blacklinesafety G7 Multi-Gas Detector with Pump User Guide
- June 12, 2024
- blacklinesafety
Table of Contents
blacklinesafety G7 Multi-Gas Detector with Pump
Product Information
The Multi-Gas Pump Cartridge Guide provides information on the hardware details and configuration modes of the G7 device. It explains how to use the different modes, including normal mode, pre-entry mode, leak check mode, pump run mode, SCBA mode, and high-risk mode. The guide also provides instructions on entering and exiting modes, as well as information on mode timeout. Additionally, the guide explains how to use the G7’s pump function. It provides step-by-step instructions on sampling withthe pump, turning the pump on and off, and performing block teststo ensure the hose and pump cartridge are in working order. The guide covers both automatic and manual block tests.
Product Usage Instructions
Entering a Mode
- Press the OK button to enter G7’s Main menu.
- Use the up and down arrows to navigate to Modes and press the OK button.
- Select the desired mode.
- Select Yes to confirm the mode.
- G7’s screen will invert and the banner will display your current mode.
Exiting a Mode
- Press the OK button to enter G7’s Main menu.
- Use the up and down arrows to navigate to Modes and pressOK.
- Select Normal mode.
- Select Yes to confirm.
- G7’s screen will revert to normal mode.
Mode Timeout
Each mode (excluding normal, high risk, and pump run mode) has a timeout period. After this period has elapsed, you can extend or leave the mode. If you do not make a choice within 30 seconds, G7 will automatically return to normal mode.
Using G7’s Pump
Sample with Pump
- Attach a quick-connect to your desired length of hose.
- Attach the quick connect to the stack by lining up the guides and twisting clockwise.
Turning Pump On
- Ensure G7 is equipped with a Multi-pump cartridge and pumped modes, such as pre-entry, leak check, or pump run.
- Press the OK button to enter G7’s Main menu.
- Use the up and down arrows to navigate to Modes and press the OK button, then select the desired mode.
- Attach the hose and perform a block test by following the instructions on G7’s screen.
- G7’s screen will invert, and the pump icon in the banner will indicate that the pump is running.
Turning Pump Off
To turn off the pump, enter any non-pumped mode.
Block Tests
A block test is performed before using the pump to ensure that the hose and pump cartridge are in working order. There are two ways to perform a block test: automatic and manual.
Automatic Block Tests
G7 automatically performs a block test before entering a pumped mode. Attach the hose to the pump and follow the instructions on G7’s screen to complete the test. If the block test passes, the pump will start. If it fails, G7 will return to the previous mode, and the pump will not start. During an automatic block test, G7 is in a safe state where gas alerts will not be triggered to prevent false alerts from residual gas in the hose.
Manual Block Tests
A manual block test can be performed anytime the pump is running. To perform a manual block test:
- Ensure you are in a pumped mode with a quick connect and hose attached to your device.
- Block the end of your inlet.
- Wait for the yellow warning alarm.
- If the alarm sounds, it indicates that the block test hasfailed, and you should stop using the pump. If no alarm sounds, the block test is successful.
HARDWARE DETAILS
CONFIGURATION MODES
G7 supports six different modes, with settings defined from G7’s configuration profile in Blackline Live. Modes allow G7 to temporarily change its behavior, including enabling pump or muting gas alarms to prevent false alerts. Entering a pumped configuration mode is how the user switches G7’s operation from diffusion to pump.
Available configuration modes
Normal mode is the configuration used for
everyday operation. G7 operates using this mode
as a default.
Pre-entry mode is used for testing an environment prior to entering.
Leak check mode can be used when checking for gas leaks in a particular
area.
**Pump run mode can be used to continually run pump with no timeout.
SCBA mode is used when the device user is wearing a self-contained or
supplied air breathing apparatus.
High risk mode is used for general high-risk situations. This mode does
not have a timeout.
NOTE:** indicates a pumped mode.
USING CONFIGURATION
Entering a mode
-
Press the OK button to enter G7’s Main menu.
-
Use the up and down arrows to navigate to
Modes and press the OK button. -
Select the desired mode.
-
Select Yes to confirm the mode.
-
G7’s screen will invert and the banner will
Exiting a mode
-
Press the OK button to enter G7’s Main menu.
-
Use the up and down arrows to navigate to
Modes and press OK. -
Select Normal mode.
-
Select Yes to confirm. display your current mode.
Mode timeout
Each mode (excluding normal, high risk and pump run mode) has a timeout
period. After this period has elapsed, you can extend or leave the mode. If
you do not make a choice in 30 seconds, G7 will return to normal mode.
USING G7’S PUMP
Sample with pump
1. Attach a quick-connect to your desired length of hose.
2. Attach the quick connect to the stack by lining up the guides and twisting
clockwise.
Turning pump on
- Ensure G7 is equipped with a Multi-pump cartridge and pumped modes, such as pre-entry, leak check or pump run.
- Press the OK button to enter G7’s Main menu.
- Use the up and down arrows to navigate to Modesand press the OK button, then select the desired mode.
- Attach hose and perform a block test by followingthe instructions on G7’s screen.
- G7’s screen will invert, and the pump icon in the banner will indicate pump is running.
Turning pump off
To turn pump off, enter any non-pumped mode.
BLOCK TESTS
A block test is performed before using pump to ensure that your hose and pump
cartridge are in working order. G7 has two ways of performing a block test:
automatic and manual.Repeat as necessary
Automatic block tests
G7 will automatically perform a block test before you enter a pumped mode.
Attach hose to pump and follow the instructions on G7’s screen to complete the
test. If G7 passes the block test, pump will start. If the block test fails,
G7 will return to the previous mode and pump will not start. While performing
an automatic block test, G7 is in a safe state where gas alerts will not be
triggered. This prevents false alerts from residual gas in the hose.
Manual block tests
A manual block test can be performed any time pump is running. To perform a
manual block test:
Ensure you are in a pumped mode, with a quick connect and hose attached to
your device.
- Block the end of your inlet.
- Wait for the yellow warning alarm to sound.
- Unblock the end of your inlet.
- If the yellow warning alarm silences, your equipment is safe to use.
Yellow warning alarm
A yellow warning alarm will be triggered when G7’s pump becomes blocked. “Pump
blocked” will display in the banner. Pressand hold the up and down arrowsto
silence. The alarm will return after two minutes until the block is resolved.
PUMP OPTIONS MENU
When a Multi-gas pump cartridge is attached to G7, a new menu option called
Pump options will appear in G7’s main menu, under Settings.
From this menu, you can customize the following:
Sample timer
From this menu, you can enable or disable the sample timer. Thesample timer
will count down from the Pump details screen and beep when one sample has been
pumped to and read by G7.
Hose length
Using the up, down and OK buttons, you can set the hose length you are using
on your device in meters or feet.
PUMP STATUS SCREEN
When a Multi-gas pump cartridge is attached to G7, a new secondary menu screen
called Pump status is available. To access this screen, press the down arrow
from G7’s Gas status screen.
From this screen, you can see the following:
-
Pump status
Off or on -
Hose length
An estimate of the hose attached to pump, set from the Pump options menu -
Flow rate
Measured in real time -
Sample timer
Available if enabled from the pump options menu. This is automatically calculated using your hose length.
LEARN MORE
Visit support.blacklinesafety.com for technical support and access to the following materials.
- Technical User Guide
- Video training tutorials
- Datasheets
- Certifications
- Accessory products
CUSTOMER CARE
Please contact us for warranty activation, technical support, advance replacement and account setup assistance.
- North America (24 hours)
- Toll Free : 1-877-869-7212
- support@blacklinesafety.com
- Europe/United Kingdom (8am-5pm GMT)
- +44 1787 222684
- eusupport@blacklinesafety.com
- International (24 hours)
- +1-403-451-0327
- support@blacklinesafety.com