Sefram DAS60 6 Channel Handheld Data User Manual

June 12, 2024
Sefram

DAS60 6 Channel Handheld Data

Product Information

Product Name: DAS60

Product Type: Data Acquisition Solution

User’s Manual: DAS60 User’s Manual

Publication Date: June 2022

Model Number: M0060001A/5

Manufacturer: SEFRAM

Manufacturer Contact Information:

  • Commercial Service: 04-77-59-01-01, Email:
    sales@sefram.com

  • After-sales Service: Email: sav@sefram.com

  • Technical Support: Email: support@sefram.com, Fax: +33 (0)4 77
    57 23 23, Website: www.sefram.com

Copyright: Sefram, 2022. All rights
reserved.

Warranty:

  • The instrument is guaranteed for two years for labor and parts
    against any manufacturing defect and/or functioning hazard.

  • The standard supplies provided with the appliance (cables,
    outlets, charger…), the consumables (batteries…), and the
    optional supplies (suitcases…) are guaranteed for 3 months
    against any manufacturing defect.

  • LCD screens and touchpads are guaranteed only for normal use.
    The guarantee does not cover wearing, accidental breaks, or any
    abnormal use.

  • The factory options integrated into the appliance are
    guaranteed for the same duration as the appliance itself.

Important Note: Any intervention by the user or
any third party without prior authorization by the company voids
the guarantee.

Product Usage Instructions

In Case of Failure:

In case of failure of your appliance, please contact the
technical support:

  • Technical Support: support@sefram.com

Note: Some advice and technical help are
available through the provided email address.

DATA ACQUISITION SOLUTION
DAS60 USER’S MANUAL

June 2022

M0060001A/5

DAS60 User’s Manual

Thank you for buying a SEFRAM recorder and for your trust in our company. The main goal of our various teams (research department, production, commercial, after-sales…) is to meet your requirements as much as possible by designing or updating high tech products.

Please read cautiously this user’s manual for an optimal use of your recorder.

Our teams are available to provide you any additional information:

04-77-59-01-01

Commercial service

e-mail: sales@sefram.com

After-sales service

e-mail: sav@sefram.com

Technical support

e-mail: support@sefram.com

Fax: +33 (0)4 77 57 23 23

Web: www.sefram.com

Copyright Sefram, 2022. All rights reserved. Any total or partial copy of this document needs the authorization of Sefram.
3 / 134

DAS60 User’s Manual
WARRANTY
Your instrument is guaranteed for two years for labor and parts against any manufacturing defect
and/or functioning hazard. This guarantee extends from the delivery date and ends 730 calendar days
later.
In case of guarantee contract, this will cancel or replace these guarantee conditions here above. The guarantee conditions by SEFRAM are available on the website www.sefram.com. The general guarantee conditions should prevail on the following conditions that they sum up. This guarantee does not cover the result of any abnormal use, handling mistake or mistake in the storage conditions outside the defined range. In case of application of the guarantee, the user shall return, at its own expenses, the relevant appliance
to our factory:
SEFRAM Instruments & Systèmes After-Sales Service
32, rue Edouard MARTEL BP 55
42009 SAINT-ETIENNE CEDEX 2
And add a description of the observed breakdown to the appliance. The standard supplies provided with the appliance (cables, outlets, charger…), the consumables (batteries …) and the optional supplies (suitcases…) are guaranteed for 3 months against any manufacturing defect.
Such items as a LCD screen or a touchpad are guaranteed only for a normal use. The guarantee does not cover wearing, accidental breaks or consecutive to a shock or any abnormal use.
The factory options integrated to the appliance are guaranteed for the same duration as the appliance itself.
In case of replacement or repair of the product, the remaining guarantee duration shall be: – The remaining duration of the guarantee if the appliance is still under guarantee – If the guarantee duration is less than 90 days, the replaced part is guaranteed for 90 days
Any replacement part becomes the property of the user and the exchanged parts become the property of SEFRAM. In case of intervention by an insurance company, the product becomes the property of the insurance company upon its exclusive request. Else, it shall remain property of the user. The guarantee covers exclusively the materials manufactured and provided by SEFRAM.
Any intervention by the user or any third party without prior authorization by the company voids the guarantee.
4 / 134

DAS60 User’s Manual
The user shall be responsible for the return of its appliance to our site. Hence, it shall provide for a conditioning that shall correctly protect the appliance while shipping. It shall subscribe, at its own expenses, any insurance required for the transport. The SEFRAM company reserves the right to refuse any product wrongly conditioned and not to take in charge any break consecutive to the transport. Particular case of the battery: There is a Li- ion battery as a standard equipment of this appliance. It shall not be transported outside the appliance. In no case shall the user replace it. Its replacement in the factory is necessary to check the charge system and the protective securities.
What to do in case of malfunction? In case of malfunction or for any advice for use, please contact the technical support by SEFRAM Instruments & Systèmes: A technician shall answer you and give you any information required to solve your problem.
What to do in case of failure? In case of failure of your appliance, please contact the technical support:
Some advice! Some technical help!
SEFRAM Instruments & Systèmes commits itself to help you on the phone about the use of your appliance. Please call or Technical Support:

Or e-mail: We thank you for your trust.

support@sefram.com

5 / 134

DAS60 User’s Manual METROLOGY
The meteorological conditions of your measurement instrument are defined in the specifications of this notice. Climate and environmental conditions restrict the specifications of your recorder. SEFRAM checks the characteristics of each appliance one by one on an automatic bench during its manufacture. The adjustment and control are guaranteed under conditions of the ISO9001 certification by facilities in connection with the COFRAC (or equivalent in the context of ILAC reciprocity). The specified characteristics are considered stable for a period of 12 months from the first use under normal conditions of use. We recommend a check after 12 months and max. 24 months of use, then every 12 months after 24 months. For any check of the characteristics, the following average climate conditions shall be maintained (23°C+3°C ­ 50(+20)%RH). The recorder should have been working for 0,5 hour before check. We recommend that you have this control made by our after-sales service (Service Après-Vente) for the best service and preservation of the measuring quality of your instrument. When a recorder returns to SEFRAM, maximum service is provided with internal updating according to the required adjustments and software updates. In case of shift in the characteristics, your instrument shall be adjusted to recover its original characteristics.
6 / 134

DAS60 User’s Manual
TABLE OF CONTENT
TABLE OF CONTENT ………………………………………………………………………………………………………..7 1. IMPORTANT INFORMATION……………………………………………………………………………………11
1.1. Specific cautions……………………………………………………………………………………………………11 1.2. Security instructions ………………………………………………………………………………………………11 1.3. Conformity and limits of the device …………………………………………………………………………12 2. PRESENTATION ……………………………………………………………………………………………………….13 2.1. General…………………………………………………………………………………………………………………13 2.2. Description……………………………………………………………………………………………………………14 2.3. Touch screen…………………………………………………………………………………………………………15 2.4. MAIN MENU……………………………………………………………………………………………………….19 2.5. Update of the internal software ……………………………………………………………………………….20 3. SETUP and PRECAUTIONS of USE ……………………………………………………………………………21 3.1. Loading of the recording paper ……………………………………………………………………………….21 3.2. Precautions for storage of the recordings ………………………………………………………………….21 3.3. Setup ……………………………………………………………………………………………………………………22 3.4. Configuration at power- up………………………………………………………………………………………22 3.5. CONNECTION TO THE MEASUREMENT NETWORK…………………………………………23 3.6. Routine maintenance ……………………………………………………………………………………………..24 3.7. Battery …………………………………………………………………………………………………………………24 3.8. To recharge the battery inside the device: …………………………………………………………………25 3.9. Offset calibration …………………………………………………………………………………………………..25 3.10. Default calibration ……………………………………………………………………………………………..26 3.11. Screen brightness ……………………………………………………………………………………………….26 3.12. Locking of the recorder……………………………………………………………………………………….26 3.13. User mode …………………………………………………………………………………………………………26 4. USE …………………………………………………………………………………………………………………………..27 4.1. Description of the various modes …………………………………………………………………………….27 4.2. Setup menu …………………………………………………………………………………………………………..28 4.3. Channels Menu ……………………………………………………………………………………………………..29 4.4. Validation of the channels ………………………………………………………………………………………36

4.5. « Direct display » key

………………………………………………………………………………….. 37

4.6. « Trigger » menu …………………………………………………………………………………………………..41

4.7. « Memory output » key ………………………………………………………………………………………….42

4.8. « Record » key ………………………………………………………………………………………………………43

7 / 134

DAS60 User’s Manual
4.9. Screen copy…………………………………………………………………………………………………………..44 5. TRIGGERS ………………………………………………………………………………………………………………..45
5.1. Analog channel trigger …………………………………………………………………………………………..46 6. MATHEMATICAL CALCULATIONS…………………………………………………………………………50
6.1. Definitions ……………………………………………………………………………………………………………50 6.2. Types of calculations ……………………………………………………………………………………………..51 7. DIRECT MODE………………………………………………………………………………………………………….54 7.1. Launching the printing……………………………………………………………………………………………54 7.2. Configuration of print…………………………………………………………………………………………….56 7.3. Relaunching the printing ………………………………………………………………………………………..57 7.4. Writing information……………………………………………………………………………………………….57 8. MEMORY MODE ………………………………………………………………………………………………………58 8.1. Setup and start of the data acquisition ………………………………………………………………………58 8.2. Sampling period…………………………………………………………………………………………………….60 8.3. Internal memory, blocks …………………………………………………………………………………………60 8.4. Trigger position …………………………………………………………………………………………………….60 8.5. Double Trigger mode……………………………………………………………………………………………..61 8.6. Recording……………………………………………………………………………………………………………..62 8.7. Memory output ……………………………………………………………………………………………………..64 9. TEMPLATE MODE ……………………………………………………………………………………………………65 9.1. Setup and launching of the data acquisition ………………………………………………………………65 9.2. Creation of the template …………………………………………………………………………………………67 10. FILE MODE…………………………………………………………………………………………………………….69 10.1. Setup and triggering of the data acquisition …………………………………………………………..69 10.2. Restrictions ……………………………………………………………………………………………………….70 11. POWER ANALYSIS ………………………………………………………………………………………………..71 11.1. General ……………………………………………………………………………………………………………..71 11.2. Power analysis mode installation………………………………………………………………………….72 11.3. Power analysis setup …………………………………………………………………………………………..74 11.4. Real time display of the signals ……………………………………………………………………………75 11.5. Acquisition parameters ……………………………………………………………………………………….77 12.6. Record …………………………………………………………………………………………………………………..78 11.7. Measurement process………………………………………………………………………………………….79 12. FILE MANAGEMENT……………………………………………………………………………………………..80 12.1. General ……………………………………………………………………………………………………………..80 12.2. Management of the configuration files ………………………………………………………………….81 12.3. Operating software……………………………………………………………………………………………..85
8 / 134

DAS60 User’s Manual
13. INPUTS / OUTPUTS………………………………………………………………………………………………..88 13.1. Additional Input/Output Connectors……………………………………………………………………..88 13.2. Logic inputs ………………………………………………………………………………………………………89 13.3. Alarm outputs…………………………………………………………………………………………………….89 13.4. Power supply output …………………………………………………………………………………………..90 13.5. Exvoltage box input/output interface…………………………………………………………………….91
14. ETHERNET INTERFACE ………………………………………………………………………………………..93 14.1. Ethernet Interface……………………………………………………………………………………………….93 14.2. WIFI interface ……………………………………………………………………………………………………94 14.3. FTP transfer ………………………………………………………………………………………………………95 14.4. Managing with VNC:………………………………………………………………………………………….96 14.5. Programming language ……………………………………………………………………………………….97 14.6. Standard instructions…………………………………………………………………………………………..99 14.7. Indication of the status of the device …………………………………………………………………..100 14.8. Dictionary of programming ……………………………………………………………………………….106 14.9. Error messages …………………………………………………………………………………………………119
15. TECHNICAL SPECIFICATIONS ……………………………………………………………………………120 15.1. Isolated inputs ………………………………………………………………………………………………….120 15.2. Additional inputs / outputs…………………………………………………………………………………123 15.3. Inputs with platinum resistance ………………………………………………………………………….123 15.4. Triggers …………………………………………………………………………………………………………..124 15.5. Memory data acquisition……………………………………………………………………………………124 15.6. Data acquisition to file ………………………………………………………………………………………124 15.7. Power analysis …………………………………………………………………………………………………125 15.8. Printer option …………………………………………………………………………………………………..126 15.9. Communication interface…………………………………………………………………………………..126 15.10. Display ……………………………………………………………………………………………………………127 15.11. Environment conditions …………………………………………………………………………………….127 15.12. Power supply – battery ………………………………………………………………………………………127 15.13. Electromagnetic compatibility, Security………………………………………………………………128 15.14. Accessories ……………………………………………………………………………………………………..130
16. APPENDS ……………………………………………………………………………………………………………..131 16.1. Information about the input calibers ……………………………………………………………………131 16.2. Measurement accuracy for thermocouple …………………………………………………………….132 16.3. Instant measurement accuracy as a function of the filters ………………………………………134 16.4. Accuracy class ­ class index………………………………………………………………………………134
9 / 134

DAS60 User’s Manual 10 / 134

DAS60 User’s Manual
1. IMPORTANT INFORMATION
Please read cautiously the following instructions before using your recorder.
1.1. Specific cautions
Do not use the product for any other use than planned. Use normalized cords for the connection of the device to the measurement points. To prevent any risk of electric shock, never plug or unplug the measurement cords when linked to the mains. Never use in a wet environment. Never use in an explosive atmosphere. In case of malfunction or for the maintenance of the device, only a qualified personal shall be allowed to work on it. In such a case, it is necessary to use Sefram spare parts. Never cover the device: there is live voltage.
1.2. Security instructions
For correct use of the device, it is necessary that the users abide by the security instructions and the instructions of use described in this manual. Specific warnings are provided all along this manual. If required, there are warning symbols on the device Symbols and definitions Symbols in this handbook:
Warning: potential danger for the user
Attention: potential danger for the device and/or the connected equipment
Remark: important information.
11 / 134

Symbols on the device:

DAS60 User’s Manual

Danger (high voltage): immediate corporal danger

Attention: See the handbook. Risk of damage to the material connected to the instrument or to the instrument itself

Ground: accessible parts linked to the ground of the device.

1.3. Conformity and limits of the device
The DAS60 recorder is in conformity with CEI 61010-1 (2001-02).
It includes high-potting (IR) with two primary insulation levels (IP). Thanks to them, the channels 1 to 6 are insulated from each other and from the carter. This architecture allows independent floating measurements within the limits of the admissible maximum voltages. See chapter “Technical specifications”.

IR
CH1
Entrées
IR
CH2
Entrées

Opto-coupler

DC/DC

IP

IP

IP

DC/DC

DC/DC

Opto-coupler

accessible parts:
USB Ethernet Logical inputs Pt100 Alarms Power Carter

IR

IR

IP: Main or additional insulation IR: Reinforced insulation

Casing of the DAS60

Attention: Never apply a voltage higher than the maximum admissible voltage between the channels and from the ground of the carter.

The use of the device at frequencies > 100Hz and of the voltages> 50V, requires imperatively the connection of the earthing protection or power supply for safety reasons.
12 / 134

2. PRESENTATION

DAS60 User’s Manual

2.1. General
DAS60 is a programmable recorder that make it possible to measure and record voltages, currents, temperatures, etc. on 6 channels and on 16 logical channels. You can also record from Pt100 or Pt1000 probes on dedicated channels.
An integrated thermal printer option makes it possible to print signals in real time.
They can work in several modes: Direct, for paper printing (option) Memory, for acquisition on internal quick memory File, for acquisition on internal SD card Template, for acquisition on a pre-recorded template Analysis of network
The “operator ­ recorder” dialog is made easier thanks to an intuitive interface on a wide touch screen. The measurement parameters are easy to program.

13 / 134

2.2. Description

DAS60 User’s Manual

Connection:
1- A RJ45 connector for the ETHERNET 10/100BaseT interface 2- two USB connectors for (female) USB stick 3- a SUB-D 25-pin connector for the 16 logic inputs and the alarm outputs 4- 6 inputs through pin plugs according to the device 5- 2-way Pt100 (screw terminal block) 6- Power/Battery charger (on the upper side) 7- Power on button and light 8- Battery charging LED 9- Ground connector
Isolated inputs have 2 security terminals on each input: · 1 red terminal: input « + » · 1 black terminal: input « – »

4

7

86

3

9

5

2

1

14 / 134

2.3. Touch screen
Description of the title bar

DAS60 User’s Manual

1

2

34 5

6 7

8

9

Various zones on the screen:
1- Current mode (direct mode, memory mode…) 2- Status of acquisition (pending acquisition, stop) 3- Status of the battery 4- Direct access to display 5- Access to the display of the memory 6- Access to the recording 7- Screenshot button 8- Date and time 9- Access to the main menu
Note: By pressing on the date and time, you can access to the menu to change this information.

15 / 134

DAS60 User’s Manual

Description of the channel validation window

10

11

12

7

1 2

3 4 8

9

6b

6

5

1- Change of the color of the channel (by pressing this button, you can access pre-defined

colors or design customary paint mixes:

2- Change the thickness of the line for the selected channel. You can also change this

thickness with the

(thinner) and

(thicker) keys

3- Checkbox to validate a threshold

4- Current threshold value or change of the threshold value

5- Information about the channel (caliber, thresholds, offset…)

6- Logic inputs : real time status of the logic inputs

6b- Logic inputs : validation of the logic inputs

7- Access to the parameters of the various channels

8- Temperature channels through Pt100 or Pt1000 type sensors

9- Function channels (they allow calculations between two channels)

10- Checkbox for validation of the channels

11- Parameter zone: you can change the name of the parameters

12- Key to change of parameter and display the current value

Note: By pressing the logic inputs, you pop up a window with all logic channels.

16 / 134

DAS60 User’s Manual
Some information is available on this window: – when clicking on a logic channel, the relevant pin on the connector turns red – when pressing the keyboard, you can change the name of the channel.

2.3.3. Description of the power analysis window

10 11

12

13

1

2 9
3
4
5 8

6 7

1- Setup of the network analysis 2- Changing the color of the various measurements (voltage, current, power…)
17 / 134

DAS60 User’s Manual
3- Measurement mode of the current (clamp, shunt, transformer) 4- Parameters of the current measurement (conversion from the measured value and the
value sent to the recorder) 5- Nominal current in the phase conductor 6- Wiring scheme 7- Logic channels (validation and parameter setting) 8- Display of the measured values 9- Phase diagram 10- Type of network 11- Parameters of voltage 12- Value of the nominal voltage 13- Frequency of the network to analyze
18 / 134

DAS60 User’s Manual
2.4. MAIN MENU

9 10 11 12 13 14

15 16 17

1
2
3
4
5
6
8 7
1- Direct mode key (direct printing of the curve) 2- Memory mode key 3- File mode key 4- Template mode key 5- Network mode key 6- Start screen to display the serial number, the software version… 7- Recording 8- Replay of the memory 9- Digital display of the values of the various inputs 10- Display of the F(t) curve 11- Display of the XY curve 12- Configuration of the logic channels 13- Adjustment of the triggers 14- Setup of the device (language, date and time, alarm outputs, update of the internal
software, network connection, recovery of the configuration…) 15- Access to the table of channel configuration:

19 / 134

DAS60 User’s Manual

Note: By pressing various parameters, you can change them. In addition, by pressing the name of the channel (A1, A2…), you directly access the parameter setting page of the channel
16- Configuration of the function channels (calculations between several channels) 17- Activation or deactivation menu of the various channels.
2.5. Update of the internal software
The internal software is regularly updated with the latest evolutions. These updates are available on our website.
http://www.sefram.com/wwwfr/F_D_SOFTWARE.asp
To update it, copy the file you receive on an USB stick. Plug it on to the USB connector at the rear of the device.

Press

, then press

to access to the update.

To launch the update process, press The internal software automatically copies the files required for the new version. Stop and restart the device at the end of the update process.

20 / 134

DAS60 User’s Manual
3. SETUP AND PRECAUTIONS OF USE
3.1. Loading of the recording paper
NOTE: Only one face of the paper is heat-sensitive. Reversing the paper at loading will result in the absence of any printing out. Procedure:
· Lay the device on the right side · Lift the extractor of the roller on the lid · Install the paper into the receptacle (“no de ref” must be on the side of the plugs) · Introduce the paper into the slot of the lid · Close the lid by firmly pressing it on both sides.
A bad paper feed could damage the motor and the thermal head. You should take great care when putting it in place. The recorder shall always receive paper, because the thermal head may be damaged in case of extended direct contact with the roller.
3.2. Precautions for storage of the recordings
In order to maintain the quality of the paper recordings, it is advisable to abide by precautions of handling: · never use laminated sleeves · store away from light, in a dry and cool place. Cardboard sleeves are advisable.
21 / 134

DAS60 User’s Manual
3.3. Setup
You turn on the recorder by pressing the button at the top of the device.
When the device is on, the ON/OFF button is lit blue.
After launching the software, the recorder displays a homepage that specifies the hardware version, then switches automatically to the « Direct display » mode (oscilloscope).

You can display this homepage any time in the main menu by pressing

.

To turn off the device, you only have to press the ON/OFF button twice.

If the device does not stop correctly, you must press the ON/OFF button 5 seconds; in this case, the configuration won’t be saved.

3.4. Configuration at power-up
At power-up, the devices start with the latest configuration when they were shut down.

If the configuration has changed at power-up, contact the after-sales service.

If the initial setup is wrong, you can reload the default configuration.

Go to the main menu. Press

, then

22 / 134

DAS60 User’s Manual
3.5. CONNECTION TO THE MEASUREMENT NETWORK
3.5.2. Voltage measurement The voltage is measured between the red and black terminals of the inputs with wires and male security pin plugs (according to CEI 1010).
3.5.3. Temperature measurement with a thermocouple The voltage resulting from thermocouple effect must be measured between the red and black terminals of the relevant input. To provide right measure values, connect the two ends of the thermocouple cord directly onto the quick fastening socket. Plug the two pin plugs onto the selected input with respect of the polarity.
Do not use welding male pin plugs: the welding would alter the thermocouple effect.

3.5.4. Temperature measurement with Pt100/Pt1000
The Pt100 probe must be connected on the « I+ » and « – » terminals. The voltage from the Pt100 must be measured between the « + » and « -» terminals with one of the two following mountings: 2 or 3 wires. The 3-wire mountings make the measurement independent from the resistance of the line.
Wiring:

2 wires

3 wires

3.5.5. .Current measurement
You can make current measurements by shunt between the red and black terminals of the relevant input.
In this case, select the “current” type among the parameters of the relevant channel. Connect the measurement wires to the terminals of the shunt.
The results are directly displayed in Amperes or milliamperes according to the caliber of the selected channel.

23 / 134

DAS60 User’s Manual
3.5.6. Connection of the ground
· If the source of the signal you have to record has low internal impedance, you shall use twisted wires. In case of high impedance, you shall use shielded wires.
· When gathering grounds from the various items on the measurement line, it is good to check that there is no voltage difference between them, in order to prevent any shortcut. If any doubt, make a measurement with a voltmeter on a low resistance (i.e. 1k) between the terminals.
3.6. Routine maintenance
Maintenance only consists in cleaning outside the device. Any other operation requires a qualified personal.
Unplug the device before any intervention.
Never let water flow through the device to prevent any electrical discharge. Periodically clean the recorder with the following the instructions:
– use water and soap to clean the front and rear boards – never use any product that contains petrol, benzine or alcohol: they would damage the
screen printings – wipe with a soft lint-free cloth – use some antistatic product to cleanse the screen.
3.7. Battery
The device is equipped with a lithium-ion battery (Li-ion). It is sent with charged battery. However, if the device has remained unused for more than one month, check its loading status and reload it if required.
Attention: Any intervention on the battery requires dismantling of the device and shall be made by a SEFRAM technician. Only use batteries provided by SEFRAM.
Security advice: · Never heat up or throw the battery pack into the fire · Never shunt parts of the battery: risk of explosion! · Do not bore the battery · Do not dismantle the battery pack · Do not reverse the polarities of the battery · This battery pack includes a protective item that shall not be damaged or removed · Do not store the pack in a place exposed to heat 24 / 134

DAS60 User’s Manual
· Do not damage the protective sheath of the pack · Do not store the device in a vehicle exposed to sunbeams.
The battery should last 200 charge-discharge cycles or 2 years.
Advice to extend the life of the battery: · Do not discharge too much · Do not store the batteries too long without using them · Store the battery at around 40% charge · Do not fully charge or discharge the battery before storage.
When the battery is practically discharged, the device closes all open files, stops correctly the software, then shuts up by itself.

Attention: When the charger is connected to the device, the metal frame is grounded.
3.8. To recharge the battery inside the device:
Connect the provided external power supply on the jack plug of the device. Connect the mains. The internal charger starts loading the battery; the green light of battery charge lights up. You can charge the device when on or off. When the battery is loaded, the light turns off automatically.
External power supply
The device can be powered by an external direct voltage power supply. It works under 15V (5A). The charger block provided when you bought the device can also work as an external power supply.

3.9. Offset calibration

You can easily calibrate the recorder inputs for the voltage and thermocouple offsets. This will remove any offsets present on different channel. To do so: · Let the device work for 20 minutes (ambient temperature 20-25°C) · Link the « + » to « – » terminals on each input (resp. the red and black terminals for the isolated
inputs) · Validate all ‘ON’ channels

· Go to the homepage of the recorder and press

.

· Select

then

.

· Press

to validate the calibration; it will last ca. 5 minutes. 25 / 134

DAS60 User’s Manual

3.10. Default calibration
You can restore the default factory calibration to correct any possible mistake in the calibration coefficients:
Go to page « Electric calibration » (see the previous section)

Press

and validate to restore the default factory calibration coefficients

and give the date of initial plant start.

3.11. Screen brightness
You can settle the screen brightness. You can access this parameter at page

then

and then

.

3.12. Locking of the recorder
You can lock the recorder to prevent any modification. You can access this parameter at page

then

and then

.

To unlock the recorder, you only have to press 5 times the key at the top right of the screen.

3.13. User mode
You can activate or disactivate the user mode. The user mode allows you to lock the configuration of the device while keeping all other menus available. The password is: Sefram.

You can access this parameter at page

then

and

then

.

26 / 134

4. USE

DAS60 User’s Manual

4.1. Description of the various modes

Selection of the functioning mode of the recorder:

DIRECT mode: · real time print out of the measured signals (Option) · use: immediate paper printing, slow protracted data acquisition · possibilities: complex triggers of the drawing, action after the end of the drawing,
simultaneous record on file
MEMORY mode: · quick acquisition into internal memory of the measured signals · use: quick short data acquisition (transitory) · possibilities: complex triggers of the data acquisition, action after the end of the data
acquisition, simultaneous record on file
TEMPLATE mode: · quick acquisition into internal memory of the measured signals · use: quick short data acquisition (transitory) for capture of non repetitive events · possibilities: complex triggers, particularly when outrunning a pre-recorded template from a
channel, action after the end of the data acquisition, simultaneous record on file
FILE mode: · quick acquisition into internal memory of the measured signals · use: quick protracted data acquisition (only restricted by the size of the saving means) · possibilities: complex triggers, action after the end of the data acquisition, acquisition of
very big amounts of data.
NETWORK ANALYSIS mode: · single-phase network analysis · three-phase analysis, Aaron mode, Star and Delta. (two-wattmeter method) . · Data acquisition on internal memory.

27 / 134

DAS60 User’s Manual
4.2. Setup menu

General configuration of the device, monitoring of the alarm outputs, network address TCP/IP, calibration of the channels, update of the internal software.

12

34

11 5
10

9

6

8

7

1- Language: selection of the language of the device 2- Screen intensity: increase/diminution of the backlighting of the LCD screen, adjustment of

the delay with the keys

(decrease) and

(increase)

3- Default setup: initialization of the device in default configuration

4- Load on disk: loading a configuration recorded on the internal storage means or on an USB

stick

5- Save on disk: saving the configuration on the internal storage or on an USB stick

ATTENTION: the current configuration shall be lost

6- Network: Change of the IP parameters and of the network options 7- Alarm A, B, C or D: use of the alarm outputs A, B, C or D (0-5V outputs)
· None: no condition monitors the contact; it remains always open · Trigger: control by a combination of analog or logic channels on several thresholds (see
chapter Triggers) · Paper error: control by paper shortage or opening of the door of the printing block
(Option). 8- Software update: update of the internal software (see chapter Presentation) 9- Additional option: change of various parameters:
· Bip keyboard: activation or deactivation of the Bip keyboard · Electric calibration: calibration of the channel offsets, reset to factory offsets 10- Graphic options: max position of the bargraph: scrolling direction of the bargraph of each channel on screen:
28 / 134

DAS60 User’s Manual · Right: max. value of the right channel · Left: max. value of the left channel 11- Date change: changing the date and time of the device
4.3. Channels Menu
By pressing the menu « channels », you access the table of all available analog and temperature channels of your recorder.
This table gives you a quick insight of the various adjustments of the channels. You can change these adjustments directly in this table by pressing the parameter you want to update. You can access the setting of parameters channel by channel by pressing the name of the channel at the top of the table (A1, A2…) or by pressing on the bargraph of the channel (right side of the screen).
29 / 134

Analog channels
14 13 12

DAS60 User’s Manual

1234

56

7

8

9 10

11

1- Number of the channel 2- Name of the channel: give a name to the channel (max. 26 characters) 3- Type: selection of the type of measurement made on the input
· voltage, current, frequency, thermocouple or counter · direct, RMS, V/t or V.dt · value of the shunt for current measure (sensor 4-20mA automatic) · selection of the type in thermocouple, compensation, unit
4- Type of external sensor 5- Filter: positioning of a filter on the input
· 10kHz, 1kHz, 100Hz, 10Hz for analog filters · 1Hz, 10s, 100s or 1000s for the digital filters (according to the type of signal)
6- Change of the color of the channel

7- Change of the thickness of the signal with the keys

(decrease) and

(increase)

8- Validation and adjustment of the trigger thresholds 1 and 2

Note: A warning message pops up when the programmed analog thresholds are out of the measurement range.

9- Function: makes it possible to append a mathematical calculation function to the relevant channel
30 / 134

DAS60 User’s Manual
· None: no function · Change of unit: changes the unit of the measures made on the channel; you can
program a couple of dots X1, Y1 and X2, Y2 for scaling · Calculation: available mathematical functions, associated parameters and unit o aX+b : identical to the change of units, but instead of giving a couple of dots, you give the intercept (b) and the slope (a). o a |X|+b: absolute value o a X2+b : square o a ln(X)+b : Naperian logarithm o a Sqrt (X+c)+b : square root o aExp(cx) +b : natural exponential o a (1/X) +b : reverse. 10- Copy to channel: makes it possible to copy the adjustment of a given channel onto one or several other channels 11- Display window for the thresholds, caliber, zero offset… 12- Adjustment of the position: position of the zero on screen or on the paper from -100 to 100%: i.e. in RMS mode, it is interesting to have zero at the minimum value (-100%); you will get the max. value = caliber 13- Adjustment of zero: Zero (or center or offset) is the central value of the measurement. Note: When using a mathematical function or a scale change, zero is the zero of the requested unit. 14- Adjustment of the caliber: the caliber is the measurement range corresponding to the full width of the screen, where the channel is plotted.
31 / 134

Type of measurement V/t :

DAS60 User’s Manual

1

2
3
1- V/ t : Put on the « derivative » measurement mode. Every period t, the recorder divide the voltage variation by t and get a value in Volt/second.
2- V : Select the measurement range of the input signal from +-500µV to +-500V. The range must be larger than the input signal. As example for a signal varying from -7V to +7V select +-10V
3- t : Select the measurement period from 200µs to 10min It’s the time between 2 derivative calculations. Take a period compatible with the signal speed: a period too short give a derivative with noise.
Example: Channel 1: triangle signal, -5V to + 5V, period 5s Channel 2: use measurement type V/t with the same signal
V +-10V t 50ms Span 20V/s
The signal (Channel 1) is varying linearly at +4V/s and then at -4V/s The derivative (Channel 2) give 4V/s and -4V/s

32 / 134

Type of measurement V.dt:

DAS60 User’s Manual

1
2
3
1- V.dt :Put on the measurement type « integral » If U is the voltage measurement, every period dt the recorder calculate Udt and add this to the previous calculation: this sum or integral is expressed in Vs (Voltsecond). The integral is reset by the user and/or every start of measurement.
2- V : Select the measurement range of the input signal from +-500µV to +-500V. The range must be larger than the input signal.
3- dt :Calculation period from 200µs to 10min.
Example: Channel 1: square signal, period 10s, 2s at 5V, 8s at 0V . Channel 2: use measurement type V.dt with the same signal
V +-10V dt 200 µs Span 200Vs
The signal « integral » (channel 2) begin at 0Vs and then increase of 10Vs (integration of 5V during 2s) every signal (Channel 1) period.

33 / 134

DAS60 User’s Manual Additional functions between channels There are function channels on this device. You can activate up to 4 function channels (FA, FB, FC, FD). These channels make it possible to make calculations between channels or calculations of average values… For example, if you want to calculate a value of power, select the parameters of the A function channel
You get this page above. Then, you make the calculation you want by pressing the formula bar and create the calculation you want in the following window:

For the calculation of power, you will have: a(x*y) + b with:

x = channel A1

operator = *

y = channel A2

coefficient a = 1

unit W coefficient b = 0
34 / 134

DAS60 User’s Manual
Logic channels
The 16 logic channels can be recorded simultaneously with the analog channels. To set the parameters of the logic channels, you can press the « logic channels » menu on the main window or directly press the logic channels part at the bottom of each configuration window:

You access the configuration page of the logic channels: 1

23

1- Selection of the name of each channel on screen 2- Selection of the color of each channel on screen 3- Logic validity: validation of the data acquisition and plotting of the logic channels
35 / 134

DAS60 User’s Manual
4.4. Validation of the channels
You can validate or not the channels to record or print out in the various configuration menus of the recorder. In order to validate or not the record or the printing of a channel, you must check (to validate) or uncheck (not to validate) the checkbox right to the bargraph of the channel.
36 / 134

DAS60 User’s Manual
4.5. « Direct display » key
Real time on-screen display of the measurements
1 2
3
4
5
6
7
1- Screen configuration: configuration of the display of measurements on screen. Selection of the mode of graphic display: · F(t) · XY · Digital measures (full screen) · Full screen to display the reticule on the whole screen (mode f(t)) · Colors to customize the display (background, reticule, cursor) · Graphic option · Custom reticule
2- Channels validation: Validation of the channels to display If the logic channels are actives, you shall select whether the diagram of the logic channels should be distinct from the analog channels or not (see chapter 4.3.3).
3- Freeze screen: freeze the measurements on screen to make measurements by cursors, make calculations, save or print the measures on screen (1000 dots) in F(t) mode. You get access to: · Relaunch ­ relaunches the sweep · Time Cursors displays the vertical cursors (2) to make measurements on the screen; move the cursor by selecting it with your fingers or the mouse. · Voltages Cursors displays the horizontal cursors (2) to make amplitude measurements on the screen; proceed like for the time cursors to move them. You can also change the caliber / zero to expand and move your measurement on screen.
4- Autoset: This function allows you to have automatically the caliber of the active channels and the time base compatible with the frequency of the signal.
5- Time base: Allows you to settle the time base.
37 / 134

DAS60 User’s Manual 6- Math: Makes it possible to display values of mathematical calculations on screen. 7- Access to the channel configuration page:
1 2
3
4
5 6 7 8 9
1- Screen setup: configuration of display of measurements on screen · Selection of the graphic display mode: · F(t) · XY · Digital measures (full screen) · Full screen to display the reticule on the whole screen (mode f(t)) · Colors to customize the display (background, reticule, cursor). · Graphic option · Custom reticle
2- Channels validation: Allows you to activate or disactivate the channels. 3- Math: Displays values of mathematical calculations on screen. 4- Selection of the channel to modify (with the « + », « – » and central keys). 5- Modification of the range of the selected channel (with the « + », « – » and central keys). 6- Modification of the position of zero (with the « + », « – » and central keys). 7- Modification of the position of the selected channel (with the « + », « – » and central keys). 8- Time base: Allows you to adjust the time base. 9- Back to the previous page.
38 / 134

XY display

DAS60 User’s Manual

The XY display mode makes it possible to display the validated channels in real time one against the other. One of the channels defines the horizontal axis; the other channels provide dots on the vertical axis.

You can access the XY mode via the main menu by pressing
1 2 3 4 5

1- Screen setup: configuration of the display of measurements on screen. Selection of the mode of graphic display: · F(t) · XY · Digital measures (full screen) · Full screen to display the reticule on the whole screen (mode f(t)) · Colors to customize the display (background, reticule, cursor) · Graphic option · Custom reticule
2- Plot: starts or stops of the plot 3- Delete: erases the screen 4- X channel: selection of the channel on the horizontal axis (sweep) 5- Y channel: the Y channels are the valid channels
39 / 134

DAS60 User’s Manual Digital display The Digital display mode displays the digital values of all channels in real time on screen. When pressing one of the channels, you access directly to the parameter page of the corresponding channel.
40 / 134

DAS60 User’s Manual
4.6. « Trigger » menu
Programming the start and stop conditions for the paper print in Direct mode, and the data acquisition conditions for the channels in Memory, File and Template modes. Selection of the actions after data acquisition or plot and validation of the save in real time.
The program of triggers is different according to the pending mode (Direct, Memory, File or Template). See the chapter of the pending mode for more details.
41 / 134

DAS60 User’s Manual
4.7. « Memory output » key
Display on screen of the data acquisitions available from the internal memory or from files of the internal memory or an USB stick.
This function includes the same commands as the « Direct display » function. The « Read block or file » command allows you to select the memory block (zone inside the internal memory) or the file to display. When pressing this key, the following window pops up:
· Block number: number of the memory block to display · Read file: selection of the file to display · Write data to file: save the current display into a file
When the size of the data acquisition to display is big, it may take time to recover and display the dots. The display is made in 2 phases:
42 / 134

DAS60 User’s Manual
· A quick phase to display the frame of the data acquisition: some dots may not appear · A slower phase where all data acquisition dots are displayed: the advancement percentage is
displayed at the bottom of the screen.
4.8. « Record » key

This key has various effects according to the current mode of the device.
1- Direct mode: launches the paper printing if the trigger is in manual Start; else, the device waits for the start condition.
2- Memory mode: launches the data acquisition into internal memory; the device waits for the start condition
3- Template mode: launches the data acquisition into internal memory; the device waits for the start condition
4- File mode: launches the data acquisition into file; the device waits for the start condition
5- Network analysis mode: real time data acquisition directly into the internal memory (there is no trigger)
In all cases, to force the record to start or stop when the conditions registered in the recorder are not

met, you only have to press

(to force the record) and

(to force it to stop).

In Memory, Template or File mode, the device switches automatically to display of the current data acquisition.

the top of the screen, with:

shows up at

· the pending block number, if relevant · the current sampling rate · the status of the data acquisition (stand-by, pending xx%, …) · the opening of a save file, if relevant · a bargraph specifying the percentage of data acquisition achieved and the percentage of data
acquisition displayed

43 / 134

DAS60 User’s Manual
4.9. Screen copy
You can create a file copying the content of the LCD screen:

You only have to press

at the top right of the screen.

The file will be created either onto the USB stick, if there is one, or into the internal memory. The name of the file will be bmpxxxxx.bmp (incremental name). In the internal memory, files are saved in the « FolderBMP» directory.

You can either copy this directory into an USB stick, or delete it. You can also use a ftp link to recover these files.

44 / 134

DAS60 User’s Manual
5. TRIGGERS
This chapter describes all possible triggers of the device. They are used by:
– the « configuration » menu, with the A and B alarms – the « triggers » menu, with the start and stop parameters of the recording. Edge/Level trigger · Edge trigger: you need a change of status
o Example: Channel A1, rising edge, threshold = 0V: triggers only if the signal status changes from negative to positive
· Level trigger: no need to pass the threshold o Example: Channel A1, high level, threshold = 0V: triggers only if the signal is positive
· Alarms only have level triggers. Analog / Logic channels: trigger only from analogic or logic channels Single / Multiple threshold:
· triggering from only one threshold, or · complex triggering from several channels and several thresholds; see description below. Such complex triggering is only possible with analogic channels. You can setup these triggers thanks to the « trigger » menu in the homepage.
45 / 134

DAS60 User’s Manual
5.1. Analog channel trigger
By pressing Start or Stop, you open the setup window for the trigger of the start or stop of recording. After selection of the trigger on Analog Channels, the following line on screen allows you to set the trigger condition. It depends on the selection of single or multiple threshold.

Single threshold

1

234

5

1- Channel: selection of the channel, on which the trigger threshold is applied 2- Threshold 1 / Threshold 2: selection of the threshold to settle; each channel is tested with
reference with 2 thresholds, i.e. you can program a start condition on the channel A1 and the threshold 1, and a stop condition on this same channel A1 and the threshold 2. 3- Higher / Lower: trigger when the channel value is more or less than the set threshold. 4- Threshold value: threshold value selected in real value (taking the current unit and scale of the configuration of the selected channel into account) 5- Edge: Selection of the active edge of the channel in reference to the threshold.

46 / 134

DAS60 User’s Manual

Multiple thresholds

After selection of a trigger on a set of logic channels, the window allows you to set the trigger under

several conditions.

Then, by pressing the various thresholds in front of their respective channels, you open the

threshold parameter window.

