HANNA instruments BL983313 EC Process Mini Controller Instruction Manual
- June 12, 2024
- HANNA instruments
Table of Contents
BL983313 EC
Process Mini Controller
Instruction Manual
EC Process Mini Controller Series
- BL983313
- BL983317
- BL983320
- BL983322
- BL983327
TDS Process Mini Controller Series
- BL983315
- BL983318
- BL983319
- BL983321
- BL983324
- BL983329
Dear Customer,
Thank you for choosing a Hanna Instruments ® product.
Please read this instruction manual carefully before using this instrument as
it provides the necessary information for correct use of this instrument as
well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e‑mail us at
tech@hannainst.com.
Visit www.hannainst.com for more information about
Hanna Instruments and our products.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner,
Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
Preliminary Examination
Remove the instrument and accessories from the packaging and examine it
carefully.
For further assistance, please contact your local Hanna Instruments office or
email us at tech@hannainst.com.
Each instrument is supplied with:
- Mounting brackets
- Transparent cover
- 12 VDC power adapter (BL9833XX‑0 only)
- Quick reference guide with instrument quality certificate
Note : Save all packing material until you are sure that the instrument works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.
GENERAL SAFETY & INSTALLATION RECOMMENDATIONS
Procedures and instructions detailed in this manual may require special precautions to ensure personnel safety.
- Electrical connection, installation, start‑up, operation and maintenance must be carried out by specialized personnel only. The specialized personnel must have read and understood the instructions in this manual and should adhere to them.
- User serviceable connections are clearly labeled on the back panel.
- Before powering the controller, verify wiring has been done properly.
- Always disconnect the instrument from power when making electrical connections.
- A clearly marked disconnect switch must be installed in the vicinity of the instrument to ensure that the electrical circuit is completely de‑energized for service or maintenance.
GENERAL DESCRIPTION & INTENDED USE
Hanna Instruments EC and TDS process conductivity mini controller series are
compact panel mount units designed to conveniently measure the electrolytic
conductivity of a process stream.
BL9833XX-Y series configuration
XX | 1 3 | 15 | 17 | 18 | 19 | 20 | 21 | 22 | 24 | 27 | 29 |
---|---|---|---|---|---|---|---|---|---|---|---|
Y | 0 (12 VDC) | 1 (115 or 230 VAC) | 2 (115 or 230 VAC, 4‑20 mA output) |
Intended applications
Quality control of water produced from reverse osmosis, ion exchange,
distillation processes, cooling towers; process control of source water, rinse
water, drinking water, boiler water, and of other industrial, agriculture‑
specific applications
Main Features
- Option to select manual or automatic dosing mode
- Dry contact dosing relay, active when the reading is above/below a programmable setpoint (model dependent)
- Programmable overdosing timer, stops dosing if setpoint is not reached within a specified time interval
- 4‑20 mA galvanic isolated output with external dosing disable contact (BL9833XX‑2 only)
- Temperature compensated readings from 5 to 50 °C (41 to 122 °F)
- Internal fuse protected dosing contacts
- Large, clear LCD and LED operational indicator
- Splash‑resistant transparent cover
Controller Specifications
| B1983313 1| B1983317 1| B1983320 1| B1983322|
BL983327| 81983315| 81983318| 1319833191| 81983321|
181983324| BL983329
---|---|---|---|---|---|---|---|---|---|---|---
Type| EC| TDS
s Unit| PS/01| mS/cm| PS/cm| {6/cm| mS/cm| m9/1 (pR)| 9/1 Opt)| n19/1 4P41)|
n19/1 (pR)| n19/1 (1)011)| n19/1 (ppm)
1 Range| 0-1999| 0.00-10.00| 0.0-199.9| 0.00 —19.99| 0.00-10.00| 0.0-199.9|
0.00-10.00| 0-1999| 0.00-19.99| 0.0 —49.9| 0-999
” Resolution| 1| 0.01| 0.1| 0.01| 0.01| 0.1| 0.01| 1| 0.01| 0.1| 1
-
TDS Factor| —| —| —| —| —| 0.5| 0.5| 0.65| 0.5| 0.5| 0.5
A«rnocy| -±2 % F.S. at 25 °C (77 °F)
Temperature compensation| automatic , from 5 to 50°C (41 to 122 °F), with 0 = 2 W°C
Calibration| manual, with collimation trimmer
Output| galvanic isolated 4-20 mA output; atrium ±0.2 mA; 500 0 maximum load (819833)0(2 only)
Adjustable setpoint| covais measure range
Relay doses when
measurement is| > setpoint| < setpoint| > setpoint| < setpoint| > set point
I Dosing Contact| maximum 2 A (internal fuse protection), 250 VAC o 30 VD(
Overtime| Dosing relay is disabled if setpoint is not reaped within the set time interval. Timer adjustable between oprox. 5 to 30 minutes, or disabled by jumper.
External disable input| Normally Open: enable/Closed: disable dosing (B19833XX-2 only)
12 VD( °dopier| BL983313.0| BL983317-0| BL983320-0| 8L983322-0| BL983327-0| BL983315.0| BL983318.0| BL983319-0| 8L983321-0| 8L9833240| BL983329-0
It- 115/230 VAC| 8L983313•1| 8L983317-1| 8L983320-1| 8L983322-1| 8L983327-1| BL983315.1| BL983318.1| 8L983319-1| 8L983321-1| 8L983324-1| 8L983329-1
115/230 VAC with a. 4-20 mA output| BL983313-2| BL983317-2| BL983320-2| 8L983322-2| 8L983327-2| BL983315.2| N/A| BL983319-2| N/A| N/A| BL983329-2
Input| 10 VA for 115/230 VAC, 50/60 Hz models; 3 W for 12 VDC models; fuse p Acted; installation categoi II.
g HI7632-00| | •| | | •| | •| | | |
at HI7634-00| •| | •| •| | •| | •| •| •| •
Dimensions| 83 x 53 x 92 mm (3.3 x 2.1 x 3.6″)
Weight| 12 VDC models, 200 g (7.1 oz); 115/230 VAC models 300 g (10.6 oz -
Sold separately.
PROBE SPECIFICATIONS
HI7632‑00 and HI7634‑00 probes are sold separately.
| | HI7632-00| HI7634-00
---|---|---|---
Type| Two-pole Amperometric| •
NTC sensor| 4.7 KC)| •| –
9.4 KC)| –| •
Cell constant| 1 cm-‘| •
Materials| PVC body; AN 316 electrodes| •
Temperature| 5 to 50 °C (41 to 122 °F)| •
Maximum pressure| 3 bar| •
Probe length| 64 mm (2.5″)| •
Connection| 1/2″ NPT thread| •
Cable length| 2 m (6.6′)| •
4 m (13.1′)| –| •
5 m (16.41| –| •
_ 6 m (19.7″)| •|
Probe Dimension
Probe Wiring
Easy access to the controller terminals enables quick wiring.
Probe low voltage connections are made to the color coded terminal on the
left.
Note : Calibrate the probe prior to measurement.
Functional Description
6.1. FRONT PANEL
-
LCD
-
Dosing switch
• OFF (dosing disabled)
• AUTO (automatic dosing, setpoint value)
• ON (dosing enabled) -
MEAS key (measurement mode)
-
SET key (configure display value)
-
SET trimmer (adjust setpoint value)
-
CAL trimmer
-
LED operational indicator
• Green – measurement mode
• Orange‑Yellow – active dosing
• Red (blinking) – alarm condition
6.2. REAR PANEL
-
Probe connection terminal, low voltage connections
-
Power supply terminal
• BL9833XX‑1 & BL9833XX‑2, line voltage connections, 115/230 VAC
• BL9833XX‑0, low voltage connections, 12 VDC -
Relay contact acts as a switch for driving the dosing system
-
Jumper for enabling (jumper inserted) or disabling (jumper removed) the overtime control
-
Trimmer for overtime setting (approx. from 5 to 30 minutes)
-
External control for dosing system disabling (BL9833XX‑2)
-
4‑20 mA output contacts (BL9833XX‑2)
INSTALLATION
7.1. UNIT MOUNT
WARNINGS
All external cables connected to the rear panel should be fitted with cable
lugs.
A clearly marked disconnect switch (max. 6A) must be installed in the vicinity
of the instrument to ensure that the electrical circuit is completely
de‑energized for service or maintenance.
7.2. REAR PANEL CONNECTIONS
Probe terminal
- Follow color code to connect the probe.
Power supply termina l
-
BL9833XX‑0
Connect the 2 wires of a 12 VDC power adapter to the +12 VDC and GND terminals. -
BL9833XX‑1 & BL9833XX‑2Connect a 3‑wire power cable paying attention to the correct contacts:
-
earth (PE)
-
ine (L), 115 VAC or 230 VAC
-
neutral (N1 for 115 V or N2 for 230 V)
Dosing Contact
- Dosing contact (NO) output drives the dosing system as per configured setpoint.
Overtime feature (system control)
-
This feature is provided to set the maximum continuous time the relay is running pump or valve, by adjusting the trimmer (from approx. 5 minutes minimum, to approx.
30 minutes maximum). -
When the set time expires, dosing stops, the LED operational indicator turns red (blinking), and “TIMEOUT” message is displayed. To exit, set the dosing switch to OFF then Auto.
-
Remove the jumper from the rear panel to disable the feature. Note : Ensure the dosing switch (front panel) is on Auto for the Overtime feature to be enabled.
External Disabling Contact (NO)
- Normally Open: dosing is enabled.
- Closed: dosing stops, the LED indicator turns red (blinking) and the “HALT” warning message is displayed.
Note : If the dosing switch is ON, dosing continues even with the external disabling contact closed.
OPERATIONS
Hanna® EC and TDS mini controller series are intended to be used to control industrial processes. Relays and White or Brown 50/60Hz; 10 VA Outputs are used to interact with valves or pumps to monitor a process.
CALIBRATION
- If the instrument is not in measurement mode, press MEAS key.
- Immerse the probe in calibration solution. See below table for recommended calibration solutions.
- Shake briefly and allow reading to stabilize.
- Adjust the CAL trimmer until the LCD displays the nominal value given here:
Series | Calibration Solution | Read Value |
---|---|---|
EC | BL983313 | 1413 µS/cm (HI7031) |
BL983317 | 5.00 mS/cm (HI7039) | 5.00 mS |
BL983320 | 84 µS/cm (HI7033) | 84.0 µS |
BL983322 | custom calibration solution about 13 µS/cm or higher | EC solution |
value
BL983327| 5.00 mS/cm (HI7039)| 5.00 mS
TDS| BL983315| 84 µS/cm (HI7033)| 42.0 ppm
BL983318| 6.44 ppt (HI7038)| 6.44 ppt
BL983319| 1413 µS/cm (HI7031)| 919 ppm
BL983321| custom calibration solution about 13 ppm or higher| TDS solution
value
BL983324| 84 µS/cm (HI7033)| 42.0 ppm
BL983329| 1413 µS/cm (HI7031)| 706 ppm
8.2. SETPOINT CONFIGURATION
General: a set point is a threshold value that will trigger control if the measurement value crosses it.
- Press the SET key. The LCD displays the default or previously configured value along with the ”SET“ tag.
- Use a small screwdriver to adjust the SET trimmer to the desired setpoint value.
- After 1 minute the instrument resumes measure mode. If not, press the MEAS key.
Note : The setpoint has a typical hysteresis value comparable to the instrument’s accuracy.
8.3. MONITORING
Best practices
- Ensure wiring is done correctly.
- Ensure setpoint value is configured correctly.
- Ensure probe calibration.
- Select dosing mode.
Procedure
- Immerse (or install) the probe in the solution to be monitored.
- Press the MEAS key (if necessary). The LCD displays the measured value.
• LED indicator lights up Green indicating instrument is in measurement mode and dosing is not active.
• LED indicator lights up Orange/Yellow indicating dosing in progress.
8.4. PROBE MAINTENANCE
Regular cleaning and correct storage is the best way to maximise probe’s life.
- Immerse the tip of the probe in HI7061 Cleaning Solution for 1 hour.
- If a more thorough cleaning is required, brush the metal pins with very fine sandpaper.
- After cleaning, rinse the probe with tap water and recalibrate the meter.
- Store the probe clean and dry.
ACCESSORIES
Ordering Codes | Description |
---|---|
HI7632-00 | EC/TDS probe for high range mini controllers with 2 m (6.6’) cable |
HI7632-00/6 | EC/TDS probe for high range mini controllers with 6 m (19.7’) |
cable
HI7634-00| EC/TDS probe for low range mini controllers with 2 m (6.6’) cable
HI7634-00/4| EC/TDS probe for low range mini controllers with 4 m (13.1’)
cable
HI7634-00/5| EC/TDS probe for low range mini controllers with 5 m (16.4’)
cable
HI70031P| 1413 µS/cm conductivity standard solution, 20 mL sachet (25 pcs.)
HI7031M| 1413 µS/cm conductivity standard solution, 230 mL
HI7031L| 1413 µS/cm conductivity standard solution, 500 mL
HI7033M| 84 µS/cm conductivity standard solution, 230 mL
HI7033L| 84 µS/cm conductivity standard solution, 500 mL
HI70038P| 6.44 g/L (ppt) TDS standard solution, 20 mL sachet (25 pcs.)
HI70039P| 5000 µS/cm conductivity standard solution, 20 mL sachet (25 pcs.)
HI7039M| 5000 µS/cm conductivity standard solution, 250 mL
HI7039L| 5000 µS/cm conductivity standard solution, 500 mL
HI7061M| Cleaning solution for general use, 230 mL
HI7061L| Cleaning solution for general use, 500 mL
HI710005| Power adapter, 115 VAC to 12 VDC, US plug
HI710006| Power adapter, 230 VAC to 12 VDC, European plug
HI710012| Power adapter, 230 VAC to 12 VDC, UK plug
HI731326| Calibration screwdriver (20 pcs.)
HI740146| Mounting brackets (2 pcs.)
CERTIFICATION
All Hanna® instruments conform to the CE European Directives. Disposal of
Electrical & Electronic Equipment. The product should not be treated as
household waste. Instead, hand it over to the appropriate collection point for
the recycling of electrical and electronic equipment, which will conserve
natural resources.
Ensuring proper product disposal prevents potential negative consequences for
the environment and human health. For more information, contact your city,
your local household waste disposal service, or the place of purchase.
RECOMMENDATIONS FOR USERS
Before using this instrument, make sure that it is entirely suitable for your
specific application and for the environment in which it is used. Any
variation introduced by the user to the supplied equipment may degrade the
instrument’s performance.
For your and the instrument’s safety do not use or store the instrument in
hazardous environments.
WARRANTY
The mini controllers are warranted for a period of two years against defects
in workmanship and materials when used for their intended purpose and
maintained according to instructions. This warranty is limited to repair or
replacement free of charge. Damage due to accidents, misuse, tampering, or
lack of prescribed maintenance is not covered. If service is required, contact
your local Hanna Instruments ® office.
If under warranty, report the model number, date of purchase, serial number
and the nature of the problem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments office,
first obtain a Returned Goods Authorization (RGA) number from the Technical
Service department and then send it with shipping costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
MANBL983313 09/22
Documents / Resources
|
HANNA instruments BL983313 EC Process Mini
Controller
[pdf] Instruction Manual
BL983313, BL983317, BL983320, BL983322, BL983327, BL983313 EC Process Mini
Controller, EC Process Mini Controller, Process Mini Controller, Mini
Controller, Controller
---|---