GREISINGER GDH 200- 07 Digital Fine Manometer Instruction Manual

June 13, 2024
GREISINGER

GREISINGER-logo

GREISINGER GDH 200- 07 Digital Fine Manometer

GREISINGER-GDH-200-07-Digital-Fine-Manometer-product

Product Information

  • Product Name: Manometer (Over- /Under- And Difference Pressure)
  • Model Number: H59.0.01.6C-03
  • Operating Manual: GDH 200-07
  • WEEE-Reg.-No.: DE93889386

Contact Information:

Product Usage Instructions

General Note
Read the operating manual carefully and familiarize yourself with the device’s operation before using it. Keep the operating manual within easy reach near the device for reference in case of any doubts. All mounting, start-up, operating, maintenance, and removal procedures should be performed by qualified and specially trained personnel who have thoroughly read and understood the manual.

Safety

Intended Use
The device must be used only for its intended purpose and under unsuitable conditions. Adhere to the safety requirements specified below. Handle the device carefully and by its technical specifications. Avoid throwing or striking the device and protect it from dirt.

Safety Signs and Symbols
The following signs are used in this manual to indicate warnings:

  • DANGER: This symbol warns of imminent danger, including the risk of death, serious injuries, and significant property damage if not observed.
  • Caution: This symbol warns of possible dangers or dangerous situations that may cause damage to the device or the environment if not observed.
  • Note: This symbol indicates processes that may indirectly affect the operation or provoke unforeseen reactions if not observed.

Safety Guidelines
This device has been designed and tested in accordance with safety regulations for electronic devices. To ensure trouble-free operation and reliability, it is essential to adhere to the standard safety measures and special safety advice provided in the manual.

  1. Trouble-free operation and reliability can only be guaranteed if the device is used within the specified climatic conditions stated in the specifications. When transporting the device from a cold to a warm environment, allow it to adjust to the ambient temperature to prevent condensation-related failures before attempting a new start-up.
  2. Do not use this device in potentially explosive areas, as it may increase the risk of deflagration, explosion, or fire due to sparking.
  3. If there is any risk involved in operating the device, immediately switch it off and mark it accordingly to prevent accidental re-starting. Operator safety may be at risk if there is visible damage to the device if it is not functioning as specified, or if it has been stored under unsuitable conditions for an extended period.

Safety

Intended Use
The safety requirements (see below) have to be observed. The device must be used only according to its intended purpose and under suitable conditions. Use the device carefully and according to its technical data (do not throw it, strike it, …) Protect the device from dirt.

Safety signs and symbols
Warnings are labeled in this document with the following signs:

  • Caution! This symbol warns of imminent danger, death, serious injuries, and significant damage to property at non-observance.
  • Attention! This symbol warns of possible dangers or dangerous situations that can provoke damage to the device or environment at non-observance.
  • Note! This symbol points out processes that can indirectly influence operation or provoke unforeseen reactions at non-observance.

Safety guidelines
This device has been designed and tested by the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advice given in this manual are adhered to when using the device.

  1. 1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under “Specification”. If the device is transported from a cold to a warm environment, condensation may cause a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
  2. DANGER : This device must not be used in potentially explosive areas! The usage of this device in potentially explosive areas increases the danger of deflagration, explosion, or fire due to sparking. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid restarting.
  3. DANGER : Operator safety may be at risk if:
    • there is visible damage to the device
    • the device is not working as specified
    • the device has been stored under unsuitable conditions for a longer time. In case of doubt, please return the device to the manufacturer for repair or maintenance.
  4. Do not use these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage.

Product Specification

Scope of supply
The scope of supply includes:

  • Device, incl. 9V battery block
  • Operation manual

Operation and maintenance advice

Battery operation:
If the symbol “BAT“ is displayed on the left side of the display, the battery is weak, measuring can be continued for a certain time. If “bAt“ is displayed in the main display the battery is used up and needs to be replaced. Measuring is no longer possible. The battery has to be taken out when storing the device above 50 °C. We recommend taking out the battery if the device is not used for a longer period.