1

2

3

1- One of the thresholds (or): the first achieved condition activates the trigger 2- All thresholds (and): all conditions must be simultaneously achieved to validate the trigger 3- Slope (ou): trigger on the slopes of the signals; the first achieved condition validates the
trigger

Example: The trigger displayed above is: Trigger if
Channel 1 decreasing and equal to threshold S1, value 1V OR Channel 2 decreasing and equal to threshold S1, value -1V

47 / 134

DAS60 User’s Manual The same interpretation is valid for the AND trigger (all thresholds). Parasites You can trigger a parasite with a periodic 50Hz signal: trigger when the shape of the signal radically changes:
· parasite impulsion >200µs · sudden change of frequency · sudden change of shape
48 / 134

DAS60 User’s Manual Trigger on Logic Channels After selection of the trigger on Logic Channels, the window allows you to set the trigger condition.
The 16 logic channels can be used in the trigger word: · either active at status 0 (less than 1.6 volt) · or active at status 1 (more than 4.0 volts) · or non used X · AND / OR: the logic function AND / OR is applied to each channel · Type: edge or level: the trigger is applied on an edge (change of value) or a level
49 / 134

DAS60 User’s Manual
6. MATHEMATICAL CALCULATIONS
You can apply mathematical calculations on your data acquisitions. You can access them with the « Direct display » function.

6.1. Definitions
Press the « Mathematical calculations » key. 3

21

4

1- Add: add a mathematical calculation 2- Remove: remove one of the displayed mathematical calculations 3- Channel: selection of the channel on which the calculation is applied 4- Function: selection of the calculation function applied in the formula. You can change it by
pressing directly on the type, which opens a window with all available calculations.
18 various mathematical formulas are available.
50 / 134

DAS60 User’s Manual You can display up to 5 simultaneous calculations on screen. The display is made in windows above the diagrams, where are specified: · the number of the channel (with its color) · the type of calculation · the value of calculation Under « Direct display », calculations are made in real time and the display is updated every 300ms. The calculation is made on the 1000 points displayed on screen. Hence, the time resolution is 0.1%.
6.2. Types of calculations
51 / 134

Drawing
T1 T2

DAS60 User’s Manual

Math function

Calculation

Minimum

Maximum

Observations
The lowest negative voltage bottom
The highest positive voltage peak

Peak to Peak

Max ­ Min

Low High

The most frequent value below the median
The most frequent value above the median

Amplitude

High ­ Low

On positive oscillation
On negative oscillation

Max – High 100 Amplitude Low – Min 100 Amplitude

Frequency Period
Rising edge

1 Period

Average frequency

Average duration of

Duration of N complete periods a complete cycle

calculated on as

N

many periods as

possible

T1 = 10% Amplitude T2 = 90% Amplitude
Trise = T2 ­ T1

52 / 134

T1 T2

DAS60 User’s Manual

Falling edge

T1 = 90% Amplitude T2 = 10% Amplitude
Tfall = T2 ­ T1

Width of positive impulsion

Measurement of the time of 1st positive pulse. It is measured at
50% of amplitude

Width of negative impulsion

Measurement of the time of 1st negative pulse. It is measured at
50% of amplitude

Positive cyclic ratio

positive pulse duration period

Negative cyclic ratio

negative pulse duration period

Average

Aver

=

1 N

N
Vi
i =1

N : total number of dots

Calculation on the whole graphic range

RMS

( ) RMS =

1 N N i=1 Vi

2

Calculation on the whole graphic range

53 / 134

DAS60 User’s Manual
7. DIRECT MODE
This chapter describes the Direct mode of the recorder, designed to print out the measurements made on the channels in real time on heat-sensitive paper.
This mode does not exist on models without integrated printer
You can start and stop the print out under various conditions. You can activate a simultaneous save of the measurements on file.
7.1. Launching the printing

Press

or

.

Programming the start and stop conditions of the print. Selection of the actions after print and validation of the save in real time.
1

7 2 3 4 5 6

1- Mode: selection of the printing mode · F(t): printing a graph as a function of time · Digital: printing digital values of the measurements as a table
2- Paper speed: printing speed You can change it from 10mm/min to 20mm/h.
3- Print configuration: see chapter 4.2 4- Start: start condition of the print

54 / 134

DAS60 User’s Manual

· Manual: with the

key

· Trigger: with a channel, with a combination of analog and logic channels

(see chapter 5)

· Stand-by: after a delay or at a precise date and time

5- Stop: stop condition of the print

· Manual: with the

key

· Trigger: with a channel, with a combination of analog and logic channels

(see chapter 5)

· Stand-by: after a delay or at a precise date and time

· Paper length: after printing a programmed length of paper

6- After acquisition: action after the end of the print; only possible if the start and stop triggers are not Manual · Stop: no action · Relaunch: waiting for the trigger Start · Save to file: saving or not of the data acquisition into a file

7- Save: simultaneous record of the measurements into a file · Simultaneous save: saving or not the simultaneous save · File: record to a file o Name of the file: directory and name of the saving file o Acquisition period: sampling rate of the channels o Length of the max file.

See chapter File mode to know the restrictions of the recording on file.

· If an initial scroll speed has been programmed, the paper will scroll at that speed while waiting for a trigger.

· The operator can force a trigger any time by pressing start trigger.

when waiting a

· When printing, pressing

will stop the printer.

55 / 134

DAS60 User’s Manual
7.2. Configuration of print

Press

.

1

2

3

4

Definition of the specific characteristics of the paper printing:
1- Reticule: setup of the reticule on paper · none, all 5mm, 10mm or per integer number of divisions · fine or wide, which means with or without subdivisions · access to the « Modification of diagrams » to setup the display of channels on paper (see chapter Diagrams)
2- Name of the plot: setup of a title (80 characters max.) for printing. It will appear on top of the paper.
3- Boundaries: possibility to note the boundaries of each channel (same as on screen) at the end of the print.
4- Notes: possibility to write notes while printing · without, at the beginning of the printing, on alarm, on a length of paper

56 / 134

DAS60 User’s Manual
7.3. Relaunching the printing
In relaunch mode, as soon as stop conditions happen, the device waits for new start conditions for the print out. To terminate the cycle, you only have to force the interruption of the print by pressing during the print.
7.4. Writing information
At the beginning of each printing, the date and time and the programmed title appear at the top of the paper. The number of each channel appears on the plot.
57 / 134

DAS60 User’s Manual
8. MEMORY MODE
This chapter describes the Memory mode designed to record in real time the measurement data made on the channels into the internal memory.
You can start and stop the data acquisition under various conditions. You can simultaneously save measurements on file.

8.1. Setup and start of the data acquisition

Press

.

Setup of the data acquisition into internal memory.

1

5

2

3
4 6

7

1- Blocks: parting of the internal memory in blocks · 1, 2, 4, 8, 16, 32, 64 or 128 · Erase blocks: erasing of all blocks: the current block is #1
2- Acquisition period: sampling rate of the channels · internally paced from 1µs to 10min · externally paced by the logic channel 16

3- Start: start condition of the data acquisition

58 / 134

DAS60 User’s Manual

· Manual: with the

key

· Trigger: with a channel, with a combination of analog and logic channels

· (see chapter 5)

· Stand-by: after a delay or at a precise date and time

· Automatic: immediate; automatic stop when the block is full

4- Stop: stop condition of the data acquisition · Automatic: when the block is full · Trigger: with a channel, with a combination of analog and logic channels (see chapter 5) · Trigger mode: Double Trigger; see the following chapter.

5- Real time save: simultaneous record of the data acquisition into a file · directory and name of the saving file · maximum length of the file

6- After acquisition: action after the end of the print; only possible if the start and stop triggers are not Manual · Stop: no action · Relaunch: waiting for the start trigger · Save to a file: Saving or not the data acquisition into a file

7- View of the active channels, of the recording time, of the pre-trigger…

Pre-trigger: definition of the position of the Start trigger in the data acquisition – data acquisition before and after trigger (before or after Start), and disabling of the trigger.

You can access this menu when setting a trigger by pressing

59 / 134

DAS60 User’s Manual « Impossible » shows up when there is not enough place for saving data: you should reduce the sampling rate or the number of channels
See chapter File mode to know the restrictions of the recording on file.

8.2. Sampling period
When the frequency of the input signals is too high to record them in Direct mode, you must record the measurement signals into memory at high sampling rate: it is the Memory mode. Sampling consists in recording the instant value of a signal at regular intervals and storing each of these values into a memory. To get adequate acquisition of a periodic signal, you need to choose a sampling period compatible with the frequency of the signals present at the input terminals of the recorder.
A 10-dot per signal period definition is minimal for a satisfactory plot.

8.3. Internal memory, blocks
The available memory in the recorder can be shared into blocks that will be filled one after the other at each data acquisition.
The acquisition depth is worth N / (B V) N= total memory amount B = number of blocks V = number of channels to scan
Example: memory 32 Mwords, 16 blocks and 3 channels ‘ON’, we shall have: N = 32 554 432 / (16
3)= 699 050 words/channel.
Any new data acquisition goes automatically into the first free block (in increasing order). When all blocks are full, any new data acquisition will shift the previous blocks from the N position into the N-1 position; the 1st block will be lost. The new data acquisition will take the place of the last block.

8.4. Trigger position
The data acquisition into a memory block relies on the concept of « circular buffer »:

60 / 134

DAS60 User’s Manual

· the buffer is filled from the starting condition on
· the buffer is filled as long as the stop condition is not valid; when the buffer is full, it fills from the initial position of the buffer zone (block)
Thus, the memory block always contains the last N samples.

The operator can select when the samples in buffer will be frozen; to do so, you will select a delay between the trigger start and the beginning of the effective buffering. This delay may be positive or negative: thus, buffered samples may be before, after or on both sides of the start moment.

You can also not inhibit the trigger: you will be able to test the trigger during the pre-trigger phase, in case the trigger should happen before the end of this phase. In case of repetitive signals, the trigger should be inhibited.

Delay to memory input: -50% from the 1st position of buffer

Delay to memory input: +50% from the 1st position of buffer

8.5. Double Trigger mode
In this mode, you will use a start trigger and a stop trigger. Measurements between these two triggers shall be registered.
Data acquisition will stop automatically: – when the stop condition is valid, or – when the memory block is full.

An information window will sum up the general configuration of data acquisition:

channels and functions between channels validated for data acquisition

number of points per channel (as a function of the number of blocks)

total data acquisition time (as a function of the acquisition speed)

– position of the trigger (as a function of pre-trigger phase)

61 / 134

8.6. Recording

DAS60 User’s Manual

You launch the data acquisition by pressing

after pressing

.

At the top left of the screen appear: – the number of the pending block, if required – the current sampling rate – the status of data acquisition (waiting for trigger, filling xx%…) – the opening of a save file, if required – a bargraph showing the percentage of data acquisition made and the percentage of data acquisition displayed

Pressing

stops the pending data acquisition process.

If the acquisition time of the record is less than 2 minutes, you will see the whole data acquisition. Then, you won’t be able to leave this page: only when the acquisition stops can you change of menu.

62 / 134

DAS60 User’s Manual For longer acquisition time, you can zoom onto a part of the data or change of page. When changing of menu page, you can return to data acquisition by pressing

Additional menu keys: Display:

· Total

: the whole memory is displayed and refreshed during the data acquisition.

· Partial

: only a part of the memory is displayed; the screen is frozen; only the

bargraph and the filling rate show the status of data acquisition; you have access to time

cursors and zooms.

You cannot visualize a block else than the current one, draw a plot or save the data acquisition into a file as long as the current data acquisition is not finished.

63 / 134

DAS60 User’s Manual
8.7. Memory output
You can access the « memory output » mode by pressing Display of the data acquisitions available in the internal memory or in files. Possibility to start the paper printing of the data acquisitions.
1
2
3 4 5 6 7
8
9 10
1- Read blocks or file: selection of the memory block or of the file to display; if there is no valid block, the only menu key available will be « Load file »
2- Screen setup: type of display, diagrams; see chapter Use 3- Time cursors: vertical cursors, zoom 4- Voltage cursors: horizontal cursors 5- Zoom between cursors: allows a zoom between the positioned cursors 6- Zoom + & Zoom -: allow zooming or not onto a part of the data acquisition 7- Plotting setup: configuration of the data acquisition drawing; « Launch plot » key 8- Total display of the data acquisition 9- Validation of the buffered channels to display 10- Cursor for scrolling along the data acquisition.
The upper window displays: · the number of the displayed block · the sampling rate of the block · the date of trigger · the number of points per channel in the block · a bargraph showing the filling level of the block and the position of the trigger
64 / 134

DAS60 User’s Manual
9. TEMPLATE MODE
This chapter describes the Template mode designed to record in real time the measurement data made on the channels into the internal memory. You can start the data acquisition under various conditions It stops when the measurement values go out of a range defined by a previous data acquisition called Template. A simultaneous save of the measurements onto a file can be activated.
9.1. Setup and launching of the data acquisition
Press

Setup of the data acquisition as a template into internal memory.

12

3

4 5 6

65 / 134

DAS60 User’s Manual
8- Blocks: parting of the internal memory in blocks · 1, 2, 4, 8, 16, 32, 64 or 128 · Erase blocks: erasing of all blocks: the current block is #1
1- Acquisition period: sampling speed of the channels · internally paced from 1µs to 20min · externally paced by the logic channel 16
2- Real time save: simultaneous record of the data acquisition into a file · directory and name of the saving file · maximum length of the file
3- Start: start condition of the data acquisition

· Manual: with the

key.

· Trigger: with a channel, with a combination of analog and logic channels

(see chapter 5)

· Stand-by: after a delay or at a precise date and time

– Automatic: immediate; automatic stop when the block is full

4- Template channels: channels to compare with the template; access to the setup of the

template

5- After acquisition: action after the end of the print

· Save file: save the data acquisition into a file if the Real Time save has not been

validated (or is impossible)

· Draw: plotting the data acquisition

· Stop: no other action

· Relaunch: waiting for the Start trigger

· Change setup: loading a configuration and waiting for the Start trigger

« Impossible » shows up when there is not enough place for saving data: you should reduce the sampling rate or the number of channels

See chapter File mode to know the restrictions of the recording on file.

66 / 134

DAS60 User’s Manual
9.2. Creation of the template
Make a data acquisition into memory.

Press

, then select

.

The device displays the current template. It is marked as 2 curves (a min. curve and a max. curve) around the channel selected as a reference for the template.

The 2 curves are saved in internal ROM.

