NATIONAL INSTRUMENTS PXI-5650 Analog Signal Generator Installation Guide
- June 12, 2024
- NATIONAL INSTRUMENTS
Table of Contents
NATIONAL INSTRUMENTS PXI-5650 Analog Signal Generator
Product Information: PXI-5650
The PXI-5650 is an I/Q modulator manufactured by NI (National Instruments). It is designed to provide high-performance modulation capabilities for various applications.
Product Usage Instructions
Initial Setup:
Before starting the calibration procedure, ensure that you have installed
the NI-RFSG software version 1.6 or later on your calibration system. This
software includes all the necessary VIs (Virtual Instruments) for calibration.
If you don’t have the latest version, you can download it from ni.com/idnet.
Verification:
Refer to the user manual for detailed instructions on how to perform the
calibration procedure for the NI PXIe-5611 I/Q modulator. The manual provides
performance tests to verify if the instrument is performing within the
published specifications. Make sure to follow the instructions carefully to
ensure accurate calibration results.
Password:
If prompted for a password during the calibration procedure, please refer
to the user manual or contact NI support for assistance.
Note: If you are using the NI PXIe-5673 or NI PXIe-5673E, please refer to their respective calibration procedures available at ni.com/manuals.
PXI-5650
CALIBRATION PROCEDURE
NI PXIe-5611
Conventions
bold italic monospace monospace bold monospace italic
The following conventions are used in this document:
The » symbol leads you through nested menu items and dialog box options to a
final action. The sequence File»Page Setup»Options directs you to pull down
the File menu, select the Page Setup item, and select Options from the last
dialog box.
This icon denotes a tip, which alerts you to advisory information.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid
injury, data loss, or a system crash.
Bold text denotes items that you must select or click in the software, such as
menu items and dialog box options. Bold text also denotes parameter names.
Italic text denotes variables, emphasis, a cross-reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a
word or value that you must supply.
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This
font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames, and extensions.
Bold text in this font denotes the messages and responses that the computer
automatically prints to the screen. This font also emphasizes lines of code
that are different from the other examples.
Italic text in this font denotes text that is a placeholder for a word or
value that you must supply.
Software Requirements
To calibrate the NI PXIe-5611, you must install NI-RFSG version 1.6 or later
on the calibration system. NI-RFSG includes all the VIs necessary for
calibration. You can download the latest version of NI-RFSG at ni.com/idnet.
NI-RFSG supports programming the calibration procedures in LabVIEW.
Calibration VIs are accessible in LabVIEW from the Functions palette. Refer to
Table 1 for file locations.
NI PXIe-5611 Calibration Procedure
2
ni.com
For more information about calibration VIs, refer to the NI RF Signal Generators Help, accessible at Start»All Programs»National Instruments»NI- RFSG»Documentation»NI RF Signal Generators Help.
Table 1. Calibration File Locations (NI-RFSG 1.6 or Later)
File Name and Location
Description
IVIBinniRFSG.dll
NI-RFSG driver containing the entire NI-RFSG API, including calibration
functions.
Folder containing NI-RFSG calibration API VIs. You can access calibration
functions from the NI-RFSG calibration section of the LabVIEW function
palette.
Documentation Requirements
You might find the following documentation helpful as you write the
calibration procedure: · NI PXIe-5673 Specifications · NI PXIe-5673E
Specifications · NI PXIe-5673 Calibration Procedure · NI PXIe-5673E
Calibration Procedure · NI 5450 Calibration Procedure · NI PXI-5650/5651/5652
Calibration Procedure · NI PXIe-5650/5651/5652 Calibration Procedure · NI RF
Signal Generators Getting Started Guide · NI RF Signal Generators Help,
including LabVIEW VI programming
references
These documents are installed with NI-RFSG. You can also download the latest
versions of documentation at ni.com/manuals.
This calibration procedure calibrates the NI PXIe-5611 I/Q modulator as a
stand-alone device. To calibrate the NI PXIe-5673 or NI PXIe-5673E, refer to
their calibration procedures at ni.com/manuals.
© National Instruments Corporation
3
NI PXIe-5611 Calibration Procedure
Password
The default password for password-protected operations is NI.
Calibration Interval
The measurement accuracy requirements of your application determine how often
you should calibrate your device. NI recommends that you perform a complete
calibration for the NI 5611 at least once a year. You can shorten this
calibration interval based on the accuracy demands of your application.
Calibration Overview
Calibration involves verification, and, if necessary, adjustment, and
reverification of the NI 5611.
Verification is the process of testing to ensure that the accuracy of the
device. Perform verification post-adjustment to determine if the adjustment
was successful.
Adjustment is the process of measuring and compensating for device performance
to improve the measurement accuracy. Performing an adjustment updates the
calibration date.
Perform a complete calibration to guarantee successful adjustment and
performance of the NI 5611 for a one-year calibration interval. After
adjustment, repeat verification to ensure that the device meets the
calibration test limits.
After calibrating the NI 5611 as a stand-alone device, NI strongly recommends
that you calibrate the NI 5673, which includes the NI 5611. To calibrate the
NI PXIe-5673 or NI PXIe-5673E, refer to their calibration procedures at
ni.com/manuals.
NI PXIe-5611 Calibration Procedure
4
ni.com
Figure 1 shows the programming flow for a complete calibration.
Verify
Adjust (Calibration Dates and Temperatures Updated)
Verify
Document Pre-Adjustment
Results
Document Post-Adjustment
Results
Yes
Meets
No
Test Limits?
Calibration/ Verification Complete
Review Verification/ Adjustment Procedure
or Return Device
Figure 1. Complete Calibration Programming Flow
© National Instruments Corporation
5
NI PXIe-5611 Calibration Procedure
Test Equipment
Table 2 lists the equipment required to calibrate the NI 5611. If you do not have the recommended equipment, select a substitute calibration standard using the specifications listed in Table 2.
Table 2. Required Equipment Specifications for NI 5611 Calibration
Required Equipment Spectrum analyzer
Power meter
Power splitter
NI PXIe-5450 arbitrary waveform generator NI PXI-5652 RF signal generator 6 dB
attenuator
50 terminator (included in the NI 5673 kit) Adapter SMA(m)-to-N(f) SMA torque
wrench NI PXIe-5673 Cable Accessory Kit
Recommended Equipment
Specifications
Rohde & Schwarz (R&S) FSU Spectrum Analyzer with high-frequency preamplifier option (B23)
Frequency range: 50 MHz to 6.6 GHz Noise floor: <152 dBm/Hz to 6.6 GHz
Anritsu ML2438A and a MA2472 diode sensor
Range: 50 dBm to +10 dBm Frequency range: 50 MHz to 6.6 GHz Accuracy: 0.5%
Aeroflex/Weinschel 1593 SWR: 1.25 Amplitude tracking: <0.25 dB
NI PXIe-5450
This device must be fully calibrated before calibrating the NI 5611.
NI PXI-5652 Anritsu 41KB-6 NI 778353-01
This device must be fully calibrated before calibrating the NI 5611.
Frequency range: DC to 12 GHz SWR: 1.1
—
S.M. Electronics SM4241 VSWR: 1.15 : 1
NI 187106-01 NI 780567-01
1 N · m
Matched-length I/Q semi rigid SMA cable (4) LO semi-rigid SMA cable (1) RF
flexible SMA cable (1)
SMA adapter plug/plug
Huber+Suhner
VSWR: 1.05 : 1
32_N-SMA-50-1/11-_NE
BNC (m) to BNC
—
—
(m) cable (36 inches)
NI PXIe-5611 Calibration Procedure
6
ni.com
Test Conditions
Follow these guidelines to optimize the connections and the environment during
calibration:
· Verify that the NI 5611, NI 5450, and NI 5652 are properly connected as
indicated in Chapter 5, NI 5673 RF Vector Signal Generator or Chapter 6, NI
5673E RF Vector Signal Generator in the NI RF Signal Generators Getting
Started Guide.
· Verification limits are defined assuming the same NI 5611, NI 5450, and NI
5652 is used during adjustment and verification.
· Keep cabling as short as possible. Long cables act as antennae, picking up
extra noise that can affect measurements.
· Verify that all connections, including front panel connections and screws,
are secure.
· Ensure that the PXI Express chassis fan speed is set to HI, the fan filters
are clean, and the empty slots contain filler panels. For more information,
refer to the Maintain Forced-Air Cooling Note to Users document available at
ni.com/manuals.
· Keep relative humidity between 10% and 90%, noncondensing.
· Maintain an ambient temperature of 23 °C ±5 °C.
· Allow a warm-up time of at least 30 minutes after the chassis is powered on.
The warm-up time ensures that the NI 5611 is at a stable operating
temperature.
· Perform a self-calibration on the NI PXIe-5450 signal generator.
· Each verification procedure requires a delay to be inserted in between
configuring all devices and acquiring the measurement. This delay may need to
be adjusted depending on the instruments used but should always be at least
1,000 ms for the first iteration, 1,000 ms when the power level changes, and
100 ms for each other iteration.
Calibration Procedures
The calibration process includes the following steps:
1. Initial Setup–Install the device and configure it in Measurement &
Automation Explorer (MAX). Refer to Chapter 5, NI 5673 RF Vector Signal
Generator or Chapter 6, NI 5673E RF Vector Signal Generator in the NI RF
Signal Generators Getting Started Guide for more information about how to
configure your device.
2. Verification–Verify the existing operation of the device. This step
confirms whether the device is operating within its specified range prior to
adjustment.
© National Instruments Corporation
7
NI PXIe-5611 Calibration Procedure
3. Adjustment–Perform an external adjustment of the device that adjusts the
calibration constants of the device. The adjustment procedure automatically
stores the calibration date on the EEPROM to allow traceability.
4. Reverification–Repeat the verification procedure to ensure that the device
is operating within its specifications after adjustment.
These procedures are described in more detail in the following sections.
Initial Setup
Refer to the NI RF Signal Generators Getting Started Guide for information about how to install the software and hardware and how to configure the device in MAX.
Characterizing the Power Splitter
Several procedures in this document require using a splitter that has been
characterized to remove error from future measurements. Complete the following
steps to characterize a splitter using an NI 5652, power meter, and spectrum
analyzer, as shown in Figure 2.
1. Disconnect the NI 5611 LO IN front panel connector from the NI 5652 RF OUT
front panel connector.
2. Connect the power meter power sensor to the NI 5652 RF OUT front panel
connector.
3. Generate a tone with the NI 5652 with the following NI-RFSG property
settings:
· Frequency (Hz): 50 MHz
· Power Level (dBm): 10 dBm
4. Use the power meter to measure the output power.
5. Repeat steps 3 through 4 for every frequency from 50 MHz to 6.6 GHz in 5
MHz increments, including endpoints. Store the resulting measurements.
6. Disconnect the power sensor from the NI 5652 RF OUT front panel connector.
Caution If you use a cable, it should be as short as possible. The cable should always be used with the splitter for subsequent verification procedures in this document that require a splitter.
7. Connect the power splitter input port to the NI 5652 RF OUT front panel connector without a cable, if possible.
NI PXIe-5611 Calibration Procedure
8
ni.com
Caution Remember the port to which the power sensor is connected. The power
sensor must be connected to the same port for subsequent procedures in this
document.
8. Connect one available port of the power splitter to the power sensor
without a cable, if possible.
9. Connect one available end of the power splitter to the 6 dB attenuator.
10. Connect the other available port of the 6 dB attenuator to the spectrum
analyzer RF INPUT front panel connector as shown in Figure 2. 11. Configure
the reference level of the spectrum analyzer to 30 dBm.
The spectrum analyzer is used only for termination.
© National Instruments Corporation
9
NI PXIe-5611 Calibration Procedure
1
2
3
5 6
7
4
8
1 NI 5652 RF Signal Generator 2 Splitter 3 Power Sensor 4 Attenuator
5 Power Meter 6 CH A Connector 7 Spectrum Analyzer 8 RF INPUT Connector
Figure 2. Power Splitter Characterization Setup
12. Repeat steps 3 through 4 for every frequency from 50 MHz to 6.6 GHz in 5
MHz increments, including endpoints. Store the resulting measurements.
13. Subtract the power measurements from step 5 from the measurements in step
12 for each corresponding frequency. Store the resulting calculations. The
result is an array representing the loss through the splitter for the entire
needed frequency range in 5 MHz increments.
NI PXIe-5611 Calibration Procedure
10
ni.com
Note This array is later used to compensate measurements and correct for splitter loss. If you do not find the exact frequency needed in the splitter correction array, use the loss corresponding to the next closest frequency.
You have successfully characterized your splitter.
Verification
This section describes the steps you must follow to verify successful
adjustment of the NI 5611.
Verification tests the following NI 5611 specifications: · LO filter response
· LO path gain · Modulation impairments · Modulation bandwidth and impairments
· RF path gain · Intermodulation products
Verification of the NI 5611 is complete only after you have successfully
completed all tests in this section.
Verifying LO Filter Response
Complete the following steps to verify the LO filter response of an NI 5611
module using a spectrum analyzer. 1. Connect the NI 5611 LO OUT front panel
connector to the spectrum
analyzer RF INPUT front panel connector. 2. Connect the PXI Express chassis 10
MHz REF IN rear panel connector
to the spectrum analyzer REF OUT rear panel connector. 3. Generate a signal
with the following NI-RFSG property settings:
· Frequency (Hz): 85 MHz · Power Level (dBm): 0 dBm · Generation Mode: CW · LO
OUT Enabled: Enabled 4. Use the spectrum analyzer to measure the mean
fundamental tone power using the following spectrum analyzer parameter
settings: · Center Frequency: Frequency from step 3 · Reference level: +5 dBm
· Frequency span: 0 Hz
© National Instruments Corporation
11
NI PXIe-5611 Calibration Procedure
· Resolution bandwidth: 500 Hz
· Sweep time: 5 ms
5. Use the spectrum analyzer to measure the mean tone power at the 2nd
harmonic frequency using the following spectrum analyzer parameter settings:
· Center Frequency: Frequency from step 3 × 2 · Reference level: +5 dBm
· Frequency span: 0 Hz
· Resolution bandwidth: 500 Hz
· Sweep time: 5 ms
6. Calculate the power level of the harmonic in dBc by subtracting the
measured fundamental power from step 4 from the measured harmonic power from
step 5.
7. Compare the calculated power level from step 6 to the corresponding limit
in Table 1.
8. Repeat steps 3 through 7 for every frequency from 85 MHz to 2.2 GHz in 5
MHz increments, including endpoints.
Table 3. LO Filter Response Verification Test Parameters
Harmonic 2nd (2F)
2nd Harmonic Upper Limits (dBc)
F 100 MHz 100 MHz < F 2.2 GHz
21
31
If the results are within the selected test limit, the device has passed this portion of the verification.
Verifying LO Path Gain
Complete the following steps to verify the LO path gain of the NI 5611 module
using a power meter. 1. Connect the NI 5611 LO OUT front panel connector to
the power
meter. 2. Generate a signal with the following NI-RFSG property settings:
· Frequency (Hz): 85 MHz · Power Level (dBm): 0 dBm · Generation Mode: CW · LO
OUT Enabled: Enabled 3. Use the power meter to measure the LO OUT power.
NI PXIe-5611 Calibration Procedure
12
ni.com
4. Compare the measured LO OUT output power to the limit in Table 4.
5. Repeat steps 2 through 4 for every frequency from 85 MHz to 6.6 GHz in 10
MHz increments, including endpoints.
Table 4. LO Path Gain Verification Upper Test Limits
LO OUT Power (dBm) 0
As Found LO OUT Power Accuracy Test Limits (dB)
±2.0
As Left LO OUT Power Accuracy Test Limits (dB)
±1.0
If the results are within the selected test limit, the device has passed this portion of the verification.
Verifying Modulation Impairments
Complete the following steps to verify the modulation impairments of the NI
5611 using a spectrum analyzer. 1. Connect the NI 5611 RF OUT front panel
connector to the spectrum
analyzer RF INPUT front panel connector. 2. Connect the PXI Express chassis 10
MHz REF IN rear panel connector
to the spectrum analyzer 10 MHz REF OUT rear panel connector on the spectrum
analyzer. 3. Generate a single-sideband tone with a +1 MHz offset from the
carrier signal with the following NI-RFSG property settings: · Frequency (Hz):
85 MHz · Power Level (dBm): 0 dBm · Generation Mode: Arb Waveform · Reference
Clock Source: ClkIn The image appears at 84 MHz, and the carrier leakage at 85
MHz. 4. Use the spectrum analyzer to measure the mean power of the RF output
using the following spectrum analyzer parameter settings: · Center frequency:
Frequency in step 3 + 1 MHz · Reference level: +5 dBm · Frequency span: 0 Hz ·
Resolution bandwidth: 500 Hz · Sweep time: 5 ms · Reference clock source:
External
Note Zero-span mode shortens test time by avoiding unnecessary frequency sweeping. To obtain the average power, convert the zero-span trace data to linear volts, perform a mean calculation, and then convert the linear volts back to dBm.
© National Instruments Corporation
13
NI PXIe-5611 Calibration Procedure
5. Use the spectrum analyzer to measure the mean power of the image using the
following spectrum analyzer parameter settings: · Center frequency: Frequency
in step 3 1 MHz · Reference level: +5 dBm · Frequency span: 0 Hz ·
Resolution bandwidth: 500 Hz · Sweep time: 5 ms · Reference clock source:
External
6. Use the spectrum analyzer to measure the mean power of the carrier using
the following spectrum analyzer parameter settings: · Center frequency:
Equivalent to the frequency in step 3 · Reference level: +5 dBm · Frequency
span: 0 Hz · Resolution bandwidth: 500 Hz · Sweep time: 5 ms · Reference clock
source: External
7. Calculate the image suppression ratio (ISR) and carrier suppression ratio
(CSR) according to the following formulas and save the values: ISR = Measured
Image Power Measured RF Output Power CSR = Measured Carrier Leakage Power
Measured RF Output Power
8. Repeat steps 3 though 7 for each LO frequency from 85 MHz to 6.6 GHz in 10
MHz steps, including endpoints.
Note With the baseband set to 1 MHz, 0 dBm single-sideband tone, the RF output is always 1 MHz above the LO frequency, the image is 1 MHz below LO, and the carrier leakage is at LO frequency.
9. Use Table 5 to compare the ISR and CSR from step 7 for each LO frequency.
Table 5. Modulation Impairments Verification Upper Test Limits
LO Frequency (MHz)
85 to 400
400 to 2,500
ISR (dBc) Test Limit
43
50
CSR (dBc) Test Limit
44
44
NI PXIe-5611 Calibration Procedure
14
ni.com
Table 5. Modulation Impairments Verification Upper Test Limits
LO Frequency (MHz)
ISR (dBc) Test Limit
CSR (dBc) Test Limit
2,500 to 5,500
46
44
5,500 to 6,600
43
41
If the results are within the selected test limit, the device has passed this portion of the verification.
Verifying Modulation Bandwidth and Impairments
Complete the following steps to verify the modulation bandwidth performance
and modulation impairments of the NI 5611 using a spectrum analyzer. 1.
Connect the NI 5611 RF OUT front panel connector to the spectrum
analyzer RF INPUT front panel connector. 2. Connect the PXI Express chassis 10
MHz REF IN rear panel connector
to the spectrum analyzer 10 MHz REF OUT rear panel connector on the spectrum
analyzer. 3. Generate an I/Q tone at 50 MHz offset from the carrier with the
following NI-RFSG property settings: · Frequency (Hz): 200 MHz · Power Level
(dBm): 0 dBm · Generation Mode: Arb Waveform · Reference Clock Source: ClkIn
4. Use the spectrum analyzer to measure the mean output power of the RF output
using the following spectrum analyzer parameter settings: · Center frequency:
frequency in step 3 + offset in step 3 · Reference level: +5 dBm · Frequency
span: 0 Hz · Resolution bandwidth: 500 Hz · Sweep time: 5 ms · Reference clock
source: External 5. Use the spectrum analyzer to measure the mean output power
of image using the following spectrum analyzer parameter settings: · Center
frequency: frequency in step 3 offset in step 3 · Reference level: +5 dBm ·
Frequency span: 0 Hz · Resolution bandwidth: 500 Hz
© National Instruments Corporation
15
NI PXIe-5611 Calibration Procedure
· Sweep time: 5 ms · Reference clock source: External 6. Calculate the ISR
according to the following formula and save the values: ISR = Measured Image
Power Measured RF Output Power
7. Repeat steps 3 through 6 for every offset frequency from 50 MHz to +50
MHz in 5 MHz increments while keeping the LO frequency fixed.
Tip Set the NI-RFSG Generation Mode property to CW to generate a 0 Hz offset. The ISR with a 0 Hz frequency offset cannot be measured, and the 0 Hz data point can be ignored.
8. Calculate the relative modulation bandwidth flatness by subtracting the RF
output power measured at 0 Hz offset from the RF output powers measured at
each of the other offset frequencies.
9. Compare each value calculated in step 8 to the limits in Table 6 and
compare only the ISR values calculated for offset frequencies between 10 MHz
and 10 MHz in step 5 to the limits in Table 7.
Note Values calculated for offset frequencies less than 10 MHz and greater than 10 MHz are not specified and do not have test limits.
10. Repeat steps 3 through 9 for each LO frequency in Table 6.
Table 6. Modulation Bandwidth Impairment Verification Test Limits
LO Frequency (GHz)
0.2
Modulation Bandwidth Test Limits (dB)
Maximum
Minimum
3
3
2.4
3
3
4.0
3
3
5.8
3
3
NI PXIe-5611 Calibration Procedure
16
ni.com
Table 7. Modulation Bandwidth Impairment Verification Test Limits for Offset Frequencies Between 10 MHz and +10 MHz
LO Frequency (GHz)
Test Limits ISR (dBc)
0.2
41
2.4
48
4.0
45
5.8
41
If the results are within the selected test limit, the device has passed this portion of the verification.
Verifying RF Path Gain
Complete the following steps to verify the RF path gain of the NI 5611 using a
power meter, rubidium frequency standard, spectrum analyzer, power splitter,
and 6 dB attenuator.
Note The attenuator is placed in front of the spectrum analyzer to improve the spectrum analyzer return loss. The power splitter must be characterized so that you can account for its loss. Refer to the Characterizing the Power Splitter section for more information about how to characterize your power splitter.
© National Instruments Corporation
17
NI PXIe-5611 Calibration Procedure
5
1
34
2
6
7
8 10
9 2
11
1 NI 5450 Arbitrary Waveform Generator 2 Connection from NI PXI Express
REF IN Rear Panel Connector to Spectrum Analyzer REF OUT Rear Panel Connector
3 NI 5611 I/Q Modulator 4 NI 5652 RF Signal Generator 5 Power Sensor
6 Power Splitter 7 Attenuator 8 Power Meter 9 Spectrum Analyzer 10 CH A Connector 11 RF INPUT Connector
Figure 3. Connecting Hardware to Verify RF Path Gain
1. Connect the input port of the power splitter to the NI 5611 RF OUT front
panel connector.
2. Connect the PXI Express chassis 10 MHz REF IN rear panel connector to the
spectrum analyzer 10 MHz REF OUT rear panel connector.
Caution The port used to connect the power splitter to the power sensor in step 8 of the Characterizing the Power Splitter section must be the same port used to connect the power splitter to the power meter in step 3 of this section.
NI PXIe-5611 Calibration Procedure
18
ni.com
3. Connect one available port of the power splitter to the power meter.
Caution The port used to connect the power splitter to the 6 dB attenuator in
step 9 of the Characterizing the Power Splitter section must be the same port
used to connect the power splitter to the power meter in step 4 of this
section.
4. Connect one available port of the power splitter to the 6 dB attenuator.
5. Connect one available port of the 6 dB attenuator to the spectrum
analyzer RF INPUT front panel connector. 6. Disconnect the NI 5450 CLK IN
front panel connector from the
NI 5652 REF IN/OUT front panel connector. 7. Generate a single-sideband tone
with a +1 MHz offset from the carrier
with the following NI-RFSG property settings: · Frequency (Hz): 85 MHz · Power
Level (dBm): 5 dBm · Generation Mode: Arb Waveform · Reference Clock Source:
ClkIn 8. Use the power meter to measure power splitter-compensated output
power. 9. Repeat steps 7 through 8 for every frequency from 85 MHz to 6.6 GHz
in 20 MHz increments, including endpoints. 10. Set the NI-RFSG Power Level
property to 0 dBm, and repeat steps 7 through 9. 11. Set the NI-RFSG Power
Level property to 30 dBm, and repeat steps 7 through 9, measuring the power
using both the power meter and spectrum analyzer. Use the following parameter
settings for the spectrum analyzer: · Center frequency: frequency in step 7 +
1 MHz · Reference level: 30 dBm · Frequency span: 0 Hz · Resolution
bandwidth: 500 Hz · Sweep time: 20 ms · Reference clock source: External ·
Attenuation: 5 dB · Preamplifier: On 12. Use the power meter measurements to
calculate a correction for the spectrum analyzer inaccuracies and attenuator
loss.
© National Instruments Corporation
19
NI PXIe-5611 Calibration Procedure
13. Set the NI-RFSG Power Level property to 60 dBm, and repeat steps 7
through 9 using only the spectrum analyzer and its measured correction to
measure the power accuracy using the following spectrum analyzer parameter
settings: · Center frequency: frequency in step 7 + 1 MHz · Reference level:
30 dBm · Frequency span: 0 Hz · Resolution bandwidth: 500 Hz · Sweep time: 20
ms · Reference clock source: External · Attenuation: 5 dB · Preamplifier: On
14. Set the NI-RFSG Power Level property to 90 dBm and repeat steps 7
through 9 using only the spectrum analyzer and its measured correction to
measure the power accuracy using the following spectrum analyzer parameter
settings: · Center frequency: frequency in step 7 + 1 MHz · Reference level:
30 dBm · Frequency span: 0 Hz · Resolution bandwidth: 500 Hz · Sweep time:
200 ms · Reference clock source: External · Attenuation: 5 dB · Preamplifier:
On
15. Compare the measured values from steps 7 through 14 to the corresponding
limits in Table 8.
Table 8. RF Path Gain Verification Test Limits
Output Power (dBm) 5
Power Accuracy Test Limits (dB) ±0.75
0
±0.75
30
±0.75
60
±0.75
90
±0.75
If the results are within the selected test limit, the device has passed this portion of the verification.
NI PXIe-5611 Calibration Procedure
20
ni.com
Adjustment
Following the adjustment procedure automatically updates the calibration date
and temperature in the EEPROM of the NI 5611. The calibration date is also
updated when you call the niRFSG Initialize External Calibration and niRFSG
Close External Calibration VIs.
Adjustment involves the following NI 5611 procedures: · LO filter response ·
LO path gain · Modulation impairments · RF path gain
Adjusting LO Filter Response
This adjustment yields a reduced LO second harmonic. Complete the following
steps to adjust the LO filter response of the NI 5611 using a spectrum
analyzer.
1. Connect the NI 5611 LO OUT front panel connector to the spectrum analyzer
RF INPUT front panel connector.
2. Connect the PXI Express chassis 10 MHz REF IN rear panel connector to the
spectrum analyzer 10 MHz REF OUT rear panel connector on the spectrum
analyzer.
3. Call the niRFSG Initialize External Calibration VI.
4. Call the niRFSG Initialize LO Filter Calibration VI.
Note The Filter to Calibrate parameter on the niRFSG Initialize LO Filter Calibration VI is by default set to All Filters. If you set the Filter to Calibrate parameter to another filter setting, you must repeat steps 4 through 8 for each filter.
5. Call the niRFSG 5611 Configure LO Filter Calibration VI. 6. Use the
spectrum analyzer to measure the mean output power using the
following spectrum analyzer parameter settings: · Center frequency: frequency
specified by the Frequency to
Measure (Hz) parameter of the niRFSG 5611 Configure LO Filter Calibration VI ·
Reference level: +5 dBm · Frequency span: 0 Hz · Resolution bandwidth: 500 Hz
· Sweep time: 5 ms
© National Instruments Corporation
21
NI PXIe-5611 Calibration Procedure
7. Call the niRFSG 5611 Adjust LO Filter Calibration VI, and pass the value
measured in step 6 to the Measured LO OUT Power (dBm) parameter.
8. Repeat steps 5 through 7 in a loop until the LO Filter Calibration
Complete parameter returns a value of TRUE.
9. Call the niRFSG Close External Calibration VI to close the session and
store the results to the EEPROM on the NI 5611.
Adjusting LO Path Gain
This adjustment yields accurate power levels reaching both the internal
modulators and the LO OUT connector on the NI 5611. Complete the following
steps to calibrate the LO path gain on the NI 5611 using a power meter. 1.
Connect the NI 5611 LO OUT front panel connector to the power
meter. 2. Call the niRFSG Initialize External Calibration VI. 3. Call the
niRFSG 5611 Initialize LO Gain Calibration VI.
Note The Filter to Calibrate parameter on the niRFSG Initialize LO Gain
Calibration VI is by default set to All Filters. If you set the Filter to
Calibrate parameter to another filter setting, you must repeat steps 3 through
7 for each filter.
4. Call the niRFSG 5611 Configure LO Gain Calibration VI. 5. Use the power
meter to measure the output power at the frequency
specified by the Frequency to Measure (Hz) parameter of the niRFSG 5611
Configure LO Gain Calibration VI. 6. Call the niRFSG 5611 Adjust LO Gain
Calibration VI, and pass the value measured in step 5 to the Measured LO OUT
Power (dBm) parameter. 7. Repeat steps 4 through 6 until the LO Gain
Calibration Complete parameter returns a value of TRUE. 8. Call the niRFSG
Close External Calibration VI to close the session and store the results to
the EEPROM on the NI 5611.
