QUICK REFERENCE GUIDE
HI510 Smart Probes

HI510 Smart Probes

HANNA Instruments HI510 Smart ProbesPlease scan the QR code to download the user manual or follow the link.
https://manuals.hannainst.com/HI510 Thank you for choosing Hanna Instruments!
Hanna Instruments digital probes allow shared management of settings between probe and HI510 process controller. The probe manages temperature compensation and buffer calibration measurement settings. Smart technology enables probe optimization for
different temperature applications and ranges for all supported parameters.
Data captured on the probe is automatically updated on the controller via digital input.

Supported Probe Series and Configurations

HI10 – pH & TEMPERATURE

XX 6 PTFE junction
16 Ceramic junction
Y Sensor Glass
1 Low Temperature Titanium Matchim

to 176.0 °F)
3| High Temperature| 0.00 to 14.00 pH| 0.0 to 100.0 °C (32.0 to 212.0 °F)
4| Fluoride Resistant| 0.00 to 10.00 pH| —5.0 to 60.0 °C (23.0 to 140.0 °F)

HI20 – ORP & TEMPERATURE

XX 4 PTFE junction
14 Ceramic junction
Y Sensor Type mV range
1 Platinum ± 2000 mV
2 Gold

HI7630 – CONDUCTIVITY & TEMPERATURE

Y| 2| Two-electrode cell conductivity, AISI 316 stainless steel, silicone cell constant k —– 0.1/cm| EC 0.000 pSAm to 30.00 mS/cm
TDS 0.000 mg/L to 15.00 g/L (TDS factor 0.5)
RES 34 a • cm to 99.99 Ma • cm
Temperature 0.0 to 50.0 °C (32.0 to 122.0 °F)
---|---|---|---
4| Four-ring conductivity, platinum on glass, cell constant k —– 1.0/cm| EC 0.0 pS/cm to 999.9 mS/cm
TDS 0.0 mg/L to 400.0 g/L (TDS factor 0.5)
RES 1.00 0 • cm to 9.99 MO • cm
Seawater Salinity 400.0 WoNaCI, 42 psu, 80 ppt
Temperature 0.0 to 100.0 °C (32.0 to 212.0 °F)

HI7640 – DISSOLVED OXYGEN & TEMPERATURE

Y| 1| Galvanic sensor| Concentration  0.00 to 50.00 mg/L (ppm)
Saturation 0.0 to 500.0 %
Temperature −5.0 to 50.0°C (23.0 to 122.0 °F)
---|---|---|---
5| Optical sensor| Concentration  0.00 to 50.00 mg/L (ppm)
Saturation 0.0 to 500.0 %
Temperature −5.0 to 50.0°C (23.0 to 122.0 °F)
8| Smart probe, with RS485 connection
---|---
ZZ| 00 supplied with DIN connector (without cable)
05, 10, 15, 25, 50 fixed cable length (in meters)

Probe wiring

  1. Disconnect the controller from power. Run the probe cable through the conduit opening.
  2. Connect the probe leads to the removable terminal connector marked PROBE. Follow lead markings (–/+) to ensure correct wiring position for output leads.
  3. Carefully put the wired terminal connector into place on the board.
  4. Position excess cable through cable gland before tightening the nut.
  5. Remove the ground screw and hardware located below the PROBE connector. Attach the ground lead ( ****).

HANNA Instruments HI510 Smart Probes - fig 1 Probe cabling color code

Marking Attached Cable Patch Cable Functionality
GREEN BLACK 0 V
B WHITE WHITE RS485 D-
A  YELLOW BLUE RS485 D+
+ BROWN RED 5 V
**** GREEN-YELLOW GREEN-YELLOW PROTECTIVE GROUND

Probe connection

Align the pins and key then push the plug into the socket. Rotate the collar to lock in place. ****All Hanna instruments conform to the CE European Directives and our production facilities are ISO 9001 certified. Probes are warrantied for six months (HI7640-58, two years) against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. For technical information e‑mail us attech@hannainst.com or view our contact list at www.hannainst.com.

Hanna Instruments Inc.,
584 Park East Drive,
Woonsocket, RI 02895 USA
www.hannainst.com
Please retain for future use.
QR510-P
05/22

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

HANNA instruments User Manuals

Related Manuals