GIMA M44740EN Manual Hospital Bed User Manual
- June 13, 2024
- GIMA
Table of Contents
GIMA M44740EN Manual Hospital Bed
Product Information
- Product Name: V2K5C (Gima 44740)
- Manufacturer: Jiangsu Saikang Medical Equipment Co.,Ltd
- Manufacturing Location: Zhangjiagang City, Jiangsu Province, China
- Distributors:
- SUNGO Europe B.V. – Amsterdam, Netherlands
- Gima S.p.A – Gessate (MI), Italy
All serious accidents concerning the medical device supplied by us must be reported to the manufacturer and competent authority of the member state where your registered office is located.
PART LISTS
INSTALLATION
CASTERS INSTALLATION
Lay a protective pad and put the bed body upside down; Remove the screw as
shown below, and use a 17-inch hex wrench.
BED FRAME INSTALLATION
Place the bed upright and install the bed plate.
SIDE GUARDRAILS INSTALLATION
Place the bed upright and install the guardrail; (The picture may not exactly
match the real product, this is only for installation instructions) Remove the
screw as shown below, and use a 14-inch hex wrench.
HEADBOARD & FOOTBOARD ASSEMBLY METHOD
(The picture does not exactly match the real product, only for the
installation instructions) The vertical rod used for the head of the bed is
φ21.7×154, and the vertical rod used for the end of the bed is φ21.7×184. The
installation method is the same as the following figure:
IV POLE INSTALLATION
Female part installation: Press the plastic hook (female part) down on the
Φ16mm stainless steel pipe. Male part installation: Place the plastic hook
(male part) directly above the female part in a cross shapes, and then aligns
the buckles on the male part with the slots on the female part and press down.
After hearing a “click”, check whether the end of the buckle is tight, if it
is tight, then complete the installation, if not, continue to press down until
the buckle is tight. Insert the installed IV pole vertically into the IV pole
prevision. (The arrow in the picture shows the IV pole prevision).
USE METHOD
- IV pole
- Head board
- Guardrail
- Bed surface board
- Foot board
- Leg frame
- Drainage hook
- Shows holder
- Caster
USE METHOD OF CASTERS
- When there is a patient on the bed, the guardrail should be raised before moving the bed.
- Except when the bed must be moved, the caster brake must be stepped on during the rest of the time. Need to depress all four wheel brakes.
This bed is not suitable for patients weighing less than 40 kg. The maximum recommended patient weight is168 kg. The safe working load is calculated as follows:
- Maximum patient weight …………………..…… 168kg
- Mattress ………………………………………….. 7kg
- Accessories (including attached loads) …….…25kg
- TOTAL …………………………………………… 200kg
Lock the caster: After pushing the bed to the desired position, press the
ON button with your foot and lock the caster when you hear a “click”, as shown
in the figure below.
Unlock the caster: Press the OFF button with your foot and unlock the
castor when you hear a “click”.
USE METHOD OF GUARDRAILS
- Raising and lowering the guardrail
- Raise the guardrail: Hold the red gun of the guardrail and lift it up. When you hear the sound of “click”, it means the guardrail is locked.
- Lower the guardrail: Hold the guardrail red gun and slowly lay down.
USE METHOD OF DRAINAGE HOOKS
The drainage hooks as shown in the figure.
USE METHOD OF IV POLE
Adjusting the height of the IV pole:
- Loosen the outer sleeve counter clock carefully, insert the inner tube of the IV pole into the outer sleeve, and then adjust the length of the extension rod up and down;
- Adjust the required length, and then tighten the outer sleeve clock carefully. The maximum bearing of each hook is 2Kg, a total of 8kg. (Height Adjustment :890–1540mm)
USE METHOD OF CRANK HANDLES
- Back up and down: Pull out the manual crank handle and turn clock wisely to lift the back and turn counter clock wisely to lower the back.
- Leg up and down: Pull out manual crank handle and turn it counter clock wisely to lift the leg and turn it clock wisely to lower the leg.
SPARE PART REPLACEMENT AND MAINTENANCE
Use 2pcs 17-inch Allen wrenches to unscrew the telescopic rocker knot, take out the bolt, and replace the rocker. When the replacement is completed, install the bolts in the order shown in the figure, and use the wrench to tighten the knot.
FAULTS EXCLUSION
BED-MAINTENANCE
- In order to use the bed safely, safety inspections must be done regularly. Conduct a comprehensive inspection every six months, to ensure that there is no looseness in the connection parts and the function of up-down for the bed can be operated normally.
- At the joints of the frequently moving parts of the bed, an appropriate amount of lubricating oil needs to be added during daily inspection; if serious wear is found, it should be replaced immediately to ensure safety, add lubricated parts as show.
- Avoid hitting the bed and scratching the coating on the bed with sharp objects.
- Please avoid inhaling moisture into the bed and mattress, which will cause rust, abnormal noise and bacteria breeding. If water adheres, wipe it off immediately.
- When the bed frame ages and it reaches a certain servicing time, which the metal part of the bed and the plastic parts of the headboard or footboard, guardrail can be recycle.
BED-CLEANING
- When cleaning, please wring out the cloth soaked in neutral detergent diluted with water and wipe it, then wring out the cloth soaked in water to wipe off the remaining detergent ingredients, and finally use a dry wipe dry with a cloth.
- Do not use volatile substances (thinner, volatile agent, gasoline, etc.), which may cause chemical reactions and damage the bed.
- When using a disinfectant to clean, be sure to dilute it according to its specified concentration before use. Depending on the composition of the disinfectant, it may corrode metal parts, resin parts, etc., causing undesirable phenomena such as discoloration and deformation. Therefore, the content of the disinfectant is recommended as follows:
- 0.05~0.2% ammonium chloride disinfectant
- 0.05~0.2% Chlorinated phenyl disinfectant
- 0.05% diclofenac ethane solution
- 0.05~0.2% sodium hypochlorite disinfectant
Do not use smoking sterilizers, autoclaves, and do not use methyl (phenol) to clean the headboard and foot of the bed. It may cause corrosion, discoloration and deterioration.
Note: When cleaning the bed or changing the bedclothes, please pay attention to the corners, edges and screws of the frame to prevent scratches.
MATTRESS-CLEANING AND MAINTENANCE
- When using it, avoid scratching the fabric with sharp-angled tools or knives, cover it with sheets or cleaning pads, and clean it frequently to protect it from dryness.
- Clean the mattress regularly with a vacuum cleaner, Mattress cover can be washed in 100°C.
- Expose the mattress often to keep the sleep fresh and comfortable, but be careful not to expose the mattress to too long.
- If the fabric is accidentally stained, please refer to the corresponding recommended cleaning methods in the instructions, do not use corrosive chemicals to clean the fabric.
- If accidentally the chewing gum or other jellies get on the fabric, do not wipe it forcibly. Use an ice pack to allow it harden and gently peel it off.
- If the thread is found to be loose due to wear and tear, do not pull it off by hand, use scissors to cut it together.
- If your order does not have a mattress, please refer to the contract configuration.
CASTER-DETECTION AND MAINTENANCE
- Regularly check the casters to avoid hitting or overloading, which will cause damage to the casters.
- Grease the casters or wheels on a regularly, and the frequency of addition should be determined according to the specific actual use conditions. In general, lubricating grease is added every six months.
- After cleaning the caster equipment, pay attention to adding grease. The selection of lubricating grease should be based on the actual use environment, special high temperature and special low temperature environment, and the corresponding lubricating grease should be selected.
- It is forbidden to force the casters to brake when braking, which will cause damage to the brake system.
- The wear of the tire surface of the wheel can be detected by visual inspection. Some thread spun yarn and other accumulations may be wrapped around the wheel, remove the bolts and nuts of the wheel, clean up the debris, and check whether the bearing of the wheel is damaged. If the parts are not damaged, can be reassembled and used.
- Judging whether the caster is replaced according to the time used, wear and tear, you can replace the wheels and make sure that the axles bolt and nuts are well put together and connected. If possible, use a lock washers or locknuts to help you with this process.
- For the casters equipped with brakes, it is necessary to regularly check whether the brakes are working properly. Check the brakes once a day or before each use. For a hospital bed equipped with multiple brake casters, only one brake caster can be locked at a time which will allow you to try and push the bed also check whether one of the brake performance of each caster is good. If the function of the brake fails due to wear or damage to the wheels, immediately replace the wheels and then retest the brake system.
- If the mechanism of the brake system of the caster is damaged, and the brake needs to be repaired or replaced, please contact our after-sales service department or authorized dealer provide by our company. Every time the brakes are replaced, the brake performance of the casters must be retested.
GUARDRAIL-MAINTENANCE
- Please check whether the side rails are installed correctly to prevent the lying patient from shifting and rollingdown.
- Please always check the screws of the total position of the guardrail and the wire to prevent loosening and falling off.
- Please use the red plastic parts of the guardrail switch correctly, and do not use brute force prevent it from damaging the guardrail.
SPARE PART LIST
Note: Due to product updates, parts of the original product may not be available for life. (The company ensures that the sold products can provide the parts that need to be replaced within 6 years after the end of their production)
PACKAGING, TRANSPORT, STORAGE
- The packaging of manual bed is carried out according to the contract or product standard.
- Avoid rushing out, violent vibration, and protection from sunlight and rain during the transportation of the manual bed.
- The manual moving bed should be stored:
- Ambient temperature: -10°C ~ +40°C.
- Relative humidity: ≤ 85%.
- Atmospheric pressure: 70 kPa ~ 106 kPa.
Explanation of symbols:
Note! Check the file that comes with the bed Don’t use the device in
direct contact with the skin
Explanation of symbols:
AFTER-SALES SERVICE
- Please keep the files that comes with the machine and invoices of this product properly, and you need to present these materials when the company performs warranty and maintenance for the product.
- If there is any problem in the process of using, please contact our company in time, so that our company can provide you with accurate and fast technical support and maintenance services in a timely manner.
- Do not disassemble the internal device of this product by yourself to avoid unnecessary damage. If you find that the use is affected due to quality problems, please contact our after-sales service department or our authorized dealer.
- From the date of purchase, if it is indeed damaged or does not work normally due to quality problems within one year, the company will provide to the users a free repairing of the product
- Lifetime service from manufacturer: JIANGSU SAIKANG MEDICAL EQUIPMENT CO LTD.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Jiangsu Saikang Medical Equipment Co.,Ltd
No. 35 Lehong Road, Modern Agriculture Demonstration Park, Zhangjiagang City,
Jiangsu Province, China
Made in China
SUNGO Europe B.V.
Olympisch Stadion 24, 1076DE Amsterdam, Netherlands
Gima S.p.A.
Via Marconi, 1 – 20060 Gessate (MI) Italy
gima@gimaitaly.com
export@gimaitaly.com
www.gimaitaly.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>