microtech 132050X22 Bore Gauge With Carbide Contacts User Manual
- June 13, 2024
- microtech
Table of Contents
microtech 132050X22 Bore Gauge With Carbide Contacts
Product Information
The MICROTECH Bore Gauge with Carbide Contacts is a precision measuring tool designed for accurately determining the internal diameter of holes or bores. It is calibrated according to ISO 17025:2017 and ISO 9001:2015 standards, ensuring reliable and accurate measurements. The bore gauge comes in various models with different ranges, depths, rod travels, and carbide contacts.
Specifications:
Item No | Type | Range | Depth | Rod Travel | Carbide Contacts | Accuracy |
---|---|---|---|---|---|---|
132010X00 | DIAL | 6-10 mm | 60 mm | res.10 m *X – 1 m | 1.5 m | mm |
132018X00 | DIAL | 10-18 mm | 110 mm | res.1 m *X – 8 m | 4 m | mm |
132050X00 | DIAL | 18-50 mm | 130 mm | res.0.1 m *X – 7 m | 2000 m | mm |
132160X00 | DIAL | 50-160 mm | 150 mm | 500 m | 4 m | mm |
132250X00 | DIAL | 160-250 mm | 260 mm | 500 m | 4 m | mm |
132450X00 | DIAL | 250-450 mm | 260 mm | 500 m | 4 m | mm |
132900X00 | DIAL | 6-1000 mm | 440 mm | 500 m | 8 m | mm |
132951X00 | DIAL | 130-1500 mm | 440 mm | – | – | mm |
Bore Gauge Sets:
- 132160X00: 18-160 mm
- 132450X00: 18-450 mm
- 132990X00: 6-1000 mm
Indicators:
- Analog: 0.01 mm
- Digital: 0.001 mm
- Computerized: 0.0001 mm
Product Usage Instructions
- Clean oil from the measuring surfaces of the bore gauge and indicator.
- If the bore gauge body consists of two or more parts, attach them together.
- Be careful not to use excessive force when screwing the body toavoid damaging the rods and threads.
- For zero setting, ensure that the measurement line of the bore gauge coincides with a diametrical cross-section of the settingring.
- Slightly shake the bore gauge in the axial plane and determinethe largest reading on the indicator, which represents the size ofthe ring gauge.
- Combine the zero mark of the scale with an arrow by turning the ring and use a clamp to fixate the ring.
- To increase accuracy, ensure that the position of the bore gauge during measurement is the same as during the setup.
- Do not subject the measuring rod to sudden shocks and avoid applying excessive force if it gets jammed.
- Avoid contact between the indicator and oil emulsion.
- Do not turn the indicator when it is fixed in the tool holder by a sleeve.
IMPORTANT: Follow these instructions carefully to ensure proper usage and avoid damage to the MICROTECH Bore Gauge.
MODIFICATIONS
**Item No***
|
Range
|
Depth
|
Rod travel
|
Carbide contacts
| DIAL| DIGITAL| COMPUTERIZED
---|---|---|---|---|---|---|---
res.10 μm
*X – 1
| res. 1 μm
*X – 8
| res.0.1 μm
*X – 7
Accuracy
mm| mm| mm| | μm| μm| μm
132010X00| 6-10| 60| | •| ±8| ±3| ±3
132018X00| 10-18| 110| | •| ±8| ±3| ±3
132050X00|
18-50
| 130|
1,5
| •| ±12| ±3| ±3
132050X11| 500| •| ±12| ±3| ±3
132050X22| 1000| •| ±12| ±3| ±3
132050X44| 2000| •| ±12| ±3| ±3
132160X00|
50-160
| 150|
4
| •| ±15| ±3| ±3
132160X11| 500| •| ±15| ±3| ±3
132160X22| 1000| •| ±15| ±3| ±3
132160X44| 2000| •| ±15| ±3| ±3
132250X00| 160-250| 260| 4| •| ±15| ±5| ±5
132250X22| 1000| •| ±15| ±5| ±5
132250X44| 2000| •| ±15| ±5| ±5
132450X00| 250-450| 260| 6| •| ±22| ±8| ±8
132450X22| 1000| •| ±22| ±8| ±8
132450X44| 2000| •| ±22| ±8| ±8
132900X00| 400-1000| 440|
8
| •| ±30| ±12| ±12
132200X22| 1000| •| ±30| ±12| ±12
132900X44| 2000| •| ±30| ±12| ±12
132951X00| 130-1500| 440| •| ±25| ±8| ±8
BORE GAUGES SETS
132160X00| 18-160| | •| ±15| ±3| ±3
132450X00| 18-450| | •| ±22| ±8| ±8
132990X00| 6-1000| | •| ±30| ±12| ±12
INDICATORS
OPERATION INSTRUCTIONS
Clean oil from the measuring surfaces of bore gauge and indicator. Attach bore
gages body components to each other (if the body consists of two or more
parts). To avoid damaging the rods and threads DO NOT use excessive force when
screwing body. For the zero setting position make sure that the measurement
line of bore gauge coincided with a diametrical cross-section of the setting
ring. Slightly shaking bore gages in the axial plane, determine the largest
reading of indicator, which is the size of ring gauge. Combine the zero mark
of the scale with an arrow by turning ring. Use a clamp to fixate a ring.In
order to increase the accuracy of setting up the position of the bore gauge
must be the same as in the measurement.
For the zero setting position:
- by gauge blocks – the appropriate size block fix in a clamp between two outsets and enter between them bore gauge. Slightly shaking bore gauge in two mutually perpendicular directions, to determine the lowest reading device, which corresponds to the block size.
- by micrometer – fix micrometer and enter between micrometer measuring surface bore gauge. Slightly shaking bore gauge in two mutually perpendicular directions, to determine the lowest reading device, which corresponds to micrometer indication.
IMPORTANT!
Do not make sudden shock to the measuring rod, not put more effort in the case
of jamming. Avoid contact with the indicator and oil emulsion. Do not turn
indicator when it is fixed in the tool holder by a sleeve
Change without prior notice