Temperature calibration in AKODATA devices
Process to follow when an AKODATA device requires calibration.

  •  AKO-59810/1: Temperature sensor inside (Ambient temperature)

  •  AKO-59820/1: Temperature sensor inside and outside (Ambient temperature + outside temperature)
    The following is necessary for calibration:

  •  Access to the akonet.cloud platform

  • A standard temperature master gauge

Step 1
Confirm that the AKODATA device is operating correctly on the akonet.cloud platform. Confirm that the measurement is enabled.
Step 2
Locate the AKODATA device inside the facility and ensure it is not near any air currents or elements that could affect the temperature.
Step 3
Place the Master Gauge as near as possible to the sensor element to be calibrated.AKO 5981H032 Temperature And Moisture Monitor -
Temperature

Step 4
Avoid sudden changes in temperature around the sensor element for 30 minutes.
Step 5
Force a temperature reading and its upload to the akonet.cloud platform.

Step 6
Temperature reading of the sensor element via the akonet.cloud platform.AKO
5981H032 Temperature And Moisture Monitor - Step 6

Step 7
Temperature reading measured by the Master Gauge.
Step 8
Compare the two values.

AKO logo AKO ELECTROMECÁNICA , S.A.L.
Avda. Requites, 30-38
08812 Sant Pere de Rabes.
Barcelona Spain.
www.ako.com
We reserve the right to supply materials that might vary
slightly to those described in our Technical Sheets.
Updated information is available on our website.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals