HANNA Instruments HI510 Smart Probes User Guide
- October 30, 2023
- HANNA instruments
Table of Contents
QUICK REFERENCE GUIDE
HI510 Smart Probes
HI510 Smart Probes
Please scan
the QR code to download the user manual or follow the link.
https://manuals.hannainst.com/HI510 Thank you for choosing Hanna
Instruments!
Hanna Instruments digital probes allow shared management of settings between
probe and HI510 process controller. The probe manages temperature compensation
and buffer calibration measurement settings. Smart technology enables probe
optimization for
different temperature applications and ranges for all supported parameters.
Data captured on the probe is automatically updated on the controller via
digital input.
Supported Probe Series and Configurations
HI10 – pH & TEMPERATURE
XX | 6 | PTFE junction |
---|---|---|
16 | Ceramic junction | |
Y | Sensor Glass | |
1 | Low Temperature | Titanium Matchim |
to 176.0 °F)
3| High Temperature| 0.00 to 14.00 pH| 0.0 to 100.0 °C (32.0 to 212.0 °F)
4| Fluoride Resistant| 0.00 to 10.00 pH| —5.0 to 60.0 °C (23.0 to 140.0 °F)
HI20 – ORP & TEMPERATURE
XX | 4 | PTFE junction |
---|---|---|
14 | Ceramic junction | |
Y | Sensor Type | mV range |
1 | Platinum | ± 2000 mV |
2 | Gold |
HI7630 – CONDUCTIVITY & TEMPERATURE
Y| 2| Two-electrode cell conductivity, AISI 316 stainless steel, silicone cell
constant k —– 0.1/cm| EC 0.000 pSAm to 30.00 mS/cm
TDS 0.000 mg/L to 15.00 g/L (TDS factor 0.5)
RES 34 a • cm to 99.99 Ma • cm
Temperature 0.0 to 50.0 °C (32.0 to 122.0 °F)
---|---|---|---
4| Four-ring conductivity, platinum on glass, cell constant k —– 1.0/cm| EC
0.0 pS/cm to 999.9 mS/cm
TDS 0.0 mg/L to 400.0 g/L (TDS factor 0.5)
RES 1.00 0 • cm to 9.99 MO • cm
Seawater Salinity 400.0 WoNaCI, 42 psu, 80 ppt
Temperature 0.0 to 100.0 °C (32.0 to 212.0 °F)
HI7640 – DISSOLVED OXYGEN & TEMPERATURE
Y| 1| Galvanic sensor| Concentration 0.00 to 50.00 mg/L (ppm)
Saturation 0.0 to 500.0 %
Temperature −5.0 to 50.0°C (23.0 to 122.0 °F)
---|---|---|---
5| Optical sensor| Concentration 0.00 to 50.00 mg/L (ppm)
Saturation 0.0 to 500.0 %
Temperature −5.0 to 50.0°C (23.0 to 122.0 °F)
8| Smart probe, with RS485 connection
---|---
ZZ| 00 supplied with DIN connector (without cable)
05, 10, 15, 25, 50 fixed cable length (in meters)
Probe wiring
- Disconnect the controller from power. Run the probe cable through the conduit opening.
- Connect the probe leads to the removable terminal connector marked PROBE. Follow lead markings (–/+) to ensure correct wiring position for output leads.
- Carefully put the wired terminal connector into place on the board.
- Position excess cable through cable gland before tightening the nut.
- Remove the ground screw and hardware located below the PROBE connector. Attach the ground lead ( ****).
Probe cabling color code
Marking | Attached Cable | Patch Cable | Functionality |
---|---|---|---|
— | GREEN | BLACK | 0 V |
B | WHITE | WHITE | RS485 D- |
A | YELLOW | BLUE | RS485 D+ |
+ | BROWN | RED | 5 V |
**** | GREEN-YELLOW | GREEN-YELLOW | PROTECTIVE GROUND |
Probe connection
Align the pins and key then push the plug into the socket. Rotate the collar to lock in place. ****All Hanna instruments conform to the CE European Directives and our production facilities are ISO 9001 certified. Probes are warrantied for six months (HI7640-58, two years) against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. For technical information e‑mail us attech@hannainst.com or view our contact list at www.hannainst.com.
Hanna Instruments Inc.,
584 Park East Drive,
Woonsocket, RI 02895 USA
www.hannainst.com
Please retain for future use.