HTC PM-6175 Manometer User Manual

June 3, 2024
HTC

HTC PM-6175 Manometer

 Safety

Read the following safety information carefully before attempting to operate or service the meter.
Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired.

Environment conditions:

  1. RH≤90%(Non-Condensation)
  2. Operating Temperature-0~40℃

Maintenance and cleaning

  1. Repairs or servicing not covered in this manual should only be performed by qualified personnel.
  2. Wipe the unit with a dry soft cloth. Do not use solvents or detergents on this instrument.

Safety Symbol
Comply with EMC

Introduction

This product is aimed to measure different pressure values, here are some units for Choice: inH2O, psi ,bar, mbar , kPa, inHg, mmHg ,ozin2 ,,ftH2O, cmH2O ,kgcm2.equipped with simple Tube connection. It is widely used in measurement of ventilation and air conditioning system, clean room or any gas pressure system troubleshooting .Easy operation and clear reading.

Error Warning
When internal diagnostic test fails, an error message will appear on the LCD display.

  • error1: max pressure
  • error2: below min pressure

Specification

unit

| PM- 6102| PM- 6105| PM- 6115
---|---|---|---
Range| Resolution| Range| Resolution| Range| Resolution
inH2O| ±55.40| 0.01| ±138| 0.1| ±415.2| 0.1
Psi| ±2.000| 0.001| ±5.000| 0.001| ±15.00| 0.01
Mbar| ±137.8| 0.1| ±344.7| 0.1| ±1034| 1
KPa| ±13.79| 0.01| ±34.47| 0.01| ±103.4| 0.1
inHg| ±0.407| 0.001| ±1.018| 0.001| ±30.54| 0.01
mmHg| ±103.4| 0.1| ±258.5| 0.1| ±775.7| 0.1
ozin²| ±32.00| 0..01| ±80| 1| ±240.0| 0.1
ftH2O| ±4.612| 0.001| ±11.53| 0.01| ±34.60| 0.01
cmH2O| ±140.6| 0.1| ±351.5| 0.1| ±1055| 1
kgcm²| ±0.140| 0.001| ±0.351| 0.001| ±1.055| 0.001
bar| ±0.137| 0.001| ±0.344| 0.001| ±1.034| 0.001
Maximum Pressure Psi| 10 Psi| 20 Psi| 100 Psi
Accuracy| ±0.3%FS
Linearity/Hysteresis| ±0.1%FS
Repeatability| ±0.25%FS
Pressure connector| 1/8〞(3.18mm)quick connector
Working temperature| 0~40℃/32~104℉
Storage temperature| -10~60℃/14~140℉
Over range indication| ERR1 or ERR2
Work current| 9V battery
unit| PM- 6130| PM- 6175
---|---|---
Range| Resolution| Resolution| Range
inH2O| ±830| 1| ±2078| 1
Psi| ±30.00| 0.01| ±75.00| 0.01
Mbar| ±2608| 1| ±5168| 1
KPa| ±206.8| 0.1| ±517.1| 0.1
inHg| ±61.08| 0.01| ±15.00| 0.01
mmHg| ±1551| 1| ±3878| 1
ozin²| ±480.0| 0.1| ±1200| 0.1
ftH2O| ±69.2| 0.01| ±173.0| 0.01
cmH2O| ±2109| 1| ±5272| 1
kgcm²| ±2.109| 0.001| ±5.250| 0.001
bar| ±2.068| 0.001| ±5.138| 0.001
Maximum Pressure Psi| 100Psi| 100Psi
accuracy| ±0.3%FS
Linearity/Hysteresis| ±0.1%FS
Repeatability| ±0.25%FS
Pressure connector| 1/8〞(3.18mm)quick connector
Working temperature| 0~40℃/32~104℉
Storage temperature| -10~60℃/14~140℉
Over range indication| ERR1 or ERR2
Work current| 9V battery

Meter Description

  1. Air intake
  2. LCD Display
  3. Power key
  4. Backlight key
  5. Unit key
  6. Hold button
  7. record key

LCD display

  1. Data record
  2. Record mode
  3. Max Min, Avg
  4. Readings
  5. Unit
  6. Low battery indication
  7. Memory time

HTC-PM-6175-Manometer-02

Operation steps
  1. Open battery cover and install a 9-volt battery in the battery compartment. Press power key to start measuring
  2. press UNITS key ,choosing proper unit
  3. connect Tube
  4. read the testing values on LCD display
    Note :when the meter is on, no gas pressure input, the meter will not necessarily be zero, you need to reset by yourself.

Buttons function

  1. Key ;it can turn on and turn off the meter ,press“ ”, open manometer, the screen will display about 2 seconds, and then enter normal measurement model.
  2. Key ;after power on the meter ,pressing this button can turn on and turn off the backlight.
  3. Key ;in measurement mode ,press“ ”, it can review Max ,Min, Avg
  4. values. In this mode, other button functions are locked not operated. Press  ”for three seconds ,the meter exits record model.
  5. key ;after power on the meter ,press“ convert the units of measurement:bar,mmHg,Ozin2,Kgcm2,psi,inH2O,Kpa,ft H2O,inHg,cmH2O,mbar.
  6. key;
  7. After the meter power on, press“ ”button once ,the screen will  be locked, press“ ”again ,it will cancel this function.
  8. After power on the meter ,no gas pressure input ,press“ ”for three seconds,LCD screen display“—-”,the meter reset.
  9. Automatically power on/off; the meter will automatically turn off after approx. 25 minutes of inactivity. It can save battery’s life. If you do not want to turn off, first press ,then press  ,”ON “will appear on the LCD display, it indicates that power off function is disabled.

Notes

  1. Do not use it in high temperature and humidity。
  2. Remove the battery when the meter is to be stored for long periods of time to avoid battery leakage.
  3. Replace new batteries when the low battery icon ” ” appears.

Accessories

  1. Manual
  2. Canvas ba
  3. 6F22 9V battery
  4. ∮4.5mm mental gas mouth 2 PCS
  5. ∮5.0mm*30mm soft Tube 2PCS
  6. ∮5.0mm*30mm hard Tube 2PCS

www.htcinstrument.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals