GREISINGER G1910-02 Compact Co2 Monitor With Alarm User Manual

June 13, 2024
GREISINGER

G1910-02 Compact Co2 Monitor With Alarm

Product Information

The G 1910-02 is a compact CO2 monitor with an alarm. It is
designed to measure and monitor the levels of carbon dioxide (CO2)
in the surrounding environment. The monitor features an NDIR CO2
sensor that ensures accurate and reliable measurements.

Scope of Delivery

  • Compact CO2 monitor
  • Alarm unit
  • Battery pack
  • User manual

Functional Description

The CO2 monitor is equipped with an NDIR (Non-Dispersive
Infrared) CO2 sensor that measures the concentration of CO2 in
parts per million (ppm). The monitor has an alarm unit that alerts
the user when the CO2 levels exceed a predefined threshold.

Product Usage Instructions

Commissioning

  1. Open the package and remove all the components.

  2. Insert the battery pack into the designated slot on the
    monitor.

  3. Ensure that the battery pack is fully charged before
    proceeding.

  4. Turn on the CO2 monitor by pressing the power button.

Configuration

  1. Once the monitor is turned on, you can configure the settings
    by accessing the menu using the navigation buttons.

  2. Use the navigation buttons to navigate through the menu
    options.

  3. Select the desired option by pressing the enter button.

  4. Adjust the settings according to your preferences or specific
    requirements.

  5. Save the configuration changes and exit the menu.

Measurement Basics

The CO2 monitor measures the concentration of CO2 in the
surrounding environment. The measurement is displayed in parts per
million (ppm) on the monitor’s screen. The alarm unit will be
triggered if the CO2 levels exceed the predefined threshold.

Operation and Maintenance

It is important to follow these operating and maintenance
notices:

  • Keep the CO2 monitor clean and free from dust or debris.

  • Do not expose the monitor to extreme temperatures or
    humidity.

  • Regularly check the battery level and replace it when
    necessary.

Error and System Messages

If any error or system messages appear on the monitor’s screen,
refer to the user manual for troubleshooting steps or contact
customer support for assistance.

Disposal

When disposing of the CO2 monitor, follow local regulations and
guidelines for electronic waste disposal. Do not dispose of the
monitor in regular household waste.

Technical Data

For detailed technical specifications, refer to the user
manual’s technical data section.

Spare Parts and Accessories

If you require any spare parts or additional accessories for the
CO2 monitor, refer to the spare parts and accessories section in
the user manual or contact customer support for assistance.

Operating manual

EN

G 1910-02

Compact CO2 monitor with alarm

Table of contents

1 1.1 1.2 1.3 1.4 1.5 1.6
2 2.1 2.2 2.3 2.4 2.5
3 3.1 3.2
4 4.1 4.2 4.3 4.4
5 5.1 5.1.1 5.2 5.2.1 5.2.2 5.2.3 5.2.4 5.2.5
6 6.1 6.1.1 6.1.2
7 7.1 7.2 7.2.1 7.2.2 7.2.3 7.3
8
9
10
11

Table of contents
About this documentation ………………………………………………………………………………………………………….. 4 Foreword ………………………………………………………………………………………………………………………………………. 4 Purpose of the document………………………………………………………………………………………………………………. 4 Legal notices …………………………………………………………………………………………………………………………………. 4 Correctness of content ………………………………………………………………………………………………………………….. 4 Layout of this document ………………………………………………………………………………………………………………… 4 Further information………………………………………………………………………………………………………………………… 5
Safety …………………………………………………………………………………………………………………………………………… 6 Explanation of safety symbols……………………………………………………………………………………………………….. 6 Foreseeable misuse ……………………………………………………………………………………………………………………… 6 Safety instructions…………………………………………………………………………………………………………………………. 7 Intended use …………………………………………………………………………………………………………………………………. 8 Qualified personnel ……………………………………………………………………………………………………………………….. 8
Description ………………………………………………………………………………………………………………………………….. 9 Scope of delivery…………………………………………………………………………………………………………………………… 9 Functional description……………………………………………………………………………………………………………………. 9
The product at a glance…………………………………………………………………………………………………………….. 10 The G 1910-02 ……………………………………………………………………………………………………………………………. 10 Display elements …………………………………………………………………………………………………………………………. 10 Operating elements……………………………………………………………………………………………………………………… 10 Connections ………………………………………………………………………………………………………………………………… 11
Operation …………………………………………………………………………………………………………………………………… 12 Commissioning ……………………………………………………………………………………………………………………………. 12
Explanation……………………………………………………………………………………………………………………….. 12 Configuration……………………………………………………………………………………………………………………………….. 12
Explanation……………………………………………………………………………………………………………………….. 12 Opening the configuration menu ………………………………………………………………………………………. 12 Configuring parameters of the configuration menu…………………………………………………………… 13 Call-up of the expanded settings menu……………………………………………………………………………… 15 Configure parameters of the expanded settings menu ………………………………………………………. 16
Measurement Basics…………………………………………………………………………………………………………………. 18 NDIR CO2 sensor ………………………………………………………………………………………………………………………… 18
Explanation……………………………………………………………………………………………………………………….. 18 Design……………………………………………………………………………………………………………………………….. 18
Operation and maintenance ……………………………………………………………………………………………………… 19 Operating and maintenance notices ……………………………………………………………………………………………. 19 Battery…………………………………………………………………………………………………………………………………………. 19
Charge status display………………………………………………………………………………………………………… 19 Charging the batteries ……………………………………………………………………………………………………….. 19 Rechargeable battery replacement …………………………………………………………………………………….. 20 CO2 adjustment …………………………………………………………………………………………………………………………… 21
Error and system messages …………………………………………………………………………………………………….. 24
Disposal……………………………………………………………………………………………………………………………………… 26
Technical data……………………………………………………………………………………………………………………………. 27
Spare parts and accessories ……………………………………………………………………………………………………. 28

2 / 30

B-H90.0.01.DB2-2.1

Table of contents

12

Service ………………………………………………………………………………………………………………………………………..29

12.1

Manufacturer ………………………………………………………………………………………………………………………………..29

12.2

Repairs processing……………………………………………………………………………………………………………………….29

12.3

Sales offices …………………………………………………………………………………………………………………………………29

B-H90.0.01.DB2-2.1

3 / 30

1 | About this documentation
1

G 1910-02
About this documentation

1.1 Foreword
Read this document carefully and familiarise yourself with the operation of the product before you use it. Keep this document ready to hand and in the immediate vicinity of the product so that it is available to the personnel/user for reference at all times in case of doubt. The product was developed according to the state of the art and fulfils the requirements of the applicable European and national Directives. All corresponding documents are available from the manufacturer. Only technically qualified persons are permitted to carry out commissioning, operation, maintenance and decommissioning. The qualified personnel must have carefully read and understood the operating manual before beginning any work.
1.2 Purpose of the document
­ It provides important information for operating safely and efficiently with the product.
­ In addition to the quick reference guide with all relevant legal and safety content in hard copy, this document is a detailed reference option for the product.
1.3 Legal notices
The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse, disregarding this document, disregarding safety notices, assignment of inadequately qualified technical personnel and arbitrary modifications of the product. Only carry out the maintenance and service tasks on this product that are described in this documentation. In the process, adhere to the specified steps. For your own safety, only use original spare parts and accessories of the manufacturer. We assume no liability for the use of other products and resulting damage. This document is entrusted to the recipient for personal use only. Any impermissible transfer, duplication, translation into other languages or excerpts from this operating manual are prohibited. The manufacturer assumes no liability for print errors.
1.4 Correctness of content
The contents of this document were checked for corrected and are subject to a continuous correction and updating process. This does not rule out potential errors. In the event that errors are discovered or in case of suggestions for improvement, please inform us immediately via the indicated contact information in order to help us make this document even more user-friendly.
1.5 Layout of this document
Description
Each chapter is explained at the beginning in the description.

4 / 30

B-H90.0.01.DB2-2.1

G 1910-02

About this documentation | 1

Prerequisite
All mandatory prerequisites are then listed for each step.

Instruction
Tasks to be carried out by the personnel / user are represented as numbered instructions. Adhere to the sequence of the specified instructions.

Representation
Shows an illustrative instruction or a configuration of the product.

Formula
Some instructions include a formula for a general understanding of a configuration, programming or a setting of the product.

Outcome of an action
Result, consequence or effect of an instruction.

Emphases
In order to simplify legibility and provide a clearer overview, various sections / information are emphasised.
­ 1234 Display elements ­ Mechanical controls ­ Product functions ­ Product labels ­ Cross-reference [ 4] ­ Foot notes

1.6 Further information
Software version of the product: ­ V1.6 or later

B-H90.0.01.DB2-2.1

5 / 30

2 | Safety

G 1910-02
2 Safety
2.1 Explanation of safety symbols
DANGER
This symbol warns of imminent danger, which can result in death, severe bodily injury, or severe property damage in case of non-observance.
CAUTION
This symbol warns of potential dangers or harmful situations, which can cause damage to the device or to the environment in case of non-observance.
NOTE
This symbol indicates processes, which can have a direct influence on operation or can trigger an unforeseen reaction in case of non-observance.
2.2 Foreseeable misuse
The fault-free function and operational safety of the product can only be guaranteed if applicable safety precautions and the device-specific safety instructions for this document are observed. If these notices are disregarded, personal injury or death, as well as property damage can occur.
DANGER
Incorrect area of application!
In order to prevent erratic behaviour of the product, personal injury or property damage, the product must be used exclusively as described in the chapter Description [ 9] in the operating manual.
­ Do not use in safety / Emergency Stop devices! ­ The product is not suitable for use in explosion-prone areas! ­ The product must not be used for diagnostic or other medical purposes on pa-
tients! ­ The product is not intended to come into direct contact with food! ­ Not suitable for use with requirements on functional safety, e.g. SIL!

6 / 30

B-H90.0.01.DB2-2.1

G 1910-02

Safety | 2

DANGER

Danger due to elevated CO2 concentration
The product is not suitable for use as personal protective equipment with elevated CO2 levels. However, it can indicate an elevated CO2 value. The measured value appears in the display as a % or ppm value.

G 1910-02 G 1910-20 CO2 concentration

%

ppm

20

10

4.0

3.0

2,0

20,000

1.0

10,000

0.5

5000

0.1 .. 0.2 1000 .. 2000

0.04

400

Effect
Death within a few seconds Loss of consciousness, death, dizziness, vomiting, headaches, reduced blood flow to brain IDLH – immediate danger to life and health Normal exhalation concentration, elevated breathing and pulse rate
Possible shortness of breath TWA ­ Maximum for working conditions Recommended maximum value in public areas Fresh air

Product is not permitted for the area
Expanded measuring range. The product can be used conditionally
Area of application of the product with specified accuracy
The values are guideline values. Depending on the health condition and duration of exposure, problems can also occur below the indicated concentrations under certain circumstances.

2.3 Safety instructions
NOTE
This product does not belong in children’s hands!

B-H90.0.01.DB2-2.1

7 / 30

2 | Safety

G 1910-02
2.4 Intended use
The product is designed exclusively for measurements in ambient air and environments with slightly elevated CO2 concentrations in areas that are not harmful to the health. It is designed to be carried on the body for mobile use. The user can be warned optically and acoustically of elevated CO2 concentrations based on variable alarm limits. Example applications for this are:
­ Use as a monitor for recording of the mean value weighted over 8 hours (TWA) or 15 minutes (STEL).
­ Monitoring of air quality.
2.5 Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have adequate knowledge of the measuring process and the significance of the measurements. This document makes a valuable contribution to this. The instructions in this document must be understood, observed and followed. In order to avoid any risks arising from interpretation of the measurements in the concrete application, the user must have additional expertise. The user is solely liable for damages/danger resulting from misinterpretation due to inadequate expertise.

8 / 30

B-H90.0.01.DB2-2.1

G 1910-02
3 Description

Description | 3

3.1 Scope of delivery
Please check to ensure the completeness of the product after opening the package. You should find the following components:
­ Quick reference guide
­ Handheld measuring device, ready for operation, including rechargeable batteries
­ Micro USB to USB type A connection cable
3.2 Functional description
The product offers precision, speed and reliability in a compact, ergonomic housing. It is distinguished by an illuminated 3-line display. The product can be switched on, switched off and configured and the measurements and parameters can be adjusted and held with the operating elements. The product is equipped with an integrated optical carbon dioxide sensor. In addition to display of the currently measured CO2 value, the mean value weighted over 8 hours (TWA) or 15 minutes (STEL) can be output. An integrated two-stage alarm also issues an optical signal and an acoustic signal as a warning when the adjusted limits are exceeded.
­ A pre-alarm issues a text warning or the backlight flashes with a brief horn when the limit is exceeded depending on the setting.
­ A pre-alarm issues a text warning or the backlight flashes rapidly with a continuous horn when the limit is exceeded depending on the setting.

B-H90.0.01.DB2-2.1

9 / 30

4 | The product at a glance
4
4.1

G 1910-02
The product at a glance
The G 1910-02

10 / 30

LCD Display

Sensor openings

Micro USB socket

Front view

4.2 Display elements
Display

Charge status display

Evaluation of the charge status

Unit display

Display of units or type of mode, min/max/hold

Main display Auxiliary display Bar graph

Measurement of the current CO2 value Display of the mean value Visualisation of the CO2 value

4.3 Operating elements
On / Off button

Press briefly Long press

Switch on the product Activate / deactivate lighting Switch off the product Reject changes in a menu

Up / Down button

Press briefly Long press Both simultaneously

Display of the min/max value
Change value of the selected parameter
Reset the min/max value of the current measurement Rotate display, overhead display

B-H90.0.01.DB2-2.1

G 1910-02

The product at a glance | 4

Function key

Press briefly Long press, 2s

Freeze measurement (Hold) Return to measurement display Call up next parameter
Start menu configuration, (ONF appears in the display

4.4 Connections

Micro USB socket

Charging the batteries

B-H90.0.01.DB2-2.1

11 / 30

5 | Operation

G 1910-02
5 Operation

5.1 Commissioning

Description
Prerequisite Instruction Outcome of an action

5.1.1 Explanation
The product is switched on with the On/Off button. It may be necessary to configure the product after switching on. See Configuration [ 12].
A self-test is conducted after switching on. The following increasing CO2 values 1000, 2000, 4000 and 8000 are shown in the display. TEST is shown in the secondary display. If an alarm is active and is in one of these relevant areas, it is triggered.
If —- appears in the display after the self-test, the sensor is not ready for measurement. If batteries are drained and the product is not used for an extended time, it can take up to 30 seconds until the measurement starts. If no measurement is received by the sensor within 30 seconds, the product issues an alarm.
­ The rechargeable batteries have been charged via the micro USB socket.
­ Press the On/Off button.
Information about the configuration of the product appears in the display.

TEST

Self-test

An automatic self-test is carried out during system

start-up. The bar indicator for the threshold value is

displayed and the alarm is tested if it has been acti-

vated.

POFF

Automatic shut- Automatic shut-off activated. The product is

off

switched off if no buttons have been pressed after

the adjusted time

­ The product is now ready for measurement.

5.2 Configuration

Description
Prerequisite Instruction

5.2.1 Explanation
The following steps describe how to adapt the product for your purposes.
NOTE
There are various configuration parameters available depending on the product version and configuration. These can vary depending on the product version and configuration.
5.2.2 Opening the configuration menu
In order to configure the product, you must first open the Configuration menu. The menu is opened as shown in the illustration. 1. Press the Function key for 2 seconds to open the Configuration menu. 2. (ONF appears in the display. Release the function key.

12 / 30

B-H90.0.01.DB2-2.1

Representation

G 1910-02

Operation | 5

3. By briefly pressing the Function key, you can scroll through the parameters. Select the parameter you would like to configure.
4. When you have selected the desired parameter, change the parameter to the desired value with the Up button and the Down button.
5. The changes are saved after running through the entire Configuration menu. STOR appears in the display. The Configuration menu can be exited from any arbitrary parameter by pressing and holding the Function key for 2 seconds. The changes made up that point are saved.

Call up menu Next parameter Change value Save changes Discard changes

Outcome of an action

2s

Press: Single 2s

step

Hold: Fast change

The Configuration menu is closed after the last parameter.

2s
Product is switched off

NOTE

If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product.

Description Prerequisite Instruction
Outcome of an action
Representation

5.2.3 Configuring parameters of the configuration menu
The following representation shows the available parameters and various configuration options.
­ The Configuration menu is open. See Opening the configuration menu [ 12].
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the Up button and Down button.
3. The available configuration options are listed for each parameter in the following representation.
The changed value is saved and the Configuration menu is closed. STOR appears in the display. If necessary, the product is restarted automatically in order to adopt the changed values.

NOTE

The configuration is closed if no button is pressed for 2 minutes. Any changes made up to that point are not saved. C.END appears in the display.

Parameter Values

Meaning

B-H90.0.01.DB2-2.1

13 / 30

5 | Operation

G 1910-02

Input

INP
% PPm
Alarms

CO2 measurement in % CO2 measurement in ppm

AL. OFF ON
BEEP L,TE
AL.1 0.000 .. AL.2 0 .. AL2

AL.2 Mean value

AL.1 .. 1.000 AL1 .. 10000

No active alarm
Alarm alerting via text display, acoustic signal and flashing of the backlighting
Alarm alerting via text display and acoustic signal
Alarm alerting via text display and flashing of the backlighting
Depending on the setting of the parameter value INP
Min. alarm limit in % or ppm; a pre-alarm is triggered when the value is exceeded When the pre-alarm is triggered, it can be muted for 5 minutes. To do this, press any key. The display shows (LR AL.1.
Depending on the setting of the parameter value INP
Max. alarm limit in % or ppm; the main alarm is triggered when the value is exceeded

LCD.2 8 H STEL OFF
Shut-off time

Time weighted over 8 hours, mean value TWA Time weighted over 15 minutes, mean value STEL Mean value determination deactivated

POFF

OFF

No automatic shut-off

0:15 0:30 1:00 4:00 Automatic shut-off after a selected time in hours

12:00

and minutes, during which no buttons have been pressed

14 / 30

B-H90.0.01.DB2-2.1

Description Prerequisite Instruction
Representation

Backlight L,TE

G 1910-02

Operation | 5

OFF

Backlight deactivated

0:15 0:30 1:00 4:00 Automatic shut-off of the backlight after a selected time in minutes and seconds, during which no buttons have been pressed

ON

No automatic shut off of the backlight

Factory settings IN,T
NO YES

Use current configuration
Reset product to factory settings. IN,T DONE appears in the display

5.2.4 Call-up of the expanded settings menu
In order to configure the product, you must first open the Expanded settings menu. The menu is opened as shown in the illustration.
­ The product is switched off. 1. Press and hold the Down button. 2. Press the On/Off button to switch on the product. 3. Release the On/Off button after 1 second and then the Down button in order to call
up the Expanded settings menu. The display shows the first parameter. 4. By briefly pressing the Function key, you can scroll through the parameters. Select
the parameter you would like to configure. 5. When you have selected the desired parameter, change the parameter to the de-
sired value with the Up key and the Down key. 6. In order to save the new parameter value, press and hold the Function key for
longer than 2 seconds.
Call up menu

Outcome of an action

Hold

1s

Release

Release

The Expanded settings menu is closed after the last parameter.

B-H90.0.01.DB2-2.1

15 / 30

5 | Operation
Description Prerequisite Instruction Representation

G 1910-02

NOTE
If the Expanded settings menu is processed completely, the changes are saved automatically. STOR appears in the display. However, it can be exited at any time by pressing the Function key for 2 seconds. The changes made up that point are also saved. If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product.

5.2.5 Configure parameters of the expanded settings menu
The following representation shows the available parameters and various configuration options.
­ The Expanded settings menu is called up. See Call-up of the expanded settings menu [ 15].
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the Up key and Down key.
3. The available configuration options are listed for each parameter in the following representation.

Parameter Values

Meaning

Adjustment

[AL

OFF

No adjustment

2.PT

2-point adjustment

1.PT

1-point adjustment

KARD

Basic sensor adjustment
Only 0 ppm or 0.000 % can be selected as a setpoint value for the adjustment, such as, for example on nitrogen or 400 ppm or 0.040 % for adjustment on clean ambient air.

Gradient adjustment specification

[.SL

350 .. 10,000 0.035 .. 1.000

Setpoint value in ppm or in % for 1.PT or 2.PT adjustment

0 / 400 0.000 / 0.040

Setpoint value in % or in ppm for KARD adjustment

16 / 30

B-H90.0.01.DB2-2.1

G 1910-02

Operation | 5

Outcome of an action

NOTE
The setpoint value of 400 ppm or depending on selection 0.040 % can be used for a simple adjustment in clean ambient air. Otherwise, use the expected value of the test gas according to the analysis certificate or display value of the reference device.
The changed value is saved and the Expanded settings menu is closed. STOR appears in the display.
NOTE
If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product.

B-H90.0.01.DB2-2.1

17 / 30

6 | Measurement Basics
6

G 1910-02
Measurement Basics

6.1 NDIR CO2 sensor

IR radiation source

6.1.1 Explanation
The sensors are based on non-dispersive infrared sensor technology, NDIR. This is the most widely used sensor technology for CO2 measurement. The principle of the NDIR sensor is that an IR light source is focusses so that an optical beam is created, which passes through the existing gas, including CO2. After an optical band pass filter is passed, an IR sensor measures the level of IR light, wherein the CO2 values in the optical path are displayed.
6.1.2 Design
1. Gas inlet 2. Gas outlet 3. IR lamp 4. Optical filter 5. Thermopile detector
IR detectors
A sensor module with single-channel detectors is used for the product. A micro light bulb is used as an IR radiation source. It emits a broadband spectrum. The sensor is long-lasting and maintenance-free. In order to be able to maintain the specified accuracy for many years to come, regular adjustment must be carried out depending on the required accuracy.

18 / 30

B-H90.0.01.DB2-2.1

G 1910-02

Operation and maintenance | 7

7 Operation and maintenance

7.1 Operating and maintenance notices
CAUTION
Damage to the sensor
A sensitive optical sensor is installed in the product. The sensor parameters can change due to impact or falling. This can result in incorrect measurements.
­ Protect the product from impact and falling! ­ The measurements must be checked after the product falls or is jarred. If the val-
ues deviate, a basic sensor adjustment must be carried out!
NOTE
The product must be handled with care and used in accordance with the technical data. Do not throw or strike.
NOTE
If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed or recharged regularly. This prevents leaks from the rechargeable batteries and increases the life of the rechargeable batteries.
7.2 Battery

7.2.1 Charge status display
If the battery status display blinks, the batteries are depleted and must be recharged. However, the device will still operate for a certain length of time.
If the BAT display text appears in the main display, the rechargeable battery voltage is no longer adequate for operation of the product. Now the rechargeable battery is fully depleted.
7.2.2 Charging the batteries
­ Charge via Micro USB socket with the accompanying cable. ­ Must be operated on a USB port or USB mains adapter with an output voltage of
4.75 V .. 5.25 V, which can deliver a current of 500 mA. ­ The charging process is visualised in the charge status display. ­ If the charging process has concluded BAT FVLL is shown in the display. ­ Charging at room temperature in a range of 0 .. 40 °C is permitted. ­ The device can heat up during charging. Max. up to 50 °C. ­ Charging time approx. 8 hours. ­ The charging time can take longer in ambient temperatures above 30 °C. ­ The rechargeable battery temperature is monitored. At temperatures below 0 °C
and above 50 °C, the charging is interrupted.

B-H90.0.01.DB2-2.1

19 / 30

7 | Operation and maintenance

G 1910-02

­ In order to protect the batteries, the charging process is not started when the charge cable is plugged in, if fully charged batteries are detected based on the voltage monitor.

Description Prerequisites Instruction
Outcome of an action

7.2.3 Rechargeable battery replacement

DANGER
Danger of explosion!
Using damaged or unsuitable rechargeable batteries can generate heat, which can cause the rechargeable batteries to crack and possibly explode!
­ Only use high-quality and suitable NiMH rechargeable batteries!
CAUTION
Damage!
If the rechargeable batteries have different charge levels, leaks and thus damage to the product can occur.
­ Use new, high-quality rechargeable batteries! ­ Do not use different types of rechargeable batteries! ­ Remove depleted rechargeable batteries and dispose of them at a suitable collec-
tion point!
NOTE
Read the following handling instructions before replacing rechargeable batteries and follow them step by step. If disregarded, the product could be damaged or the protection from moisture could be diminished. Proceed as follows to replace the rechargeable batteries.
­ The product is switched off.

Fig. 1: Batteriewechsel The product is now ready for use again.

1. Unscrews the Phillips screws (A)and remove the cover.
2. Carefully replace the two Mignon AA batteries (B). Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
3. The O-ring (C) must be undamaged, clean and positioned at the intended depth. In order to facilitate assembly and avoid damage, a suitable grease can be applied.
4. Fit the cover on evenly. The O-ring must remain at the intended depth!
5. Tighten the Phillips screws (A).

20 / 30

B-H90.0.01.DB2-2.1

Description Prerequisite

G 1910-02

Operation and maintenance | 7

7.3 CO2 adjustment
In order to improve the accuracy, the carbon dioxide sensor can be adjusted. In order to conduct a CO2 adjustment, proceed as follows.
1-point adjustment
Used for optimisation of the accuracy at the adjustment point. The best possible accuracy can be achieved even with elevated CO2 concentrations
2-point adjustment
Used for optimisation of the accuracy for extreme requirements over a wide measuring range beginning from 0 ppm.
Basic sensor adjustment
Used to reset the sensor in case of deviations beyond the normally expected deviations. The values of all prior adjustments are reset and cannot be restored. A plausibility check of the carbon dioxide concentrations does not take place. Depending on the setting, adjustment to 0 ppm CO2 test gas or fresh outside air can take place. Ensure that the setting of the gradient compensation setpoint [.SL was entered correctly in the Expanded settings menu.
If reliable values can no longer be displayed, we recommend performing basic sensor adjustment of the zero point at 0 ppm CO2 with nitrogen and a possibly necessary additional gradient compensation via 1-point adjustment.
If incorrect values are displayed after the adjustment, e.g. ERR.2 at 0 ppm CO2, the adjsutment must be repeated.

NOTE

Adjustment can be carried out in clean ambient air or with test gases (optionally available gas extraction device recommended). 1-point adjustment at any arbitrary point above 350 ppm and 2-point adjustment at 0 ppm and an arbitrary point above 350 ppm are both possible.
For automatic adjustment, open the Adjustment menu.
­ The product is switched on.
­ Clean ambient air or test gas for gradient correction
­ Test gas 0 ppm CO2 for zero point adjustment for 2-point adjustment or basic sensor adjustment at 0 ppm
­ Gas extraction device, if applicable

B-H90.0.01.DB2-2.1

21 / 30

7 | Operation and maintenance Instruction

G 1910-02

22 / 30

1. If you would like to perform a adjustment with test gas, connect the extraction device to the product first.
2. Press the Function key for 4 seconds to start the adjustment. [ONF appears in the display first, then [AL.
3. 1-point, 2-point or basic sensor adjustment is started. This depends on what was adjusted in the Expanded settings menu.
1-point adjustment 1. [.SL appears in the display for 1-point adjustment. 2. The product determines a stable value first. If the measured value is outside of the
value range integrated in the product, the display flashes briefly and an acoustic signal is issued every 10 s. 3. If a stable correct value is reached, the display flashes briefly, an acoustic signal sounds and the bar indicator blinks. 4. You can change the value of the gradient compensation with the up key and down key. Otherwise, confirm the preadjusted value with the function key. The 1-point adjustment is finished.
2-point adjustment 1. [.OF appears in the display for 2-point adjustment. 2. Allow the test gas to flow at about 0.5 l/min at 0 ppm of CO2. The product deter-
mines a stable value first. If the measured value is outside of the value range integrated in the product, the display flashes briefly and an acoustic signal is issued every 10 s. 3. If a stable correct value is reached, the display flashes briefly, an acoustic signal sounds and the bar indicator blinks. [.SL appears in the display. 4. Remove the test gas and connect a second test gas to the extraction device. The product determines a stable value first. If the measured value is outside of the value range integrated in the product, the display flashes briefly and an acoustic signal is issued every 10 s.
B-H90.0.01.DB2-2.1

G 1910-02

Operation and maintenance | 7

5. If the display flashes briefly, an acoustic signal sounds and the bar display blinks, a stable correct value has been achieved.
6. You can change the value of the gradient compensation with the up key and down key. Otherwise, confirm the preadjusted value with the function key. Remove the extraction device. The 2-point adjustment is finished

Basic sensor adjustment
1. KARD appears in the display for basic sensor adjustment.
2. Depending on the specification [.SL, let the test gas with 0 ppm CO2 flow in at approx. 0.5 l/Min or move the product into clean ambient are with 400 ppm CO2. The product determines a stable value first. If the displayed measurement is outside of the measuring range of the sensor, the product cannot perform stability recognition. It must be ensured that the carbon dioxide concentration at the sensor is stable before the adjustment is started.
3. If a stable correct value is reached, the display flashes briefly, an acoustic signal sounds and the bar indicator blinks.
4. Remove the extraction device, if applicable.
NOTE
Current test gases normally have accuracies of ± 2 %. This tolerance must be taken into consideration with the measurement uncertainty. The specifications on the analysis certificate must always be observed.

Outcome of an action

NOTE
For information about the available adjustment settings, refer to Configuring parameters of the configuration menu [ 13].
After the adjustment is finished [AL DONE is displayed. Then, the current measurement is shown in the display again.
If the adjustment is not completed successfully an error message is displayed. [AL ERR. appears in the display. See Error and system messages [ 24]. Confirm the error message pressing the Function key. The product restarts. The values of the last correctly performed adjustment are restored. Configure parameters of the expanded settings menu [ 16]

B-H90.0.01.DB2-2.1

23 / 30

8 | Error and system messages

G 1910-02

8 Error and system messages

Display —-
—-

Meaning

Possible causes Remedy

No signal from the Sensor not ready Wait the start-up time of the

sensor

CO2 concentration sensor

Measurement far too high

Place the product in clean

outside of the meas- Defective sensor ad- outdoor air

uring range

justment

Perform sensor adjustment

Sensor defect

Send in for repair

The display value Sensor not ready could not be deter- Sensor defect mined

Wait the start-up time of the sensor
Send in for repair

No display, Rechargeable batunclear char- tery depleted acters or no System error response when but- Product is defective tons are pressed

Rechargeable battery depleted
Error in the product
Product is defective

Charge battery
Replace rechargeable battery
Send in for repair

BAT LO

Rechargeable battery depleted

Rechargeable battery discharged
Rechargeable battery defect

Charge battery
Replace rechargeable battery

(AL ERR.1

Zero point adjustment defective

Incorrectly meas- Expose sensor to a test gas ured CO2 concentra- with 0 ppm CO2 tion for adjustment

(AL ERR.2 (AL ERR.3 (AL ERR.5 ERR.1
ERR.2 ERR.7

Defective gradient compensation

Incorrectly meas- Expose sensor to a test gas ured CO2 concentra- with known CO2 concentration for adjustment tion
Incorrect CO2 con- Enter correct value centration

Time for stability recognition exceeded

Stability recognition Provide a consistent flow

lasts longer than 10 with a constant CO2 concen-

minutes

tration

Measuring range ex- Measurement too Stay within allowable meas-

ceeded

high

urement range

Defective sensor ad- Perform sensor adjustment

justment

Send in for repair

Sensor defect

Measuring range is Defective sensor ad- Perform sensor adjustment

undercut

justment

Send in for repair

Sensor defect

Sensor error

Defective sensor ad- Perform sensor adjustment

justment

Send in for repair

Sensor defect

24 / 30

B-H90.0.01.DB2-2.1

ERR.T SYS ERR STAB

G 1910-02

Error and system messages | 8

Temperature error
System error
No measurement change within 2 minutes

Permissible temper- Charging of battery only at 0

ature range during .. 40 °C

charging exceeded Bring product to room tem-

or undercut

perature and restart charg-

ing process

Error in the product Sensor defect

Switch product on/off
Replace rechargeable batteries
Send in for repair

Product in extremely Place the product in clean

constant environ- outdoor air

ment

Perform sensor adjustment

Sensor defect

Send in for repair

B-H90.0.01.DB2-2.1

25 / 30

9 | Disposal

G 1910-02
9 Disposal
Separation by material and recycling of device components and packaging must take place at the time of disposal. The valid regional statutory regulations and directives applicable at the time must be observed.
NOTE
The device must not be disposed of with household waste. Return it to us, freight prepaid. We will then arrange for the proper and environmentally- friendly disposal. Private end users in Germany have the possibility of dropping off the product at the municipal collection centre. Please dispose of empty batteries at the collection points intended for this purpose.
NOTE
Fill in the return form available from the information base online at www.ghm- group.de and sent it in with the product.

26 / 30

B-H90.0.01.DB2-2.1

G 1910-02
10 Technical data

Technical data | 10

Measuring range (specified accuracy) Measuring range (not specified)

0 .. 2000 ppm

0.000 .. 0.200 % 0.000 .. 1.000 %

Accuracy Measuring cycle Display Additional functions
Adjustment Housing

± 70 ppm ± 3 % measurement 2 seconds 3-line segment LCD, additional symbols, illuminated (adjustable white, permanent illumination) Min/Max/Hold TWA calculation / STEL 2-stage alarm (optical and acoustic) 1-point, 2-point and basic sensor adjustment Break-proof ABS housing

Protection rating Dimensions LWH [mm] and weight
Operating conditions Storage temperature Current supply

IP30 108 54 28 mm without measuring cell or kink protection 180 g, incl. battery and measuring cell 0 to 50 °C; 0 to 85 % r.h. (non-condensing) -20 to 70 °C 2*AA-NiMH batteries (included in delivery)

Current requirement/ Rechargeable battery life Battery indicator
Plug connector Auto-power-OFF function Directives and standards

approx. 50 mA, approx. 60 mA with lighting Life approx. 24 hours with NiMH batteries (without backlight) charging time of about 8 hours
4-stage charge status indicator, Charge indicator for low charge level: “BAT LO”
Micro USB socket (not a data connection)
The device switches off automatically if this is activated
The devices conform to the following Directives of the Council for the harmonisation of legal regulations of the Member States: 2014/30/EU EMC Directive 2011/65/EU RoHS Applied harmonised standards: EN 61326-1:2013 Emission limits: Class B Immunity according to Table 2 Additional errors: < 1 % FS EN 50581:2012

The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.

B-H90.0.01.DB2-2.1

27 / 30

11 | Spare parts and accessories

G 1910-02

Article Contact

11 Spare parts and accessories
A selection of spare parts and accessories for this product is listed below.

Number Name

Description

410355

NiMH rechargea- NiMH replacement rechargeable battery ble battery

411907 GKK 1002

G1000 series case, 235 x 185 x 48 mm

476698 GZ-18

Gas cylinder with 12 l test gas: 5000 ppm CO2

476699 GZ-19

Adjustment set for G 1910: MiniFlo valve for 12 l gas cylinders, quick-connect adapter 4 mm, closing screw, gas-bezel

479767 GZ-20

Gas connection for G 1910: Quick-connect adapter 6 mm, spare closing screw, gas bezel

481885

G 1000 table base and wall suspension

Plastic holder with connection possibility for USB cable

A complete list of all accessories- and spare parts is available in our product catalogue or on our home page. We can also provide further information by phone.

Internet:www.greisinger.de Tel: +49 94029383-52

28 / 30

B-H90.0.01.DB2-2.1

Contact

G 1910-02
12 Service
12.1 Manufacturer
If you have any questions, please do not hesitate to contact us:
GHM Messtechnik GmbH GHM GROUP – Greisinger Hans-Sachs-Str. 26 93128 Regenstauf | GERMANY Email: info@greisinger.de | www.greisinger.de WEEE reg. no. DE 93889386

Service | 12

12.2 Repairs processing
Defective products are repaired professionally and quickly in our service centre.

Open hours and contact

Monday to Thursday from 8:00 to 16:00 Friday from 8:00 to 13:00 GHM Messtechnik GmbH GHM GROUP – Greisinger Hans-Sachs-Str.26 Service Centre 93128 Regenstauf | GERMANY Tel: +49 94029383-39 Fax: +49 94029383-33 service@greisinger.de

NOTE

Fill in the return form available from the information base online at www.ghm- group.de and sent it in with the product.

North Sales Office Post code:
Email: Tel: Fax:

12.3 Sales offices
00000 ­ 25999 | 27000 ­ 34999 37000 ­ 39999 | 98000 ­ 99999 vertrieb-nord@ghm- messtechnik.de +49 4067073-0 +49 4067073-288

B-H90.0.01.DB2-2.1

29 / 30

12 | Service
West Sales Office Post code:
Email: Tel: Fax:
South Sales Office Post code: Email: Tel: Fax:

G 1910-02
26000 ­ 26999 | 35000 ­ 36999 40000 ­ 69999 vertrieb-west@ghm-messtechnik.de +49 2191 9672-0 +49 2191 9672-40
70000 ­ 97999 vertrieb-sued@ghm-messtechnik.de +49 9402 9383-52 +49 9402 9383-33

30 / 30

B-H90.0.01.DB2-2.1

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

GREISINGER User Manuals

Related Manuals