NI PXIe-5611 Calibration Procedure
22
ni.com
Adjusting Modulation Impairments
This adjustment measures and corrects the NI 5611 RF fundamental, carrier, and
image tones until the carrier and image tones are sufficiently suppressed.
Complete the following steps to calibrate the modulation impairments of the NI
5611 using a spectrum analyzer. 1. Connect the NI 5611 RF OUT front panel
connector to the spectrum
analyzer RF INPUT front panel connector. 2. Connect the PXI Express chassis 10
MHz REF IN rear panel connector
to the spectrum analyzer 10 MHz REF OUT rear panel connector. 3. Call the
niRFSG Initialize External Calibration VI. 4. Call the niRFSG 5611 Initialize
Impairment Calibration VI. 5. Call the niRFSG 5611 Configure Impairment
Calibration VI.
Note The default behavior of the niRFSG 5611 Configure Impairment Calibration
VI is to calibrate all filters. If you choose to calibrate a single filter,
repeat the entire RF path calibration procedure for each of the eight filters
to complete a full calibration.
6. Use the spectrum analyzer to measure the mean tone power using the
following spectrum analyzer parameter settings: · Center frequency: Value
specified by the Frequency to Measure (Hz) parameter of the niRFSG 5611
Configure LO Filter Calibration VI · Reference level: Value specified by the
Spectrum Analyzer Reference Level (dB) parameter of the niRFSG 5611 Configure
Impairment Calibration VI · Frequency span: 0 Hz · Resolution bandwidth: 500
Hz · Sweep time: 5 ms
7. Call the niRFSG 5611 Adjust Impairment Calibration VI, and pass the
measurement from step 6 to the Measured RF OUT Power (dBm) parameter of the
niRFSG 5611 Adjust RF Gain Calibration VI.
8. Repeat steps 5 through 7 until the Impairment Calibration Complete
parameter returns a value of TRUE.
9. Call the niRFSG Close External Calibration VI to close the session and
store the results to the EEPROM on the NI 5611.
© National Instruments Corporation
23
NI PXIe-5611 Calibration Procedure
Adjusting RF Path Gain
This adjustment yields accurate power levels out of the NI 5611 RF OUT front
panel connector. Complete the following steps to properly connect the devices,
as shown in Figure 4, to adjust RF path gain.
4
1
23
5
Reverse Pwr +30 dBm MAX
TTL
NI PXIe-5450 400 MS/s I/Q Signal Generator
NI PXIe-5611 I/Q Vector Modulator
ACCESS
ACTIVE
CLK IN
CH 0 +
ACCESS
ACTIVE
85 MHz 6.6 GHz RF OUT
I+ I+
5 Vp-p MAX, 50
CLK OUT
50
CH 0
PLS MOD I I
0.7 Vp-p NOM
PFI 0
50
CH 1 +
LO IN Q+ Q+
LVT T L
PFI 1
LVT T L
50
CH 1
LO OUT Q Q
50 ESD SENSITIVE
RF PORTS ±5 VDC MAX
+18 dBm MAX
+18 dBm MAX
7 8
9
6
10
1 NI 5450 Arbitrary Waveform Generator 2 NI 5611 I/Q Modulator 3 NI 5652 RF Signal Generator 4 Power Sensor 5 Power Splitter
6 Attenuator 7 Power Meter 8 CH A Connector 9 Spectrum Analyzer 10 RF INPUT Connector
Figure 4. Connecting Hardware to Adjust RF Path Gain
1. Connect the input port of the power splitter to the NI 5611 RF OUT front panel connector.
NI PXIe-5611 Calibration Procedure
24
ni.com
2. Connect the PXI Express chassis 10 MHz REF IN rear panel connector to the
spectrum analyzer 10 MHz REF OUT rear panel connector.
3. Connect one available port of the power splitter to the power meter. 4.
Connect one available port of the power splitter to the 6 dB attenuator. 5.
Connect one available port of the 6 dB attenuator to the spectrum
analyzer RF INPUT front panel connector.
Note For best results, do not use any cables between the NI 5611, power
splitter, and power sensor. Connect these devices directly together.
6. Call the niRFSG Initialize External Calibration VI. 7. Call the niRFSG
5611 Initialize RF Gain Calibration VI. 8. Call the niRFSG 5611 Configure RF
Gain Calibration VI, and set the
Instrument Transition Threshold (dBm) parameter to the power level equal to
the bottom of the power meter’s accurate range.
Note The default behavior of the niRFSG 5611 Configure RF Gain Calibration VI
is to calibrate all filters. If you choose to calibrate a single filter,
repeat the entire RF path calibration procedure on each of the eight filters
to complete a full calibration.
9. Configure the spectrum analyzer using the following spectrum analyzer
parameter settings: · Center frequency: value specified by the Frequency to
Measure (Hz) parameter of the niRFSG 5611 Configure RF Gain Calibration VI ·
Reference level: value specified by the Spectrum Analyzer Reference Level (dB)
parameter of the niRFSG 5611 Configure Impairment Calibration VI · Frequency
span: 0 Hz · Resolution bandwidth: 500 Hz · Sweep time: 20 ms · Attenuation: 5
dB · Preamplifier: On
10. Using the instrument(s) specified by the Measurement Instrument parameter
of the niRFSG 5611 Configure Impairment Calibration VI, measure the mean power
of the generated tone at the frequency specified by the Frequency to Measure
(Hz) parameter of the niRFSG 5611 Configure RF Gain Calibration VI.
© National Instruments Corporation
25
NI PXIe-5611 Calibration Procedure
Note If the instrument specified is Power Meter & Spectrum Analyzer, then you
must measure the power with the power meter and the spectrum analyzer
simultaneously. These measurements are used to correct the spectrum analyzer’s
inaccuracy.
11. Compensate only the power meter measurement for the loss of the power
splitter by subtracting the loss measured in the Characterizing the Power
Splitter section from the power meter measurement in step 10 in this section.
12. Call the niRFSG 5611 Adjust RF Gain Calibration VI, and pass the
measurements from step 11 to the Power Meter Measurement (dBm) and/or Spectrum
Analyzer Measurement (dBm) parameter.
13. When the spectrum analyzer measures a level of 110 dBm, increase the
sweep time for each measurement to 200 ms. The sweep time can return to 20 ms
when the Measurement Instrument parameter of the niRFSG 5611 Configure RF Gain
Calibration VI returns a value of Power Meter. The sweep time should alternate
several times between 20 ms and 200 ms as these conditions are met.
Note Only the instrument indicated by the Measurement Instrument parameter is
actually used. Values passed for the inactive instrument are ignored. The
power meter is used for higher power measurements, and the spectrum analyzer
is used for lower power measurements.
14. Repeat steps 8 through 12 until the RF Gain Calibration Complete
parameter returns a value of TRUE. Additional averaging is required on the
spectrum analyzer to minimize the effects of noise at the lower power levels.
After a level of 90 dBm or less is measured, increase the sweep time to 200
ms for the remaining measurements of the procedure.
15. Call the niRFSG Close External Calibration VI to close the session and
store the results to the EEPROM on the NI 5611.
You have finished adjusting the NI 5611. Repeat the Verification section to
reverify the performance of the NI 5611 after adjustments.
NI PXIe-5611 Calibration Procedure
26
ni.com
Appendix A: Calibration Utilities
NI-RFSG provides a full complement of calibration utility VIs. You can use
utility VIs to retrieve information about adjustments performed on the NI
5611, change the calibration password, and store small amounts of information
in the onboard EEPROM. The Calibration Utility palette includes the following
VIs: · niRFSG Change External Calibration Password VI · niRFSG Close External
Calibration VI · niRFSG Get External Calibration Last Date And Time VI ·
niRFSG Initialize External Calibration VI · niRFSG Set External Calibration
Last Date and Time VI
Calibration VI References
The VIs used in this procedure, including all calibration VIs and functions,
are documented in the NI-RFSG VI Reference, which you can access from the NI
RF Signal Generators Help at Start»All Programs»National Instruments»NI-
RFSG»Documentation.
© National Instruments Corporation
27
NI PXIe-5611 Calibration Procedure
Where to Go for Support
The National Instruments Web site is your complete resource for technical
support. At ni.com/support you have access to everything from troubleshooting
and application development self-help resources to email and phone assistance
from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the Council
of the European Communities using the manufacturer’s declaration of
conformity. This system affords the user protection for electromagnetic
compatibility (EMC) and product safety. You can obtain the DoC for your
product by visiting ni.com/certification. If your product supports
calibration, you can obtain the calibration certificate for your product at
ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac
Expressway, Austin, Texas, 78759-3504. National Instruments also has offices
located around the world to help address your support needs. For telephone
support in the United States, create your service request at ni.com/support
and follow the calling instructions or dial 512 795 8248. For telephone
support outside the United States, contact your local branch office:
Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020,
Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech
Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511,
France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972
3 6393737, Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793,
Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90
76 60, Poland 48 22 328 90 10, Portugal 351 210 311 210, Russia 7 495 783
6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11
805 8197, Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56
2005151, Taiwan 886 02 2377 2222, Thailand 662 278 6777, Turkey 90 212 279
3031, United Kingdom 44 (0) 1635 523545
LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate
logo, and the Eagle logo are trademarks of National Instruments Corporation.
Refer to the Trademark Information at ni.com/trademarks for other National
Instruments trademarks. Other product and company names mentioned herein are
trademarks or trade names of their respective companies. For patents covering
National Instruments products/technology, refer to the appropriate location:
Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patent Notice at ni.com/patents.
© 20092010 National Instruments Corporation. All rights reserved.
375067B-01
Mar10
References
- Test and Measurement Systems, a part of Emerson - NI
- Test and Measurement Systems, a part of Emerson - NI
- Product Certifications - NI
- Software and Driver Downloads - NI
- Instrument Driver Network (IDNet) - NI
- Product Documentation - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- NI Trademarks and Logo Guidelines - NI
- Software and Driver Downloads - NI
- Product Certifications - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- PXI-5650 National Instruments Analog Signal Generator | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>