HTC CAL-900 Get Sound Level Meter User Manual
- June 4, 2024
- HTC
Table of Contents
- CAL-900 Get Sound Level Meter
- APPLICATION & FEATURES
- PARAMETER
- **FRONT PANEL DESCRIPTIONS
- **OPERATION
- OPERATION PROCEDURE
- BATTERY REPLACEMENT
- CARE OF THE INSTRUMENT
- EFFECTS OF ATMOSPHERIC PRESSURE AND TEMPERATURE
- ACCESSORIES
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
CAL-900 Get Sound Level Meter
User Manual
CAL-900 Get Sound Level Meter
SOUND LEVEL CALIBRATOR
CAL-900
This Sound Level Calibrator is small in size, light in weight, easy to
carry. Although complex and advanced, it is convenient to use and operate. Its
ruggedness will allow many years of use if proper operating techniques are
followed. Please read the following instructions carefully and always keep
this manual within easy reach.
APPLICATION & FEATURES
- Handy sound source for quick and easy calibration of sound level meters and sound measuring systems. The Calibrator employs solid state integrated circuitry that provides accurate and stable performance. Sensitivity calibration of microphones. Suitable for field and laboratory use. Applies display and key operation, simple to use.
PARAMETER
Sound pressure level 94 dB and 114 dB Accuracy: ±0.3 dB (20°C, 760 mm Hg)
Frequency of 1000±0.01% Hz allows calibration with A, B, C or D weighting
networks or linear. Extremely low influence of static pressure. Conforms to
IEC 942 class 1 Calibration of 1 ” and 1/2 ” microphones Temperature Range:
-10 to +50°C operating Storage (with batteries removed) -40 to +65°C
Temperature Coefficient: 0 to 0.01 dB/°C Altitude Effects: Approximately 0.1
dB decrease for every 2000 feet increase in altitude from sea level to 12,000
feet elevation, or comparable atmospheric pressure change (approximately every
50 mm of Hg decrease).
Power Source: 2×1.5 AAA(UM-4) battery
Size: 137x51x51mm
Weight: about 200g 3
**FRONT PANEL DESCRIPTIONS
**
****3-1 Transducer Cap
3-2 Microphone Adaptors
3-3 114dB Key
3-4 94dB Key
3-5 Power Key
3-6 Display
**OPERATION
**
The ND9 Calibrator is designed to check the accuracy of many types of sound
instruments, not only Landtek manufactured equipment. Our instruments commonly
use the standard 1/2-inch diameter ceramic microphone which fits directly into
the calibrator coupler cavity. 2
When testing an instrument with a 1-inch microphone, the Transducer Cap must
first be taken out. This keeps a close tolerance fit around the microphone
head. Be sure the microphone fits down inside the adaptor and rests on the
lower rim. This rim supports the microphone and forms the necessary inner
seal.
OPERATION PROCEDURE
5.1 Press the Power Key to power on the calibrator. 1000 Hz tone should be
heard. Press the 114dB Key or the 94dB Key to select 94 dB or 114 dB position.
5.2 Turn on the sound level meter which is to be calibrated.
5.3 Carefully insert the microphone into the calibrator coupler. Be sure the
microphone is down inside the coupler resting flush on the lower coupler rim.
5.4 When calibration has been made, carefully remove the microphone and turn
the calibrator to OFF.
BATTERY REPLACEMENT
6.1 When the battery voltage is too low, the battery symbol ` ‘ will appear on
the lower left quarter of the display. It is necessary to replace the
batteries.
6.2 Slide the Battery Cover away from the instrument and remove the batteries.
6.3 Install the batteries correctly into the case.
CARE OF THE INSTRUMENT
7.1 Immediately clean any spilled materials from the Instrument and wipe dry.
If spillage is corrosive, use a suitable cleaner to remove it and to
neutralize corrosive action.
7.2 Remember to turn off the Instrument when not using it.
7.3 Avoid prolonged exposure or usage in areas subject to temperature and
humidity extremes, vibration, mechanical shock, dust, corrosive fumes, and
strong electrostatic and electromagnetic interference.
7.4 Be sure the transducer cap is firmly in place.
7.5 If the Instrument has not been used for 30 days, check the battery for
leakage and replace if necessary.
7.6 When the Instrument is not in use, store it in a room free from
temperature extremes, dust, corrosive fumes, mechanical vibration, or shock.
If the storage time is expected to exceed 30 days, remove the battery.
EFFECTS OF ATMOSPHERIC PRESSURE AND TEMPERATURE
For any one location, the effects of normal variations of atmospheric pressure are usually negligible. But most calibrators including the Nd9 are affected by altitude. The transducer diaphragm within the calibrator creates the sound as it vibrates against the air. When the air is thinner (at higher elevations) a lower sound level is produced. The ND9 is calibrated to produce 94dB at sea level. When the unit is operated above sea level a slightly lower sound level is emitted depending on altitude. For each 2000 feet of elevation above sea level the ND9 produces 0.1dB less than the 94 dB rating. As an example, the calibrator will only emit 93.7dB at an elevation of 6,000 feet. Therefore, a sound level meter should be set at 93.7dB, not at the rated 94dB. The effects of temperature are less than ±0.05dB/°C (reference is 23°C).
ACCESSORIES
Carrying case ………….1PC
Operation manual………1PC