1 2

3 4

1- Channel: allows the selection of the reference channel 2- DX & DY: allow you to build the min. and max. curves 3- Begin & End: limits of the memory extent where the trigger test shall be made 4- Save Go/NoGo: as soon as the template is correct, you can save it into the internal ROM.
You cannot save a template on a file. To save a template on a file, save the block and the channel in order to recreate the template for reuse.

67 / 134

DAS60 User’s Manual

Use of the template The comparison with the template is the stop criterion of the data acquisition. This comparison shall

be made with the « Template Channels » selected in the

page.

The information window at the bottom of the page

sums up the general

configuration of the data acquisition:

· channels and functions between validated channels for data acquisition

· channels and functions between channels compared to the template

· number of points per channel (as a function of the number of blocks)

· total data acquisition time (as a function of the acquisition speed)

· position of the trigger (as a function of pre-trigger phase)

The comparison to the template is only possible when the channels are validated for data acquisition. An error message shows up if you have selected « Template Channels » that have note been validated.
The data acquisition is made as above (see chapter Memory Mode).
When the memory block is full, the device checks that all measured points on the template channels are between the 2 min. and max. reference curves constituting the template.
If all points are between the curves, the data acquisition restarts (in the same block). If not, the data acquisition stops.

68 / 134

DAS60 User’s Manual
10. FILE MODE
This chapter describes the File Mode designed to record in real time the measures made on the channels. The start and stop of the plot can be triggered under various conditions.
10.1. Setup and triggering of the data acquisition

Press

.

Setup of the data acquisition into a file. 1

2
3 4

1- File name: directory and name of the data acquisition file · Directory of the data acquisition file · Name of the data acquisition file
2- Acquisition period: sampling speed of the channels · internally paced from 1µs to 20min · externally paced by the logic channel 16
3- Start: start condition of the data acquisition

· Manual: with the

key

· Trigger: with a channel, with a combination of analog and logic channels (see chapter 5)

· Stand-by: after a delay or at a precise date and time

· Automatic

69 / 134

DAS60 User’s Manual 4- Stop: stop condition of the data acquisition
· Automatic: when the disk is full · Trigger : with a channel, with a combination of analog and logic channels (see chapter 5) Trigger mode Double Trigger; see chapter Memory mode.
10.2. Restrictions
Real time data acquisitions to a file are limited by the transfer rate between the inputs and the saving means of the internal memory or USB sticks. It is advisable no to create too big files, because reading them may be very long. It takes ca. 3 minutes to read a 200MB file. The minimum transfer rate to the flash disk is 500 Ksamples/s (2µs) To know the number of channels you can record at a given acquisition period, calculate: · number of channels = 500 000 x acquisition period
example: at 2µs, the number of channels is max. 1.
70 / 134

DAS60 User’s Manual
11. POWER ANALYSIS
11.1. General
The recorder also includes a network analysis function. It allows power and harmonics measurements. Measured values can be viewed in real time or delayed. You can also record the available logic channels and channels.

You can access the power analysis menu with the

key.

The setup page shows up. It lets you access various parameters: · Description of network and wiring · Selection of the type of analysis and of the desired values

You can also access other menus:

· Visualization (Oscilloscope, Digital, Harmonics):

key.

· Acquisition:

key.

Attention: The connections of the inputs on the installation to test shall be made by a duly authorized personal.

In power analysis mod the DAS60 should be strictly connected to its external power supply and connected to the socket : this for security reasons ensuring earth connection.

71 / 134

11.2.

DAS60 User’s Manual
Power analysis mode installation

In the main menu, change the mode by pressing page.

. You directly access the setup

Select the basis configuration of the analyzer: · single-phase · Star · Triangle · Three phase mode Aaron: Only 2U and 2I are used. The 3rd phase is re-calculated (method of the double-wattmeter).

The

key allows you to set up the channels in a basis configuration:

· Frequency : select the frequency the closer of your measurement: generally 50/60Hz · Voltage measurement: « Direct » or « Transformer ». If the network voltage is measured
through a transformer, select this setup and enter the transformation ratio · The device automatically selects the right caliber; you only have to specify the nominal voltage
of the networks. Beware: if this value is too different from the real voltage, this may lead to uncertainty and measurements over the caliber · Current measurement: select the current type of input (clamp, shunt, transformer). In each configuration, you have to enter the associated parameters. · For “clamp”: primary current, secondary voltage (these are current / voltage clamps) · For “transformer” (transformer + shunt): primary and secondary currents and value of the shunt · For “shunt”: value of the shunt, in ohms · You also have to set the nominal current and voltage · Data acquisitions are synchronized with the calculation of frequency of channel 1.

72 / 134

DAS60 User’s Manual Wiring and check: Direct display of the network wiring. Any wrong wiring may lead to wrong results. The colors of the channels match the colors of the respective inputs. The voltage inputs U1, U2 and U3 are connected to the terminals 1, 3 and 5. The current inputs I1, I2 and I3 are connected to the terminals 2, 4 and 6.
Fresnel diagram: You can also check if the wiring is right by displaying the Fresnel diagram. The length of the arrows is not proportional to the efficient value of the currents: the user will have to check that what he sees is coherent. The position of the voltage arrows is calculated with reference to the voltage V1. Attention: the angles are those of the voltages/currents representative either of the shape factor or of the fundamental values. You can check whether the connection is direct or reverse. Intensity vectors are calculated in reference to the corresponding voltages.
Attention: even if nothing abnormal appears, there may be mistakes. In any case, the wirings must be made with caution by a qualified personal.
Warning messages: Messages may appear in red any time at the top of the screen: · Impossible to synchronize the signal: the frequency is wrong, the signal is too weak or too noisy · Caliber exceeded: The voltage or the current are too high for the theoretical nominal value: you
must change this value or use another probe.
73 / 134

DAS60 User’s Manual
11.3. Power analysis setup
Once the device is set up, you have to set the type of measurement you want to make by selecting
the touch You can measure: voltage, current, power, power factor, energy, frequency, harmonics, and not used channels. A maximum of 14 parameters can be selected with check boxes.

Recordable parameters · Number of observed values: until 14. · By clicking on the parameter you get:
· You can select:

74 / 134

DAS60 User’s Manual

· Voltage and Current:

o RMS

RMS : root mean square value

o Fund

fund : RMS value of the fundamental

o Average

dc : average value (DC offset)

o Peak

cr : maximum value between the max and min peaks

o Peak factor

fcr : ratio between the peak value and the efficient value

(1,414 for a sinusoidal signal)

o Total harmonic distortion THD: ratio of the total presence of harmonics to the

fundamental component of the signal

o Distortion factor DF: DF% is the ratio of the total presence of harmonics to the true

efficient value of the signal.

· Power:

o P : Active

o Q : Reactive

o S : Apparent

o Energy: cumulative value; you will find an initialization at page « Display »

· Energy:

Measurement on a phase E1, E2, E3 or all phases.

Integration Value, a reset is in the page “Real time display”

· Power factor:

o PF : power factor

o Cos phi : cosine between the fundamental components of the signals U1 and I1,

· Frequency: the calculation of frequency is made at the input U1.

· Harmonics:

Harmonic measurement on a voltage or current on a given rank (2 to 50)

· Other channel:

We can measure standard channels not used by the power analysis (example channel 5

and 6 in ARON mod) and the 2 PT00 channels.

For voltage, current and power, you can select each line (ex. U1, P2, I2) or the global value of the network (Ut, It, Pt)

For each parameter, you can change the following values: · Validity: / : allows you or not to display a parameter as a graph; the parameters will always be recorded; only the graphic display is changed.
· Color: color of the displayed parameter in the graph; for the additional channels and the logic channels, the colors are as set up for each of them .
· Range: the range is defined by the nominal values of the voltages and currents. For example, for a nominal voltage of 230V, the range shall be 400V centered on 200V (hence, between 0 and 400V); the minimum and maximum values may be changed for finer display. For example, you will be able to display the signal U1 between 220V and 240V.

11.4. Real time display of the signals

By pressing

, you have access to the following windows:

75 / 134

DAS60 User’s Manual
You have access to this page even during data acquisition. In case of energy measurement the reset is made by pressing Each of these windows can be invalidated by pressing .
· Fresnel diagram: display of the signal vector and the RMS values of the inputs and the frequency.
· Oscilloscope: This mode allows you to see shape of the voltage and current signals on a least one period. You can change the screen number and the position of the displayed channels.
· Harmonic display: This window is updated ca. once a second. 76 / 134

DAS60 User’s Manual o Selection of the channels: you select either all the voltage channels (U1, U2, U3) or all
current channels (I1, I2, I3) or all network channels (U1, U2, U3 + I1, I2, I3) or only one channel (these channels are also used to record harmonics. o Sampling frequency: 200µs o Number of points: 2048 o A cursor allows you to display the value of the selected harmonic · Parameters: display of the values of each parameter. At any moment, you can copy the screen into a bitmap file (on a hard disk or a USB stick) by pressing .
11.5. Acquisition parameters
To access the trigger and acquisition parameters you click on
You should first set the sampling period (200µs mini), the directory of the file and the maximum size of the file. The start conditions are:
· Manual : The acquisition starts the pressing · Trigger:
Acquisition starts on a particular level of a record parameter You select the parameter and the condition of trigger
77 / 134

DAS60 User’s Manual “Inrush” mod: the trigger occurs when the current is over a threshold with hysteresis.

· Waiting: the acquisition starts on a date or after a delay
· Automatic : the acquisition starts automatically after pressing
The stop conditions are: · Awaiting: the acquisition stops at a date or after a delay · Automatic: the acquisition stops when the programmed acquisition time is reached.
Next… After an acquisition we can stop or start a new acquisition.

12.6. Record

You launch the data acquisition by pressing

.

We wait the Trigger condition if programmed

The real time acquisition is made in the disk

The trigger can be forced by pressing
During the data acquisition, you cannot change the parameters of data acquisition, but you can change the display parameters (zoom, color, display validity…).
78 / 134

DAS60 User’s Manual You can also display signals in scope mode or display the parameters in digital mode.

To stop the data acquisition, press

.

11.7. Measurement process

Efficient values:

X rms =

x N 2 k =1 k N

Average values:
Active powers: Apparent powers:

X mean

=

1 N

N
xk
k =1

P

=

1 N

N
vk
k =1

  • ik

S =U*I

Reactive powers:

Q = S2 – P2

Power factor: Peak factor: Total harmonic distortion:

FP = P S
Fc = Xpeak Xrms

THD =

X

2 t

X2 fund

X2 fund

Distortion factor:
Total active power: Total reactive power:

DF =

X

2 t

X

2 fund

X

2 t

PT = P1 + P2 + P3

QT = Q1 + Q2 + Q3

Total apparent power:

ST = PT2 + QT2

Total voltage:

UT =

U

2 1

+U

2 2

U

2 3

3

Total intensity:

IT = I1 + I2 + I3

Power factor:

FPT

=

Pt St

Energy: cumulative value of power P; this value is reset in the Visualization page.

Attention: the maximum value of energy during the data acquisition is restricted to E= Pn*H, where

Pn = nominal power

H = 12 hours.

79 / 134

DAS60 User’s Manual
12. FILE MANAGEMENT
12.1. General
For all possible files, the device has an internal flash disk available and may accept an USB stick, which makes it possible:
· to save and load the total configuration of the recorder · to save or restore a data acquisition.
The names of the setup files have a « cnf » exvoltage. The names of the data acquisition files have a « rec » exvoltage.
You can create directories and save the files into them.

1

2

7

3

8

4

5

6
1- HD / USB stick: read/write into the internal flash disk or an USB stick (if it was connected at the start of the device)
2- Copy of a file to USB: copy the selected file or directory to an USB peripheral 3- Type the name of a new file 4- Create directory: creation of a new directory; you can type its name with the
alphanumerical keyboard displayed on screen 5- Erase: erases the selected directory or file 6- Close the window 7- Selection of the directory window: Allows you to select the directory to work with 8- Selection of the file window: Allows you to select the file to work with
80 / 134

DAS60 User’s Manual
It is highly advisable to work in a directory and not at the root of the flash disk When erasing a directory, all files in this directory will be erased too.

12.2. Management of the configuration files

The configuration file management page is linked to the

key.

There name bear the “cnf” exvoltage.

1
2 3
4
1- Reset: sets up the device in standard configuration 2- Load from disk: loads a configuration from a file in the internal flash disk or an USB stick 3- Save to disk: saves a configuration into a file in the internal flash disk or an USB stick 4- Save ASCII file to disk : saves a configuration into ASCII file in the internal flash disk or an
USB stick

Save setup files
Press « Save to disk » You can type the name of the file with the alphanumerical keyboard displayed on screen.
Advice: Do not save files to the root of the disk; create directories for a more efficient management.

Load setup files Press « Load from disk » Select the directory and then the file to load, and click « Load ».
Attention: you will lose the pending configuration.

81 / 134

DAS60 User’s Manual
Save data acquisitions There are 2 ways to save data acquisitions into the device:
– in Manual mode after a data acquisition – in Real Time during a data acquisition.

Manual:

Available under every mode; you can access it from the

function after stopping

the sweep and pressing

. The following window pops up:

1 2 3 4 5

1- Read block: lire un bloc mémoire 2- Selection of the block to read 3- Read a recorded file 4- Write data into a file: Record the data displayed on screen into another file. 5- Close the window
Real time:
Available in Direct, Memory and Template modes after validation of « Real Time Save » in the triggers page:

As soon as the record is launched, the device saves the samples into the file. The file closes when the data acquisition is done.
82 / 134

DAS60 User’s Manual
If the data acquisition is waiting for a start trigger, the device records into the file all points before this trigger. When the data acquisition stops, the file will include more points than the memory block that was used for the data acquisition.
In both cases ­ Manual or Real Time record ­ the date of the file is the date when the file was closed.
Name of the files: the name of each file is made of 12 characters plus one 4-cipher number. The device will automatically increase this number at each record.
Only the validated channels (i.e. in ON position; see chapter Use) will be recorded into the file.
Before writing into the file, a pop-up reminds you the saved channels and the number of points per channel.
The file is made of: · a header, including the information later required to recover the file (the setup of each channel,
the sampling rate, etc.). · the N samples (1 sample = 2 bytes per channel).

Recovery of the data acquisition files

Press

from the main menu, then

.

The following window pops up:

Select

. The following window pops up:

83 / 134

DAS60 User’s Manual Select the place and the name of the file to load and display on screen, then press
84 / 134

DAS60 User’s Manual
12.3. Operating software
Data acquisition files may be transferred to a PC computer.
The SeframViewer software is provided on a CD-ROM with the device. It makes it possible to display the recorded files or convert them as xls or txt files.
It works under WINDOWS XP, 2000, Vista, Seven, 8 with Framework 2.0
· Transfer of files from the recorder to the PC: o USB peripheral o Thanks to the FTP protocol.

Transfer via FTP
Using SeframPilot: The ftp transfer is made directly by SeframPilot: if the IP address is for DAS60, validate with

Using Windows Explorer: You can create a favorite icon: · Go to Tools Connect to a network driver · Click « Open a storage session or a network server » · Under « Internet network address », type ftp:// and the IP address of the recorder. You can give a name to this link that will always be active among « network favorites ».
You get a direct access to the directories of your recorder, which contain the data acquisition files: main directory on the internal flash disk of your device.
Once selected the directory where you record your data acquisitions, you can rename, move, copy and delete them.

You cannot read a file in real time while recording it.

You cannot delete a file while the recorder reads it.

Transfer your files to your computer to use them with the SeframViewer or FLEXPRO software.

85 / 134

DAS60 User’s Manual Display under SeframViewer Launch SeframViewer (or double- click on a file) Open a.REC file. You can select:
· the channels to display · the display mode f(t) or xy · the autocalibration of the channels. Your data acquisition file shows up on screen. SeframViewer functions are available.
Refer to the user’s manual included in the software to discover all functions of SeframViewer by clicking the last icon: « Help ».
86 / 134

DAS60 User’s Manual You can also directly create a result file (text or Excel file) by launching the following line in the Windows® command mode: (see Menu, sub-menu Help Options) Example: C:SeframViewer monfichier.rec /x : direct creation of an Excel file C:SeframViewer monfichier.rec /t : direct creation of a text file. Excel® export directly launches Excel in the explorer.
Managing with SeframPilot You can manage the recorder directly from SeframPilot. This software allows you to: · Manage setups (creation and change setup files) · Start and stop records · Load data blocks · Load files thanks to Windows Explorer through FTP · Launch SeframViewer · Real time display of data in F(t) format
87 / 134

DAS60 User’s Manual
13. INPUTS / OUTPUTS
13.1. Additional Input/Output Connectors
The connector is at the rear (female SUB-D 25pin).

pin 13 25 12 24 11 23 10 22 9 21 8 20 7 19 6 18 5 17 4 16 3 15 2 14 1

I/O

Name of the signals

I

Logic channel 1

I

Logic channel 2

I

Logic channel 3

I

Logic channel 4

I

Logic channel 5

I

Logic channel 6

I

Logic channel 7

I

Logic channel 8

I

Logic channel 9

I

Logic channel 10

I

Logic channel 11

I

Logic channel 12

I

Logic channel 13

I

Logic channel 14

I

Logic channel 15

I

Logic channel 16

Ground

Ground

Ground

O

9-15V 0.2A power

Ground

Alarm C

Alarm D

O

Alarm A

O

Alarm B

The power supply ground is the mechanical ground of the carter of the SUB-D25 connector.

88 / 134

DAS60 User’s Manual
13.2. Logic inputs
Non-connected inputs are according to the following circuit:

Non-connected inputs are 0V potential (level 0).
Number of logic channels: from 1 to 16. TTL level: 3.3V (protected up to 24V)
To create a rising edge, you only have to place a connection between the power input and output of the connector. Similarly, to create a falling edge, you only have to remove this connection. You can also use an output signal TTL 3.3V.
Use
Plotting and display: These channels are plotted on the right or left side of the paper according to the selected format. They are displayed at the top or the bottom of the screen according to the selected position. They are numbered from the right to the left. Each channel is plotted between two dotted lines that show their limits. The height of reticules is min. 2mm for the plot.
External clock: You can use the logic channel 16 (see figure “connector”) as a data acquisition clock for the sampling in memory or file mode. The data acquisition can then be made up to 500kHz.
Triggers: You can use the logic channels to trigger the plotting and the data acquisitions (start and stop). (AND or OR mode) See chapter « Trigger ».

13.3. Alarm outputs
Available contacts and outputs on the rear connector A, B, C and D. All outputs are TTL 5V.
When the device is powered off, A, B, C and D outputs have a 5k impedance

89 / 134

DAS60 User’s Manual Use Signal for internal events to the outside toward the recorder (triggers on logic or analog channels, status of the printing block…) See chapter « Use ».
13.4. Power supply output
There is a power supply output restricted to 0.2A (see figure “connector”). The voltage is the one of the 9-12V battery as a function of the charge and 15V is the charger is connected. The ground of this power supply is the mechanical ground. Thus, you can use this output to provide power to a sensor or an electronic circuit in order to manage the logic inputs.
The power supply may disappear in case of overload (> 0.2A). In this case, you must turn off the device a few minutes before turning it on again.
90 / 134

DAS60 User’s Manual
13.5. Exvoltage box input/output interface
The exvoltage box input/output interface option (code 984405500) makes it possible to:
· Convert an alternating voltage (example 230V 50Hz) into a logic signal 1 · Easy connection of the inputs of the 16 logic channels · Insulation of the 16 logic inputs (250V=~ between channels, 250V=~ between channels and
ground) · Easy connection of the alarm outputs by a screw terminal · Power supply of an external accessory with 3.3V, 5V or 12V by a screw terminal.
Connection of the logic channels according to the signal voltage:
· 0 to 250V =~
red
connection 0 – 250V=~ logic channel
black
by insulated pin plug between red and black terminals max. usable voltage: 250V continuous or alternating typical trip threshold (AC or DC): 48V frequency: 45 to 440Hz undetected low threshold (AC or DC): 0 to 10V detected high threshold (AC or DC): 60V to 250V insulation: 250V=~ between channel and ground
· 0 to 48V=~
1 1 2 3
connection 0 – 48V logic channel
by a screw terminal between the points 1 and 3 of the screw terminal box max. usable voltage: 48V (AC or DC) frequency: 45 to 440Hz typical trip threshold (AC or DC): 9V undetected low threshold (AC or DC): 0 to 2V detected high threshold (AC or DC): 10V to 48V insulation: 50V=~ between channel and ground
91 / 134

· 0 to 10V=~
1
1 2 3

DAS60 User’s Manual

connection 0 – 10V logic channel 1
by a screw terminal between the points 1 and 2 of the screw terminal box max. usable voltage: 10V continu ou alternatif frequency: 45 à 440Hz typical trip threshold (AC or DC): 2.2V undetected low threshold (AC or DC): 0 to 1V detected high threshold (AC or DC): 3V to 10V insulation: 50V=~ between channel and ground
Response time: To detect the alternating current, the signal of the logic channels is rectified and filtered. Typical delay for a rising signal: 10ms Typical delay for a falling signal: 50ms
Use: Link the box to the input/output connector of the device with the 25-pin connector provided as an option. Connect the logic channels to record with respect of the maximum acceptable voltages.
Power supply / Alarms:

max current 3.3V:

200mA

max current 5V:

200mA

max current 12V:

200mA

Alarm A contact 1 & 2: open contact in case of alarm, insulation 50Vmax / ground

Alarm B: logic output 0V / 5V not isolated

Alarme C: logic output 0V / 5V not isolated

92 / 134

DAS60 User’s Manual
14. ETHERNET INTERFACE
14.1. Ethernet Interface
General You can use the recorder remotely via its Ethernet interface with the TCP-IP protocol. Connect the recorder to your network with a straight-through cable onto the 10/100 BASE-T (RJ45) connector at the rear of the device.
You can use the recorder on an Ethernet network 10 Mbit/s or 100Mbit/s.
In case of a network with BNC cable, you must use an external hub to convert the BNC signal into RJ45 signal (use a straight-through cable).
If your PC has no network, you can also use a crossover cable directly between the PC and the recorder.
Crossover cable

The recorder uses the TCP/IP protocol to dialog with the PC. Thus, you can give it an IP address with a subnet mask. Ask your network administrator an IP address and a corresponding mask, then, when you have set

up the recorder with the are taken into account.

key, turn off and on the device, so that the changes

If you don’t have any network administrator: Check that the TCP/IP driver is correctly installed on your device. Case of a PC under Win XP or other: From Windows Explorer, go to “Work station”, “Configuration panel”, “network connection”, and then go to the properties of the network card. Check that the TCP/IP protocol has been installed, then check the IP address and the subnet mask.

You have to take great care and define an IP address and a subnet mask that are compatible with the ones of the remote computer. The port that the recorder uses is Port #23.

For example, in I mode, you could have:

PC Recorder

IP=192 135.20.00 IP=192.135.20.01

mask = 255.255.255.0 mask = 255.255.255.0

93 / 134

DAS60 User’s Manual Refer to the definition of the classes of IP addresses. Programmation: You can create your own software under Visual Basic, Visual C++, etc. by using, for example, the Winsock.dll driver by Microsoft. Then, you only have to send the commands as per the following paragraphs to the recorder.
14.2. WIFI interface
Use only the USB stick provided by SEFRAM. Turn the device on with the USB stick on a relevant port. In the « Setup » menu, press the « Network » key. Once the Wi-Fi validated, press the « Wi-Fi networks » key.
(if you don’t see your network, press the Ethernet/Wi-Fi keys several times) When your network appears, select it and validate. The commands are automatically positioned. You only have to type the password. In case of connection problem, contact your network administrator.
In the setup page you will get the IP addresses of the Ethernet and Wi-Fi networks.
94 / 134

14.3. FTP transfer

DAS60 User’s Manual

Transfer your files to your computer to save them or view them with the SeframViewer software.

Using of the Windows file explorer

Filezilla: freeware (https://filezilla-project.org/)

Using an external browser (Internet Explorer, Mozilla, Opera, Chrome, etc.)

Launch your browser and type the connection address (see § Setup): « Ftp:// » followed by the IP address of your recorder Then, press the Entry key of your computer.
You will get direct access to the directories of your recorder, including the data acquisition files: HD: root directory of the internal drive disk of your device. Once you have selected the folder where you recorded your data acquisitions, you can rename them, move them, copy them or erase them. Transfer your files to your computer to use them with the provided SeframViewer software or FLEXPRO.

You cannot read a file in real time while recording it.

95 / 134

14.4.

DAS60 User’s Manual
Managing with VNC:

This piece of software allows you to manage the recorder remotely.

Changing the password
In the « Setup » page, press VNC. You can then change the password of VNC. The default password is « sefram ».

External software VNC Viewer
You are submitted a link to easily download this software: http://www.realvnc.com/download/viewer/. Select the.exe file.
You only have to follow the instructions on site to install this software properly. At the end of setup, you will have the possibility to select a version for your VNC viewer (you may have to pay for some of them). Select the version you want according to the applications you need.
Just open the « VNC viewer » file where you recorded it.
On the VNC Server line: you must type the IP address of your recorder and the password (the default password is « sefram »), then click « Connect »
You will be able to control your recorder from your computer.

96 / 134

DAS60 User’s Manual
14.5. Programming language
Format of the reception messages
In all the following examples, the space character is pictured as a blank space.
Exchanges from a controller to the recorder are made as messages made of a chain of ASCII characters (and possibly of binary bytes) with a message termination at the end. Syntax of a receipt message

Message unit: if the message includes several message units, they are separated with a semicolon “;” and possibly preceded and/or followed by one or several “filling” characters in ASCII code (0 to 32, in decimal format except 10 and 13).
The message termination (TERM) for the Ethernet link is: – LF: Line Feed (10 in decimal)
The message termination may be preceded with one or several “filling” characters under ASCII code (0 to 32, in decimal format except 10 and 13).
Example of messages made with 3 message units: MESSAGE 1; MESSAGE 2 ; MESSAGE 3 TERM CHANNEL 1; TYPE:VOLTAGE DC;:CALDEC ? TERM
Syntax of a message unit
A message unit (ex. :REAR:SETUP 1) is made of several fields:
– Header: For the command messages (ex. :REAR:SETUP 1) or query messages (ex. :REAR ?), it is made of a chain of characters (simple header) or several chains separated by “:” (compound header). A chain includes from 1 to 12 alphanumeric characters or “_” (code ASCII 95 in decimal format). Recommended chain length: 4 characters. A header chain always starts with an alphabetic character. It may be preceded by a colon “:” (compound header) or end by a question mark “?” (query message).
Query messages must be followed by the termination.

– Header separator:

97 / 134

DAS60 User’s Manual
One or several ASCII characters (0 to 32, decimal except 10 and 13).
– One or several pieces of data: (ex. :SPEED 1,MMS), alphanumeric, digital or compound of various characters or binary bytes.
– Data separator: A comma “,” possibly followed and/or preceded by one or several “filling” characters in ASCII code (0 to 32, decimal except 10 and 13).
Data: There are several types of data:
– Alphanumeric data: 1 to 12-character words that can be alphabetic (upper case or lower case), digital or “
” (95d) coded under ASCII. Words always start with an alphabetic character. For example, for a non-digital parameter: S1M.
– Decimal digital data: Made of a mantissa and possibly of an exponent, and shown as a chain of ASCII characters starting with a cipher or a sign (+ or -). They are of NR1 (integer), NR2 (decimal) or NR3 type (with exponent) or of a combination of these three types.
– Text: Any chain of ASCII 7-bit characters between brackets (“) or apostrophes (‘). ex. “Channel 1”
Formats of emission messages Exchanges from a recorder to a controller to the recorder are made as messages made of a chain of ASCII characters (and possibly of binary bytes) with a message termination at the end. The format of emission messages is identical to the reception messages, although with a more rigid structure.
The syntax of an emission message is: Message unit + message termination.
Message unit: If the message includes several message units, they shall be separated by a semicolon “;”.
Message termination: – LF: Line Feed (10 in decimal format)
Syntax of a message unit:
A message unit (ex. TYP:THE J,COMP) is made of several fields:
– One header: (ex.TYP:THE) made of only one (simple header) or several (compounds header) chains of 1 to 12 alphanumeric characters (upper scale only or digital or “_” (code ASCII 95 in decimal format).
98 / 134

DAS60 User’s Manual
A header chain always starts with an alphabetic character. In a compound header, chains of characters are separated by “:” (ex. TYP:THE).
– Header separator: “space” (32d) only.
– One or several pieces of data: (ex. J,COMP) alphanumeric, digital or compounds of any characters or binary bytes.
– Data separator: Comma “,”.
Data: There are several types of data:
– Alphanumeric data: 1 to 12-character words that can be alphabetic (upper case only), digital or “_” (95d) coded under ASCII (example: J).
– Decimal digital data: Made of a chain of ASCII characters starting with a cipher or a sign (+ or -). They are of NR1 (integer), NR2 (decimal) or NR3 type (with exponent). For example, for a digital character: -25.02.
– Text data: Any chain of ASCII 7-bit characters between brackets (“) or apostrophes (‘). Example : “A”.
– Any chain ASCII characters ends with the message termination.

14.6. Standard instructions
All these instructions start with an asterisk “*”.

*IDN ?

REQUEST TO IDENTIFY AN DEVICE

answer by the device: 4 pieces of data separated by ‘,’:

– the trademark of the device

– the name of the device followed by _nn where nn is the number of inputs of the recorder

– the serial number of the device (0 if unknown)

– the software version number as x.xx x

*OPT ?

REQUEST TO IDENTIFY OPTIONS OF AN DEVICE

answer by the device: n pieces of data separated by ‘;’:

– number of cards

– number of channels per card

*RST

RESET OF AN DEVICE

action: setup of a recorder in a fixed configuration (voltage inputs, 10V caliber, centered on 0V…)

99 / 134

DAS60 User’s Manual

*REM

PENDING PROGRAMMING PART (REMOTE) (facultative)

*LOC

RETURN TO LOCAL MODE

*CLS

ERASING THE STATUS LOGS

action: the device resets the status logs.

*ESE

VALIDATION OF THE STANDARD EVENT BITS OF A DEVICE

*ESE is followed by a number between 0 and 255

action: changes the log of validation of the standard events and updates the ESB bit in the status log

of service request (see following paragraph).

*ESE ?

INTERROGATION OF THE CONTENT OF THE VALIDATION LOG OF THE

STANDARD EVENTS OF A DEVICE

answer by the device: NR1 number between 0 and 255 (see following paragraph).

*ESR ?

INTERROGATION OF THE CONTENT OF THE STATUS LOG OF THE

STANDARD EVENTS OF A DEVICE

answer by the device: NR1 number between 0 and 255.

All events are erased and the log is reset to zero (see following paragraph).

*SRE

VALIDATION OF THE SERVICE REQUESTS OF A DEVICE

*SRE is followed by a number between 0 and 63 or between 128 and 191.

action: the device changes the validation log of the service request (see following paragraph).

*SRE ?

INTERROGATION OF THE VALIDATION LOG OF THE SERVICE REQUESTS

OF A DEVICE

answer by the device: NR1 number between 0 and 63 or between 128 and 191 (see following

paragraph).

*STB ? READING OF THE SERVICE REQUEST LOG OF A DEVICE answer by the device: NR1 number between 0 and 255: status word with bit 6 MSS (Master Summary Statut) (see following paragraph)

14.7. Indication of the status of the device
Structure of the status data Here is the pattern of the structure of status data: they keep you informed about the changes of status in the device (reconnection, printing start…). Overview of the structures of status data of the recorder:
100 / 134

DAS60 User’s Manual
4 logs are used: – the service request log (STB) associated with its validation log. – the standard event log (ESR) associated with its validation log. The bits #0, 1, 2 and 7 of the STB log are available for use as sum-up messages specific to the device. Each of these bits can be associated to a structure of data with a defined pattern, which will manage the events in the device that may lead to a service request. The user may set up the recorder, so that it activates the bit #6 of the service request log in case of one or several specific events. Status of these logs at power-up: The content of the STB, ESR and alarm logs is systematically reset at power-up (except the bit #7 of the ESR that specifies a power-up).
101 / 134

Service request logs

DAS60 User’s Manual

Status log: It includes the status word of the device. You can read this status word thanks to the request with the “STB?” instruction. In this case, the bit #6 is MSS (Master Summary Status) that results from the logic operations as pictured in the figure below. Indeed, MSS is 1 when at least one of the other bits is 1 both in the status log and in the validation log.
Composition of the STB log: The bit #6 (value 64) contains the sum-up message “MSS” (reading with “
STB?”).
The service request takes place in the following cases: – one bit of the status log of service request switches from 0 to 1, while the corresponding bit in
its validation log is 1, and vice versa – the bit #5 of the validation log of service request is 1; a standard event happens in the following
conditions: – one bit of the status log of standard events switches from 0 to 1, while the corresponding bit
in its validation log is 1 – one bit of the validation log of standard events switches from 0 to 1, while the
corresponding bit of its status log is 1 – the bit #0 of the validation log of service request is 1; a standard event happens in the
following conditions: – a bit of the status log of alarms switches from 0 to 1, while the corresponding bit in its
validation log is 1 – a bit of the validation log of alarms switches from 0 to 1, while the corresponding bit of its
status log is 1.
The bit #5 (ESB: Event Status Bit, value 32) contains the sum-up message of the status log of standard events (see the detail of these bits in the description of this log). Its status specifies whether one or several authorizes events have happened in the status log of standard events since its latest reset (an event is authorized if the corresponding bit in the validation log of events is 1).
The bit #4 (MAV: Message AVailable, value 16) contains the sum-up message of the output queue. Its status specifies whether a message or data from the device are ready for emission via the interface (ex: answer to a query instruction).
102 / 134

DAS60 User’s Manual
The bits #7, 3, 2, 1 and 0 are used to receive sum-up messages as defined by the device. In the case of the recorder, the bit #0 is used while the bits #1, 2, 3, 7 always remain 0. The bit #0 contains the sum-up message of the status log of alarms (see the detail of these bits in the description of this log). Its status specifies whether one or several authorized events have happened in the status log of alarms since its latest reset.
Validation log: The status word is associated to a validation log that makes it possible to manage the service request by allowing it only for some specific cases. When a bit is 1, it allows that the status 1 of the bit of same rank in the status log (STB) leads to the activation of the bit #6 of the same log. Writing the validation byte is made with the command SRE, where is the sum of the binary values of the bits 0 to 5 and 7. Reading the validation byte is made with the instruction SRE?. The answer is sent back under decimal format (NR1).

Log of standard events
Refer to the overview of the structures of status data. The structure of the logs of standard events is appended to the bit #5 of the log of service requests.

Status log: This log contains some specific standard messages, whose meaning is exposed hereafter. You can read its content with the command *ESR? Reading the log erases it.

The bits of the log of status of events are appended to specific events:

  • bit #7:

POWER UP (value 128)

It shows that the device has been powered up

  • bit #6:

REQUEST FOR USE (value 64)

Not used. Positioned to 0

  • bit #5:

INSTRUCTION ERROR (value 32)

This bit shows that an unknown or wrong instruction has been sent to the recorder

  • bit #4:

EXECUTION ERROR (value 16)

Not used. Positioned to 0

  • bit #3:

DEVICE-LINKED ERROR (value 8)

Not used. Positioned to 0

  • bit #2:

REQUEST ERROR (value 4)

This bit shows that the output queue is full and that the data is or may be lost

  • bit #1:

CONTROL REQUEST (value 2)

Not used. Positioned to 0

  • bit #0:

OPERATION COMPLETED (value 0)

Not used. Positioned to 0

An event is authorized if the corresponding bit in the validation log of events is 1.

Validation log: It manages the status log of standard events: When a bit of this log is 1, it authorizes that the status 1 of the bit of same rank in the status log of standard events leads to the switch to 1 of the bit #5 in the status log of service requests (STB). You can write in this log with the command *ESE where is the sum of the binary values of the validation log.
103 / 134

DAS60 User’s Manual You can read its content with the command “*ESE?”

Log of alarms
Refer to the overview of the structures of status data. The structure of the log of alarms is appended to the bit #0 of the log of service requests.

Log of status: This log contains some messages specific to the recorder, whose meaning is exposed hereafter. You can read its content with the command SRQ_TYPE? Reading the log erases it.

The bits of the status log of alarms are appended to specific events:

– bit #7:

TRIGGER DATA ACQUISITION MEMORY (value 128)

This bit shows that the trigger condition of a data acquisition into memory has been

achieved.

– bit #6:

END OF DATA ACQUISITION MEMORY (value 64)

This bit shows that a data acquisition into memory has been achieved.

– bit #5:

START OF DATA ACQUISITION MEMORY (value 32)

This bit shows that a data acquisition into memory has been launched.

– bit #4:

Not used (value 16)

– bit #3:

END OF PAPER (value 8)

This bit shows that there is no more paper in the printer.

– bit #2:

END OF WRITING (value 4)

This bit shows that a text writing is achieved: cartridge, text programmed with the

instruction WRIte (see dictionary of programming)…

– bit #1:

END OF PLOTTING (value 2)

This bit shows that a printing is achieved.

– bit #0:

START OF PRINTING (value 1)

This bit shows that a printing has started.

An event is authorized is the corresponding bit in the validation log of events is 1.

Validation log: It manages the status log of alarms: When a bit of this log is 1, it authorizes that the status 1 of the bit of same rank in the status log of alarms leads to the switch to 1 of the bit #0 of the status log of service requests (STB). You can write into this log with the command *SRQ_ENABLE where is the sum of the binary values of the validation log. You can read the log with “SRQ_ENABLE?”

Use of the status data structure
Before use, it is advisable to send the instruction CLS to the recorder to reset the status logs. The user must first set which events to detect by authorizing them in the validation log:
– with the instruction “SRQ_ENABLE n” for events in connection with the alarm logs – with the instruction “
ESE n” for events in connection with the logs of standard events – with the instruction “*SRE n” for events in connection with the log of service requests.

Example:

104 / 134

DAS60 User’s Manual

Programming a request of service for: a start or an end of the paper print, an instruction error, the

presence of data at the output of the recorder, with the instructions:

SRQ_ENABLE 3

(Bit #0 and 1 to 1)

*ESE 32

(Bit #5 to 1)

*SRE 49

(Bit #0, 4 and 5 to 1)

The controller should regularly read the log of service requests with the instruction “*STB?”.

Switching of bit #6 (MSS) to 1 shows that an authorized event happened.

Reading the status word makes it possible to discover the type of event that happened. In the case of

a standard or specific event, you have to read the associated status log with the instructions

“*ESR?” or “SRQ_TYPE?” to precisely identify the event.

A standard event appeared. The instruction “*ESR?” is sent: Answer by the recorder: 160 (bit #7 and 5 to 1) Two events are shown (power up and instruction error) but the request of service has been caused by the instruction error only (the only authorized event in the validation log).

105 / 134

DAS60 User’s Manual
14.8. Dictionary of programming
In the following table, it is not compulsory to send the lower case characters of the headers and parameters. As a rule, digital parameters are integers (NR1); parameters specified “decimal” may be NR1, NR2 or NR3.

Configuration
HEADER MODE MODE ? PAGe
ALArm
ALArm:DEF ALArm ? DATe
DATe ? HOUrs

PARAMETERS P1 Definition of the mode of use of the appliance P1=DIRect, MEMory, FILE, GONOgo,POWer Returns the mode P1 Displays a screen P1 = SETUP : Config
CHAN : Channel N (see command :CHAN ) TRigger : Trigger CHArt: Paper SCOpe : Visu direct REPLay : Memory output P1 Definition of the alarm to change P1=A,B
P1
P2=NO, Trigger, RECtr or ERRor Returns the alarms P1,P2,P3 changes the current date
P1 = day (from 1 to 31) P2 = month (from 1 to 12) P3 = year (from 0 to 99) Returns the date
P1,P2,P3 Changes the current time
P1 = hour (from 0 to 23)

EXAMPLES MODE FILE
:CHAN 3;:SCREEN CHAN Displays the channel 3
ALARM:VAL A,TR;TR:CH 1,S1,EDGEP the trigger is then set by the command TRig: (see chapter 16.5.8)
DAT 11,12,10 December 11th 2010
HOURS 10,6,0 10 hours and 6 minutes

106 / 134

DAS60 User’s Manual

Parameters of the channels

HEADER CHAnnel
CHAnnel ? VALid
VALid ? NAMe
NAMe ? TYPe:VOLtage
TYPe:SHUNT

PARAMETERS P1 Allows you to set the CHANNEL input you can change with the commands

EXAMPLES CHAN 3 We chose to change the channel 3

P1 = selection of the input 1, 2, PT1, PT2, A, B, C, D.

Returns the number of the selected input and its value. P1,P2
Definition of the authorization status of each channel P1 = ALL for all channels, or
1, 2 etc. for each channel LOG for the logic channels P2 = ON or OFF Returns the validity of all inputs P1
Allows to change the name of the CHANNEL input

VALid ALL,OFF; VALid 1 ON; VALid LOG,ON Only the channel 1 and logic channels are authorized
CHAN 3; NAM ‘four1’

P1 = name (26 char. Max.) between two ‘ or ”
Returns the name of the channel P1

TYPE:THERM K,COMP

Change of the channel to Voltage P1= DC , RMS, DERIVE, INTEGRE P1,P2 Change of the channel to SHUNT
P1 = DC or RMS P2 = S1M,S10M,S01,S1,S10,S50 (for 1mOhm … 50 Ohm)

Use of a thermocouple K with compensation

TYPe:FREQ TYPe:PT100 ou PT1000

Change of the channel to FREQUENCYCOUNTER P1,P2
Change of the type of channel to PT100 P1= W2,W3 for 2 wires or 3 wires P2= Value of the resistance (if W2)

TYPe:THErmo TYPe:COUNTer

P1,P2
Change of the type of channel to Thermocouple P1= Thermocouple = J, K, T, S, B, E, N, C, L P1 Change of the type of channel to COUNTER P1= decision threshold (in volt)

TYP:COUNT 1.4
the command resets the counter

TYPe ? UNIT
UNIT ?

Returns the type of channel P1 Temperature unit in thermocouple and PT100 P1: CEL,FAR,KEL Returns the temperature unit of the channel

UNIT CEL Unit Celsius

107 / 134

DAS60 User’s Manual

HEADER FILter
FILter ? RANGE
RANGE ? THREshold
THREshold ?

PARAMETERS P1
Definition of the filter of the selected channel from the CHANNEL command
P1 = WOUT,F10kHz,F1kHz,F100Hz,F10Hz, F1Hz,F10S,F100S or F1000S Returns the filter of the selected input
P1,P2,P3 Changes the caliber and the center of the input :CHAN P1 = Caliber in ISO units (Volts or °C) in real time P2 = Center in ISO units in real time P3 = Position in percentage Returns the caliber and the center of the selected input P1,P2,P3 Definition of the thresholds P1 = SI or S2 P2 = ON or OFF (validation of the plot) P3 = Value of the threshold Returns the values of the 2 thresholds

Return of the instant values:

EXAMPLES FILTER 10Hz
RANGE 12,3,0 caliber = 12 Volts centered around 3 Volts
:THRES S1,ON,10 threshold S1 = 10 Volts

EN-TETE RDC ?

PARAMETRES Returns the values of all channels and of the logic channels or of the parameters in network analysis

EXEMPLES

108 / 134

DAS60 User’s Manual

Functions of the channels and between channels

HEADER FUNCMATH
FUNCMATH ? COEFf COEFf ? UNITFunction
UNITFunction ? FUNCXY
FUNCXY ? RDUnit
FUNCTion FUNCTIon ?

PARAMETERS P1 Allows the selection of a mathematical function for the CHANNEL input P1 = Type of function : NONe, UNIT, AX, ABSX, SQRX, SQROOTX, LOGX, EXPX,AINVX,ADVDT,AINTV (none, change of unit, ax+b, a|x|+b, ax²+b, …) Returns the function of the CHANNEL function P1,P2 Definition of the coefficients of the function P1 = A, B, C or X1, X2, Y1, Y2 Returns the values of the coefficients of the function of the CHANNEL input P1 Definition of the unit of the function P1 = name of the unit (max. 6 characters) between two ” or ‘ . Returns the name of the unit of the function P1,P2,P3 Additional function between channels P1 = Number of the channel 1 (from 1 to PT2) P2 = Operateur PLUS,MINUS,MULT,DIV P3 = Number of the channe 1 (from 1 to PT2) Returns the function P1
Selection of the measurement unit: P1: ISO unit of the channels NORMe: standard unit between 0 and 10000 P1 Validity of the functions in general P1 = ON or OFF Returns the validity of the functions

EXAMPLES CHAN 2; FUNCTION LOGX; The channel 2 = aLog(x)+b
:COEF A,2;COEF B,0 A = 2 B = 0
UNITF ‘DB’
CHAN FB;FUNCXY 1,PLUS,2
RDU ISO;ONOFF ALL,OFF;ONOFF 1,ON;ONOFF 3,ON;RDU ISO;DRC ?

109 / 134

DAS60 User’s Manual

Paper
HEADER DIRECTPLOT DIRECTPLOT ? SPEed
SPeed ? TEXTSpeed
TEXTSpeed ? GRATicule
GRATicule ? CHART:TITle CHART:TITle ? CHART:DATe CHART:DATe ? CHART:BOUndary P1 CHART:BOUndary ? ANNOte
ANNOte ? ANNOte:TYpe
ANNOte:TYpe ?

PARAMETERS

EXAMPLES

P1

DIRECTPLOT FT

Set-up of the transcription mode on paper in direct mode

P1 = FT, TEXTe, XY

Returns the paper mode

P1,P2

SPEED 10,MM_S

Definition of the paper scroll speed

Speed 10 mm/s

P1 = speed value:

1,2,5,10,20 for P2 = MM_H or MM_M,

1,2,5,10,20,25,50,100,200 for P2 = MM_S

P2 = units

MM_S (mm/second)

MM_M (m m /m inute)

MM_H (mm/hour)

Returns the status of the command SPEED or

SPEED:EXT P1,P2

TEXTSpeed 2,SEC

Set-up of the paper period text mode texte

P1 varies from 1 to 500

P2 is Sec or MIn or HOurs

Returns the period in text mode

P1,P2

GRAT G5,Coarse

Set-up of the reticule on the paper

P1=WOUT,G5,G10 or DIV defines the type of reticule

P2=FIne or COarse

Returns the reticule

P1

CHART:TITLE “OVEN 12”

Set-up of the title of the acquisition

P1 = message between apostrophes

Returns the title

P1

CHART:DAT ABS

Set-up of the type of date on the paper

P1 = ABSolue or RELative

Returns the command

P1

CHART:BOU WITH

Sets up if writing the boundaries at the end of the plot Writing the boundaries

P1 = WITH or WOUT

Returns the command

P1,P2

ANNOT LEN,20

Set-up of the annotation mode

P1 = Wout, STart, ALarm or LEngth

Annotation every 20cm from

10 to 50cm

P2 is worth the number of alarm (from 1 to 2) or the

length of paper

Returns the command

P1,P2,P3

ANNOT:TYpe NAME, NUM,

Writing the names of the channels

VALUE

P1 = NONAME or NAME writing the names of the

channels

P2 =NONUMber, NUMber writing the names of the

channels

P3= NO, VALue, RANge, SCAle MINm ax Set-up of the

type of annotation to write

Returns the command

110 / 134

Triggerings
EN-TETE START:MANual START:TRIG START:WAIt
START:DATe
START:AUTO START ? STOP:MANual STOP:TRIG STOP:WAIt
STOP:DATe
STOP:LENGth
STOP:AUTO STOP ?

DAS60 User’s Manual

PARAMETRES Manual trigger (start or stop) Trigger on a combination of thresholds (see chapter 8.3)
P1,P2,P3 Trigger on a waiting time P1 = number of hours waiting (0 to 23) P2,P3 = minutes, seconds (0 to 59) P1,P2,P3,P4,P5,P6
Trigger at a date
P1 = day (from 1 to 31) P2 = month (from 1 to 12) P3 = year (from 0 to 99) P4 = hour (frm 0 to 23) P5,P6 = minute, second (from 0 to 59) Automatic trigger (except in DIRECT mode) Returns the start command Manual stop (direct mode) Trigger on a combination of thresholds (see chapter 8.3) P1,P2,P3 (Hour, Minute, Second) Trigger on a waiting time (see START:WAIT) only in DIRECT mode P1,P2,P3,P4,P5,P6 (Day, Month, Year, Hour, Minute, Second) Trigger at a date only in DIRECT mode P1 End of trigger at a given length of plot (only in DIRECT mode)
P1 = length of plot x 10cm Automatic stop (memory or file mode) Returns the command of end of acquisition

EXEMPLES START:MANUAL start:trig;:trig:chan A1,S1,POS
START:WAIt 0,2,10 waiting 2min 10sec
SEQ START;SEQ:DATE 3,10,06,15,30,10 starting on 3/10/06 at 15:30:10

111 / 134

DAS60 User’s Manual

Triggers

TRIG:LOG P1

P1,P2
Selection of trigger on logic channels P1 = sets the 16 values of the triggers; add a message delimiter (brackets)

TRIG:LOG “XXXXXXXXXXXXXXX1″, AND
trigger on logic channel VL1

TRIG:CHan P1,P2,P3 TRIG:COm P1

P2 = AND or OR P1 = Number of the channel (1,2, etc.) P2 = Threshold (S1 or S2) P3 = POS or NEG
for rising edge or falling edge Selection of the type of com plex trigger

P1 = OR, AND or DELta means: one of the thresholds (OR) all thresholds (AND)
slope (DELta) TRIG:COm:DELta P1,P2 Selection of the slope

TR:CH 1,S1,EDGEP
Trigger on rising edge of channel 1 (threshold 1)
TRIG:CO DEL;CO:DEL 2,S;RESET; ADD 1,S1,POS; ADD 2,S1,NEG
2 thresholds (S1 on 1 and S1 on 2)

P1 = value (from 1 to 500) P2 = MICro, MILIsec, Sec, Min, Hours

TRIG:COm:REset

ON removes all channels

TRIG:COm:ADD P1,P2,P3
TRIG ?

Add a threshold to the trigger
P1 = Number of channel (1, 2 etc.) P2 = Threshold (S1 or S2) P3 = POS or NEG for rising edge or falling edge Returns the value of the selected trigger

The programmed trigger depends on the latest sent command (alarm, trigger start, stop, etc.).

112 / 134

Memory mode
HEADER MEMSpeed
MEMSpeed:EXT MEMSpeed ? MEMBloc
MEMBloc ? POSTrig
POSTrig ? MEM:CONT
MEM:CONT ? FILE:NAMe
FILE:NAMe ? FILE:LENGth
FILE:LENGth ?

DAS60 User’s Manual

PARAMETERS P1,P2 Set-up of the sampling period P1 = Period (from 1 to 500) P2 = MICro,MILIsec,Sec,MIn,HOurs is the unit Use of an external clock Returns the data acquisition speed P1 Set-up of the number of blocks P1 = 1,2,4,8,16…128 Returns the number of blocks and the validation of each block P1,P2 Set-up of the triggering position in the data acquisition
P1 = varies from -100 to +100 in % P2 = ON or OFF : inhibition of the trigger during the pretrigger Returns the position of the trigger P1,P2 Set-up of the aftermath P1 = PLot ,NOPlot P2 = FIle ,NOFile : save to a file
P1,P2 Name of the save file P1 = BINary, TEXTe file format P2 = name of the file (12 characters max.)
P1,P2 Limit of the number of samples P1 = from 0 to 1000 (0: no limit) P2 = KSample or Msample Returns the limit to the length of the file

EXAMPLES MEMSpeed 10,MICro 10 µsec period
MEMBLOC 4 4 blocks :MEMBLOC 4,2 : 4 blocks, incl. 2 valid ones :STOP:AUTO;POSTRIG 0 Acquisition after triggering
:FILE:NAME BIN,”FileO”;LENG LIM;LENG:LIMIT 10,MS

113 / 134

DAS60 User’s Manual

Relaunchs, real time save

HEADER REARm
REARm:SETup

PARAMETERS P1 définition du réarmement manuel P1 = SINgle, AUTo, SETup P1

REARm ? SAVE
SAVE ? SAVE:MEM

Number of the set-up to change P1 = 1 to 15 Returns the type of relaunch P1 Real time recording P1 = NO, DISk or MEMOry NO : no record DISK : record to HD or USBKey MEMOry (direct mode only)
P1,P2 Set-up of the trigger for saving to memory in direct mode

P1 = DIRect, TRIG or MANual P2 = CONt,NOCont relaunch

EXAMPLES REARm SINGLE
REARM SETUP; REAR:SETUP 2 Go to configuration 2
SAVE DISK
SAVE MEM;SAVE:MEM TRIG,NOC;:TRIG:CHAN A2,S1,POS

Launching plots and data acquisitions

HEADER RECord

PARAMETERS P1 Start or stop of the plot (or of the data acquisition into memory)
P1= ON : Start OFF : Stop TRIG : forcing the trigger
TRIGREC : forcing the mix data acquisition trigger

EXAMPLES RECORD ON
In direct mode, the plot will be effective when the start condition is chieved; you may force the start trigger with RECORD TRIG, and the stop trigger with RECORD OFF.

RECord ? WRIte

Returns the command status and the percentage of data acquisition into memory P1 Printing a message on paper (for appliances with a printer)

WRITE ‘RECORDER’

Allows you to write an annotation while recording to disk

P1 = message (max. 50 characters) between brackets (“) or apostrophes (‘)

114 / 134

Diagrams
GRID
GRID ? GRID:LOG
GRID:LOG ? GRID:LENGth
GRID:LENGth ? GRID:CHAnnel
GRID:CHAnnel ? COLOR
DEFLOG

DAS60 User’s Manual

P1,P2

GRID:LOG 50,5,UP;:GRID

Set-up of the diagrams

2,SEPLOGON

P1 = Number of the diagram

The logic channels are at the

P2 = SEPLOGON or SEPLOGOFF : separate logic channels top and are 50mm tall on 2 100mm screens

Returns the set-up of all diagrams
P1,P2,P3 Set-up of the diagrams for logic channels P1 = Number of logic channels P2 = Heigth of the logic channels P3 = UP or DOWN : position of the logic channels Returns the set-up of all diagrams P1,P2,P3 Set-up of each diagram
P1 = number of the diagram P2 = min value (0 to max); max = 250 or 200 according to the appliance
P3 = valeur max (0 to max) Returns the set-up of all diagrams P1,P2,P3 Set-up of the position of a channel P1 = Number of the channel P2 = Number of the diagram: from 1 to max P3 = Thickness of the line: 1 to 8 Returns the set-up of the selected channel P1,P2,P3
Color of each channel P1 = value of red (from 0 to 100) P2 = value of green P3 = value of blue P1,P2,P3,P4,P5 Set-up of the logic channels P1 = Number of the logic channel from 1 to 16 P2 = value of red (from 0 to 100) P3 = value of green P4 = value of blue

GRID:LENG 1,0,100 Diagram 1 from 0 to 100mm
1 GRID:CHA 4,3,2 Channel 4 in the diagram #3 with line thickness 2
CHAN 2,COLOR 100,100,100

115 / 134

Direct display
SCREEN SCREEN:FT
SCREEN:XY
SCREEN:TIMEBASE
SCREEN:RUN SCREEN:RUN ? SCREEN:TRIG

DAS60 User’s Manual

P1 Set-up of the display m ode P1 = FT, TEXT or XY P1,P2,P3 Set-up in F(T) P1:VER or HOR for vertical or horizontal P2:BOUNON or BOUNOFF to display the boundaries P3:FULLON or FULLOFF for full-screen display or not P1,P2 Set-up in Y

SCREEN FT
PAGE SCOPE;SCREEN FT;:SCREEN VER,BOUNON,FULLON vertical full-screen display
SCREEN:XY 3,2

P1 = X channel of 1,2, etc. P2 = CHANNEL Y turns to ALL for all channels ON or 1,2 for

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals