Product instructions
FDTH3400
PN:6004213F001-03A
Description Version: V3.0
Revision date: April 2021
The production date is shown on the outer package.
For American please refer to “℉”,
For European please refer to “℃”.
Before using your device, please read these instructions and warnings
completely.
FCC ID :PONFDTHB34
Product Description
Thank you very much for choosing our products!
This product is a high-tech electronic thermometer which is used to measure
the temperature of human body by collecting the change of the temperature and
energy of human body heat source.
This product has 8 seconds rapid measurement mode and direct measurement mode,
so that you can easily and quickly know the health status of yourself and your
family anytime and anywhere.
Product name:Digital Clinical Thermometer
Product model: FDTH3400
Product structure and components
The thermometer mainly consists of plastic shell, thermistor, PCBA, buzzer
chip, LCD display,and battery.
Main appearance structure diagram description:
Display Description:
Product specifications
Product name | Digital Clinical Thermometer |
---|---|
Product model | FDTH3400 |
Power supply voltage | d.c. 3.0V, one CR1632 battery |
Measurement type | Adjusted mode / Direct mode |
Measuring part | Axillary / Oral / Rectal |
Reference site | Oral cavity |
Measuring range | 32.00℃ ~ 42.90℃ (89.60℉~109.22℉) |
Measuring accuracy
| ±0.10℃: between 35.00℃~42.00℃ (±0.18℉ between 95.00℉~107.60℉)
±0.20℃(or±0.36℉): outside this ranges
Display resolution| 0.01℃(0.01℉)
Repeatability error| ≤0.20℃ (±0.36℉)
Measurement time
| Adjusted mode:Less than 8s (for water bath)
Direct mode:Less than 50s (for water bath)
High body temperature hint| ≥37.60℃(99.68℉)
Operation environment
| Temperature:5℃~40℃ (41℉~104℉)
Relative humidity:15%~95%RH,No condensing Atmospheric pressure:70 kPa~106 kPa
Transportation and storage environment| Temperature:-25℃~55℃ (-13℉~131℉)
Relative humidity:15%~95%RH,No condensing Atmospheric pressure:70 kPa~106 kPa
Temperature unit| °C/°F
Memory group| Last time memory display
Battery life| 1 year/1000 measurement
Product size| 28mm×13mm×131mm
Net Product| 17g(Battery free)
Period of validity| 5 years (excluding vulnerable parts)
Software version| V1.0
Grade of waterproof| IP22
Electric shock| Internally powered ME equipment
Operational mode| Continuous operation
Applied part| Type BF applied part,including the whole unit.
*Statement: If stored or used outside the temperature, humidity or atmospheric pressure range specified by the manufacturer, the product may not achieve the claimed performance.
Intended use and scope of application
4.1 This product is used to measure the temperature of human body by
collecting the change of the
temperature and energy of Axillary,Oral,Rectal heat source.
4.2 Expected patients:and can be used for body temperature measurement in
infants, children and adults.
4.3 Intended user : Doctor, nurse or regular user. ( We recommend adults to
operate the thermometer
instead of babies and children.)
4.4 Expected Operating environment:Hospitals, clinics and home Settings
Contraindications
5.1 Do not use local lesions such as inflammation, trauma and postoperative.
5.2 It is forbidden for those who are allergic to stainless steel and ABS
plastic.
Notes
6.1 Measurement
- This product is only for monitoring and self-testing body temperature, can not be used for disease diagnosis, patients only through the results of self-judgment and treatment is very dangerous, so please follow the guidance of the doctor.
- To ensure the accuracy of the measurement, please set the thermometer in the normal working environment for more than 30 minutes, so that it reaches the equilibrium state before measuring.
- Please do not measure under strong electromagnetic interference environment (working microwave oven, induction cooker, mobile phone nearby, etc.), this may cause measurement is not correct or can not be measured.
- Participants were measured after taking medications that change their body temperature (e.g. aspirin, acetaminophen, ibuprofen), which may cause measurement bias.
- The subjects may cause temperature fluctuations after intense exercise, crying, eating, etc. It is recommended to keep quiet for 30 minutes before measurement.
- When measuring, please keep the subject quiet and steady, according to the measurement method shown in the specification, so as not to cause the measurement deviation caused by improper operation.
- Continuous measurement, the measurement results may have a small error, which is a normal phenomenon, it is recommended that the maximum continuous measurement in a unit of time 3 times.
- The human body temperature fluctuates from time to time, and the body temperature collected from different parts will also vary. When observing the changes in body temperature, it is recommended to measure and observe the body temperature in the same part multiple times.
- Different types and brands of temperature measuring products do not have comparative value. It is recommended to use the same temperature measuring product to observe temperature changes in order to better grasp the temperature changes.
6.2 About Products
- Except for measurement of Axillary, oral and rectal temperature, don ’ t use this product for other purposes.
- Please do not force collision, fall, stampede or shake this product, may cause product failure or cause measurement deviation.
- Do not disassemble, repair or modify the product.
- Do not use in electromagnetic interference environment, which may affect the measurement results.
- Please keep this product out of reach of children to prevent children from swallowing batteries or small parts. If swallowed, please contact the doctor immediately.
- Please keep the product in accordance with the product transportation / storage conditions to avoid use problems or measurement deviations.
- Do not put in direct sunlight, high temperature and humidity, dust, near the fire, vulnerable to vibration shock place custody.
- The thermometer is not used for a long time, the battery should be removed to prevent battery leakage.
- Do not place this product in a place with electric shocks.
- Do not insert the metal probe end of the thermometer into the power socket. This may cause electric shock.
- Do not throw the thermometer and its accessory batteries into the fire, which may cause an explosion.
- Please dispose of waste or residue at the end of the product’s service life in accordance with local laws and regulations.
- The normal use of this product is the human body temperature measurement mode, if you need to enter the correction mode must contact the company’s after-sales service agencies.
Installation and use
7.1 Installation of batteries
- This product uses One CR1632 button battery and has been built into the thermometer body when it leaves the factory.(Note: The built-in battery in the thermometer is a trial battery. This battery may not provide the same life as the new battery.)
- To replace the battery, turn the phillips screwdriver counterclockwise to remove the screw on the battery cover.
- gently remove the battery cover, then take out the old battery.
- Follow the battery polarity mark in the battery compartment and put the new battery into the battery compartment. When installing, pay attention to the positive and negative polarity.
- Reinstall the battery cover as is, and use a Phillips screwdriver to turn clockwise to tighten the fixing screws.
7.2 Measurement process
7.2.1 Adjusted mode / Direct mode
-
Tap the “power key”, the thermometer will display in full screen, and then display the effective memory temperature value measured last time.
-
If adjusted mode was used last time, then this measurement is also adjusted mode.
If direct mode was used last time, this measurement will also be in direct mode.
Note: In direct mode, the screen displays the “ ” direct mode icon. -
When the icon “℃” or”℉” on the screen begins to blink, place the temperature sensor of the thermometer in the following test parts:
Measuring part| Measuring method| Graphic
---|---|---
Axillary|
• Raise your arm,Place the probe tip under the armpit parallel to the arm,
pressing the arm against the body.
• Make sure the metal probe of the thermometer is completely wrapped around
the armpit.
|
---|---|---
a Oral| • Place the thermometer in one of the two locations under the tongue,
to the left or right of the root of the tongue. The measuring probe must be in
good contact with the tissue.
• Close your mouth and breathe evenly through the nose to prevent the
measurement from being influenced by inhaled/exhaled air.|
b Rectal| • This is the most reliable measuring method, and is especially
suitable for infants and small children.
• Lay the baby on its back on a hard surface and lift its legs to an angle of
approximately 90°. Hold them by the ankles in order to prevent their body from
turning while taking the temperature.
• Carefully insert the measuring probe of the thermometer 2 to 3cm into the
anal aperture.Until the thermometer’s metal probe is completely covered.|
a For some sensitive people, nausea may occur when mouth temperature is
measured;
b After measuring rectal temperature, there may be a slight feeling of
distension in the anus.These are all normal phenomena, and a rest may relieve
them. If you are concerned that the measurement cannot be done correctly,
please consult a professional physician for the measurement.
c To obtain more accurate measurement data, it is recommended to measure the
oral and rectal temperature in direct mode.
-
Under the direct mode,after the thermometer detects the temperature of the human body, the measurement results will be displayed about 8s later with a prompt sound.
In direct mode, when the temperature value of human body is stable or the test time reaches 5min, the thermometer will display the measurement result with prompt sound.
-
Remove the battery of the thermometer and power it on again each time. The Adjusted mode is entered by default for the next measurement.
7.2.2 The mode switch
- In the off state, press and hold “power key”for 5 seconds and release it, then it will automatically enter the measurement state different from that used last time.
- If the adjusted mode was used last time, it will automatically change to direct mode and start temperature measurement.
7.3 Display and Sound Tips
After the measurement is completed, the following tips may appear in the
thermometer:
Screen display | Sound Tips | Representation of information |
---|---|---|
Temperature : 32.00℃~37.59℃ (89.60℉~99.67℉) | “beep”~ | Target temperature |
values for this measurement
Temperature : 37.60℃~42.90℃ (99.68℉~109.22℉)| “beep beep beep”…”beep beep
beep”…”beep beep beep”…”beep beep beep”…”beep beep beep”| Target temperature
values for this measurement
Err| “beep “…”beep”…”beep “… ” beep”| Abnormal temperature rise rule (e.g.,
measurement process shedding)
Hi| “beep “…”beep”…”beep “… ” beep”| The measurement temperature is higher
than the maximum value of the thermometer range
Lo| “beep “…”beep”…”beep “… ” beep”| The measured temperature is below the
minimum value of the thermometer range
7.4 Power off function
If the thermometer does not perform any operation within 30 seconds or press
“power key” again, it will enter the shutdown state directly.
7.5 Temperature unit setting function
-
When the thermometer is off, hold down“power key” and do not release it. The product will start. Wait for 5 seconds after the screen turns off, and then enter the temperature unit setting interface.
-
Hold down the“power key” and do not release it. The unit of ℃or℉will be automatically switched,every 2s.
When the required unit is displayed, release the hand, that is, select the current unit, and the thermometer immediately enters the state to be measured.
7.6 Low Voltage Prompt Function
- The thermometer automatically detects the battery voltage every time it is turned on. When the battery voltage is insufficient but can continue to use, the“ ” icon appears on the display screen, but does not affect the operation.
- When the battery voltage is too low to continue to use, the display screen appears separately “ ”icon and flashes, about 8 s after forced shutdown, thermometer need to replace the new battery to continue to use.
7.7 Initialization function
Remove the battery and repower the thermometer, and it will automatically
restore to the factory setting.
Body Temperature
The normal body temperature of people is a range value, the normal body
temperature of different people will be different, and the personal body
temperature will also change at different times.
The following are the reference values for the definition of fever in various
parts of the human body:
*Data sources: ISO 80601-2-56:2017/AMD 1:2018, for information only.
It is dangerous to make self-diagnosis and treatment according to the
measurement results. Please consult a professional doctor with the measurement
results.
Troubleshooting instructions
Fault Description | Possible causes | Solution |
---|
The screen does not react when the power is on or Equipped with batteries.
| ****
Battery runs out
| Please replace the new battery and re-open
Battery installation error| Please install the battery correctly
according to the positive and negative pole mark in the battery bin.
After reloading the battery, the screen remains blank| Please contact
after sale
All ICONS on the screen flicker| Hardware failure| Please
contact after sale
The battery icon appears on the screen and flashes| Too low
battery voltage, Causing failure to boot.| Please replace the new
battery and restart
Battery icon appears on the screen| Battery power is running out|
**** Please replace the new battery as soon as possible
| ****
Abnormal temperature rise rule (e.g., measurement process shedding)
| Please restart the measurement. The thermometer should be properly placed at the measuring site according to the instructions.
| The measurement temperature is higher than the maximum value of the thermometer range| Please check the measurement target is correct and re- measure again.
| The measured temperature is below the minimum value of the thermometer range| Please check the measurement target is correct and re-measure again.
Maintenance and maintenance
-
Please use a soft, dry cloth to gently wipe the dirt on the mainframe. If the surface is more serious dirty, you can dip some alcohol to wipe.Cleaning is recommended every 1 month.
Caution: Do not wash with water or use cleaners, diluents or volatile oils containing abrasives. -
After the thermometer measuring probe is soiled, please use a soft cloth dipped in more than 75% concentration of medical alcohol to cover the measuring probe end and gently wipe it. If the wipe is still not clean, please contact customer service.
-
If the thermometer measuring probe, button, display and other parts are damaged, please contact the after-sales service.
-
Do not put in direct sunlight, high temperature and humidity, dust, near the fire, vulnerable to vibration shock place custody.
-
When the thermometer is not used for a long time, the battery should be removed to prevent battery leakage.
-
When the “ ” icon appears on the thermometer display screen, it indicates that the battery power is about to run out. Please replace the battery as soon as possible.
-
We do not authorize any organization or individual to carry out maintenance, please do not disassemble and adjust the thermometer.
-
The thermometer has been tested and calibrated at the time of manufacture. If you have any questions about the function or measurement accuracy of the product, please contact the dealer or manufacturer.
List of accessories
Use only original accessories to check whether the received accessories are complete.
Quantity | Components |
---|---|
1 pcs | FDTH3400 host |
1 pcs | CR1632 batteries (Placed inside the mainframe) |
1 pcs | User Manual |
The technical data should be combined with the user manual.
In addition to specifying the use of CR1632 batteries, replacing existing components with non-manufacturer supplied components may cause measurement errors and reduce safety.
Final disposal
At the end of the product lifecycle, do not throw this product into the normal
household garbage, but bring it to a collection point for the recycling of
electronic equipment.
Waste Electrical and Electronic Equipment can have potentially harmful effects
on the environment.
Incorrect disposal can cause harmful toxins to build up in the air, water and
soil and can be harmful to human health.
Information on critical parts
Name | Model | Supplier |
---|---|---|
Thermistor | XWTH002 | A041 |
IC | XWIC036 | A102 |
Thermometer casing | ABS | D004 |
Explanation of standardized symbol
| Complies with the European Medical Device Directive (93/42/ EEC), Notified
Body is SGS Belgium NV
---|---
| **** Authorized representative in the European Community.
| **** Attention: see Instructions for use!
| Caution! Consult accompanying documents.
| Type BF applied parts
| Batch code
| Serial number
| Manufacturer information:
IP22
| IP code of the device: this device’s grade of against ingress of solid foreign objects .
| **** Disposal in accordance with Directive 2002/96/EC (WEEE).
| **** Medical Device
| **** Catalogue number
| **** Unique Device Identifier
| **** Temperature limit
| **** Humidity limitation
| **** Date of manufacture
Electromagnetic compatibility information
15.1 FCC STATEMENT
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This transmitter must not be co-located or operated in conjunction with any
other antenna or transmitter.
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment and meets the FCC radio frequency (RF) Exposure
Guidelines. This equipment has very low levels of RF energy that are deemed to
comply without testing of specific absorption ratio (SAR).
15.2 WARNING:
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
Not use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this the FDTH3400 could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the FDTH3400, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
electromagnetic emission
– for all EQUIPMENT AND SYSTEMS
– electromagnetic emission
The FDTH3400 is intended for use in the electromagnetic environment specified below. The customer or the user of
FDTH3400 should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment –
guidance
RF emissions CISPR 11| Group 1| The FDTH3400 uses RF energy only for its
internal function. There for, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The FDTH3400 is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2| N/A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
| ****
N/A
lectromagnetic immunity
– for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The FDTH3400 is intended for use in the electromagnetic environment specified
below. The customer or the user of the FDTH3400 should assure that it is used
in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment -guidance
Electrostatic discharge (ESD) IEC 61000-4-2
| ****
± 8 kV contact
± 2 kV, ± 4 kV, ±
8 kV,± 15 kV air
| ****
± 8 kV contact
± 2 kV, ± 4 kV, ±
8 kV,± 15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %. If ESD
interfere with the operation of equipment, counter measurements such as wrist
strap, grounding shall be considered.
Electrostatic transient / burst IEC 61000-4-4| ± 2 kV for power supply lines|
**** N/A| Mains power quality should be that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
| ± 1 kV differential mode
± 2 kV common mode| N/A| Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11| 0 % UT (100 % dip in UT) for 0,5 cycle
0 % UT (100 % dip in UT) for 1 cycles
70 % UT (30 % dip in UT) for 25/30cycles 0 % UT (100 % dip in| N/A|
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the FDTH3400 requires continued operation during
power mains interruptions, it is recommended that the FDTH3400 be powered from
an uninterruptible power supply or a battery.
| UT) for 250/300 cycles| |
---|---|---|---
Power frequency (50/60Hz) magnetic field IEC 61000-4-8|
30 A/m
| ****
30 A/m
| Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE:UT is the a. c. mains voltage prior to application of the test level.
electromagnetic immunity – for EQUIPMENT and SYSTEM that are not LIFE- SUPPORTING
electromagnetic immunity
The FDTH3400 is intended for use in the electromagnetic environment specified
below. The customer or the user of the FDTH3400 should assure that it is used
in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
**** Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3| ****
**** 3 Vrms
150 kHz to 80 MHz 6 V in ISM and amateur radio bands between 0,15 MHz and 80
MHz.a 10 V/m 80 MHz to 2.7 GHz 385MHz-5785MHz
Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communication equipment (Refer to table 9 of IEC 60601-1-2:2014)
| ** N/A
**10 V/m
80 MHz to 2.7 GHz 385MHz-5785MHzTest specifications for ENCLOSURE PORT
IMMUNITY to
RF wireless communication equipment (Refer to table 9 of IEC 60601-1-2:2014)|
Portable and mobile RF communications equipment should be used no closer to
any part of the FDTH3400, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
where p is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,b should be less than the compliance level in each
frequency range.c
| | | Interference may occur in the vicinity of equipment marked with the
following symbol:
---|---|---|---
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is
affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80
MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and
80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz
to 7,3 MHz,
10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to
21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0
MHz.
b Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the FDTH3400 is used exceeds
the applicable RF compliance level above, the FDTH3400 should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary,such as such as reorienting or relocating the
FDTH3400.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -for EQUIPMENT and
SYSTEMS
Recommended separation distances between
portable and mobile RF communications equipment and the FDTH3400
The FDTH3400 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the FDTH3400 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the FDTH3400 as recommended below, according to the maximum output power of the communications equipment
Rated maximum output of transmitter
(W)
| Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz outside ISM and amateur radio bands
| 150 kHz to 80 MHz in ISM and amateur radio bands
| **** 80 MHz to 800 MHz
| **** 800 MHz to 2.7 GHz
0.01| N/A| N/A| 0.035| 0.07
0.1| N/A| N/A| 0.11| 0.22
1| N/A| N/A| 0.35| 0.7
10| N/A| N/A| 1.1| 2.21
100| N/A| N/A| 35| 70
For transmitters rated at a maximum output power not listed above the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 : At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Standard List
FDTH3400 complies with the following standards
IEC 60601-1| Medical electrical equipment Part 1: General requirements for
basic safety and essential performance
---|---
IEC 60601-1-2| Medical electrical equipment — Part 1-2: General requirements
for basic safety and essential performance – Collateral standard:
Electromagnetic compatibility – Requirements and tests
IEC 60601-1-11
| Medical electrical equipment – Part 1-11: General requirements for basic
safety and essential performance – Collateral Standard: Requirements for
medical electrical equipment and medical electrical systems used in the home
healthcare environment
ISO 10993-1| Biological evaluation of medical devices – Part 1: Evaluation and
testing within a risk management process
EN ISO 15223-1| Medical device –symbols to be used with medical device labels,
labeling and information to be supplied –Part 1; General requirements
ISO 80601-2-56| Medical electrical equipment – Part 2-56:Particular
requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
Warranty
We provide one year warranty starting from the date of purchase. Please refer to the followings situations that are excluded from the free repair services within the warranty period.
- All damages caused by disassembly and repair of the device by yourselves.
- All damages caused by dropping the device during usage, or transport.
- All damages caused by improper usage of the device and not following the instructions on the user manual.
Please contact after-sales service and support and enclose your product
purchase receipt while claiming for warranty services.
Location of purchase:
Contact number:
Date of purchase:
Famidoc Technology Co., Ltd.
Add.: No. 212 Yilong Road, Hexi Industrial Zone,Jinxia, Changan Town, Dongguan
523853, Guangdong
Province, P.R. China.
Tel.: +86-769-89272488
Fax: +86-769-89272498
Website: www.famidoc.com
Name: Shanghai International Holding Corp. GmbH (Europe)
Add: Eiffestrasse 80, 20537 Hamburg, Germany
References