MSA ALTAIR 4XR Multi Gas Detector User Manual
- June 13, 2024
- MSA
Table of Contents
ALTAIR 4XR Multi Gas Detector
Product Information: The product is a Multigas Detector called
ALTAIR 4XR. It is designed to be used by trained and qualified
personnel for performing hazard assessments. The device can detect
various gases including toxic gases, oxygen concentration, and
combustible gases. It has a calibration certification and conforms
to safety regulations. Product Usage Instructions: 1. Correct Use:
The ALTAIR 4XR Multigas Detector should only be used by trained and
qualified personnel. It is important to perform a hazard assessment
before using the device. 2. Safety Regulations: Read and observe
the operating manual carefully. Follow the safety instructions and
guidelines provided. Take into account the national regulations
applicable in your country for safe use. 3. Environmental Factors:
Consider the environmental conditions before using the device. 4.
Turning ON and Fresh Air Setup: Follow the instructions to turn on
the device and set up fresh air. Fresh Air Setup (FAS) is explained
in detail in the manual. 5. Measurement Mode: Understand the
measurement mode for normal operation. 6. Device Setup: Learn how
to set up the device according to your requirements. 7. Bluetooth
Operation: Familiarize yourself with the Bluetooth operation
features of the device. 8. Data Logging: Understand how to log data
using the device. 9. Function Tests: Perform function tests as
instructed in the manual. 10. Bump Test: Learn how to perform a
bump test on the device. 11. Calibration: Follow the instructions
for calibrating the device. 12. Time of Day Testing: Understand how
to test the time of day using the device. 13. Maintenance: Follow
the maintenance procedures provided in the manual. 14.
Troubleshooting: Refer to the troubleshooting section in case of
any issues. 15. Live Maintenance Procedure – Replacing and Adding a
Sensor: Learn how to replace or add a sensor to the device. 16.
Cleaning: Follow the cleaning instructions to maintain the device.
17. Storage: Understand how to store the device properly. 18.
Technical Data: Refer to the technical data section for detailed
specifications. 19. Ordering Information: Find the necessary
information for ordering additional accessories or parts. 20.
Appendix: Explore additional features and setup options explained
in the appendix section. Note: It is important to read and observe
the entire operating manual for safe and proper use of the ALTAIR
4XR Multigas Detector.
Operating Manual
ALTAIR® 4XR
Multigas Detector
US
Order No.: 10175896/08 Print Spec: 10000005389 (EO)
CR: 800000057847
MSAsafety.com
WARNING! Read this manual carefully before using or maintaining the device.
The device will perform as designed only if it is used and maintained in
accordance with the manufacturer’s instructions. Otherwise, it could fail to
perform as designed, and persons who rely on this device could sustain serious
injury or death. The warranties made by MSA with respect to the product are
voided if the product is not installed and used in accordance with the
instructions in this manual. Please protect yourself and your employees by
following the instructions. Please read and observe the WARNINGS and CAUTIONS
inside. For additional information relative to use or repair, call
1800-MSA-2222 during regular working hours. For countries of Russian
Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector
will be delivered with a passport document that includes valid approval
information. On the CD with manual instruction attached to the gas detector
the user will find the documents “Type Description” and “Test Method” –
appendixes to Pattern Approval Certificate of Measuring instrument, valid in
the countries of use. MSA is a registered trademark of MSA Technology, LLC in
the US, Europe and other Countries. For all other trademarks visit
https://us.msasafety.com/Trademarks. This product incorporates Bluetooth®
wireless technology. The Bluetooth word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc., and any use of such marks by MSA is
under license. Other trademarks and trade names are those of their respective
owners. Versions of this product manufactured July 2022 and later may not
contain Bluetooth wireless technology. This will be indicated by the front
case of the detector not utilizing a Bluetooth logo. All references in this
manual to Bluetooth will not pertain to these versions of the device. This
device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including
interference that may cause undesired operation. You are cautioned that
changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment. English:
This device complies with RSS-210 of the Industry Canada Rules. Operation is
subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. French: Ce
dispositif est conforme à la norme CNR-210 d’Industrie Canada applicable aux
appareils radio exempts de licence. Son fonctionnement est sujet aux deux
conditions suivantes: (1) le dispositif ne doit pas produire de brouillage
préjudiciable, et (2) ce dispositif doit accepter tout brouillage reçu, y
compris un brouillage susceptible de provoquer un fonctionnement indésirable.
1000 Cranberry Woods Drive Cranberry Township, PA 16066 USA Phone
1-800-MSA-2222
For your local MSA contacts, please go to our website www.MSAsafety.com
© MSA 2022, 2023. All rights reserved
Contents
1
Calibration Certification and Conformance
5
1.1 Calibration Certification
5
1.2 Conformance Statement
5
2
Safety Regulations
5
2.1 Correct Use
5
2.2 Liability Information
6
2.3 Safety and Precautionary Measures to be Adopted
6
2.4 Warranty
8
3
Description
10
3.1 Overview
10
3.2 Device Hardware Interfaces
10
3.3 On-Screen Indicators
12
3.4 Battery Care
13
3.5 Viewing Additional Pages
15
3.6 Sensor Missing Alarm
17
3.7 Sensor End of Life Warning
17
3.8 Sensor End of Life Indicator
17
3.9 Monitoring Toxic Gases
17
3.10 Monitoring Oxygen Concentration
18
3.11 Monitoring Combustible Gases
19
4
Operation
19
4.1 Environmental Factors
19
4.2 Turning ON and Fresh Air Setup
20
4.2.1 Fresh Air Setup (FAS)
23
4.3 Measurement Mode [Normal Operation]
24
4.4 Device Setup
24
4.5 Bluetooth Operation
28
4.6 Data Logging
30
4.7 Function Tests
30
4.8 Bump Test
30
4.9 Bump LED
31
4.10 Calibration
31
4.11 Time of Day Testing
34
5
Maintenance
34
5.1 Troubleshooting
35
5.2 Live Maintenance Procedure – Replacing and Adding a Sensor
35
5.3 Cleaning
37
5.4 Storage
37
5.5 Scope of Delivery
37
6
Technical Data
38
6.1 Factory-set Alarm Thresholds and Setpoints
39
6.2 Performance Specification
40
6.3 XCell Sensor Patents
43
7
Ordering Information
45
8
Appendix
47
8.1 Start Up Sequence (Power on)
47
8.2 Fresh Air Setup (FAS)
48
US
ALTAIR 4XR
3
8.3 Reset Screen Controls
49
8.4 Bump Test
51
8.5 Options Setup
52
8.6 Sensor Setup
53
8.7 Calibrations
54
8.8 Alarm Setup
55
8.9 Time and Date Setup
56
8.10 Bluetooth Setup
56
4
ALTAIR 4XR
US
1 Calibration Certification and Conformance
1 Calibration Certification and Conformance
1.1 Calibration Certification
All applicable inspections, testing, and calibrations were performed using
NIST traceable equipment, where available, in accordance with MSA’s ISO 9001
Certified Quality System. Each material, component, and/or instrument must be
installed, operated and maintained in strict accordance with its labels,
cautions, warnings, instructions, and within the limitations stated in the
supplied instruction manual. Routine calibration checks, equipment
inspections, and applicable preventative maintenance measures must be
performed to verify that the materials, components, and/or instruments are
operating properly. Failure to perform these tasks on a routine basis, or
suggested intervals, with specified equipment or methods, may result in
inaccurate readings.
1.2 Conformance Statement
MSA certifies that the materials, components, and/or instruments delivered in
this shipment conform to all applicable specifications. The items delivered
have been processed through the appropriate approved document controlled
procedures for Receiving, Manufacturing and Inspection. The materials,
components, and/or instruments were inspected, tested, and calibrated, as
applicable, per the associated drawings, standards requirements, and/or
specifications, and were deemed acceptable by appropriate authorized
personnel.
2 Safety Regulations
2.1 Correct Use
The ALTAIR 4XR Multigas Detector is for use by trained and qualified
personnel. It is designed to be used when performing a hazard assessment to:
· Assess potential worker exposure to combustible and toxic gases and vapors
as well as low level of oxygen. · Determine the appropriate gas and vapor
monitoring needed for a workplace.
The ALTAIR 4XR Multigas Detector can be equipped to detect:
· Combustible gases and certain combustible vapors · Oxygen-deficient or
oxygen-rich atmospheres · Oxygen for monitoring inertization applications. The
device is suitable and certified for the measurement of the
oxygen concentration in gas mixtures for inertization according to EN 50104
but without alarm function. · Specific toxic gases for which a sensor is
installed.
NOTE: While the device can detect up to 30% oxygen in ambient air, it is
approved for use only up to 21% oxygen.
The ATEX directive is only valid until 25% Vol O2.
It is imperative that this operating manual be read and observed when using
the device. In particular, the safety instructions, as well as the information
for the use and operation of the device, must be carefully read and observed.
Furthermore, the national regulations applicable in the user’s country must be
taken into account for a safe use.
WARNING! · This product is supporting life and health. Inappropriate use,
maintenance or servicing may affect the function of the
device and thereby seriously compromise the user’s life. · Before use the
product operability must be verified. The product must not be used if the
function test is unsuccessful,
it is damaged, a competent servicing/maintenance has not been made, genuine
MSA spare parts have not been used. · Alternative use, or use outside this
specification will be considered as non-compliance. This also applies
especially to unauthorized alterations to the product and to commissioning
work that has not been carried out by MSA or authorized persons. Failure to
follow these warnings can result in serious personal injury or death.
US
ALTAIR 4XR
5
2 Safety Regulations
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
WARNING! This is a class A product in accordance with CISPR 22. In a domestic
environment, this product may cause radio interference, in which case the user
may be required to take adequate measures. Failure to follow this warning can
result in serious personal injury or death.
This Class A digital apparatus complies with Canadian ICES-003.
2.2 Liability Information MSA accepts no liability in cases where the product
has been used inappropriately or not as intended. The selection and use of
this product must be under the direction of a qualified safety professional
who has carefully evaluated the specific hazards of the jobsite where it will
be used and who is completely familiar with the product and its limitations.
The selection and use of this product and its incorporation into the safety
scheme of the jobsite is the exclusive responsibility of the employer. Changes
and modifications not expressly approved by the manufacturer will void the
user’s authority to operate the equipment.
2.3 Safety and Precautionary Measures to be Adopted
WARNING! Carefully review the following safety limitations and precautions
before placing this device in service. Do not alter or modify the device.
Failure to follow this warning can result in serious personal injury or death.
WARNING! All device readings and information must be interpreted by someone
trained and qualified in interpreting device readings in relation to the
specific environment, industrial practice and exposure limitations. Failure to
follow this warning can result in serious personal injury or death.
Check Function Before each day’s use, check the function of the device ( 4.7
Function Tests). MSA recommends carrying out a routine inspection prior to
each day’s use. Bluetooth Operation If your device contains Bluetooth
functionality, Bluetooth operation is dependent upon signal availability of
the wireless service(s) necessary to maintain the communication link. Loss of
wireless signal will prevent communication of alarms and other information to
linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs. Perform a Bump Test
6
ALTAIR 4XR
US
2 Safety Regulations
Bump test frequency is often stipulated by national or corporate regulations;
however, bump testing before each day’s use is generally the accepted best
safety practice and is therefore MSA’s recommendation. The device must pass
the bump test. If it fails the test, perform a calibration before using the
device.
Perform a bump test (see 4.8 Bump Test) more frequently if the device is
subjected to physical shock or high levels of contaminants. Also, perform a
bump test more frequently if the tested atmosphere contains the following
materials, which may desensitize the combustible gas sensor and reduce its
readings:
· Organic silicones · Silicates · Lead-containing compounds · Hydrogen
sulphide exposures over 200 ppm or exposures over 50 ppm for one minute.
Check minimum concentration of a combustible gas
The minimum concentration of a combustible gas in air that can ignite is
defined as the Lower Explosive Limit (LEL). A combustible gas reading of “XXX”
indicates the atmosphere is above 100 % LEL or 5.00 % vol CH4, and an
explosion hazard exists. Move away from hazardous area immediately.
Observe atmosphere
Do not use the device to test for combustible or toxic gases in the following
atmospheres as this may result in erroneous readings:
· Oxygen-deficient or oxygen-rich atmospheres · Reducing atmospheres · Furnace
stacks · Inert environments · Atmospheres containing combustible airborne
mists/dusts.
Use the device only to detect gases/vapors for which a sensor is installed.
The device is suitable and certified for the measurement of the oxygen
concentration in gas mixtures for inertization according to EN 50104 but
without alarm function.
Make sure >10% oxygen is present for accurate combustible readings with the
catalytic sensor.
Not to be used for gases having a flash point in excess of 38 °C (100 °F)
Do not use the device to test for combustible gases in atmospheres containing
vapors from liquids with a high flash point (above 38 °C, 100 °F), as this may
result in erroneously low readings.
Physical shock
Recheck calibration if device is subjected to physical shock.
Sensor maintenance
Do not block sensor openings as this may cause inaccurate readings. Do not
press on the face of the sensors, as this may damage them and cause erroneous
readings. Do not use compressed air to clean the sensor holes, as the pressure
may damage the sensors.
Observe Proper Time for Display Stabilizing
Allow sufficient time for device to display accurate reading. Response times
vary based on the type of sensor being utilized.
Observe proper battery maintenance
Use only battery chargers made available by MSA for use with this device;
other chargers may damage the battery pack and the device. Dispose of in
accordance with local health and safety regulations.
US
ALTAIR 4XR
7
2 Safety Regulations
Use of the MSA GALAXY® GX2 Automated Test System is an alternate approved
method for charging ALTAIR 4XR devices.
As the battery pack ages, there will be a reduction in usable device runtime.
Risk of explosion: Do not recharge device in hazardous area.
Be Aware of Environmental Conditions
A number of environmental factors may affect the sensor readings, including
changes in pressure, humidity and temperature. Pressure and humidity changes
also affect the amount of oxygen actually present in the atmosphere.
Be Aware of the Procedures for Handling Electrostatically Sensitive
Electronics
The device contains electrostatically sensitive components. Do not open or
repair the device without using appropriate electrostatic discharge (ESD)
protection. The warranty does not cover damage caused by electrostatic
discharges.
Be Aware of the Product Regulations
Follow all relevant national regulations applicable in the country of use.
Be Aware of the Warranty Regulations
The warranties made by MSA The Safety Company with respect to the product are
voided if the product is not used and maintained in accordance with the
instructions in this manual. Please protect yourself and others by following
them. We encourage our customers to write or call regarding this equipment
prior to use or for any additional information relative to use or service.
Special Conditions for Safe Usage
· In the event of combustible sensor overrange, the device will enter a Lock
Alarm state, which must be reset in a fresh air environment. To reset this
alarm, cycle power off and on, in fresh air. Keep the device in the fresh air
environment until LEL or CH4 readings have stabilized and then follow the
Fresh Air Setup and Zero Calibration instructions contained in this manual.
· The RF radiation power used to activate the RFID tag antenna shall not
exceed 6 Watts for EPL Group I applications or 2 Watts for EPL Group IIC
applications.
2.4 Warranty
ITEM Chassis and electronics All sensors unless otherwise specified XCell EX-H Sensor In-Box Accessories including Replacements
WARRANTY PERIOD* Four years Four years One year Two years
*Period begins date of received shipment.
Specific battery run-time over temperature is not warranted.
This warranty does not cover filters, fuses, etc. As the battery pack ages, there will be a reduction in usable device run time. Certain other accessories not specifically listed here may have different warranty periods. This warranty is valid only if the product is maintained and used in accordance with Seller’s instructions and/or recommendations.
The Seller shall be released from all obligations under this warranty in the event repairs or modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the Seller to any affirmation, representation or warranty concerning this product. Seller makes no warranty concerning components or accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
8
ALTAIR 4XR
US
2 Safety Regulations
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY
DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR
PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser’s sole and exclusive remedy for breach
of the above warranty, for any tortuous conduct of Seller, or for any other
cause of action, shall be the replacement at Seller’s option, of any equipment
or parts thereof, which after examination by Seller is proven to be defective.
Replacement equipment and/or parts will be provided at no cost to Purchaser,
F.O.B. Seller’s Plant. Failure of Seller to successfully replace any
nonconforming equipment or parts shall not cause the remedy established hereby
to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will
seller be liable to purchaser for economic, special, incidental or
consequential damages or losses of any kind whatsoever, including but not
limited to, loss of anticipated profits and any other loss caused by reason of
non-operation of the goods. This exclusion is applicable to claims for breach
of warranty, tortuous conduct or any other cause of action against seller.
US
ALTAIR 4XR
9
3 Description
3 Description
3.1 Overview
Figure 1 Device overview
1
Datalink Communication port
8
Display
2
Bump LED (green/red) and Fault LED (-yellow)
9
Alarm LEDs (4)
3 Sensor Inlets
10 Belt Clip
4 Horn
11 Charging Connection
5 p Button
12 Screws (4)
6 q Button
13
Charge LED (red/green/orange)
7
Button
14 Bluetooth Status LED
The device monitors gases in ambient air and in the workplace.
It is available with a maximum of three sensors, which can display readings
for four separate gases (one Two-Tox Sensor can detect two toxic gases with a
single sensor).
While the device can detect up to 30% oxygen in ambient air, it is approved
for use only up to 21% oxygen.
The alarm levels for the individual gases are factory-set and can be changed
through the Setup Menu. These changes can also be made through the GALAXY GX2
Test Stand or MSA Link software. Ensure that the latest version of the GALAXY
GX2 firmware or MSA Link software has been downloaded from MSA’s website
www.msasafety.com.
If gas is present during Fresh Air Setup, the device will fail and enters Normal Operation mode.
3.2 Device Hardware Interfaces
Device operation is dialog driven from the display with the aid of the three
function buttons ( 3.2 Device Hardware Interfaces).
The device has three buttons for user operation. Each button can function as a
“soft key”, as defined on the display directly above the button.
10
ALTAIR 4XR
US
3 Description
Button Definitions
Button q p
Description
The button is used to turn device on or off and to confirm user action
selections. When the p button and the button are pressed simultaneously for
device start-up, the Options Setup Mode displays.
The q button is used to move forward through data screens in measuring mode,
or as page back and to decrease the values in set-up mode. Holding this button
for 3 seconds while in Normal Operation mode activates the InstantAlert alarm.
The p button is used to reset peak, STEL TWA and acknowledge alarms (where
possible) or access calibration in measuring mode. It is also used as page up
or to increase the values in set-up mode.
LED Definitions
LED GREEN/RED(Bump LED)
RED(Alarm LED) YELLOW (Fault LED)
RED/GREEN/ORANGE (Charge LED)
BLUE (Bluetooth Status)
Description After a successful Bump Test, the green LED will flash every 15
seconds for a 24-hour period.
When the device fails a bump test or when the 24-hour period expires, the red
LED flashes every 15 seconds.
This option can be turned off through the MSA GALAXY GX2 Test Stand or MSA
Link software. The red alarm LEDs are visual indications of an alarm condition
or any type of error in the device. The yellow fault LED is a visual
indication of an device fault condition. This LED turns on with the following
conditions:
· Device memory error · Sensor Missing · Sensor Error The charge LED is a
visual indication of charge status.
· RED: device is charging · GREEN: charge is complete · ORANGE: problem
detected during charging If your instrument supports Bluetooth, the blue LED
is a visual indication of the Bluetooth connection status.
· Off = Bluetooth OFF or Undiscoverable · Fast Flash = Discoverable Mode ·
Slow Flash = Connected
Vibrating Alarm
The device is equipped with a vibrating alarm.
Backlight
The backlight automatically activates when any front panel button is pressed. The backlight remains on for the duration of the user-selected timeout. This duration can be set through MSA GALAXY GX2 Test Stand or MSA Link software.
Horn
The horn provides an audible alarm.
US
ALTAIR 4XR
11
3 Description
Operating Beep This operating beep activates every 30 seconds by momentarily
beeping the horn and flashing the alarm LEDs under the following conditions:
· Operating beep is enabled · Device is in Normal Operation mode · Device is
not in battery warning · Device is not in gas alarm · Audible and visual
options are enabled
3.3 On-Screen Indicators
Figure 2 Display
1 Graphic Symbols 2 Gas Type
3 Gas Concentration
Alarm Symbol Indicates alarm state
Motion Alert Indicates Motion Alert is active Bump Check Symbol Indicates successful bump or cal
Indicates required interaction
Battery Condition Indicates the battery charge level Sensor Labels
Calibration Setup Cal Gas Cylinder Indicates cal gas must be applied
12
ALTAIR 4XR
US
3 Description No Gas Cylinder Indicates cal gas should not be applied and
device must be exposed to fresh air. Hourglass Indicates user should wait
Minimum Indicates a minimum value or low alarm PEAK Symbol Indicates a
PEAK reading or high alarm STEL Symbol Indicates a STEL alarm
TWA Symbol Indicates a TWA alarm Sensor life Symbol Indicates the end of
sensor life Bluetooth Symbol – Indicates enabled Bluetooth operation (if
applicable)
3.4 Battery Care Battery Life Indicator The battery condition icon is continuously displayed in the upper right-hand corner of the display. As the battery is depleted, battery icon segments blank until only the battery icon outline remains. Each indicator segment represents approximately 25 % of the total battery capacity. Battery Warning
WARNING! If battery warning alarm activates while using the device, leave the
area immediately as the end of battery life is approaching. Failure to follow
this warning can result in serious personal injury or death.
The nominal run-time of the device at room temperature is 24 hours. The
runtime with an active Bluetooth connection is approximately 22 hours. The
runtime of the device at -20 °C/- 4 °F will be approximately 14 hours.
Actual run-time will vary depending on ambient temperature and alarm
conditions.
The alarm levels for the individual gases are factory-set and can be changed
through the Setup Menu.
A Low Battery Warning indicates that a nominal 30 minutes of operation remain
before the battery’s charge is depleted.
The duration of remaining device operation during a Low Battery Warning depends on ambient temperatures.
When the device goes into Low Battery Warning, the:
· battery life indicator flashes · alarm sounds
US
ALTAIR 4XR
13
3 Description
· alarm LEDs flash
· display shows “LOW BATT” and · device repeats this warning every 60 seconds
and continues to operate until it is turned off or battery shutdown
occurs. Battery Shut Down
WARNING! If the battery shutdown alarm activates, stop using the device as it
no longer has enough power to indicate potential hazards, and persons relying
on this device for their safety could sustain serious personal injury or
death.
The device goes into battery shutdown mode 60 seconds before final shutdown
(when the batteries can no longer operate the device):
· On the display flashes “BATT ALARM” and · Alarm sounds and lights flash;
alarm cannot be silenced, · No other pages can be viewed; after approximately
one minute, the device -automatically turns off. When battery shutdown
condition occurs: 1. Leave the area immediately. 2. Recharge the battery.
Battery Charging
WARNING! Risk of explosion: Do not recharge device in hazardous area. Failure
to follow this warning can result in serious personal injury or death.
WARNING! · Use of any charger, other than the charger supplied with the
device, may damage or improperly charge the batteries. · The charger is
capable of charging a completely depleted pack in less than four hours in
normal, room-temperature
environments. Failure to follow these warnings can result in serious personal
injury or death.
Allow very hot or cold devices to stabilize for one hour at room temperature
before attempting to charge.
To Charge the Device · Ensure the charger connector is firmly seated in the
charge port on the back of the device. · The battery symbol will scroll
through a progressively increasing number of segments and the charge LED will
be red until 90 % of full charge has been obtained. Then the battery symbol
will remain fully illuminated and the charge LED will be green while the
battery is trickle charged to its full capacity. · If a problem is detected
during charging, the battery symbol will flash and the charge LED will be
orange. Disconnect the device from the power module and reconnect it to the
power module to reset the charge cycle. · During periods of non-use, the
charger may remain connected to the device/battery pack. · Minimum and maximum
ambient temperature to charge the device is 10 °C (50 °F) and 35 °C (95 °F),
respectively. · For best results, charge the device at room temperature 23 °C
(73 °F).
14
ALTAIR 4XR
US
3 Description
3.5 Viewing Additional Pages The Main Screen appears at device turn-on.
Optional displays can be viewed by pressing the q button to move to: Bump Mode
1. To select the Bump mode, press the button. 2. To move forward, press the q
button. 3. To move backward to the Main page, press the p button. Peak
Readings (PEAK page) The peak icon ( 3.3 On-Screen Indicators) shows the
highest levels of gas recorded by the device since turn-on or since peak
readings were reset. To reset the peak readings: 1. Access the PEAK page. 2.
Press the p button. Minimum Readings (MIN page) This page shows the lowest
level of oxygen recorded by the device since turn-on or since the MIN reading
was reset. The minimum icon (3.3 On-Screen Indicators) appears on the display.
To reset the MIN reading: 1. Access the MIN page. 2. Press the p button. Short
Term Exposure Limits (STEL page)
WARNING! If the STEL alarm activates, leave the contaminated area immediately;
the ambient gas concentration has reached the preset STEL alarm level. Failure
to follow this warning will cause over-exposure to toxic gases and persons
relying on this product for their safety could sustain serious personal injury
or death. Failure to follow this warning can result in serious personal injury
or death.
The STEL icon ( 3.3 On-Screen Indicators) appears on the display to show the
average exposure over a 15 minute period. When the amount of gas detected by
the device is greater than the STEL limit:
· Alarm sounds · Alarm LEDs flash · STEL icon flashes. To reset the STEL: 1.
Access the STEL page. 2. Press the p button. The STEL alarm is calculated over
a 15 minute exposure. STEL calculation examples: Assume the device has been
running for at least 15 minutes:
US
ALTAIR 4XR
15
3 Description
15 minute exposure of 35 ppm: (15 minutes x 35 ppm) 15 minutes
= 35 ppm
10 minute exposure of 35 ppm and 5 minutes exposure of 5 ppm: (10 minutes x 35 ppm) + (5 minutes x 5 ppm) = 25 ppm 15 minutes
This page can be de-activated through MSA GALAXY GX2 Test Stand or MSA Link. Time Weighted Average (TWA page)
WARNING!
If the TWA alarm activates, leave the contaminated area immediately; the
ambient gas concentration has reached the preset TWA alarm level. Failure to
follow this warning will cause over-exposure to toxic gases and persons
relying on this product for their safety could sustain serious personal injury
or death.
The TWA icon ( 3.3 On-Screen Indicators) appears on the display to show the
average exposure since the device was turned on or since the TWA reading was
reset. When the amount of gas detected is greater than the eight-hour TWA
limit:
· Alarm sounds · Alarm LEDs flash · TWA icon flashes.
To reset the TWA:
1. Access the TWA page. 2. Press the p button.
The TWA alarm is calculated over an eight-hour exposure.
TWA calculation examples:
1 hour exposure of 50 ppm: (1 hour x 50 ppm) + (7 hours x 0 ppm) = 6.25 ppm 8
hours
4 hour exposure of 50 ppm and 4 hour exposure of 100 ppm:
(4 hours x 50 ppm) + (4 hours x 100 ppm)
= 75 ppm
8 hours
12 hour exposure of 100 ppm: (12 hours x 100 ppm) 8 hours
= 150 ppm
This page can be deactivated through the MSA GALAXY GX2 Test Stand or MSA Link software. Time Display Current time appears on the display in a 12-hour format by default. A 24-hour format can be selected using MSA GALAXY GX2 Test Stand or MSA Link. Date Display
16
ALTAIR 4XR
US
3 Description
Current date appears on the display in the format: MMM-DD-YYYY.
Discoverable Mode page
Allows the user to put the device into Bluetooth discoverable mode in order to
pair with another device. This page can be deactivated through the BT SETUP
page.
Motion Alert Activation To activate or deactivate the Motion Alert feature,
press the p button while the Motion Alert Activation page is displayed. When
the Motion Alert feature is active, the Motion Alert symbol ( 3.3 On-Screen
Indicators) will flash every 3 seconds. The device will enter pre-alarm when
no motion is detected for 20 seconds. This condition can be cleared by moving
the device.
After 30 seconds of inactivity, the full Motion Alert alarm is triggered. This
alarm can only be cleared by pressing the p button.
3.6 Sensor Missing Alarm
The device enters the Sensor Missing alarm if the device detects that the
sensor is not properly installed in the device or is not functional.
If a sensor is detected as missing, the following occurs:
· “SENSOR ERROR” displays. · The flag above the sensor detected as missing
flashes on the display. · Alarm sounds and lights flash. · Yellow Fault LED is
on solid. · If there is a sensor error at startup, the device shuts off in 60
seconds.
3.7 Sensor End of Life Warning
If a sensor is nearing its end of life, the device will warn the user
following a sensor calibration. The sensor is still fully functional at this
point, but the warning gives the user time to plan for a replacement sensor
and minimize downtime. The symbol will be continuously displayed. For more
details, see 4.10 Calibration.
3.8 Sensor End of Life Indicator
If the device cannot calibrate one or more sensors, the device will display
“SPAN ERR” followed by the Alarm symbol and symbol to indicate end of sensor
life. For more details, see 4.10 Calibration.
3.9 Monitoring Toxic Gases
The device can monitor the concentration of the following toxic gases in
ambient air:
· Carbon Monoxide (CO) · Hydrogen Sulfide (H2S) · Sulfur Dioxide (SO2) ·
Nitrogen Dioxide (NO2) The device displays the gas concentration in parts per
million (PPM) or mg/m3 on the Normal Operations page until another page is
selected or the device is turned off.
WARNING!
If an alarm activates while using the device, leave the area
immediately.Remaining in the area under such circumstances can cause serious
personal injury or death.
The device has four gas alarms for each toxic gas: · HIGH Alarm
US
ALTAIR 4XR
17
3 Description
· LOW Alarm · STEL Alarm · TWA Alarm
If the gas concentration reaches or exceeds the alarm set point, the device:
· backlight turns on · triggers a vibrating alarm · displays and flashes the
Alarm symbol and either the Minimum icon (LOW alarm) or the Maximum (PEAK)
icon
(HIGH alarm) · enters an alarm state.
When a gas reading exceeds the full scale range of the sensor, the device
displays “+ + +” in place of the actual reading.
When a gas reading falls below a defined under-range threshold, the device
displays ” – – -” in place of the actual reading.
3.10 Monitoring Oxygen Concentration The device monitors the oxygen
concentration in ambient air. The alarm set points can be set to activate on
two different conditions:
· Enriched – oxygen concentration > 20.8% or · Deficient – oxygen
concentration < 19.5%.
While the device can detect up to 30% oxygen in the ambient air, it is
approved for use only up to 21% oxygen-content.
WARNING! If an alarm activates while using the device, leave the area
immediately.Remaining in the area under such circumstances can cause serious
personal injury or death.
When the alarm set point is reached for either of the above conditions:
· an alarm sounds · alarm LEDs flash · a vibrating alarm triggers · device
displays and flashes the Alarm icon and either the Minimum icon (Enriched
alarm) or the Maximum icon
(Deficient alarm) ( 3.3 On-Screen Indicators) along with the corresponding
oxygen concentration.
The LOW alarm (oxygen deficient) is latching and will not automatically reset
when the O2 concentration rises above the LOW set point. To reset the alarm
press the p button. If the alarm condition still exists, the p button only
silences the alarm for five seconds.
False oxygen alarms can occur due to changes in barometric pressure (altitude)
or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at the temperature
and pressure of use. Be sure that the device is in known fresh air before
performing a calibration.
When a gas reading exceeds the full scale range of the sensor, the device
displays “+ + +” in place of the actual reading.
When a gas reading falls below a defined under-range threshold, the device
displays ” – – -” in place of the actual reading.
18
ALTAIR 4XR
US
4 Operation
3.11 Monitoring Combustible Gases The device can monitor these concentrations
in ambient air:
· Methane · Combustible gases The device displays the combustible gas
concentration in % LEL or % CH4 on the Measuring page until another page is
selected or the device is turned off.
WARNING! If an alarm activates while using the device, leave the area
immediately.Remaining in the area under such circumstances can cause serious
personal injury or death.
The device has two alarm set points: · HIGH Alarm · LOW Alarm
If the gas concentration reaches or exceeds the alarm set point, the device: ·
backlight turns on · a vibrating alarm triggers · displays and flashes the
Alarm symbol and either the Minimum icon (LOW alarm) or the Maximum (PEAK)
icon (HIGH alarm) · enters an alarm state.
When a gas reading reaches 100% LEL or 5.00% CH4, the device enters a Lock
Alarm state: the combustible sensor shuts down and displays “XXX” in place of
the actual reading. This state can only be reset by turning the device off and
on in a fresh air environment. When a gas reading falls below a defined under-
range threshold, the device displays ” – – -” in place of the actual reading.
WARNING! · A combustible gas reading of “100” or “5” indicates the atmosphere
is above 100% LEL or 5.00% vol CH4
respectively, and an explosion hazard exists. Move away from contaminated area
immediately.
· In such cases, the device enters a Lock Alarm state. Failure to follow this
warning can result in serious personal injury or death.
Check your national standard values for 100 % LEL (EN60079-20-1).
4 Operation
Device operation is dialog driven from the display with the aid of the three
Function buttons (see Figure 1 Device overview).
4.1 Environmental Factors A number of environmental factors may affect the gas
sensor readings, including changes in pressure, humidity and temperature.
Pressure and humidity changes affect the amount of oxygen actually present in
the atmosphere.
Pressure Changes If pressure changes rapidly (e.g., stepping through airlock),
the oxygen sensor reading may temporarily shift and possibly cause the device
to go into alarm. While the percentage of oxygen may remain at or near 20.8
Vol %, the total amount of
US
ALTAIR 4XR
19
4 Operation
oxygen present in the atmosphere available for respiration may become a hazard
if the overall pressure is reduced by a significant degree.
Humidity Changes
If humidity changes by any significant degree (e.g., going from a dry, air
conditioned environment to outdoor, moisture laden air), oxygen readings can
be reduced by up to 0.5 %, due to water vapor in the air displacing oxygen.
The oxygen sensor has a special filter to reduce the effects of humidity
changes on oxygen readings. This effect will not be noticed immediately, but
slowly impacts oxygen readings over several hours.
Temperature Changes
The sensors have built-in temperature compensation. However, if temperature
shifts dramatically, the sensor reading may shift. Zero the device at the work
site temperature for the least effect.
4.2 Turning ON and Fresh Air Setup
Device operation is dialog driven from the display with the aid of the three
Function buttons (see Figure 1 Device overview).
For more information, see the flow charts in 8 Appendix.
1. Turn the device on with the button.
The device performs a self test and then goes to Fresh Air Setup:
· all display segments are activated · audible alarm sounds · alarm LEDs light
· vibrating alarm is activated.
During the self test, the device checks for missing sensors. In the case of a
missing sensor, the device displays the Sensor Missing screen and alarms until
it is turned off. Otherwise, the turn-on sequence continues.
The device displays:
· Alarm & display self test · Manufacturer name · Device name · Software
version · FCC ID · IC ID · Bluetooth software version (if activated) · Sensor
discovery · Combustible gas type · Toxic gas units · Alarm set points (PEAK,
MIN) (STEL, TWA) · Calibration values · Date and time display · Last cal date
(if activated) · CAL due date (if activated) · Fresh Air Setup option.
Screen Displays during Startup
During the startup sequence, all automatic page display timeouts are preset to a range from two to four
seconds.
20
ALTAIR 4XR
US
Several sequences and screens occur during start up: Device Self Test The
device performs a self test.
Device Name and Software version Software version and device name display.
4 Operation
Combustible Gas Type
Name of Combustible Gas Type displays, e.g. BUTANE.
NOTE: Combustible gas type can be changed manually through the SENSOR SETUP
menu or the MSA GALAXY GX2 Test Stand or MSA Link software. Toxic Gas Units
Name of Toxic Gas Units displays (ppm or mg/m3).
NOTE: Toxic units can only be modified through the MSA GALAXY GX2 Test Stand
or MSA Link software.
US
ALTAIR 4XR
21
4 Operation
Alarm Set points Alarm set points for all installed and activated sensors
display. LOW alarm set points display, followed by HIGH alarm set points.
NOTE: Alarm set points can be changed manually through the -Setup menu or the
MSA GALAXY GX2 Test Stand or MSA Link software.
STEL and TWA Set points The preset STEL and TWA values for installed and
activated sensors display.
Calibration Values The preset calibration values for installed and activated
sensors display.
Time and Date The date displays in a month, day and year format. NOTE: In the
event that the battery is fully discharged, the time and date reset. At
startup, the user is prompted to enter the time and date. If the time and date
information is missing, they are reset to (Jan-01-2016) with time stamp
(00:00).
22
ALTAIR 4XR
US
Last CAL Date and CAL Due
NOTE: These display options can be set by MSA GALAXY GX2 Test Stand or MSA
Link software. If these options are not set, these screens are not displayed.
· By default, Last Cal is activated. · By default, Cal Due is deactivated.
4 Operation
Fresh Air Setup (FAS) The FAS screen is prompted.
4.2.1 Fresh Air Setup (FAS) The FAS is for automatic Zero calibration of the
device. The Fresh Air Setup has limits. The zero of any sensor that is outside
of these limits will not be adjusted by the FAS command.
WARNING! Do not activate the Fresh Air Setup unless you are certain you are in
fresh, uncontaminated air; otherwise, inaccurate readings can occur which can
falsely indicate that a hazardous atmosphere is safe. If you have any doubts
as to the quality of the surrounding air, do not use the Fresh Air Setup
feature. Do not use the Fresh Air Setup as a substitute for daily calibration
checks. The calibration check is required to verify span accuracy. Failure to
follow this warning can result in serious personal injury or death.
If a battery charging cycle is interrupted before it is completed (4 hours for
a fully discharged battery), allow the device’s internal temperature to
stabilize for 30 minutes before performing a Fresh Air Setup.
Figure 3 Fresh Air Setup
If this option is enabled, the device displays “FAS?”, prompting the user to perform a Fresh Air Setup.
1. Press the p button to bypass the Fresh Air Setup.
· The Fresh Air Setup is skipped and the device goes to the Measuring page (Main page).
2. Press the button within 10 seconds to perform the Fresh Air Setup.
· The device starts the FAS.
· The screen shows a No Gas Symbol, a blinking hourglass, and all enabled gas sensor readings.
US
ALTAIR 4XR
23
4 Operation · At the end of the FAS Calibration, the device displays “FAS
PASS” or “FAS ERR” along with the flags of the sensors that were outside of
the FAS limits. All sensors that are within the FAS limits will be zeroed.
4.3 Measurement Mode [Normal Operation] In Normal Operation mode, the user can
check the Minimum and Peak readings prior to clearing the STEL and TWA values
or performing a Span and Zero Calibration. The following options pages can be
executed from the Normal Operation screen: Bump Page This page allows the user
to perform a bump test.
Peak Page This page shows the peak readings for all sensors.
Min Page This page shows the minimum reading for the oxygen sensor.
STEL Page This page shows the calculated STEL readings of the device.
TWA Page This page shows the calculated TWA readings of the device.
Time / Date Page This page shows actual time and date settings of the device.
Discoverable Mode This page allows the user to put the device into Bluetooth
discoverable mode in order to pair with another device.
Motion Alert (optional) This page allows the Motion Alert Feature to be
-activated or deactivated.
Using the three device buttons, the user can navigate through each sub-menu in a top/down sequence.
Refer to Figure 1 Device overview and 8 Appendix for detailed instructions on navigating through these screens.
4.4 Device Setup
The setup menus can be accessed only when the device is turned on while pressing and holding the p button.
24
ALTAIR 4XR
US
4 Operation
This mode can only be activated at device turn-on. The operation is as
follows:
1. Press and hold the p button while turning the device on. 2. Use the p and
q buttons to enter the setup password. The default password is “672”. 3. Press
button to enter the setup menus.
After two incorrect password entries, device enters Normal Operation mode.
Correct password: device continues/beeps once.
The password can be changed through the MSA GALAXY GX2 Test Stand or MSA Link
software.
In the Setup mode: · Press the button to store chosen value or go to the next
page. · Press p button to increase values by one or toggle a selection on or
off. · Press and hold p button to increase values by 10. · Press q button to
decrease values by one or toggle a selection on or off. · Press and hold q
button to decrease values by 10.
The following options are available by pressing the q and p buttons: · Sensor
Setup (SENSOR SETUP) · Calibration Setup (CAL SETUP) · Alarm Setup (ALARM
SETUP) · Setup Time and Date (TIME DATE) · Bluetooth Setup (BT SETUP) · EXIT
4.4.1 Sensor Setup Each sensor can be turned on or off. For more information,
see the flow charts in 8.6 Sensor Setup.
Figure 4 Sensor Setup 1. To bypass this setup, press the q or p button; otherwise, continue as follows. 2. Press the button to enter the submenu. 3. Use the q or p button to change the option and confirm with the button. 4. Repeat this procedure for all other sensors. 5. After setting the last sensor, continue to Calibration Setup.
US
ALTAIR 4XR
25
4 Operation
4.4.2 Calibration Setup The user can change and set the calibration values for
each sensor. It is also possible to
· select whether the Cal Due screen is displayed · set the number of days
until the next calibration is due · select whether the device password screen
is displayed to protect Bump and Calibration operations For more information,
see the flow charts in 8.7 Calibrations.
Figure 5 Calibration Setup 1. To bypass this setup, press the q or p button;
otherwise, continue as follows. 2. Press the button to enter the submenu. The
calibration gas concentration of the first sensor is shown. 3. Press the q or
p button to change the value. 4. Press the button to store the value. Setup
screen for the next sensor is displayed. 5. Repeat the procedure for all other
sensors. After setting the last sensor, the user is prompted to set CALDUE. 6.
Press the q or p button to enable or disable CALDUE. 7. Confirm with the
button. 8. If CALDUE is set on, press the q or p button to select the number
of days 9. Confirm with the button. 10. Press the q or p button to enable or
disable the calibration password (CAL PW) screen. 11. Confirm with the button.
12. After confirmation, continue to Alarm setup.
4.4.3 Alarm Setup
The user can switch all alarms on or off and change the alarm set points for
each sensor. For more information, see the flow charts in 8.8 Alarm Setup.
See 6.1 Factory-set Alarm Thresholds and Setpoints for alarm adjustment
limits. The value of the High Alarm can only be set to a value that is higher
than the Low Alarm set point.
26
ALTAIR 4XR
US
4 Operation
Figure 6 Alarm Setup 1. To bypass this setup, press the q or p button;
otherwise, continue as follows. 2. Press the button to enter the submenu. 3.
Set alarms on or off by pressing the q or p button. 4. Press the button to
confirm the selection. LOW ALARM settings for the first sensor display. 5.
Press the q or p button to change the value. 6. Press the button to store the
value. HIGH ALARM settings for the first sensor display. 7. Press the q or p
button to change the value. 8. Press the button to store the value. STEL ALARM
settings (for toxic sensors only) display. 9. Press the q or p button to
change the value. 10. Press the button to store the value. TWA ALARM settings
(for toxic sensor only) for display. 11. Press the q or p button to change the
value. 12. Press the button to store the value. 13. Repeat the procedure for
all other sensors. 14. After setting the last sensor, continue to Time and
Date setup.
60% LEL or 3.0% volume of methane is the maximum High Alarm set point that can
be programmed.
The combustible alarm can be turned OFF by the user in the instrument setup.
When the combustible alarm is turned OFF, the only indicator to the user that
the combustible alarm is turned OFF occurs during power up of the instrument
at which a startup screen will indicate that the combustible alarm is turned
OFF. When turned ON, the combustible high alarm is latching.
The combustible alarm can be silenced momentarily by pressing the p button.
However, if the gas concentration causing the alarm is still present, the
device will go back into alarm.
4.4.4 Setup Time and Date
This submenu is for setting date and time.
For more information, see the flow charts in 8.9 Time and Date Setup.
US
ALTAIR 4XR
27
4 Operation
Figure 7 Date and Time Setup 1. To bypass this setup, press the q or p button;
otherwise, continue as follows. 2. Press the button to enter the submenu. 3.
Set month by pressing the q or p button. 4. Press the button to confirm month.
5. Repeat this procedure for day, year, hours and minutes. By default, time is
displayed in 12 hour format. The BT SETUP screen will be displayed next.
4.4.5 Enable Bluetooth Operation Some devices are configured with a Bluetooth
capable communications feature.
1. Press the p button to enable or disable the Bluetooth communications
device (ON/OFF). NOTE: Only devices with the Bluetooth icon on the label are
capable of Bluetooth pairing. For devices that do NOT support Bluetooth, the
device will go into a Bluetooth Error if setup is attempted. The instrument
must be placed back in “Bluetooth OFF” mode to operate.
2. Press the button to accept the setting and return to the EXIT? page.
4.5 Bluetooth Operation
NOTE: Versions of this product manufactured July 2022 and later may not
contain Bluetooth wireless technology. This will be indicated by the front
case of the detector not utilizing a Bluetooth logo. All references in this
manual to Bluetooth will not pertain to these version of the device.
The Bluetooth communication device must be enabled for any Bluetooth functions
to operate. See 4.4 Device Setup.Compatible Bluetooth host with appropriate
software is required for proper operation.
Bluetooth Security
28
ALTAIR 4XR
US
4 Operation
The Bluetooth connection is encrypted and secured with a unique six digit pin
that must be double confirmed on both device and Bluetooth host at the time of
pairing.
Discovery Mode
This device mode is used to enable a Bluetooth host to pair with the device
for the first time or if a different Bluetooth host was connected with the
device previously.
Note that the device will automatically enter discovery mode for five minutes
at device turn on if Bluetooth has been enabled. Discovery mode will also be
entered for 5 minutes following a disconnection.
To manually enter Discovery Mode: 1. Page down through the menu pages in
Normal Operation mode using the q button until the Discovery Mode page is
displayed. 2. Press the p button until the display shows dSCVRY YES. 3. Press
the button to enter Discovery Mode. The blue LED will blink rapidly indicating
that the device is in Discovery Mode.
Connecting the Device to a Bluetooth Host for the First Time
1. Ensure that the device is on and in Discovery Mode 2. On the Bluetooth
host, locate the Bluetooth device list. Select “A4X-xxxxxxxx” from the list.
Both the device and Bluetooth host will display a unique six digit security
code to ensure that the correct devices are being paired. 3. After verifying
that the six digit codes match, confirm the pairing request on the device by
pressing the q button. 4. Confirm on the Bluetooth host as well.
Bump Pairing the Device to a Bluetooth Host
This device has an integrated RFID chip to facilitate a faster Bluetooth
pairing process with a Bluetooth host that supports a RFID or NFC reader with
appropriate software. Simply align the RFID or NFC reader of the Bluetooth
host directly over the approval label on the back of the device.The device and
Bluetooth host should become paired and connected.
Connecting the Device to a Bluetooth Host
If this was the last device connected to the Bluetooth host, the Bluetooth
host can connect to the device whether or not the device is in discovery mode
as long as Bluetooth is enabled. The six digit code confirmation will not be
displayed.
The device will only recall the last Bluetooth host it was paired with. If
connecting to another Bluetooth host, the device must be placed into discovery
mode to be detected.
Disconnecting the Device from a Bluetooth Host
The device does not have a disconnect feature as this would be initiated by
the Bluetooth host. Use the Bluetooth host functions to purposefully
disconnect the device from the Bluetooth host.
Device Configuration over Bluetooth Connection
The device has the ability to receive updates to device settings over the
Bluetooth connection. The user must successfully pair the device and Bluetooth
host confirming that the six digit security code matches both on the device
and the Bluetooth host. After a configuration change has been initiated, the
user must confirm the request on the device by pressing the q button.
Evacuation Alert over Bluetooth Connection
The device has the ability to receive an evacuate message over the Bluetooth
connection. The user must successfully pair the device and Bluetooth host
confirming that the six digit security code matches both on the device and the
Bluetooth host. Once connected, an evacuate message sent to the device will
send the device into alarm while displaying EVAC on the
US
ALTAIR 4XR
29
4 Operation
display. Press the p button to silence the Evacuation alert and confirm the
alert was received. Press the p button a second time to reset the Evacuation
alert once in a safe area.
4.6 Data Logging This device’s data logs can be downloaded via the GALAXY GX2
Automated Test Stand or via a PC using MSA Link software. Connecting device to
PC
1. Switch ON the device and align the Datalink Communication port on the
device to the IR interface of the PC. 2. Start the MSA Link software on the PC
and start the connection by clicking the connect icon.
4.7 Function Tests Alarm Test Turn on the device. Verify that:
· all LCD segments are activated momentarily · alarm LEDs flash · horn sounds
briefly · vibrating alarm triggers briefly.
4.8 Bump Test
WARNING! Perform a Bump Test before each day’s use to verify proper device
operation. Failure to perform this test can result in serious personal injury
or death.
Bump test frequency is often stipulated by national or corporate regulations; however, bump testing before each day’s use is generally the accepted best safety practice and is therefore MSA’s recommendation.
This test quickly confirms that the gas sensors are functioning. Perform a
full calibration periodically to ensure accuracy and immediately if the device
fails the Bump Test. The Bump Test can be performed using the procedure below
or automatically using the GALAXY GX2 Test Stand.
CSA requires (per 22.2 NO. 152) that combustible sensor sensitivity be tested
before each day’s use on a known concentration of methane equivalent to 25 to
50 % of full scale concentration. ACCURACY MUST BE WITHIN 0 to +20 % OF
ACTUAL. Correct accuracy by performing the calibration procedure described in
4.10 Calibration.
Equipment
See accessory section for ordering information for these components.
· Calibration Check Gas Cylinder · 0.25 liters/min. Flow Regulator · 1/8″ ID
Superthane Ester Tubing · Calibration Cap
Performing a Bump Test
1. Connect the regulator to the Calibration Check Gas Cylinder.
2. From the Normal Operation screen press the q button to display “BUMP
TEST?”.
30
ALTAIR 4XR
US
4 Operation
3. Verify the gas concentrations displayed match the Calibration Check Gas
Cylinder. If they do not, adjust the values through the Calibration Setup menu
as described in 4.4 Device Setup.
4. Attach the calibration cap (see 4.10.2 Attach the Calibration Cap
5. Press the button to start the Bump Test. If calibration lockout option is
selected, enter password. The hourglass will flash and the sensors will
respond to the gas.
6. Open the pressure reducer valve on test gas cylinder.
7. Close the valve after bump testing.
After the Bump Test completes, the device momentarily displays “BUMP PASS” or
“BUMP ERROR” along with the label of any sensor that failed before returning
to Normal Operation mode. If the device fails the Bump Test, perform a
calibration as described in 4.10.2 Attach the Calibration Cap.
The symbol will be displayed in the Normal Operation mode and the Bump LED
will flash green for 24 hours after a successful Bump Test.
4.9 Bump LED
The device is equipped with a green Bump LED. The green LED flashes every 15
seconds under the following conditions:
· the Bump LED feature is enabled · after a successful Bump Test (for 24
hours) · device is in Normal Operation mode · device is not in Low Battery
warning or alarm
4.10 Calibration
The device can be calibrated either manually using this procedure or
automatically using the GALAXY GX2 Test Stand. Refer to 8.7 Calibrations.
Calibration must be performed using a flow regulator with a flow rate set to
0.25 liters per minute.
If a battery charging cycle is interrupted before it is completed (4 hours for
a fully discharged battery), allow the device’s internal temperature to
stabilize for 30 minutes before performing a Calibration.
NOTE: The GALAXY GX2 Test Stand is not a CSA certified calibration method.
Under normal circumstances MSA recommends calibration at least every six
months, however, many European countries have their own guidelines. Please
check your national legislation.
US
ALTAIR 4XR
31
4 Operation
4.10.1 Fresh Air Setup and Zero Calibration
To skip the Zero procedure and move directly to the Calibration Span
procedure, push the p button. If no button is pushed for 30 seconds, the
device prompts user to perform a Span calibration before returning to the
Normal Operation mode.
1. Press and hold the p button in Normal Operation mode for three seconds. 2.
If calibration lockout option is selected, enter password.
ZERO CAL? screen is then displayed.
· If calibration lockout option is NOT selected: ZERO screen displays.
3. With the device exposed to fresh air, press the button to confirm the ZERO
CAL? screen. A sensor refresh and Zero Calibration will occur.
Alternatively press the q button to execute a Fresh Air setup (FAS). See 4.2
Turning ON and Fresh Air Setup for more details.
After Zero calibration completes, the device momentarily displays “ZERO PASS” or “ZERO ERR” along with the flag of any sensor that failed.
During zero calibration the O2 sensor is also span calibrated to 20.8% O2 fresh air, adjusting the calibration curve as needed. During span calibration, the O2 sensor’s accuracy is checked against a known oxygen gas concentration without adjusting the calibration curve.
4.10.2 Attach the Calibration Cap
Attach the calibration cap to the device:
1. Insert tab on calibration cap into slot on device.
2. Press calibration cap as shown until it seats onto device.
32
ALTAIR 4XR
US
4 Operation
3. Press both side tabs down onto device until they snap into place.
4. Ensure that the calibration cap is properly seated.
5. Connect one end of the tubing to the calibration cap.
6. Connect other end of tubing to the cylinder regulator (supplied in the calibration kit).
4.10.3 Span Calibration
To skip the Span procedure, push the p button. If no button is pushed for 30 seconds, the device returns to Normal Operation mode.
1. Once the Zero is set, the SPAN CAL? screen displays.
2. Connect the regulator to the calibration check gas cylinder.
3. Connect the appropriate calibration gas to the device.
4. Attach the calibration cap (see 4.10.2 Attach the Calibration Cap)
5. Open the pressure reducer valve on test gas cylinder.
6. Press the button to calibrate (span) the device.
SPAN calibration starts.
7. Close the valve after SPAN calibration.
· After the SPAN calibration completes, the device momentarily displays “SPAN PASS” or “SPAN ERR” along with the label of any sensor that failed then returns to Normal Operation mode.
If a sensor is nearing its end of life, this “SPAN PASS” indication will be
followed by the end of sensor life warning ( symbol). The symbol, and the gas
type of the sensor nearing its end of life, will blink for 15 seconds when the
device returns to Normal Operation mode. When in Normal Operation mode, the
symbol is continuously displayed.
Finishing Calibration
US
ALTAIR 4XR
33
5 Maintenance
1. Close the valve on the regulator. 2. Remove the calibration cap.
The calibration procedure adjusts the span value for any sensor that passes
the ca-libration test; sensors that fail calibration are left unchanged. Since
residual gas may be present, the device may briefly go into an exposure alarm
after the ca-libration sequence is completed.
Autocalibration Failure
If the span calibration is unsuccessful:
· If the device cannot calibrate one or more sensor(s), it goes to the SPAN
ERR page and remains in alarm until the p button is pressed.
· A sensor life indicator is displayed (Alarm symbol and symbol) to show the
sensor has reached its end of life and should be replaced.
This occurs if the span calibration is unsuccessful twice.
· The device will remain in alarm state until the p button is pressed. · The
Alarm symbol and symbol will remain on the display until a successful
calibration or sensor in question is
replaced.
A span calibration can fail for many reasons besides sensor at the end of his
life. If a span calibration failure occurs, items such as remaining gas in the
calibration cylinder, gas expiration date, security of the calibration cap,
etc. should be verified and calibration should be repeated prior to replacing
the sensor.
4.11 Time of Day Testing
This feature permits the device to be automatically calibrated on a user-
defined interval. The most common use of this feature allows the user to
configure the ALTAIR 4XRand the GALAXY GX2 System to automatically calibrate a
device prior to the start of work-shift. See the GALAXY GX2 Operating Manual
(“Automated Testing Features” section) for a complete description of how to
configure the GALAXY GX2 for this mode.
On ALTAIR 4XR devices, the following settings must be configured using either
MSA Link software or the GALAXY GX2 Instrument Setup page:
· For automated calibration testing, Calibration Due must be enabled and a
non-zero Calibration Interval must be entered for automatic calibration
testing.
· For automated bump testing, Bump Due must be enabled and a non-zero Bump
Interval must be entered for automatic bump testing.
Carefully follow all GALAXY GX2 set-up directions as described in the GALAXY
GX2 Operating Manual for proper setup.
5 Maintenance
If an error occurs during operation, use the displayed error codes to
determine appropriate next steps. The device should be regularly checked and
maintained by qualified personnel.
WARNING! · Repair or alteration of the device beyond the procedures described
in this manual by anyone other than a person
authorized by MSA, could cause the unit to fail to perform properly. Use only
genuine MSA replacement parts when performing any maintenance procedures
described in this manual. · Substitution of components can seriously impair
performance of the unit, alter intrinsic safety characteristics or void agency
approvals. Failure to follow these warnings can result in serious personal
injury or death.
34
ALTAIR 4XR
US
5 Maintenance
Refer to EN 60079-29-2 (Guide for the selection, installation, use and maintenance of apparatus for the detection and measurement of combustible gases or oxygen) and EN 45544-4 (Guide for the selection, installation, use and maintenance of electrical apparatus used for the direct detection and direct concentration measurement of toxic gases and vapors).
5.1 Troubleshooting Problem ERROR TEMP
ERROR EE ERROR PRG ERROR RAM ERROR BT ERROR UNK
LOW BATT
BATT ALARM
ERROR CHARGE
SENSOR ERROR Device does not turn on
- or (flashing)*
Description
Reaction
Temperature is below -40°C (-40 °F) Return device to normal
or above 75°C (167 °F).
temperature range and recalibrate.
EEPROM Memory error Flash Memory error RAM Memory error Bluetooth error
Unknown error Battery warning repeats every 60 seconds. Battery is completely
discharged.
Charge error
Contact MSA Contact MSA Contact MSA Contact MSA Contact MSA Contact MSA Remove from service as soon as possible and recharge battery Device is no longer sensing gas. Remove from service and recharge battery. Device must be between 10° C and 36° C to charge.
Missing Sensor Low battery Sensor Warning
Contact MSA if problem persists Verify if sensor is properly -installed Charge device Sensor is near the end of its life
Sensor Alarm
Sensor has reached the end of its life and cannot be calibrated. Replace sensor and recalibrate.
*Sensor Warning and Sensor Alarm indicate that the device detects a loss of sensor sensitivity during calibration. In addition to true loss of sensor sensitivity, the device may issue a Sensor Warning or Sensor Alarm if calibration gas is expired, misapplied or not used during the span calibration. Ensure the quality of the calibration gas and the calibration gas delivery system. Recalibration may remove the Sensor Warning or Sensor Alarm indication.
5.2 Live Maintenance Procedure – Replacing and Adding a Sensor
WARNING! Remove and reinstall sensors carefully, ensuring that the components
are not damaged; otherwise device intrinsic safety may be adversely affected,
wrong readings could occur, and persons relying on this product for their
safety could sustain serious personal injury or death.
WARNING!
Before handling the PC board, ensure you are properly grounded; otherwise,
static charges from your body could damage the electronics. Such damage is not
covered by the warranty. Grounding straps and kits are available from
electronics
US
ALTAIR 4XR
35
5 Maintenance
suppliers. Failure to follow this warning can result in serious personal injury or death.
To add a sensor to an device that is not already equipped with a full array of sensors, remove the sensor plug from in front of the formerly unused sensor housing.
While device case is open, do not touch any internal components with metallic/conductive objects or tools. Damage to the device can occur.
1. Verify that the device is turned off. 2. Remove the four case screws, and
remove the case front while carefully noting the orientation of the sensor
gasket. 3. Gently lift out and properly discard the sensor to be replaced.
a. Using fingers only, gently remove the toxic, combustible, or oxygen sensor
by gently rocking it while pulling it straight from its socket.
4. Carefully align the new sensor contact pins with the sockets on the
printed circuit board and pressing it firmly in place. a. Ensure tab on sensor
aligns with groove at top of holder. b. Insert the toxic sensor by placing it
in the left-hand position of the sensor holder. c. Insert the O2 sensor by
placing it in the right-hand position of the sensor holder. d. Insert the
combustible sensor by placing it in the middle position of the sensor holder.
e. If any sensor is not to be installed, ensure that a sensor plug is
installed properly in its place.
5. Reinstall the case front. 6. Reinstall the screws.
a. Tighten each screw with a torque value of 6.00 in-lbs (+/- 0.25 in-lbs) or
0.678 Nm (+/- 0.028 Nm) to maintain the device’s IP 68 rating.
7. Turn on the device.
If the sensor replaced is the same as the previous sensor: If the sensor replaced is not the same as the previous sensor or this sensor channel was deactivated:
· The device starts up normally.
· The device automatically senses that a new sensor is installed and displays
the “SENSOR DSCVRY” screen.
· The device automatically senses the difference and displays “SENSOR CHANGE”.
· “ACCEPT?” appears on the display.
Accept the change with q button or reject with p button.
Go into the sensor setup and turn on the appropriate sensor ( 4.4.1 Sensor Setup).
8. Calibrate the device after the sensors have stabilized.
WARNING!
Calibration is required after a sensor is installed; otherwise, the device
will not perform as expected and persons relying on this product for their
safety could sustain serious personal injury or death.
Allow sensors to stabilize at room temperature for at least 30 minutes before calibration ( 4.10 Calibration).
36
ALTAIR 4XR
US
5 Maintenance
5.3 Cleaning
· Routine Cleaning: Clean the exterior of the device regularly using only a
damp cloth. Do not use clea-ning agents, as many contain silicone, which will
damage the combustible sensor.
· Dust and Dirt Exposure: Use a dry, soft bristled brush to remove any dust or
dirt that has accumulated on the apparatus, especially at the sensor openings.
If there is a buildup of dust or dirt particles remaining in the sensor area
after brushing, use a vacuum to remove remaining particles, but maintain at
least a 1/2 inch (1.3 cm) distance from the gas detector.
· Water Exposure: If the apparatus is exposed to water, turn the device sensor
side down and gently shake water off the sensor area. Any remaining water can
be removed with a clean dry cloth.
5.4 Storage
When not in use, store the device in a safe, dry place between 18 °C (64 °F)
and 30 °C (86 °F). After storage, always recheck device calibration before
use.
5.5 Scope of Delivery
Pack the device in its original shipping container with suitable padding. If
the original container is unavailable, an equivalent container may be
substituted.
US
ALTAIR 4XR
37
6 Technical Data
6 Technical Data
Weight Dimensions (L x W x H) Alarms Volume of audible alarm Display Battery
type
Device run time (Bluetooth OFF) Charging time
Warm up time Temperature range
Humidity range
Atmospheric pressure range Ingress protection Measuring methods
7.9 oz./224 g (device with battery and clip) 4.4 x 3.00 x 1.37 inches/112 x 76
x 33 mm without fastening clip
Four gas alarm LEDs, a charge status LED, an audible alarm and a vibrating
alarm 95 dBA at 30 cm typical
LCD display Rechargeable Li polymer battery. Must not be charged in Ex area.
24 hours at 25 °C (77 °F)
4 hours The maximum safe area charging voltage Um = 6.7 VDC 2 min 14 to 104 °F
(-10 to 40 °C) Normal operating range -40 to 140 °F (-40 to 60 °C) Extended
operating range 50 to 95 °F (10 to 35 °C) While charging battery -40 to +140
°F (-40 to 60 °C) Intrinsic safety ambient temperature range (ATEX, IEC) -40
to +129 °F (-40 to 54 °C) Intrinsic safety ambient temperature range (CSA,
ETL) 15 % 90 % relative humidity, non-condensing, 5 % 95 % RH intermittent
800 to 1200 mbar
IP 68 (2 meters for 45 minutes)
Combustible gases: Catalytic sensor Oxygen: Electrochemical sensor Toxic
gases: Electrochemical sensor
Measuring range
Combustible 0-100% LEL
0-5.00% Vol.CH4 H2S-LC 0-100 ppm
O2* 0-30% Vol.
NO2 0-50 ppm
CO 0-1999 ppm 0-1999 mg/m3 CO-H2 0-1999 ppm
H2S 0-200 ppm
0-284 mg/m³ SO2 0-20 ppm
- Certified for 0 – 25 vol.% O2
The technical and performance specifications for the specialized EX-H and EX-M sensor are the same as for the standard EX sensor.
38
ALTAIR 4XR
US
6 Technical Data
The ppm to mg/m³ conversion is calculated at 20° C (68 °F) and at atmospheric pressure.
6.1 Factory-set Alarm Thresholds and Setpoints
Check the monitor or calibration certificate for exact alarm levels as they vary depending on national or corporate regulations.
Sensor EX (% LEL) EX-H (% LEL) EX-M (% vol) O2 (% vol) H2S (ppm) H2S-LC(ppm)
CO (ppm) NO2 (ppm) SO2 (ppm)
Sensor EX (% LEL) EX-H (% LEL) EX-M (% vol) O2 (% vol) H2S (ppm) H2S-LC (ppm)
CO (ppm) NO2 (ppm) SO2 (ppm)
Sensor
EX (% LEL) EX-H (% LEL) EX-M (% vol) O2 (% vol) H2S (ppm) H2S-LC (ppm) CO
(ppm) NO2 (ppm) SO2 (ppm)
LOW alarm 10 10 0.5 19.5 10 5 25 2 2
HIGH alarm 20 20 1.0 23.0 15 10 100 5 5
STEL —-15 10 100 5 5
TWA —-10 1 25 2 2
Min. alarm set point 5 5 0.1 5 5 1 10 .5 .5
Max. alarm set point 60 60 3.0 24 175 70 1700 47.5 17.5
Default Calibration Setpoint 58 58 2.5 15.0 20 20 60 10 10
Min. Calibration Setpoint 5 5 0.1 5.0 5 1 10 1 1
Max. Calibration Setpoint 100 100 5.0 30.0 200 70 1700 50 20
US
ALTAIR 4XR
39
6 Technical Data
Feature Instrument password Cal Due days Combustible gas type Gas measurement
units
· Combustible sensor · Toxic sensor Periodic log collection rate Backlight
timeout
Options 000 – 999 1 – 180 Methane, butane, propane, pentane, nonane, hydrogen,
combustible
· % LEL or % CH4 · ppm or mg/m3
Off, 15 sec, 30 sec, 1 min, 3 min, 5 min, 10 min, 15 min Off, 10 sec, 30 sec,
1 min, 3 min, 5 min, 10 min, On
6.2 Performance Specification Combustible Gas
Range Resolution Reproducibility
Response time
0 to 100 % LEL or 0 to 5 % CH4
1 % LEL or 0.05 % vol CH4
3 % LEL, 0 % to 50 % LEL reading or 0.15 % CH4, 0.00 % to 2.50 % CH4 (normal
temperature range)
5 % LEL, 50 % to 100 % LEL reading or 0.25 % CH4, 2.50 % to 5.00 % CH4 (normal
temperature range)
5 % LEL, 0 % to 50 % LEL reading or 0.25 % CH4, 0.00 % to 2.50 % CH4 (extended
temperature range)
8 % LEL, 50 % to 100 % LEL reading or 0.4 % CH4, 2.50 % to 5.00 % CH4
(extended temperature range)
90 % of final reading in less than or equal to 15 sec (Pentane) and 10 sec
(methane) (normal temperature range)
Combustible Gas Cross Reference Factors for General Purpose Calibration Using Calibration Cylinder (P/N 10053022)
Combustible Gas
Acetone Acetylene Butane Cyclohexane Diethylether Ethane Ethanol Ethylene
Gasoline n-Hexane Hydrogen Isobutane Isopropyl Alcohol
Methane Calibration 1.45 Vol % CH4 Set 33 % LEL 1.09 1.07 1.37 1.94 1.43 1.27 1.16 1.09 1.63 1.86 0.98 1.63 1.55
Pentane Simulant Calibration 1.45 Vol % CH4Set 58 % LEL
0.62 0.61 0.79 1.11 0.82 0.73 0.66 0.62 0.93 1.06 0.56 0.93 0.88
40
ALTAIR 4XR
US
6 Technical Data
Combustible Gas
Methane Methanol Nonane Nonane with EX-H sensor Pentane Propane Propylene
Toluene Xylene Xylene with EX-H sensor
Methane Calibration 1.45 Vol % CH4 Set 33 % LEL 1.00 0.93 4.48 3.03 1.90 1.39 1.14 2.09 4.83 3.57
Pentane Simulant Calibration 1.45 Vol % CH4Set 58 % LEL
0.57 0.53 2.56 1.73 1.00 0.79 0.65 1.19 2.76 2.04
NOTE: Gas in the table above indicated with a will have significantly longer
response times, especially at lower concentration levels.
Response notes
1. Some compounds may reduce the sensitivity of the combustible gas sensor
by poisoning or inhibiting the catalytic action or by polymerizing on the
catalytic surface.
2. Multiply the displayed % LEL value by the conversion factor above to get
the true % LEL. 3. These conversion factors should be used only if the
combustible gas is known. 4. All factors are based on the IEC 100 % LEL levels
a. i.e. Methane 100% LEL = 4.4 Vol %, b. Pentane 100% LEL = 1.1 Vol % c.
Propane 100% LEL = 1.7 Vol % 5. These conversion factors are typical.
Individual units may vary by ±25 % from these values. 6. The results are
intended for guidance only. For the most accurate measurements, a device
should be calibrated using the gas under investigation. 7. The conversion
factors for the standard EX and the specialized EX-H and EX-M sensor are the
same except for EXH and Nonane and o-Xylene. The conversion factors for these
two vapors are therefore especially mentioned in the table.
Oxygen
The oxygen sensor has built-in temperature compensation. However, if
temperature shifts dramatically, the oxygen sensor reading may shift. Zero the
device at a temperature within 30 °C (86 °F) of the work place temperature for
the least effect.
Range Resolution Reproducibility Response time (90% of final reading) Sensor Cross-Sensitivity
0 to 30 vol.% O2* 0.1 vol.% O2 0.7 vol.% O2 for 0 to 30 vol.% O2 90 % of final reading, 16 seconds typical (normal temperature range) The oxygen sensor has no common cross-sensitivities.
- Certified for 0 – 25 vol.% O2 Carbon Monoxide
Data is presented as the indicated output in PPM which will result from the application of the test gas.
US
ALTAIR 4XR
41
6 Technical Data
Range Resolution Reproducibility
Response time
Test Gas Applied Hydrogen Sulfide (H2S) Carbon Monoxide (CO) Nitric Oxide (NO)
Nitrogen Dioxide (NO2) Sulfur Dioxide (SO2) Chlorine (Cl2) Hydrogen Cyanide
(HCN) Ammonia (NH3) Toluene Isopropanol Hydrogen (H2) Hydrogen Sulphide Range
Resolution
Reproducibility
Response time
Test Gas Applied Hydrogen Sulfide (H2S) Carbon Monoxide (CO) Nitric Oxide (NO)
Nitrogen Dioxide (NO2) Sulfur Dioxide (SO2) Chlorine (Cl2) Hydrogen Cyanide
(HCN) Ammonia (NH3)
42
0 – 1999 ppm (0 – 1999 mg/m3) CO 1 ppm (1.2 mg/m3) CO for 0 to 1999 ppm ±5 ppm
(5.8 mg/m3) CO or 10 % of reading, whichever is greater (normal temperature
range) ±10 ppm (11.6 mg/m3) CO or 20 % of reading, whichever is greater
90% of final reading in less than or equal to 15 seconds (normal temperature
range)
Concentration (PPM) Applied 40 100 50 11 9 10 30 25 53 100 100
CO Channel % Cross-sensitivity 0 100 84 0 -4 0 -5 0 0 -8 48
0 – 200 ppm (0 to 284 mg/m3) H2S 1 ppm (1.4 mg/m3) H2S,
for 3 to 200 ppm (4,3 to 284 mg/m3) H2S ±2 ppm (2.8 mg/m3) H2S or 10 % of
reading, whichever is greater (normal temperature range) 0 to 100 ppm (0 to
142 mg/m3) H2S, ±5 ppm (7,1 mg/m3) H2S or 10 % of reading, whichever is
greater
90 % of final reading in less than or equal to 15 seconds (normal temperature
range)
Concentration (PPM) Applied 40 100 50 11 9 10 30 25
H2S Channel % Cross-sensitivity 100 1 25 -1 14 -14 -3 -1
ALTAIR 4XR
US
6 Technical Data
Test Gas Applied Toluene Isopropanol Hydrogen (H2)
Concentration (PPM) Applied 53 100 100
Hydrogen Sulphide Low Concentration (H2S-LC)
H2S Channel % Cross-sensitivity 0 -3 0
Range Resolution Reproducibility
Response time (typical)
0 – 100 ppm H2S
0.1 ppm H2S
±0.2 ppm H2S or 10 % of reading, whichever is greater (normal temperature
range)
±0.5 ppm H2S or 20 % of reading, whichever is greater (extended temperature
range)
90% of final reading < 15 seconds (normal temperature range)
Nitrogen Dioxide
Range Resolution Reproducibility
Response time (typical)
0 – 50 ppm NO2
0.1 ppm NO2
±1 ppm NO2 or 10 % of reading, whichever is greater (normal temperature range)
±2 ppm NO2 or 20 % of reading, whichever is greater (extended temperature
range)
90 % of final reading < 20 seconds (normal temperature range)
Sulfur Dioxide Range Resolution Reproducibility
Response time (typical)
0 – 20 ppm SO2
0.1 ppm SO2
±1 ppm SO2 or 10 % of reading, whichever is greater (normal temperature range)
±2 ppm SO2 or 20 % of reading, whichever is greater (extended temperature
range)
90 % of final reading < 20 seconds (normal temperature range)
Hydrogen Resistant CO (CO-H2-RES) Range Resolution Reproducibility
Response time (typical) Hydrogen resistance
0 – 2000 ppm CO 1 ppm CO ±5 ppm CO or 10 % of reading, whichever is greater (normal temperature range) ±10 ppm CO or 20 % of reading, whichever is greater (extended temperature range) 90% of final reading < 20 seconds (normal temperature range) <5%
6.3 XCell Sensor Patents
· US 8826721
· US 7959777
US
ALTAIR 4XR
43
6 Technical Data
· US 8702935 · US 8790501 · Additional patents pending
44
ALTAIR 4XR
US
7 Ordering Information
Description
Sensor Replacement Kits EX EX-H EX-M O2 CO/H2S CO/H2S-LC CO-H2/H2S CO/NO2 SO2
/H2S-LC Repair Kits Front Housing with dust filters
· Charcoal with Bluetooth without Bluetooth
· Phosphorescent with Bluetooth without Bluetooth
LCD assembly Stainless steel suspender clip, screws, & sensor gasket Stainless
steel suspender clip kit Replacement Belt Clip Screw, Package of 50 Power
Supplies Power Supply with Charge Connector
· North America · Global Charging cradle with power supply · North America ·
Europe · Australia Vehicle charging cradle Calibration Pressure reducer valve
0.25 l/min Calibration assembly (cap, tube, connector) Calibration Cylinder
58L 4-gas mix (1.45% CH4, 15% O2, 60 ppm CO, 20 ppm H2S) · North America ·
Europe
Part Number
10106722 10121211 10121212 10106729 10106725 10121213 10121214 10121217
10121215
10178360 10236170
10178361 10236169 10179265 10110062 10208107 10193731
10092233 10092936
10087368 10086638 10089487 10095774
467895 10089321
10045035 10053022
US
ALTAIR 4XR
7 Ordering Information 45
7 Ordering Information
Description
Accessories ALTAIR Pump Probe
· North America · ATEX/IEC · China JetEye IR adapter with USB connector
Part Number
10152669 10152668 10150031 10082834
GALAXY GX2 and additional accessories are available on request.
46
ALTAIR 4XR
US
8 Appendix
8.1 Start Up Sequence (Power on)
8 Appendix
US
ALTAIR 4XR
47
8 Appendix 8.2 Fresh Air Setup (FAS)
48
ALTAIR 4XR
US
8.3 Reset Screen Controls
8 Appendix
US
ALTAIR 4XR
49
8 Appendix
50
ALTAIR 4XR
US
8.4 Bump Test
8 Appendix
US
ALTAIR 4XR
51
8 Appendix 8.5 Options Setup
52
ALTAIR 4XR
US
8.6 Sensor Setup
8 Appendix
US
ALTAIR 4XR
53
8 Appendix 8.7 Calibrations
54
ALTAIR 4XR
US
8.8 Alarm Setup
8 Appendix
US
ALTAIR 4XR
55
8 Appendix 8.9 Time and Date Setup
8.10 Bluetooth Setup
56
ALTAIR 4XR
US
References
- MSA Safety | Global
- MSA Safety | Global
- MSA Safety | Global
- Trademark Information | MSA Safety | United States
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>