Configuration of the Device
To configure the instrument proceed like follows:

  1. Switch off the instrument.

  2. Press the ‘Mode‘ key while switching on the instrument, and keep the ‘Mode‘ key pressed until ‘P.oF‘ appears (after about 3 seconds).

  3. Auto Power Off Time „P.oF“ The auto power off time is entered in minutes. If no key is pressed during measuring, the instrument switches itself off automatically after the entered period.

  4. Press the up‘ or ‘down‘ key, and the currently selected power-off time will be displayed (off, 1..120min)

  5. Enter the desired time by pressing the ‘up‘ or ‘down‘ key. (factory setting: 20)
    Possible input:
    off 1…120: The auto power-off function is deactivated (permanent operation) auto power-off time in minutes.

  6. Confirm the value by pressing a left key, ‘Uni‘ appears in the display Display Unit „Uni“

  7. Press the up‘ or ‘down‘ key, and the currently selected unit will be displayed: mbar = hPascal (‘hPA‘), mmHg (‘nHG‘), PSI (‘PSI‘) or Pascal (‘PA‘)

  8. Enter the desired unit by pressing the ‘up‘ or ‘down‘ key. (factory setting: hPa)

  9. Confirm the value by pressing the left key, ‘dyn‘ appears in the display
    Measuring static/fast varying pressures “dyn“ (= dynamic)

  10. Press the‘ up‘ or ‘down‘ key, and the currently selected measuring mode will be displayed (see Chapter 5: “The Pressure Measuring”)

    • dyn = off standard measuring of static pressures, optimum battery life
    • dyn = on measuring of dynamic pressures
  11. Enter the desired mode by pressing the ‘up‘ or ‘down‘ key. (factory setting: off)

  12.  Confirm with the left key: values will be stored, and the instrument will restart (segment test). End of configuration.

Please note: If during the configuration no key is pressed within 60 seconds, the configuration will be aborted. Eventually made changes won’t be stored!

The Pressure Measuring
The device measures the difference pressure between the two pressure ports: (“+“ = higher pressure, “-“ = lower pressure) An automatic zero point adjustment is performed when switching on the device! If a pressure is connected when switching on, the high accuracy cannot be reached! As long as no pressure is connected additionally the zero-point is automatically stabilized. If a fast varying pressure should be measured (chimneys, etc.) the “dyn“ measuring should be activated (see below). This function is filtering the variations. Attention: “dyn“- measuring increases the power consumption more than 10 times.

  • Over pressure measuring: Connect to the “+“-port.
  • Under pressure measuring: Connect to “the -“ port, under pressure ,will be displayed without a negative sign.

Notes to Special functions

  • watch MIN value (Lo): press key ‘Mode‘ shortly once display changes between ‘Lo‘ and MIN value
  • watch MAX value (Hi): press key ‘Mode‘ shortly once again display changes between ‘Hi‘ and MAX value
  • restore current value: press key ‘Mode‘ shortly once again current value is displayed
  • clear MIN-/MAX- value: press key ‘Mode‘ for 2 seconds MIN and MAX values are cleared. The display shows shortly ‘CLr‘.

Zero-Function
Using the zero-function, the display will be set to zero. This is mainly intended to correct position dependency and drift of the sensor. Disconnect both pressure ports before zeroing

  • Press the “zero“-key for 2 seconds -> the display will be set to 0,
  • Press “zero“ for 5 seconds -> Zero function will be reset.
  • Relative measuring can also be performed by this function (= “tare“).

Scale Adjustment
The scale adjustment is intended to be used to compensate errors of the internal pressure sensor. The display value is given by the following formula: Display = measured value * ( 1 + scale adjustment/100)

To adjust a measuring scale proceed like follows:

  1. Switch off the device.
  2. Press the ‘Zero‘ key while switching on the instrument, and keep the ‘Zero‘ key pressed until ‘SCL‘ appears (after about 3 seconds).
  3. Press the‘ up‘ or ‘down‘ key, and the currently selected scale adjustment appears. (factory setting: off=0%)
  4. Choose the desired value by pressing ‘up‘ or ‘down‘ key. (max. input range: ±5.00%, off = 0%). The input is displayed in %.
    • Example : scale adjustment is 1.00  scale is increased by 1.00%  Scale = 101% At a measured value of 100 (without offset correction) the instrument would show 101
  5. Store the value by pressing the left key. The instrument will restart (segment test).

Please note : If during the changing of the scale adjustment, no key is pressed within 60 seconds, the input will be aborted. Eventually made changes won’t be stored!!

System messages

  • Er. 1 = measuring range has been exceeded
  • Er. 2 = meas. value has fallen below perm. range
  • Er. 3 = display range has been exceeded (>1999)
  • Er. 4 = meas. value has fallen below displayable range (<-1999)
  • Er. 7 = System fault – the device has detected a system fault (defective or far outside the allowable ambient temperature range)
  • = Sensor error or value could not be calculated
  • „bAt “ = battery is used up => replace battery

Specification

  • Measuring Range:
    • 0.00 …19.99 mbar, resolution 0.01 mbar
    • 20.0…199.9 mbar, resolution 0.1 mbar;
    • 0.00 …19.99 mmHg, resolution 0.01 mmHg
    • 20.0…150.0 mmHg, resolution 0.1 mmHg;
    • 0.000…1.999 PSI, resolution 0.001 PSI; 0 …1999
    • Pa, resolution 1 Pa;
    • Measuring range for underpressure: same as above; just switch connections -> display without neg. sign.
    • Because automatic offset adjustment is not suitable for pressures smaller than ±0.04mbar (4Pa). A suitable instrument for such measurings: GMH3110 + GMSD 2,5 (0,001mbar or 0,1Pa resolution!)
  • Max. Overpressure : 500 mbar
  • Measuring medium : noncorrosive gases
  • Accuracy : (at nominal temperature = 25°C, ±1 digit, offset adjustment)
    • 20.0 …199.9mbar:
    • ±0.2%FS hysteresis and linearity
    • ±0.4%FS temperature dependency 0 to 50°C
    • 0.00 …19.99mbar:
    • ±1%FS hysteresis and linearity
    • ±2%FS temperature dependency 0 to 50°C
  • Pressure Connections: 2 metal pressure ports for connection to 6 x 1 mm tubes at the top of the device (4mm inner tube Ø) standing out approx. 11mm „+“ = higher pressure, „-“ = lower pressure
  • Measuring Frequency: 1 measuring per second (at ‘dyn‘-measuring integrating, for quickly changing pressures)
  • Display : approx. 13 mm high, 3½-digit LCD
  • Operation Elements: 3 keys for ON/OFF, min-/max-value display, zero setting
  • Min-/Max-Value Memory: Min and max measured values are stored
  • Zero Function: The display value is set to zero per keypress, identical to an autom. zero adjustment
  • Scale: digital scale correction for pressure measuring
  • Ambient Conditions: -25 to 50°C; 0 to 80% RH. (not condensing)
  • Storage Temperature: – 25 to 70°C
  • Power Supply : 9V-battery (in the scope of supply)
  • Power Consumption: approx. 250μA at standard measuring of static pressures, 3mA at ‘dyn‘-measuring
  • Battery Life: Standard zinc carbon battery more than 1200 hours at standard measuring!
  • Battery Change Indicator: automatically if the battery is used up: “BAT”
  • Auto-Off-Function: when the Auto-Off Function is activated, the device switches automatically off, if the keypad is not attended for a longer time (selectable 1…120min).
  • Housing : impact-resistant ABS, transparent panel, front side IP65
  • Dimensions: approx. 106 x 67 x 30 mm (L x W x D) without pressure ports
  • Weight: approx. 135g, incl. battery
  • EMC : The device corresponds to the essential protection ratings established in the Regulations of the Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EC). Additional fault: <1%

Reshipment and Disposal

Reshipment

DANGER : All devices returned to the manufacturer have to be free of any residual measuring media and other hazardous substances. Measuring residuals at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment, especially for fully functional devices. Please make sure that the device is protected in the package by enough packing materials.

Disposal instructions
Batteries must not be disposed of in the regular domestic waste but at the designated collecting points. The device must not be disposed of in the unsorted municipal waste! Send the device directly to us (sufficiently stamped), if it should be disposed of. We will dispose of the device appropriate and environmentally sound.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

GREISINGER User Manuals

Related Manuals