NATIONAL INSTRUMENTS PXIe-6570 32-Channel Digital Pattern Instrument User Guide
- June 12, 2024
- NATIONAL INSTRUMENTS
Table of Contents
- COMPREHENSIVE SERVICES
- SELL YOUR SURPLUS
- OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
- GETTING STARTED GUIDE
- Verifying the System Requirements
- Unpacking the Kit
- Installing the Software
- Installing the PXIe-6570
- Connecting Signals
- Configuring the PXIe-6570 in MAX
- Operating the Digital Pattern Instrument
- Troubleshooting
- Where to Go Next
- Worldwide Support and Services
COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series.
We work out the best solution to suit your individual needs.
Sell For Cash Get Credit Receive a Trade-In Deal
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.
Bridging the gap between the manufacturer and your legacy test system.
1-800-915-6216
www.apexwaves.com
sales@apexwaves.com
All trademarks, brands, and brand names are the property of their respective owners.
GETTING STARTED GUIDE
PXIe-6570
32-Channel Digital Pattern Instrument
Note Before you begin, install and configure your chassis and controller.
This document explains how to install, configure, and test the PXIe-6570. The PXIe-6570 is a 32-channel digital pattern instrument.
Verifying the System Requirements
To use the NI-Digital Pattern Driver and Digital Pattern Editor, your system
must meet certain requirements.
Refer to the product readme, which is available online at
ni.com/manuals or on the NI-Digital Pattern Driver
and Digital Pattern Editor Media (NI part number 785249-03), for more
information about minimum system requirements, recommended system, and
supported application development environments (ADEs).
Unpacking the Kit
Notice To prevent electrostatic discharge (ESD) from damaging the module, ground yourself using a grounding strap or by holding a grounded object, such as your computer chassis.
-
Touch the antistatic package to a metal part of the computer chassis.
-
Remove the module from the package and inspect it for loose components or other signs of damage.
Notice Never touch the exposed pins of connectors.
Note Do not install a module if it appears damaged in any way. -
Unpack any other items and documentation from the kit.
Store the module in the antistatic package when the module is not in use.
Kit Contents
The kit contains the following items:
- PXIe-6570 Digital Pattern Instrument
- PXIe-6570 Safety, Environmental, and Regulatory Information
- PXIe-6570 Getting Started Guide (this document)
Note You can download the NI-Digital Pattern Driver and Digital Pattern Editor at ni.com/downloads or purchase the NI-Digital Pattern Driver and Digital Pattern Editor media (NI part number 785249-03) separately.
Installing the Software
You must be an Administrator to install NI software on your computer.
-
Install an ADE, such as LabVIEW or Microsoft Visual Studio.
-
Insert the software media USB or download the NI-Digital Pattern Driver and Digital Pattern Editor at ni.com/downloads. The installer should open automatically after inserting the software media USB.
If the installation window does not appear, navigate to the drive, double- click it, and double-click autorun.exe. -
Follow the instructions in the installation prompts.
a) Select .NET Framework 4.0 Languages Support or .NET Framework 4.5 Languages Support in the NI-Digital Pattern Driver installer to install .NET support for the NI-Digital Pattern Driver software.
Note Windows users may see access and security messages during installation. Accept the prompts to complete the installation. -
When the installer completes, select Restart in the dialog box that prompts you to restart, shut down, or restart later.
Digital Pattern Editor
Use the Digital Pattern Editor to view, create, modify, and debug pin and channel maps, specifications, levels, timing, pattern, source waveform, and capture waveform files. You can also use the Digital Pattern Editor to configure the state of the digital pattern instrument.
The Digital Pattern Editor is installed when you install NI-Digital Pattern Driver. For more information, refer to the Digital Pattern Help.
Documentation for the Digital Pattern Instrument, NI-Digital Pattern Driver, and Digital Pattern Editor is available from the Start menu by navigating to NI Digital Pattern Documentation in the National Instruments folder. The primary help file is digipat.chm, and all other help files in the folder are embedded in digipat.chm.
Related Information
Digital Pattern Instrument Examples on page 11
Operating the Digital Pattern Instrument on page 10
You can operate the PXIe-6570 through use of the Digital Pattern Editor or the supported ADE of your choice.
Digital Pattern Help
NI-Digital Pattern Driver
NI-Digital Pattern Driver LabVIEW API
Use the NI-Digital VIs and properties in LabVIEW to configure and control
the digital pattern instrument. The NI-Digital Pattern Driver LabVIEW API is
installed by default when you run the installer. For more information, refer
to the Digital Pattern Help.
NI-Digital Pattern Driver .NET API
You can optionally install and use the NI-Digital Pattern Driver .NET API
to configure and control the digital pattern instrument. For more information,
refer to the Digital Pattern Help.
Documentation for the Digital Pattern Instrument, NI-Digital Pattern Driver, and Digital Pattern Editor is available from the Start menu by navigating to NI Digital Pattern Documentation in the National Instruments folder. The primary help file is digipat.chm, and all other help files in the folder are embedded in digipat.chm.
NI-Digital Pattern Driver C API
You can use the NI-Digital Pattern Driver C API dynamically linked library
to configure and control the digital pattern instrument. Add a reference to
C:\Program Files (x86)\IVI Foundation\IVI\Bin\niDigital_32.dll for 32-bit
development or to C:\Program Files\IVI Foundation\IVI\Bin\niDigital_64.dll for
64-bit development.
Related Information
Digital Pattern Instrument Examples on page 11
Operating the Digital Pattern Instrument on page 10
You can operate the PXIe-6570 through use of the Digital Pattern Editor or the supported ADE of your choice.
Digital Pattern Help
Associated Drivers
NI-DCPower
The NI-DCPower driver is required to use NI-DCPower modules within the
Digital Pattern Editor. For more information about using NI-DCPower modules
within the Digital Pattern Editor, refer to the Digital Pattern Help. Visit
ni.com/downloads to download NI-DCPower.
NI-Sync
The NI-Sync driver is required to support operations using the PXIe-6674T
timing and synchronization module, such as combining comparison results across
digital pattern instruments. For more information about using the PXIe-6674T
with digital pattern instruments, refer to the Digital Pattern Help. Visit
ni.com/downloads to download NI-Sync.
Installing the PXIe-6570
1. Ensure the AC power source is connected to the chassis before installing
the module.
The AC power cord grounds the chassis and protects it from electrical damage
while you install the module.
2. Power off the chassis.
3. Inspect the slot pins on the chassis backplane for any bends or damage
prior to installation. Do not install a module if the backplane is damaged.
4. Remove the black plastic covers from all the captive screws on the module
front panel.
5. Identify the slot you will use in the chassis.
The PXIe-6570 is a two-slot module with two PXI Express backplane connectors.
The module must be installed into two adjacent PXI Express-compatible slots.
If you plan to use the optional PXIe-6674T Timing and Synchronization Module,
reserve the PXI Express System Timing Slot for its placement in the chassis.
The following figure shows the symbols that indicate the slot types.
Figure 1. Chassis Compatibility Symbols
- PXI Express System Controller Slot
- PXI Peripheral Slot
- PXI Express Hybrid Peripheral Slot
- PXI Express System Timing Slot
- PXI Express Peripheral Slot
6. Touch any metal part of the chassis to discharge static electricity.
7. Place the module edges into the module guides at the top and bottom of the
chassis. Slide the module into the slot until it is fully inserted.
Figure 2. PXIe-6570 Installation
- PXI Express Chassis
- Ejector Handle in Downward (Unlatched) Position
- PXI Express Two-Slot Module
- Captive Screw
8. Secure the module front panel to the chassis using the front-panel
mounting screws.
Note Tightening the top and bottom mounting screws increases
mechanical stability and also electrically connects the front panel to the
chassis, which can improve the signal quality and electromagnetic performance.
9. Cover all empty slots using EMC filler panels or fill using slot blockers
to maximize cooling air flow, depending on your application.
10. Power on the chassis.
Connecting Signals
PXIe-6570 Front Panel
The PXIe-6570 front panel has a single 68-pin VHDCI Digital Data and Control (DDC) connector. The following figure shows the front panel and pinout of the PXIe-6570 digital pattern instrument.
Figure 3. PXIe-6570 Front Panel
Table 1. PXIe-6570 Digital Data and Control Connector Pins
Pins
| Signal Name| Signal Type|
Signal Description
---|---|---|---
1, 3, 5, 7, 9, 11, 13, 15, 17, 19, 21, 23, 25, 27, 29, 31, 35, 37, 39, 41, 43,
45, 47, 49, 51, 53, 55, 57, 59, 61, 63, 65| DIO <0..31>| Data| Bidirectional
PPMU-capable digital I/O data channels 0 through 31.
2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 22, 24, 28, 32, 34, 36, 38, 40, 42, 44,
46, 48, 50, 52, 54, 56, 58, 62, 66, 68 | GND| Ground| Instrument ground. Also
default ground reference if DUT Ground Sense (DGS) is not connected.
26| CAL MEASURE| Analog| Resource for external calibration.
33| CAL SENSE| Analog| Resource for external calibration.
64| DGS| Ground| Optional DGS for improved accuracy at higher currents in some
configurations.
CAL GND| Analog| Resource for external calibration.
67| CAL FORCE| Analog| Resource for external calibration.
30, 60| RESERVED| N/A| These terminals are reserved for future use. Do not
connect to these pins.
Note The 32 digital I/O data channels on the digital pattern instrument are split into banks for PPMU operation efficiency: DIO<0..7>, DIO<8..15>, DIO<16..23>, DIO<24..31>. PPMU and frequency measurements run in parallel when you take measurements on channels in different banks.
LED Status Indicators
Status indicators on the front panel of the PXI-6570 provide feedback about
instrument operation.
Use the following table to determine the PXIe-6570 state using the Access LED status indicator.
Table 2. Access LED Status Indicator
Color
|
Indications
---|---
Off| Instrument not yet functional.
Amber| Instrument being accessed by software.
Green| Instrument ready to be programmed.
Use the following table to determine the PXIe-6570 state using the Active LED status indicator.
Table 3. Active LED Status Indicator
Color
|
Indications
---|---
Off| Instrument is not waiting for a trigger, not bursting a pattern, and not
experiencing an error. During PPMU operations, the Active LED remains off.
Amber| Instrument awaiting Start trigger.
Green| Instrument received Start trigger and is bursting a pattern.
Red| Instrument error condition.
Possible error conditions:
- The instrument has detected an unlocked condition on a PLL that might result in incorrect data. To clear this error condition, reset the instrument programmatically or through Measurement & Automation Explorer (MAX).
- The instrument has been disabled because it exceeded its overall power limit. To re-enable the instrument, reset it programmatically or through MAX.
- The instrument has been disabled because it exceeded its overall temperature limit. To re-enable the instrument, cool the instrument to an acceptable range and resolve the environmental condition that caused the shutdown. Reset the instrument programmatically or through MAX, or power cycle the instrument.
Note You will receive an error if self-test, self-calibration, or external calibration failed on the instrument, but the LED will remain green. The error will persist until self-test or calibration succeeds or the instrument is reset.
Connecting to a Device Under Test
To connect a device under test (DUT) to the PXIe-6570, you can create a custom device interface board to access the signals of the DUT. For more information, refer to the Interfacing to the Digital Pattern Instrument or Digital Waveform Instrument using the VHDCI Connector application note. Go to ni.com/info and enter the info code rdinwa to locate the application note.
Note If you plan to use the PXIe-6570 as part of an integrated
system, such as a Semiconductor Test System (STS), refer to the system
documentation for guidance on device interface board design and connection
guidelines. For system documentation, contact your NI sales engineer.
To connect a custom device interface board to the VHDCI Digital Data and
Control connector on the PXIe-6570, use a mating connector for the VHDCI cable
from NI.
PXIe-6570 Accessories
Description
|
NI Part Number
---|---
NI CB-2162 Connector Block and Prototyping Board| 778592-01
SCB-68 HSDIO Shielded Connector Block for R Series and HSDIO Devices|
782914-01
SMB-2163 SMB Breakout Box for High-Speed Digital Devices| 778747-01
NI SHC68-C68-D4 Shielded Single-Ended Cable for High-Speed Digital Devices|
781013-01 – 0.55 m
196275-01 – 1 m
152870-01 – 1 m
Low Leakage
781293-01 – 2 m
SHC68-H1X38 High-Speed Digital Flying-Lead Cable, 1.5 m| 192681-1R5
VHDCI connectors| 780389-01
778914-01
780390-01
NI-Digital Pattern Driver and Digital Pattern Editor Media| 785249-03
Note For more information on recommended accessories for use within a system such as STS, contact your NI Sales Engineer.
Configuring the PXIe-6570 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX informs other programs about which NI hardware products are in the system and how they are configured. MAX is automatically installed with NI-Digital Pattern Driver.
-
Launch MAX.
-
In the configuration tree, expand Devices and Interfaces to see the list of installed NI hardware.
Installed modules appear under the name of their associated chassis. -
Expand your Chassis tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note If you do not see your module listed, pressto refresh the list of installed modules. If the module is still not listed, power off the system, ensure the module is correctly installed, and restart. -
Record the identifier MAX assigns to the hardware. Use this identifier when programming the PXIe-6570.
-
Self-test the hardware by selecting the item in the configuration tree and clicking Self-Test in the MAX toolbar.
The MAX self-test of the PXIe-6570 performs a basic verification of hardware resources and lasts several minutes.
Operating the Digital Pattern Instrument
You can operate the PXIe-6570 through use of the Digital Pattern Editor or the supported ADE of your choice.
Table 4. PXIe-6570 Interactive Use Options
Application Software
| Location|
Description
---|---|---
Digital Pattern Editor| In the Start menu, navigate to NI Digital
Pattern Editor in the National Instruments folder to launch the Digital
Pattern Editor.| Use the Digital Pattern Editor to view, create, modify, and
debug pin and channel maps, specifications, levels, timing, pattern, source
waveform, and capture waveform files. You can also use the Digital Pattern
Editor to configure the state of the digital pattern instrument.
Table 5. PXIe-6570 Programming Options
Application Programming Interface (API)
| Location|
Description
---|---|---
NI-Digital Pattern Driver LabVIEW API|
- LabVIEW 2015— Functions» Instrument I/O»Instrument Drivers»NI-Digital palette
- LabVIEW 2016 and later— Functions»Measurement I/O»NI-Digital palette or Functions»Instrument I/O» Instrument Drivers»NI-Digital palette
| Use the NI-Digital VIs and properties in LabVIEW to configure and control
the digital pattern instrument. The NI-Digital Pattern Driver LabVIEW API is
installed by default when you run the installer.
NI-Digital Pattern Driver .NET API| You can use the NI-Digital Pattern .NET
class library by adding a reference to NationalInstruments.
ModularInstruments.NIDigital.Fx40 or NationalInstruments.
ModularInstruments.NIDigital.Fx45 and any dependent class libraries from
within the Solution Explorer in Visual Studio.| You can optionally install and
use the NI-Digital Pattern Driver .NET API to configure and control the
digital pattern instrument.
Digital Pattern Instrument Examples
To locate programming and operating examples, refer to the following table.
Table 6. NI-Digital Pattern Driver Example Locations
Application Software or Application Development Environment (ADE)
|
Location
---|---
Getting Started| Use the getting started examples, located by default in the
\National Instruments\NI-Digital \Examples\Getting Started directory, to help
you learn key concepts or to serve as a starting point for applications you
create.
Digital Pattern Editor| In the Start menu, navigate to NI Digital
Pattern Editor in the National Instruments folder to launch the Digital
Pattern Editor. In the Welcome window, click Examples or click the
Learning tab to launch the Learning window, which includes examples of common
usages of digital patterns.
LabVIEW| Use the NI Example Finder to locate LabVIEW examples. In LabVIEW,
select Help»Find Examples to launch the NI Example Finder. Find examples
specific to NI-Digital by selecting Hardware Input and Output»Modular
Instruments»NI-Digital Pattern Driver.
Microsoft Visual Studio| The Microsoft .NET examples are located in the
\National Instruments\NI-Digital \Examples\DotNET 4.x directory. In the
Start menu, navigate to NI Digital Pattern Examples in the National
Instruments folder.
Troubleshooting
If an issue persists after you complete a troubleshooting procedure, contact NI technical support or visit ni.com/support.
What Should I Do if the PXIe-6570 Doesn’t Appear in MAX?
- In the MAX configuration tree, expand Devices and Interfaces.
- Expand the Chassis tree to see the list of installed hardware, and press
to refresh the list. - If the module is still not listed, power off the system, ensure that all hardware is correctly installed, and restart the system.
- Navigate to the Device Manager.
Operating System| Description
---|---
Windows 10| Right-click the Start button, and select Device Manager.
Windows 7| Select Start»Control Panel»Device Manager. - Verify the PXIe-6570 appears in the Device Manager.
a) Under an NI entry, confirm that a PXIe-6570 entry appears.
Note If you are using a PC with a device for PXI remote control system, under System Devices , also confirm that no error conditions appear for the PCI-to-PCI Bridge.
b) If error conditions appear, reinstall NI-Digital Pattern Driver and the PXIe-6570.
What Should I Do if the PXIe-6570 Fails the Self-Test?
- Restart the system.
- Launch MAX, and perform the self-test again.
- Power off the chassis.
- Reinstall the failed module in a different slot.
- Power on the chassis.
- Perform the self-test again.
Why Is the ACCESS LED Off When the Chassis Is On?
The LEDs may not light until the module has been configured in MAX. Before proceeding, verify that the PXIe-6570 appears in MAX.
If the ACCESS LED fails to light after you power on the chassis, a problem may exist with the chassis power rails, a hardware module, or the LED.
Notice Apply external signals only while the PXIe-6570 is powered on. Applying external signals while the module is powered off may cause damage.
-
Disconnect any signals from the module front panels.
-
Power off the chassis.
-
Remove the module from the chassis and inspect it for damage. Do not reinstall a damaged module.
-
Install the module in a different chassis slot from which you removed it.
-
Power on the chassis.
Note If you are using a PC with a device for PXI remote control system, power on the chassis before powering on the computer. -
Verify that the module appears in MAX.
-
Reset the module in MAX and perform a self-test.
Where to Go Next
Located online at ni.com/manuals Located using the NI Example Finder
EXPLORE| | DEVELOP + DEBUG| | PROGRAM
the Digital Pattern Editor, application development environment (ADE) and
instrument specifications.
Digital Pattern Help
Getting Started with LabVIEW
National Instruments .NET Support
PXIe-6570 Specifications
| |
patterns and configuration files for the digital pattern instrument.
Develop pin and channel map files, specifications files, pin levels, time sets, and pattern files using the Digital Pattern Editor.
Debug with pin view pane, History RAM results, Shmoo plots, and digital scope.
| |
custom applications with NI-Digital Pattern Driver LabVIEW or .NET API.
NI-Digital Pattern Driver
NI-Digital Pattern Examples
Digital Pattern Help
| | | |
DISCOVER
more about your products through ni.com.
Support
ni.com/support
Digital Instruments
ni.com/digital-instruments
Services
ni.com/services
NI Community
ni.com/community
*This item is also installed with the driver software.
Worldwide Support and Services
The NI website is your complete resource for technical support. At ni.com/support , you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for information about the services NI offers.
Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information.
Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2016—2018 National Instruments. All rights reserved.
376528B-01 June 20, 2018
References
- Test and Measurement Systems, a part of Emerson - NI
- NI Community - National Instruments
- Digital Instruments - NI
- Software and Driver Downloads - NI
- Using Info Codes - NI
- Product Documentation - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- NI Trademarks and Logo Guidelines - NI
- Software and Driver Downloads - NI
- Using Info Codes - NI
- Product Documentation - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- PXIe-6570 National Instruments PXI Digital Pattern Instrument | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>