PeakTech 3690 5 In 1 Digital Multitester Instruction Manual
- June 12, 2024
- PeakTech
Table of Contents
Unser wert ist messbar…
3690 5 In 1 Digital Multitester
Instruction Manual
PeakTech® 3690
Operation manual
„5 in 1“ Digital-Multitester
Safety precautions
This product complies with the requirements of the following directives of the
European Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2014/35/EU (low voltage), 2011/65/EU (RoHS).
Overvoltage category III 600V; pollution degree 2.
CAT I: For signal level, telecommunication, electronic with small
transient over voltage
CAT II: For local level, appliances, main wall outlets, portable
equipment
CAT III: Supplied from a cable under earth; fixed installed switches,
automatic cut-off or main plugs
CAT IV: Units and installations, which are supplied overhead lines, which
are stand in a risk of persuade of a lightning, i.e. main-switches on current
input, overvoltage-diverter, current use counter.
To ensure safe operation of the equipment and eliminate the danger of serious
injury due to short-circuits (arcing), the following safety precautions must
be observed.
Damages resulting from failure to observe these safety precautions are exempt
from any legal claims whatever.
General:
- Read these operating instructions carefully and make them available to subsequent users.
- It is essential to observe the warning notices on the device, do not cover or remove them.
- Pay attention to the use of the multimeter and only use it in the suitable overvoltage category.
- Familiarize yourself with the functions of the measuring device and its accessories before you carry out the first measurement.
- Do not operate the measuring device unsupervised or only protected against unauthorized access.
- Use the multimeter only for the purpose of its determination and pay particular attention to warning notices on the device and information on the maximum input values.
Electric safety:
- Voltages over 25 VAC or 60 VDC are generally considered dangerous voltages.
- Only work on dangerous voltages by or under the supervision of qualified personnel.
- When working on dangerous voltages, wear suitable protective equipment and observe the relevant safety rules.
- Do not exceed the maximum permissible input values under any circumstances (risk of serious injury and / or destruction of the device)
- Pay special attention to the correct connection of the test leads depending on the measuring function in order to avoid a short circuit in the device. Never apply a voltage in parallel to the current sockets (A, mA, µA).
- Current measurements are always carried out in series with the consumer, i.e. with the supply line disconnected.
- Remove the test probes from the measurement object before changing the measuring function.
- Never touch the bare test probes during the measurement, only hold the test leads by the handle behind the finger guard. If applicable, discharge any capacitors before measuring the circuit to be measured.
- The thermocouple for temperature measurements is made of conductive material. To avoid electric shock, never connect it with a live conductor.
Measurement environment:
- Avoid any proximity to explosive and flammable substances, gases and dust. An electric spark could lead to an explosion or deflagration – danger to life!
- Do not carry out measurements in corrosive environments, the device could be damaged or contact points inside and outside the device could corrode.
- Avoid working in environments with high interference frequencies, high-energy circuits or strong magnetic fields, as these can negatively affect the multimeter.
- Avoid storage and use in extremely cold, humid or hot environments, as well as long-term exposure to direct sunlight.
- Only use devices in damp or dusty environments in accordance with their IP protection class.
- If no IP protection class is specified, only use the device in dust-free and dry indoor rooms only.
- When working in damp or outside areas, pay particular attention to completely dry handles on the test leads and test probes.
- Before starting the measuring operation, the device should be stabilized at the ambient temperature (important when transporting from cold to warm rooms and vice versa)
Maintenance and Care:
- Never use the device if it is not completely closed.
- Before each use, check the device and its accessories for damage to the insulation, cracks, kinks and breaks. If in doubt, do not take any measurements.
- Change the battery when a battery symbol is displayed to avoid incorrect rdg.s.
- Switch off the multimeter before changing batteries or fuses and also remove all test leads and temperature probes.
- Replace defective fuses only with a fuse that corresponds to the original value. Never short-circuit a fuse or fuse holder.
- Charge the battery or change the battery as soon as the battery symbol lights up. Insufficient battery power can lead to inaccurate measurement results. Electric shocks and physical damage can result.
- If you are not going to use the device for a longer period of time, remove the battery from the compartment.
- Have maintenance and repair work on the multimeter carried out only by qualified specialists.
- Do not lay the device upside down on the workbench or work surface to avoid damaging the control elements.
- Clean the housing regularly with a damp cloth and a mild cleaning agent. Do not use any caustic abrasives.
- Do not make any technical changes to the device.
Cleaning the cabinet
Clean only with a damp, soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment
1.1. Input limits
DCV | 600V DC/AC rms |
---|---|
ACV | 600V DC/AC rms |
µA/mA DC/AC | 500mA/600V |
10 A DC/AC | 10A/ 600V |
Resistance | 600V DC/AC rms |
Capacity | 600V DC/AC rms |
Frequency | 600V DC/AC rms |
Duty Cycle | 600V DC/AC rms |
Temperature | 600V DC/ACrms |
Diode – and Continuity-Test | 600V DC/AC rms |
1.2. Multimeter Safety
Be sure to follow the WARNINGS in this manual. Erroneous use may put human
bodies in danger.
The following legend applied to this manual:
| Dangerous voltage (take care not to get an electric shock in voltage
measurement)
---|---
| Ground (allowable applied voltage range between the input terminal and
earth)
| Refer to the instruction manual (very important description for safe
use)
| Direct current (DC)
| Alternating current (AC)
| Replace fuses with amp/voltrating shown
| Double insulation (Protection class II)
Features
- 14 positions “easy to use” rotary switch for function and range selection
- 15 mm high contrast multiline LCD with backlight
- Automatic overrange indication with the “OL” displayed
- Automatic polarity indication on DC ranges
- All ranges fully protected
- Diode testing & Audible continuity test
- Sound-Level Meter in dBC weighting
- Lux-Meter up to 40.000 Lux
- Humidity Meter with integrated sensor
- Temperature Meter for air temperature
- Temperature measurement with supplied wire probe
- High precision multimeter functions
- Complies with latest safety standards
Front Panel Description
-
Humidity & Temperature: Humidity Sensor and Semiconductor Sensor
-
LCD display: 3 4/5 digits LCD display
-
Function switch
-
V/Hz%/Ω/Cap/°C input jack
-
COM input jack
-
uA/mA input jack
-
10A input jack
-
Microphone: Electric condenser microphone inside.
-
Photo Detector: Long life silicon photo diode inside.
-
Hz/% button
-
Hold button
-
Backlight button
-
MODE button
-
Range button
The button to select AC or DC measurement when in Voltage, Ω ranges. -
REL button
The relative measurement feature allows you to make Measurements relative to a stored reference value. A reference voltage, current, Capacitor, etc. can be stored and measurements made in comparison to that value. The displayed value is the difference between the reference value and the measured value.
- Perform the measurement as described in the operating instructions.
- Press the REL button to store the reading in the display and the “REL” indicator will appear on the display.
- The display will now indicate the difference between the stored value and the measured value.
- Press the REL button to exit the relative mode. 16. NCV indication lamp
Specifications
Accuracies are: (% of reading + no. of digits) guaranteed for 1 year, 23°C +/-
5°, less than 75% RH
4.1. DC Voltage
Range | Resolution | Accuracy |
---|---|---|
400 mV | 0,1 mV | **** |
+/-1,2% rdg. + 5 dgt.
4 V| 1 mV
40 V| 10 mV
400 V| 100 mV| +/-1,8% rdg. + 4 dgt.
600 V| 1 V
Overload protection: 600V DC / ACrms
Input Impedance: 10MΩ
4.2. AC Voltage
Range | Resolution | Accuracy |
---|---|---|
400 mV | 0,1 mV | +/-1,8% rdg. + 20 dgt. |
4 V | 1 mV | +/-1,2% rdg. + 5 dgt. |
40 V | 10 mV | |
400 V | 100 mV | +/-1,8% rdg. + 4 dgt. |
600 V | 1 V | +/-2,2% rdg. + 5 dgt. |
Overload protection: 600V DC / ACrms
Frequency Range: 50 – 400Hz
Input Impedance: 10MΩ
4.3. DC Current
Range | Resolution | Accuracy |
---|---|---|
400 µA | 0,1 µA | **** +/-1,2% rdg. + 2 dgt. |
4000 µA | 1 µA | |
40 mA | 10 µA | |
400 mA | 100 µA | +/-1,5% rdg. + 2 dgt. |
10 A | 10 mA | +/-2,2% rdg. + 5 dgt. |
Overload protection: µA/mA
Ranges: 500mA/600V 10A
Range: 10A/ 600V
4.4. AC Current
Range | Resolution | Accuracy |
---|---|---|
400 µA | 0,1 µA | +/-1,5% rdg. + 3 dgt. |
4000 µA | 1 µA | |
40 mA | 10 µA | |
400 mA | 100 µA | +/-1,8% rdg. + 3 dgt. |
10 A | 10 mA | +/-2,2% rdg. + 6 dgt. |
Overload protection: µA/mA Ranges: 500mA/600V
10A Range: 10A/ 600V fuse (quick acting)
Frequency Range: 50 – 400 Hz
4.5. Resistance
Range | Resolution | Accuracy |
---|---|---|
400 Ω | 0,1 Ω | +/-1,8% rdg. + 5 dgt. |
4 kΩ | 1 Ω | +/-1,5% rdg. + 2 dgt. |
40 kΩ | 10 Ω | |
400 kΩ | 100 Ω | |
4 MΩ | 1 kΩ | +/-2,0% rdg. + 2 dgt. |
40 MΩ | 10 kΩ | +/-2,5% rdg. + 2 dgt. |
Test voltage: ca. 0,28V
Overload protection: 600V DC/ACrms
4.6. Capacitance
Range | Resolution | Accuracy |
---|---|---|
50 nF | 10 pF | +/-5,5% rdg. + 25 dgt. |
500 nF | 0,1 nF | +/-3,5% rdg. + 7 dgt. |
5 µF | 1 nF | |
50 µF | 10 nF | |
100 µF | 100 nF | +/-5,0% rdg. + 7 dgt. |
Overload protection: 600V DC/ACrms
4.7. Frequency
Range | Resolution | Accuracy |
---|---|---|
5.000 Hz | 1 mHz | +/-1,5% rdg. + 3 dgt. |
50.00 Hz | 10 mHz | |
500.0 Hz | 0,1 Hz | |
5.000 kHz | 1 Hz | |
50.00 kHz | 10 Hz | |
500.0 kHz | 100 Hz | |
10.00 MHz | 1 kHz | +/-1,5% rdg. + 4 dgt. |
Sensivity:
0,5Vrms ≤ 1MHz
3Vrms >1MHz
Overload Protection: 600V DC/ACrms
4.8. Diode and Continuity Test
Range | Description | Test condition |
---|---|---|
Displaying approximate forward voltage of diode | Forward DC current: 1mA |
Forward DC voltage: 1,5V
| Built-in buzzer will sound if resistance is lower than 50Ω| Open circuit voltage: ~ 2,8V DC
Overload protection: 600V DC/ACrms
4.9 Duty
Range | Resolution | Accuracy |
---|---|---|
0,1 … 99,9 % | 0,1 % | +/-3,0% |
4.10. Relative Humidity (on RH and humidity display)
Range | Resolution | Accuracy |
---|---|---|
33 … 99 % | 1 % RH | +/-3% + 5,5% RH |
Operating temperature: 0°C … 50°C Sampling Period: ~20s.
4.11. Temperature (room temperature)
Range | Resolution | Accuracy |
---|---|---|
0°C … 50°C | 0,1°C | +/-3% rdg + 3°C |
Sampling Period: ~20s.
4.12. Temperature (thermocouple at main display)
Range | Resolution | Accuracy |
---|---|---|
°C | 0,1°C | -20°C… 400°C |
1°C | -20°C…1300°C | +/-3% rdg. + 3°C |
Overload protection: 600V DC/ACrms
4.13. Sound Level (dB)
Range | Resolution | Accuracy |
---|---|---|
35 – 100 dB | 0,1 dB | +/-5,5 dB at 94dB, 1kHz sine wave |
Typical instrument frequency range: 30Hz ~ 10kHz
Weighting: C
Time response: Fast
4.14. Luminance (LUX)
Range | Resolution | Accuracy |
---|---|---|
4000 Lux | 1 Lux | +/-5,0% rdg. + 10 dgt. |
40000 Lux (x10Lux) | 10 Lux |
Repeatability : +/- 2%
Calibrated to standard lamp with 2856 K
Sensor: Silicon Photo-Diode
General Characteristics
Display| 15mm LCD display, 3999 counts (3 3/4 digits) with automatic
polarity indication
---|---
Overrange indication| “OL” Figure only in the display
Common mode voltage| max. 600V DC / 600V ACrms
Reading rate time| 2,5 readings per sec. (approx.)
Temperature for guaranteed accuracy| 23° C ± 5° C
Operating Temperature| 0° C…40° C, 32° F… 104° F; <70% RH
Storage Temperature| -10° C…50° C, 14° F…122° F; <80% RH
Power Supply| 9V-battery (NEDA 1604)
Low Battery Indication| Batterysymbol appears on the left of display
Size (WxHxD)| 78 (B) x 170 (H) x 48 (T) mm
Weight| 335 g incl. Holster
**** Accessories| test leads, holster, battery, operation manual and
temperature probe
Operation
6.1. Preliminary Note
- Check the batteries by turning to any position. If the batteries are weak, a sign will appear on the left of display. If this does not appear on the display proceed as below. See “Maintenance” if the batteries have to be replaced. Please use only 4mm-safety test leads to ensure immaculate functions.
- The warning sign next to the test leads jack is for warning that the input voltage or current should not exceed the indicated values. This is to prevent damage to the internal circuitry.
- The function switch should be set to the range which you want to test before operation.
6.2. DC Voltage Measurement
- Connect the black test lead to the COM jack and the red test lead to the V/Ω – jack.
- Set the Function switch to mV or V
- Select DC by pressing the “SELECT” button.
- Connect the test leads across the source or load under measurement.
Note:
- If the voltage range is not known beforehand set the function switch to the highest range and work down.
- When only the figure “OL” is displayed, overrange is being indicated and you must be set to a higher range.
- CAUTION: Do not apply more than 600V to the input. Indication is possible at higher voltages but there is danger of damaging the internal circuitry.
- Use extreme caution to avoid contact with high tension circuits when measuring high voltage.
6.3. AC Voltage Measurement
- Connect the black test lead to the COM jack and the red test lead to the V/Ω- jack.
- Set Function Switch to V ~.
- Connect the test leads across the source of load under measurement.
Note:
- If the voltage range is not known beforehand set the function switch to the highest range and work down.
- Caution: Do not apply more than 600Vrms to the input. Indication is possible at higher voltages but there is danger of damaging the internal circuitry.
- Use extreme caution to avoid contact with high tension circuits when measuring high voltage.
6.4. DC Current Measurement
- Connect the black test lead to the COM jack and the red test lead to the µA/mA- jack for a max. of 400mA. For a maximum of 10A, move the red test lead to the 10A jack.
- Set the Function switch to the µA/mA or 10A range to be used and connect the test leads in series with the load under measurement
- Select DC by pressing the “MODE” button.
Note:
- If the current range is not known beforehand, set the FUNCTION switch to the highest range and work down.
- When only the figure “OL” is displayed overrange is being indicated and the FUNCTION switch must be set at higher range.
- Caution: The maximum input current is 400mA, or 10A depending upon the jack used. Excessive current will blow the fuse which must be replaced.
6.5. AC Current measurement
- Connect the black test lead to the COM jack and the red test lead to the µA/mA jack for a max. of 400mA. For a maximum of 10A, move the red test lead to the 10 A jack.
- Set the function switch to the µA/mA or 10A range to be used and connect the test leads in series with the load under measurement.
- Select DC by pressing “MODE”-button.
Note:
- If the current range is not known beforehand, set the FUNCTION switch to the highest range and work down.
- When only the figure “OL” is displayed overrange is being indicated and the FUNCTION switch must be set at higher range.
- Caution: The maximum input current is 400mA, or 10A depending upon the jack used. Excessive current will blow the fuse which must be replaced.
6.6. Resistance Measurement
- Connect the black test lead to the COM jack and the red test lead to the V/Ω – jack. (Note: The polarity of the red test lead is “+”)
- Set the Function switch to the “ ” -function.
- Select Ω -function by pressing “MODE”-button.
- Connect the test leads across the resistance under measurement.
Note:
- If the resistance value being measured exceeds the maximum voltage of the range selected, an overrange indication will be displayed (“OL”). Select a higher range. For resistance approx. 1 MΩ and above, the meter may take a few seconds to stabilise. This is normal for high resistance readings.
- When the input is not connected, i. e. at open circuit, the figure “OL” will be displayed for the overrange condition.
- When checking in-circuit resistance, be sure the circuit under test has all power removed and that all capacitors are fully discharged.
6.7. Capacitance measurements
Caution!
Turn off power and discharge the capacitor before attempting a capacitance
measurement. Use the DCV function to confirm that the capacitor is discharged.
- Set the Function switch to “CAP”-position.
- Connect the red test lead to the V/Ω -input jack and the black test lead to the COM-input jack.
- Connect the test leads to the capacitor. Observe polarity when measuring polarized capacitors.
- Read the capacitance directly from the display. A shorted capacitor will indicate an overrange. An open capacitor will indicate near zero on all ranges.
6.8. Diode Measurement
- Connect the black test lead to the COM jack and the red test lead tothe V/Ω – jack. (Note: the polarity of the red test lead is “+”)
- Set the function switch to the “ “- range.
- Select Diode-function by pressing “MODE”-button.
- Connect the test leads across the diode under measurement.
Note:
- When the input is not connected, i. e. at open circuit, the figure “OL” will be displayed for the overrange condition.
- There is 1mA Current flow through the device under test.
- The meter displays the forward voltage drop in millivolts, and overload when the diode is reversed.
6.9. Audible Continuity Test
- Connect the black test lead to the COM jack and the red test lead to the V/Ω- jack.
- Set the Function switch to ” ” range and connect the test leads across the resistance under measurement.
- Select continuity-function by pressing “MODE”-button.
- Buzzer sounds if the resistance between two prods is less than approx. 50Ω.
Note:
- When the input is not connected, i. e. at open circuit, the Figure “OL” will be displayed for the overrange indication.
- The circuit to be tested must be in power off status during the continuity check.
6.10 frequency measurements:
ATTENTION!
Do not perform measurements on circuits with voltages above 250V DC / AC rms.
Exceeding this voltage value, the risk of serious injury exists due to
electric shock and / or the risk of damage to the device.
- Turn the Function / Range switch (2) in the required frequency measuring position.
- Connect red test lead to the V//CAP/Hz/Temp. – input, black lead to the COM input.
- Connect test leads across the circuit to be measured or the component to be measured and read the frequency in the LCD display.
- For accurate frequency measurements a measuring line is recommended with BNC connectors.
6.11. Measuring Sound Level
- Set the function to „dBC“ Position.
- Face the microphone to sound source in a horizontal position.
- The C-weighting curve is nearly uniform over the frequency range from 30Hz to 10.000kHz, thus giving an indication of overall Sound Level.
- The fast response is suitable to measure shout bursts and peak values from sound source.
- The sound level will be displayed.
- Note: Strong wind (over 10m/sec.) striking the microphone can cause misreading for measurement in windy locations.
6.12. Measuring Light
- Set the function switch to „Lux“-position and set the range to desired („lux“ or „10 x lux“) range.
- Face the photo detector to light source in a horizontal position.
- Read the illuminance nominal from the LCD-display.
- Overrange: If the instrument only display one „OL“ in the LCD the input signal is too strong, and a higher range should be selected.
Locations | LUX |
---|---|
Office | |
Conference, Reception room | 200 – 700 |
Clerical work | 700 – 1500 |
Typing drafting | 1000 – 2000 |
Factory | |
Packing work, Entrance passage | 150 – 300 |
Visual work at production line | 300 – 750 |
Inspection work | 750 – 1500 |
Electronic parts assembly line | 1500 – 3000 |
Hotel | |
Public room, Cloakroom | 150 – 200 |
Reception, Cashier | 200 – 1000 |
Store | |
Indoors Stairs Corridor | 150 – 200 |
Show window, Packing table | 750 – 1500 |
Forefront of show window | 1500 – 3000 |
Hospital | |
Sickroom, Warehouse | 100 – 200 |
Medical Examination room | 300 – 750 |
Operating room, Emergency Treatment | 750 – 1500 |
School | |
Auditorium, Indoor Gymnasium | 100 – 300 |
Class room | 200 – 750 |
Laboratory Library Drafting room | 500 – 1500 |
6.13. Measuring Humidity
- Turn on the meter
- The display will show the humidity reading value (%RH) directly.
- When the tested environment humidity value changed, it needs a few minutes to get the stable „%RH“ reading.
Warning!
Don´t expose the humidity sensor to direct sunlight.
Don´t touch or manipulate the humidity sensor.
6.14. Measuring Temperature
- Set the function switch to “1°C or 0,1°C” position.
- Connect the black plug of temperature probe the COM-jack and red plug to the “V/ Ω” jack”.
- Touch the end of the temperature sensor to the area or surface of the object to be measured. The display will show the temperature reading value (°C) directly.
Warning!
When function switch on temperature range, never attempt a voltage
measurement.
6.15. Non-Contact AC voltage test (NCV):
- Set the function switch to the ON position
- Remove the meter and face the NCV detector to ACV source.
- If source voltage in 50-600V the NCV indicate lamp will light.
Maintenance
7.1. Installing the battery
Your meter requires a 9V-battery for power. The battery-symbol appears when
the battery voltage drops to certain limits. For correct operation, replace
the battery as soon as possible. Continued use with a low battery will lead to
errors in readings.
WARNING!
To avoid electric shock, disconnect all leads from any equipment before you
remove or install the battery.
Follow these steps to install the battery.
- Turn off the power and disconnect all test leads.
- Remove the back cover by unscrewing the two screws of the battery compartment and pull it off.
- Remove the used battery.
- Place the new battery inside the insulation capsule and snap it onto place.
WARNING!
Do not discard the provided battery insulation capsule. If you do not use this
insulation capsule properly, it might cause damage or injury.
WARNING!
Do not operate the meter until you replace the batteries and close the battery
compartment cover.
Notification about the Battery Regulation
The delivery of many devices includes batteries, which for example serve to
operate the remote control. There also could be batteries or accumulators
built into the device itself. In connection with the sale of these batteries
or accumulators, we are obliged under the Battery Regulations to notify our
customers of the following:
Please dispose of old batteries at a council collection point or return them
to a local shop at no cost. The disposal in domestic refuse is strictly
forbidden according to the Battery Regulations. You can return used batteries
obtained from us at no charge at the address on the last side in this manual
or by posting with sufficient stamps.
Contaminated batteries shall be marked with a symbol consisting of a crossed-
out refuse bin and the chemical symbol (Cd, Hg or Pb) of the heavy metal which
is responsible for the classification as pollutant:
- “Cd” means cadmium.
- “Hg” means mercury.
- “Pb” stands for lead.
You can obtain further information about the Battery Regulations from the
Bundesministerium für Umwelt, Naturschutz und Reaktorsicherheit (Federal
Ministry of Environment, Nature Conservation and Reactor Safety).
7.2. Replacing the fuse
WARNING!
To avoid electric shock, disconnect all the test probes before removing the
fuse. Replace only with the same type of fuse. Service should be performed
only by qualified personnel.
CAUTION!
For continued protection against fire or other hazard, replace only with fuse
of the specified voltage and current ratings.
Follow these steps to replace the fuse:
-
Turn ON/OFF button to turn the meter off and disconnect the test probes.
-
Remove the back cover by unscrewing the 6 screws and pulling off the back cover.
-
Remove the blown fuse.
-
Install the new fuse in the fuse compartment.
FF 500mA/600V; 5 x 20mm
F 10A/600V; 6,3 x 32mm -
Replace the battery compartment and secure it with the screws.
WARNING!
Do not operate your meter until the back cover is in place and fully closed.
All rights, also for translation, reprinting and copy of this manual or parts
are reserved. Reproductions of all kinds (photocopy, microfilm or other) only
by written permission of the publisher.
This manual is according the latest technical knowing. Technical alterations
reserved.
Misprints and errors are reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications. We recommend to
calibrate the unit again, after 1 year. ® © PeakTech 03/2023/Po/Th/Lie/Ehr
PeakTech Prüf- und Messtechnik GmbH – Gerstenstieg 4 –
DE-22926 Ahrensburg / Germany
+49 (0) 4102 97398-80 +49 (0) 4102 97398-99
info@peaktech.de
www.peaktech.de
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>