ChoiceMMed MD100C Card ECG Monitor User Manual

June 4, 2024
ChoiceMMed

Card ECG Monitor
User Manual
MD100C

ChoiceMMed MD100C Card ECG Monitor - Back1

Introduction

Thank you for purchasing Card ECG Monitor. Please read this manual carefully before using the device. Failure to follow the instructions may cause measurement abnormalities or equipment damage.
ChoiceMMed (the “Company”) assumes no responsibility for personal injury or device damage sustained by or through the use of this product. No part of these instructions may be copied, reproduced, or translated into another language without the prior written consent of the Company. The Company reserves the right to revise this guide at any time without prior notice.
Intended use
The Card ECG Monitor is a handheld non-invasive device intended for checking electrocardiogram (ECG) and heart rate and for use by healthcare professionals, patients with known or suspected heart conditions, and health- conscious individuals in hospitals, hospital-type facilities, and home care. It is not intended for pediatric use.
Contraindication
Contraindicated in persons with pacemakers.
Product Features

  • Pocket size, easy to carry
  • Support real-time and offline mode measurement

Precautions for Use

  1. Before use, carefully read the manual.
  2. The operation of the device may be affected by the use of an electrosurgical unit (ESU).
  3. Do not use the device in an MRI or CT environment.
  4. Do not use the device in situations where alarms are required. The device has no alarms. It is not for continuous monitoring.
  5. Do not use the device in an explosive atmosphere.
  6. Do not use the device with a cardiac pacemaker or defibrillator.
  7. Do not use the device near flammable anesthetics, or near pressurized oxygen such as in a hyperbaric chamber, ultraviolet sterilizer, or oxygen tent.
  8. Do not use the device near strong electromagnetic forces.
  9. Do not use the device in areas with extreme temperatures or humidity. Avoid using the device immediately after a significant change in temperature or humidity.
  10. Do not expose the device to static electricity. Disperse static electricity from your body before handling the device.
  11. Do not operate the device after it has been immersed in liquid.
  12. The device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.
  13. Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components.
  14. This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device.
  15. Portable and mobile RF communications equipment can affect medical electrical equipment. The portable and mobile RF communications equipment should be used no closer than 30cm (12 inches) to any part of the device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  16. This equipment is not intended for use during patient transport outside the healthcare facility. Do not take measurements in a moving vehicle.
  17. The patient is an intended operator. The patient can safely use all functions of the device.
  18. It may be unsafe to -use accessories, detachable parts, and materials not described in the instructions for use -interconnect this equipment with other equipment not described in the instructions for use -disassemble, repair, or modify the equipment
  19. The material that contact with the patient’s skin has passed the ISO10993-5 Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity.
  20. Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  21. The use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  22. Do not take measurements when there are drops of water on your skin, such as from sweat or after bathing.
  23. The manufacturer will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist the user’s appropriately trained personnel to repair those parts of the equipment designated by the manufacturer to be repairable.

Device description

Front

Back

Charging the device
Connect the device with the power adapter by the attached magnetic charging cable:

  1. Connect the standard USB plug to a power adapter and then contact the magnetic connector with the magnetic suction charging interface of the device.
  2. Plug the power adapter into a power outlet.

Notes:

  • The screen will display “Charging” in the charging process.
  • Low batteries may affect the HR measurement accuracy. Charge the device when the screen displays .
  • The device will turn off when the battery power is too low for normal operation.
  • Charge the battery every six months if the unit is not required for long periods of time.
  • The adapter should meet medical-grade regulations.

Using the Lanyard

  1. Thread the thinner end of the lanyard through the hanging hole.
  2. Thread the thicker end of the lanyard through the threaded end and tighten.

ChoiceMMed MD100C Card ECG Monitor - Lanyard

Warnings!

  • Keep the device away from young children. Small items such as the lanyard are choking hazards.
  • Please notice that the lanyard may cause strangulation due to excessive length.

Take a measurement

Follow these instructions carefully for use. Failure to comply with any of these instructions will result in an inaccurate measurement.
Power on

| Press the Power Button to turn the unit on. The display will show the current date and time first and then the ECG symbol
ChoiceMMed MD100C Card ECG Monitor - icon5
---|---

ECG measurement

ChoiceMMed MD100C Card ECG Monitor - Back1| 1. Place two fingers from each hand on the two electrodes (you may hold the device as the figure).
---|---
2. The device will start a 30s’ recording after a countdown of 3s. Hold still until it completes.
3. Read the data (The measurement will be saved automatically).
ChoiceMMed MD100C Card ECG Monitor - icon3

Real-time ECG measurement by APP & Data Transmission
Connect the unit with your choice APP via Bluetooth.
Real-time measurements will be transferred to the APP automatically. You can view the result analysis from the APP as well as historical records and trends.
You can upload offline measurements to the APP and view the result analysis.
Check records
Press the Power Button for 2 seconds to check measurement records. The latest 20 ECG measurements can be saved.
Power off
Press the Power Button until the battery indicator appears, then the device will power off.
Notes:

  • Do not take measurements with overly wet skin.
  • Do not take a recording if the electrodes are dirty. Clean them first.
  • Too dry skin will cause a weak ECG signal, please measure after moisturizing skin.
  • When measuring ECG, keep your fingers and arms in a relaxed position for the full duration of the measurement. If the device senses poor contact-the device will indicate “Leadoff” or if the measurement is interrupted by a hand adjustment, you will need to repeat the recording.
  • Make sure your left and right fingers are placed on the right electrodes for ECG measuring.
  • Make sure the electrodes are in direct contact with your skin. DO NOT press firmly.
  • The device will automatically shut off in 8 seconds with no operation.
  • After each use, carefully disinfect the contact electrodes only with a disinfectant swab or equivalent.

CAUTION!

  • The device is for monitoring purposes only. It is not intended to substitute for a hospital diagnostic device and should not be used as a basis for starting or modifying medical treatment. Do NOT make a diagnosis or treatment decisions by yourself according to the measurement and analysis results. Always consult your physician if you have any questions or if you believe you have abnormal measurements.
  • The values displayed by the device are the ones derived at the time of measurement. Medical conditions can change suddenly. If you notice any change in your condition, consult your physician, regardless of the measured results.
  • The device is not intended for simultaneously recording and transmitting a user’s Heart Rhythm.

maintenance and storage

  • Do not drop or hit this device.
  • Do not disassemble, repair, or modify the unit.
  • Do not expose the device to strong shocks or vibrations.
  • Keep out of reach of small children and people who cannot express their consent.
  • This device does not require calibration during the expected life cycle.
  • This device is designed to be compliant with the rules and regulations where it is sold and will be labeled as required.
  • Any change or modification to this device, not expressly approved by the manufacturer, will void the user’s authority to operate the equipment.
  • The life of the device is 5 years.
  • The environment temperature for transport or storage of the packaged device is -25°C ~70°C/-13°F~158°F, and the humidity is ≤93%, with no condensation.
  • Avoid extreme changes in temperature and humidity. Do not use this device in locations subject to high or low temperatures or humidity. Use it at a temperature within 5°C~40°C/ 41°F~104°F and ≤80% RH.
  • Do not store the device in ambient conditions: exposed to direct sunlight, with high temperatures and/or high humidity, that is wet or damp or where water may get on the device, that is dusty near fires or open flames, or exposed to strong vibration, or strong electromagnetic fields.

Clean and disinfect the device

  • The applied parts touching the patient’s body are required to be disinfected once after each use.
  • Use a medical disinfectant swab only to clean the device electrodes. Do not use water, or substances such as benzene, gasoline, paint thinner, concentrated alcohol, or detergents. Do not clean other parts of the device with any liquid.
  • Excessive disinfection may cause damage to the device and is therefore not recommended for this device unless otherwise indicated in your hospital’s servicing schedule.
  • Do not pour or spray liquids onto the device and do not allow any liquid to enter any openings in the device. Allow the device to dry thoroughly before reuse.
  • Do not sterilize this device in an autoclave, ultraviolet sterilizer, or gas sterilizer (EOG, formaldehyde, high-density ozone, etc.). The device is not intended for sterilization.

All the maintenance above can be performed by the patient.

Troubleshooting

Problem Cause Solution
The device can not be powered on 1. The battery power is exhausted.
2. The device might be broken. 1. Charge the device.

2. Please contact the local service center.
The device fails to measure heart rate| 1. The electrodes are not making good contact with your fingers.
2. The user does not keep still during the measurement.
3. Electromagnetic interference.
4. The signal is too weak.| 1. Place the electrode correctly.
2. Keep motionless and avoid moving when measuring.
3. keep away from electromagnetic interference.
4. Try to take another measurement.
Unstable ECG waveform| 1. The skin is dry.
2. The electrodes are not making good contact with your body.
3. Too nervous.| 1. If your skin is very dry, use a moisturizer.
2. Place the electrode correctly.
3. Please relax and keep calm.
The battery can not charge.| 1. No power supply.
2. The battery is damaged.| 1. Charge the device correctly and check the power source.
2. Contact the local service center.

Symbol definition

Symbol Meaning Symbol Meaning

HR

| Heart rate (bpm: beat per minute)| | Low battery voltage
| Type CF applied part| | Attention

IP22

| Protected against dripping water| | Follow instructions for use
| Serial Number| | Date of manufacture
****| Manufacturer’s information| | Fragile or handle with care
ChoiceMMed MD100C Card ECG Monitor - icon7| Keep dry or keep away from rain| | Storage temperature and relative humidity
| Conformity to WEEE Directive| |

Technical specifications

Notes: Specifications may be changed without prior notice.
Weight: ≈80g
Dimension: 80mm(L) X 38mm(W) X 8mm(H)
Heart Rate
Sampling Rate: 500Hz
ECG Bandwidth: 0.67Hz~40Hz
Measurement Range: 30bpm~240bpm
Resolution: 1bpm
Accuracy: 30bpm~100bpm; ±2bpm, 101bpm~240bpm; ±4bpm
Power requirement
Power supply: one built-in rechargeable lithium battery (3.7V/150mAh) The battery will last up to 2 months if it is used for 5 measurements in one day.
Environment requirements
Operating environmental conditions: 5°C~40°C (41°F~104°F), ≤80% (no condensation) Storage/transport environmental conditions: -25°C~70°C (-13°F~158°F, ≤93% (no condensation) Atmosphere pressure: 70kPa~106kPa
Notes: When the ambient temperature is 20°C, it is required 6 hours for the equipment to warm from the minimum storage temperature or 4 hours to cool from the maximum
storage temperature between uses until it is ready for its intended use.
Classification
Type of protection: Internally powered equipment Degree of protection: CF type (the application part are electrodes) According to the degree of protection against ingress of water: IP22 According to the mode of operation: Continuous operation
Packing list

  • One User Manual
  • One charging cable

electromagnetic compatibility

The device conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. Table
1: Electromagnetic Emissions Limits and Compliance

Emissions Test Compliance
RF Emissions CISPR 11 Group 1, Class B

Note: Harmonic Emissions (IEC 61000-3-2), and Voltage Flicker Emissions (IEC 61000-3-3) are not applicable.

Table 2: Electromagnetic Immunity

Immunity Test Compliance
Electrostatic Discharge (ESD) IEC 61000-4-2 ±8 kV contact ±2 kV, ±4 kV, ±8

kV, ±15 kV air
Rated power Frequency Magnetic Fields IEC 61000-4-8| 30 A/m 50Hz and 60 Hz
Radiated RF IEC 61000-4-3| 80 MHz – 2.7 GHz| 10 V/m 80% AM 1kHz
380 – 390 MHz| 27 V/m Pulse mod. 18Hz
430 – 470 MHz| 28 V/m FM±5Hz deviation 1kHz sine
704 – 787 MHz| 9 V/m Pulse mod. 217Hz
800 – 960 MHz| 28 V/m Pulse mod. 18Hz
1.7 – 1.99 GHz| 28 V/m Pulse mod. 217Hz
2.4 – 2.57 GHz| 28 V/m Pulse mod. 217Hz
5.1 – 5.8 GHz| 9 V/m Pulse mod. 217Hz
Note: Electrical Fast Transients (IEC 61000-4-4), Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), and Conducted Immunity (IEC 61000-4-6) are not applicable.

FCC Declaration
FCC ID: WWIMD100C
Please take attention that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operation.

Model: MD100C
Version: Ver1.0
Issue date: May 19, 2022
All rights reserved

Beijing Choice Electronic Technology Co., Ltd.
2nd Floor, 3rd Floor and Room 410-412 4th Floor,
No. 2 Building, No. 9 Shuangyuan Road, Shijingshan District,
100041 Beijing, PEOPLE’S REPUBLIC OF CHINA

Documents / Resources

| ChoiceMMed MD100C Card ECG Monitor [pdf] User Manual
MD100C, WWIMD100C, MD100C Card ECG Monitor, Card ECG Monitor
---|---

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals