely. Increase the speed only
gradually to the desired shaking frequency, especially with open vessels, and
observe the fluid movements. Lower the shaking frequency or reduce the total
load on the platform when the device starts to vibrate during operation.
Always use the necessary and appropriate personal protective equipment!
➜ Always note the even and homogeneous distribution of the vessels in the
adapter: ➜ Align all vessels in such way that the outer ring is placed in the
center. ➜ Group vessels of a different diameter and leave a space of at least
one attachment between vessels with a different diameter:
Operation
Switching the device on/off
Use the main switch (on the side of the device’s housing) to switch the device
on or off. When switched on, the button lights up green.
Continuous operation
➜ After switching on the device, press the two arrow keys on the control panel
simultaneously (see section „Mechanical design“ on page 27).
– The value [Time] is set to [000].
➜ Press the [Start/Stop] button.
– The device is started in continuous operation, the LEDs of the buttons
[Start] and [Speed] light up green.
➜ Define the desired shaking frequency (motor speed) in a range of 150 – 2,000
rpm by means of the speed controller.
– The set value is shown on the display.
– The shaking frequency can be adjusted at any time during operation.
➜ Press the [Start/Stop] button again to deactivate the shaking operation
Timer mode
In timer mode, a specific period of time can be set for a continuous shaking
movement.
➜ After switching on the device, set the desired running time in a range of 1
– 99 minutes (see section „Mechanical design“ on page 27) by means of the two
arrow keys :
– Tap the [up/down] button to increase/reduce the set value gradually.
– Press and hold the [up/down] button to increase/reduce the display value in
fast forward.
– Press the two buttons simultaneously to set the value directly to [000].
➜ After having set the desired time, press the [Start/Stop] button.
– The device is started in Timer mode with the last set shaking frequency, the
LEDs of the buttons [Start] and [Time] light up green, the shaking frequency
is shown on the display with four digits.
– The shaking frequency can be adjusted at any time during operation: press
the button [Select] to switch to the shaking frequency display.The LED [Speed]
lights up, the LED [Time] flashes.
– The expired time is shown on the display with three digits.
Press the [Start/Stop] button to interrupt the shaking operation if necessary.
The timer stops and continues as soon as the shaking operation is restarted
with [Start/Stop].
In case of a mains interruption, the timer will be reset and has to be
adjusted again!
- The timer value can be adjusted at any time during operation. The initial setting remains until the value is adjusted again or the device is switched off.
- After the functional sequence of the set time has expired, an acoustic signal sounds and the shaking movement stops.
Troubleshooting
Failure | Cause | Possible remedy |
---|
Display remains dark after switching on
Shaking movement does not start| No mains voltage| Check if the on/off switch
is in position ON
Check the power supply cord for damages/correct connection
Check device fuses
Check building fuses Power supply cord defective
On/off switch in position OFF
Display defective| Contact Heidolph Sales or Service
Shaking movement stops| Timer expired| Restart timer or change operating mode
Motor overheat protection has tripped| Wait about 20 minutes and reduce the
load
Electrical fault (no engine noise)| Contact Heidolph Sales or Service
Mechanical defect (engine noise audible)| Contact Heidolph Sales or Service
In case of recurring errors please contact the responsible sales department or our technical service (see section „Contact details“ on page 34).
Appendix
Technical specifications
Hei-SHAKE Multi Reax|
---|---
Dimensions (W × H × D)| 270 × 172 × 410 mm
Weight| approx. 10 kg
Permissible load| max. 1.5 kg
Motion| circular vibrating
Orbit| 3 mm
Operating modes| Continuous operation, Timer mode
Speed range| 150 – 2,000 rpm
Rotation speed indicator| digital
Drive| Asynchronous motor
Overheat protection| self-resetting
Protection class| IP30 (acc. to EN 60529)
Protection class| I
Acoustic pressure| < 45 dB (A), in accordance with IEC 61010
Rated voltage| 230 V / 50 Hz, 110 V / 60 Hz
Power input| 50 W
Operating and storage temperature| 5 °C – 31 °C at up to 80 % rel. humidity 32
°C – 40 °C at up to 50 % rel. humidity (decreasing linearly)
Installation altitude| up to 2,000 m asl
Scope of delivery
Item | Quantity | Product no. |
---|---|---|
Hei-SHAKE Multi Reax | 1 | 545-10000-00 |
Power supply cord | 1 | country specific |
Attachment for 26 vessels up to ∅ of 16 mm | 1 | 549-016001-00 |
Attachment for twelve vessels with ∅ of 16 – 32 mm | 1 | 549-03000-00 |
Operating instructions | 1 | 01-005-004-28 |
Warranty registration | 1 | 01-006-002-78 |
Accessories
Detailed information on the available accessories for your device variant
can be found on our website at www.heidolph-instruments.com.
In case of need, contact an authorized dealer or our technical service, see
„Contact details“ on page 34.
Device service
DANGER
Switch the device’s main switch off and disconnect it from the power supply
before carrying out maintenance work, cleaning, or repairs. When cleaning,
avoid the penetration of liquids. Before replacing the fuses, switch off the
device and disconnect the power supply cord.
Always replace the two device fuses in pairs with original manufacturer fuses.
Further information on available accessories can be found on our website at
www.heidolph.com!
After fuse replacement, check the device for a safe condition according to IEC
61010-1.
General cleaning instructions
CAUTION
Clean the device’s surfaces with a soft, lint-free and only slightly moistened
cloth. Never use any aggressive or abrasive cleaning agents or aids.
Repairs
Repairs to the device may only be carried out by authorized experts!
Unauthorized repairs during the warranty period will result in the loss of the
warranty claim.
The owner is solely liable for damage caused by unauthorized repairs.
If repairs are required, contact an authorized dealer or our technical
service, see section „Contact details“ on page 34.
Include the completed certificate of decontamination with every device return,
see „Certificate of decontamination“ on page 35.
Maintenance
The device contains no user-serviceable components. If necessary (in the event
of abnormal operating behavior such as excessive noise or heat generation, for
example), contact our technical service, see section „Contact details“ on page
34.
Disposal
- When disposing of the device, observe the provisions of the WEEE Directive 2012/19/EU and its transposition into national law in the country of use.
- When disposing of portable batteries, observe the provisions of the European Battery Directive 2013/56/EU and their implementation in national law in the country of use.
- Check the device and all components for residues of substances that are hazardous to health, the environment and biohazardous before disposal.
- Properly remove and dispose of residues of substances that are hazardous to health, the environment and biohazardous!
Contact details
Heidolph Instruments North America
Phone: 1-866-650-9604
E-mail: service@heidolph.com
www.heidolphNA.com
Heidolph Instruments United Kingdom
Phone: 01799 – 5133-20
E-mail: service@radleys.co.uk
www.heidolph-instruments.co.uk
Local distributors
To find your local distributor please visit
www.heidolph.com
Warranty statement
Heidolph Instruments provides a three-year warranty against material and
manufacturing defects.
Glass and wear parts, transportation damage, and damage resulting from
improper handling or non-intended use of the product are excluded from the
warranty.
The warranty period for registered products begins on the date of purchase.
Register the product with the enclosed warranty card or on our homepage
www.heidolph.com.
For non-registered products, the warranty period begins with the date of the
serial production (to be determined by the serial number). In the event of
material or manufacturing defects, the product will either be repaired or
replaced free of charge within the warranty period.
Certificate of decontamination
Enclose the certificate of decontamination, duly completed, with your device
return. Submissions without a certificate of decontamination cannot be
processed!
CERTIFICATE OF DECONTAMINATION IN CASE OF RETURNS
Please fill in the required fields.
Note: The sender must package the goods properly and appropriately for
transport.
Heidolph Instruments GmbH & Co. KG
Walpersdorfer Straße 12 91126 Schwa Bach, Germany
Phone: +49 (0) 9122 9920-380
Fax: +49 (0) 9122 9920-19
Email: service@heidolph.de
SENDER
Name………….
Company/institution…………………….
Address………………………………
ZC/City………………………..
Country……………………
Email…………….
First name………..
Department…………
Workgroup…………….
Phone……………….
DEVICE DETAILS
Article number……….. Serial no………………..
Ticket number………..
Reason for sending in……….
Has the device been cleaned, decontaminated/disinfected?
If yes, which measures were carried out?
Yes No (Please mark as applicable)
Does this device pose a risk to people and/or the environment due to the
processing of substances that are hazardous to health, the environment and/or
are biohazardous?
If yes, with which substances did the device come into contact?
Yes No (Please mark as applicable)
LEGALLY BINDING DECLARATION
The principal/consignor is aware that they are liable to the agent/consignee
for losses or damage incurred due to incomplete and incorrect information.
Date……….
Signature……..
Company stamp……..
EU Declaration of Conformity
EU-Konformitaetserklaerung EU Declaration of conformity Schüttelund
Mischgeräte Shakers & Mixers
We, Heidolph Instruments GmbH & Co. KG,
Heidolph Instruments GmbH & Co. KG Walpersdorfer Straße 12 **91126 Schwa
Bach / Deutschland
** hereby declare, that the product (from serial number on 200066855)
designated below is in compliance with the basic requirements of all
applicable EU-directives stated below with regard to design, type of model
sold and manufactured by us. This certificate will be invalid if the product
is modified without the prior written consent and agreement of the
manufacturer.
Reax 2 | 541-21001-XX |
---|---|
Reax top | 541-10000-XX |
Reax control | 541-11000-XX |
Reax 20/4 | 541-20004-XX |
Reax 20/8 | 541-20008-XX |
Reax 20/12 | 541-20012-XX |
Tetramer 100 | 544-11100-XX, |
Duomo 1030 | 543-32105-XX, |
Polymer 1040 | 543-42105-XX, |
Polymer 2040 | 542-40005-XX, |
Multi Reax | 545-10000-XX |
Vibram ax 100 | 544-21200-XX |
Rotana 120 | 544-41200-XX |
Miramax 1000 | 544-12200-XX |
Anima 1010 | 543-12310-XX |
Anima 2010 | 542-10020-XX |
544-11200-XX | |
--- | --- |
543-32205-XX, | 543-32210-XX |
543-42110-XX, | 543-42205-XX, |
542-40010-XX | 543-42210-XX |
Tetramer 101 | 544-11300-XX |
--- | --- |
Vibramax 110 | 544-31200-XX |
Incubator 1000 | 549-90010-XX |
Promax 1020 | 543-22332-XX |
Promax 2020 | 542-20020-XX |
Maschinenrichtlinie / Machinery Directive 2006/42/EG EMV-Richtlijnen /
Electromagnetic Compatibility Directive 2014/30/EU Delegate (RoHS-)
Richtlijnen / Delegated (RoHS) Directive 2015/863/EU Angewandte
(harmonisierte) Normed / (Harmonized) Standards applied: EN ISO 12100:2010, EN
61326-1:2013, EN 61010-1:2010, EN IEC 63000:2018
Person Authorized to compile the technical file: Jorge Zyl – Heidolph
Instruments GmbH & Co. KG, Walpersdorfer Straße 12, 91126 Schwa Bach / Germany
Schwa Bach, 13.08.2020
Wolfgang Jaenicke Geschäftsführer
Managing Director Jörg Ziel
Qualitätsmanager
Quality Manager
Declaration of Conformity
In accordance with UK Government guidance
This declaration is issued under the sole responsibility of the manufacturer,
Heidolph Instruments GmbH & Co. KG
Walpersdorfer Straße 12 91126 Schwa Bach / Germany
Product: Shaker & Mixer
Model:
Reax top /control | 541-1XXXX-XX |
---|---|
Reax 20/4…12 | 541-200XX-XX |
Reax 2 | 541-21001-XX |
Duomo 1030 | 543-32XXX-XX |
Unimog X010 | 54X-1XXX0-XX |
Promax X020 | 54X-2XXXX-XX |
Polymer X040 | 54X-4XXXX-XX |
Tetramer 1XXX | 544-1XXXX-XX |
Vibram ax 1XX | 544-X1200-XX |
Rotana 120 | 544-41200-XX |
Multi Reax | 545-10000-XX |
Incubator 1000 | 549-90010-XX |
Description:
Shaker & Mixer with different working direction The object of the declaration
described above is in conformity with the relevant UK Statutory Instruments
(and their amendments):
2008 No. 1597 | The Supply of Machinery (Safety) Regulations |
---|---|
2008 2016 No. 1091 | The Electromagnetic Compatibility Regulations 2016 |
2012 No. 3032 | The Restriction of the Use of Certain Hazardous Substances in |
Electrical and Electronic Equipment Regulations 2012
and complies with the following technical standards :
EN ISO 12100:2010, EN 61326-1:2013, EN 61010-1:2010, EN IEC 63000:2018 UK
Authorised Representative (for authorities only) ProductIP ( UK ) Ltd. 8.
Northumberland Av.
London WC2N 5BY
Signed for and on behalf of Heidolph Instruments GmbH & Co. KG Walpersdorfer
Straße 12, 91126 Schwa Bach / Germany
Schwa Bach, 01.09.2021
Wolfgang Jaenicke Managing Director Jörg Ziel Quality Manager
Declaration of Conformity In accordance with UK Government guidance
This declaration is issued under the sole responsibility of the
manufacturer, Heidolph Instruments GmbH & Co. KG Walpersdorfer Straße 12 91126
Schwa Bach / Germany
Product: Shaker & Mixer
Model:
Reax top /control | 541-1XXXX-XX |
---|---|
Reax 20/4…12 | 541-200XX-XX |
Reax 2 | 541-21001-XX |
Duomo 1030 | 543-32XXX-XX |
Anima X010 | 54X-1XXX0-XX |
Promax X020 | 54X-2XXXX-XX |
Polymer X040 | 54X-4XXXX-XX |
Tetramer 1XXX | 544-1XXXX-XX |
Vibram ax 1XX | 544-X1200-XX |
Rotana 120 | 544-41200-XX |
Multi Reax | 545-10000-XX |
Incubator 1000 | 549-90010-XX |
Description:
Shaker & Mixer with different working direction
The object of the declaration described above is in conformity with the
relevant UK Statutory Instruments (and their amendments):
2008 No. 1597 | The Supply of Machinery (Safety) Regulations |
---|---|
2008 2016 No. 1091 | The Electromagnetic Compatibility Regulations 2016 |
2012 No. 3032 | The Restriction of the Use of Certain Hazardous Substances in |
Electrical and Electronic Equipment Regulations 2012
and complies with the following technical standards : EN ISO 12100:2010, EN
61326-1:2013, EN 61010-1:2010, EN IEC 63000:2018
UK Authorised Representative (for authorities only) ProductIP ( UK ) Ltd.
8. Northumberland Av. London WC2N 5BY
Signed for and on behalf of Heidolph Instruments GmbH & Co. KG Walpersdorfer
Straße 12, 91126 Schwabach / Germany
Schwabach, 01.09.2021
Wolfgang Jaenicke
Managing Director Jörg Ziel
Quality Manager
China RoHS DECLARATION OF CONFORMITY
Heidolph Instruments GmbH & Co.KG has made reasonable efforts to ensure
that hazardous materials and substances may not be used in its products.
In order to determine the concentration of hazardous substances in all
homogeneous materials of the subassemblies, a “Product Conformity Assessment”
(PCA) procedure was performed. As defined in GB/T 26572 the “Maximum
Concentration Value” limits (MCV) apply to these restricted substances:
• Lead (Pb): | 0.10% |
---|---|
• Mercury (Hg): | 0.10% |
• Cadmium (Cd): | 0.01% |
• Hexavalent chromium (Cr(VI)): | 0.10% |
• Polybrominated biphenlys (PBB): | 0.10% |
• Polybrominated diphenyl ether (PBDE): | 0.10% |
Environmental Friendly Use Period (EFUP)
EFUP defines the period in years during which the hazardous substances
contained in electrical and electronic products will not leak or mutate under
normal operating conditions.
During normal use by the user such electrical and electronic products will not
result in serious environmental pollution, cause serious bodily injury or
damage to the user’s assets.
The Environmental Friendly Use Period for Heidolph Instruments GmbH & Co.KG
products is 25 years.
This table is created according to SJ/T 11364-2014.
MATERIAL CONTENT DECLARATION FOR Heidolph Instruments GmbH & Co. KG
PRODUCTS
Hazardous substances
Part name| Pb| Hg| Cd| Cr(VI)| PBB| PBDE| EFUP
Packaging| 0| 0| 0| 0| 0| 0|
Plastic housing / parts| 0| 0| 0| 0| 0| 0
Battery| 0| 0| 0| 0| 0| 0
Glass| 0| 0| 0| 0| 0| 0
Electrical and electronic parts| X| X| X| 0| 0| 0|
Controller / measuring device| X| 0| X| 0| 0| 0
Metal housing /parts| X| 0| 0| 0| 0| 0
Motor| X| 0| 0| 0| 0| 0
Accessories| X| 0| 0| 0| 0| 0
China RoHS Certification
Note: Table applies to all products. Some of the components or parts
listed above may not be part of the enclosed product.
Indicates that the above mentioned hazardous substance contained in all
homogeneous materials of the part is below the required limit as defined in
GB/T 26572.
Apart from the disclosures in the above table, the subassemblies are not
intentionally manufactured or formulated with lead (Pb), mercury (Hg), cadmium
(Cd), hexavalent chromium (CrVI), polybrominated biphenyls (PBB), and
polybrominated diphenyl ethers (PBDE).
Products manufactured by Heidolph Instruments GmbH & Co.KG may enter into
further devices or can be used together with other appliances .
With these products and appliances in particular, Heidolph Instruments GmbH &
Co.KG will not take responsibility for the EFUP of those products and
appliances.
Schwabach, 06.08.2021
Wolfgang Jaenicke
Chief Executive Officer CEO Jörg Ziel Quality
Manager
© Heidolph Instruments GmbH & Co. KG
Doc-ID: 01-005-004-28-4 – Ed.: 2023-07-11
Subject to change without notice. The printed version of this document is not
regularly updated.
The latest issue of this document can be found by visiting our homepage.
References
- Heidolph Instruments : Start
- ments.com
- P65Warnings.ca.gov
- Heidolph EMEA - research made easy
- Heidolph Instruments : Start
- Heidolph NA - research made easy
- Heidolph NA - research made easy