SPECIFICATION

Item No

| Range| Resolution| Accuracy| Protection class| Indication| Data output| Calibration
---|---|---|---|---|---|---|---
| mm| inch| mm| inch| μm| | | Wireless|

ISO 17025

110180027

| 0 25| 0-1”| 0,0001| 0.00001”| ±2| IP-65| computerized
MICS system with 1,54” touchscreen| Long range|

110181057

| 0-50| 0-2”| ±3| IP-65| Long range|

TECHNICAL DATA

Parameters

LED display| color 1,54 inch
Resolution| 240×240
Indication system| MICS 3.0
Power supply| Rechargeable Li-Pol battery
Battery capacity| 450mAh
Charging port| micro-USB
Case material| Aluminium
Buttons| Switch (Multifunctional), Reset
Wireless data transfer| Long range

Clean the oil from the measuring surfaces of micrometer by cloth.
Turn on micrometer by pressing SWITCH ON button.
Check the setting display to zero. Contact with each other (or with setting gauge) measuring surfaces of the anvil and the
spindle, using the ratchet and take a measurement.
Rotation the micrometer screw in the direction of convergence measuring surfaces using the ratchet.
For ABSOLUTE zero setting push RESET button (this zero position will be saved also after switch off micrometer)
For zero setting, press the ZERO button on Touchscreen.

MAIN INFO

  1. PRECISION MEASUREMENTS
    MICROTECH computerized micrometer with 0.1 resolution can be used for high precisoт measurements in manufacturing, automotive medical industry and for calibration
    Main Info

  2. COLOR GO/NOGO INDICATION
    Easy and quick limits control.
    It’s possible setup max and min value throw menu.
    Color indication reduce possibility of error during control
    Main Info

  3. INTERNAL MEMORY
    Computerized micrometer has memory manger with possibility saving data to folders. Separate Data’s or Folders can be transfer Wireless to MICROTECH Software (Windows, Android,iOS)
    Main Info

  4. FORMULA MODE
    In Formula mode we use Ax2 +Bx+C
    Main using of this function is:
    – Preset function
    – Magnification of measured value
    For customers request Formula can be updated
    Main Info

  5. ERROR COMPENSATION
    It’s possible to make mathematical error compensation. Just re-calibrate micrometer with gauge block set and save results. Micrometer will indicate measuring results include re-calibration.
    Main Info

  6. TEMPERATURE COMPENSATION
    Temperature compensation function gives possibility recalculate details sizes with high or low temperatures to normal (20C conditions). Available recalculation for 7 materials
    Main Info

ABSOLUTE ZERO SETTING

For ABSOLUTE ZERO setting push RESET button (this zero position will be saved also after switch off micrometer)

For ZERO setting, press the ZERO button on Touchscreen.

Absolute Zero Setting

CHARGING

Charging

For charging connect USB cable Charging process indicating on Battery status (switch on device)

WIRELESS DATA TRANSFER

Activate WIRELESS data transfer in WIRELESS menu

Wireless Data Transfer

Connect instrument with MICS indication system to Tablet or PC Send data to Tablet or PC:

  • by Touchscreen
  • by Multifactional button pushy (activated in Wireless menu)
  • by Timer (activated in Timer menu)
  • from Internal Memory

MEMORY

For save measuring data to internal caliper’s memory touch data area on screen or short button push. You can view saved data throw menu or send Wireless connection to Windows PC, Android or iOS devices.

Basic STATISTIC functions:
MAX – maximum saved value
MIN – minimum saved value
AVG – average value
D – variance VAR(x)

Memory

MEMORY settings

Memory

CONNECTION

Connection

Connection

Icon| Icon| Icon
---|---|---
| |

SAVING DATA TO XLS. CSV FORMATS

SAVING DATA TO XLS. CSV FORMATS

SAVING DATA TO XLS. CSV FORMATS

TRANSFER DATA TO CAD, SPC OR OTHER SOFTWARE

SAVING GRAPH

SAVING GRAPH

TRANSFER DATA TO EXCEL OR OTHER TABLE EDITORS

Symbol VOICE MODE

Symbol VOICE MODE

QUALITY CONTROL RESULTS

DATA TRANSFER DISTANCE FOR INTELLIGENT INSTRUMENTS

  • All MICROTECH Intelligent instruments has internal memory. You can collect data to memory and when it will be neccesary transfer all data to the peripheral device with no data loosing.
  • It’s possible transfer data, like in Wireless instruments up to 15 m or up 50 m with external antenna
  • USB port can be used to transfer data directly to Windows PC

USE TOUCHSCREEN TO OPERATE MENU

Menu Navigation

MEMORY WITH STATISTIC

Menu Navigation

MENU CONFIGURATION

Menu Navigation

MICS SYSTEM FUNCTIONS

LIMITS GO/NOGO

MENU SETTING

Mics System Functions

Colour indication limits

MAIN SCREEN INDICATION

Mics System Functions

MAX/MIN

MENU SETTING

Mics System Functions

Indication and saving MAX or MIN values

MAIN SCREEN INDICATION

Mics System Functions

FORMULA

MENU SETTING

Mics System Functions

FORMULA Ax2 +Bx+C

EXAMPLE OF FORMULA USING

Standard measuring A=0, B=1, C=0
PRESET 50mm A=0, B=1, C=50
Value magnification x3 A=0, B=3, C=0
Square value A=1, B=0, C=0

TIMER

MENU SETTING

Mics System Functions

Saving data to memory or sending Wireless by Timer

MATHEMATICAL ERROR COMPENSATION

MENU SETTING

Mathematical correction points with big error to normal values

Mics System Functions

use gauge blocks or measuring machine for points correction

TEMPERATURE COMPENSATION

MENU SETTING

Re-calculation detail size to calibration conditions (20°C)

Mics System Functions

7 types of material:

  • glass, quartz, solid alloys
  • granite, graphite
  • Fe, NI, Fe alloys
  • stainless steel
  • cooper, Cu alloys
  • alluminium, al alloys
  • acrylic sheet

RESOLUTION

MENU SETTING

Mics System Functions

MAIN SCREEN INDICATION

Selecting resolution of value on main screen and memory

EXTRA (Axis mode)

MENU SETTING

Mics System Functions

Icon
WIRELESS CONNECTION

MENU SETTING

Activating or disactivating wireless data transfer

Mics System Functions

USB CONNECTION

MENU SETTING

Mics System Functions

Connect USB cable to PC
Activate USB connection mode (DON’t activate mode without USB cable connection) Select setting of data transfer You can direct transfer data to any Windows app

PIN & RESET 

MENU SETTING

Activating PIN to device setting

RESET to factory settings

DISPLAY SETTING

MENU SETTING

Mics System Functions

DISPLAY ROTATION, BRIGHTNESS, SLEEPING MODE

Mics System Functions
MEMORY SETTINGS

MENU SETTING

Mics System Functions

Activating or disactivating saving values to memory

Standard – saving values to memory OFF – disable saving to memory

LINK TO SOFTWARE

MENU SETTING

QR Link to MICROTECH web site page with MDS Software download

  • Android, iOS, Windows versions
  • Free and Pro versions
  • Manuals

CALIBRATION DATE

MENU SETTING

Calibration date info

Mics System Functions

DEVICE INFO

MENU SETTING

Information about device

  • device name
  • firmware version
  • MAC adress for easy connection

DIMESIONAL DRAWINGS

Microtech Dimesional Drawings
Dimension

*Dimensions for 25mm range

CUSTOMERS SUPPORT

MICROTECH
innovative measuring instruments
61001, Kharkiv, Ukraine, str. Rustaveli, 39
tel.: +38 (057) 739-03-50
www.microtech.ua
tool@microtech.ua

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals